K Number
K213075

Validate with FDA (Live)

Date Cleared
2022-01-20

(119 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Blue Nitrile Examination Gloves Powder Free are disposable devices intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

Blue Nitrile Examination Gloves Powder Free are equivalent to the Class I patient examination gloves bearing the product code LZA (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. They are made from Nitrile (NBR)100%. These gloves are blue in color and are powder free. The product is non-sterile, ambidextrous and single use only.

Blue Nitrile Examination Gloves Powder Free with sizes X-Small, Medium, Large and X-Large are included in the submission.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Blue Nitrile Examination Gloves Powder Free, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

Characteristic / Test MethodAcceptance Criteria (Standard)Reported Device Performance
Dimensions - Length (ASTM D6319-19)Min 230 mm for all sizesX-Small: 250 mm, Small: 241 mm, Medium: 240 mm, Large: 243 mm, X-Large: 240 mm (All > 230 mm)
Dimensions - Width (ASTM D6319-19)X-Small: 70+/-10 mm, Small: 80+/-10 mm, Medium: 95+/-10 mm, Large: 110+/-10 mm, X-Large: 120+/-10 mmX-Small: 80 mm, Small: 82 mm, Medium: 93 mm, Large: 103 mm, X-Large: 110 mm (All within specified ranges)
Physical Properties - Tensile Strength (Before Ageing) (ASTM D6319-19)14 MPa Min for all sizesX-Small: 24.5 MPa, Small: 22.7 MPa, Medium: 25.3 MPa, Large: 26.4 MPa, X-Large: 25.5 MPa (All > 14 MPa)
Physical Properties - Tensile Strength (After Ageing) (ASTM D6319-19)14 MPa Min for all sizesX-Small: 23.8 MPa, Small: 24.3 MPa, Medium: 31.9 MPa, Large: 30.3 MPa, X-Large: 27.4 MPa (All > 14 MPa)
Physical Properties - Ultimate Elongation (Before Ageing) (ASTM D6319-19)500% Min for all sizesX-Small: 620%, Small: 580%, Medium: 530%, Large: 540%, X-Large: 540% (All > 500%)
Physical Properties - Ultimate Elongation (After Ageing) (ASTM D6319-19)400% Min for all sizesX-Small: 590%, Small: 560%, Medium: 530%, Large: 540%, X-Large: 530% (All > 400%)
Thickness (Palm) (ASTM D6319-19)0.05 mm min for all sizesX-Small: 0.101 mm, Small: 0.089 mm, Medium: 0.100 mm, Large: 0.098 mm, X-Large: 0.094 mm (All > 0.05 mm)
Thickness (Finger) (ASTM D6319-19)0.05 mm min for all sizesX-Small: 0.121 mm, Small: 0.103 mm, Medium: 0.116 mm, Large: 0.121 mm, X-Large: 0.114 mm (All > 0.05 mm)
Powder Residue (ASTM D6124-06)≤ 2 mg/gloveX-Small: 1.07 mg/glove, Small: 1.49 mg/glove, Medium: 0.62 mg/glove, Large: 0.56 mg/glove, X-Large: 1.60 mg/glove (All ≤ 2 mg/glove)
Water Tight (ASTM D5151-19)AQL 2.5Passes AQL 2.5
Biocompatibility - Primary Skin Irritation (ISO 10993-10:2010)Under the condition of study not an irritantUnder the condition of study, not an irritant
Biocompatibility - Dermal Sensitization (ISO 10993-10:2010)Under the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizer
Biocompatibility - In vitro Cytotoxicity (ISO 10993-5:2009)Under the conditions of the study, non-cytotoxicUnder the conditions of the study, the device is cytotoxic. (Note: This is a discrepancy. The table in section F states "Similar" and describes it as "non-cytotoxic at 50%, 25%, 12.5% and 6.25% extracts." However, the table in section G, which details the bench test data, explicitly states "the device is cytotoxic." This is a significant finding that would typically require further explanation or clarification in a submission. Assuming the 'bench test data' result is the final documented finding for this specific test.)
Biocompatibility - Acute Systemic Toxicity Test (ISO 10993-11:2017)Under the conditions of study, the device extracts do not pose a systemic toxicity concernUnder the conditions of study, the device extracts did not reveal any systemic toxicity

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes for each test in the provided tables. However, it indicates that the tests were conducted according to established ASTM and ISO standards, which typically specify sample size requirements.

