K Number
K213049

Validate with FDA (Live)

Date Cleared
2022-10-16

(389 days)

Product Code
Regulation Number
868.2775
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is a nerve stimulation device designed to be used by an anesthetist during:

  1. General anesthesia, for the purpose of establishing the efficacy of a Neuromuscular Blocking Agent using non-invasive surface electrodes.

  2. Regional anesthesia, for the purpose of:

i. nerve mapping using the non-invasive Nerve Mapping Probe (supplied)

ii. nerve locating using invasive needles (not supplied)

Device Description

The STIMPOD NMS450 is a quantitative Neuromuscular Transmission (NMT) Monitor which provides real-time quantitative feedback utilizing tri-axial accelerometery.

The STIMPOD NMS450 is also a precision nerve locating tool used for localizing specific neural pathways. Localization of nerves by electrical stimulation involves connecting the nerve stimulator to a conducting needle through which local anaesthetics can be injected. The distance of the needle (cathode) from the nerve can be estimated by establishing the minimum threshold current required, to facilitate a neuromuscular response.

The STIMPOD NMS450 currently makes use of 3 dimensional acceleromyography (AMG) as the measurement technology of choice for objective Neuromuscular Transmission (NMT) monitoring in surgery. Although AMG is still the de-facto standard in industry, the use of electromyography (EMG) has gained renewed interest in NMT monitoring applications where the introduction of robotic surgeries, new surgical techniques, and changes in workflow, for instance, made the use of AMG impracticable.

In order to address this steadily increasing market demand for EMG sensor technology in NMT monitoring applications, Xavant identified the need to design and develop a dedicated NMT Monitoring Cable (EMG) that could be used interchangeably with the existing NMT Monitoring Cable (AMG) already in use with the STIMPOD NMS450.

The NMT Monitoring Cable (EMG) was specifically designed for use with the STIMPOD NMS450 to assist anaesthesiologists in theatre with monitoring the efficacy of Neuromuscular Blocking Agents (NMBAs).

The NMT Monitoring Cable (EMG) was designed with an intelligent EMG signal processing module (henceforth referred to as the µEMG) on one end of the cable for connection to an EMG electrode, and with a standard FireWire connector on the other end for connection to the STIMPOD NMS450.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the STIMPOD NMS450 Nerve Stimulator, focusing on its substantial equivalence to a predicate device (STIMPOD NMS450, K102084) and a reference device (Senzime, Tetragraph, K190795) with the addition of a new NMT Monitoring Cable (EMG) accessory.

However, the document does not provide specific acceptance criteria or an explicit study proving the device meets those criteria with detailed performance metrics. Instead, it states that "The subject device met or exceeded the stated acceptance criteria for each performance test." It then lists the types of performance tests conducted.

Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor can I detail specific sample sizes, ground truth establishment, or multi-reader multi-case study results, as this information is not present in the provided text.

Based on the available information, here's what can be extracted:

Acceptance Criteria and Reported Device Performance:

The document broadly states that the device met or exceeded acceptance criteria, but it does not provide the specific criteria or quantitative performance results.

Acceptance Criteria CategoryReported Device Performance
Functional PerformanceMet or exceeded
Electrical Safety (IEC 60601-1)Met or exceeded
Electromagnetic Compatibility (IEC 60601-1-2, IEC 60601-2-40)Met or exceeded
Biocompatibility (ISO 10993-1)Met or exceeded
UsabilityMet or exceeded
Battery LifeMet or exceeded
Tensile Strength of ElectrodeMet or exceeded
Stimulation Performance of Electrode (Charge Density, Current Density, Power Density)Met or exceeded

Study Information (Based on available text):

  1. Sample size used for the test set and the data provenance:
    This information is not provided in the document. The text only lists the types of performance tests conducted, but not the specific methodologies, sample sizes, or data provenance (e.g., country of origin, retrospective/prospective nature).

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This information is not provided in the document. The performance tests mentioned are primarily technical and engineering-focused (e.g., electrical safety, biocompatibility), which typically rely on standardized measurements rather than expert-established ground truth in the context of clinical interpretation.

  3. Adjudication method for the test set:
    This information is not provided in the document.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This information is not applicable/provided. The device is an electrical peripheral nerve stimulator for monitoring neuromuscular blockade, not an AI-assisted diagnostic imaging tool with human readers. No MRMC study or AI-assistance evaluation is mentioned.

  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
    This information is not applicable/provided in the context of an algorithm's performance. The device itself (STIMPOD NMS450 with the new EMG cable) performs measurements and stimulation; its "performance" is evaluated through the technical tests listed.

