(84 days)
Sheath with integrated stopcocks are used in conjunction with a hysteroscope to permit direct viewing of the cervical canal and the uterine cavity to perform diagnostic and surgical procedures.
The subject device is a reusable sheath (Part WA4777A) and stopcock (Part WA47778A), provided non-sterile and labeled for reprocessing via cleaning and steam sterilization. The sheath is a rigid instrument made from stainless steel. The sheath has a working length of 206.7 ± 0.15 mm. The sheath has a single lumen shaped to allow for both, insertion of one 3 mm hysteroscope as well as a 5 Fr instrument channel for the instrument and irrigation inflow. There are three ports, one for irrigation, one for instruments, and one for the hysteroscope. There is no outflow channel for the irrigation fluid; the irrigation fluid flows out between the sheath and the cervix channel. A hysteroscope can be inserted into the hysteroscope channel of the sheath from the proximal end to provide an endoscopic image during the procedure. The endoscopic image can be viewed using the ocular or by connecting a compatible camera head. Compatible surgical instruments can be inserted into the instrument channel of the sheath. The irrigation stopcock made from stainless steel and polyether ether ketone (PEEK) can be used to control the inflow of the irrigation fluid. The instrument stopcock controls the passage of the surgical instruments.
The provided text describes the 510(k) premarket notification for a "Sheath and Stopcock Accessory" and does not contain information about an AI/ML powered device, therefore the request cannot be fully answered. This document purely focuses on a traditional medical device, specifically a reusable sheath and stopcock for hysteroscopes.
However, based on the information provided regarding the "Sheath and Stopcock Accessory," here's a breakdown of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria Category | Specific Criteria/Tests | Reported Device Performance |
|---|---|---|
| Biocompatibility | Cytotoxicity (ISO 10993-5:2009) | Non-cytotoxic |
| Sensitization (ISO 10993-10:2010) | Non-sensitizing | |
| Intracutaneous Irritation (ISO 10993-10:2010) | Non-irritating | |
| Mechanical Performance | Mechanical resistance of inflow interface, scope interface, instrument interface | Demonstrated comparable mechanical performance |
| Resistance of sheath tube | Demonstrated comparable mechanical performance | |
| Resistance of supporting ring of optic guide plate | Demonstrated comparable mechanical performance | |
| Resistance against repeated assembly/disassembly | Demonstrated comparable mechanical performance | |
| Irrigation inflow and outflow (comparable flow during operation) | Bench tests demonstrated comparable flow could be achieved during operation | |
| Electrical & Thermal Safety | Heat hazard assessment (AAMI/ANSI ES60601-1:2005/(R)2012 and updates) | Evaluated to demonstrate not a heat hazard, compatible with energized devices |
| Safety with electrocautery devices (AAMI/ANSI IEC 60601-2-2:2017) | Evaluated to demonstrate not a heat hazard, compatible with energized devices | |
| Basic endoscopic safety (IEC 60601-2-18:2009) | Evaluated to demonstrate not a heat hazard, compatible with energized devices | |
| Reprocessing/Sterilization | Reprocessing instructions and method validation (FDA guidance, March 17, 2015) | Validated and documentation provided |
2. Sample size used for the test set and the data provenance
The document does not specify exact sample sizes for each test set. It mentions "comparative validation testing" for mechanical performance and "risk analysis" for electrical/thermal safety, and "method validation testing" for reprocessing. The provenance of the data is from internal testing conducted by Olympus Winter & Ibe GmbH. No country of origin for the data is explicitly mentioned, but the manufacturer is based in Hamburg, Germany. The studies are assumed to be prospective experiments or tests designed specifically for this regulatory submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided because the device is a physical medical instrument, not an AI/ML diagnostic tool requiring human expert interpretation for ground truth. The "ground truth" for this device is established through objective physical, chemical, and electrical measurements and adherence to recognized standards.
4. Adjudication method for the test set
Not applicable, as this is not an interpretive diagnostic device requiring adjudication of human-labeled data.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device, not an AI/ML system, so no MRMC study or AI assistance evaluation was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a medical device, not an AI/ML system.
7. The type of ground truth used
The ground truth for this device's performance evaluation is based on objective physical measurements, chemical analyses (for biocompatibility), and adherence to established international and national standards (ISO, AAMI/ANSI, IEC, FDA guidance documents). For example, "non-cytotoxic" is a measurable outcome, and "comparable flow" is determined by direct measurement during bench testing.
8. The sample size for the training set
Not applicable. There is no "training set" as this is not an AI/ML powered device.
9. How the ground truth for the training set was established
Not applicable. There is no "training set" as this is not an AI/ML powered device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 22, 2021
Olympus Winter & Ibe GmbH % Jonathan Gilbert Regulatory Affairs Consultant for OCA Olympus Corporation of the Americas (OCA) 3500 Corporate Parkway Center Valley, PA 18034
Re: K213207
Trade/Device Name: Sheath and Stopcock Accessory Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH Dated: September 28, 2021 Received: September 29, 2021
Dear Jonathan Gilbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213207
Device Name Sheath and Stopcock Accessory
Indications for Use (Describe)
Sheath with integrated stopcocks are used in conjunction with a hysteroscope to permit direct viewing of the cervical canal and the uterine cavity to perform diagnostic and surgical procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
- 1 General Information
- Olympus Winter & Ibe GmbH Applicant Kuehnstr. 61 * 22045 Hamburg * Germany Establishment Reqistration No: 9610773
- 트 Official Correspondent Jon Gilbert fbo Christina Flores Manager, Regulatory Affairs 800 West Park Drive Westborough, MA 01581 Email: Christina.Flores@olympus-osta.com Phone: (508) 808-3341 Establishment Registration No.: 3003790304
- I Manufacturer Olympus Winter and Ibe GmbH Kuehnstr. 61 * 22045 Hamburg * Germany Establishment Registration No.: 9610773
2 Device Identification
| • | Device Trade Name: | Sheath and Stopcock Accessory |
|---|---|---|
| --- | -------------------- | ------------------------------- |
- Common Name: Hysteroscope (And Accessories)
- 트 Regulation Number: 21 CFR 884.1690
- Regulation Name: Hysteroscope and accessories ll
- Regulatory Class:
■
- I Product Code: HIH (Hysteroscope and accessories)
{4}------------------------------------------------
3 Predicate Device
The Sheath and stopcock accessory is considered substantially equivalent to:
| Table 1: Predicate device | ||
|---|---|---|
| Predicate device | Manufacturer | 510(k) No |
| TROCARS, CANNULAE, | Karl Storz | K943713 |
| SHEATHS, OBTURATORS | Endoskop GmbH |
|---|---|
| --------------------- | --------------- |
The predicate device has not been subject to a design-related recall.
