K Number
K213058
Date Cleared
2022-02-08

(139 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

O'Star Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

O'Star Nitrile Examination Gloves Powder Free is a Class I device bearing the product code LZA (21CFR 880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. They are made from acrylonitrilebutadiene copolymer dispersion. These gloves are blue in color having Finger Texture, Ambidextrous and are powder free. The product is non-sterile.

AI/ML Overview

This document (K213058) is a 510(k) Premarket Notification for O'Star Nitrile Examination Gloves Powder Free. It primarily focuses on demonstrating substantial equivalence to a predicate device (K210276) through non-clinical performance and biocompatibility testing. The data provided is for the verification and validation of the physical properties and biocompatibility of the gloves, not for an AI/algorithm-based medical device.

Therefore, many of the requested items regarding AI/algorithm performance (e.g., sample size for training set, number of experts for ground truth, MRMC studies, AI assistance effect size) are not applicable to this type of medical device submission.

Here's the breakdown of the information provided in the document concerning the acceptance criteria and the study proving the device meets them:

1. A table of acceptance criteria and the reported device performance:

The document provides two tables in Section VII. PERFORMANCE DATA.

Table 2: Performance Testing Summary - This table details the physical and barrier properties of the gloves.

SI No.TestsProposed Device actual DataAcceptance CriteriaResult
1.Dimension
Length, Width and Thickness
ASTM D6319-19 /ASTM D412-16/ASTM D3767-03-(Reapproved2020) Standard Specification for Nitrile Examination Gloves for Medical ApplicationLength
S: 247.2mm
M: 243.8mm
L: 245.5mm
XL: 247.2mm
Width
S: 90mm
M: 103.7mm
L: 104.2mm
XL: 110.2mm
Thickness
S: Palm 0.055mm, Finger 0.052mm
M: Palm 0.052mm, Finger 0.053mm
L: Palm 0.053mm, Finger 0.053mm
XL: Palm 0.057mm, Finger 0.053mmLength
M: 230mm min
(Note: Specific length criteria for S, L, XL are not explicitly listed in the "Acceptance Criteria" column, but based on the "Pass" result and the general ASTM standard, it implies they met standard)
Width
S: 80mm±10
M: 95mm±10
L: 110mm±10
XL: 120mm±10
Thickness
Palm: 0.05 mm min
Finger: 0.05 mm minPass
2.Physical property
Tensile strength and Ultimate Elongation
ASTM D6319-19/ASTM D573-04 (Reapproved 2019) Standard Specification for Nitrile Examination Gloves for Medical ApplicationTensile strength
Before Aging:
S: 14.23 Mpa, M: 15.85 Mpa, L: 14.92 Mpa, XL: 15.32 Mpa
After Aging:
S: 14.07Mpa, M: 14.09 Mpa, L: 14.13 Mpa, XL: 14.81 Mpa
Ultimate Elongation
Before Aging:
S: 702%, M: 935%, L: 768%, XL: 831%
After Aging:
S: 694%, M: 836%, L: 728%, XL: 785%Tensile strength
Before Aging: 14Mpa Min for all sizes
After Aging: 14Mpa Min for all sizes
Ultimate Elongation
Before Aging: 500% Min for all sizes
After Aging: 400%Min for all sizesPass
3.Detection of Holes in Medical Gloves
ASTM D6319-19 /ASTM D5151-19 Standard Test Method for Detection of Holes in Medical GlovesAQL=2.5AQL=2.5Pass
4.Powder Free Residue
ASTM D6124-06(Reapproved2017) Standard Test Method for Residual Powder on Medical GlovesResidual powder content
S: 0.03 mg
M: 0.07 mg
L: 0.01 mg
XL: 0.05 mg≤2 Mg/Glove Max for all sizesPass

Table 3: Biocompatibility Test Summary - This table details the biological safety of the gloves.

SI. NoTest PerformedProposed DeviceAcceptance CriteriaResult
1.Invitro CytotoxicityUnder the conditions of the study, cytotoxic to L-929 cells. Additional Testing was performed to determine if this was a systemic toxicity concern.Under the conditions of the study, not cytotoxic.Pass
2.Skin SensitizationUnder the conditions of the study not a sensitizer.Under the conditions of the study not a sensitizer.Pass
3.Skin IrritationUnder the condition of study not an irritant.Under the condition of study not an irritant.Pass
4.Acute Systemic ToxicityUnder the conditions of the study, the device extracts do not pose a systemic toxicity concern.Under the conditions of the study, the device extracts do not pose a systemic toxicity concern.Pass

2. Sample sizes used for the test set and the data provenance:

  • Sample Size for Test Set: The document does not explicitly state the sample sizes (number of gloves) for each physical and biocompatibility test performed. However, for a 510(k) submission, implied sample sizes for such tests typically follow the requirements of the referenced ASTM and ISO standards. These standards specify sample sizes for testing batches of gloves to ensure compliance.
  • Data Provenance: The manufacturer's address is listed as "Khac Niem Industrial Cluster, Khac Niem Ward, Bac Ninh City, Bac Ninh Province, Vietnam." This implies the testing was likely conducted in Vietnam or by labs accredited by Vietnamese or international standards, reflecting the country of origin of the data. The data is retrospective, as it was generated from pre-market testing to support the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This question is not applicable as this submission is for a physical medical device (gloves), not an AI/algorithm. Ground truth in this context is established by laboratory measurements according to established international standards (ASTM, ISO).

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This question is not applicable for a physical device. Testing results are based on objective measurements and established standards, not human adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This question is not applicable as this submission is for a physical medical device (gloves), not an AI/algorithm. No human readers or AI assistance are involved in the performance evaluation of examination gloves.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This question is not applicable as this submission is for a physical medical device (gloves), not an AI/algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the performance of these gloves is established by objective laboratory measurements against predefined international standard specifications (ASTM D6319-19, ASTM D412-16, ASTM D3767-03, ASTM D573-04, ASTM D5151-19, ASTM D6124-06) for physical properties (dimensions, tensile strength, elongation, hole detection, powder residue) and biological testing standards (ISO 10993-1:2018) for biocompatibility (cytotoxicity, skin sensitization, skin irritation, acute systemic toxicity).

8. The sample size for the training set:

This question is not applicable as this submission is for a physical medical device (gloves), not an AI/algorithm. There is no "training set" in the context of device testing for physical properties unless referring to batch manufacturing quality control, which is implied by the adherence to standards.

9. How the ground truth for the training set was established:

This question is not applicable as this submission is for a physical medical device (gloves), not an AI/algorithm.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.