K Number
K213944
Device Name
HealthOST
Manufacturer
Date Cleared
2022-04-22

(126 days)

Product Code
Regulation Number
892.1750
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
HealthOST is an image processing software that provides qualitative and quantitative analysis of the spine from CT images to support clinicians in the evaluation and assessment of musculoskeletal disease of the spine. The HealthOST software provides the following functionality: -Labelling of T1-L4 vertebrae -Measurement of height loss in each vertebra (T1-L4) -Measurement of the mean Hounsfield Units (HU) in volume of interest within vertebra (T11-L4) HealthOST is indicated for use in patients aged 50 and over undergoing CT scan for any clinical indication, that includes at least four vertebrae in the T1-L4 portion of the spine (for vertebral height loss) and T11-L4 (for bone attenuation) portions of the spine. The device is indicated for FBP-reconstructed images only.
Device Description
HealthOST is an image processing software that provides qualitative and quantitative analysis of the spine from CT images to support clinicians in the evaluation and assessment of musculoskeletal disease of the spine. HealthOST does this by analyzing CT scans of patients aged 50 and above being performed for any clinical indication and providing the following outputs for each analyzed vertebra - 1. The vertebral name, e.g. "L3" - 2. The percentage of vertebral height loss, calculated as: Height loss = 1 - (minimum(anterior or middle) vertebral height measurement / posterior vertebral height measurement) from three vertebral height measurements placed at the anterior, middle and posterior aspects of the vertebral body, at the point nearest to the center of the vertebral body. The height loss measurement is provided for every complete vertebra in the range of T1-L4. Height loss above the height loss display threshold will be indicated to the user. The height loss display threshold is configurable per installation from a predefined list, with the default set at >20% height loss. In addition, vertebrae without height loss but with significant height deviation from neighboring vertebrae will be displayed to the user. - 3. The mean hounsfield unit vertebral bone attenuation, calculated by taking the mean HU from a volume of interest of the trabecular bone. This measurement is provided for every complete vertebrae in the range of T11-L4. HU bone attenuation below a set bone attenuation display threshold will be indicated to the user. The bone attenuation display threshold is configurable per installation from a predefined list. The following modules compose the HealthOST software: Data input and validation: Following retrieval of a study, the validation feature assessed the input data (i.e. age, modality, view, etc.) to ensure compatibility for processing by the algorithm. HealthOST algorithm: Once a study has been validated, the algorithm analyzes the CT for analysis and quantification. IMA Integration feature: The study analysis and the results of a successful study analysis is provided to IMA. Error codes feature: In the case of a study failure during data validation or the analysis by the algorithm, an error is provided to the system. Use of axial scans for analysis is only intended as a "back up" in cases that a sagittal scan is not available in the study.
More Information

Not Found

Yes
The document explicitly states that the device utilizes an "artificial intelligence algorithm" and compares its performance to a predicate device that also uses an "artificial intelligence algorithm."

No.
This device is an image processing software that provides quantitative and qualitative analysis of the spine from CT images to assist clinicians in evaluating and assessing musculoskeletal disease. It does not provide any treatment or therapy.

Yes

Explanation: The "Intended Use / Indications for Use" section states that HealthOST is designed "to support clinicians in the evaluation and assessment of musculoskeletal disease of the spine." This aligns with the definition of a diagnostic device, which assists in identifying or characterizing a disease or condition. The software provides qualitative and quantitative analysis (labeling vertebrae, measuring height loss, and Hounsfield Units) that clinicians use to evaluate and assess musculoskeletal disease.

Yes

The device is described as "image processing software" and its functionality is solely based on analyzing and processing CT images. There is no mention of any accompanying hardware component that is part of the device itself.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: HealthOST is an image processing software that analyzes CT images of the spine. It does not analyze biological samples from the patient.
  • Intended Use: The intended use is to support clinicians in the evaluation and assessment of musculoskeletal disease of the spine by providing qualitative and quantitative analysis of CT images. This is a diagnostic aid based on imaging, not on in vitro testing.

