K Number
K213936

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-07-14

(210 days)

Product Code
Regulation Number
866.3215
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DiaSorin LIAISON® MeMed BV® is an automated in vitro diagnostic semi-quantitative assay that uses chemiluminescent immunoassay (CLIA) technology to measure three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. The LIAISON® MeMed BV® assay is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for seven days or less. The LIAISON® MeMed BV® assay generates a numeric score that falls within discrete interpretation ranges based on the increasing likelihood of bacterial infection. The assay has to be performed on the automated LIAISON® XL Analyzer.

The DiaSorin LIAISON® MeMed BV® Control Set is intended for use as assayed quality control to monitor the performance of the DiaSorin LIAISON® MeMed BV® assay. The performance characteristics of the LIAISON® controls have not been established for any other assays or instrument platforms different from the automated LIAISON® XL Analyzer. The control set is intended for in vitro diagnostic use in a professional laboratory only.

Device Description

The LIAISON MeMed BV assay consists of three individual chemiluminescence immunoassay (CLIA) for quantitative determination of TRAIL, IP-10, and CRP. The LIAISON MeMed BV test result is a score between 0 and 100 derived from computational integration of the measurements of the three proteins TRAIL, IP-10, and CRP, where low scores are indicative of viral infection and high score of bacterial infection. All three reagent packs must be the same lot and present at the same time on the same instrument used for sample testing. All three reagent packs are individually calibrated and quality controlled. Specimens are to be assigned to the MMBV assay protocol where all three reagent packs will be utilized to provide combined results and a final score.

The TRAIL reagent pack uses a monoclonal antibody for capture of TRAIL and a polyclonal antibody for the detection of TRAIL. The assay incubates sample, calibrator or control with assay buffer and paramagnetic particles coated with a monoclonal antibody that specifically recognizes the TRAL. Following the incubation, an isoluminol conjugated polyconal antibody that recognizes TRAIL is then added to the reaction and incubated. The unbound conjugate is removed with a wash step. Starter reagents are then added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of TRAL present in the calibrators, controls or samples. The result of the TRAIL reagent pack is only used to calculate a final LIAISON MeMed BV Score and should not be used individually for diagnosis.

The IP-10 reagent pack uses a monoclonal antibody for the capture of IP-10 and a polyclonal antibody for the detection of IP-10. The assay incubates sample, calibrator or control with assay buffer and paramagnetic particles coated with a monoclonal antibody that specifically recognizes the IP-10. Following the incubation, an isoluminol conjugated polyclonal antibody that recognizes IP-10 is then added to the reaction and incubated. The unbound conjugate is removed with a wash step. Starter reagents are then added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of IP-10 present in the calibrators, controls or samples. The result of the IP-10 reagent pack is only used to calculate a final LIAISON MeMed BV Score and should not be used individually for diagnosis.

The CRP reagent pack uses monoclonal antibodies for capture and detection of CRP. First the patient serum sample is pre-diluted 1:196 with assay buffer. The assay incubates the pre-diluted sample, calibrator or control with assay buffer and paramagnetic particles coated with a monoclonal antibody that specifically recognizes the CRP. Following the incubation, an isoluminol conjugated monoclonal antibody that recognizes CRP is then added to the reaction and incubated. The unbound conjugate is removed with a wash step. Starter reagents are then added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of CRP present in the calibrators, controls or samples. The result of the CRP reagent pack is only used to calculate a final LIAISON MeMed BV Score and should not be used individually for diagnosis.

AI/ML Overview

The provided document describes the FDA clearance (K213936) for the DiaSorin LIAISON® MeMed BV® assay, which aids in differentiating bacterial from viral infections. Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly listing predefined acceptance criteria with numerical targets. However, the performance data presented implies a set of internal acceptance criteria related to statistical significance, agreement with expert adjudication, and reproducibility.

Given the information provided, here's a table summarizing the reported device performance, with implicit acceptance criteria derived from the study's conclusions:

