K Number
K214041
Manufacturer
Date Cleared
2022-12-30

(372 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AESCULAP Aicon™ Container is a reusable rigid sterilization container intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device during transport and until used. This container is compatible for use in the following sterilization modalities in the configurations listed below:

  • Ethylene Oxide (EtO)
  • STERIS V-PRO maX/maX 2: Lumen, Non-Lumen, Flex
  • STERIS V-PRO 60: Lumen, Non- Lumen, Flex
  • STERRAD 100NX: Standard, Express, Flex, Duo
  • STERRAD NX: Standard, Advanced
  • STERRAD 100S
  • STERIZONE VP4
  • PreVac Steam

The AESCULAP Aicon™ consists of two different container styles: a solid bottom container and enhanced drying system (EDS) container.

Device Description

The AESCULAP Aicon Container is a reusable container system intended for sterilization and storage of other medical devices. This container system is compatible for use in the following sterilization modalities:

  • Ethylene Oxide
  • STERIS VPRO maX/ maX 2: Lumen, Non- Lumen, Flex,
  • STERIS VPRO 60: Lumen, Non-Lumen, Flex
  • STERRAD 100NX: Standard, Express, Flex, Duo
  • STERRAD NX: Standard, Advanced
  • STERRAD 100S
  • STERIZONE VP4
  • PreVac Steam

The containers are perforated and made from anodized aluminum and utilize a single-use filter.

AI/ML Overview

This document describes the premarket notification (510(k)) for the Aesculap AICON™ Container, a reusable rigid sterilization container. The following information is extracted from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Performance TestingPurposeAcceptance CriteriaReported Device Performance
Sterilization EfficacyTo determine sterilization effectiveness of test device after processing in a sterilization cycle.A sterility assurance level (SAL) of 10-6 will be achieved post sterilization using the BI overkill method and half cycle validation. Biological indicators must be negative for growth after incubation period.Pass
Dry TimeTo determine the proper drying time required for the Aesculap Aicon Container.The system shall demonstrate an average pre and post sterilization weight difference of less than 0.2% within five 5 minutes of cycle completion using final validated parameters and be free of visible moisture following a cooling period.Pass
Microbial Aerosol ChallengeTo analyze the package integrity and microbial barrier properties of the test device.The container load maintains sterility after exposure to a defined amount of aerosol microorganisms. No presence of growth after incubation period.Pass
Simulated UseTo determine the effective sterilization of flexible scopes when used with the test device.A minimum of 1.0 x 10^6 spores contained within organic soil representative of actual use conditions are killed during defined sterilization cycle.Pass
Material CompatibilityTo assess effects of full use cycles on device components and their intended functionality.No impact to functionality at the completion multiple sterilization cycles.Pass
CytotoxicityTo determine the potential of a test device to cause cytotoxicity.Testing completed in accordance with ISO 10993-5: 2009 to demonstrate no significant cytotoxic reaction after exposure to sterilant. Using the ISO Elution Method, the response to the article is not greater than 2 (mild reactivity).Pass
Sterility MaintenanceTo demonstrate that a processed test device can maintain a sterile barrier for a defined period of time.Sterility of container contents is maintained after processing for 365 days under conditions which simulate hospital sterile package handling and storage conditions. Biological indicators must be negative for growth after incubation period.Pass

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample sizes used for each specific test or the data provenance (e.g., country of origin, retrospective/prospective). It generally refers to "testing conducted in accordance with applicable FDA guidance and standards." For example, for "Sterilization Efficacy" and "Sterility Maintenance," it mentions using "Biological Indicators" (BI), implying an appropriate number of BIs were used to validate the SAL.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This information is not provided in the document. The tests described are primarily in vitro/laboratory-based performance tests for a sterilization container, rather than clinical studies requiring expert interpretation of medical images or patient outcomes.

4. Adjudication Method for the Test Set:

This information is not applicable and therefore not provided. The tests described are objective, measurable performance tests with clear pass/fail criteria, not subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done. Such studies are typically conducted for devices that involve human interpretation of diagnostic or screening results, which is not the function of a sterilization container. The studies performed are directed at the physical and biological effectiveness of the container in sterilization and maintaining sterility.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

This question is not applicable to a physical medical device like a sterilization container. The performance tests are inherently "standalone" in the sense that they evaluate the device itself.

7. The Type of Ground Truth Used:

The ground truth for the performance tests is established by:

  • Sterility Assurance Level (SAL) of 10^-6: This is a widely accepted standard for sterilization efficacy, and is verified using biological indicators (containing resistant bacterial spores).
  • Weight difference and visible moisture: For dry time, objective measurements and visual inspection define the ground truth for adequate drying.
  • Absence of microbial growth: For microbial aerosol challenge and sterility maintenance, the absence of bacterial growth from cultures of the container contents confirms sterility.
  • Killing of spores in organic soil: For simulated use, the elimination of a specified number of spores in a challenging environment serves as the ground truth for effective sterilization of internal lumens.
  • Functional integrity: For material compatibility, maintained functionality after multiple cycles serves as the ground truth.
  • Cytotoxicity (mild reactivity): Defined by ISO 10993-5: 2009 standards, where a response not greater than 2 (mild reactivity) indicates acceptable biocompatibility.

8. The Sample Size for the Training Set:

This information is not provided and is not applicable. The device is a physical sterilization container, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

This information is not provided and is not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

December 30, 2022

Aesculap Inc. Tim Stoudt Associate Director- Regulatory Affairs 3773 Corporate Parkway Center Valley, Pennsylvania 18034

Re: K214041

Trade/Device Name: Aesculap AICON™ Container Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: November 30, 2022 Received: December 2, 2022

Dear Tim Stoudt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray III -S

Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K214041

Device Name AESCULAP Aicon™ Container

Indications for Use (Describe)

The AESCULAP Aicon™ Container is a reusable rigid sterilization container intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device during transport and until used. This container is compatible for use in the following sterilization modalities in the configurations listed below:

  • · Ethylene Oxide (EtO)
  • STERIS V-PRO maX/maX 2: Lumen, Non-Lumen, Flex
  • · STERIS V-PRO 60: Lumen, Non- Lumen, Flex
  • · STERRAD 100NX: Standard, Express, Flex, Duo
  • STERRAD NX: Standard, Advanced
  • · STERRAD 100S
  • STERIZONE VP4
  • PreVac Steam

The AESCULAP Aicon™ consists of two different container styles: a solid bottom container and enhanced drying system (EDS) container.

