(431 days)
The Konig Mogen Clamp is an instrument used in a medical procedure to compress the foreskin of the penis during the circumcision of a male infant or child only.
The Konig Mogen Clamp is an instrument used in a medical procedure to compress the foreskin of the penis during the circumcision of a male infant or child only.
The Konig Mogen Clamp consists of an arm with a locking mechanism in order facilitate compression on the foreskin. The locking mechanism consists of a lock connected to a lock bar using a pin. The clamp is fixed with a spot welding point, which allows for a maximum opening of 2.5 mm. The edges of the closing area are chamfered and rounded to avoid tissue injury.
The Konig Mogen Clamp is a reusable instrument that is provided non-sterilyzed by the user. The Konig Mogen Clamp will be validated for use in pre-vacuum steam sterilization.
The provided text describes a 510(k) premarket notification for a medical device called the Konig Mogen Clamp. This submission is for demonstrating substantial equivalence to a predicate device, not for establishing clinical efficacy or safety from scratch through the rigorous testing typically performed for novel devices or AI/software as a medical device (SaMD).
Therefore, the document does not contain the information requested about acceptance criteria, study details (sample size, data provenance, expert ground truthing, MRMC studies, standalone performance, training set details) that would be relevant for an AI/SaMD device.
The "study that proves the device meets the acceptance criteria" in this context refers to non-clinical bench testing for physical performance, biocompatibility, and cleaning/sterilization validation, as detailed in the "Summary of Non-Clinical Testing" section.
Here's a breakdown based on the provided document, highlighting why most of the requested information is not applicable:
1. Table of Acceptance Criteria and Reported Device Performance:
The document provides a summary of non-clinical tests. The "Acceptance Criteria" for this type of device are typically met by passing standard international and ASTM (American Society for Testing and Materials) tests for surgical instruments. The "Reported Device Performance" is stated as "Test articles met the acceptance criteria" and "The Konig Mogen Clamp was non-toxic," and "No growth at the end of the incubation period." Specific numerical results or detailed acceptance ranges are not provided in this summary.
| Study | Description | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|---|
| Biocompatibility | ISO 10993-5 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro CytotoxicityISO 10993-17:2002 Biological evaluation of medical devices — Part 17: Establishment of allowable limits for leachable substances | Must demonstrate non-toxicity and allowable limits for leachable substances as per ISO 10993 standards. | "The Konig Mogen Clamp was non-toxic." (Implies compliance with ISO 10993) |
| Cleaning & Sterilization | ISO 17665-1:2006 Moist Heat - Req. for the development, validation, and routine control of a sterilization process for medical devices.ISO 15883-2:2006 Req. and tests for washer-disinfectors employing thermal disinfection for surgical instruments (Used for the cleaning and thermal disinfection for reusable surgical instruments)ANSI/AAMI ST79: 2017 Guide to steam sterilization and sterility assurance in health care facilities | Must be effectively cleaned and sterilized to achieve sterility assurance level (SAL). No microbial growth after sterilization. | "No growth at the end of the incubation period." (Confirms successful sterilization to required standards) |
| Performance Testing | Custom Grip TestASTM E322-12 Standard Test Method For Analysis Of Low-Alloy Steels And Cast Irons By Wavelength Dispersive X-Ray Fluorescence SpectrometryASTM F1089-10 Standard Test Method for Corrosion of Surgical InstrumentsASTM E18 - 20 Standard Test Methods for Rockwell Hardness of Metallic Materials | Material composition, grip force, corrosion resistance, and hardness must meet specified ASTM standards for surgical instruments to ensure durability and proper function. | "Test articles met the acceptance criteria." (Confirms compliance with specified ASTM standards and custom grip test requirements, although specific values are not provided) |
2. Sample size used for the test set and the data provenance:
This document describes non-clinical bench testing for a physical surgical instrument. The "test set" would refer to the number of physical clamps subjected to the various physical, chemical, and sterilization tests. The document does not specify the exact number of units tested.