Performance MetricImplicit Acceptance Criteria (based on stated conclusions)Reported Device Performance (LIAISON MeMed BV)
Clinical Agreement (Primary Endpoint)Significant trend between SCORE and likelihood of bacterial infection; high percentage of patients in outer bins.Significant trend demonstrated between LIAISON MeMed BV SCORE and increasing likelihood of bacterial infections across SCORE bins. High percentage of patients found in outer bins (Bin 1 and 5).
Likelihood Ratio for Bacterial Basis (Bin 5: High Bacterial)High Likelihood Ratio indicating strong association with bacterial infection.13.00 (7.09-25.83)
Likelihood Ratio for Bacterial Basis (Bin 1: Viral)Low Likelihood Ratio indicating strong association with viral/non-infection.0.043 (0.002-0.180)
Clinical Agreement (Secondary Endpoint)Significant trend between SCORE and likelihood of bacterial infection; high percentage of patients in outer bins.Significant trend demonstrated between LIAISON MeMed BV SCORE and increasing likelihood of bacterial infections across SCORE bins. High percentage of patients found in outer bins (Bin 1 and 5).
Likelihood Ratio for Bacterial Basis (Bin 5: High Bacterial)High Likelihood Ratio.52.97 (19.90-214.87)
Likelihood Ratio for Bacterial Basis (Bin 1: Viral)Low Likelihood Ratio.0.051 (0.003-0.214)
Method Correlation (Primary Endpoint)High overall agreement and high agreement in outer bins with predicate device.Overall Agreement: 79.3% (95% CI: 74.2% - 83.6%). Bin 1: 91.8%; Bin 5: 96.7%.
Method Correlation (Secondary Endpoint)High overall agreement and high agreement in outer bins with predicate device.Overall Agreement: 79.0% (95% CI: 73.6% - 83.5%). Bin 1: 92.2%; Bin 5: 100%.
Reproducibility (SCORE)Acceptable coefficient of variation (CV) across laboratories and within-laboratory.Reproducibility CV for Score: KC1 (2.24): 0.688 (N/A); KC2 (98.3): 0.491 (N/A); MMBV-PREC3 (55.0): 2.765 (N/A); MMBV-PREC4 (7.84): 1.072 (N/A).
Matrix Equivalence (Fresh vs. Frozen Serum)Strong correlation (slope ~1, intercept ~0, high R-squared) between fresh and frozen samples.SCORE: Slope 1.00, Intercept 0.00, R-squared 0.9843, R 0.992.
Limit of Blank (LoB)Quantifiable low limit.CRP: 0.024 mg/L; IP-10: 0.578 pg/mL; TRAIL: 5.33 pg/mL.
Limit of Detection (LoD)Quantifiable low limit.CRP: 0.067 mg/L; IP-10: 4.31 pg/mL; TRAIL: 7.03 pg/mL.
Limit of Quantitation (LoQ)Quantifiable low limit.CRP: 1.0 mg/L; IP-10: 100 pg/mL; TRAIL: 15.0 pg/mL.
Cross-ReactivityNo significant cross-reactivity with specified substances.Testing performed; implies no significant cross-reactivity observed (conclusion not explicitly stated but implied by study inclusion).
Interfering SubstancesNo interference observed for specified substances."No interference was observed for substances."

2. Sample Sizes Used for the Test Set and Data Provenance

  • Clinical Agreement Test Set: 285 serum samples.
    • Provenance: Collected at hospital admission, Emergency Department, and Urgent Care Centers. Patients ranged in age from 5 months to 92 years. The origin country is not explicitly stated, but the mention of "21 international experts" suggests a multi-national or at least internationally adjudicated dataset. The study is retrospective in the sense that samples were likely collected before the adjudication process and testing with the device.
  • Method Correlation Test Set:
    • Primary Endpoint analysis: 285 clinical samples (the same as the clinical agreement test set).
    • Secondary Endpoint analysis: 257 clinical samples (28 of the 285 were excluded).
  • Expected Values Test Set: 150 serum samples from apparently healthy asymptomatic adults.
    • Provenance: Collected in the Southwestern U.S.
  • Reproducibility Study Test Set: 4 serum samples (1 normal, 1 viral, 1 bacterial, 1 equivocal) plus kit controls.
  • Cross-Reactivity Study Test Set: Two (2) serum samples (one low, one high SCORE) for each cross-reactant being tested.
  • Interfering Substances Study Test Set: Two (2) serum samples (one low, one high SCORE) for each interfering substance being tested.
  • Matrix Equivalence Study Test Set: 43 fresh serum samples from individual patients, with 20 spiked for range coverage.
  • Limit of Blank, Detection, Quantitation Test Set: Varied based on particular limit, ranging from 4-8 serum/spiked matrix samples for LoD/LoQ studies and 5 calibrator matrix samples for LoB.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: A pool of 21 international experts.
  • Qualifications of Experts: Clinicians with at least 7 years of relevant clinical experience. Each panel comprised at least three experts.

