(73 days)
Intended for making dental plates, bite plates, frame-works, clasps,personal trays, appliances, occlusal splints and night guards.
Flexible Block is a circular solid (disc) or rectangular solid (block) made from polyamide through high-temperature injection molding, and is used in a CAD/CAM milling machine for production of dental plates, bite plates, frame-works, clasps, personal trays, appliances, occlusal night guards. These blocks are available in a variety of shapes and sizes for different milling systems and in a variety of dental shades.
Here's a breakdown of the acceptance criteria and study information for the Flexible Block device, based on the provided FDA 510(k) summary:
This device is a material (dental polymer blanks) used in a CAD/CAM milling machine to create dental restorations, not a diagnostic AI device. Therefore, many of the typical AI acceptance criteria and study components (like expert ground truth, adjudication, MRMC studies, standalone algorithm performance, and training set details) are not applicable. The "acceptance criteria" here refer to the product's physical, chemical, and biological properties meeting established standards for similar dental materials.
1. A table of acceptance criteria and the reported device performance
| Comparison Item / Acceptance Criteria | New Device (Flexible Block) Performance | Predicate Device (Perflex Biosens K150454) |
|---|---|---|
| Material Composition | Polyamide | Super polyamide, thermoplastic compound |
| Tensile strength (based on ASTM D638) | ≥ 30 Mpa (The document states "Tensile strength: ≥30Mpa" as a reported property, implying this is the performance and implicitly the acceptance criteria for this specific property based on comparison to the predicate and relevant standards) | (Not explicitly stated for predicate in table, but overall comparison indicates compliance with relevant standards) |
| Elongation at break (based on ASTM D638) | ≥ 40% (Similar to tensile strength, this is a reported property implying acceptance criteria) | (Not explicitly stated for predicate) |
| Water absorption (based on ASTM D570) | ≤ 5% (Similar to tensile strength, this is a reported property implying acceptance criteria) | (Not explicitly stated for predicate) |
| Impact strength (based on ASTM D256) | ≥ 20 KJ/m² (Similar to tensile strength, this is a reported property implying acceptance criteria) | (Not explicitly stated for predicate) |
| Flexural strength (based on ASTM D790) | ≥ 30 Mpa (Similar to tensile strength, this is a reported property implying acceptance criteria) | (Not explicitly stated for predicate) |
| Flexural modulus (based on ASTM D790) | ≥ 1000 Mpa (Similar to tensile strength, this is a reported property implying acceptance criteria) | (Not explicitly stated for predicate) |
| Biocompatibility (ISO 10993 standards) - Non genotoxicity (ISO 10993-3) - Non cytotoxicity (ISO 10993-5) - Non sensitization or irritation (ISO 10993-10) - Non systemic toxicity (ISO 10993-11) | Met (Biocompatibility tests performed and "satisfy the ISO 10993 standards," including systemic toxicity, skin irritation, skin sensitization, in vitro cytotoxicity, in vitro mammalian cell TK gene mutation test, in vitro mammalian chromosome aberration test, and bacterial reverse mutation study) | NA (Predicate "Flexible Block" stated as "NA" for specific 10993 tests in the comparison table, but overall assumed biocompatible as a marketed device.) |
| Regulatory Classifications | Class II | Class II |
| Indications for Use | Similar (Intended for making dental plates, bite plates, frame-works, clasps, personal trays, appliances, occlusal splints and night guards.) | Similar (for making dental plates, bite plates, frame-works, clasps, personal trays, appliances, occlusal splints and night guards) |
| Contraindications | Same | Same |
| Labeling | Similar | Similar |
| Target Population | Dental patients | Dental patients |
| Anatomical Site | On teeth | On teeth |
| Where Used | Hospital, dental clinic, and relevant places | Hospital, dental clinic, and relevant places |
| Human Factors | Dental professional | Dental professional |
| Design | Circular solid (disc) or rectangular solid (block) of polyamide. | Circular solid (disc) or rectangular solid (block) of super polyamide and thermoplastic compound. |
| Cautions | Similar | Similar |
| Standards Met | Same (implicitly, by meeting specific ASTM and ISO standards) | Same (implicitly, by meeting specific ASTM and ISO standards) |
| Sterility | Non-sterile | Non-sterile |
| Chemical Safety | Similar | Similar |
Study to prove acceptance criteria:
The study performed was a series of bench tests to evaluate the physical, chemical, and biological properties of the Flexible Block device.
