K Number
K212928
Device Name
Lorals
Date Cleared
2022-04-22

(220 days)

Product Code
Regulation Number
884.5300
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Lorals are used as a barrier when engaging in oral/anal sex to help reduce the transmission of bodily fluids, harmful pathogens, and sexually transmitted infections.

Device Description

Lorals are wearable, single-use, natural rubber latex oral dams that are placed over the vagina and/or anus during oral sexual contact. Lorals are designed similar to underwear, with a region that covers the genital region and a waistband to keep the dam in place, all made from natural rubber latex. The dams are available in two designs, Shortie and Bikini, which provide the same coverage of the genital region, but have different waistband heights, based on user preference. Each design is available in 3 different gusset sizes. The dams are scented with vanilla fragrance and are provided in black only.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called "Lorals," which are oral dams. This submission is for demonstrating substantial equivalence to a predicate device, not for proving an AI device meets acceptance criteria. Therefore, most of the requested information regarding acceptance criteria and study details for an AI device is not applicable to this document.

However, I can extract the information related to the performance testing conducted for the Lorals device, which can be thought of as "acceptance criteria" for a non-AI medical device in the context of demonstrating substantial equivalence.

Here's the relevant information:

1. A table of acceptance criteria and the reported device performance:

Acceptance Criteria (Test)StandardReported Device Performance (Outcome)
Dimensional TestingISO 29942:2011Results acceptable
Tensile Testing and Elongation at BreakISO 29942:2011Results acceptable
Tear Resistance and Tearing ForceISO 29942:2011Results acceptable
Freedom from Holes and Visual DefectsISO 29942:2011Results acceptable
Packaging Integrity Testing(Implied by ISO 29942:2011 and shelf-life support)Results acceptable
CytotoxicityISO 10993-5:2009Results acceptable
Irritation TestingISO 10993-10:2010Results acceptable
SensitizationISO 10993-10:2010Results acceptable
Acute Systemic ToxicityISO 10993-11:2017Results acceptable

2. Sample size used for the test set and the data provenance:

  • Sample size: Not explicitly stated. The document mentions "Performance testing was performed on the subject device per the FDA recognized standard ISO 29942:2011" and similar for biocompatibility. These standards typically specify sample sizes, but the exact numbers used for Lorals are not in this summary.
  • Data provenance: Not specified. It's bench testing, so it's not patient data and would be prospective in nature.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This relates to a human-in-the-loop or AI evaluation, which is not described for this device. The testing is laboratory-based physical and biocompatibility testing.

4. Adjudication method for the test set:

  • Not applicable. This relates to interpretation of clinical data by multiple experts, which is not relevant here.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI device.

7. The type of ground truth used:

  • For physical properties (tensile strength, tear resistance, freedom from holes, dimensions, etc.): Objective measurements defined by the ISO 29942:2011 standard.
  • For biocompatibility: Laboratory test results evaluated against the criteria established in the ISO 10993 series of standards.

8. The sample size for the training set:

  • Not applicable. This section is for an AI device.

9. How the ground truth for the training set was established:

  • Not applicable. This section is for an AI device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.

April 22, 2022

Brazen Goods Inc. dba Lorals % Erin Gontang Regulatory Specialist AcKnowledge Regulatory Strategies LLC 2251 San Diego Avenue, B-257 San Diego, CA 92110

Re: K212928 Trade/Device Name: Lorals Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: MSC, Dated: March 21, 2022 Received: March 22, 2022

Dear Erin Gontang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212928

Device Name Lorals

Indications for Use (Describe)

Lorals are used as a barrier when engaging in oral/anal sex to help reduce the transmission of bodily fluids, harmful pathogens, and sexually transmitted infections.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

DATE PREPARED

April 21, 2022

MANUFACTURER AND 510(k) OWNER

Brazen Goods Inc. 106 1/2 Judge John Aiso St. #305, Los Angeles, CA 90012, USA Telephone: Official Contact: Melanie Cristol, Founder & CEO

REPRESENTATIVE/CONSULTANT

Erin A. Gontang, Ph.D. Allison C. Komiyama, Ph.D., RAC RQM+ Corp. Telephone: +1 (412) 816-8253 Email: egontang@rqmplus.com akomiyama@rqmplus.com Website: www.rqmplus.com

DEVICE INFORMATION

Proprietary Name/Trade Name:Lorals
Common Name:Oral Dam
Regulation Number:21 CFR 884.5300
Regulation Name:Condom
Regulatory Class:II
Product Code:MSC (Barrier, Std, Oral Sex)

PREDICATE DEVICE IDENTIFICATION

Lorals are substantially equivalent to the following predicate:

510(k) NumberPredicate Device Name / ManufacturerPrimary Predicate
K091769Trust Dam/Line One Laboratories, Inc.