4. Adjudication Method for the Test Set

  • Method: Expert adjudication.
    • Process: Panelists for each subject adjudication were drawn from the pool of 21 international experts. Each panel comprised at least three experts who independently adjudicated the etiologic label for each patient. The etiologic label was determined as bacterial, viral, or non-infectious. The adjudication was based on anonymized patient data. Critically, the adjudicators were blinded to the MeMed BV result (for the primary endpoint) and to CRT and PCT results (for the primary endpoint). For the secondary endpoint, adjudicators were un-blinded to PCT and CRP results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This document describes the analytical and clinical performance of an in vitro diagnostic assay (a laboratory test that measures protein levels and computes a score), not an AI-assisted diagnostic tool that human readers interpret. Therefore, an MRMC comparative effectiveness study involving human readers assisting with AI is not applicable and was not performed. The device itself is the "AI" (computational algorithm) that processes biomarker data.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, the performance presented for the "LIAISON MeMed BV" assay is a standalone performance of the device (assay and its integrated computational scoring). The "SCORE" is generated by the device's algorithm based on the measured protein levels (TRAIL, IP-10, and CRP), without a human in the loop for the scoring itself. The clinical utility is then evaluated by comparing this score to expert adjudication.

7. The Type of Ground Truth Used

  • Type of Ground Truth: Expert consensus (physician expert adjudication). Specifically, physicians were forced to make a bacterial, viral, or non-infectious diagnosis.

8. The Sample Size for the Training Set

  • The document does not specify the sample size used for training the algorithm (the "SCORE" computation). This document focuses on the validation of the device for FDA clearance. Typically, details of the training dataset are considered proprietary or part of the algorithm development process prior to clinical validation.

9. How the Ground Truth for the Training Set was Established

  • Since the training set size and details are not provided, information on how its ground truth was established is also not available in this document. It is typical for such algorithms to be trained on large, well-characterized datasets, often with similar expert-adjudicated ground truth, but these details are not part of the premarket notification summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

DiaSorin Inc. Emily Peterson Principal Regulatory Affairs Specialist 1951 Northwestern Avenue Stillwater, Minnesota 55082

July 14, 2022

Re: K213936

Trade/Device Name: LIAISON MeMed BV, LIAISON MeMed BV Control Set Regulation Number: 21 CFR 866.3215 Regulation Name: Device To Detect And Measure Non-Microbial Analyte(s) In Human Clinical Specimens To Aid In Assessment Of Patients With Suspected Sepsis Regulatory Class: Class II Product Code: QPS Dated: December 15, 2021 Received: December 16, 2021

Dear Emily Peterson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Noel J. Gerald, Ph.D. Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213936

Device Name LIAISON MeMed BV LIAISON MeMed BV Control Set

Indications for Use (Describe)

The DiaSorin LIAISON® MeMed BV® is an automated in vitro diagnostic semi-quantitative assay that uses chemiluminescent immunoassay (CLIA) technology to measure three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. The LIAISON® MeMed BV® assay is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for seven days or less. The LIAISON® MeMed BV® assay generates a numeric score that falls within discrete interpretation ranges based on the increasing likelihood of bacterial infection. The assay has to be performed on the automated LIAISON® XL Analyzer.

The DiaSorin LIAISON® MeMed BV® Control Set is intended for use as assayed quality control to monitor the performance of the DiaSorin LIAISON® MeMed BV® assay. The performance characteristics of the LIAISON® controls have not been established for any other assays or instrument platforms different from the automated LIAISON® XL Analyzer. The control set is intended for in vitro diagnostic use in a professional laboratory only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

SUBMITTED BY:Emily PetersonDiaSorin Inc.1951 Northwestern AvenueStillwater, MN 55082-0285Phone (651) 351-5628Fax (651) 351-5669Email emily.peterson@diasorin.com
NAME OF DEVICE:
Trade Name:LIAISON® MeMed BV®LIAISON® MeMed BV® Control Set
Common Names/Descriptions:LIAISON® MeMed BV®LIAISON® MeMed BV® Control Set
Classification Names:Immunoasssay for Host Biomarkers oInfection
Product Code:QPSQCH
Predicate Device:MeMed BV®

INTENDED USE:

The DiaSorin LIAISON MeMed BV is an automated in vitro diagnostic semi-quantitative assay that uses chemiluminescent immunoasay (CLIA) technology to measure three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. The LIAISON MeMed BV assay is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for seven days or less. The LIAISON MeMed BV assay generates a numeric score that falls within discrete interpretation ranges based on the increasing likelihood of bacterial infection. The assay has to be performed on the LIAISON XL Analyzer.

The DiaSorin LIAISON" MeMed BV Control Set is intended for use as assayed quality control to monitor the performance of the DiaSorin LIAISON MeMed BV assay. The performance characteristics of the LIAISON controls have not been established for any other assays or instrument platforms different from the automated LIAISON *XL Analyzer. The control set is intended for in vitro diagnostic use in a professional laboratory only.

