K Number
K213411
Device Name
iSR’obot Mona Lisa 2.0
Date Cleared
2021-12-22

(64 days)

Product Code
Regulation Number
892.1560
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
iSR'obot Mona Lisa 2.0 is a user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a third-party needle or a probe, during image-guided diagnostic and interventional procedures in conjunction with the guidance of transrectal ultrasound involving the prostate gland in a clinical setting. Examples of such procedures include, but are not limited to, image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, and soft tissue ablations. The iSR'obot Mona Lisa 2.0 provides 2D and 3D visualization of Ultrasound images and the ability to fuse and register these images with those from other imaging modalities such as Magnetic Resonance, etc. It also provides the ability to display a simulated image of an insertion tool on a computer monitor screen, the target organ, and the current and projected future path of the insertion tool taking into account patient movement. Other software features include multiplanar reconstruction, segmentation, image measurements, 2D/3D image registration, and reporting.
Device Description
The iSR'obot Mona Lisa 2.0 is user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a needle or a probe, during image-guided diagnostic and interventional procedures involving the prostate gland in a clinical setting. The device displays the 2D live image feeds from commercially available ultrasound systems and constructs 3D ultrasound image stacks. The system allows the importation of an MRI image to create a model of patient prostate by providing fusion between Ultrasound and Magnetic Resonance Imaging (MRI). The system is compatible with commercially available ultrasound systems, transrectal ultrasound bi-plane probes, and commercially available needle devices. Other software features include multi-planar reconstruction, segmentation, image measurements and 2D/3D image registration.
More Information

No
The summary describes image processing, fusion, and guidance features, but does not mention AI or ML.

No
The device is described as an accessory to guide physicians in the planning and positioning of insertion tools, not for providing therapy itself. Its function is for imaging, guidance, and measurement during diagnostic and interventional procedures.

No.
The device is described as an accessory to guide physicians in planning and positioning insertion tools for diagnostic and interventional procedures, rather than performing the diagnosis itself. Its functions include visualization, image fusion, and displaying tool paths, which aid in diagnostic procedures but do not constitute a diagnostic function.

No

The device is described as a "stereotaxic accessory" intended to guide physicians in the "positioning of insertion tools." While it includes significant software features for image processing and guidance, the description of performance testing involving "accurately position[ing] the needle" and "adjustment of the generated prostate models and planned needle position to compensate for the patient movement" strongly implies the presence of hardware components that interact with the patient and insertion tools for physical guidance and positioning, which would not qualify it as a software-only device.

Based on the provided information, the iSR'obot Mona Lisa 2.0 is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
  • iSR'obot Mona Lisa 2.0 Function: The iSR'obot Mona Lisa 2.0 is a stereotaxic accessory used to guide physicians in the planning and positioning of insertion tools (like needles or probes) during image-guided procedures involving the prostate. It uses imaging modalities (Ultrasound and MRI) and software features to assist in these procedures.
  • No Sample Analysis: The device does not analyze any biological samples taken from the patient. Its function is to provide guidance and visualization during a procedure performed directly on the patient.

Therefore, while the device is used in diagnostic and interventional procedures, its role is as a surgical/procedural guidance system, not as a diagnostic test performed on a sample outside the body.

N/A

Intended Use / Indications for Use

iSR'obot Mona Lisa 2.0 is a user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a third-party needle or a probe, during image-guided diagnostic and interventional procedures in conjunction with the guidance of transrectal ultrasound involving the prostate gland in a clinical setting. Examples of such procedures include, but are not limited to, image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, and soft tissue ablations.

The iSR'obot Mona Lisa 2.0 provides 2D and 3D visualization of Ultrasound images and the ability to fuse and register these images with those from other imaging modalities such as Magnetic Resonance, etc. It also provides the ability to display a simulated image of an insertion tool on a computer monitor screen, the target organ, and the current and projected future path of the insertion tool taking into account patient movement. Other software features include multi-planar reconstruction, segmentation, image measurements, 2D/3D image registration, and reporting.