The following tables identify the load configurations.

Sterilization CycleContainer SizesCompatible Lumens
Ethylene Oxide (EtO)(130°F, 60 minute exposure50-70% RH, concentration 735mg/L)Full Size(2) Stainless Steel Lumens (≥3.0mm ID x ≤ 400mm L)
Three-Quarter Size(2) Stainless Steel Lumens (≥3.0mm ID x ≤ 400mm L)
Half Size(2) Stainless Steel Lumens (≥3.0mm ID x ≤ 400mm L)
STERIS V-PRO maX /maX 2Lumen CycleFull Size(6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L)
Three-Quarter Size(6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L)
Half Size(6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L)(6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L)
(12) Stainless Steel Lumens (≥1mm ID x ≤ 150mm L)
STERIS V-PRO Max/ Max 2Non-Lumen CycleFull SizeNon lumened stainless steel instruments
Three-Quarter SizeNon lumened stainless steel instruments
Half SizeNon lumened stainless steel instruments
STERIS V-PRO maX / maX 2Flex CycleFull Size(1) Flexible endoscope with a light cord (if not integral toendoscope) without any additional load. The flexible endoscope may be single lumenthat is ≥1mm ID x ≤1050mm L
Three-Quarter Size(1) Flexible endoscope with a light cord (if not integral toendoscope) without any additional load. The flexible endoscope may be single lumenthatis ≥1mm ID x ≤1050mm L
Half Size(2) Flexible endoscopes with a light cord (if not integral to

Table 1. AESCULAP Aicon™ Container Lumen Configurations

{3}------------------------------------------------

endoscope) without any additional load. The flexible endoscope may be single lumenthat is ≥1mm ID x ≤1050mm L or two lumens with one ≥ 1 mm ID and ≤ 990 mm Land the other ≥ 1 mm ID and ≤ 850 mm LOR(1) flexible endoscope with a light cord (if not integralto endoscope) and additional non-lumened instruments. The flexible endoscope canhave:a single lumen that is ≥ 1mm ID and ≤ 1050 L or two lumens with one ≥ 1 mm ID and< 990 L and the other ≥ 1 mm ID and ≤ 850 mm L
Full Size(6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L)
(6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L)
STERIS VPRO 60Three-Quarter Size(6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L)
Lumen Cycle(6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L)
Half Size(12) Stainless Steel Lumens (≥1mm ID x ≤ 150mm L)
STERIS VPRO 60Full SizeNon lumened stainless steel instruments
Non-Lumen CycleThree-Quarter SizeNon lumened stainless steel instruments
Half SizeNon lumened stainless steel instruments
Full Size(1) flexible surgical endoscope or bronchoscope with alight cord (if not integral to scope) and single lumen (≥1mm ID x ≤990mm L)
STERIS VPRO 60Three-Quarter Size(1) flexible surgical endoscope or bronchoscope with alight cord (if not integral to scope) and single lumen (≥1mm ID x ≤990mm L)
Flex Cycle
Half Size(1) flexible surgical endoscope or bronchoscope with alight cord (if not integral to scope) and single lumen or dual lumens(≥1mm ID x ≤990mm L)
Full Size(5) Stainless-Steel Lumens (≥1.0mm ID x ≤500mm L)
(5) Stainless-Steel Lumens (≥0.7mm ID x ≤400mm L)
STERRAD 100NXThree-Quarter Size(5) Stainless-Steel Lumens (≥1.0mm ID x ≤500mm L)
Standard Cycle(5) Stainless-Steel Lumens (≥0.7mm ID x ≤400mm L)
Half Size(5) Stainless-Steel Lumens (≥0.7mm ID x ≤250mm L)
(5) Stainless-Steel Lumens (≥0.7mm ID x ≤200mm L)
Full Size(5) Stainless Steel Lumens (≥3.0mm I.D. x ≤ 400mm L)
STERRAD 100SThree-Quarter Size(5) Stainless Steel Lumens (≥3.0mm I.D. x ≤ 400mm L)
Half Size(5) Stainless Steel Lumens (≥2.0mm ID x ≤ 250mm L)
Full Size(10)Stainless-steel lumens (≥2.5mm ID x ≤ 400mm L)
STERRAD NXThree-Quarter Size(10)Stainless-steel lumens (≥2.5mm ID x ≤ 400mm L)
Standard Cycle
Half Size(10) Stainless-steel lumens (≥2mm ID x ≤ 200mm L)
Full Size(5) Stainless-steel lumens (≥2.5mm ID x ≤500mm L)
(5) Flexible Lumens (≥1mm ID x ≤850mm L)
STERRAD NXAdvanced CycleThree-Quarter Size(5) Stainless-steel lumens (≥2.5mm ID x ≤500mm L)
Half Size(5) Flexible Lumens (≥1mm ID x ≤850mm L)(10) Stainless-steel lumens (≥2mm ID x ≤ 200mm L)
STERRAD 100NXFull SizeNon lumened stainless steel instruments
Express CycleThree-Quarter SizeNon lumened stainless steel instruments
Full SizeNon lumened stainless steel instruments
STERRAD 100NXFull Size(1) Flexible Lumen (≥1mm ID x ≤ 850mm L)
DUO CycleThree-Quarter Size(1) Flexible Lumen (≥1mm ID x ≤ 850mm L)
Half Size(1) Flexible Lumen (≥1.mm ID x ≤ 650mm L)
Full Size(1) Flexible Lumen(≥1.5mm ID x ≤ 650mm L)
STERRAD 100NXThree-Quarter Size(1) Flexible Lumen (≥1.5mm ID x ≤ 650mm L)
Flex Cycle
Half Size(1) Flexible lumen (≥1.mm ID x ≤ 650mm L)
Full Size(4) Stainless Steel Lumens (≥1.mm ID x ≤200mm L)
PreVac Steam with dry time
(270°F, 4 min Exposure)Three-Quarter(4) Stainless Steel Lumens (≥1.0mm ID x ≤200mm L)
Half Size(4) Stainless Steel Lumens (≥1.0mm ID x ≤200mm L)
PreVac Steam without dry timeFull Size(4) Stainless Steel Lumens (≥1mm ID x ≤ 125 mm L)
(270°F, 4 min Exposure)Three-Quarter Size(4) Stainless Steel Lumens (≥1mm ID x ≤ 125 mm L)
PorousHalf Size(4) Stainless Steel Lumens (≥1mm ID x ≤ 125 mm L)
Prevac cycles without dry times arePrevac cycles without dry times areImmediate Use Sterilization Cycles(IUSS).