- Sample Size: Not specified for non-clinical tests. Medical device bench testing for performance and cleaning/sterilization typically involves a representative sample size of devices, often in the range of 3-30 units, depending on the test and statistical considerations.
- Data Provenance: Not applicable in the sense of patient data. The tests are performed on manufactured devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. "Ground truth" in the context of this device's testing refers to objective measurements against engineering standards (e.g., material composition, hardness, successful sterilization, resistance to corrosion) rather than expert interpretation of patient data. These tests are performed by technicians or engineers.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. This refers to establishing consensus among human readers for diagnostic tasks, which is not relevant for this type of device's non-clinical testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This type of study is relevant for AI or diagnostic imaging devices that assist human readers. The Konig Mogen Clamp is a manual surgical instrument.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This refers to the performance of an AI algorithm, not a manual surgical instrument.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this device's non-clinical testing is based on:
- Pre-defined engineering specifications and ASTM/ISO standard requirements (e.g., chemical composition, hardness values, corrosion resistance, maximum opening, sterilization success).
- Objective measurements obtained through laboratory testing (e.g., spectral analysis for material, physical force measurements, microbial growth assays).
8. The sample size for the training set:
Not applicable. This device is a mechanical instrument, not an AI/machine learning model that requires a training set.
9. How the ground truth for the training set was established:
Not applicable, for the same reason as #8.
In summary: The provided FDA 510(k) clearance document details the substantial equivalence of a manual surgical instrument and, therefore, focuses on non-clinical performance, safety (biocompatibility), and sterilization validation against established industry standards. It does not involve the types of clinical studies, AI performance metrics, or ground truth establishment relevant for an AI/SaMD device that the questions are implicitly structured to address.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 18, 2022
Medline Industries, Inc. Jennifer Mason Regulatory Affairs Principal Three Lakes Drive Northfield, IL 60093
Re: K212911
Trade/Device Name: Konig Mogen Clamp Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: HFX Dated: October 7, 2022 Received: October 11, 2022
Dear Jennifer Mason:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark J. Antonino -S
Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K212911
Device Name Konig Mogen Clamp
Indications for Use (Describe)
The Konig Mogen Clamp is an instrument used in a medical procedure to compress the foreskin of the penis during the circumcision of a male infant or child only.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Medline. The logo consists of the word "MEDLINE" in white, stacked vertically, with a stylized white starburst symbol above it. The background is a solid dark blue color. The logo is simple and clean, with a modern and professional look.
Medline Industries, Inc. Three Lakes Drive Northfield. IL 60093
K212911 Page 1 of 5
510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]
Submitter / 510(k) Sponsor
Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093 Registration Number: 1417592
Contact Person
Jenny Ly Regulatory Affairs Specialist Phone: 224-234-7457 Email: jly(@medline.com
Summary Preparation Date September 10, 2021
Type of 510(k) Submission Traditional
Device Name / Classification
Name of Device: Konig Mogen Clamp Proprietary Name: Konig Mogen Clamp Common Name: Circumcision Clamp Classification Name: Clamp, Circumcision Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Product Code: HFX Classification Panel: Obstetrics/Gynecology Regulatory Class: Class II Regulation #: 21 CFR 884.4530
Predicate Device
Medicon Mogen Circumcision Clamp K100916
Device Description
The Konig Mogen Clamp is an instrument used in a medical procedure to compress the foreskin of the penis during the circumcision of a male infant or child only.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, sans-serif font, stacked on top of a white, stylized starburst shape. The background is a solid dark blue color. The logo is simple and clean, with a focus on the company name and a visual element that suggests energy or innovation.
Medline Industries, Inc. Three Lakes Drive Northfield. IL 60093
K212911 Page 2 of 5
The Konig Mogen Clamp consists of an arm with a locking mechanism in order facilitate compression on the foreskin. The locking mechanism consists of a lock connected to a lock bar using a pin. The clamp is fixed with a spot welding point, which allows for a maximum opening of 2.5 mm. The edges of the closing area are chamfered and rounded to avoid tissue injury.
The Konig Mogen Clamp is a reusable instrument that is provided non-sterilyzed by the user. The Konig Mogen Clamp will be validated for use in pre-vacuum steam sterilization.
