K Number
K213189
Date Cleared
2022-03-01

(153 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Wrist-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age.

Device Description

The Wrist-type Fully Automatic Digital Blood Pressure Monitor is consist of sphygmomanometer main body and cuff. The mainbody contains ABS housing, LCD, keys, measurement control module, pneumatic control module, power supply module. The Wrist-type Fully Automatic Digital Blood Pressure Monitor is automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment. The device with an inflatable cuff wrapping around the patient's wrist and the cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minutes. The cuff can measure pressure range from 0 to 300mmHg, and the pulse rate range from 30 to 180 beats/min.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are derived from the ISO 81060-2:2018+AMD2020 standard for clinical investigation of automated noninvasive sphygmomanometers.

Acceptance Criteria (from ISO 81060-2:2018+AMD2020)Reported Device Performance (Model DBP-2208, Cuff 13.5-21.2cm)Reported Device Performance (Model DBP-2242, Cuff 12.5-23cm)
Method 1:
Average difference for Systolic Blood Pressure ≤ ±5 mmHg0.84 mmHg0.87 mmHg
Standard deviation for Systolic Blood Pressure ≤ 8 mmHg5.25 mmHg5.22 mmHg
Average difference for Diastolic Blood Pressure ≤ ±5 mmHg0.21 mmHg0.20 mmHg
Standard deviation for Diastolic Blood Pressure ≤ 8 mmHg4.92 mmHg4.93 mmHg
Method 2:
Average difference for Systolic Blood Pressure ≤ ±5 mmHg0.84 mmHg0.87 mmHg
Standard deviation for Systolic Blood Pressure ≤ Maximum permissible standard deviation (based on mean difference, from Table 1 in document)4.52 mmHg (<6.95mmHg)4.48 mmHg (<6.887mmHg)
Average difference for Diastolic Blood Pressure ≤ ±5 mmHg0.21 mmHg0.20 mmHg
Standard deviation for Diastolic Blood Pressure ≤ Maximum permissible standard deviation (based on mean difference, from Table 1 in document)4.36 mmHg (<6.879 mmHg)4.38 mmHg (<6.95 mmHg)

Conclusion: Both tested models (DBP-2208 and DBP-2242) met all specified acceptance criteria according to both Method 1 and Method 2 of ISO 81060-2:2018+AMD2020.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 85 patients for each clinical investigation (one for cuff 13.5-21.2cm using DBP-2208 as representative, and one for cuff 12.5-23cm using DBP-2242 as representative). Each group comprised 43 females and 42 males.
  • Data Provenance: The document does not explicitly state the country of origin. It describes clinical studies conducted on "patients" and "subjects" indicating prospective data collection. "All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document states that a "manual Mercury Sphygmomanometer was used as a reference device" to establish the ground truth. It does not explicitly mention the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience"). However, for a clinical study involving a mercury sphygmomanometer as the reference, it is implied that trained medical professionals would perform these measurements. The standard ISO 81060-2 typically requires measurements by qualified observers.

4. Adjudication Method for the Test Set

The adjudication method described is the "Same wrist sequential method." This means the device under test and the reference device (manual Mercury Sphygmomanometer) measurements were taken sequentially on the same wrist. The document does not specify a separate "adjudication" process involving multiple experts reviewing discrepancies, but rather a direct comparison to the reference device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study focuses on the accuracy of the device itself compared to a reference standard, not on how human readers' performance might improve with AI assistance. This is a standalone device accuracy study.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, this was a standalone performance study. The "Wrist-type Fully Automatic Digital Blood Pressure Monitor" is an automated device designed to measure blood pressure. The clinical investigations directly assessed the device's accuracy against a manual reference without human intervention in the measurement process (other than proper placement and initiation).

7. Type of Ground Truth Used

The ground truth used was established by a manual Mercury Sphygmomanometer. This is considered a gold standard for blood pressure measurement in clinical studies.

8. Sample Size for the Training Set

The document does not provide any information about a training set. This is a validation study for a medical device's accuracy, not an AI/machine learning algorithm that typically requires a training set. The device determines blood pressure using an oscillometric method based on pre-programmed algorithms and hardware.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set for an AI/machine learning algorithm is mentioned in the document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

March 1, 2022

Joytech Healthcare Co., Ltd Dandan Huang Regulatory Engineer No.365, Wuzhou Road, Yuhang Economic Development Zone Hangzhou, Zhejiang 311100 China

Re: K213189

Trade/Device Name: Wrist-type Fully Automatic Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN Dated: January 20, 2022 Received: January 26, 2022

Dear Dandan Huang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213189

Device Name

Wrist-type Fully Automatic Digital Blood Pressure Monitor Including DBP-2101, DBP-2202, DBP-2116, DBP-2206, DBP-2220, DBP-2242, DBP-2141, DBP-2152, DBP-2253, DBP-2228, DBP-2229, DBP-2127, DBP-2261, DBP-2160

Indications for Use (Describe)

The Wrist-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

JOYTEC-I健玮拓

510(k) Summary

The assigned 510(k) number is:

Subjectter's Identification: 2.1

Name: JOYTECH Healthcare Co., Ltd..