{4}------------------------------------------------

(270°F, 3 min Exposure)Full SizeNon-lumened instruments
Non-PorousThree-Quarter SizeNon-lumened instruments
Prevac cycles without dry times areImmediate Use Sterilization Cycles(IUSS).Half SizeNon-lumened instruments
STERIZONE VP4Load 8(Aicon Validated Load 1)Full Size(1) Dual channel flexible scope (≥1.5mm ID x ≤850mm L) and(≥1.5mm ID x ≤989mm L)(1) Single channel flexible scope (≥1.5mm ID x ≤850mm L)
Three-Quarter Size(1) Dual channel flexible scope (≥1.5mm ID x ≤850mm L) and(≥1.5mm ID x ≤989mm L)(1) Single channel flexible scope (≥1.5mm ID x ≤850mm L)
Half Size(1) Dual channel flexible scope (≥1.5mm ID x ≤850mm L) and(≥1.5mm ID x ≤989mm L)(1) Single channel flexible scope (≥1.5mm ID x ≤850mm L)
STERIZONE VP4Load 4(Aicon Validated Load 2)Full Size(1) Dual channel semi rigid scope (≥0.7mm ID x 500mm L; ≥1.1mm ID x ≤500mm L)(1) Dual channel semi rigid scope (≥1.1 mm ID x 500mm L; ≥1.1 mm ID x ≤500mm L)(2) Stainless Steel Lumens (≥2mm ID x ≤250mm L)(2) Stainless Steel Lumens (≥3mm ID x ≤400mm L)
Three-Quarter Size(1) Dual channel semi rigid scope (≥0.7mm ID x 500mm L; ≥1.1mm ID x ≤500mm L)(1) Dual channel semi rigid scope (≥1.1 mm ID x 500mm L; ≥1.1 mm ID x ≤500mm L)(2) Stainless Steel Lumens (≥2mm ID x ≤250mm L)(2) Stainless Steel Lumens (≥3mm ID x ≤400mm L)
Half Size(1) Stainless Steel Lumen (≥0.7 mm ID x ≤300mm L)(1) Stainless Steel Lumen (≥1.0 mm ID x ≤300mm L)(2) Stainless Steel Lumens (≥2.0 mm ID x ≤250mm L)(2) Stainless Steel Lumens (≥3.0 mm ID x ≤300mm L)
STERIZONE VP4Load 7(Aicon Validated Load 3)Full SizeNon-Lumened Stainless Steel Instruments (including hinges, luerlocks)
Three-Quarter SizeNon-Lumened Stainless Steel Instruments (including hinges, luerlocks)
Half SizeNon-Lumened Stainless Steel Instruments (including hinges, luerlocks)