Product codes (comma separated list FDA assigned to the subject device)

IYO, ITX, LLZ, OIJ

Device Description

The iSR'obot Mona Lisa 2.0 is user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a needle or a probe, during image-guided diagnostic and interventional procedures involving the prostate gland in a clinical setting. The device displays the 2D live image feeds from commercially available ultrasound systems and constructs 3D ultrasound image stacks. The system allows the importation of an MRI image to create a model of patient prostate by providing fusion between Ultrasound and Magnetic Resonance Imaging (MRI). The system is compatible with commercially available ultrasound systems, transrectal ultrasound bi-plane probes, and commercially available needle devices. Other software features include multi-planar reconstruction, segmentation, image measurements and 2D/3D image registration.

Workstation

The workstation is connected to the ultrasound system via a standard cable and hardware connector and displays the 2D live image feed in a format that is compatible with the iSR 'obot Mona Lisa 2.0. Using the application software, the urologist or physician defines the apex and base limits of the prostate gland and the robotic arm moves the ultrasound probe within those limits to capture multiple 2D slices of the prostate gland to construct the 3D image stack. The urologist or physician may refine the constructed 3D image stack by indicating and confirming the planned lesion core or tumor location, anatomical markers within or around the prostate gland, and the prostate gland contour. In addition, the application software can utilize previously acquired images of the patient's prostate, which may include other image modalities like magnetic resonance images and register to this 3D image stack. The completed procedure information can be stored on a location selected by the urologist or physician, such as in the workstation or in a PACS server. The stored procedure information can be used for future review with the patient or examination. Previously stored biopsy procedural plan can be recalled and its 3D model may be aligned and registered to current 3D model of the prostate to facilitate planning of the interventional procedure.

Robotic Navigation Module

The robotic navigation module comprises of a robotic arm and bed rail stabilizer. The stabilizer is a mechanical device and is able to lock and release to position the robotic arm. One end of the stabilizer is first attached to the bed rail while the other end is used to mount the robotic arm so that the robotic arm is able to be positioned close to the patient's perineum while the patient is in a lithotomy position. The robotic arm is a motorized mechanical structure and has two key functions, which is to hold and move the ultrasound probe of a commercially available ultrasound system to display 2D live ultrasound image feeds and to orientate its needle guidance mechanism to facilitate insertion of a commercially available needle (based on a planned and simulated needle trajectory) by the urologist or physician. The movement of the ultrasound probe and the needle guidance mechanism is motorized. As a result, the workstation is able to construct and display a 3D image stack and rendered surface model of the prostate.

During the procedure, the real time 2D ultrasound image is visible on the iSR 'obot Mona Lisa 2.0 display. After the robotic arm has orientated its trajectory according to the planned location, the urologist or physician manually inserts the needle into the prostate via the needle guidance mechanism. The 2D live ultrasound image may be marked up to record the actual locations where the needle has been inserted. The system is able to pivot and facilitate re-insertion of the needle if the marked up actual location is distant from the planned location. The urologist or physician is able to observe prostate shift on live ultrasound image, re-register the prostate model and needle location plan to the live ultrasound prostate image.

Needle Guide Holder

The needle guide holder is designed to be used with the iSR obot Mona Lisa 2.0. It is a component installed on the robotic arm functioning as a channel for insertion tools to go through. The robotic arm has a motorized mechanism to close and open the needle guide holder in order to hold and release the insertion tools. The needle guide holder may be manufactured to fit insertion tools with varying diameters.

Disposables

The iSR 'obot Kit is intended to be used with the iSR'obot Mona Lisa 2.0 as a component of a system for performing transperineal prostate procedure. The kit consists of probe sheath, drape and biopsy papers. The probe sheath functions as a sheath to a third-party ultrasound probe for ultrasound scanning of the prostate organ.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Ultrasound, Magnetic Resonance Imaging (MRI)

Anatomical Site

prostate gland

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Physicians (urologist or physician) in a clinical setting.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-clinical System Performance Testing (system level testing using phantom)

  • Testing conducted to demonstrate that after the system has been calibrated, the system is able to accurately position the needle.
  • Testing conducted to demonstrate the adjustment of the generated prostate models and planned needle position to compensate for the patient movement during the procedure.
  • Testing conducted to demonstrate multi-needle positioning

