(228 days)
The DiaSorin LIAISON® Calprotectin assay is an in vitro diagnostic chemiluminescent immunoassay (CLIA) intended for the quantitative measurement, in human stool, of fecal calprotectin, a neutrophilic protein that is a marker of mucosal inflammation. The LIAISON® Calprotectin assay can be used as an aid in the diagnosis of inflammatory bowel diseases (IBD), specifically Crohn's disease and ulcerative colitis, and as an aid in differentiation of IBD from irritable bowel syndrome (IBS). Test results are to be used in conjunction with information obtained from the patients' clinical evaluation and other diagnostic procedures.
The test has to be performed on the LIAISON® Analyzer Family.
The DiaSorin LIAISON® Q.S.E.T. Device Plus (Quantitative Stool Extraction and Test) is intended for use in the preparation of human stool specimens for testing in the LIAISON® Calprotectin assay.
The LIAISON® Calprotectin assay is a sandwich assay that uses 2 monoclonal antibodies for capture and detection of calprotectin. The LIAISON® Calprotectin assay must be run on the LIAISON® Analyzer family, a fully automated system with continuous loading.
Calprotectin is first extracted from human stool samples with LIAISON® Q.S.E.T. Buffer using either the weigh method, the LIAISON® Q.S.E.T. Device or the LIAISON® Q.S.E.T. Device Plus. The assay incubates extracted sample, calibrator, control, or calibration verifiers with assay buffer and paramagnetic particles coated with a monoclonal antibody that specifically recognizes the calprotectin heterocomplex. Following incubation, a wash cycle is performed to remove any unbound material. An isoluminol conjuqated monoclonal antibody that recognizes calprotectin is then added to the reaction and incubated. The unbound conjugate is removed with a second wash step. Starter reagents are then added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of calprotectin present in the calibrators, controls or samples.
All assay steps and incubations are performed by the LIAISON® XL Analyzer. The analyzer software automatically calculates the concentration of calprotectin in the sample. This concentration is expressed in ug/g.
The Q.S.E.T. Device Plus differs from its predicate Q.S.E.T. Device in that it is provided ready to use, and comes prefilled with the same extract buffer as required for use with the Q.S.E.T. Device, eliminating the need for the user to prepare the buffer and add it to the device themselves. In addition, minor changes to the shape and design of the tube were made.
The provided text is related to the FDA 510(k) premarket notification for the DiaSorin LIAISON® Calprotectin assay and the LIAISON® Q.S.E.T. Device Plus. None of the information requested in your prompt (Acceptance criteria, Study proving device meets criteria, Sample size, data provenance, expert numbers, etc. for AI/clinical studies) is present in the document.
The document describes an in vitro diagnostic (IVD) chemiluminescent immunoassay (CLIA) for fecal calprotectin, intended as an aid in diagnosing Inflammatory Bowel Diseases (IBD) and differentiating it from Irritable Bowel Syndrome (IBS). It also details a device for stool sample preparation.
The performance data included in the document specifically refers to the analytical performance of the IVD device and its sample preparation component, such as:
- Precision/Reproducibility: This section details the reproducibility of the LIAISON® Q.S.E.T. Device Plus extraction using five (5) stool samples with 90 measurements per sample (6 replicates over 5 days by 3 operators). It also shows sampling reproducibility using 5 human stool specimens, sampled by 3 operators with 5 replicates per specimen per operator on 3 lots of devices, totaling 225 sampling events.
- Reagent Stability: Mentions stability of the LIAISON® Q.S.E.T. Device Plus at 2-8°C for 12 months.
- Specimen Stability: Provides stability data for stool specimens under various storage conditions (refrigerated, room temperature, frozen, freeze/thaw cycles) and for sample extracts.
Therefore, I cannot provide the requested information regarding acceptance criteria for an AI/clinical study, the study setup to prove meeting those criteria, sample sizes for test/training sets in an AI context, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance, as these concepts are not applicable to the analytical validation described for this IVD device.
