K Number
K213478

Validate with FDA (Live)

Device Name
BD Pen Needle
Date Cleared
2022-06-28

(242 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Becton Dickinson Pen Needle is intended for use with pen injector devices for subcutaneous injection of drugs.

Device Description

The BD Pen Needle has the same intended use/indications for use, technological characteristics, and principles of operation as the predicate device cleared under K162516. The pen needle assembly consists of a double-ended cannula that is assembled into an injection-molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen-injector device. This allows the Non-Patient (NP) end of the cannula to penetrate through the rubber septum of the pen injector cartridge. The Patient and NP ends of the cannula are lubricated using silicone-based lubes for ease of injection and rubber septum penetration. An injection-molded inner shield is assembled over the Patient end of the cannula to protect the point from damage and accidental needle sticks. This needle assembly is inserted into a protective injection-molded outer cover and sealed with a peel-away (tear drop) label to provide a sterile barrier and tamper evidence. The outer cover is also used to remove the hub and cannula from the pen. The peel-away tear-drop label is pre-printed with information, which includes the lot number and needle gauge / length. The BD Pen Needle is a single-use disposable device and is provided sterile (gamma irradiation sterilization). It is non-toxic and non-pyrogenic.

AI/ML Overview

This document is an FDA 510(k) summary for a Becton Dickinson (BD) Pen Needle (K213478). It describes the device, its intended use, and comparative testing against a predicate device (K162516) to demonstrate substantial equivalence.

Based on the provided text, the device in question is a BD Pen Needle, a medical device used with pen injector devices for the subcutaneous injection of drugs. The information focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving that the device meets specific acceptance criteria as an AI/ML-driven diagnostic or treatment device with performance metrics like sensitivity, specificity, etc.

Therefore, the requested information regarding AI/ML acceptance criteria, study design (sample size, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance), training set details, and specific metrics (sensitivity, specificity, AUC) is not applicable to this document.

This 510(k) summary demonstrates substantial equivalence by:**

  • Comparing Indications for Use: The subject device has the identical indication for use to the predicate device.
  • Comparing Technological Characteristics: The devices share the same fundamental scientific technology (single lumen needle).
  • Demonstrating Performance Testing: Non-clinical tests were conducted based on ISO standards and BD internal requirements to show that the design, materials, and manufacturing process changes do not raise new questions of safety or effectiveness.

Here's a breakdown of the relevant information provided in the document:

1. A table of acceptance criteria and the reported device performance:

The document does not present acceptance criteria in the form of specific performance metrics (e.g., sensitivity, specificity, F1-score) and reported performance values as would be typical for an AI-driven device. Instead, it relies on demonstrating equivalence through adherence to relevant standards and direct comparison with a predicate device.

The "acceptance criteria" are implied by the standards and internal test requirements met, and the "reported device performance" is the successful compliance with these tests.

Characteristic / TestImplied Acceptance (Met if not stated otherwise)Reported Device Performance
Functional Performance (ISO 11608-2:2012, ISO 9626:2016, BD internal tests)Compliance with standards and internal requirements.Validated design; supporting evidence provided.
Biocompatibility (ISO 10993-1:2018)Compliance with ISO 10993-1:2018 (e.g., Cytotoxicity, Sensitization, Intracutaneous reactivity, Irritation, Material Mediated Pyrogenicity, Genotoxicity, Implantation and Particulate Analysis, Chemical Characterization, Toxicological Risk Assessment).Testing conducted according to standard; results are presumed to have met criteria (not explicitly listed).
Particulate Matter (USP<788>)Meet USP<788> acceptance criteria.Met the USP acceptance criteria.
Sterility (ISO 11137-2:2013)Minimum SAL of 10^-6.Validated gamma irradiation; SAL of 10^-6 achieved.
Endotoxin (LAL assay)Meet endotoxin limit requirement.Requirement met; product is non-Pyrogenic.
Sterile barrier testing (Tear drop label)Successful performance of tear drop label removal force and seal integrity.Testing performed; presumed to meet criteria.
Packaging Integrity (ASTM D4169-16)Meet requirements for product protection and sterility maintenance.All packaging deemed acceptable.
Shelf-life / StabilitySupport 5-year shelf-life (sterility and performance).Accelerated stability testing conducted to validate 5-year shelf-life.
Intended UseIdentical to predicate.Identical to predicate.
Sterilization MethodIdentical to predicate.Identical to predicate (gamma irradiation).
SAL 10^-6Identical to predicate.Identical to predicate (YES).
Tamper Evident FeatureIdentical to predicate.Identical to predicate (YES, tear-drop label).
Non-pyrogenic and Endotoxin FreeIdentical to predicate.Identical to predicate (YES).