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K203659

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K161109, K163502, K162474

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

Biobot Surgical Pte Ltd % Lim Yan Shin Regulatory Affairs 79 Ayer Rajah Crescent, #04-05 Singapore. Singapore 139955 SINGAPORE

December 22, 2021

Re: K213411

Trade/Device Name: iSR'obot Mona Lisa 2.0 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, ITX, LLZ, OIJ Dated: October 15, 2021 Received: October 19, 2021

Dear Lim Yan Shin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

1

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb, Ph.D. Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known)

K213411

Device Name iSR'obot Mona Lisa 2.0

Indications for Use (Describe)

iSR'obot Mona Lisa 2.0 is a user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a third-party needle or a probe, during image-guided diagnostic and interventional procedures in conjunction with the guidance of transrectal ultrasound involving the prostate gland in a clinical setting. Examples of such procedures include, but are not limited to, image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, and soft tissue ablations.

The iSR'obot Mona Lisa 2.0 provides 2D and 3D visualization of Ultrasound images and the ability to fuse and register these images with those from other imaging modalities such as Magnetic Resonance, etc. It also provides the ability to display a simulated image of an insertion tool on a computer monitor screen, the target organ, and the current and projected future path of the insertion tool taking into account patient movement. Other software multiplanar reconstruction, segmentation, image measurements, 2D/3D image registration, and reporting.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Image /page/3/Picture/1 description: The image shows the logo for Biobot. The word "biobot" is written in lowercase letters. The "bio" portion of the word is in orange, and the "bot" portion of the word is in blue. The "bio" portion of the word is also slightly larger than the "bot" portion of the word.

510(k) Summarv

K213411

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).

Date:October 15, 2021
Submitter:Biobot Surgical Pte Ltd
79 Ayer Rajah Crescent,
#04-05, Singapore 139955
Primary Contact:Lim Yan Shin, Regulatory Affairs
Biobot Surgical Pte Ltd
Cell: +65 97263992
E-mail: yanshin.lim@ziggroup.com.sg
Secondary Contact:Georgiann Keyport, Regulatory Consultant, U.S.
Email: gkeyport@canopyregulatory.com
Cell: 952-994-8267

Product Identification

Device Trade Name:iSR'obot Mona Lisa 2.0
Common / Usual Name:System, image processing, radiological
Classification Names:892.1560 Ultrasonic pulsed echo imaging system
892.1570 Diagnostic ultrasonic transducer
Product Code:IYO, ITX, LLZ, OIJ
Manufacturer / DesignBiobot Surgical Pte Ltd
Location:79 Ayer Rajah Crescent,
#04-05, Singapore 139955

Device Description

The iSR'obot Mona Lisa 2.0 is user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a needle or a probe, during image-guided diagnostic and interventional procedures involving the prostate gland in a clinical setting. The device displays the 2D live image feeds from commercially available ultrasound systems and constructs 3D ultrasound image stacks. The system allows the importation of an MRI image to create a model of patient prostate by providing fusion between Ultrasound and Magnetic Resonance Imaging (MRI). The system is compatible with commercially available ultrasound systems, transrectal ultrasound bi-

4

Image /page/4/Picture/1 description: The image shows the logo for Biobot. The word "biobot" is written in blue, with the "b" and "i" connected. Above the word is an orange oval shape, with a small dot on the upper left side. The logo is simple and modern.

plane probes, and commercially available needle devices. Other software features include multi-planar reconstruction, segmentation, image measurements and 2D/3D image registration.

Workstation

The workstation is connected to the ultrasound system via a standard cable and hardware connector and displays the 2D live image feed in a format that is compatible with the iSR 'obot Mona Lisa 2.0. Using the application software, the urologist or physician defines the apex and base limits of the prostate gland and the robotic arm moves the ultrasound probe within those limits to capture multiple 2D slices of the prostate gland to construct the 3D image stack. The urologist or physician may refine the constructed 3D image stack by indicating and confirming the planned lesion core or tumor location, anatomical markers within or around the prostate gland, and the prostate gland contour. In addition, the application software can utilize previously acquired images of the patient's prostate, which may include other image modalities like magnetic resonance images and register to this 3D image stack. The completed procedure information can be stored on a location selected by the urologist or physician, such as in the workstation or in a PACS server. The stored procedure information can be used for future review with the patient or examination. Previously stored biopsy procedural plan can be recalled and its 3D model may be aligned and registered to current 3D model of the prostate to facilitate planning of the interventional procedure.