  6. The type of ground truth used:
    For the technical performance tests mentioned (functional, electrical safety, EMC, biocompatibility, battery life, tensile strength, stimulation parameters), the ground truth would be established by engineering standards, specifications, and physical measurements rather than clinical ground truth like pathology or outcomes data. The document states that the device "met or exceeded the stated acceptance criteria," implying these criteria are based on defined technical standards and specifications.

  7. The sample size for the training set:
    This information is not applicable/provided. The STIMPOD NMS450 is a hardware device with an accessory. There is no mention of a "training set" in the context of machine learning or AI models, as this is not an AI/ML powered device.

  8. How the ground truth for the training set was established:
    This information is not applicable/provided as there is no mention of a training set or AI/ML components.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".

October 16, 2022

Xavant Technology (Pty) Ltd Roche Van Rensburg Chief Executive Officer Unit 102 Tannery Industrial Park, 309 Derdepoort Road Silverton Pretoria, Gauteng 0184 South Africa

Re: K213049

Trade/Device Name: STIMPOD NMS450 Nerve Stimulator Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: Class II Product Code: BXN Dated: September 1, 2022 Received: September 12, 2022

Dear Roche Van Rensburg:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

James J. Lee, Ph.D. Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthsia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213049

Device Name STIMPOD NMS450 Nerve Stimulator

Indications for Use (Describe)

This device is a nerve stimulation device designed to be used by an anesthetist during:

  1. General anesthesia, for the purpose of establishing the efficacy of a Neuromuscular Blocking Agent using non-invasive surface electrodes.

  2. Regional anesthesia, for the purpose of:

i. nerve mapping using the non-invasive Nerve Mapping Probe (supplied)

ii. nerve locating using invasive needles (not supplied)

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a logo for Xavant Technology. The logo features a stylized letter X with a circular orbit around it. The word "AVANT" is in red, and the word "TECHNOLOGY" is in gray below it.

This 510(k) Summary was prepared in accordance with the requirements of 21CFR807.92

5.1 Date Prepared

08 September, 2022

5.2 Submitter's Information

Company Name:XAVANT TECHNOLOGY (PTY) LTD
Street Address:Unit 102 Tannery Industrial Park309 Derdepoort RoadSilverton
City:Pretoria
State/Province:Gauteng
Country:South Africa
Telephone:+27(0) 12 743 5959
Fax:+27(0) 86 547 0026
Contact Person:Roche Janse van Rensburg
Contact Title:Chief Executive Officer
Contact Email:compliance@xavant.com

5.3 Trade Name, Common Name, Classification

Type of 510(k) submissionTraditional
Trade Name:STIMPOD NMS450 Nerve Stimulator
Common Name:Battery Powered Peripheral Nerve Stimulator
Classification Name:Battery Powered Nerve Stimulator
per 21 CFR § 868.2775
Regulatory Class:Class II
Product Code:BXN
Panel:Anesthesiology
Reason for SubmissionNew Accessory

{4}------------------------------------------------

5.4 Identification of Predicate/Reference Device(s)

PREDICATE DEVICE
Xavant Technology, STIMPOD NMS450(K102084)
This predicate device has not been subject to a design-related recall.
REFERENCE DEVICE
Senzime, Tetragraph(K190795)
This reference device has not been subject to a design-related recall.

5.5 Description of the Device

The STIMPOD NMS450 is a quantitative Neuromuscular Transmission (NMT) Monitor which provides real-time quantitative feedback utilizing tri-axial accelerometery.

The STIMPOD NMS450 is also a precision nerve locating tool used for localizing specific neural pathways. Localization of nerves by electrical stimulation involves connecting the nerve stimulator to a conducting needle through which local anaesthetics can be injected. The distance of the needle (cathode) from the nerve can be estimated by establishing the minimum threshold current required, to facilitate a neuromuscular response.

The STIMPOD NMS450 currently makes use of 3 dimensional acceleromyography (AMG) as the measurement technology of choice for objective Neuromuscular Transmission (NMT) monitoring in surgery. Although AMG is still the de-facto standard in industry, the use of electromyography (EMG) has gained renewed interest in NMT monitoring applications where the introduction of robotic surgeries, new surgical techniques, and changes in workflow, for instance, made the use of AMG impracticable.

In order to address this steadily increasing market demand for EMG sensor technology in NMT monitoring applications, Xavant identified the need to design and develop a dedicated NMT Monitoring Cable (EMG) that could be used interchangeably with the existing NMT Monitoring Cable (AMG) already in use with the STIMPOD NMS450.

The NMT Monitoring Cable (EMG) was specifically designed for use with the STIMPOD NMS450 to assist anaesthesiologists in theatre with monitoring the efficacy of Neuromuscular Blocking Agents (NMBAs).