Table 2. AESCULAP Aicon™ Container Configurations

SterilizationMethodContainer SizeStandardContainerEDSContainerContainerLid Part #Total LoadedContainer (lbs)
EtO(130°F, 60 minute exposure50-70% RH, concentration 735mg/L)Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131JJ41025
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ43025
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231JJ42025
Half Size 10"JJ240JJ241
STERIS VPROmaX/ maX 2 LumenFull Size 4"JJ110JJ111
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131JJ41019.65
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ43019.65
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221JJ42013.8
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ11125
STERIS VPROmaX / maX 2 Non-LumenFull Size 6"JJ120JJ121JJ410
Full Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ31125
Three-Quarter Size 6"JJ320JJ321JJ430
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221JJ420
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ111(1) Flexible endoscope witha light cord (if not integral toendoscope) without anyadditional load. The flexibleendoscope may be singlelumen that is ≥1mm ID x<1050mm L
Full Size 6"JJ120JJ121JJ410
Full Size 8"JJ130JJ131
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ430
Three-Quarter Size 10"JJ340JJ341
STERIS VPROmaX / maX 2 FlexHalf Size 4"JJ210JJ211JJ420(2) Flexible endoscopeswith a light cord (if notintegral toendoscope) without anyadditional load. Theflexible endoscope may besingle lumen that is ≥1mmID x ≤1050mm L or twolumens with one ≥ 1 mmID and ≤ 990 mm L andthe other ≥ 1 mm ID and< 850 mm LOR(1) flexible endoscopewith a light cord (if notintegral to endoscope) andadditional non-lumenedinstruments (24lbs). Theflexible endoscope canhave: a single lumen that is≥ 1mm ID and ≤ 1050 L ortwo lumens with one≥ 1 mm ID and ≤ 990 Land the other ≥ 1 mm IDand ≤ 850 mm L
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
STERIS VPRO 60 LumenFull Size 4"JJ110JJ111JJ41011
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ43011
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ21111
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231JJ420
Half Size 10"JJ240JJ241
STERIS VPRO 60 Non-LumenFull Size 4"JJ110JJ111
Full Size 6"JJ120JJ121JJ41012
Full Size 8"JJ130JJ131
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ43012
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ22112
Half Size 8"JJ230JJ231JJ420
Half Size 10"JJ240JJ241
STERIS VPRO 60 FlexFull Size 4"JJ110JJ111(1) flexible surgicalendoscope or bronchoscopewith a light cord (if notintegral to scope) and singlelumen (≥1mm ID x<990mm L )
Full Size 6"JJ120JJ121JJ410
Full Size 8"JJ130JJ131
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321JJ430(1) flexible surgicalendoscope or bronchoscopewith a light cord (if notintegral to scope) and singlelumen (≥1mm ID x<990mm L )
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211JJ420(1) flexible surgicalendoscope or bronchoscopewith a light cord (if notintegral to scope) and single
Half Size 6"JJ220JJ221lumen (≥1mm ID x≤990mm L )
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
STERRAD 100NX StandardFull Size 4"JJ110JJ11121.4
Full Size 6"JJ120JJ121JJ410
Full Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ31121.4
Three-Quarter Size 6"JJ320JJ321JJ430
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
STERRAD 100SHalf Size 4"JJ210JJ211
Half Size 6"JJ220JJ221JJ42013.8
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ11113.9
Full Size 6"JJ120JJ121JJ410
Full Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ31113.9
Three-Quarter Size 6"JJ320JJ321JJ430
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221JJ42013.9
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
STERRAD NX StandardFull Size 4"JJ110JJ11110.7
Full Size 6"JJ120JJ121JJ410
Full Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ31110.7
Three-Quarter Size 6"JJ320JJ321JJ430
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221JJ42010.7
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
STERRAD NX AdvancedFull Size 4"JJ110JJ111JJ41010.7
Full Size 6"JJ120JJ121
Three-Quarter Size 4"JJ310JJ311JJ43010.7
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221JJ42010.7
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121JJ41010.7
Full Size 8"JJ130JJ131
STERRAD 100NX ExpressThree-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321JJ43010.7
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231JJ42010.7
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121
STERRAD 100NX DUOFull Size 8"JJ130JJ131JJ41013.2
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321JJ43013.2
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231JJ4206.0
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121JJ41014.7
Full Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
STERRAD 100NX FlexThree-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321JJ43014.7
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221JJ42010.7
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
Prevac Steam270°F, 4 min exposureFull Size 4"JJ110JJ11125
Full Size 6"JJ120JJ121JJ41024
With silicone mat and lid cover*30 minute dry timeFull Size 8"JJ130JJ13123
Full Size 10"JJ140JJ14122
Without silicone mat and lidThree-Quarter Size 4"JJ310JJ311JJ43023
Three-Quarter Size 6"cover*
20 minute dry timeThree-Quarter Size 8"JJ330JJ33120
Three-Quarter Size 10"JJ340JJ34118
Half Size 4"JJ210JJ21118
Half Size 6"JJ220JJ22118
Half Size 8"JJ230JJ231JJ42015
Half Size 10"JJ240JJ24113
Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131JJ41025
PreVac Steam without dry time270°FFull Size 10"JJ140JJ141
3 min Exposure(Non-porous)Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ43025
ANDThree-Quarter Size 10"JJ340JJ341
Prevac Steam without dry time270°FHalf Size 4"JJ210JJ211
4 min Exposure(Porous)Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231JJ42025
Half Size 10"JJ240JJ241
Prevac cycles without dry timesare Immediate Use SterilizationCycles (IUSS).
Full Size 4"JJ110JJ111JJ410Based on Load Configuration
Full Size 6"JJ120JJ121in Table 1
Full Size 8"JJ130JJ13113.8lbs
STERIZONE VP4Based on Sterizone ValidatedLoad 8Three-Quarter Size 4"JJ310JJ311JJ430
(Aicon Validated Load 1)Three-Quarter Size 6"JJ320JJ321Based on Load Configuration
Three-Quarter Size 8"JJ330JJ331in Table 1
Three-Quarter Size 10"JJ340JJ34113.8lbs
Half Size 4"JJ210JJ211JJ420
Half Size 6"JJ220JJ221Based on Load Configuration
Half Size 8"JJ230JJ231in Table 1
Half Size 10"JJ240JJ2415.8lbs
Full Size 4"JJ110JJ111
Based on Load Configuration
Full Size 6"Full Size 8"JJ120JJ121JJ410in Table 1
JJ130JJ13111.7lbs
Three-Quarter Size 4"JJ310JJ311
STERIZONE VP4Based on Sterizone ValidatedLoad 4Three-Quarter Size 6"JJ320JJ321Based on Load Configuration
(Aicon Validated Load 2)Three-Quarter Size 8"JJ330JJ331JJ430in Table 1
Three-Quarter Size 10"JJ340JJ34111.7lbs
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221Based on Load Configuration
JJ420in Table 1
Half Size 8"JJ230JJ2315.8lbs
Half Size 10"JJ240JJ241
Full Size 4"Full Size 6"JJ110JJ111
Full Size 8"JJ130JJ131JJ41025
STERIZONE VP4Based on Sterizone ValidatedLoad 7Three-Quarter Size 4"JJ310JJ311
(Aicon Validated Load 3)Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ43025
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211JJ42025
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

Aesculap AICON Containers may be used with commercially available Aesculap accessories according to the table below:

AccessoriesEtOSteris V-PROSystemsSTERRADSystemsSTERIZONEVP4PREVACSTEAM
Baskets, trays, indicator cards,tamper evident locks, faceplateholders, and tags.YesYesYesYesYes
Instrument Organization System(Silicone and Stainless Steel racks,brackets, holders, and clamps)YesNo Silicone PiecesNo Silicone PiecesNo Silicone PiecesYes
Silicone matsNoNoNoNoYes*
Lid CoversNoNoNoNoYes*
  • Use of silicone mats and/or lid covers may require a longer drying time.

Silicone mats and lid covers are compatible with PreVac steam only. Silicone pieces should not be used with Ethylene Oxide, Steris V-Pro Systems, STERRAD Systems, or STERIZONE VP4 Systems.

Healthcare Facilities should follow steam sterilization cycle guidelines as stated in AAM ST79, Comprehensive guide to steam sterilization and sterility assurance in health care facilities.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

& Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."

{11}------------------------------------------------

510(k) SUMMARY K214041

Aesculap Aicon™ Container

November 30, 2022

COMPANY:Aesculap, Inc.3773 Corporate Parkway Center Valley, PA 18034Establishment Registration Number: 2916714
CONTACT:Tim Stoudt tim.stoudt@bbraunusa.com(484) 523-4581610-791-6882 (fax)
TRADE NAME:Aesculap Aicon™ Container
COMMON NAME:Sterilization Container Wrap
CLASSIFICATION NAME:Wrap, Sterilization
REGULATION NUMBER:880.6850
PRODUCT CODE:KCT
DEVICE CLASS:Class II per 21 CFR 880.6850
PREDICATE DEVICE
TRADE NAME:Aesculap JJ Series Container System
510(k) NUMBER:K192056
COMMON NAME:Sterilization Container Wrap
CLASSIFICATION NAME:Wrap, Sterilization
REGULATION NUMBER:880.6850
PRODUCT CODE:KCT
DEVICE CLASS:Class II per 21 CFR 880.6850

DEVICE DESCRIPTION

The AESCULAP Aicon Container is a reusable container system intended for sterilization and storage of other medical devices. This container system is compatible for use in the following sterilization modalities:

  • Ethylene Oxide
  • · STERIS VPRO maX/ maX 2: Lumen, Non- Lumen, Flex,
  • STERIS VPRO 60: Lumen, Non-Lumen, Flex
  • STERRAD 100NX: Standard, Express, Flex, Duo

{12}------------------------------------------------

  • STERRAD NX: Standard, Advanced
  • · STERRAD 100S
  • · STERIZONE VP4
  • · PreVac Steam

The containers are perforated and made from anodized aluminum and utilize a single-use filter.