Robotic Navigation Module

The robotic navigation module comprises of a robotic arm and bed rail stabilizer. The stabilizer is a mechanical device and is able to lock and release to position the robotic arm. One end of the stabilizer is first attached to the bed rail while the other end is used to mount the robotic arm so that the robotic arm is able to be positioned close to the patient's perineum while the patient is in a lithotomy position. The robotic arm is a motorized mechanical structure and has two key functions, which is to hold and move the ultrasound probe of a commercially available ultrasound system to display 2D live ultrasound image feeds and to orientate its needle guidance mechanism to facilitate insertion of a commercially available needle (based on a planned and simulated needle trajectory) by the urologist or physician. The movement of the ultrasound probe and the needle guidance mechanism is motorized. As a result, the workstation is able to construct and display a 3D image stack and rendered surface model of the prostate.

5

Image /page/5/Picture/1 description: The image shows the logo for Biobot Analytics. The logo features the word "biobot" in a sans-serif font. The "bio" portion of the word is in a gradient orange color, while the "bot" portion is in blue. The logo has a modern and clean design.

During the procedure, the real time 2D ultrasound image is visible on the iSR 'obot Mona Lisa 2.0 display. After the robotic arm has orientated its trajectory according to the planned location, the urologist or physician manually inserts the needle into the prostate via the needle guidance mechanism. The 2D live ultrasound image may be marked up to record the actual locations where the needle has been inserted. The system is able to pivot and facilitate re-insertion of the needle if the marked up actual location is distant from the planned location. The urologist or physician is able to observe prostate shift on live ultrasound image, re-register the prostate model and needle location plan to the live ultrasound prostate image.

Needle Guide Holder

The needle guide holder is designed to be used with the iSR obot Mona Lisa 2.0. It is a component installed on the robotic arm functioning as a channel for insertion tools to go through. The robotic arm has a motorized mechanism to close and open the needle guide holder in order to hold and release the insertion tools. The needle guide holder may be manufactured to fit insertion tools with varying diameters.

Disposables

The iSR 'obot Kit is intended to be used with the iSR'obot Mona Lisa 2.0 as a component of a system for performing transperineal prostate procedure. The kit consists of probe sheath, drape and biopsy papers. The probe sheath functions as a sheath to a third-party ultrasound probe for ultrasound scanning of the prostate organ.

Predicate Device NamePredicate 510(k) Submission Reference
iSR'obot Mona Lisa 1.0K203659 (Primary)
iSR'obot MRI-US FusionK161109 (Reference)
iSR'obot Biopsy KitK163502 (Reference)
ArtemisK162474 (Reference)

Predicate Device Information and Comparison

Intended Use / Indications for Use

iSR'obot Mona Lisa 2.0 is a user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a third-party needle or a probe, during image-guided diagnostic and interventional procedures in conjunction with the guidance of transrectal ultrasound involving the prostate gland in a clinical setting.

6

Image /page/6/Picture/1 description: The image shows the logo for Biobot. The word "biobot" is written in a sans-serif font, with "bio" in a bright orange color and "bot" in a dark blue color. The "bio" portion of the logo is set against a larger orange oval shape, giving the impression of a stylized sun or planet.

Examples of such procedures include, but are not limited to, image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, and soft tissue ablations.

The iSR 'obot Mona Lisa 2.0 provides 2D and 3D visualization of Ultrasound images and the ability to fuse and register these images with those from other imaging modalities such as Magnetic Resonance, etc. It also provides the ability to display a simulated image of an insertion tool on a computer monitor screen, the target organ, and the current and projected future path of the insertion tool taking into account patient movement. Other software features include multi-planar reconstruction, segmentation, image measurements, 2D/3D image registration, and reporting.