Therefore, HealthOST falls under the category of medical imaging software, not In Vitro Diagnostics.

No
The letter does not explicitly state that the FDA has reviewed and approved or cleared a PCCP for this specific device.

Intended Use / Indications for Use

HealthOST is an image processing software that provides qualitative analysis of the spine from CT images to support clinicians in the evaluation and assessment of musculoskeletal disease of the spine. The HealthOST software provides the following functionality:
-Labelling of T1-L4 vertebrae
-Measurement of height loss in each vertebra (T1-L4)
-Measurement of the mean Hounsfield Units (HU) in volume of interest within vertebra (T11-L4)

HealthOST is indicated for use in patients aged 50 and over undergoing CT scan for any clinical indication, that includes at least four vertebrae in the T1-L4 portion of the spine (for vertebral height loss) and T11-L4 (for bone attenuation) portions of the spine.

The device is indicated for FBP-reconstructed images only.

Product codes

JAK

Device Description

HealthOST is an image processing software that provides qualitative and quantitative analysis of the spine from CT images to support clinicians in the evaluation and assessment of musculoskeletal disease of the spine.

HealthOST does this by analyzing CT scans of patients aged 50 and above being performed for any clinical indication and providing the following outputs for each analyzed vertebra

    1. The vertebral name, e.g. "L3"
    1. The percentage of vertebral height loss, calculated as:
      Height loss = 1 - (lmin(anterior or middle)vertebral_height_measurement / lposterior_vertebral_height_measurement)
      from three vertebral height measurements placed at the anterior, middle and posterior aspects of the vertebral body, at the point nearest to the center of the vertebral body.

    The height loss measurement is provided for every complete vertebra in the range of T1-L4. Height loss above the height loss display threshold will be indicated to the user. The height loss display threshold is configurable per installation from a predefined list, with the default set at >20% height loss. In addition, vertebrae without height loss but with significant height deviation from neighboring vertebrae will be displayed to the user.

    1. The mean hounsfield unit vertebral bone attenuation, calculated by taking the mean HU from a volume of interest of the trabecular bone. This measurement is provided for every complete vertebrae in the range of T11-L4. HU bone attenuation below a set bone attenuation display threshold will be indicated to the user. The bone attenuation display threshold is configurable per installation from a predefined list.
      The following modules compose the HealthOST software:

Data input and validation: Following retrieval of a study, the validation feature assessed the input data (i.e. age, modality, view, etc.) to ensure compatibility for processing by the algorithm.

HealthOST algorithm: Once a study has been validated, the algorithm analyzes the CT for analysis and quantification.

IMA Integration feature: The study analysis and the results of a successful study analysis is provided to IMA.

Error codes feature: In the case of a study failure during data validation or the analysis by the algorithm, an error is provided to the system.

Use of axial scans for analysis is only intended as a "back up" in cases that a sagittal scan is not available in the study.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Yes

Input Imaging Modality

CT

Anatomical Site

Spine (T1-L4 vertebrae, T11-L4 for bone attenuation)

Indicated Patient Age Range

50 and over

Intended User / Care Setting

Clinicians

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

The validation data-set included a truthed and enriched sample of 150 anonymized CT scans including a minimum of four vertebrae between T1 to L4 from two healthcare institutions composed of multiple clinical sites in the US and OUS. The sample included sufficient representation from across the disease spectrum for the two key measurement parameters provided by the device, namely vertebral height, and mean trabecular bone attenuation (measured in Hounsfield Units). Ground truth measurements were determined by the three US board-certified radiologists.

Summary of Performance Studies

The HealthOST device performance was evaluated in a stand-alone retrospective study of its performance compared to the established ground truth and respective to the predicate device. The validation data-set included 150 cases, that included 1425 vertebrae.

