Nitrylex® Classic Powder Free Nitrile Blue Examination Gloves Tested for use with Chemotherapy Drugs are disposable devices intended for medical propose that are worn on the examiner's hands to prevent contamination between patent and examiner.
The proposed device was tested for use with Chemotherapy Drugs as per ASTM D678-05 Standard Practice for Assessment Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. Chemotherapy Drugs Permeation.
Nitrylex® Classic Powder Free Nitrile Blue Examination Gloves Tested for use with Chemotherapy Drugs are disposable devices intended for medical propose that are worn on the examiner's hands to prevent contamination between patent and examiner.
Here's a breakdown of the acceptance criteria and the study information for the Nitrylex® Classic Powder Free Nitrile Blue Examination Gloves, based on the provided FDA 510(k) clearance document:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Chemotherapy Drugs | Concentration | Acceptance Criteria (Breakthrough Detection Time) | Reported Device Performance (Breakthrough Detection Time) |
|---|---|---|---|
| Bleomycin Sulfate | 15 mg/ml (15,000 ppm) | >240 min. | >240 min. |
| Busulfan | 6 mg/ml (6,000 ppm) | >240 min. | >240 min. |
| Carboplatin | 10 mg/ml (10,000 ppm) | >240 min. | >240 min. |
| Cisplatin | 1 mg/ml (1,000 ppm) | >240 min. | >240 min. |
| Cyclophosphamide (Cytoxan) | 20.0 mg/ml (20,000 ppm) | >240 min. | >240 min. |
| Cytarabine HCl | 100 mg/ml (100,000 ppm) | >240 min. | >240 min. |
| Cytovene | 10 mg/ml (10,000 ppm) | >240 min. | >240 min. |
| Dacarbazine | 10 mg/ml (10,000 ppm) | >240 min. | >240 min. |
| Daunorubicin HCl | 5 mg/ml (5,000 ppm) | >240 min. | >240 min. |
| Docetaxel | 10 mg/ml (10,000 ppm) | >240 min. | >240 min. |
| Doxorubicin HCl | 2.0 mg/ml (2,000 ppm) | >240 min. | >240 min. |
| Epirubicin HCl | 2 mg/ml (2,000 ppm) | >240 min. | >240 min. |
| Etoposide | 20.0 mg/ml (20,000 ppm) | >240 min. | >240 min. |
| Fludarabine | 25 mg/ml (25,000 ppm) | >240 min. | >240 min. |
| Fluorouracil | 50.0 mg/ml (50,000 ppm) | >240 min. | >240 min. |
| Gemcitabine | 38 mg/ml (38,000 ppm) | >240 min. | >240 min. |
| Idarubicin HCl | 1 mg/ml (1,000 ppm) | >240 min. | >240 min. |
| Ifosfamide | 50 mg/ml (50,000 ppm) | >240 min. | >240 min. |
| Irinotecan | 20 mg/ml (20,000 ppm) | >240 min. | >240 min. |
| Mechlorethamine HCl | 1 mg/ml (1,000 ppm) | >240 min. | >240 min. |
| Melphalan | 5 mg/ml (5,000 ppm) | >240 min. | >240 min. |
| Methotrexate | 25 mg/ml (25,000 ppm) | >240 min. | >240 min. |
| Mitomycin C | 0.5 mg/ml (500 ppm) | >240 min. | >240 min. |
| Mitoxantrone | 2 mg/ml (2,000 ppm) | >240 min. | >240 min. |
| Oxaliplatin | 2 mg/ml (2,000 ppm) | >240 min. | >240 min. |
| Paclitaxel | 6.0 mg/ml (6,000 ppm) | >240 min. | >240 min. |
| Rituximab | 10 mg/ml (10,000 ppm) | >240 min. | >240 min. |
| Trisenox | 1 mg/ml (1,000 ppm) | >240 min. | >240 min. |
| Vincristine Sulfate | 1 mg/ml (1,000 ppm) | >240 min. | >240 min. |
| Vinorelbine | 10 mg/ml (10,000 ppm) | >240 min. | >240 min. |
| Fentanyl Citrate Injection | 100 mcg/2 ml | >240 min. | >240 min. |
| Drugs Not Recommended for Use | |||
| Carmustine (BCNU) | 3.3 mg/ml | Not met (>240 min.) | 34.8 minutes |
| Thiotepa (THT) | 10.0 mg/ml | Not met (>240 min.) | 47.4 minutes |
The device meets the acceptance criteria for all listed chemotherapy drugs by demonstrating a breakthrough detection time of greater than 240 minutes, with the explicit exception of Carmustine and Thiotepa, for which a warning is issued.
2. Sample Size Used for the Test Set and Data Provenance
The document states that the proposed device was "tested for use with Chemotherapy Drugs as per ASTM D678-05 Standard Practice for Assessment Resistance of Medical Gloves to Permeation by Chemotherapy Drugs."
- Sample Size: The document does not explicitly state the number of gloves (samples) tested for each chemotherapy drug. However, ASTM D6978-05 generally specifies a minimum of three individual test specimens per drug for permeation testing.
- Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. It implies the data was collected specifically for this submission through testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This section is not applicable to this type of device and study. The "ground truth" here is objective scientific measurement (chemical breakthrough time) performed by laboratory testing according to a recognized standard (ASTM D6978-05), not expert interpretation of medical images or patient outcomes.