Technology Characteristics Compared to Predicate Devices

iSR'obot Mona Lisa 2.0 employs the same fundamental scientific technology (design, function and specifications) as that of its legally marketed predicate device, iSR'obot Mona Lisa 1.0 (K203659).

Similarities in technology characteristics include:

  • Platform-hosted motorized devices and are able to provide 2D and 3D views of . the prostate gland;
  • Use the same technology to acquire a transrectal ultrasound image to plan and . guide a needle for diagnostic procedure;
  • Control the trajectory and depth for needle placement via needle guiding • mechanism;
  • . Fusion of Magnetic Resonance and Ultrasound images for mapping planning information of the prostate gland
  • . Re-positioning of subsequent needle insertion for the same target lesion after the user has indicated the actual needle landing position from the prior insertion.

Modifications to iSR'obot Mona Lisa 2.0 include:

  • The iSR'obot Mona Lisa 2.0 is intended for image-guided diagnostic and . interventional procedures while the legally marketed predicate device is intended only for image-guided diagnostic procedure;
  • . Visualization of simulated images in an interventional procedure of 1) the lesions and their margins and 2) treatment zone;
  • . Needle guide holder, which acts as a channel for insertion tool positioning, is detachable and can be cleaned and sterilized for subsequent use, and enables multi-needle insertion trajectory planning for interventional procedures;

7

Image /page/7/Picture/1 description: The image shows the logo for Biobot Analytics. The logo features the word "biobot" in a stylized font. The "bio" portion of the word is in orange, while the "bot" portion is in blue. The "o" in "bio" is shaped like an oval.

  • . Patient movement adjustment function for re-alignment of the prostate model with the live-ultrasound image in the event that prostate shifts due to patient movement during a procedure;
  • . Connectivity to picture archiving and communication system (PACS) to download DICOM images and download/upload of procedure case; and
  • . Provide an offline post-procedure review

Other changes include minor user interface variations such as graphics user interface (GUI) design. These differences do not significantly affect the function or use of the device, nor do they raise new or additional safety risks.

These changes are being implemented as a product improvement effort and not due to a corrective action or field action.

iSR'obot Mona Lisa 2.0 has the following technological characteristics, similar to the Artemis (K162474):

  • . Same intended use for insertion tool planning and navigation for ultrasound imageguided prostate diagnostic and interventional procedures.
  • . Connecting to standard ultrasound system to display and acquire 2D live ultrasound images of the prostate gland
  • Perform other viewing and imaging-processing function such as image segmentation, image multi planar reconstruction, image overlays, and measurements
  • . Image registration and elastic fusion of ultrasound image with previously created MRI prostate model and/or prior biopsy plan on application software
  • . Generate 2D/3D model of a prostate model and trajectory of insertion tools
  • . Plan insertion tool positions on the fused prostate model by physicians and mark the actual position of insertion tools
  • . Connectivity to PACS to download DICOM images and download/upload of procedure case
  • Artemis has a needle guide mechanism attachment similar to the iSR'obot Mona . Lisa 2.0. which allows insertion of third-party insertion tools

The differences between iSR'obot Mona Lisa 2.0 and Artemis (K162474) are that:

  • . Anatomical Access - iSR'obot Mona Lisa 2.0 is intended for insertion tools positioned transperineally while Artemis is intended for insertion tools positioned transperineally and transrectally procedure.

8

Image /page/8/Picture/1 description: The image shows the logo for Biobot. The logo consists of the word "biobot" in blue, with the "bio" portion of the word in a larger font size than the "bot" portion. To the left of the word "biobot" is an orange oval shape that is slightly tilted to the left.

  • . Positioning of insertion tools - iSR'obot Mona Lisa 2.0 has an 8 axis motorized robotic arm that navigates the needle guide holder based on the clinician's insertion tool position plan. The Artemis has a semi-robotic arm (mechanical arm with encoders) that clinicians may control to track and visualize the insertion tool as it is being positioned. Despite the slight difference, the clinician is in control throughout and manually inserts the insertion tool for both devices.
  • Adjustment of prostate model in the event of patient movement iSR'obot ● Mona Lisa 2.0 provides an adjustment panel for users to manually adjust the prostate model to align with the ultrasound live image of the shifted prostate. Artemis tracks the position and orientation of the ultrasound probe, and users mark landmark features on the ultrasound live image to enable its software to automatically adjust the prostate model to align with the ultrasound live image of the shifted prostate.
  • Information for active surveillance - Artemis is intended to be used for patients in active surveillance to keep track of previous procedures information and outcomes. iSR'obot Mona Lisa 2.0 allows previous procedures information to be accessed for patients under active surveillance, but the histopathological outcome is not currently captured in the software.