The HealthOST device demonstrated an overall agreement for vertebral naming of 91.49% (95% CI: [89.91%, 92.89%]) exceeding the stated performance goal. The method comparison analysis demonstrated 95% limits of agreement (LOA) for the HealthOST height loss of [-7.98, 7.36], within the stated performance goal and 91.86% of the differences between HealthOST and the GT lie within the ground truthers LOA's, which is substantially equivalent to the predicate device (K193267). There was 91.28% overall agreement between HealthOST and the ground truth in differentiating height loss above and below 20%, exceeding the stated performance goal. Finally, HealthOST demonstrated 95% limits of agreement (LOA) for the bone attenuation of [-20.83, 18.29] which exceeded the stated performance goal however was superior to the [-21.17, 17.19] LOA demonstrated between the ground truthers. In addition, 96.06% of the differences between HealthOST and the GT lie within the ground truthers LOA's. The reported LOA's by the subject devices are similar to those reported by the predicate device (K193267). All CT data across, slice thickness, slice increment, exposure, KVP and manufacturers were well supported by the HealthOST device.

Specifically, the dataset included 33 axial cases (22.45%), of which 14 has slice thickness between 0-1.5mm, and 19 cases had slice thickness 1.5-3.1 mm. The 16 of the axial scans had a slice increment of 0-1.5mm and 17 scans had a slice increment of 1.5-3.1mm. For axial scans, the HealthOST device demonstrated an overall agreement for vertebral naming of 92.19% (95% CI [89.31% .95.07%]) exceeding the stated performance goal. Regarding the performance of HealthOST on height loss in axial scans, 94.81% of the differences between HealthOST and the GT lie within the ground truthers LOA's, and there was an overall agreement of 91.89% between the HealthOST and ground truth in differentiating height loss above and below 20%. In axial scans, HealthOST demonstrated 95% limits of agreement (LOA) for the bone attenuation of [-15.78, 12.21]. Finally, 96.06% of the differences between HealthOST and the GT lie within the ground truthers LOA's.

A secondary testing dataset from an additional, representative US data source demonstrated device performance across all metrics was generalizable to US populations.

Key Metrics

  • Overall agreement for vertebral naming: 91.49% (95% CI: [89.91%, 92.89%])
  • 95% limits of agreement (LOA) for HealthOST height loss: [-7.98, 7.36]
  • Percentage of differences between HealthOST and GT for height loss within ground truthers LOA’s: 91.86%
  • Overall agreement between HealthOST and ground truth in differentiating height loss above and below 20%: 91.28%
  • 95% limits of agreement (LOA) for bone attenuation: [-20.83, 18.29]
  • Percentage of differences between HealthOST and GT for bone attenuation within ground truthers LOA’s: 96.06%

For axial scans:

  • Overall agreement for vertebral naming: 92.19% (95% CI [89.31% .95.07%])
  • Percentage of differences between HealthOST and GT for height loss within ground truthers LOA’s: 94.81%
  • Overall agreement between HealthOST and ground truth in differentiating height loss above and below 20%: 91.89%
  • 95% limits of agreement (LOA) for bone attenuation: [-15.78, 12.21]
  • Percentage of differences between HealthOST and GT for bone attenuation within ground truthers LOA’s: 96.06%

Predicate Device(s)

K193267

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.

0

April 22, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of a human figure, while the right side features the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

NanoxAI Ltd. % Shlomit Cymbalista Head of Regulatory Affairs Shefayim Commercial Center, PO Box 25 Shefavim. 6099000 ISRAEL

Re: K213944

Trade/Device Name: HealthOST Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-ray System Regulatory Class: Class II Product Code: JAK Dated: March 20, 2022 Received: March 23, 2022

Dear Shlomit Cymbalista:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk, Ph.D Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K213944

Device Name HealthOST

Indications for Use (Describe)

HealthOST is an image processing software that provides qualitative analysis of the spine from CT images to support clinicians in the evaluation and assessment of musculoskeletal disease of the spine. The HealthOST software provides the following functionality:

-Labelling of T1-L4 vertebrae

-Measurement of height loss in each vertebra (T1-L4)

-Measurement of the mean Hounsfield Units (HU) in volume of interest within vertebra (T11-L4)

HealthOST is indicated for use in patients aged 50 and over undergoing CT scan for any clinical indication, that includes at least four vertebrae in the T1-L4 portion of the spine (for vertebral height loss) and T11-L4 (for bone attenuation) portions of the spine.