  • Provenance: This is a non-clinical device (examination gloves). The data provenance refers to laboratory bench testing. The location of the testing is not explicitly stated, but the applicant's address is in Thailand. Given the nature of bench testing, the "country of origin of the data" is the laboratory where the tests were performed, which is not specified beyond being conducted according to international standards. The data is prospective in the sense that these tests were performed specifically for this submission to demonstrate compliance.

3. Number of Experts and Qualifications for Ground Truth

Not applicable. This device is a physical product (examination glove) and its performance is evaluated through standardized physical, chemical, and biocompatibility bench tests, not through expert interpretation of medical images or observations. Therefore, there is no "ground truth" established by medical experts in the traditional sense for an AI/medical imaging device.

4. Adjudication Method

Not applicable for this type of device and testing. Adjudication methods like 2+1 or 3+1 are used for human-in-the-loop evaluations, typically with medical image interpretation, to resolve discrepancies among experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is not an AI/medical imaging device that would involve human readers or AI assistance in interpretation. It's a non-clinical evaluation of a physical product's safety and performance characteristics.

6. Standalone Performance Study

Yes, a standalone performance study was done. The entire "BENCH TEST DATA" section (pages 7-8) and "BIOCOMPATIBILITY DATA" section (page 8) outline the algorithm/device-only performance against predefined acceptance criteria from international standards (ASTM and ISO). There is no "human-in-the-loop" component described for these tests.

7. Type of Ground Truth Used

The "ground truth" for this device's performance relies on established international standards and specifications:

  • ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves for Medical Application)
  • ASTM D5151-19 (Standard Test Method for Detection of Holes in Medical Gloves)
  • ASTM D6124-06 (Reapproved 2017) (Standard Test Method for Residual Powder on Medical Gloves)
  • ISO 10993-10:2010(E) (Biological Evaluation of Medical Devices - Part 10, Tests for Irritation and Skin Sensitization)
  • ISO 10993-5:2009(E) (Biological Evaluation of Medical Devices - Part 5, Tests for In Vitro Cytotoxicity)
  • ISO 10993-11:2017(E) (Biological Evaluation of Medical Devices - Part 11, Tests for Systemic Toxicity)

The device's physical and biological properties are measured and compared directly against the quantitative thresholds and qualitative outcomes specified in these standards.

8. Sample Size for the Training Set

Not applicable. This is not an AI device that requires a training set. The "device" here refers to the physical product (gloves).

9. How the Ground Truth for the Training Set was Established

Not applicable, as no training set is involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 20, 2022

Thai Rubber Gloves Co., Ltd. Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114, Aurora, Illinois 60504

Re: K213075

Trade/Device Name: Blue Nitrile Examination Gloves Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: December 20, 2021 Received: December 20, 2021

Dear Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213075

Device Name Blue Nitrile Examination Gloves Powder Free

Indications for Use (Describe)

Blue Nitrile Examination Gloves Powder Free are disposable devices intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY K213075

AS REQUIRED BY: 21CFR§807.92(C)

A. APPLICANT INFORMATION

510(K) Owner's NameTHAI RUBBER GLOVES CO., LTD.
Address680 MOO 2 BANBUENG – KLAENG RD.,NONGYAI CHONBURI 20190 THAILAND.
Phone+66-2-1306356, +66-38-168613
Fax+66-2-1306357
E-mailchalongkwan@thaitex.cominfo@thairubbergloves.com
Contact PersonMiss Chalongkwan Wongsasuthikul
DesignationManaging Director
Contact Number+66-2-1306356
Contact Emailchalongkwan@thaitex.com
Date Submitted13 September 2021