Add .: No. 365. Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100 Zhejiang, China.

Contact Person: Yunhua Ren

Phone: +86-571-81957767

Fax: +86-571-81957750

Email: renyh(@sejoy.com

Name of the Device: 2.2

Trade Name: Wrist-type Fully Automatic Digital Blood Pressure Monitor

Including all models DBP-2101, DBP-2202, DBP-2116, DBP-2206, DBP-2208, DBP-2220, DBP-2242, DBP-2141, DBP-2152, DBP-2253, DBP-2229, DBP-2127, DBP-2127, DBP-2261, DBP-2160

Common Name: Blood Pressure Monitor

Classification name: Non-invasive blood pressure measurement System

21 CFR 870-1130, Class II, 74-DXN.

2.3 Classification Information:

Regulation Number: 870.1130 Product Code: DXN Device Class: II Panel: 74 Cardiovascular

2.4 Predicate Device Information:

The Wrist-type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the Predicate device of Wrist-type blood pressure monitor DBP-2261(K173024)

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for JOYTEC-I. The logo is in blue and red. The text is in a sans-serif font. The logo is simple and modern.

manufactured by JOYTECH Healthcare Co., Ltd..

2.5 Device Description:

The Wrist-type Fully Automatic Digital Blood Pressure Monitor is consist of sphygmomanometer main body and cuff. The mainbody contains ABS housing, LCD, keys, measurement control module, pneumatic control module, power supply module.

The Wrist-type Fully Automatic Digital Blood Pressure Monitor is automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment. The device with an inflatable cuff wrapping around the patient's wrist and the cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minutes. The cuff can measure pressure range from 0 to 300mmHg, and the pulse rate range from 30 to 180 beats/min. The detail comparisons between Wrist-type series are listed in table below:

FeaturesABCDEFGH (mm)I (cm)J(mm)KLMNO
Models
DBP-2101Y120Memories×1NNNNN76×67.5×2913.5-21.530×36NNNYY
DBP-2202Y30 Memories×4YNNNN79×64×2913.5-21.542×34NNNYY
DBP-2116Y120Memories×1YYNNN79×66×2813.5-21.545×30NNNYY
DBP-2229Y30 Memories×4YYNYY76×67.5×28.513.5-21.545×30NNNYY
DBP-2228Y30 Memories×4YNNNN76×67.5×28.513.5-21.545×30NNNYY
DBP-2127Y120Memories×1NNNNN76×67.5×28.513.5-21.545×30NNNYY
DBP-2206Y60 Memories×2YYNYY77×64×32.513.5-21.545×30NONYO
DBP-2208Y60 Memories×2YYYYN77×64×32.513.5-21.549×38OONYO
DBP-2220Y60 Memories×2YYNYY77×64×3213.5-21.549×38OONYO
DBP-2141Y120Memories×1YYNYY84×64×2913.5-21.545×30NONYO
DBP-2242Y60 Memories×2YYNYY84×64×2913.5-21.5or12.5-23cm(Optional)49×38OONYO

Table 2.1 Characteristics of Wrist-type Blood Pressure Monitor

{5}------------------------------------------------

DBP-2152Y60 Memories×2YYNYY77×64×3213.5-21.545×30NONYO
DBP-2253Y60 Memories×2YYYYN77×64×32.513.5-21.549×38OONYO
DBP-2261Y60 Memories×2YYNYY85×62×2513.5-21.543×40OONYO
DBP-2160Y60 Memories×2YYNYY85×62×2513.5-21.530.6×45OONYO

Note:

A = Powered by AAA Batteries

B= Memory Size

C= Time & Date

D = WHO (World Health Organization) Classification Indicator

E = Results Average in Three way

F = Irregular Heartbeat Detection

G = Last 3 Results Average

H= Outside Dimension (L x W x H in unit mm)

I = Cuff Size

J = LCD Size (Viewing Area in unit mm)

K= Backlight

L= Voice function

M= PC Interface

N=Low voltage detection

O=Beep

Y= Yes

N = No

O= Optional function depending on clients' needs

The devices are all designed and manufactured according to AAMVANSI/IEC80601-2-30:2018 medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.