{4}------------------------------------------------

KIT DESCRIPTION:

The LIAISON MeMed BV assay consists of three individual chemiluminescence immunoassay (CLIA) for quantitative determination of TRAIL, IP-10, and CRP. The LIAISON MeMed BV test result is a score between 0 and 100 derived from computational integration of the measurements of the three proteins TRAIL, IP-10, and CRP, where low scores are indicative of viral infection and high score of bacterial infection. All three reagent packs must be the same lot and present at the same time on the same instrument used for sample testing. All three reagent packs are individually calibrated and quality controlled. Specimens are to be assigned to the MMBV assay protocol where all three reagent packs will be utilized to provide combined results and a final score.

The TRAIL reagent pack uses a monoclonal antibody for capture of TRAIL and a polyclonal antibody for the detection of TRAIL. The assay incubates sample, calibrator or control with assay buffer and paramagnetic particles coated with a monoclonal antibody that specifically recognizes the TRAL. Following the incubation, an isoluminol conjugated polyconal antibody that recognizes TRAIL is then added to the reaction and incubated. The unbound conjugate is removed with a wash step. Starter reagents are then added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of TRAL present in the calibrators, controls or samples. The result of the TRAIL reagent pack is only used to calculate a final LIAISON MeMed BV Score and should not be used individually for diagnosis.

The IP-10 reagent pack uses a monoclonal antibody for the capture of IP-10 and a polyclonal antibody for the detection of IP-10. The assay incubates sample, calibrator or control with assay buffer and paramagnetic particles coated with a monoclonal antibody that specifically recognizes the IP-10. Following the incubation, an isoluminol conjugated polyclonal antibody that recognizes IP-10 is then added to the reaction and incubated. The unbound conjugate is removed with a wash step. Starter reagents are then added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of IP-10 present in the calibrators, controls or samples. The result of the IP-10 reagent pack is only used to calculate a final LIAISON *MeMed BV *Score and should not be used individually for diagnosis.

The CRP reagent pack uses monoclonal antibodies for capture and detection of CRP. First the patient serum sample is pre-diluted 1:196 with assay buffer. The assay incubates the pre-diluted sample, calibrator or control with assay buffer and paramagnetic particles coated with a monoclonal antibody that specifically recognizes the CRP. Following the incubation, an isoluminol conjugated monoclonal antibody that recognizes CRP is then added to the reaction and incubated. The unbound conjugate is removed with a wash step. Starter reagents are then added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of CRP present in the calibrators, controls or samples. The result of the CRP reagent pack is only used to calculate a final LIAISON "MeMed BV" Score and should not be used individually for diagnosis.

SUBSTANTIAL EQUIVALENCE:

The DiaSorin LIAISON MeMed BV is substantially equivalent in principle and performance to the MeMed Diagnostics Ltd. MeMed BV assay which was FDA cleared on September 01, 2021-510k K210254.

{5}------------------------------------------------

The following comparison of the use, technology, and performance presented in Table 1 supports the Statement of Equivalence between the LIAISON® MeMed BV assay to the MeMed Diagnostics Ltd. MeMed BV test system. The differences (presented in Table 2) in technology do not raise additional concerns regarding safety and effectiveness. Safety and effectiveness are demonstrated to be substantially equivalent.

Table of Similarities
FeatureCandidate DeviceLIAISON® MeMed BV®Predicate DeviceMeMed BV® System - K210254
Intended UseThe DiaSorin LIAISON® MeMed BV® is anautomated in vitro diagnostic semi-quantitativeassay that uses chemiluminescent immunoassay(CLIA) technology to measure three non-microbial (host) proteins (TRAIL, IP-10, and CRP)in adult and pediatric serum samples and isintended for use in conjunction with clinicalassessments and other laboratory findings as anaid to differentiate bacterial from viralinfection. The LIAISON® MeMed BV® assay isindicated for use in patients presenting to theemergency department or urgent care centerand with samples collected at hospitaladmission from patients with suspected acutebacterial or viral infection, who have hadsymptoms for seven days or less. The LIAISON®MeMed BV® assay generates a numeric scorethat falls within discrete interpretation rangesbased on the increasing likelihood of bacterialinfection.The MeMed BV® test is an automated semi-quantitative immunoassay that measures threenon-microbial (host) proteins (TRAIL, IP-10, andCRP) in adult and pediatric serum samples andis intended for use in conjunction with clinicalassessments and other laboratory findings as anaid to differentiate bacterial from viralinfection. The MeMed BV® is indicated for use inpatients presenting to the emergencydepartment or urgent care center and withsamples collected at hospital admission frompatients with suspected acute bacterial or viralinfection, who have had symptoms for less thanseven days. The MeMed BV® test generates anumeric score that falls within discreteinterpretation bins based on the increasinglikelihood of bacterial infection
ResultsSemi-QuantitativeSame
Measurandthree non-microbial (host) proteins (TRAIL, IP-10, and CRP)Same
ConjugateantibodyspecificitiesAnti-TRAIL - polyclonal antibodyIP-10 – polyclonal antibodyCRP - monoclonal antibodySame
Sample typeHuman serumSame
AntigenRecombinant TRAIL, IP-10 and CRPSame
MeasurementSystemPhotomultiplier(flash chemiluminescence reader)Photomultiplier(flash chemiluminescence reader)
OutputSCORE resultSame
Table of Differences
FeatureCandidate Device LIAISON® MeMedBV®Predicate DeviceMeMed BV® System – K210254
Conjugate antibodyspecificitiesConjugated to anIsoluminol derivativeConjugated toAlkaline phosphatase
Sample size200 μL100 μL
Assay ProcedureAutomated –LIAISON® XL AnalyzerAutomated –MeMed Key Analyzer
CalibrationTwo levels (low and high) ofcalibrators for each Analyte3 calibrators containing high, medium andlow concentration of each analyte
Calibration Interval3 weeks2 weeks