4. Adjudication Method for the Test Set
This section is not applicable. As mentioned above, the "truth" is determined by objective laboratory measurements and standardized methodology, not through human adjudication of differing opinions.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., radiologists reading images) and the impact of AI on their performance. This device is an examination glove, and its performance is measured through physical and chemical interaction, not human interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This concept is not applicable to this device. A "standalone" study (algorithm only) refers to the performance of an AI algorithm independent of human input. This device is a physical product (gloves) whose performance is evaluated through material testing against chemical permeation, not an AI algorithm.
7. The Type of Ground Truth Used
The ground truth used is objective laboratory measurement of chemical permeation. Specifically, it is the "Breakthrough Detection Time" as measured according to the ASTM D6978-05 standard. This is a direct physical measurement of the glove's resistance to different chemotherapy drugs.
8. The Sample Size for the Training Set
This section is not applicable. This device is a physical product and its performance is evaluated through direct physical/chemical testing, not by training an algorithm on a dataset. Therefore, there is no "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set was Established
This section is not applicable for the same reason as point 8.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 21, 2022
Mercator Medical (Thailand) LTD % Kevin Walls Principal Consultant Regulatory Insight, Inc. 33 Golden Eagle Lane Littleton. Colorado 80127
Re: K213548
Trade/Device Name: Nitrylex Classic Powder Free Nitrile Blue Examination Gloves Tested for Use with Chemotherapy Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: January 31, 2022 Received: February 1, 2022
Dear Kevin Walls:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213548
Device Name
Nitrylex® Classic Powder Free Nitrile Blue Examination Gloves Tested for use with Chemotherapy Drugs
Indications for Use (Describe)
Nitrylex® Classic Powder Free Nitrile Blue Examination Gloves Tested for use with Chemotherapy Drugs are disposable devices intended for medical propose that are worn on the examiner's hands to prevent contamination between patent and examiner.
The proposed device was tested for use with Chemotherapy Drugs as per ASTM D678-05 Standard Practice for Assessment Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. Chemotherapy Drugs Permeation.
The following chemicals have been tested with the proposed device.
| Test Chemotherapy Drugs | Concentration | Breakthrough Detection(Time In Minutes) |
|---|---|---|
| Bleomycin Sulfate | 15 mg/ml (15,000 ppm) | >240 min. |
| Busulfan | 6 mg/ml (6,000 ppm) | >240 min. |
| Carboplatin | 10 mg/ml (10,000 ppm) | >240 min. |
| Cisplatin | 1 mg/ml (1,000 ppm) | >240 min. |
| Cyclophosphamide (Cytoxan) | 20.0 mg/ml (20,000 ppm) | >240 min. |
| Cytarabine HCl | 100 mg/ml (100,000 ppm) | >240 min. |
| Cytovene | 10 mg/ml (10,000 ppm) | >240 min. |
| Dacarbazine | 10 mg/ml (10,000 ppm) | >240 min. |
| Daunorubicin HCl | 5 mg/ml (5,000 ppm) | >240 min. |
| Docetaxel | 10 mg/ml (10,000 ppm) | >240 min. |
| Doxorubicin HCl | 2.0 mg/ml (2,000 ppm) | >240 min. |
| Epirubicin HCl | 2 mg/ml (2,000 ppm) | >240 min. |
| Etoposide | 20.0 mg/ml (20,000 ppm) | >240 min. |
| Fludarabine | 25 mg/ml (25,000 ppm) | >240 min. |
| Fluorouracil | 50.0 mg/ml (50,000 ppm) | >240 min. |
| Gemcitabine | 38 mg/ml (38,000 ppm) | >240 min. |
| Idarubicin HCl | 1 mg/ml (1,000 ppm) | >240 min. |
| Ifosfamide | 50 mg/ml (50,000 ppm) | >240 min. |
| Irinotecan | 20 mg/ml (20,000 ppm) | >240 min. |
| Mechlorethamine HCl | 1 mg/ml (1,000 ppm) | >240 min. |
| Melphalan | 5 mg/ml (5,000 ppm) | >240 min. |
| Methotrexate | 25 mg/ml (25,000 ppm) | >240 min. |
| Mitomycin C | 0.5 mg/ml (500 ppm) | >240 min. |
| Mitoxantrone | 2 mg/ml (2,000 ppm) | >240 min. |
| Oxaliplatin | 2 mg/ml (2,000 ppm) | >240 min. |
| Paclitaxel | 6.0 mg/ml (6,000 ppm) | >240 min. |
| Rituximab | 10 mg/ml (10,000 ppm) | >240 min. |
| Trisenox | 1 mg/ml (1,000 ppm) | >240 min. |
| Vincristine Sulfate | 1 mg/ml (1,000 ppm) | >240 min. |
| Vinorelbine | 10 mg/ml (10,000 ppm) | >240 min. |
| Fentanyl Citrate Injection | 100 mcg/2 ml | >240 min. |
{3}------------------------------------------------
Please note that the following drugs showed break through detected in less than 240 minutes:
- Carmustine (BCNU), 3.3 mg/ml: 34.8 minutes
- Thiotepa (THT) 10.0 mg/ml: 47.4 minutes
Warning-Not for Use with Carmustine and Thiotepa
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.