2.6 Intended Use:

The Wrist-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age.

2.7 Comparison of Technological Characteristics with predicate device:

The PCB and Main chip of blood pressure monitor device inside were changed based on the marketed products which applied to K170666 including models DBP-2101, DBP-2202, DBP-2116, DBP-2206, DBP-2208, DBP-2220, DBP-2242, DBP-2141, DBP-2152, DBP-2228, DBP-2229, DBP-2127 and marketed products which applied to K173024 including models

{6}------------------------------------------------

DBP-2261, DBP-2160 manufacturing by JOYTECH Healthcare Co., Ltd. , which is considered to be significant change affectiveness and safety. Therefore it had to resubmitted 510(k) application. As for intended use, operation principle, features and specification have been keep the same, the device DBP-2261(K173024) was selected as predicate device. The detail comparison of technical characteristic as table below:

Comparison itemSubject device in present applicationPredicate device K173024(Model:DBP-2261)Comparison result / Explanation
The trade nameWrist-type Fully Automatic Digital Blood Pressure MonitorWrist-type Fully Automatic Digital Blood Pressure MonitorIdentical
ManufacturerJOYTECH Healthcare Co., Ltd.JOYTECH Healthcare Co., Ltd.Identical
Recommended classification regulation21CFR 870.1130,Noninvasive Blood Pressure Measurement System21CFR 870.1130,Noninvasive Blood Pressure Measurement SystemIdentical
Regulatory classIIIIIdentical
Panel74 Cardiovascular74 CardiovascularIdentical
Product codeDXNDXNIdentical
Indications for useThe Fully Automatic Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 year of age.The device is intended for used by adults and adolescents age 12 through 21 years of age to measure the systolic and diastolic blood pressure and pulse rate.Identical
Environment of UseHome and Clinical environment (OTC)Home and Clinical environment (OTC)Identical
Measuring principleOscillometric methodOscillometric methodIdentical
Measurement typeDetermined during deflationDetermined during deflationIdentical
Cuff locationWristWristIdentical
Measuring rangeSystolic Pressure: 60mmHg~260 mmHgDiastolic Pressure:30mmHg200 mmHgPulse:30180 Beats/minSystolic Pressure: 60mmHg~260 mmHgDiastolic Pressure:30 mmHg200 mmHgPulse:30180 Beats/minIdentical
Max cuff pressure300 mmHg300 mmHgIdentical
Cuff circumferenceOther models: Fits wristcircumference 13.5-21.5cmDBP-2242 Fits wristcircumference 13.5-21.5cm and12.5-23cm (optional)Fits wrist circumference13.5-21.5cmSimilarNote1
Table 2.2 The comparison of subject device and Predicate device
---------------------------------------------------------------------------

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for JOYTECH. The logo is in blue and red. The text is in English and Chinese.

JOYTECH Healthcare Co., Ltd..
Pulse: ± 5%Pulse: ± 5%Identical
InflationBy air pumpBy air pumpIdentical
Pressure releaseBy solenoid valveBy solenoid valveIdentical
Operating Temp. &humidityTemp.: 10°C40°CHumidity: 1593%RHAtmospheric:800hPa~1060hPaTemp.: 10°C40°CHumidity: 1593%RHAtmospheric:700hPa~1060hPaSimilarNote 2
Storage Temp.& humidityTemp.: -25°C~55°CHumidity:≤93% RHTemp.: -25°C~70°CHumidity:≤93% RHSimilarNote 3
Pressure sensorSilicon integrate pressuresensorSilicon integrate pressure sensorIdentical
Air PumpDC3V Micro air pumpDC3V Micro air pumpIdentical
Solenoid valveDC3V solenoid-controlledvalveDC3V solenoid-controlledvalveIdentical
Main ChipSZC900uPD78F0485wDifferentNote 4
Display componentDevice LCDDevice LCDIdentical
PCBDBP-2101: BP01SNPCBDBP-2202: BP02SNPCBDBP-2206: BP06SNPCBDBP-2208, DBP-2220:BP08SNPCBDBP-2116: BP16SNPCBDBP-2228, DBP-2229,DBP-2220: BP20SNPCBDBP-2261, DBP-2160:BP61SNPCBDBP-2242, DBP-2141,DBP-2152, DBP-2253:BSP22SNPCBBP61PCBSimilarNote 5
Supply power source2*1.5V AAA batteries2*1.5V AAA batteriesIdentical
PC InterfaceNoNoIdentical
Ingress Protection RatingIP22IP22Identical
MaterialABS housing and Nylon FleececuffABS housing and Nylon FleececuffIdentical
SterilizationNot applicableNot applicableIdentical
Device Functions
Irregular heart beatIrregular heart beat is displayedon the LCD (some models seetable 2.1)Irregular heart beat is displayedon the LCD.SameThere is nonew risks fordeleting thefunction.
WHO classificationindicatorNone: DBP-2101, DBP-2202,DBP-2228, DBP-2127YesSameThere is no