Table 1:

{6}------------------------------------------------

Table 2·

PERFORMANCE DATA:

CLINICAL AGREEMENT

A total of 285 serum samples from patients presenting with signs and symptoms suggestive of acute bacterial or viral infection were included in the study. Study populations included samples taken at hospital admission, Emergency Department and Urgent Care Centers. Patients ranged in age from 5 months to 92 years of age with 54.7% females (156) and 45.3% males (129).

The most common clinical syndrome was respiratory tract infections (67%) which included both upper and lower respiratory tract infections.

A primary and secondary endpoint analysis were performed to establish the diagnostic performance of the LIAISON MeMed BV test for differentiating bacterial from viral infection in patients with suspected acute bacterial or viral infection.

Primary Endpoint Analysis: The results of the LIAISON "MeMed BV" assay were compared to physician expert adjudication. Expert adjudication was used as the comparator method. Expert adjudication was determined through the process in which physicians were forced to make a bacterial, viral, or noninfectious diagnosis with categorization of patients. Panelists for each subject adjudication were drawn from a pool of 21 international experts, who were clinicians with at least 7 years of relevant clinical experience. Each panel comprised at least three experts who independently adjudicated the etiologic label for each patient. The etiologic label of the adjudicator was based on anonymized patient data. Importantly, the adjudicators were blinded to the MeMed BV result. The adjudicators were blinded to CRT and PCT results for the primary endpoint.

The performance of the LIAISON MeMed BV assay in differentiating between bacterial and viral infection /non-infection was assessed using two (2) statistical tests.

  • A Cochran-Armitage test (Agresti 2010) of the slope of a weighted linear regression of the proportion of bacterial samples in each bin and SCORE range, respectively.

  • The likelihood ratio of the bin number/SCORE range is the ratio of the proportion of all bacterial patients who fall in that bin to the proportion of nonbacterial patients who fall in that bin.

{7}------------------------------------------------

The analysis comprised 285 Apollo specimens characterized by expert adjudicated diagnosis, by either having a bacterial infection, or not having a bacterial infection, by having either no infection or a viral one. Each patient was also scored using the LIAISON MeMed BV assay and the predicate assay MeMed BV . The subjects were distributed in five (5) bins corresponding to the degree of suspicion for bacterial infection. The predicate device SCORE value for each was also analyzed.

A significant trend is demonstrated between the LIAISON MeMed BV SCORE and the increasing likelihood of bacterial infections across the SCORE bins. In addition, a high percentage of patients are found in the outer bins (bin 1 and 5) representing a very high likelihood of viral or bacterial infection, respectively.

Expert Adjudication Label
LIAISON® MeMed BV®SCORE BINNReferenceBacterial(N)ReferenceViral orNon-Infectious(N)%Patients each Bin%ReferenceBacterial Patients%ReferenceViral orNon-InfectiousPatientsLikelihood Ratio ofBacterial Basis Each Bin(95% CI)
BIN 590 ≤ score ≤ 100(Bacterial)33221111.666.733.313.00(7.09-25.83)
BIN 465 < score < 90(Mod Bacterial)268189.130.869.22.89(1.27-5.95)
BIN 335 ≤ score ≤ 65(Equivocal)256198.824.076.02.05(0.79-4.51)
BIN 210 < score < 35(Mod Viral)4814716.82.197.90.14(0.01-0.60)
BIN 10 ≤ score ≤ 10(Viral)153115253.70.7099.30.043(0.002-0.180)
TOTAL28538247100

Primary Endpoint Analysis

LIAISON * MeMed BV results were also analyzed according to a secondary endpoint format in which the adjudicators were un-blinded to PCT and CRP results. For the secondary objective cohort, each patient that did not receive a final adjudication outcome was assigned an indeterminate comparator method outcome.