The device is indicated for FBP-reconstructed images only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Image /page/3/Picture/0 description: The image shows the logo for Nanox AI. The logo consists of a stylized butterfly-like symbol on the left, with the text "NANOXAI" to the right. The butterfly symbol is composed of blue and yellow elements, creating a gradient effect. The text "NANOX" is in blue, while "AI" is in yellow, matching the colors of the butterfly symbol.

510(K) Summary - HealthOST Nanox AI Ltd.

510(k) Number - K213944

Applicant's Name:

Nano-X AI Ltd. Shefayim Commercial Center PO Box 25 Shefayim, 6099000 ISRAEL Telephone: +972-9-8827795 Fax: +972-9-8827795

Date Prepared: April 21, 2022

Device Trade Name: HealthOST

Regulation Number: 21 CFR 892.1750

Regulation Name and Product Code:

JAK - Computed tomography x-ray system

Regulatory Class:

Class II, Radiology

Predicate Device:

The HealthOST device is substantially equivalent to the following Predicate Device:

Proprietary NamePredicate Device:
AI-Rad Companion (Musculoskeletal)
Premarket NotificationK193267
Classification NameComputed tomography x-ray system.
Regulation Number21 CFR 892.1750
Product CodeJAK
Regulatory ClassII

Performance Standards:

4

Image /page/4/Picture/0 description: The image contains the logo for Nanox.AI. The logo consists of a stylized graphic to the left of the company name. The graphic is a symmetrical design with yellow on the top and blue on the bottom. The company name is written in blue, except for the AI which is written in yellow.

No performance standards have been established for such device under Section 514 of the Federal Food, Drug, and Cosmetic Act.

Intended Use/Indication for Use:

HealthOST is an image processing software that provides qualitative and quantitative analysis of the spine from CT images to support clinicians in the evaluation and assessment of musculoskeletal disease of the spine. The HealthOST software provides the following functionality:

-Labelling of T1-L4 vertebrae

-Measurement of height loss in each vertebra (T1-L4) -Measurement of the mean Hounsfield Units (HU) in volume of interest within vertebra (T11-L4)

HealthOST is indicated for use in patients aged 50 and over undergoing CT scan for any clinical indication, that includes at least four vertebrae in the T1-L4 portion of the spine (for vertebral height loss) and T11-L4 (for bone attenuation) portions of the spine.

The device is indicated for FBP-reconstructed images only.

Device Description:

HealthOST is an image processing software that provides qualitative and quantitative analysis of the spine from CT images to support clinicians in the evaluation and assessment of musculoskeletal disease of the spine.

HealthOST does this by analyzing CT scans of patients aged 50 and above being performed for any clinical indication and providing the following outputs for each analyzed vertebra

    1. The vertebral name, e.g. "L3"
    1. The percentage of vertebral height loss, calculated as:

$$\begin{array}{c} \textbf{Heig}{\textit{h}\textit{t}\textit{best}[\forall\emptyset]} \ = \textbf{1} \ -\frac{l{\textit{min}}(\textit{anterior}\textit{ or }\textit{mid}d\textit{d}de){\textit{ver}\textit{et}}\textit{relar}\textit{h}\textit{right}\textit{measurement}}{l{\textit{posterior}\textit{ }\textit{ver}\textit{et}}\textit{relar}\textit{h}\textit{ }\textit{e}\textit{h}\textit{ }\textit{measurement}} \end{array}$$

from three vertebral height measurements placed at the anterior, middle and posterior aspects of the vertebral body, at the point nearest to the center of the vertebral body.