4 Device Description
The subject device is a reusable sheath (Part WA4777A) and stopcock (Part WA47778A), provided non-sterile and labeled for reprocessing via cleaning and steam sterilization.
The sheath is a rigid instrument made from stainless steel. The sheath has a working length of 206.7 ± 0.15 mm. The sheath has a single lumen shaped to allow for both, insertion of one 3 mm hysteroscope as well as a 5 Fr instrument channel for the instrument and irrigation inflow. There are three ports, one for irrigation, one for instruments, and one for the hysteroscope. There is no outflow channel for the irrigation fluid; the irrigation fluid flows out between the sheath and the cervix channel. A hysteroscope can be inserted into the hysteroscope channel of the sheath from the proximal end to provide an endoscopic image during the procedure. The endoscopic image can be viewed using the ocular or by connecting a compatible camera head. Compatible surgical instruments can be inserted into the instrument channel of the sheath. The irrigation stopcock made from stainless steel and polyether ether ketone (PEEK) can be used to control the inflow of the irrigation fluid. The instrument stopcock controls the passage of the surgical instruments.
- Indications for Use 5
The subject device has the following indications for use:
Sheath with integrated stopcocks are used in conjunction with a hysteroscope to permit direct viewing of the cervical canal and the uterine cavity to perform diagnostic and surgical procedures.
The subject and predicate device have the same intended use, to allow medical equipment to be delivered to procedural sites. There are no intended use concerns with the subject device.
-
Comparison of Technological Characteristics 6
The technological characteristics of the subject device are considered substantially equivalent to the predicate and reference device. -
6.1 Comparison of subject and predicate device
The subject and predicate device share the following characteristics: -
Same general indications for use
{5}------------------------------------------------
- . Similar cross-sectional and shaft design
- Same patient contacting materials ●
- Comparative testing showed that the subject device has equivalent performance ● in respect to the irrigation fluid outflow
The following differences to the predicate device exist:
- Subject device does not incorporate an inner sheath ●
- The working length and lumen diameters are different .
- Subject device has no irrigation holes at distal end, since irrigation fluid outflow is . not inside the shaft between inner and outer sheath
As stated above, the subject and predicate devices have similar design characteristics and show comparable performance. The differences in technological characteristics do not raise different questions of safety and effectiveness.
7 Performance Data
The following performance data was provided to support a determination of substantial equivalence.
7.1 Biocompatibility testing
The patient contacting materials of the sheath and stopcock have been tested for biocompatibility in compliance with the FDA quidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", dated September 2020:
- Cytotoxicity according to ISO 10993-5:2009; -
- Sensitization according to ISO 10993-10:2010; and -
- -Intracutaneous Irritation according to ISO 10993-10:2010
The results showed that the subject device was non-cytotoxic, non-sensitizing, and nonirritating.
7.2 Performance Testing Bench
To demonstrate substantial equivalence and address mechanical performance as recommended in the FDA guidance document "Hysteroscopes and Gynecologic Laparoscopes", dated March 7, 1996, mechanical resistance of the inflow interface, scope interface and instrument interface, resistance of the sheath tube, resistance of the supporting ring of the optic quide plate, resistance against repeated assembly/disassembly, and irrigation inflow and outflow were compared during comparative validation testing.
These bench tests demonstrated that comparable flow could be achieved during operation and addressed the mechanical performance recommendations outlined in the guidance.
7.3 Electrical and Thermal Safety
A risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971:2007.
{6}------------------------------------------------
The subject device was evaluated for thermal and electrical safety, to demonstrate the subject device is not a heat hazard and compatible with energized devices, according to the following standards:
- Heat hazard assessment per AAMI/ANSI ES60601-1:2005/(R)2012 and -A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012;
- Safety with electrocautery devices per AAMI/ANSI IEC 60601-2-2:2017; and -
- Basic endoscopic safety per IEC 60601-2-18:2009 -
7.4 Reprocessing/ Sterilization
Reprocessing instructions and method validation testing for the subject device was conducted and documentation provided as recommended by the FDA guidance document "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling", dated March 17, 2015.
- 8 Conclusion
The performance data summarized above demonstrate that the subject device is as safe and effective as the predicate device. The subject device is substantially equivalent to the predicate device.
§ 884.1690 Hysteroscope and accessories.
(a)
Identification. A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for hysteroscope accessories that are not part of a specialized instrument or device delivery system; do not have adapters, connectors, channels, or do not have portals for electrosurgical, laser, or other power sources. Such hysteroscope accessory instruments include: lens cleaning brush, cannula (without trocar or valves), clamp/hemostat/grasper, curette, instrument guide, forceps, dissector, mechanical (noninflatable), and scissors. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.