K Number
K213662
Date Cleared
2022-03-28

(126 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large. The subject device is non-sterile.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the device, Disposable nitrile rubber protective gloves (Tested for Use with Chemotherapy Drugs), meets those criteria.

Here's the information extracted and formatted as requested:

1. Table of acceptance criteria and the reported device performance:

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D6319Physical Dimensions TestLength (mm): S:≥220; M/L/XL:≥230; Width (mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10; Thickness (mm): Finger: ≥0.05; Palm: ≥0.05Length (mm): ≥ 230/Pass; Width (mm): S: 85-89 /Pass; M: 94-97/ Pass; L: 100-104/ Pass; XL:110-113/ Pass; Thickness (mm): Finger: 0.16-0.18/Pass; Palm: 0.11-0.12/Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125/Pass
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 2.0mg0.15-0.19mg/Pass
ASTM D412Physical properties (Tensile Strength)Before Aging: Tensile Strength ≥14MPa; Ultimate Elongation ≥500%; After Aging: Tensile Strength ≥14MPa; Ultimate Elongation ≥400%Before Aging: 16.5-33.1MPa/Pass; 500-585%/Pass; After Aging: 14.1-28.3MPa/Pass; 482-555%/Pass
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, device extract is cytotoxic (Note: This is a discrepancy identified in the document, but the device was still cleared based on other testing)
ISO 10993-11Acute systemic toxicityNon-acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivo / Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant/ Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer./ Pass
ASTM D6978-05Chemotherapy Drug Permeation (Breakthrough Detection Time)-Carmustine (BCNU) 3.3 mg/ml: 33.5 Minutes; Cisplatin 1.0 mg/ml: > 240 Minutes; Cyclophosphamide 20.0 mg/ml: > 240 Minutes; Dacarbazine 10.0 mg/ml: > 240 Minutes; Doxorubicin HCl 2.0 mg/ml: > 240 Minutes; Etoposide 20.0 mg/ml: > 240 Minutes; Fluorouracil 50.0 mg/ml: > 240 Minutes; Paclitaxel 6.0 mg/ml: >240 Minutes; Thio Tepa 10.0 mg/ml: 62.6 Minutes. Warning: Please do not use with Carmustine (BCNU) and Thiotepa.

2. Sample size used for the test set and the data provenance:

  • Sample size for Watertightness Test (ASTM D5151): 125 gloves (indicated by "0/125/Pass").
  • Data provenance: Not explicitly stated, but based on the context of a 510(k) submission, this would be data generated by the manufacturer from laboratory testing (bench testing) of their product. It is inherently retrospective as the testing precedes the submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This device is a medical glove, and the acceptance criteria are based on standardized physical and chemical performance tests rather than expert interpretation of medical images or patient data. The "ground truth" is defined by the objective measurements and pass/fail criteria of the ASTM and ISO standards.

4. Adjudication method for the test set:

  • Not applicable. The tests are objective measurements against defined standards, not subjective assessments requiring expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI-powered diagnostic device, but a medical glove. MRMC studies are used for evaluating diagnostic performance with and without AI assistance for human readers of medical images.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI algorithm.

7. The type of ground truth used:

  • The ground truth is established by objective measurements against established international standards (ASTM and ISO) for physical properties, barrier integrity, biocompatibility, and chemical permeation.

8. The sample size for the training set:

  • Not applicable. This is not a machine learning or AI device that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable. No training set is used for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 28, 2022 Taishan Weibang Medical Equipment Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful information Technology Co., Ltd. RM.1801,No.161 Lujiazui East Rd.,Pudong Shanghai. 200120 China

Re: K213662

Trade/Device Name: Disposable nitrile rubber protective gloves (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: March 11, 2022 Received: March 17, 2022

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213662

Device Name

Disposable nitrile rubber protective gloves (Tested for Use with Chemotherapy Drugs)

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes
Carmustine (BCNU)3.3 mg/ml (3,300 ppm)33.5 Minutes
Cisplatin1.0 mg/ml(1,000 ppm)> 240 Minutes
Cyclophosphamide (Cytoxan)20.0 mg/ml(20,000 ppm)> 240 Minutes
Dacarbazine (DTIC)10.0 mg/ml(10,000 ppm)> 240 Minutes
Doxorubicin HCL (Adriamycin)2.0 mg/ml(2,000 ppm)> 240 Minutes
Etoposide (Toposar)20.0 mg/ml(20,000 ppm)> 240 Minutes
Fluorouracil (Adrucil)50.0 mg/ml(50,000 ppm)> 240 Minutes
Paclitaxel (Taxol)6.0 mg/ml(6,000 ppm)> 240 Minutes
Thio TEPA10.0 mg/ml(10,000 ppm)62.6 Minutes