2. Sample size used for the test set and the data provenance
- Sample Size: Not explicitly stated in terms of number of blocks tested for each physical property. However, it mentions "We selected our Flexible Block (Model: monolayer; Specification: 98×20mm; Shade: Clear) as the representative in biocompatibility tests," implying a single representative model was used for biocompatibility.
- Data Provenance: The tests were conducted by the manufacturer, Shandong Huge Dental Material Corporation, based in China ("No. 68 Shanhai Road, Donggang District Rizhao City, Shandong 276800 China"). This is a retrospective evaluation of a newly manufactured material against established standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. For a material device, "ground truth" is established by adherence to physical, chemical, and biological standards (e.g., ASTM, ISO). There are no human "experts" establishing ground truth in the context of diagnostic performance here. The "experts" would be the personnel in the testing laboratories who perform the tests according to the specified standards.
4. Adjudication method for the test set
- Not Applicable. There is no adjudication method in the context of evaluating a material against objective physical and chemical standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is a material device, not an AI-enabled diagnostic tool. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a material device, not an algorithm.
7. The type of ground truth used
- The "ground truth" for this device is based on established material science standards (e.g., ASTM D638, ASTM D570, ASTM D256, ASTM D790) for physical properties, and international biocompatibility standards (ISO 10993 series) for biological safety.
8. The sample size for the training set
- Not Applicable. This is a material device, not a machine learning model. There is no concept of a "training set" in this context. The manufacturing process of the material could be seen as analogous to a "training set" in a very abstract sense (optimizing the material's properties), but it does not use a data-driven training set like AI models.
9. How the ground truth for the training set was established
- Not Applicable. As there is no training set for an AI model, there is no ground truth established for it. The development of the material's properties would be based on material science principles and iterative testing against target performance specifications.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 4, 2022
Shandong Huge Dental Material Corporation Maggie Zheng Regulatory Affairs Manager No. 68 Shanhai Road, Donggang District Rizhao City, Shandong 276800 China
Re: K213994
Trade/Device Name: Flexible Block Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI, MQC Dated: December 15, 2021 Received: December 21, 2021
Dear Maggie Zheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213994
Device Name Flexible Block
Indications for Use (Describe)
Intended for making dental plates, bite plates, frame-works, clasps,personal trays, appliances, occlusal splints and night guards.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510 (k) Summary K213994
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R. 807.92.
1. Date Summary Prepared: Feb. 18, 2022
2. Submitter Information:
| Name | Shandong Huge Dental Material Corporation |
|---|---|
| Address | No. 68 Shanghai Road, Donggang District, Rizhao City, ShandongProvince, 276800, P.R. China |
| Telephone | 086-633-2277268 |
| Fax | 086-633-2277298 |
| Contact Person | Ms. Maggie Zheng |
| Contact Title | Regulatory Affairs Manager |
| zhengxy@hugedent.com |
3. Device Name
Trade name: Flexible Block Common name: Dental polymer blanks (discs or blocks) Classification name: Resin, Denture, Relining, Repairing, Rebasing (21 C.F.R. 872.3760) Regulatory Class: II Product Code: EBI, MQC
4. Substantially Equivalent Device
| Company Name | Device Name | 510 (k) NO. | SubstantiallyEquivalent (SE)Decision Date | Product code |
|---|---|---|---|---|
| PERFLEX | Perflex Biosens | K150454 | 05/26/2015 | EBI, MQC |
This predicate device have not been subject to a design-related recall.
5. Description of Device
Flexible Block is a circular solid (disc) or rectangular solid (block) made from polyamide through high-temperature injection molding, and is used in a CAD/CAM milling machine for production of dental plates, bite plates, frame-works, clasps, personal trays, appliances, occlusal
{4}------------------------------------------------
night guards. These blocks are available in a variety of shapes and sizes for different milling systems and in a variety of dental shades.
6. Intended use
Patient populations: Patients need dental restorations in dental therapy.
Target user group: Use by health care professional or dentist.
Indications for use: Intended for making dental plates, frame-works, clasps, personal trays, appliances, occlusal splints and night guards.