The NMT Monitoring Cable (EMG) was designed with an intelligent EMG signal processing module (henceforth referred to as the µEMG) on one end of the cable for connection to an EMG electrode, and with a standard FireWire connector on the other end for connection to the STIMPOD NMS450.

{5}------------------------------------------------

The µEMG makes provision for the following interfaces:

    1. a wireline communications interface to exchange information with the STIMPOD NMS450,
    1. an electrical interface to condition and sample the EMG signals picked up by the measurement contact points of an attached EMG electrode array,
    1. an electrical interface to (i) detect the stimulation signals generated by the STIMPOD NMS450, and (ii) route them to the stimulation contact points of an attached EMG electrode array,
    1. a visual interface to indicate the state (powered, connected etc.) of the device.

5.6 Substantial Equivalence Discussion

The following device comparisons are made:

    1. The STIMPOD NMS450 to the predicate device STIMPOD NMS450. STIMPOD NMS450's 510(k) clearance (K102084) means it is legally marketed;
    1. The STIMPOD NMS450 to the reference device Tetragraph Neuromuscular Transmission Monitor from Senzime AB. Tetragraph's 510(k) clearance (K190795) means it is legally marketed.

Intended Use

FDA product Code of the predicate device and that of the subject device is given as "BXN" while that of the reference device is given as "KOI" - both products codes, however, are assigned under Regulation No. 21 CFR 868.2775 which applies to "Electrical Peripheral Nerve Stimulators" with the following intended use:

"An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases".

Indications for Use

This device is a nerve stimulation device designed to be used by an anaesthetist during:

    1. General anaesthesia, for the purpose of establishing the efficacy of a Neuromuscular Blocking Agent using non-invasive surface electrodes.
    1. Regional anaesthesia, for the purpose of:
    • i. Nerve mapping using the non-invasive Nerve Mapping Probe (supplied)
    • ii. Nerve locating using invasive needles (not supplied).

The subject and predicate devices have identical Indications for Use. The subject and reference devices are compared with respect to their mutual indications only. Both the devices are indicted for monitoring the efficacy (which can be defined as "the relaxation of the patient" as indicated in the reference device Indications for Use statement) of a Neuromuscular Blocking Agent.

Intended User and Patient Population

The intended user for both the subject device and predicate device is identical:

  • The subject device and predicate device are both indicated for use by medical . professionals with knowledge of anatomy

{6}------------------------------------------------

The patient population for both the subject device and predicate device is similar:

  • . The reference device is for use with patients (excluding neonates). For the subject device, the patient population includes patients of all ages, weight, and nationality (excluding neonates for electromyography).

Substantial Equivalence Table

To demonstrate substantial equivalence with respect to intended use, technological characteristics, and principles of operation, two tables of Comparison of Characteristics are presented. The first table shows a comparison between the STIMPOD NMS450 (subject device) and the predicate device. The second table shows a comparison between the STIMPOD NMS450 (subject) device and the reference device for those characteristics where the subject device and predicate device had differences. The tables are accompanied by a discussion of all similarities and differences between the devices and therefore provide sufficient detailed information regarding the basis for the determination of substantial equivalence.