Twenty eight (28) of the 285 samples were excluded. The remaining 257 were characterized by either having a bacterial or a viral infection.

The performance of the secondary endpoint for the LIAISON MeMed BV assay in differentiating between bacterial and viral infection /non-infection was assessed using the same two (2) statistical tests.

A significant trend is demonstrated between the LIAISON MeMed BV SCORE and the increasing likelihood of bacterial infections across the SCORE bins. In addition, a high percentage of patients are found in the outer bins (bin 1 and 5) representing a very high likelihood of viral or bacterial infection, respectively

{8}------------------------------------------------

Expert Adjudication Label
Score BinLIAISON ° MeMed BVNNReferenceBacterialNReferenceViral / Non-Infectious%PatientsEachBin%ReferenceBacterialPatients% ReferenceViral / Non-InfectiousPatientsBacterial Likelihood Ratio(95% CI)
BIN 590 ≤ score ≤ 100(High Bacterial)242139.387.512.552.97(19.90-214.87)
BIN 465 < score <90(Mod Bacterial)205157.825.075.02.52(0.87-5.99)
BIN 335 ≤ score ≤ 65(Eqv)233209.013.087.01.14(0.28-3.07)
BIN 210 < score <35(Mod Viral)4204216.30.0100.00.000(0.000-0.341)
BIN 10 ≤ score ≤10(High Viral)148114757.60.7099.30.051(0.003-0.214)
TOTAL25730227100

Secondary Endpoint Analysis

METHOD CORRELATION:

The LIAISON MeMed BV and another commercially available method were directly compared to each other using the clinical samples linked to the primary and secondary endpoint formats. Overall Agreement with 95% Confidence Intervals (Wilson Approach) and individual Bin Agreements were calculated for the Primary and Secondary Endpoints and are provided below.

Primary Endpoint Analysis

LIAISON® MeMedBV® Bin ResultsPredicate Bin Results (Primary Endpoint)
HighVIRBin 1ModVIRBin 2EquivocalBin 3ModBACTBin 4HighBACTBin 5
HighVIRBin 11467000
ModVIRBin 21328700
EquivocalBin 3012850
ModBACTBin 4019151
HighBACTBin 5000429
TOTAL15948242430
% Bin Agreement146/159 (91.8%)28/48 (58.3%)8/24 (33.3%)15/24 (62.5%)29/30 (96.7%)
% Overall Agreement226/285 (79.3%)95% CI (Wilson Approach) 74.2% - 83.6%

{9}------------------------------------------------

LIAISON® MeMedBV® Bin ResultsHighVIRBin 1ModVIRBin 2EquivocalBin 3ModBACTBin 4HighBACTBin 5
HighVIRBin 11417000
ModVIRBin 21224600
EquivocalBin 3012650
ModBACTBin 4017120
HighBACTBin 5000321
TOTAL15344191821
% Bin Agreement141/153 (92.2%)24/44 (54.5%)6/19 (31.6%)12/20 (60.0%)21/21 (100%)
% Overall Agreement203/257 (79.0%)
95% CI (Wilson Approach) 73.6% - 83.5%

Secondary Endpoint Analysis

EXPECTED VALUES:

To assess the expected Reference range a study was performed with 150 serum samples collected from apparently healthy asymptomatic adults who met the following inclusion criteria of no flu-like symptoms i.e. fever, chills, headache, fatigue, muscle and joint aches or swollen lymph nodes. The subjects were 65% female (98) and 35% male (52) and ranged in age from 21 – 71. Samples were collected in the Southwestern U.S.

Based on the 95% Central interval, the following values were established following CLSI guideline C28-A3.

NMean SCOREMedian SCOREObserved 95% SCORE range
1506.94.01 to 33

REPRODUCIBILITY STUDY

A 5-day reproducibility study was conducted at DiaSorin Inc. and 2 external laboratories. This study included one (1) lot of LIAISON MeMed BV assay which includes 3 Reagent Integral kits for TRAIL, IP-10 and CRP and two (2) LIAISON MeMed BV Control Sets one (1) lot for daily quality control measurement and one (1) lot as precision samples on the LIAISON XL Analyzer at 3 sites. The study was performed for 5 days, 1 run/day and 6 reps/run for a total of 30 replicates per site. The coded panel consisted of the kit controls and 4 serum samples representing 1 normal, 1 viral, 1 bacterial, and 1 equivocal sample.