7. Summary of Physical and Chemical Properties Tests
-
Chemical Composition:
The device is made from polyamide. -
Technological characteristics:
The device has the same technological characteristics as the predicate device (Perflex Biosens). And the device is similar in sizes, shapes and shades as the predicate devices. -
Properties:
The device has comparable physical and chemical properties as the predicate device. -
Applications: ●
The device has similar indications for use as the sum of the predicate devices: Intended for making dental plates, bite plates, frame-works, clasps,personal trays, appliances, occlusal splints and night guards by CAD/CAM.
8. Technological Characteristics:
The new device, Flexible Block, has the same design, materials and chemical composition as the predicate device.
Bench tests were performed on the new device including the items listed in the table below. The test results indicated that the new device meets the pass/fail criteria and supports substantial equivalence when compared to the predicate device on physical properties.
| Comparison Items | New Device | Predicate Device |
|---|---|---|
| FlexibleBlock | Perflex BiosensK150454 | |
| 1) RegulatoryClassifications | same | same |
| 2) Indications foruse | similar | similar |
| 3) Contraindications | same | same |
| 4) Composition ofMaterials | Polyamide | super polyamide, thermoplasticcompound |
{5}------------------------------------------------
| Comparison Items | New Device | Predicate Device | ||||||
|---|---|---|---|---|---|---|---|---|
| Flexible Block | Perflex Biosens | |||||||
| K150454 | ||||||||
| 5) PhysicalProperties | Physicalparameters | Tensilestrength:≥30Mpa | Elongation atbreak:≥40% | Waterabsorption:≤5% | Impactstrength:≥20KJ/m2 | Flexuralstrength:≥30Mpa | Flexuralmodulus:≥1000Mpa | |
| Inspectionmethodbasis | ASTMD638 | ASTMD638 | ASTMD570 | ASTM D256 | ASTMD790 | ASTM D790 | ||
| 6) Labeling | similar | similar | ||||||
| 7) TargetPopulation | dental patients | dental patients | ||||||
| 8) Anatomical Site | on teeth | on teeth | ||||||
| 9) Where Used | used in hospital, dental clinic andrelevant places | used in hospital, dental clinic and relevant places | ||||||
| 10) Human Factors | dental professional | dental professional | ||||||
| 11) Design | Circular solid (disc) or rectangularsolid (block) of polyamide. | Circular solid (disc) or rectangular solid (block) ofsuper polyamide and thermoplastic compound. | ||||||
| 12) Cautions | similar | similar | ||||||
| 13) Standards Met | same | same | ||||||
| 14) Biocompatibility | ISO 10993-3Non genotoxocityISO 10993-5Non cytotoxicityISO 10993-10Non sensitization or irritationISO 10993-11Non systemic toxicity | NA | ||||||
| 15) Sterility | Non-sterile | Non-sterile | ||||||
| 16) Chemical Safety | similar | similar |
The new device has highly similar indications for use as predicate device: both devices are to make dental plate, bite plates, personal trays, appliances, occlusal splints and night guards. The difference between the indications don't raise different questions of safety and effectiveness.
9. Summary of Biocompatibility
The new device, Flexible Block, is substantially equivalent to the predicate devices that have been on the market for years without any clinical adverse events. The formulation of new device does not contain any new or non-conventional chemicals compared to the legally marketed predicate device.
We selected our Flexible Block (Model: monolayer; Specification: 98×20mm; Shade: Clear) as the representative in biocompatibility tests and those biocompatibility test reports can be used in the biological evaluation of Flexible Block.
{6}------------------------------------------------
Biocompatibility tests were performed to satisfy the ISO 10993 standards. The test items include Systemic Toxicity; Skin Irritation; Skin Sensitization; In Vitro Cytotoxicity; In Vitro Mammalian Cell TK Gene Mutation Test; In Vitro mammalian chromosome aberration test and Bacterial reverse mutation study.
10. Clinical Performance Data
Not applicable. Clinical performance testing has not been performed for the subject device as it has not been marketed in any area.
11. Summary of Substantial Equivalence
As with the comparison shown in substantial equivalence discussion, these devices are same or similar in almost all aspects. The details of physical properties are slightly different, but these two devices are in compliance with the relevant standards.
It can be seen that the minor differences between the new device and the predicate device are not of significance and do not raise questions of safety and effectiveness as compared to the predicate device. We conclude that Flexible Block is substantially equivalent to the predicate device described herein.
§ 872.3760 Denture relining, repairing, or rebasing resin.
(a)
Identification. A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.(b)
Classification. Class II.