CharacteristicSubject DevicePredicate DeviceDiscussion
Trade NameSTIMPOD NMS450STIMPOD NMS450Identical
Product CodeBXNBXNIdentical
Device ClassClass IIClass IIIdentical
ClassificationNameElectrical Peripheral NerveStimulatorElectrical Peripheral NerveStimulatorIdentical
Regulation No.886.2775886.2775Identical
ClassificationPanelAnaesthesiologyAnaesthesiologyIdentical
Intended UseAn electrical peripheral nervestimulator (neuromuscularblockade monitor) used to applyan electrical current to a patientto test the level ofpharmacological effectof anesthetic drugs and gasesAn electrical peripheral nervestimulator (neuromuscularblockade monitor) used to applyan electrical current to a patientto test the level ofpharmacological effectof anesthetic drugs and gasesIdentical
Indication forUseThis device is a nerve stimulationdevice designed to be used byan anaesthetist during:1. general anesthesia, for thepurpose of establishing theefficacy of a NeuromuscularBlocking Agent using non-invasive surface electrodes.This device is a nerve stimulationdevice designed to be used byan anaesthetist during:1. general anesthesia, for thepurpose of establishing theefficacy of a NeuromuscularBlocking Agent using non-invasive surface electrodes.Identical
CharacteristicSubject DevicePredicate DeviceDiscussion
2. regional anesthesia, for thepurpose of:i. nerve mapping using thenon-invasive NerveMapping Probe (supplied)ii. nerve locating usinginvasive needles (notsupplied)2. regional anesthesia, for thepurpose of:i. nerve mapping using thenon-invasive Nerve MappingProbe (supplied)ii. nerve locating usinginvasive needles (notsupplied)Identical
DeviceComponents1. Nerve Stimulator/si. STIMPOD NMS4501. Nerve Stimulator/si. STIMPOD NMS450Identical
2. Patient Cable/si. NMT Monitoring Cable (AMG)ii. Nerve Locating Cableiii. Nerve Mapping/Locating Cable2. Patient Cable/si. NMT Monitoring Cable (AMG)ii. Nerve Locating Cableiii. Nerve Mapping/Locating CableIdentical
3. Applied Part/si. AMG Sensor3. Applied Part/si. AMG SensorIdentical
4. Power Cable/sNone.4. Power Cable/sNone.Identical
WaveformMonophasic, constant current,square pulseMonophasic, constant current,square pulseIdentical
Pulse Width200μS200μSIdentical
StimulationPatternsWaveform: Monophasic,constant current, square pulsewith a fixed pulse width of200msPulse Repetition Frequency:i. TOF: 4 pulses at 2Hzii. TWI(Twitch): 1Hz, 2Hz,5Hziii. Tetanic Stimulus: 50Hzor 100Hziv. PTC Stimulation: 50HzTetanic stimulus for 5seconds, followed by 20TWI pulses at 1HzWaveform: Monophasic,constant current, square pulsewith a fixed pulse width of200msPulse Repetition Frequency:i. TOF: 4 pulses at 2Hzii. TWI(Twitch): 1Hz, 2Hz,5Hziii. Tetanic Stimulus: 50Hzor 100Hziv. PTC Stimulation: 50HzTetanic stimulus for 5seconds, followed by 20TWI pulses at 1HzIdentical
CurrentCharacteristicsElectrode/Skin Resistance:0-5000 ohm maximumCurrent: 5-80mA peakVoltage: 0 - 400V peak,depending on skin impedanceElectrode/Skin Resistance:0-5000 ohm maximumCurrent: 5-80mA peakVoltage: 0 - 400V peak,depending on skin impedanceIdentical
ControlMechanismDigitally controlled by amicroprocessorDigitally controlled by amicroprocessorIdentical
CharacteristicSubject DevicePredicate DeviceDiscussion
Energy TypeBattery:4 X AAA alkaline batteries, 6V,1.2AhCharger:None.Battery:4 X AAA alkaline batteries, 6V,1.2AhCharger:None.Identical
NMT MonitoringCable (AMG)Type: ReusableType: Reusable
Communications: STIMPODproprietary.Communications: STIMPODproprietary.Identical
Attachment: Integral frictionstrap.Attachment: Integral frictionstrap.
NMT MonitoringCable (EMG)Type: ReusableCommunications: STIMPODproprietary.EMG Electrode Connector:Proprietary frictional & magneticRetention mechanism.Not applicableThe predicatedevice doesnot includethe NMTMonitoringCable (EMG)as this is anew addition.Comparisonprovided withreferencedevice.
EMG ElectrodeType: Self-adhesive, Single UseEvoked Response Monitoring:Electromyography.Form Factor: 2 Stimulationcontact pads and 2 recordingcontact pads.Ambidexterity: Usable on leftand right hand.Application Site: Distalstimulation of the Ulnar Nervewith CMAP recording of theAdductor Pollicis muscle.Not applicableThe predicatedevice doesnot includethe EMGElectrode asthis is a newaddition.Comparisonprovided withreferencedevice.