INDICATIONS FOR USE

The AESCULAP Aicon™ Container is a reusable rigid sterilization container intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device during transport and until used. This container is compatible for use in the following sterilization modalities in the configurations listed below:

  • Ethylene Oxide (EtO)
  • STERIS V-PRO maX/maX 2: Lumen, Non-Lumen, Flex
  • STERIS V-PRO 60: Lumen, Non- Lumen, Flex
  • · STERRAD 100NX: Standard, Express, Flex, Duo
  • · STERRAD NX: Standard, Advanced
  • STERRAD 100S
  • STERIZONE VP4
  • PreVac Steam

The AESCULAP Aicon™ consists of two different container styles: a solid bottom container and enhanced drying system (EDS) container.

The following tables identify the load configurations.
Table 1. AESCULAP Aicon™ Container Lumen Configurations
Sterilization CycleContainer SizesCompatible Lumens
Ethylene Oxide (EtO)(130°F, 60 minute exposure50-70% RH, concentration735mg/L)Full Size(2) Stainless Steel Lumens (≥3.0mm ID x ≤ 400mm L)
Three-QuarterSize(2) Stainless Steel Lumens (≥3.0mm ID x ≤ 400mm L)
Half Size(2) Stainless Steel Lumens (≥3.0mm ID x ≤ 400mm L)
STERIS V-PRO maX /maX 2Lumen CycleFull Size(6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L)(6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L)
Three-QuarterSize(6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L)(6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L)
Half Size(12) Stainless Steel Lumens (≥1mm ID x ≤ 150mm L)
STERIS V-PRO Max/ Max 2Non-Lumen CycleFull SizeNon lumened stainless steel instruments
Three-QuarterSizeNon lumened stainless steel instruments
Half SizeNon lumened stainless steel instruments
STERIS V-PRO maX / maX 2Flex CycleFull Size(1) Flexible endoscope with a light cord (if not integral toendoscope) without any additional load. The flexible endoscope may be single lumen that is≥1mm ID x ≤1050mm L
Three-QuarterSize(1) Flexible endoscope with a light cord (if not integral toendoscope) without any additional load. The flexible endoscope may be single lumen thatis ≥1mm ID x ≤1050mm L
Half Size(2) Flexible endoscopes with a light cord (if not integral toendoscope) without any additional load. The flexible endoscope may be single lumen that is≥1mm ID x ≤1050mm L or two lumens with one ≥ 1 mm ID and ≤ 990 mm L and the other ≥1 mm ID and ≤ 850 mm LOR(1) flexible endoscope with a light cord (if not integral
to endoscope) and additional non-lumened instruments. The flexible endoscope can have:a single lumen that is ≥ 1mm ID and ≤ 1050 L or two lumens with one ≥ 1 mm ID and
≤ 990 L and the other ≥ 1 mm ID and ≤ 850 mm L
STERIS VPRO 60Lumen CycleFull Size(6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L)
STERIS VPRO 60Lumen CycleThree-QuarterSize(6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L)
Half Size(12) Stainless Steel Lumens (≥1mm ID x ≤ 150mm L)
STERIS VPRO 60Non-Lumen CycleFull SizeNon lumened stainless steel instruments
STERIS VPRO 60Non-Lumen CycleThree-QuarterSizeNon lumened stainless steel instruments
STERIS VPRO 60Non-Lumen CycleHalf SizeNon lumened stainless steel instruments
STERIS VPRO 60Flex CycleFull Size(1) flexible surgical endoscope or bronchoscope with alight cord (if not integral to scope) and single lumen (≥1mm ID x ≤990mm L)
STERIS VPRO 60Flex CycleThree-QuarterSize(1) flexible surgical endoscope or bronchoscope with alight cord (if not integral to scope) and single lumen (≥1mm ID x ≤990mm L)
STERIS VPRO 60Flex CycleHalf Size(1) flexible surgical endoscope or bronchoscope with alight cord (if not integral to scope) and single lumen or dual lumens(≥1mm ID x ≤990mm L)
STERRAD 100NXStandard CycleFull Size(5) Stainless-Steel Lumens (≥1.0mm ID x ≤500mm L)
STERRAD 100NXStandard CycleThree-QuarterSize(5) Stainless-Steel Lumens (≥0.7mm ID x ≤400mm L)
STERRAD 100NXStandard CycleHalf Size(5) Stainless-Steel Lumens (≥0.7mm ID x ≤250mm L)
STERRAD 100SFull Size(5) Stainless Steel Lumens (≥3.0mm I.D. x ≤ 400mm L)
