(85 days)
Biomet Microfixation Twist Drills are intended for drilling holes in large and small bone during orthopedic, spinal, neurosurgical, medial sternotomy, and oral and maxillofacial procedures.
Biomet Microfixation manufactures and distributes a variety of single use twist drills to aid in the implantation of Biomet Microfixation implants. Biomet Microfixation Twist Drills are intended for drilling holes in large and small bone during orthopedic, spinal, neurosurgical, medial sternotomy, and oral and maxillofacial procedures. The drills are manufactured from Stainless Steel. Some drills contain flutes along the majority of the drill length, while others contain a "stop" feature which prevents further drilling past a designated length. Additionally, some drills contain an "adjustable stop" feature which allows to surgeon to set the stop length intraoperatively. Twist Drills are distributed non-sterile and are intended for single-patient use. Cleaning should only be performed on new or uncompromised drills. The drills should be steam sterilized by the health-care facility prior to use.
This FDA 510(k) summary is for a medical device (Twist Drills), not an AI/ML algorithm. Therefore, many of the requested categories related to algorithm performance, training data, and expert review are not applicable.
Here's an analysis of the provided document based on the available information:
This document describes the FDA's decision to clear the Biomet Microfixation Twist Drills (K213208) as substantially equivalent to a predicate device (Twist Drills K062842).
1. A table of acceptance criteria and the reported device performance
The document states that "Design verification and validation testing were performed on the subject devices" and "The results from all design verification and validation tests confirmed that the subject devices met the predetermined acceptance criteria." However, the specific, quantitative acceptance criteria themselves are not provided in this summary document. The performance is reported as meeting these unspecified criteria.
Acceptance Criteria | Reported Device Performance |
---|---|
Specific acceptance criteria are not detailed in this summary document. | "The results from all design verification and validation tests confirmed that the subject devices met the predetermined acceptance criteria." |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
This document discusses performance testing for a physical medical device, not a data-driven AI/ML algorithm. Therefore, terms like "test set" and "data provenance" in the context of data analysis are not directly applicable. The document refers to "design verification and validation testing," which would involve physical samples of the drill. The sample size for these physical tests is not specified in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This is not applicable as the document describes a physical medical device, not an AI/ML diagnostic or prognostic tool that requires expert-established ground truth.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This is not applicable for a physical medical device without a diagnostic component.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable as the document describes a physical medical device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable as the document describes a physical medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For a physical medical device like a drill, "ground truth" generally refers to objective physical and mechanical properties. The validation would typically involve comparing the device's performance against predefined engineering specifications and safety standards, rather than expert consensus on diagnostic images or pathology. The document generally refers to "predetermined acceptance criteria" and "design verification and validation tests."
8. The sample size for the training set
This is not applicable as the document describes a physical medical device, not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
This is not applicable as the document describes a physical medical device.
§ 882.4310 Powered simple cranial drills, burrs, trephines, and their accessories.
(a)
Identification. Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.(b)
Classification. Class II (performance standards).