Comparison between Subject Device and Predicate Device

{7}------------------------------------------------

{8}------------------------------------------------

Comparison between Subject Device and Reference Device

CharacteristicSubject DeviceReference DeviceDiscussion
Trade NameSTIMPOD NMS450TetragraphTrade names aredifferent.
Product CodeBXNKOISimilar. No newquestions ofsafety andeffectivenessraised.
Device ClassClass IIClass IIIdentical
ClassificationNameElectrical Peripheral NerveStimulatorElectrical Peripheral NerveStimulatorIdentical
CharacteristicSubject DeviceReference DeviceDiscussion
Regulation No.886.2775868.2775Identical
ClassificationPanelAnaesthesiologyAnesthesiologyIdentical
DeviceComponents1. Patient Cable/si. NMT Monitoring Cable (EMG)1. Patient Cable/si. TetraCord EMG Patient CableSimilar. No newquestions ofsafety andeffectivenessraised.
2. Applied Part/sEMG Electrode2. Applied Part/si. TetraSense EMG ElectrodeSimilar. No newquestions ofsafety andeffectivenessraised.
NMT MonitoringCable (EMG)Type: ReusableType: ReusableBoth cables arereusable.
Communications: STIMPODproprietary.Communications: TetraGraphproprietary
EMG Electrode Connector:Proprietary frictional &magnetic Retentionmechanism.EMG Electrode Connector:Proprietary frictional Retentionmechanism.See discussionof differencesbelow.
EMG ElectrodeType: Self-adhesive, SingleUseType: SelfAdhesive, Single Use
Evoked ResponseMonitoring:Electromyography.Evoked ResponseMonitoring:ElectromyographyThe type,measuringtechnology, formfactor andapplication siteare identical.
Form Factor: 2 Stimulationcontact pads and 2 recordingcontact pads.Form Factor: 2 Stimulationcontact pads and 2 recordingcontact pads.
Ambidexterity: Usable on leftand right hand.Ambidexterity: Usable on leftand right hand.See discussionof differencesbelow.
Application Site: Distalstimulation of the Ulnar Nervewith CMAP recording of theAdductor Pollicis muscle.Application Site: Distalstimulation of the Ulnar Nervewith CMAP recording of theAdductor Pollicis muscle.

{9}------------------------------------------------

Discussion of differences:

NMT Monitoring

    1. The specific means of communication that is used by the subject device to transfer measured EMG data from the electrode can be achieved in multiple different ways, and therefor does not need to be the same as that being used by the reference device.
    1. The materials and construction of the NMT Monitoring Cable (EMG) used for the subject device differs from that of the EMG cable used for the reference device. As above, the patient cable for the two devices can be built from multiple different materials and

{10}------------------------------------------------

constructed in multiple different ways, and therefore do not need to be the same to achieve the same intended purpose.

    1. The retention mechanism used in the construction of the patient cables for both devices is different, but the retention force offered by the connector used for the subject device is better or larger than that of the connector used for the reference device.

EMG Electrode

    1. The materials used in the construction of the EMG electrodes of the two devices are different, however, both electrodes satisfy the biocompatibility requirements of IEC 10993-1.
    1. The EMG electrodes for both devices furthermore satisfy the electrical and adhesion performance requirements of AAMI EC12.
  • Finally, the EMG electrode of the subject device exhibits same or better safety and 6. performance characteristics (charge density current density and power density) at the skin/electrode interface than that of the reference device.

The comparative analysis presented above does not raise any new questions in terms of risk and performance of the subject device, and is supported by the performance data captured under paragraph 5.7.

5.7 Performance Testing

The following performance tests were conducted on the subject device in support of the claim for substantial equivalence:

    1. Functional Performance Testing.
    1. Electrical safety (in accordance with IEC 60601-1).
    1. Electromagnetic compatibility (in accordance with IEC 60601-1-2 and IEC 60601-2-40).
    1. Biocompatibility testing (in accordance with ISO10993-1).
    1. Usability testing.
    1. Battery life testing.
    1. Tensile strength testing of the electrode.
    1. Stimulation Performance Testing of the electrode (Charge Density, Current Density and Power Density)

The subject device met or exceeded the stated acceptance criteria for each performance test which were all confirmed to be the same or very similar to the performance specifications of the predicate device, and/or where relevant, to the reference device. The resultant test data therefore provided sufficient evidence to support the claim that the subject device is at least as safe and as effective as the predicate device in monitoring the muscle relaxation of a patient that has been subjected to the administration of a Neuromuscular Blocking Agent.

5.8 Clinical Performance

No clinical testing was conducted to demonstrate that the subject device is substantially equivalent to the predicate device, since the two devices were already proven to have similar

{11}------------------------------------------------

technological characteristics. There was no clinical testing conducted to demonstrate that the subject was substantially equivalent to the reference device, as the reference device has similar technological characteristics.

5.9 Conclusion (Statement of Equivalence)

A detailed comparative analysis involving the identified subject, predicate and references devices, has provided sufficient evidence in support of the following conclusion:

The subject device when compared to the predicate device and reference device has the same intended use, user profile and patient population, and the same or very similar technological characteristics. Differences in the technological characteristics of the two devices did not raise any new questions of safety and effectiveness, and the resultant test data from the performance tests, demonstrated that the subject device is at least as safe and effective as the reference device in terms of the use of EMG in NMT Monitoring applications.

There are no significant differences between the STIMPOD NMS450 Nerve Stimulator and the predicate and reference devices that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.

§ 868.2775 Electrical peripheral nerve stimulator.

(a)
Identification. An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.(b)
Classification. Class II (performance standards).