STERRAD 100SThree-QuarterSize(5) Stainless Steel Lumens (≥3.0mm I.D. x ≤ 400mm L)
STERRAD 100SHalf Size(5) Stainless Steel Lumens (≥2.0mm ID x ≤ 250mm L)
STERRAD NXStandard CycleFull Size(10)Stainless-steel lumens (≥2.5mm ID x ≤ 400mm L)
STERRAD NXStandard CycleThree-QuarterSize(10)Stainless-steel lumens (≥2.5mm ID x ≤ 400mm L)
STERRAD NXStandard CycleHalf Size(10) Stainless-steel lumens (≥2mm ID x ≤ 200mm L)
STERRAD NXAdvanced CycleFull Size(5) Stainless-steel lumens (≥2.5mm ID x ≤500mm L)(5) Flexible Lumens (≥1mm ID x ≤850mm L)
STERRAD NXAdvanced CycleThree-QuarterSize(5) Stainless-steel lumen (≥2.5mm ID x ≤500mm L)(5) Flexible Lumens (≥1mm ID x ≤850mm L)
STERRAD NXAdvanced CycleHalf Size(10) Stainless-steel lumens (≥2mm ID x ≤ 200mm L)
STERRAD 100NXExpress CycleFull SizeNon lumened stainless steel instruments
STERRAD 100NXExpress CycleThree-QuarterSizeNon lumened stainless steel instruments
STERRAD 100NXExpress CycleFull SizeNon lumened stainless steel instruments
STERRAD 100NXDUO CycleFull Size(1) Flexible Lumen (≥1mm ID x ≤ 850mm L)
STERRAD 100NXDUO CycleThree-QuarterSize(1) Flexible Lumen (≥1mm ID x ≤ 850mm L)
STERRAD 100NXDUO CycleHalf Size(1) Flexible Lumen (≥1.mm ID x ≤ 650mm L)
STERRAD 100NXFlex CycleFull Size(1) Flexible Lumen(≥1.5mm ID x ≤ 650mm L)
STERRAD 100NXFlex CycleThree-QuarterSize(1) Flexible Lumen (≥1.5mm ID x ≤ 650mm L)
STERRAD 100NXFlex CycleHalf Size(1) Flexible lumen (≥1.mm ID x ≤ 650mm L)
PreVac Steam withdry time (270°F, 4min Exposure)Full Size(4) Stainless Steel Lumens (≥1.mm ID x ≤200mm L)
PreVac Steam withdry time (270°F, 4min Exposure)Three-Quarter(4) Stainless Steel Lumens (≥1.0mm ID x ≤200mm L)
PreVac Steam withdry time (270°F, 4min Exposure)Half Size(4) Stainless Steel Lumens (≥1.0mm ID x ≤200mm L)
Size
PreVac Steamwithout dry time(270°F, 4 minExposure) PorousFull Size(4) Stainless Steel Lumens (≥1mm ID x ≤ 125 mm L)
Three-Quarter Size(4) Stainless Steel Lumens (≥1mm ID x ≤ 125 mm L)
Half Size(4) Stainless Steel Lumens (≥1mm ID x ≤ 125 mm L)
Prevac cycles without dry timeare Immediate Use SterilizationCycles (IUSS).
Prevac Steamwithout dry time(270°F, 3 minExposure)Non-PorousFull SizeNon-lumened instruments
Three-Quarter SizeNon-lumened instruments
Half SizeNon-lumened instruments
Prevac cycles without dry timeare Immediate Use SterilizationCycles (IUSS).
STERIZONE VP4Load 8(Aicon Validated Load 1)Full Size(1) Dual channel flexible scope (≥1.5mm ID x ≤850mm L) and (≥1.5mmID x ≤989mm L)(1) Single channel flexible scope (≥1.5mm ID x ≤850mm L)
Three-Quarter Size(1) Dual channel flexible scope (≥1.5mm ID x ≤850mm L) and (≥1.5mmID x ≤989mm L)(1) Single channel flexible scope (≥1.5mm ID x ≤850mm L)
Half Size(1) Dual channel flexible scope (≥1.5mm ID x ≤850mm L) and (≥1.5mmID x ≤989mm L)(1) Single channel flexible scope (≥1.5mm ID x ≤850mm L)
STERIZONE VP4Load 4(Aicon Validated Load 2)Full Size(1) Dual channel semi rigid scope (≥0.7mm ID x 500mm L; ≥1.1mm ID x ≤500mm L)(1) Dual channel semi rigid scope (≥1.1 mm ID x 500mm L; ≥1.1 mm ID x ≤500mm L)(2) Stainless Steel Lumens (≥2mm ID x ≤250mm L)(2) Stainless Steel Lumens (≥3mm ID x ≤400mm L)
Three-Quarter Size(1) Dual channel semi rigid scope (≥0.7mm ID x 500mm L; ≥1.1mm ID x ≤500mm L)(1) Dual channel semi rigid scope (≥1.1 mm ID x 500mm L; ≥1.1 mm ID x ≤500mm L)(2) Stainless Steel Lumens (≥2mm ID x ≤250mm L)(2) Stainless Steel Lumens (≥3mm ID x ≤400mm L)
Half Size(1) Stainless Steel Lumen (≥0.7 mm ID x ≤300mm L)(1) Stainless Steel Lumen (≥1.0 mm ID x ≤300mm L)(2) Stainless Steel Lumens (≥2.0 mm ID x ≤250mm L)(2) Stainless Steel Lumens (≥3.0 mm ID x ≤300mm L)
STERIZONE VP4Load 7(Aicon Validated Load 3)Full SizeNon-Lumened Stainless Steel Instruments (including hinges, luerlocks)
Three-Quarter SizeNon-Lumened Stainless Steel Instruments (including hinges, luerlocks)
Half SizeNon-Lumened Stainless Steel Instruments (including hinges, luerlocks)