The technological characteristic above are features that assists physicians to plan and insert an insertion tool. The physician has full control of the tasks to be performed. According to the risk analysis, non-clinical performance tests, device verification and validation tests, the above differences do not result in additional or increase to existing known risks for iSR'obot Mona Lisa 2.0.

Substantial Equivalence

The technological characteristics such as intended use, method of operation, general function and application of the iSR'obot Mona Lisa 2.0 are equivalent to the primary predicate iSR'obot Mona Lisa 1.0 (K203659),

iSR'obot Mona Lisa 1.0 (K203659), iSR'obot MRI-US Fusion (K161109) and iSR'obot Biopsy Kit (K163502) shall be collectively known as iSR 'obot Mona Lisa 1.0 System.

Risk analysis was conducted to evaluate the modifications and features update in iSR'obot Mona Lisa 2.0. All verification and validation activities were performed and results demonstrated substantial equivalence. Table 1 indicates the comparison between the subject device and predicate devices.

9

Image /page/9/Picture/1 description: The image shows the word "biobot" in a stylized font. The word is written in lowercase letters. The "bio" portion of the word is colored in orange, while the "bot" portion is in blue. The orange color appears to fade into the blue color.

| Table 1: Comparison Between Predicate Devices and Subject Device iSR'obot Mona Lisa

2.0
Technological
CharacteristicReference Device:Predicate Device:Submitted Device:
Eigen's Artemis
(K162474)iSR'obot Mona Lisa
1.0 System (Primary -
K203659, Reference -
K161109, K163502)iSR'obot Mona Lisa
2.0
Intended UseArtemis along with the
Needle Guide
Attachment is used for
image-guided
interventional and
diagnostic procedures
of the prostate gland. It
provides 2D and 3D
visualization of
Ultrasound (US)
images and the ability
to fuse and register
these images with those
from other imaging
modalities such as
Ultrasound, Magnetic
Resonance, Computed
Tomography, etc. It
also provides the ability
to display a simulated
image of a tracked
insertion tool such as a
biopsy needle,
guidewire or probe on a
computer monitor
screen that shows
images of the target
organ and the current
and the projected future
path of the
interventional
instrument taking into
account patient
movement. The
software also provides
a virtual grid on the live
ultrasound foriSR'obot Mona Lisa 1.0
is intended for use by
trained urologist or
physician to perform
the computer-assisted
transperineal prostate
biopsy under transrectal
ultrasound guidance.
The device serves as a
biopsy needle guide
only. It shall be used in
conjunction with a third
party ultrasound
machine and endorectal
probe that supports B-
Mode, and a third party
prostate biopsy gun and
needle. The insertion of
biopsy needle will be
done by urologist. The
patient is administered
general anesthesia and
placed in a lithotomy
position.iSR'obot Mona Lisa
2.0 is a user-controlled,
stereotaxic accessory
intended to guide
physicians in the
planning and
positioning of insertion
tools, such as a third-
party needle or a probe,
during image-guided
diagnostic and
interventional
procedures in
conjunction with the
guidance of transrectal
ultrasound involving
the prostate gland in a
clinical setting.
Examples of such
procedures include, but
are not limited to,
image fusion for
diagnostic clinical
examinations and
procedures, soft tissue
biopsies, and soft tissue
ablations.
The iSR'obot Mona
Lisa 2.0 provides 2D
and 3D visualization of
Ultrasound images and
the ability to fuse and
register these images
with those from other
imaging modalities
such as Magnetic
performing systematic
sampling of the target
organ. Other software
features include patient
data management,
multi-planar
reconstruction,
segmentation, image
measurements, 2D/3D
image registration,
reporting, and
pathology management.
Artemis is intended for
treatment planning and
guidance for clinical,
interventional and/or
diagnostic procedures.
The device is intended
to be used in
interventional and
diagnostic procedures
in a clinical setting.
Example procedures
include, but are not
limited to image fusion
for diagnostic clinical
examinations and
procedures, soft tissue
biopsies, soft tissue
ablations and
placement of fiducial
markers. Artemis is
also intended to be used
for patients in active
surveillance to keep
track of previous
procedures information
and outcomes.Resonance, etc. It also
provides the ability to
display a simulated
image of an insertion
tool on a computer
monitor screen, the
target organ, and the
current and projected
future path of the
insertion tool taking
into account patient
movement. Other
software features
include multi-planar
reconstruction,
segmentation, image
measurements, 2D/3D
image registration, and
reporting.
Product CodeLLZIYO, ITX, LLZ, OIJIYO, ITX, LLZ, OIJ
ClassIIIIII
Target
AnatomyProstateProstateProstate
Anatomy
AccessTransperineal
TransrectalTransperineal
Not availableTransperineal
Not available
Patient
populationPatients for a biopsy
procedurePatients for a biopsy
procedurePatients for a biopsy
procedure
Patients for an
interventional
procedurePatients for an
interventional
procedure
Patients in active
surveillance
Clinical UtilitySoft tissue biopsies,Soft tissue biopsiesSoft tissue biopsies
Soft tissue ablationsSoft tissue ablations
Placement of fiducial
markers
Pathology management
Patient data
management
Software
Window OSYesYesYes
Medical
Imaging
SoftwareYesYesYes
Compliance
with FDA
CybersecurityYesYesYes (Updated)
Image Display
Multi-Modality
SupportYesYesYes
General Image
2D/3D ReviewYesYesYes
3D Rendering
ViewYesYesYes
Live 2D
UltrasoundYesYesYes
Image Processing