The predicate device has not been subject to a design related recall.

DEVICE DESCRIPTION

Lorals are wearable, single-use, natural rubber latex oral dams that are placed over the vagina and/or anus during oral sexual contact. Lorals are designed similar to underwear, with a region that covers the genital region and a waistband to keep the dam in place, all made from natural rubber latex. The dams are available in two designs, Shortie and Bikini, which provide the same coverage of the genital region, but have different waistband heights, based on user preference.

{4}------------------------------------------------

Each design is available in 3 different gusset sizes. The dams are scented with vanilla fragrance and are provided in black only.

INDICATIONS FOR USE

Lorals are used as a barrier when engaging in oral/vaginal sex and oral/anal sex to help reduce the transmission of bodily fluids, harmful pathogens, and sexually transmitted infections.

COMPARISON OF INDICATIONS FOR USE AND TECHNOLOGICAL CHARACTERISTICS

The intended use of the subject device, Lorals, and the predicate device, Trust Dams (K091769), are the same. Both the subject device and the predicate device are used as a barrier when engaging in oral/vaginal sex and oral/anal sex. Table 1 below compares the technological characteristics of the subject device and the predicate device.

Lorals(Subject Device)Trust Dam(Predicate Device)
Device NameLoralsTrust Dams
ManufacturerBrazen Goods Inc.Line One Laboratories Inc
510(k) NumberK212928K091769
Regulatory ClassIIII
Regulation Number21 CFR 884.530021 CFR 884.5300
Product CodeMSC (Barrier, Std, Oral Sex)MSC (Barrier, Std, Oral Sex)
Regulation NameCondomCondom
Indications for UseLorals are used as a barrierwhen engaging in oral/vaginalsex and oral/anal sex to helpreduce the transmission ofbodily fluids, harmfulpathogens, and sexuallytransmitted infections.The Line One Trust Dam isused as a barrier whenengaging in oral/vaginal sexand oral/anal sex to helpreduce the transmission ofbodily fluids, harmfulgerms, and sexuallytransmitted diseases.
MaterialNatural rubber latexNatural rubber latex
Location of UseAnus, vagina, oral cavity(mouth)Anus, vagina, oral cavity(mouth)
LubricatedNoNo
ColorBlackYellow, Red, Green
Fragrance/FlavorVanillaBanana, Strawberry, Mint
Design & DimensionsRectangular sheet withintegrated waistbandGusset width: 163 mm,Rectangular sheetWidth: 154 mm

Table 1 Comparison of Technological Characteristics of the Subject and Predicate Devices.

{5}------------------------------------------------

188 mm, or 213 mmHeight: 250 mm
Gusset Height: 228 mm
Thickness: 0.07 mmThickness: 0.07 mm
Lorals Height: 264 mm
Waistband Width: 343 mm
Waistband Height:
Bikini: 61 mm
Shortie: 162 mm
Mechanical PropertiesTested in accordance with ISO29942:2011, including thefollowing tests:Tensile strengthElongation at breakTearing forceTear resistanceFreedom from holesTested in accordance withASTM D3492:2008
SterilityNon-SterileNon-Sterile
Shelf-life2 years5 years
Biocompatibility Tested?YesYes

The intended use of Lorals is the same as the Trust Dam. Differences in technological characteristics include the dimensions, shape, packaging, color and flavor additives, and shelf life. However, these differences between the subject and predicate device, as described in Table 1, do not raise different questions of safety or effectiveness.

SUMMARY OF NON-CLINICAL TESTING

Performance testing was performed on the subject device per the FDA recognized standard ISO 29942:2011 Prophylactic dams –Requirements and test methods. The following testing was performed:

  • Dimensional Testing
  • Tensile Testing and Elongation at Break ●
  • Tear Resistance and Tearing Force
  • Freedom from Holes and Visual Defects ●
  • Packaging Integrity Testing

The bench testing was completed as outlined in the standard to support the 2-year shelf life.

As required, biocompatibility of the device was assessed per the recommendations outlined in the FDA guidance document Use of International Standard ISO 10993-1 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process." The following testing was performed:

{6}------------------------------------------------

  • Cytotoxicity (ISO 10993-5:2009)
  • . Irritation Testing (ISO 10993-10:2010)
  • . Sensitization (ISO 10993-10:2010)
  • Acute Systemic Toxicity (ISO 10993-11:2017) ●

All test results were acceptable.

CONCLUSION

The results from the non-clinical bench testing above demonstrate that Lorals dams are as safe and effective as the predicate device and support a determination of substantial equivalence.

§ 884.5300 Condom.

(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.