The height loss measurement is provided for every complete vertebra in the range of T1-L4. Height loss above the height loss display threshold will be indicated to the user. The

5

Image /page/5/Picture/0 description: The image shows the logo for Nanox AI. The logo consists of a stylized, abstract symbol on the left and the text "NANOX AI" on the right. The symbol is composed of multiple layers of geometric shapes, with the top layers in yellow and the bottom layers in blue, creating a sense of depth and dimension. The text "NANOX AI" is written in a bold, sans-serif font, with "NANOX" in blue and "AI" in yellow, mirroring the color scheme of the symbol.

height loss display threshold is configurable per installation from a predefined list, with the default set at >20% height loss. In addition, vertebrae without height loss but with significant height deviation from neighboring vertebrae will be displayed to the user.

    1. The mean hounsfield unit vertebral bone attenuation, calculated by taking the mean HU from a volume of interest of the trabecular bone. This measurement is provided for every complete vertebrae in the range of T11-L4. HU bone attenuation below a set bone attenuation display threshold will be indicated to the user. The bone attenuation display threshold is configurable per installation from a predefined list.
      The following modules compose the HealthOST software:

Data input and validation: Following retrieval of a study, the validation feature assessed the input data (i.e. age, modality, view, etc.) to ensure compatibility for processing by the algorithm.

HealthOST algorithm: Once a study has been validated, the algorithm analyzes the CT for analysis and quantification.

IMA Integration feature: The study analysis and the results of a successful study analysis is provided to IMA.

Error codes feature: In the case of a study failure during data validation or the analysis by the algorithm, an error is provided to the system.

Use of axial scans for analysis is only intended as a "back up" in cases that a sagittal scan is not available in the study.

Performance Data:

The HealthOST was designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820.

Safety and performance of HealthOST has been evaluated and verified in accordance with software specifications and applicable performance standards through Software Development and Validation & Verification Process to ensure performance according to specifications, User Requirements and Federal Regulations and Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

The HealthOST device performance was evaluated in a stand-alone retrospective study of its performance compared to the established ground truth and respective to the predicate device. The validation data-set included a truthed and enriched sample of 150 anonymized CT scans including a minimum of four vertebrae between T1 to L4 from two healthcare institutions composed of multiple clinical sites in the US and OUS. The sample included sufficient representation from across the disease spectrum for the two key measurement parameters provided by the device,

6

Image /page/6/Picture/0 description: The image shows the logo for NanoX AI. The logo consists of a stylized graphic to the left of the text "NANOX AI". The graphic is composed of blue and yellow elements arranged in a symmetrical pattern. The text "NANOX" is in blue, while "AI" is in yellow.

namely vertebral height, and mean trabecular bone attenuation (measured in Hounsfield Units). Ground truth measurements were determined by the three US board-certified radiologists.

The objective of this study was to establish the safety, effectiveness and substantial equivalence of the HealthOST software as compared to the predicate device (AI-Rad Companion (Musculoskeletal), K193267)). The HealthOST overall performance was determined by comparing the device output measurements, to the ground truth measurements. The validation data-set included 150 cases, that included 1425 vertebrae.

The HealthOST device demonstrated an overall agreement for vertebral naming of 91.49% (95% CI: [89.91%, 92.89%]) exceeding the stated performance goal. The method comparison analysis demonstrated 95% limits of agreement (LOA) for the HealthOST height loss of [-7.98, 7.36], within the stated performance goal and 91.86% of the differences between HealthOST and the GT lie within the ground truthers LOA's, which is substantially equivalent to the predicate device (K193267). There was 91.28% overall agreement between HealthOST and the ground truth in differentiating height loss above and below 20%, exceeding the stated performance goal. Finally, HealthOST demonstrated 95% limits of agreement (LOA) for the bone attenuation of [-20.83, 18.29] which exceeded the stated performance goal however was superior to the [-21.17, 17.19] LOA demonstrated between the ground truthers. In addition, 96.06% of the differences between HealthOST and the GT lie within the ground truthers LOA's. The reported LOA's by the subject devices are similar to those reported by the predicate device (K193267). All CT data across, slice thickness, slice increment, exposure, KVP and manufacturers were well supported by the HealthOST device.