10

Image /page/10/Picture/1 description: The image shows the logo for Biobot Analytics. The logo consists of the word "biobot" in a sans-serif font, with the "bio" portion in a vibrant orange color and the "bot" portion in a darker blue color. To the left of the word "biobot" is an orange oval shape, which appears to be a stylized representation of a droplet or a cell.

11

Biobot Surgical Pte. Ltd

Image /page/11/Picture/1 description: The image shows the word "biobot" in a stylized font. The word is written in a blue color, and there is an orange oval shape behind the word. The oval shape is positioned so that it appears to be behind and slightly to the left of the word.

12

Biobot Surgical Pte. Ltd

Image /page/12/Picture/1 description: The image shows the logo for Biobot Analytics. The logo features the word "biobot" in a stylized font. The "bio" portion of the word is in orange, while the "bot" portion is in blue. There is also an orange oval shape behind the word "bio".

| Gland

SegmentationYesYesYes
Image
RegistrationYesYesYes
Rigid
RegistrationYesYesYes
Elastic
RegistrationYesYesYes
Multi-Planar
ReformattingYesYesYes
Connectivity
DICOM
Import/ExportYesNot availableYes (updated)
Review Tools
Standard Image
Viewing ToolsYesYesYes
Measurement
ToolsYesYesYes
Annotation
ToolsYesYesYes
Segmentation
ToolsYesYesYes
Reporting
ToolsYesYesYes
Image OverlaysYesYesYes (updated)
Post Procedure
ReviewYesNot availableYes
Planning & Navigation
Import Prior
PlansYesYesYes
Import/Add
TargetsYesYesYes
Plan / Mark
LocationsYesYesYes (updated)

13

Image /page/13/Picture/1 description: The image shows the logo for Biobot. The logo consists of the word "biobot" in blue, with the "bio" part of the word in a slightly darker shade of blue. To the left of the word is an orange oval shape, with a darker orange dot inside of it. The oval shape is positioned so that it appears to be behind the word "biobot".