{10}------------------------------------------------

DiaSorin LIAISON® MeMed BV® Premarket Notification

MeanRepeatabilityBetween DaysWithin LaboratoryBetween SitesReproducibility
Sample IDSD%CVSD%CVSD%CVSD%CVSD%CV
TRAIL(pg/mL)
KC 11342.5471.9%8.0866.0%8.4786.3%1.9611.5%8.7026.5%
KC 227.80.5812.1%1.6546.0%1.7536.3%0.0000.0%1.6205.8%
MMBV-PREC 11894.9962.6%3.1391.7%5.9003.1%1.2460.7%6.0303.2%
MMBV-PREC-227.90.5962.1%0.6592.4%0.8893.2%0.6912.5%1.1254.0%
MMBV-PREC-360.41.0801.8%0.9061.5%1.4102.3%0.0000.0%1.3552.2%
MMBV-PREC-470.11.5832.3%0.8331.2%1.7892.6%1.0761.5%2.0883.0%
IP-10(pg/mL)
KC 1136826.9812.0%57.9164.2%63.8934.7%191.79714.0%202.1614.8%
KC 21864.4692.4%8.9444.8%9.9995.4%26.11814.0%27.96615.0%
MMBV-PREC 181524.2953.0%11.5951.4%26.9203.3%81.23410.0%85.57810.5%
MMBV-PREC-2134833.5892.5%31.7362.4%46.2103.4%151.48411.2%158.3811.7%
MMBV-PREC-356713.5732.4%11.6302.1%17.8743.2%52.5259.3%55.4839.8%
9698*255.2362.6%117.7251.2%281.0782.9%429.2094.4%513.065.3%
CRP(mg/L)
KC 113.70.3982.9%0.5974.4%0.7185.2%0.6254.6%0.9527.0%
KC 21294.3123.3%4.8433.8%6.4845.0%3.8843.0%7.5585.9%
MMBV-PREC 114.70.2831.9%0.3822.6%0.4763.2%0.7925.4%0.9246.3%
MMBV-PREC-22014.2002.1%5.0642.5%6.5793.3%6.2293.1%9.0614.5%
MMBV-PREC-350.02.4164.8%1.6223.2%2.9105.8%2.6905.4%3.9637.9%
MMBV-PREC-42.600.0953.6%0.1284.9%0.1596.1%0.2429.3%0.29011.1%
SCORE
KC 12.240.231N/A0.666N/A0.705N/A0.000N/A0.688N/A
KC 298.30.245N/A0.279N/A0.371N/A0.321N/A0.491N/A
MMBV-PREC 11.000.000N/A0.000N/A0.000N/A0.000N/A0.000N/A
MMBV-PREC-299.00.000N/A0.000N/A0.000N/A0.000N/A0.000N/A
MMBV-PREC-355.01.914N/A1.307N/A2.317N/A1.508N/A2.765N/A
MMBV-PREC-47.840.521N/A0.358N/A0.632N/A0.865N/A1.072N/A

*IP-10 MMBV-PREC4 Dose result below assay measuring range. Results are presented for RLU's.

{11}------------------------------------------------

CROSS-REACTIVITY STUDY:

Testing was performed to determine potential cross-reactivity in the LIAISON MeMed BV assay. The CLSI Guidance "Interfering Substances" (EP07-Third Edition) was consulted in the preparation of the testing protocol. Cross-reactivity was evaluated using two (2) serum samples that included one (1) low, approximate SCORE 5 and one (1) high, approximate SCORE 95 on one (1) lot.

TRAIL Cross-reactants concentration evaluated at 50 ng/mL
4-1BB LigandLT α2/β1
CD40 LigandTNF-α
LT α1/β2TNF-β
Adiponectin
IP-10 Cross-reactants concentration evaluated at 50 ng/mL
BLC/BCA-1IL-8
ENA-78I-TAC
GCP-2NAP-2
GROαMIG
GROγSDF-1α
IFNγSDF-1β
CRP Cross-reactants concentration evaluated at 500 ng/mL
Pentraxin 2/SAPPentraxin 3/TSG-14

INTERFERING SUBSTANCES:

Testing was performed to determine whether the presence of commonly encountered endogenous and exogenous substances interfere with the LIAISON MeMed BV assay results. Potentially interfering substances were evaluated on one (1) lot of the LIAISON MeMed BV assay using two (2) serum samples that included one (1) low, approximate SCORE 5 and one (1) high, approximate SCORE 95 The CLSI Guidelines EP07-A3, Interference Testing in Clinical Chemistry, 3rd Edition was consulted in the study design the MeMed BV SCORE result was evaluated in the presence of interfering substances. The impact of selected interferents on the SCORE result was studied using serum samples that represented high (bacterial) and low (viral) SCORE. No interference was observed for substances.