{8}------------------------------------------------

JOXTEC-IIIT
---------------
Yes: other modelsnew risks fordeleting thefunction.
DBP-2101, DBP-2127,DBP-2116, DBP-2141:$120 \text{ Memories } \times 1$DBP-2202, DBP-2228,DBP-2229: $30 \text{ Memories } \times 4$Other models: $60 \text{ Memories } \times 2$2*60 MemoriesSimilarNote6
Memory functionDBP-2208 own this function,while other models none.NoneSimilarNote 7
Results Average in ThreewayDBP-2208 own this function,while other models none.NoneSimilarNote 7
Last 3 Results AverageYes: DBP-2229, DBP-2206,DBP-2220, DBP-2141,DBP-2242, DBP-2152,DBP-2261, DBP-2160Other models noneYesIdentical
BacklightOptional: DBP-2208,DBP-2220, DBP-2242,DBP-2253, DBP-2261,DBP-2160Other model noneOptionalIdentical
VoiceOptional: DBP-2206,DBP-2208, DBP-2220,DBP-2141, DBP-2242,DBP-2152, DBP-2253,DBP-2261, DBP-2160Other models noneOptionalIdentical

Note 1: The device DBP-2242 adding the fit wrist circumstance 12.5cm-23cm is not change the cuff material and device specification and feature. The accuracy and new cuff size have been verified and validated by safety and clinical test report. It cannot be bring new risks.

Note 2: The operation atmospheric pressure is relate to the altitude. An increase in altitude is accompanied by a decrease in atmospheric pressure. The storage environment of the subject device modified have been verified by IEC 60601-1 and IEC80601-2-30 standard test.

Note 3: The storage environment of the subject device have been verified by IEC 60601-2-30 standard test.

Note 4: The main chip affect the accuracy and safety of the blood pressure monitor device.The safety and accuracy of the subject device have been verified by IEC 60601-2-30 and IEC81060-2 standard test.

Note 5: The PCB affect the accuracy and safety of the blood pressure monitor device. The safety and accuracy of the subject device have been verified by IEC 60601 series , IEC 80601-2-30 and IEC81060-2 standard test.

Note 6: Memory group setting does not influence the accuracy, and the memory function have been verified by

{9}------------------------------------------------

software verification and validation report.

Note 7: Results average in three way function indicate three ways to evaluate blood pressure monitor average value. This function have been validated by Software verification and validation report.

2.8. Performance Data:

Testing information demonstrating safety and effectiveness of the device in the intended environment of use is supported by testing that was conducted.

The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices.

The following National and International Standards were utilized for testing the subject device.

Electrical Safety and performance requirements:

AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012,C1:2009(R)2012 And A2:2010/(R)2012 Medical Electrical Equipment.

IEC 80601-2-30:2018, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.

Home-used medical equipment requirements and environmental test:

IEC 60601-1-11:2015 General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

Electromagnetic compatibility requirements:

IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic - disturbances - Requirements and tests

Bio-compatibility Evaluation for patient contacting components:

ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization

Guidance Document:

The software/firmware verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "FDA guidance Use of International Standard ISO 10993."

The test result all meet or exceed the requirement of these standards.

{10}------------------------------------------------

Image /page/10/Picture/2 description: The image shows the logo for JOYTECH. The logo is in blue and red, with the word "JOYTECH" in large letters. To the right of the word "JOYTECH" are three Chinese characters. The logo is simple and modern.

2.9 Discussion of Clinical Tests Performed:

Clinical Validation:

  • ISO81060-2:2018+AMD2020 Non-invasive sphygmomanometers — Part 2: Clinical investigation of intermittent automated measurement type.

2.9.1 Fit cuff circumstance 13.5-21.2cm for all all models

In the Wrist-type fully automatic digital blood pressure clinical investigation, Model DBP- 2208 was selected as representative for testing. 85 patients (43 females and 42 males) participated in the clinical study. Same wrist sequential method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

The clinical investigate statistic results in detail as below:

1) Statistic Methods:

Eligibility criteria 1: The mean value of difference is equal or within to 土 5mmHg and the standard deviation of no greater than 8 mmHg between subject device and reference device.