The document solely focuses on the analytical performance validation of an IVD immunoassay, not on clinical performance or AI/machine learning aspects.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 26, 2022
DiaSorin Inc. Christa Blaisdell Senior Regulatory Specialist 1951 Northwestern Ave Stillwater, Minnesota 55082
Re: K213858
Trade/Device Name: LIAISON Calprotectin, LIAISON O.S.E.T. Device Plus Regulation Number: 21 CFR 866.5180 Regulation Name: Fecal Calprotectin Immunological Test System Regulatory Class: Class II Product Code: NXO Dated: December 9, 2021 Received: December 10, 2021
Dear Christa Blaisdell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ying Mao, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213858
Device Name LIAISON® Calprotectin LIAISON® Q.S.E.T. Device Plus
Indications for Use (Describe)
The DiaSorin LIAISON® Calprotectin assay is an in vitro diagnostic chemiluminescent immunoassay (CLIA) intended for the quantitative measurement, in human stool, of fecal calprotectin, a neutrophilic protein that is a marker of mucosal inflammation. The LIAISON® Calprotectin assay can be used as an aid in the diagnosis of inflammatory bowel diseases (IBD), specifically Crohn's disease and ulcerative colitis, and as an aid in differentiation of IBD from irritable bowel syndrome (IBS). Test results are to be used in conjunction with information obtained from the patients' clinical evaluation and other diagnostic procedures.
The test has to be performed on the LIAISON® Analyzer Family.
The DiaSorin LIAISON® Q.S.E.T. Device Plus (Quantitative Stool Extraction and Test) is intended for use in the preparation of human stool specimens for testing in the LIAISON® Calprotectin assay.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
| SUBMITTED BY: | Christa BlaisdellSenior Regulatory Affairs SpecialistDiaSorin Inc.1951 Northwestern AvenueP.O. Box 285Stillwater, MN 55082-0285Phone (651) 351-5866Fax (651) 351-5669Email: christa.blaisdell@diasorin.com |
|---|---|
| DATE PREPARED: | July 25, 2022 |
| NAME OF DEVICE: | |
| Trade Name: | LIAISON® CalprotectinLIAISON® Q.S.E.T. Device Plus |
| Common Names/Descriptions: | Calprotectin assay |
| Classification Names: | Fecal calprotectin immunological test system:Class II, 21 CFR: 866.5180; Immunology (82) |
| Product Code: | NXO |
| PREDICATE DEVICES : | LIAISON® Calprotectin, LIAISON® Q.S.E.T. Device(K182698) |
DEVICE DESCRIPTION:
INTENDED USE:
The DiaSorin LIAISON® Calprotectin assay is an in vitro diagnostic chemiluminescent immunoassay (CLIA) intended for the quantitative measurement, in human stool, of fecal calprotectin, a neutrophilic protein that is a marker of mucosal inflammation. The LIAISON® Calprotectin assay can be used as an aid in the diagnosis of inflammatory bowel diseases (IBD), specifically Crohn's disease and ulcerative colitis, and as an aid in differentiation of IBD from irritable bowel syndrome (IBS). Test results are to be used in conjunction with information obtained from the patients' clinical evaluation and other diagnostic procedures.
The test has to be performed on the LIAISON® Analyzer family.
The DiaSorin LIAISON® Q.S.E.T. Device Plus (Quantitative Stool Extraction and Test) is intended for use in the preparation of human stool specimens for testing in the LIAISON® Calprotectin assay.
INDICATIONS FOR USE Same as intended use.
{4}------------------------------------------------
KIT DESCRIPTION:
The LIAISON® Calprotectin assay is a sandwich assay that uses 2 monoclonal antibodies for capture and detection of calprotectin. The LIAISON® Calprotectin assay must be run on the LIAISON® Analyzer family, a fully automated system with continuous loading.
Calprotectin is first extracted from human stool samples with LIAISON® Q.S.E.T. Buffer using either the weigh method, the LIAISON® Q.S.E.T. Device or the LIAISON® Q.S.E.T. Device Plus. The assay incubates extracted sample, calibrator, control, or calibration verifiers with assay buffer and paramagnetic particles coated with a monoclonal antibody that specifically recognizes the calprotectin heterocomplex. Following incubation, a wash cycle is performed to remove any unbound material. An isoluminol conjuqated monoclonal antibody that recognizes calprotectin is then added to the reaction and incubated. The unbound conjugate is removed with a second wash step. Starter reagents are then added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of calprotectin present in the calibrators, controls or samples.