Specifically, the dataset included 33 axial cases (22.45%), of which 14 has slice thickness between 0-1.5mm, and 19 cases had slice thickness 1.5-3.1 mm. The 16 of the axial scans had a slice increment of 0-1.5mm and 17 scans had a slice increment of 1.5-3.1mm. For axial scans, the HealthOST device demonstrated an overall agreement for vertebral naming of 92.19% (95% CI [89.31% .95.07%]) exceeding the stated performance goal. Regarding the performance of HealthOST on height loss in axial scans, 94.81% of the differences between HealthOST and the GT lie within the ground truthers LOA's, and there was an overall agreement of 91.89% between the HealthOST and ground truth in differentiating height loss above and below 20%. In axial scans, HealthOST demonstrated 95% limits of agreement (LOA) for the bone attenuation of [-15.78, 12.21]. Finally, 96.06% of the differences between HealthOST and the GT lie within the ground truthers LOA's.

A secondary testing dataset from an additional, representative US data source demonstrated device performance across all metrics was generalizable to US populations.

In conclusion, this study demonstrated the HealthOST overall agreement and limits of agreement with respect to the ground truth spinal measurements and establishes its safety and effectiveness, while demonstrating substantial equivalence to the predicate device. It also validated the

7

Image /page/7/Picture/0 description: The image contains the logo for Nanox AI. The logo consists of two parts: a stylized graphic on the left and the text "NANOXAI" on the right. The graphic appears to be a symmetrical design with blue and yellow elements, possibly representing a stylized butterfly or abstract shape. The text "NANOXAI" is written in a sans-serif font, with the "NANOX" part in blue and the "AI" part in yellow.

performance of the HealthOST device across important cohorts, and applicable subsets of imaging acquisition characteristics.

Technological Characteristics Compared to Predicate Device:

We believe that the HealthOST device is substantially equivalent to the AI-Rad Companion (Musculoskeletal) K193267.

Proposed Device:Primary Predicate Device:
HealthOST DeviceAI-Rad Companion
(Musculoskeletal) (K193267)
IntendedUse/
Indications
for
UseHealthOST is an image processing
software that provides qualitative and
quantitative analysis of the spine from
CT images to support clinicians in the
evaluation and assessment of
musculoskeletal disease of the spine.
The HealthOST software provides the
following functionality:
• Labelling of T1-L4 vertebrae
• Measurement of height loss in
each vertebra (T1-L4)
• Measurement of mean
Hounsfield Units in volume of
interest within vertebra (T11-
L4)
HealthOST is indicated for use in
patients aged 50 and over undergoing
CT scan for any clinical indication that
includes at least two vertebrae in the T1-
L4 portion of the spine (for vertebral
height loss) and/or T11-L4 (for bone
attenuation) portions of the spine.
The device is indicated for FBP-
reconstructed images only.AI-Rad Companion (Musculoskeletal)
is an image processing software that
provides quantitative and qualitative
analysis from previously acquired
Computed Tomography DICOM
images to support radiologists and
physicians from emergency medicine,
specialty care, urgent care, and general
practice in the evaluation and
assessment of musculoskeletal disease.
It provides the following functionality:
• Segmentation of vertebras
• Labelling of vertebras
• Measurements of heights in each
vertebra and indication if they are
critically different
• Measurement of mean Hounsfield
value in volume of interest within
vertebra.
Only DICOM images of adult patients
are considered to be valid input
Technological
CharacteristicsProposed Device:
HealthOST DevicePrimary Predicate
Device:
AI-Rad Companion
(Musculoskeletal)
(K193267)Summary
Regulation
Product CodeJAKJAKSame
Regulation Number21 CFR §892.175021 CFR §892.1750Same
General
ModalityCTCTSame
Image formatDICOMDICOMSame
Analysis and Measurement
Detection of VertebraYesYesSame
Labeling of VertebraYesYesSame
Segmentation of
VertebraDeep-learning-based
segmentation of
vertebraDeep-learning-based
segmentation of
vertebrasSame
Measurement of
Vertebral HeightsYes, comparison with
neighboring
measurements
Application of Genant
criteria, indication if
critically differentYes, comparison with
neighboring
measurements.
Application of Genant
criteria, indication if
critically differentSame
Measurement of
Hounsfield (HU) valueHU measurements
based on segmentation
results
Indication to user if
outside reference rangeHU measurements
based on segmentation
resultsSimilar, the subject
device highlights HU
outside of reference
range. This does not
raise new questions of
safety and
effectiveness.
Reporting
Device output1. Vertebrae
label/name
  1. 3 lines representing
    the anterior, middle,
    and posterior points
    measures, together
    with relative
    measurements
  2. % Height loss and
    relative Genant
    category (20)
  3. Bone density
    measured in HU | 1. Vertebrae
    label/name
  4. 3 lines representing
    the anterior, middle,
    and posterior points
    measures, together
    with relative
    measurements
  5. Bone density
    measured in HU | Similar, new
    information does not
    raise new questions of
    safety and
    effectiveness. |