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not detail specific sample sizes for "test sets" in the context of diagnostic performance evaluation. The "testing" mentioned refers to engineering, functional, and biocompatibility tests on the physical device components and assembled product, adhering to industrial standards. Data provenance (country of origin, retrospective/prospective) is not applicable or specified for these types of tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is not an AI/ML diagnostic device requiring expert-established ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not an AI/ML diagnostic device requiring adjudication of expert interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. For this type of device, "ground truth" is defined by adherence to engineering specifications, material properties, and biological safety standards rather than diagnostic accuracy against a clinical reference.

8. The sample size for the training set

Not applicable. This is not an AI/ML algorithm that has a "training set."

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

July 11, 2022

Becton, Dickinson and Company Charlton Foo Senior Manager Regulatory Affairs 1 Becton Drive Franklin Lakes, New Jersey 07417

Re: K213478

Trade/Device Name: BD Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: FMI

Dear Charlton Foo:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter issued June 28, 2022. Specifically, FDA is updating this SE Letter as an administrative correction to correct the device name on the Indications for Use form and include the missing date information on the original SE Letter which should have been June 28, 2022.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact CAPT Alan Stevens, by phone at 301-796-6294 or email at alan.stevens@fda.hhs.gov.

Sincerely,

Alan M.
Stevens -S3

CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Becton, Dickinson and Company Charlton Foo Senior Manager, Regulatory Affairs 1 Becton Drive Franklin Lakes, New Jersey 07417

Re: K213478

Trade/Device Name: BD Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: May 23, 2022 Received: May 25, 2022

Dear Charlton Foo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{2}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/2/Picture/5 description: The image shows a digital signature. The signature is for Courtney Evans -S. The date of the signature is 2022.06.28. The time of the signature is 10:46:16 -04'00'.

For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K213478

Device Name BD Pen Needle

Indications for Use (Describe)

Becton Dickinson Pen Needle is intended for use with pen injector devices for subcutaneous injection of drugs.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K213478.510K Summary

Submitter:

Submission Correspondent:

Charlton Foo Senior Manager Regulatory Affairs Tel: 857 270 1395

Applicant:

Becton Dickinson and Company 1 Becton Drive Franklin Lakes, NJ 07417

Contact:

Mark O'Donnell Vice President, Regulatory Affairs, Medical Segment Tel: 805 452 3449

  • Date Prepared: June 28, 2022
  • Device Name: Trade Name: BD Pen Needle Common Name: BD Pen Needle Regulation Name: Hypodermic Single Lumen Needle Regulation: 21 CFR 880.5570 Device Class: Class II Product Code:

Predicate

K162516: BD Pen Needle

Device Description:

The BD Pen Needle has the same intended use/indications for use, technological characteristics, and principles of operation as the predicate device cleared under K162516. The pen needle assembly consists of a double-ended cannula that is assembled into an injection-molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen-injector device. This allows the Non-Patient (NP) end of the cannula to penetrate through the rubber septum of the pen injector cartridge. The Patient and NP

{5}------------------------------------------------

ends of the cannula are lubricated using silicone-based lubes for ease of injection and rubber septum penetration. An injection-molded inner shield is assembled over the Patient end of the cannula to protect the point from damage and accidental needle sticks. This needle assembly is inserted into a protective injection-molded outer cover and sealed with a peel-away (tear drop) label to provide a sterile barrier and tamper evidence. The outer cover is also used to remove the hub and cannula from the pen. The peel-away tear-drop label is pre-printed with information, which includes the lot number and needle gauge / length. The BD Pen Needle is a single-use disposable device and is provided sterile (gamma irradiation sterilization). It is non-toxic and non-pyrogenic.

Image /page/5/Figure/1 description: The image shows the different parts of a pen needle. The parts are labeled as the outer shield, inner shield, pen needle, and tear drop label. The image shows the parts separated from each other.