All assay steps and incubations are performed by the LIAISON® XL Analyzer. The analyzer software automatically calculates the concentration of calprotectin in the sample. This concentration is expressed in uq/q.
COMPARISON WITH PREDICATE
The LIAISON® Calprotectin assay is the same as cleared for the market by K182968, with additional specimen stability claims included within the Instructions for Use.
The Q.S.E.T. Device Plus differs from its predicate Q.S.E.T. Device in that it is provided ready to use, and comes prefilled with the same extract buffer as required for use with the Q.S.E.T. Device, eliminating the need for the user to prepare the buffer and add it to the device themselves. In addition, minor changes to the shape and design of the tube were made.
A comparison of the similarities and differences between the devices are provided in the following tables.
{5}------------------------------------------------
| Characteristic | Predicate DeviceLIAISON® Calprotectin (K182698cleared 12/26/2018) | Updated DeviceLIAISON® Calprotectin |
|---|---|---|
| Intended use/Indications forUse | The DiaSorin LIAISON®Calprotectin assay is an in vitrodiagnostic chemiluminescentimmunoassay (CLIA) intended forthe quantitative measurement, inhuman stool, of fecal calprotectin,a neutrophilic protein that is amarker of mucosal inflammation.The LIAISON® Calprotectin assaycan be used as an aid in thediagnosis of inflammatory boweldiseases (IBD), specificallyCrohn's disease and ulcerativecolitis, and as an aid indifferentiation of IBD from irritablebowel syndrome (IBS). Testresults are to be used inconjunction with informationobtained from the patients' clinicalevaluation and other diagnosticprocedures.The test has to be performed onthe LIAISON® XL Analyzer.The DiaSorin LIAISON® Q.S.E.T.Device (Quantitative StoolExtraction and Test) is intendedfor use in the preparationof human stool specimens fortesting in the LIAISON®Calprotectin assay. | The DiaSorin LIAISON®Calprotectin assay is an in vitrodiagnostic chemiluminescentimmunoassay (CLIA) intended forthe quantitative measurement, inhuman stool, of fecal calprotectin,a neutrophilic protein that is amarker of mucosal inflammation.The LIAISON® Calprotectin assaycan be used as an aid in thediagnosis of inflammatory boweldiseases (IBD), specificallyCrohn's disease and ulcerativecolitis, and as an aid indifferentiation of IBD from irritablebowel syndrome (IBS). Testresults are to be used inconjunction with informationobtained from the patients' clinicalevaluation and other diagnosticprocedures.The test has to be performed onthe LIAISON® Analyzer Family.The DiaSorin LIAISON® Q.S.E.T.Device Plus (Quantitative StoolExtraction and Test) is intendedfor use in the preparationof human stool specimens fortesting in the LIAISON®Calprotectin assay. |
| Principles of operation | Sandwich assay that uses 2monoclonal antibodies for captureand detection of calprotectin | Same |
| Materials provided | Magnetic particles (2.4 mL)Conjugate (25.0 mL)Assay Buffer (27.0 mL)Specimen Diluent (13.0 mL)Number of tests: 100Calibrator 1 (2 x 1.0 mL)Calibrator 2 (2 x 1.0 mL) | Same |
| Optional Laboratory Supplies | LIAISON® Q.S.E.T. Device(REF 319050) | LIAISON® Q.S.E.T. Device(REF 319050)LIAISON® Q.S.E.T. Device Plus(REF 319060) |
| Reagent stability(unopened, 2-8°C) | Up to the expiration date | Same |
| Reagent stability(opened, 2-8°C) | Upright or on board the Analyzerfor 56 days | Same |
| Table 1: Comparison of the LIAISON® Calprotectin assay to the predicate device | ||||||
|---|---|---|---|---|---|---|
| -- | -- | -------------------------------------------------------------------------------- | -- | -- | -- | -- |
{6}------------------------------------------------
| Characteristic | Predicate DeviceLIAISON® Calprotectin (K182698cleared 12/26/2018) | Updated DeviceLIAISON® Calprotectin |
|---|---|---|
| Calibrator stability(Room temperature) | 6 hours | Same |
| Calibrator stability(2-8°C) | 28 days | Same |
| Specimen storage stability(Room temperature) | No value provided | Stable up to 6 hours |
| Specimen storage stability(2-8°C) | 72 hours | Same |