| Navigation

TypeMechanicalElectromechanicalElectromechanical
Third-Party Devices Compatibility
Insertion Tools
(e.g needles
and probes)YesYesYes
Ultrasound
SystemsYesYesYes
Components
Hardware• Workstation
• Mechanical Arm for
positioning of
insertion tool• Workstation
• Mechanical Arm for
positioning of
insertion tool
• Stabilizer• Workstation
• Mechanical Arm for
positioning of
insertion tool
• Stabilizer
DisposablesNot availableProprietary DisposablesProprietary Disposables

Safety and Effectiveness

The labelling contains instructions for use and necessary cautions, warnings and notes to provide for safe and effective use of the device. Risk management is ensured via Biobot Surgical's design control procedures and application of risk management procedure, which is used to identify and mitigate potential hazards. These potential hazards are controlled through the product development process, verification and validation testing, systematic clinical literature review and clinical effects analysis (CEA) to ensure safe profile of iSR'obot Mona Lisa 2.0.

Nonclinical Testing and Performance Information

Biobot performed the following testing to ensure safety and effectiveness of iSR'obot Mona Lisa 2.0:

  • Design and System Verification To ensure iSR'obot Mona Lisa 2.0 meets the . specifications and its intended purpose
  • . Software Verification and Validation - To ensure the all software meets the specifications and the intended purpose. Software life cycle process is in accordance with IEC 62304:2006+A1:2015 Ed 1.1
  • Usability Testing Testing conducted in accordance with IEC 623661: . 2015+AMD1: 2020 Ed 1.1 and FDA guidance Applying Human Factors and Usability Engineering to Medical Devices

14

Biobot Surqical Pte. Ltd

Image /page/14/Picture/1 description: The image shows the logo for Biobot Analytics. The logo features the word "biobot" in a stylized font. The "bio" portion of the word is in a bright orange color and is slightly larger than the "bot" portion, which is in blue. The orange section has a gradient effect, making it appear three-dimensional.

  • . Cybersecurity Testing - Testing conducted in accordance with ANSI UL 2900-1 and ANSI UL 2900-2-1
  • . Non-clinical System Performance Testing (system level testing using phantom)
    • Testing conducted to demonstrate that after the system has been calibrated, the O system is able to accurately position the needle.
    • Testing conducted to demonstrate the adjustment of the generated prostate models o and planned needle position to compensate for the patient movement during the procedure.
    • Testing conducted to demonstrate multi-needle positioning о
  • Biocompatibility testing - Testing conducted to ensure biocompatibility of Needle Guide Holder.
    • ISO 10993-1:2018 O
    • ISO 10993-5: 2008 O
    • o ISO 10993-10: 2010
    • ISO 10993-11: 2017 O
  • . Reprocessing Validation - to validate the cleaning and sterilization procedure for Needle Guide Holder. Testing is conducted in accordance with AAMI TIR12:2010 and AAMI TIR30:2011/ (R)2016.
  • . Other Non-Clinical Tests
    • IEC 60601-1:2005 Ed 3.1 o
    • IEC 60601-1-2:2014 Ed 4 O
    • IEC 60601-1-6:2013 Ed 3.1 O
    • ISO 14971:2019 O
    • ISTA 3A 2018 O
    • ISTA 3B 2017 O
    • AAMI TIR28 o

Clinical Information

There is no pre-market clinical investigation submitted for iSR'obot Mona Lisa 2.0.

Conclusion

Comparison of the intended use, indications for use, technological characteristics, and performance specifications demonstrate the functional equivalence of iSR'obot Mona Lisa 2.0 to the predicate device iSR'obot Mona Lisa 1.0.

Based on the conformance to standards, development under Biobot's quality system, and the successful verification and non-clinical testing, iSR'obot Mona Lisa 2.0 does not raise any new safety and/or effectiveness concerns. Biobot believes that the iSR'obot Mona

15

Biobot Surgical Pte. Ltd

100

Image /page/15/Picture/1 description: The image shows the logo for Biobot Analytics. The logo features the word "biobot" in a stylized, lowercase font. The "bio" portion of the word is in orange, while the "bot" portion is in blue. The orange section appears to be a gradient, with a lighter shade at the top and a darker shade at the bottom.

Lisa 2.0 is safe and effective, and performs in a substantially equivalent manner to the predicate device iSR'obot Mona Lisa 1.0.