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 33.5 Minutes Thio-Tepa 10.0 mg/ml 62.6 Minutes

Warning: Please do not use with Carmustine (BCNU) and Thiotepa.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K213662

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: Taishan Weibang Medical Equipment Co., Ltd. Address: No.8, Jiangdong Industrial Park, Dajiang Town, Taishan, Jiangmen, Guangdong, China. Contact: Shuhui Zhou Date of Preparation: Mar.28, 2022

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 Lujiazui East Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Disposable nitrile rubber protective gloves (Tested for Use with Chemotherapy Drugs) Patient Examination Gloves Common name: Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL

3.0 Classification

Production code: LZA,LZC Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Ever Growth (Vietnam) Co., Ltd. Device: Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy

{4}------------------------------------------------

Drugs, Orange Color 510(k) number: K190860

5.0 Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large. The subject device is non-sterile.

6.0 Indication for Use_

A patient examination qlove is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs

Chemotherapy DrugConcentrationBreakthrough DetectionTime in Minutes
Carmustine (BCNU)3.3 mg/ml (3,300 ppm)33.5
Cisplatin1.0 mg/ml(1,000 ppm)> 240
Cyclophosphamide(Cytoxan)20.0 mg/ml(20,000 ppm)> 240
Dacarbazine (DTIC)10.0 mg/ml(10,000 ppm)> 240
Doxorubicin HCl2.0 mg/ml(2,000 ppm)> 240
Etoposide20.0 mg/ml(20,000 ppm)> 240
Fluorouracil50.0 mg/ml(50,000 ppm)> 240
Paclitaxel6.0 mg/ml(6,000 ppm)> 240
Thio Tepa10.0 mg/ml(10,000 ppm)62.6

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 33.5 Minutes; Thio Tepa 10.0 mg/ml 62.6 Minutes.

Warning: Please do not use with Carmustine (BCNU) and Thiotepa.

7.0 Technological Characteristic Comparison Table

{5}------------------------------------------------

ItemSubject Device(K213662)Predicate Device(K190860)Remark
Product CodeLZA,LZCLZA,LZCSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended Use/Indications for UseA patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.Same
Powdered or PoweredfreePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
SterilityNon-SterileNon-SterileSame
Labeling InformationSingle-use indication, powder free, device color, device name, glove size and quantity,Non-Sterile, a statement of standard ASTM D6978-05 compliance and a summary of the testing results.Single-use indication, powder free, device color, device name, glove size and quantity, Non-Sterile, a statement of standard ASTM D6978-05 compliance and a summary of the testing results.Same
Dimensions(mm)Length:S:≥220;M/L/XL: ≥230;Width:S: 80±10;M: 95±10;L: 110±10;XL: 120±10Length:XS/S/M/L/XL: ≥230;Width:XS:70±10;S: 80±10;M: 95±10;L: 110±10;XL: 120±10SimilarAnalysis 1
Thickness(mm)Finger: ≥0.05;Finger: ≥0.05;Same

Table1-General Comparison

{6}------------------------------------------------

Palm: ≥0.05Palm: ≥0.05
ColorantBlueWhite, OrangeDifferentAnalysis 2
PhysicalPropertiesBeforeAgingTensileStrength14MPa, minTensile Strength14MPa, minSame
UltimateElongation500% minUltimateElongation500% minSame
After AgingTensileStrength14MPa, minTensile Strength14MPa, minSame
UltimateElongation400%minUltimateElongation400%minSame
Freedom from HolesBe free from holes whentested in accordance withASTMD5151 AQL=2.5Be free from holes whentested in accordance withASTMD5151 AQL=2.5Same
Powder Content0.15-0.19 mg per glove,Meet the requirements ofASTM D6124Meet the requirements ofASTM D6124SimilarAnalysis 3
BiocompatibilityISO 10993-10;Under the conditions of thestudy, not an irritant or asensitizer.ISO 10993-10;Under the conditions of thestudy, not an irritant or asensitizerSame
ISO 10993-5Under conditions of thestudy, device extract iscytotoxic.ISO 10993-5Under conditions of thestudy, device extract is notcytotoxicDifferentAnalysis 4
ISO 10993-11;Under thecondition of acutesystemic toxicity test,the test article did notshow acute systemictoxicity in vivo.N.A./
Chemotherapy DrugsTested with MinimumBreakthroughDetection Time asCarmustine (BCNU) 3.3mg/ml: 33.5 MinutesCarmustine (BCNU) 3.3mg/ml:White:11.8 Minutes;Orange:31.6MinutesSimilarAnalysis 5
Tested per ASTM D6978Cisplatin 1.0 mg/ml: > 240MinutesCisplatin 1.0 mg/ml: >240MinutesSame