| Specimen storage stability(-20°C) | Store frozen at -20°C or below. | Stable up to 16 weeks frozen at -20°C |
| Specimen extract storagestability(Room temperature, 18-25°C) | 4 hours | Same |
| Specimen extract storagestability(2-8°C) | 6 hours | Same |
| Specimen extract storagestability(-20°C) | No value provided | Stable 7 days |
| Specimen extract storagestability(Freeze/thaw cycle) | No value provided | Stable through 1 freeze/thawcycle |
| Partial Specimen extractstorage stability(Room temperature) | NA | Stable 8 hours |
| Partial Specimen extractstorage stability(2-8°C) | NA | Stable 7 days |
{7}------------------------------------------------
| Characteristic | Predicate DeviceLIAISON® Q.S.E.T. Device(K182698 cleared 12/26/2018) | Updated DeviceLIAISON® Q.S.E.T. Device Plus |
|---|---|---|
| Intended use | The DiaSorin LIAISON® Q.S.E.T.Device Plus (Quantitative StoolExtraction and Test) is intendedfor use in the preparation ofhuman stool specimens for testingin the LIAISON® Calprotectinand/or LIAISON® Elastase-1assays. | Same |
| Physical attributes | Flat bottom tube with cap | Round bottom tube (longer andwider) |
| 2 caps: 1 for tube, 1 cap includessampling wand and funnel | 1 cap includes sample wand andfunnel | |
| Sampling wand | Wand length is longer toaccommodate larger tube | |
| Funnel to remove excess stool | Same | |
| Material: polypropylene | Same | |
| Average volume of samplecollected | 11.8 mg | 11.3 mg |
| Materials provided | Polypropylene mixing tube withblack funnel, screw cap, and whitesampling wand with blue cap. | Polypropylene mixing tube withblack funnel, screw cap, and whitesampling wand with blue cap. Thetube contains 6.0 mL of Q.S.E.T.Buffer, a buffered solutioncontaining BSA, surfactant, 0.1%ProClin®300 and 0.05%gentamicin sulfate. |
| Number of Tests | 100 devices. Each device is for 1time useProvided separately, not includedwith kit | Same |
| Method | LIAISON® Q.S.E.T. Buffer isprovided separately. User mustprepare buffer and add to device. | Device is pre-filled with LIAISON®Q.S.E.T. Buffer. Ready to Use. |
| Storage | Room temperature | 2-8°C |
Table 2: Comparison of the LIAISON® Q.S.E.T. Device Plus to the predicate device
{8}------------------------------------------------
PERFORMANCE DATA:
PRECISION/REPRODUCIBILITY:
LIAISON® Q.S.E.T. DEVICE PLUS EXTRACTION REPRODUCIBILITY
LIAISON® Q.S.E.T. Device Plus extraction reproducibility was tested using five (5) stool samples spanning the analytical measuring range of the assay. Samples were extracted using the LIAISON® Q.S.E.T. Device Plus and tested once (1) per day using six (6) replicates over five (5) days by three (3) operators, for a total of 90 measurements per sample. Each LIAISON® Q.S.E.T. Device Plus extraction was performed daily by each operator independently. CLSI document EP15-A3 was consulted in the preparation of the testing protocol.
| SampleID | N | mean | Repeatability | Between-Day | Within-Operator | Between-Operator | Total | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| µg/g | SD(µg/g) | %CV | SD(µg/g) | %CV | SD(µg/g) | %CV | SD(µg/g) | %CV | SD(µg/g) | %CV | ||
| Sample 1 | 90 | 17.2 | 0.42 | 2.5% | 1.39 | 8.1% | 1.45 | 8.4% | 1.26 | 7.3% | 1.81 | 10.5% |
| Sample 2 | 90 | 37.9 | 0.88 | 2.3% | 5.63 | 14.9% | 5.69 | 15.0% | 4.29 | 11.3% | 6.67 | 17.6% |
| Sample 3 | 90 | 120 | 2.43 | 2.0% | 12.1 | 10.1% | 12.3 | 10.3% | 6.29 | 5.2% | 12.7 | 10.6% |
| Sample 4 | 90 | 264 | 7.27 | 2.8% | 29.3 | 11.1% | 30.0 | 11.4% | 24.5 | 9.3% | 36.5 | 13.8% |
| Sample 5 | 90 | 1018 | 36.3 | 3.6% | 119 | 11.7% | 124 | 12.1% | 69.4 | 6.8% | 131 | 12.9% |
Table 6: Precision/Reproducibility
Reproducibility
LIAISON® Q.S.E.T. Device Plus sampling reproducibility was tested using 5 human stool specimens ranging from 2 – 7 on the Bristol Stool Form Scale (BSFS) which were sampled by 3 operators with 5 replicates per specimen per operator on 3 lots of devices for a total of 225 sampling events. The weight of each sample collected by the Q.S.E.T. Device Plus was determined by comparison against the empty device. Repeatability was assessed across all 5 specimens, replicates, operators, and lots.