{13}------------------------------------------------

{14}------------------------------------------------

Table 2. AESCULAP Aicon™ Container Configurations
SterilizationMethodContainer SizeStandardContainerEDSContainerContainerLid Part #Total LoadedContainer (lbs)
EtO(130°F, 60 minute exposure50-70% RH, concentration 735mg/L)Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131JJ41025
Full Size 10"JJ140JJ141
EtO(130°F, 60 minute exposure50-70% RH, concentration 735mg/L)Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ43025
Three-Quarter Size 10"JJ340JJ341
EtO(130°F, 60 minute exposure50-70% RH, concentration 735mg/L)Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231JJ42025
Half Size 10"JJ240JJ241
EtO(130°F, 60 minute exposure50-70% RH, concentration 735mg/L)Full Size 4"JJ110JJ111
STERIS VPROmaX/ maX 2 LumenFull Size 6"JJ120JJ121JJ41019.65
Full Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321JJ43019.65
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211JJ42013.8
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
STERIS VPROmaX/maX 2 Non-LumenFull Size 4"JJ110JJ111JJ41025
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311JJ43025
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211JJ42018.7
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
STERIS VPROmaX/maX 2 FlexFull Size 4"JJ110JJ111JJ410(1) Flexible endoscope witha light cord (if not integral toendoscope) without anyadditional load. The flexibleendoscope may be singlelumen that is ≥1mm ID x<1050mm L
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131
Three-Quarter Size 4"JJ310JJ311JJ430(1) Flexible endoscope witha light cord (if not integral toendoscope) without anyadditional load. The flexibleendoscope may be singlelumen that is ≥1mm ID x<1050mm L
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211JJ420(2) Flexible endoscopeswith a light cord (if notintegral toendoscope) without anyadditional load. The
Half Size 6"JJ220JJ221flexible endoscope may besingle lumen that is ≥1mmID x ≤1050mm L or twolumens with one ≥ 1 mmID and ≤ 990 mm L andthe other ≥ 1 mm ID and≤ 850 mm LOR(1) flexible endoscopewith a light cord (if notintegral to endoscope) andadditional non-lumenedinstruments (24lbs). Theflexible endoscope canhave: a single lumen that is
Half Size 8"JJ230JJ231≥ 1mm ID and ≤ 1050 L ortwo lumens with one≥ 1 mm ID and ≤ 990 Land the other ≥ 1 mm IDand ≤ 850 mm L
Half Size 10"JJ240JJ241
STERIS VPRO 60 LumenFull Size 4"JJ110JJ111JJ41011
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ43011
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221JJ42011
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
STERIS VPRO 60 Non-LumenFull Size 4"JJ110JJ111
Full Size 6"JJ120JJ121JJ41012
Full Size 8"JJ130JJ131
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321JJ43012
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ22112
Half Size 8"JJ230JJ231JJ420
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121JJ410(1) flexible surgicalendoscope or bronchoscopewith a light cord (if notintegral to scope) and singlelumen (≥1mm ID x<990mm L )
Full Size 8"JJ130JJ131
Three-Quarter Size 4"JJ310JJ311
STERIS VPRO 60 FlexThree-Quarter Size 6"JJ320JJ321JJ430
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211JJ420(1) flexible surgicalendoscope or bronchoscopewith a light cord (if notintegral to scope) and singlelumen (≥1mm ID x<990mm L )
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121JJ41021.4
STERRAD 100NX StandardFull Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ43021.4
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231JJ42013.8
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ111
STERRAD 100SFull Size 6"JJ120JJ121JJ41013.9
Full Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311JJ43013.9
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211JJ42013.9
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
STERRAD NX StandardFull Size 4"JJ110JJ111JJ41010.7
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311JJ43010.7
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211JJ42010.7
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
STERRAD NX AdvancedFull Size 4"JJ110JJ111JJ41010.7
Full Size 6"JJ120JJ121
Three-Quarter Size 4"JJ310JJ311JJ430
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331
Half Size 4"JJ210JJ211JJ42010.7
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
STERRAD 100NX ExpressFull Size 4"JJ110JJ111JJ41010.7
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131JJ430
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231JJ42010.7
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121JJ41013.2
STERRAD 100NX DUOFull Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321JJ43013.2
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221JJ4206.0
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121JJ41014.7
Full Size 8"JJ130JJ131
Full Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311
STERRAD 100NX FlexThree-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ331JJ43014.7
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211
Half Size 6"JJ220JJ221JJ42010.7
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
Prevac Steam270°F, 4 min exposureFull Size 4"JJ110JJ11125
Full Size 6"JJ120JJ121JJ41024
With silicone mat and lid cover*30 minute dry timeFull Size 8"JJ130JJ13123
Full Size 10"JJ140JJ14122
Without silicone mat and lidcover*20 minute dry timeThree-Quarter Size 4"JJ310JJ31123
Three-Quarter Size 6"JJ320JJ321JJ43022
Three-Quarter Size 8"JJ330JJ33120
Three-Quarter Size 10"JJ340JJ34118
Half Size 4"JJ210JJ21118
Half Size 6"JJ220JJ22118
Half Size 8"JJ230JJ231JJ42015
Half Size 10"JJ240JJ24113
Full Size 4"JJ110JJ111
Full Size 6"JJ120JJ121JJ41025
Full Size 8"JJ130JJ131
PreVac Steam without dry time270°F3 min Exposure(Non-porous)ANDFull Size 10"JJ140JJ141
Three-Quarter Size 4"JJ310JJ311
Three-Quarter Size 6"JJ320JJ321JJ43025
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ341
Prevac Steam without dry time270°F4 min Exposure(Porous)Half Size 6"JJ220JJ221JJ42025
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241
Prevac cycles without dry times areImmediate Use Sterilization Cycles(IUSS).
STERIZONE VP4Based on Sterizone ValidatedLoad 8(Aicon Validated Load 1)Full Size 4"JJ110JJ111JJ410Based on Load Configurationin Table 1
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ13113.8lbs
Three-Quarter Size 4"JJ310JJ311JJ430
Three-Quarter Size 6"JJ320JJ321Based on Load Configurationin Table 1
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ34113.8lbs
Half Size 4"JJ210JJ211JJ420
Half Size 6"JJ220JJ221Based on Load Configurationin Table 1
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ2415.8lbs
Full Size 4"JJ110JJ111JJ410Based on Load Configurationin Table 1
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ13111.7lbs
Three-Quarter Size 4"JJ310JJ311JJ430
STERIZONE VP4Based on Sterizone ValidatedLoad 4(Aicon Validated Load 2)Three-Quarter Size 6"JJ320JJ321Based on Load Configurationin Table 1
Three-Quarter Size 8"JJ330JJ331
Three-Quarter Size 10"JJ340JJ34111.7lbs
Half Size 4"JJ210JJ211JJ420
Half Size 6"JJ220JJ221Based on Load Configurationin Table 1
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ2415.8lbs
Full Size 4"JJ110JJ111JJ41025
Full Size 6"JJ120JJ121
Full Size 8"JJ130JJ131
STERIZONE VP4Based on Sterizone ValidatedLoad 7(Aicon Validated Load 3)Three-Quarter Size 4"JJ310JJ311JJ430
Three-Quarter Size 6"JJ320JJ321
Three-Quarter Size 8"JJ330JJ33125
Three-Quarter Size 10"JJ340JJ341
Half Size 4"JJ210JJ211JJ42025
Half Size 6"JJ220JJ221
Half Size 8"JJ230JJ231
Half Size 10"JJ240JJ241

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

{18}------------------------------------------------

{19}------------------------------------------------

{20}------------------------------------------------

Aesculap AICON Containers may be used with commercially available Aesculap accessories according to the table below:

AccessoriesEtOSteris V-PROSystemsSTERRADSystemsSTERIZONEVP4PREVACSTEAM
Baskets, trays, indicator cards,tamper evident locks, faceplateholders, and tags.YesYesYesYesYes
Instrument Organization System(Silicone and Stainless Steel racks,brackets, holders, and clamps)YesNo Silicone PiecesNo Silicone PiecesNo Silicone PiecesYes
Silicone matsNoNoNoNoYes*
Lid CoversNoNoNoNoYes*
  • Use of silicone mats and/or lid covers may require a longer drying time.