B. DEVICE IDENTIFICATION

Name of the deviceBlue Nitrile Examination Gloves Powder Free
Product proprietary or trade nameGoody Gloves / Wincare / TBG Gloves
Common or usual nameBlue Nitrile Examination Gloves Powder Free
Classification namePolymer Patient Examination Glove
Device ClassificationClass-1
Product CodeLZA
Regulation Number21 CFR 880.6250
Review PanelGeneral Hospital

C. PREDICATE DEVICE

Predicate DeviceHi-Care Thai Gloves Co. Ltd.
510(k) NumberK202384
Regulatory ClassClass 1
Product codeLZA

D. DESCRIPTION OF THE DEVICE:

Blue Nitrile Examination Gloves Powder Free are equivalent to the Class I patient examination gloves bearing the product code LZA (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. They are made from Nitrile (NBR)100%. These gloves are blue in color and are powder free. The product is non-sterile, ambidextrous and single use only.

Blue Nitrile Examination Gloves Powder Free with sizes X-Small, Medium, Large and X-Large are included in the submission.

{4}------------------------------------------------

E. INDICATION FOR USE OF THE DEVICE:

Blue Nitrile Examination Gloves Powder Free are disposable devices intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE

CHARACTERSTICSSTANDARDSDEVICE PERFORMANCEComparison
PREDICATESUBJECT
510(K) Number---K202384K213075
Name of device---Palm Care Blue NitrileExamination GlovesPowder freeBlue NitrileExamination GlovesPowder free---
Dimensions - LengthASTM D6319-19Length Min 230 mmLength > 230 mmSimilar
SizeAverage
X-Small250
Small241
Medium240
Large243
X-Large240
Dimensions - WidthASTM D6319-19Width Min 95+/-10mm(for medium size)Width 95+/-10mm(for medium size)Similar
SizeAverage
X-Small80
Small82
Medium93
Large103
X-Large110
Physical Properties-Tensile StrengthASTM D6319-19Before AgeingTensile Strengthmin 14 MPaBefore AgeingTensile Strength > 14 MPaSimilar
SizeActualvalue
X-Small24.5
Small22.7
Medium25.3
Large26.4
X-Large25.5
After AgeingTensile Strengthmin 14 MPaAfter AgeingTensile Strength > 14 MPaSimilar
SizeActualvalue
X-Small23.8
Small24.3
Medium31.9
Large30.3
X-Large27.4

{5}------------------------------------------------

CHARACTERSTICSSTANDARDSDEVICE PERFORMANCEComparison
PREDICATESUBJECT
510(K) NumberK202384K213075
Physical Properties-Ultimate ElongationASTM D6319-19Before AgeingUltimate Elongation> 500%Before AgeingUltimate Elongation > 500%Size Actual value X-Small 620 Small 580 Medium 530 Large 540 X-Large 540Similar
After AgeingUltimate Elongation> 400%After AgeingUltimate Elongation > 400%Size Actual value X-Small 590 Small 560 Medium 530 Large 540 X-Large 530Similar
ThicknessASTM D6319-19Palm min 0.05 mmFinger min 0.05 mmPalm > 0.05 mm Finger > 0.05 mmSize Palm(Actualvalue) Finger(Actualvalue) X-Small 0.101 0.121 Small 0.089 0.103 Medium 0.100 0.116 Large 0.098 0.121 X-Large 0.094 0.114Similar
Powder ResidueASTM D6319-19<2 mg/glove<2 mg/gloveSize Actual value X-Small 1.07 Small 1.49 Medium 0.62 Large 0.56 X-Large 1.60Similar
BiocompatibilityPrimary SkinIrritation- ISO 10993-10:2010(E)Under the conditionof study not anirritantUnder the condition of study notan irritantSame
Dermal Sensitization-ISO 10993-10:2010(E)Under theconditions of thestudy not asensitizerUnder the conditions of the studynot a sensitizerSame
In vitro cytotoxicity-ISO 10993-5:2009(E)Under theconditions of thestudy, noncytotoxicUnder the conditions of the study,"non-cytotoxic" at 50%, 25%,12.5% and 6.25% extractsSimilar
Acute SystemicToxicity Test- ISO10993-11:2017(E)Under theconditions of study,the device extractsdo not pose asystemic toxicityUnder the conditions of study, thedevice extracts did not reveal anysystemic toxicitySame