| Precision of Q.S.E.T. Plus sample weight | ||||||
|---|---|---|---|---|---|---|
| Sample | BSFS | Sample Weight(mg Mean ±SD) | Repeatability | Between-Operator(% CV) | Between-Lot | Within-Lab |
| 1 | 2 | 10.23 ±1.05 | 10.29 | 0.00 | 0.00 | 10.29 |
| 2 | 4 | 11.02 ±0.87 | 7.91 | 0.00 | 0.00 | 7.91 |
| 3 | 6 | 12.00 ±0.71 | 5.56 | 0.00 | 2.04 | 5.93 |
| 4 | 7 | 11.39 ±0.91 | 7.37 | 3.14 | 0.00 | 8.01 |
| 5 | 5 | 11.79 ±0.73 | 6.21 | 0.00 | 0.00 | 6.21 |
Table 8 : Sample collection performance of LIAISON® Q.S.E.T. Device Plus
| Mean Sample Weight | 11.3 mg |
|---|---|
| Median Sample Weight | 11.4 mg |
| Range | 8.5-13.5 mg |
| 95% CI | 11.2-11.4 mg |
| SD | 1.06 |
| %CV | 9.38% |
{9}------------------------------------------------
REAGENT STABILITY
Table 9: Reagent Stability
| LIAISON Q.S.E.T. Device Plus | Stability |
|---|---|
| 2 - 8°C | 12 months |
SPECIMEN STABILITY
Studies were performed to determine the stability of sample at different storage conditions. The results are provided in the tables below.
Table 10: Specimen Stability
| Stool Specimen | |
|---|---|
| Storage Condition | Stability |
| Refrigerated at 2- 8°C | 72 hours |
| Room temperature (18 - 25°C) | 6 hours |
| Frozen at -20°C | 16 weeks |
| Freeze/Thaw cycles | 3 cycles |
| Sample Extract – LIAISON® Q.S.E.T. Device/Plus Methods | |
|---|---|
| Storage Condition | Stability |
| Room temperature (18 - 25°C) | 4 hours |
| Refrigerated at 2- 8°C | 6 hours |
| Frozen at -20°C | 7 days |
| Freeze/Thaw cycles | 1 cycle |
| Refrigerated at 2- 8°C (with centrifugation) | 7 days |
| Partial Sample Extract - LIAISON® Q.S.E.T. Device Plus Methods | |
|---|---|
| Storage Condition | Stability |
| Room temperature (18 - 25°C) | 8 hours |
| Refrigerated at 2- 8°C | 7 days |
CONCLUSION:
The material submitted in this premarket notification is complete and supports a substantial equivalence decision. The labeling is sufficient and it satisfies the requirements of 21CFR 809.10.
§ 866.5180 Fecal calprotectin immunological test system.
(a)
Identification. A fecal calprotectin immunological test system is anin vitro diagnostic device that consists of reagents used to quantitatively measure, by immunochemical techniques, fecal calprotectin in human stool specimens. The device is intended forin vitro diagnostic use as an aid in the diagnosis of inflammatory bowel diseases (IBD), specifically Crohn's disease and ulcerative colitis, and as an aid in differentiation of IBD from irritable bowel syndrome.(b)
Classification. Class II (special controls). The special control for these devices is FDA's guidance document entitled “Class II Special Controls Guidance Document: Fecal Calprotectin Immunological Test Systems.” For the availability of this guidance document, see § 866.1(e).