{21}------------------------------------------------

Silicone mats and lid covers are compatible with PreVac steam only. Silicone pieces should not be used with Ethylene Oxide, Steris V-Pro Systems, STERRAD Systems, or STERIZONE VP4 Systems.

Healthcare Facilities should follow steam sterilization cycle guidelines as laid out in AAMI ST79, Comprehensive guide to steam sterilization and sterility assurance in health care facilities.

TECHNOLOGICAL CHARACTERISTICS (compared to predicate device)

The table below provides a summary of the device technological characteristics comparing the subject device and predicate device. The subject device is offered in similar sizes and is comprised of similar materials of construction as the predicate device. The AESCULAP Aicon Container has the same intended use as the Aesculap predicate device.

CharacteristicSubject Device -Aesculap AiconContainer (JJ Series)(K214041)Product Code: KCTPrimary PredicateDevice - Aesculap JJSeries Container System(K192056)Product Code: KCTReference Device-Aesculap SterilContainerS2 System(K193582)Product Code: KCTReference Device-Aesculap SterilContainerS2 System(K182414 )Product Code: KCTDiscussion
Intended UseA device intended to beused to enclose anothermedical device that is to besterilized by a health careprovider. It is intended toallow sterilization of theenclosed medical deviceand also to maintainsterility of the encloseddevice until used.A device intended to beused to enclose anothermedical device that is to besterilized by a health careprovider. It is intended toallow sterilization of theenclosed medical deviceand also to maintainsterility of the encloseddevice until used.A device intended to beused to enclose anothermedical device that is to besterilized by a health careprovider. It is intended toallow sterilization of theenclosed medical deviceand also to maintainsterility of the encloseddevice until used.A device intended to beused to enclose anothermedical device that is to besterilized by a health careprovider. It is intended toallow sterilization of theenclosed medical deviceand also to maintainsterility of the encloseddevice until used.Same
SterilizationModalities• PreVac Steam with drytime• PreVac Steam withoutdry time• Ethylene Oxide• STERIS VPRO maX/maX2: Lumen, Non-Lumen,Flex,• STERIS VPRO 60:Lumen,Non- Lumen, Flex• STERRAD: 100S, NXStandard, NX Advanced,100NX Standard, 100NXFlex, 100 NX Express,100NX Duo• PreVac Steam with drytime• Prevac Steam withoutdry time• EtO• STERIZONE VP4• STERRAD 100NX DUO• Prevac Steam• Gravity Steam• STERIS VPRO maX:Lumen, Non-Lumen, Flex• STERIS VPRO 60:Lumen,Non-Lumen, Flex• STERRAD: 100S, NXStandard,NXAdvanced,100NXStandard, 100NX FlexSimilar
MaterialContainer:AnodizedaluminumLid: Anodized aluminumGasket: SiliconeContainer:AnodizedaluminumLid: Anodized aluminumGasket: SiliconeContainer:AnodizedaluminumLid: Anodized aluminumGasket: SiliconeContainer:AnodizedaluminumLid: Anodized aluminumGasket: SiliconeSame
Filter Type &MaterialSingle use (Polypropylene)Single use (Paper)Single use (Paper orpolypropylene)Single use (Paperor polypropylene)Reusable (PTFE)Similar
ContainerDesignSolid bottom orSolid bottom with enhanceddrying system withSolid bottom orSolid bottom with enhanceddrying system withPerforated bottom withperforated lidPerforated bottom withperforated lidSimilar

{22}------------------------------------------------

Perforated LidPerforated Lid
Vent tovolumeratios$0.04mm^{-1} * 10^{-3}$to$0.26mm^{-1} * 10^{-3}$$0.04mm^{-1} * 10^{-3}$to$0.26mm^{-1} * 10^{-3}$$0.50mm^{-1} * 10^{-3}$to$3.43mm^{-1} * 10^{-3}$$0.50mm^{-1} * 10^{-3}$to$3.43mm^{-1} * 10^{-3}$Similar
Maintenanceof sterility365 days365 days360 days360 daysSimilar

PERFORMANCE DATA

Testing was conducted in accordance with applicable FDA guidance and standards. The table below outlines the testing which was performed on the subject device.

PerformanceTestingPurposeAcceptance CriteriaResults
SterilizationEfficacyTo determine sterilization effectivenessof test device after processing in asterilization cycle.A sterility assurance level (SAL) of 10-6 willbe achieved post sterilization using the BIoverkill method and half cycle validation.Biological indicators must be negative forgrowth after incubation period.Pass
Dry TimeTo determine the proper drying timerequired for the Aesculap AiconContainer.The system shall demonstrate an average preand post sterilization weight difference ofless than 0.2% within five 5 minutes of cyclecompletion using final validated parametersand be free of visible moisture following acooling period.Pass
Microbial AerosolChallengeTo analyze the package integrity andmicrobial barrier properties of the testdevice.The container load maintains sterility afterexposure to a defined amount of aerosolmicroorganisms. No presence of growth afterincubation period.Pass
Simulated UseTo determine the effective sterilization offlexible scopes when used with the testdevice.A minimum of 1.0 x $10^6$ spores containedwithin organic soil representative of actualuse conditions are killed during definedsterilization cycle.Pass
MaterialCompatibilityTo assess effects of full use cycles ondevice components and their intendedfunctionality.No impact to functionality at the completionmultiple sterilization cycles.Pass
CytotoxicityTo determine the potential of a test deviceto cause cytotoxicity.Testing completed in accordance with ISO10993-5: 2009 to demonstrate no significantcytotoxic reaction after exposure to sterilant.Using the ISO Elution Method, the responseto the article is not greater than 2 (mildreactivity).Pass
SterilityMaintenanceTo demonstrate that a processed testdevice can maintain a sterile barrier for adefined period of time.Sterility of container contents is maintainedafter processing for 365 days underconditions which simulate hospital sterilepackage handling and storage conditions.Biological indicators must be negative forgrowth after incubation period.Pass

Results of the testing demonstrates that the proposed device met the acceptance criteria for each performance test.

{23}------------------------------------------------

CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate that Aesculap Aicon Container is as safe, as effective, and performs as well as or better than the legally marketed predicate device K192056.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).