{6}------------------------------------------------

CHARACTERSTICSSTANDARDSDEVICE PERFORMANCEComparison
PREDICATEK202384SUBJECTK213075
510(K) Number---Similar
Water Tight(1000 ml)ASTM D5151-19AQL 2.5PassesPassesSimilar
Indication for Use---Blue NitrileExamination GlovesPowder Free isdisposable devicesintended for medicalpurpose that are wornon the examiner'shand to preventcontaminationbetween patient andexaminer.Blue Nitrile ExaminationGloves Powder Free aredisposable devices intendedfor medical purpose that isworn on the examiner's handto prevent contaminationbetween patient and examiner.Similar
MaterialASTM D6319-19Nitrile (NBR)Nitrile (NBR)Same
Color---BlueBlueSame
SizeASTM D6319-19Extra Small, Small,Medium, Large, ExtraLargeExtra Small, Small, Medium,Large, Extra LargeSame
Single UseMedical GloveGuidanceManual- LabelingSingle UseSingle UseSame
Sterile/non sterile---NonsterileNonsterileSame
Powder/Powder free---Powder freePowder freeSame
Label and LabelingMeets FDA'slabel andlabelingrequirementsMeets FDA's label andlabeling requirementsMeets FDA's label and labelingrequirementsSame
Manufacturer(s)---Hi-Care Thai Gloves Co.Ltd.THAI RUBBER GLOVES CO.,LTD..---

There are no significant differences between the two products and are identical in terms of intended use, materials, design and manufacturing methods. Both devices meet the ASTM standard D6319-19.

{7}------------------------------------------------

G. COMPARISON BASED ON ASSESSMENT OF NON-CLINICAL PERFORMANCE DATA

BENCH TEST DATA

TEST METHODPURPOSEACCEPTANCECRITERIARESULT
ASTM D6319-19Standard Specification for NitrileExamination Gloves for MedicalApplication.To determine thelength of the glovesMin 230 mm for allsizesX-Small : 250 mmSmall : 241 mmMedium : 240 mmLarge : 243 mmX-Large : 240 mm
ASTM D6319-19Standard Specification for NitrileExamination Gloves for MedicalApplication.To determine thewidth of the glovesX-Small : 70+/-10 mmSmall : 80+/-10 mmMedium : 95+/-10 mmLarge : 110+/-10 mmX-Large : 120+/-10 mmX-Small : 80 mmSmall : 82 mmMedium : 93 mmLarge : 103 mmX-Large : 110 mm
TESTMETHODPURPOSEACCEPTANCECRITERIARESULT
ASTM D6319-19StandardSpecification forNitrile ExaminationGloves for MedicalApplication.To determine thethickness of theglovesPalm0.05 mm minfor all sizesFinger0.05 mm minfor all sizesSizePalmFinger
X-Small0.101 mm0.121 mm
Small0.089 mm0.103 mm
Medium0.100 mm0.116 mm
Large0.098 mm0.121 mm
X-Large0.094 mm0.114 mm
ASTM D6319-19StandardSpecification forNitrile ExaminationGloves for MedicalApplication.To determine thephysical properties-Tensile strengthBefore AgeingTensile Strength14MPa Min for allsizesAfter AgeingTensile Strength14MPa Min for allsizesSizeBefore ageingAfter ageing
X-Small24.5 MPa23.8 MPa
Small22.7 MPa24.3 MPa
Medium25.3 MPa31.9 MPa
Large26.4 MPa30.3 MPa
X-Large25.5 MPa27.4 MPa
To determine thephysical properties-Ultimate ElongationBefore AgeingUltimateElongation500% Min for allsizesAfter AgeingUltimateElongation 400%Min for all sizesSizeBefore ageingAfter ageing
X-Small620%590%
Small580%560%
Medium530%530%
Large540%540%
X-Large540%530%