Eligibility criteria 2 : According the known mean value of difference to determine the standard deviation required on the basis of table below. If the standard deviation is less than the corresponding standard deviation limit, the data set is judged as acceptable.

$ x ¯ $Maximum permissible standard deviation, $s_m$ , as function of, $ x ¯ $ mmHg
0.00.10.20.30.40.50.60.70.80.9
± 0.6,956,956,956,956,936,926,916,906,896,88
± 1,6,876,866,846,826,806,786,766,736,716,68
± 2,6,656,626,586,556,516,476,436,396,346,30
± 3.6,256,206,146,096,035,975,895,835,775,70
± 4.5,645,565,495,415,335,255,165,085,014,90
± 5.4,79

Table 1 - Averaged subject data acceptance (criterion 2) in mmHg

{11}------------------------------------------------

Image /page/11/Picture/2 description: The image shows the logo for JOYTECH. The logo is in blue and red. To the right of the word JOYTECH are three Chinese characters.

2) Statistic results

Average(mmHg)Standard deviation(mmHg)
Method 1
Systolic blood pressure0.845.25
Diastolic blood pressure0.214.92
Method 2
Systolic blood pressure0.844.52
Diastolic blood pressure0.214.36

According to Table above, the statistical results are as follows:

Method 1:

Average of systolic blood pressure is 0.84 mmHg (<+5 mmHg), standard deviation of systolic blood pressure is 5.25 mmHg (<8 mmHg)

Average of diastolic blood pressure is 0.21 mmHg (<+5 mmHg), standard deviation of diastolic blood pressure is 4.92 mmHg (<8 mmHg)

Method 2:

Average of systolic blood pressure is 0.84mmHg (<+5 mmHg), standard deviation of systolic blood pressure is 4.52 mmHg (<6.95mmHg)

Average of diastolic blood pressure is 0.21 mmHg (<+5 mmHg), standard deviation of diastolic blood pressure is 4.36 mmHg (<6.879 mmHg)

After comparing, the conclusion is that the averages difference in systolic and diastolic pressure and the corresponding standard deviation fall are within the range of the standard. It meets the requirements of clinical program.

2.9.2 Fit cuff circumstance 12.5-23cm for DBP-2242

In the Wrist-type fully automatic digital blood pressure clinical investigation, Model DBP-2242 was selected as representative for testing. 85 patients (43 females and 42 males) participated in the clinical study. Same wrist sequential method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunter to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

The clinical investigate statistic result as below:

  1. The Statistic Methods are the same as the clause 2.9.1;

{12}------------------------------------------------

Image /page/12/Picture/2 description: The image shows the logo for JOYTECH. The logo is in blue and red. The text "JOYTECH" is in blue, with the "T" in red. To the right of the text is Chinese characters in blue.

  1. The Statistic Results presented as below:
Average(mmHg)Standard deviation(mmHg)
Method 1
Systolic blood pressure0.875.22
Diastolic blood pressure0.204.93
Method 2
Systolic blood pressure0.874.48
Diastolic blood pressure0.204.38

According to Table above, the statistical results are as follows:

Method 1:

Average of systolic blood pressure is 0.87 mmHg (<+5 mmHg), standard deviation of systolic blood pressure is 5.22 mmHg (<8 mmHg)

Average of diastolic blood pressure is 0.20 mmHg (<+5 mmHg), standard deviation of diastolic blood pressure is 4.93 mmHg (<8 mmHg)

Method 2:

Average of systolic blood pressure is 0.87 mmHg (<+5 mmHg), standard deviation of systolic blood pressure is 4.48 mmHg (<6.887mmHg)

Average of diastolic blood pressure is 0.20 mmHg (<+5 mmHg), standard deviation of diastolic blood pressure is 4.38 mmHg (<6.95 mmHg)

After comparing, the conclusion is that the averages difference in systolic and diastolic pressure and the corresponding standard deviation fall are within the range of the standard.It meets the requirements of clinical program.

2.10 Conclusions

This submitted wrist-type series manufactured by JOYTECH Healthcare Co., Ltd. have been found to be respectively substantially equivalent to the predicate device (Wrist-type Fully Automatic Digital Blood Pressure Monitor DBP-2261, K173024) manufactured by JOYTECH Healthcare Co., Ltd. Based on the information provided in this submission, differences between the subject device and predicate devices have been tested and documented, it has been demonstrated that the subject devices are as safe and effective as the predicate device.

In the other words, the difference between the subject and the predicate devices does not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device, and the new models as mentioned on this submission are considered substantial equivalent to the predicate device

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).