The intended use is equivalent to the predicate device in that both devices perform both qualitative and quantitative analysis of the spine for the evaluation and assessment of musculoskeletal disease using an artificial intelligence algorithm. Both devices perform labeling of vertebrae, measurement of the vertebral height, and measurement of Hounsfield Units within the vertebrae. Both devices highlight vertebrae with calculated height deviations deemed different in accordance with the

8

Image /page/8/Picture/0 description: The image contains the logo for Nanox AI. The logo consists of two stylized, mirrored shapes on the left, resembling wings or petals, with a gradient of yellow to blue. To the right of the shapes is the text "NANOXAI" in a sans-serif font, with "NANOX" in blue and "AI" in a lighter shade of yellow.

Genant criteria. This information is output to the user. Neither device is intended to be used as a standalone diagnostic tool, but provides additional information the physician can use for further clinical management. Both devices are indicated for adult CT scans in DICOM format.

Based on the above, HealthOST has the same intended use and the substantially equivalent indications for use as the predicate device (AI-Rad Companion (Musculoskeletal), K193267).

Comparison of Technological Characteristics

9

Image /page/9/Picture/0 description: The image contains the logo for Nanox AI. The logo consists of two parts: a stylized graphic on the left and the text "NANOXAI" on the right. The graphic is composed of blue and yellow elements arranged in a symmetrical, wing-like shape. The text "NANOXAI" is written in a bold, sans-serif font, with "NANOX" in blue and "AI" in yellow.

The predicate devices and the subject HealthOST device are all standalone software devices that are intended for the same purpose, i.e. to perform quantitative and qualitative analysis of the spine in CTs. Both devices are DICOM-compliant software devices incorporated into the radiology infrastructure of a clinical center. The devices all utilize an artificial intelligence algorithm and run on adult CT scans.

Both the subject device and predicate device perform segmentation and naming of the vertebrae, measurement of vertebral height and bone density as measured in HU within the vertebrae. In addition, both devices indicate if there is a critical difference in height as applied through the Genant criteria. In both devices, the procedure is performed in parallel to and in conjunction with the standard processing of image storage and availability for clinician assessment.

Nano-X AI Ltd. believes that the technological characteristics of the subject device described above raise the same types of safety and/or effectiveness questions as the predicate. The verification and validation tests, in addition to the stand-alone performance assessment of the HealthOST Device demonstrate that the HealthOST Device performance is substantially equivalent to the predicate device.

Conclusion:

10

Image /page/10/Picture/0 description: The image contains the logo for Nanox AI. The logo consists of two parts: a stylized graphic on the left and the text "NANOXAI" on the right. The graphic is composed of multiple layers of shapes, with the top layers in yellow and the bottom layers in blue, creating a layered effect. The text "NANOXAI" is written in a sans-serif font, with "NANOX" in blue and "AI" in yellow, matching the color scheme of the graphic.

Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics, and performance testing, HealthOST device raises no new questions of safety and effectiveness and is substantially equivalent to the predicate device in terms of safety, efficacy, and performance.

The results of the performance comparison study demonstrated that the HealthOST device performs as intended, similarly to the predicate device. The HealthOST device is therefore substantially equivalent to the predicate device.