{8}------------------------------------------------

TEST METHODPURPOSEACCEPTANCECRITERIARESULT
ASTM D5151-19 StandardTest Method for Detectionof Holes in Medical GlovesTo determine the holesin the glovesAQL 2.5Gloves Passes AQL 2.5
ASTM D6124-06(Reapproved 2017) StandardTest Method for ResidualPowder on Medical GlovesTo determine theresidual powder in thegloves≤ 2 mg/gloveX-Small : 1.07 mg/gloveSmall : 1.49 mg/gloveMedium : 0.62 mg/gloveLarge : 0.56 mg/gloveX-Large : 1.60 mg/glove

BIOCOMPATIBILITY DATA

TEST METHODPURPOSEACCEPTANCECRITERIARESULT
ISO 10993-10:2010(E)Biological Evaluation ofMedical Devices - Part 10,Tests for Irritation and SkinSensitization.Test done for irritation.To evaluate the testitem, for skin irritationtest in New ZealandWhite rabbits.Under the conditionof study not anirritantUnder the condition ofstudy, not an irritant
ISO 10993-10:2010(E)Biological Evaluation ofMedical Devices - Part 10,Tests for Irritation and SkinSensitization.Test done for skin sensitizationTo evaluate the testitem, for the skinsensitization in Guineapigs by maximizationtest.Under the conditionsof the study, not asensitizerUnder the conditions ofthe study, not asensitizer
ISO10993-5:2009(E)Biological Evaluation ofMedical Devices - Part 5,Tests for In VitroCytotoxicity.To evaluate the testitem, for its ability toinduce cytotoxicityusing L-929 mousefibroblast cells byElution Method.Under the conditionsof the study, non-cytotoxicUnder the conditionsof the study, the deviceis cytotoxic.
ISO 10993-11:2017(E)Biological Evaluation ofMedical Devices - Part 11,Tests for Systemic Toxicity.To evaluate the testitem, for acute systemictoxicity in Swiss AlbinoMice.Under the conditionsof study, the deviceextracts do not pose asystemic toxicityconcernUnder the conditions ofstudy, the deviceextracts did not revealany systemic toxicity

The performance test data of the non-clinical tests that support a determination of safety and equivalence is the same as mentioned above (ASTM Requirements).

The performance test data of the non-clinical tests meet following standards:

ASTM D6319-19 Standard Specification for Nitrile examination Gloves for Medical Application.

{9}------------------------------------------------

ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves

ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves

ISO 10993-10:2010(E) Biological Evaluation of Medical Devices - Part 10, Tests for Irritation and Skin Sensitization.

ISO 10993-5:2009(E) Biological Evaluation of Medical Devices - Part 5, Tests for In Vitro Cytotoxicity.

ISO 10993-11:2017(E) Biological Evaluation of Medical Devices - Part 11, Tests for Systemic Toxicity.

H. COMPARISON BASED ON ASSESSMENT OF CLINICAL PERFORMANCE DATA

Not applicable - Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

I. CONCLUSION

The conclusions drawn from the non-clinical test demonstrate that the subject devices in 510(K) submission, Goody Gloves / Wincare / TBG Gloves (Blue Nitrile Examination Gloves Powder Free) are as safe, as effective, and performs as well as or better than the legally marketed predicate device K202384.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.