Gauge (G)Length (mm)BevelWall Thickness
29G12.7mm3Thin Wall
31G8mm3Thin Wall
31G8mm5Thin Wall
31G6mm3Thin Wall
31G5mm3Thin Wall
31G5mm5Thin Wall
31G5mm3Extra Thin Wall
32G6mm3Thin Wall
32G4mm3Thin Wall
32G4mm5Thin Wall
32G4mm5Extra Thin Wall

The BD Pen Needles are offered in the following sizes:

Indication for Use:

CharacteristicsPredicate DeviceDevice Name(K162516)Subject DeviceDevice Name(K213478)
Indication for UseBecton Dickinson Pen Needleis intended for use with peninjector devices for thesubcutaneous injection ofdrugs.Becton Dickinson Pen Needleis intended for use with peninjector devices for thesubcutaneous injection ofdrugs.
Prescription Only or Over theCounterOver the CounterOver the Counter

The indications for use statement for the subject device is identical to the predicate device.

{6}------------------------------------------------

Technological Characteristics

The subject device has the same fundamental scientific technology as the predicate (K162516), which operates on the principles of a single lumen needle. Compared the predicate, the subject device consists of design, performance, method, material, and labeling changes. The submission provides supporting evidence on the performance and biocompatibility of the subject device to support product changes and provide the substantiation to the most current standards and regulations. The table below provides a sideby-side comparison of the subject device compared to the predicate.

Subject BD Pen Needles (K213478)Predicate BD Pen Needles (K162516)Comparison
ImageImage: Pen Needle with Outer Shield, Inner Shield, and Tear drop labelImage: Pen Needle with Outer Shield, Inner Shield, and Tear dropSame
General Information
Intended UseBecton Dickinson Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.Becton Dickinson Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.Same
Device Classification22Same
Product TypeHypodermic single lumen needleHypodermic single lumen needleSame
Product CodeFMIFMISame
Regulation Number880.5570880.5570Same
Single UseYESYESSame
Sterilization methodGamma irradiationGamma irradiationSame
SAL 10-6YESYESSame
Tamper Evident FeatureYES (tear-drop label)YES (tear-drop label)Same
Non-pyrogenic and Endotoxin FreeYESYESSame
Design
Subject BD Pen NeedlesPredicate BD Pen Needles (K162516)Comparison
Configurations29G x 12.7mm Thin Wall 3 bevel31G x 8mm Thin Wall 3 or 5 bevel31G x 5mm Thin Wall 3 or 5 bevel31G x 5mm Extra Thin Wall 3 bevel32G x 6mm Thin Wall 3 bevel (cleared under K182320- BD Contoured Pen Needle)32G x 4mm Thin Wall 3 bevel32G x 4mm Extra Thin Wall 5 bevel31G x 6mm Thin Wall 3 bevel32G x 4mm Thin Wall 3 bevel29G x 12.7mm Thin Wall 3 bevel31G x 8mm Thin Wall 3 or 5 bevel31G x 5mm Thin Wall 3 or 5 bevel31G x 5mm Extra Thin Wall 3 bevel32G x 6mm Thin Wall 3 bevel (cleared under K182320- BD Contoured Pen Needle)32G x 4mm Thin Wall 3 bevel32G x 4mm Extra Thin Wall 5 bevelDifferent See Comment # 1
Needle Gauge Size(s)29G, 31G, 32G29G, 31G, 32GSame
Cannula Length(s)4mm, 5mm,6mm, 8mm, and 12.7mm4mm, 5mm,8mm, and 12.7mmDifferent See Comment # 2
Wall ThicknessThin Wall or Extra Thin WallThin Wall or Extra Thin WallSame
Needle insertion methodManualManualSame
Angle of InsertionStraight 90 degreesStraight 90 degreesSame
Tip Geometry3 or 5 bevel3 or 5 bevelSame
Materials
ComponentSub-componentSubject BD Pen NeedlesPredicate BD Pen Needles (K162516)Comparison
HubPolypropylenePolypropyleneSame
Cannula/NeedleStainless Steel 304Stainless Steel 304Same
Pen NeedleHub/Cannula AdhesiveUV-cured Adhesive Dymax or LoctiteUV-cured Adhesive Dymax or LoctiteDifferent See Comment # 3
Lubricant (Patient End)Medical Grade SiliconeMedical Grade SiliconeSame
Lubricant (Non-Patient End)Medical Grade SiliconeMedical Grade SiliconeSame
Inner ShieldN/APolyethylene with or without Colorant (blue, purple, orange, pink or green) or Polypropylene with Colorant (green)Polyethylene with or without Colorant (blue, purple, pink or green) or Polypropylene with Colorant (green)Different See Comment # 4
Outer CoverN/APolyethylene or PolypropylenePolyethylene or PolypropyleneSame
Tear Drop Label and InkN/APaper with foil layer and inkPaper with foil layer and inkDifferent See Comment # 5