{7}------------------------------------------------

Cyclophosphamide(Cytoxan) 20.0 mg/ml: >240 MinutesCyclophosphamide(Cytoxan)20.0 mg/ml: > 240 MinutesSame
Dacarbazine (DTIC) 10.0 mg/ml: > 240 MinutesDacarbazine (DTIC) 10.0 mg/ml:>240 MinutesSame
Doxorubicin HCl 2.0 mg/ml:> 240 MinutesDoxorubicin Hydrochloride2.0mg/ml: >240 MinutesSame
Etoposide 20.0 mg/ml: >240 MinutesEtoposide (Toposar) 20.0mg/ml: >240 MinutesSame
Fluorouracil 50.0 mg/ml: >240 MinutesFluorouracil 50.0 mg/ml: >240MinutesSame
Paclitaxel 6.0 mg/ml: >240MinutesPaclitaxel (Taxol) 6.0mg/ml:>240 MinutesSame
Thio Tepa 62.6 mg/ml: 15.2MinutesThio-Tepa 10.0 mg/ml:White:16.9 Minutes;Orange: 72.5 MinutesSimilarAnalysis 5

Analysis 1:

The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

Analysis 2:

The color of the subject device is different with that of the predicate. The subject device was evaluated according to ISO 10993-1 standards, and there were no risks identified.

Analysis 3:

Powder Content of subject device is similar with that of the predicate, because the predicate did not publish the exact results of the powder content. But they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

Analysis 4:

Under conditions of the study, cytotoxicity of the subject device is different with that of the predicate. But under the condition of acute systemic toxicity test, the test article did not show acute systemic toxicity in vivo. So there were no risks identified.

Analysis 5:

{8}------------------------------------------------

And Breakthrough detection times of Carmustine (BCNU) and Thio Tepa of subject device are different with those of the predicate. The Chemotherapy Labeling Claims has clearly defined on the labeling. So it does not raise any new safety or performance questions.

8.0 Summary of Non-Clinical Testing

Biocompatibility Testing

The biocompatibility evaluation for Medical Examination Gloves (Tested for Use with Chemotherapy Drugs) was conducted in accordance with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization.

ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.

ISO 10993-11 Third edition 2017-09, Biological evaluation of medical devices -Part 11: Tests for systemic toxicity.

Performance Testing (Bench)

Physical performance qualities of the proposed device were evaluated per ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs.

In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • ASTM D6124-06 (Reapproved 2017), Standard Test Method for -Residual Powder on Medical Gloves
  • ASTM D5151-19, Standard Test Method for Detection of Holes in -Medical Gloves.

{9}------------------------------------------------

  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves i for Medical Application.
  • ASTM D 6978-05 (Reapproved 2019) ,Standard Practice for । Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
Test MethodPurposeAcceptance CriteriaResults
ASTM D6319Physical Dimensions TestLength(mm):S:≥220;M/L/XL:≥230;Width(mm):S: 80±10;M: 95±10;L: 110±10;XL: 120±10Thickness (mm):Finger: ≥0.05Palm: ≥0.05Length(mm):≥ 230/Pass;Width(mm):S: 85-89 /PassM: 94-97/ PassL: 100-104/ PassXL:110-113/ PassThickness (mm):Finger: 0.16-0.18/PassPalm: 0.11-0.12/Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151AQL 2.50/125/Pass
ASTM D6124Powder ContentMeet the requirements of ASTM D61242.0mg0.15-0.19mg/Pass;
ASTM D412Physical propertiesBefore Aging Tensile Strength ≥14MPaUltimate Elongation ≥500%After Aging Tensile Strength ≥14MPaUltimate Elongation ≥400%16.5-33.1MPa/Pass;500-585%/Pass;14.1-28.3MPa/Pass;482-555%/Pass;
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, device extract is cytotoxic
ISO10993-11Acute systemictoxicityNon- acute systemictoxicityUnder conditions ofthe study, did notshow acute systemictoxicity in vivo / Pass
ISO10993-10IrritationNon-irritatingUnder the conditionsof the study, not anirritant/ Pass
ISO10993-10SensitizationNon-sensitizingUnder conditions ofthe study, not asensitizer./ Pass

Table 2 - Summary of non-clinical performance testing

{10}------------------------------------------------

9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Disposable nitrile rubber protective gloves (Tested for Use with Chemotherapy Drugs) is as safe, as effective, and performs as well as or better than the legally marketed predicated device under K190860.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.