SubstanceConcentration TestedSubstanceConcentration Tested
Bilirubin (conjugated)0.4 mg/mLAzithromycin11.1 µg/L
Bilirubin (unconjugated)0.4 mg/mLCaffeine108 μg/mL
Hemoglobin10 mg/mLCetirizine HCl4.35 µg/mL
Triglycerides15 mg/mLDextramethorphan15.6 ng/mL
Cholesterol4 mg/mLDoxycycline18 µg/mL
HAMA600 ng/mLEthanol0.5% v/v
Rheumatoid Factor500 IU/mLHeparin3300 U/L
Human Serum Albumin60 mg/mLIbuprofen219 µg/mL
Biotin3600 ng/mLLevofloxacin36 µg/mL
Acetaminophen0.156 mg/mLLoratidine87 ng/mL
Acetyl Salicylic Acid0.03 mg/mLOxymetazoline HCl0.0006 ug/mL
Amoxicillin54 µg/mLPhenylephrine30 ng/mL
Ampicillin75 µg/mLPrednisolone1200 ng/L

{12}------------------------------------------------

MATRIX EQUIVALENCE STUDY:

Fresh vs Frozen Serum

Forty-three (43) fresh serum samples were collected from forty-three (43) individual patients. Twenty (20) samples were spiked with recombinant TRAIL, IP-10, and/or CRP in order to meet the requirement of samples across the range of the score, 0-100. The forty-three (43) sample values included twentyfour (24) viral score of 0-34, seven (7) equivocal score of 35-65, and twelve (12) bacterial with score of 66-100. Samples were tested fresh with validation lot 1 reagents in replicates of three (3). The samples were then frozen, thawed and tested with validation lot 1 reagents in replicates of three (3).

Mean Values for each sample were calculated for fresh and frozen samples and a Passing-Bablok regression line fitted to the data for each assay and SCORE Value. The regression shows Fresh and Frozen samples may be used interchangeably in the assay.

AnalysisSlopeInterceptR-squaredR
Fresh v.FrozenSCORE1.000.000.98430.992
CRP1.020.010.99680.998
IP-101.04-5.180.9990.999
TRAIL0.934.630.94870.974

LIMIT of BLANK, LIMIT of DETECTION AND LIMIT of QUANTITATION

The Limit of Blank, Limit of Detection and Limit of Quantitation were determined according to CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline June 2012- Second Edition.

METHOD:

Limit of Blank (LoB):

Five (5) calibrator matrix samples without analyte were tested over three (3) days with two (2) runs per day and two (2) replicates per run for a total of sixty (60) results per lot. The equation used for determination of the Limit of Blank is as follows:

LoB = mean (Blank Dose) + 1.653 * SD (Blank Dose)

The higher LoB result was used in the IFU.

Limit of Detection (LoD):

Four (4) serum and/or spiked matrix samples were tested over three (3) days with two (2) runs per day and two (2) replicates per run for a total of twelve (12) results per sample. Samples were in the approximate ranges: CRP (0.09 to 4.1 mg/L), IP-10 (0.37 to 151 pg/mL) TRAIL (3.8 to 29.8 pg/mL). The equation used for determination of the Limit of Detection is as follows:

LoD = LoB + 1.654 * pooled SD (sample dose)

The higher LoD result was used in the IFU.

Limit of Quantitation (LoQ):

Seven (7) or eight (8) serum and/or spiked matrix samples (4 of which were used in the LoD determination) were tested over three (3) days with two (2) runs per day and two (2) replicates per run for a total of twelve (12) results per sample. Samples were in the approximate ranges: CRP (0.09 to 4.1 mg/L), IP-10 (0.37 to 151 pg/mL) TRAIL (3.8 to 29.8 pg/mL). The % CV for each individual sample was

{13}------------------------------------------------

plotted against the mean dose result and the best curve fit applied. LoQ is calculated as the lowest concentration at which the regression line crosses 20% CV.

The following limits were determined for the LIAISON® MeMed BV assay.

Limit of Blank(LoB)Limit of Detection(LoD)Limit ofQuantitation(LoQ)
CRP (mg/L)0.0240.0671.0
IP-10 (pg/mL)0.5784.31100
TRAIL (pg/mL)5.337.0315.0

CONCLUSION:

The material submitted in this premarket notification supports a substantial equivalence decision. The labeling satisfies the requirements of 21CFR 809.10.

§ 866.3215 Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis.

(a)
Identification. A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended.
(2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection.
(3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability.
(4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(i) Results must demonstrate adequate device performance relative to a well-accepted comparator.
(ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population.
(iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses.
(5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
e.g., age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples (
e.g., collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.