Substantial Equivalence Table

{7}------------------------------------------------

Comment 1: New configuration 31G x 6mm and 32G x 4mm 5 bevel Thin Wall was introduced to the BD pen needle family. New length was in range of previously cleared sizes under K162516. The differences between the predicate and the subject device do not raise any new questions of safety or effectiveness.

{8}------------------------------------------------

Comment 2: New configuration 31G x 6mm was introduced to the BD pen needle family. New length was in range of previously cleared sizes under K162516. The differences between the subject device do not raise any new questions of safety or effectiveness.

Comment 3: Loctite adhesive's curing agent sourced from a new supplier. The differences between the predicate and the subject device do not raise any new questions of safety or effectiveness.

Comment 4: New blue colorant formulation introduced for 8mm pen needle and orange colorant for 31G x 6mm (orange colorant was cleared under K182320- BD Contoured Pen Needle and under K190054 – BD Insulin Syringe Device). The differences between the predicate and the subject device do not raise any new questions of safety or effectiveness.

Comment 5: A new tear-drop label ink formulation has been introduced for the BD Pen Needle. The differences between the predicate and the subject device do not raise any new questions of safety or effectiveness.

Performance Testing

Non-clinical Testing

The subject device has the same technological characteristics as the predicate device cleared under K162516. BD has validated the design of the subject device as part of its design control process in accordance with the Quality System Regulation. This testing included functional performance per ISO 11608-2:2012, ISO 9626:2016, and BD internal test requirements. Material changes were evaluated in accordance with ISO 10993-1:2018: Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process.

Functional Performance Testing

  • · ISO 11608-2: 2012 Needle-based injection systems for medical use Requirements and test methods – Part 2: Needles
  • · ISO 9626: 2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
  • BD internal test requirements .

Biocompatibility Testing

The device was categorized as surface medical device contacting intact skin involving prolonged contact duration for components involving direct contact with intact skin externally communicating medical device; direct contact with tissue; prolonged contact duration. The following testing was conducted according to ISO 10993-1:2018 - Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process

  • . Cytotoxicity
  • . Sensitization
  • Intracutaneous reactivity

{9}------------------------------------------------

  • Irritation ●
  • Material Mediated Pyrogenicity ●
  • Genotoxicity
  • Implantation and Particulate Analysis .
  • Chemical Characterization .
  • o Toxicological Risk Assessment

Particulate matter Testing was conducted in accordance with USP<788> Particulate Matter in Injection and met the USP acceptance criteria.

Sterility, Shipping and Shelf-life

  • The BD Pen Needle is sterilized using a validated gamma irradiation sterilization method . by Cobalt 60. It has been validated via the method described in ISO 11137-2:2013 -Sterilization of Health Care Product- Radiation Part 2: Establishing the Sterilization Dose. The sterilization parameters were chosen to assure the sterilization dose provides a minimum SAL of 10-6.
  • Residuals are not applicable for the gamma irradiation sterilization method. ●
  • Limulus Amebocyte Lysate (LAL) assay was used to measure the endotoxin limit the ● requirement was met. The product is non-Pyrogenic.
  • . This device is packaged in an Outer Cover with a tamper evident peel-away tear drop label. The tear drop label is the sterility barrier of the medical device.
    • o Sterile barrier testing performed on the subject device:
      • .. Tear drop label removal force
      • . Seal integrity
  • Packaging Integrity Testing under simulated shipping conditions were conducted to . satisfy the requirements in ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems. All packaging deemed acceptable for protection of product and sterility maintenance.
  • Accelerated stability testing has been conducted to validate the sterility and . performance of the BD Pen Needle device to support the shelf-life of 5 years.

{10}------------------------------------------------

Conclusion:

The modifications of the design, material and labelling to the subject device compared to the predicate device in K162516 met the requirements of the standards. The differences between the predicate (K162516) and the subject device do not raise any new or different questions of safety or effectiveness. The BD Pen Needle is substantially equivalent to the predicate BD Pen Needle (K162516).

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).