Indications for Use
The Konig Mogen Clamp is an instrument used in a medical procedure to compress the foreskin of the penis during the circumcision of a male infant or child only.
The Konig Mogen Clamp is intended for use in healthcare facilities, including hospitals, medical clinics and surgical centers, by surgeons of proper training and experience.
Summary of Technological Characteristics
| Device Characteristic | Proposed Device | Predicate Device | ComparisonAnalysis |
|---|---|---|---|
| Product Name | Konig Mogen Clamp | Medicon MogenCircumcision Clamp | Different |
| 510(k) Reference | N/A | K100916 | Different |
| Product Owner | Medline Industries, Inc. | Medicon | Different |
| Product Code | HFX | HFX | Same |
| Intended Use | The Konig Mogen Clampis an instrument used in amedical procedure tocompress the foreskin ofthe penis duringcircumcision of a maleinfant or child. | The Mogen CircumcisionClamp is an instrument usedin a medical procedure tocompress the foreskin of thepenis during circumcision ofa male infant or child. | Same |
| Regulation Number | 21 CFR 884.4530 | 21 CFR 884.4530 | Same |
| Design Features | Clamp and locking bar | Clamp and locking bar | Same |
| Design Configurations | In the open position, thejaws are designed toseparate to a maximum of2.5 mm. For safe clamping | In the open position, the jawsare designed to separate to amaximum of 2.5 mm. Forsafe clamping the Medicon | Same |
| the Konig Mogen Clampis locked with a definedclosing force. For safetyreasons, the lock screw ofthe clamp is fixed with aspot welding point inorder not to become loose. | Mogen Clamp is locked witha defined closing force. Forsafety reasons, the lockscrew of the clamp is fixedwith a spot welding point inorder not to become loose. | ||
| Materials | Stainless Steel - corrosionresistant | Stainless Steel- corrosionresistant | Same |
| PerformanceSpecifications | Standards ASTM F 899-09 and ISO 7153-1Standard Specification forStainless Steels forSurgical InstrumentsASTM E322-12 StandardTest Method For AnalysisOf Low-Alloy Steels AndCast Irons By WavelengthDispersive X-RayFluorescenceSpectrometryASTM F1089-10 StandardTest Method for Corrosionof Surgical InstrumentsASTM E18 - 20 StandardTest Methods forRockwell Hardness ofMetallic Materials | Standards ASTM F 899-09and ISO 7153-1 StandardSpecification for StainlessSteels for SurgicalInstrumentsASTM F 1089-10 and ISO13402 StandardSpecifications for resistanceagainst corrosion andautoclaving of SurgicalInstruments. | Different |
| Prescription vs. OTC | Prescription | Prescription | Same |
| Contact Durations | Surface DeviceBreached or compromisedsurfaceLimited duration | Surface DeviceBreached or compromisedsurfaceLimited duration | Same |
| Sterile vs. Non-Sterile | Non-Sterile | Non-Sterile | Same |
| Single Use vs. Reusable | Reusable | Reusable | Same |
| Sterilization Method | Steam sterilization | Steam sterilization | Same |
TARLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in a sans-serif font, with a stylized starburst symbol to the left of the text. The starburst is white, and the background is a solid blue color. The logo is simple and modern, and the colors are clean and professional.
Medline Industries, Inc. Three Lakes Drive
Northfield, IL 60093
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in a sans-serif font, with a stylized white starburst or lightning bolt symbol positioned vertically behind the text. The background is a solid dark blue color. The logo is simple and modern, with a clean design.
Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093
Summary of Non-Clinical Testing
Biocompatibility, cleaning and sterilization, performance, and physical properties testing was conducted on the Konig Mogen Clamp to demonstrate that the product meets its intended use.
| Study | Description | Results |
|---|---|---|
| Biocompatibility | ISO 10993-5 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro CytotoxicityISO 10993-17:2002 Biological evaluation of medical devices — Part 17: Establishment of allowable limits for leachable substances | The Konig Mogen Clamp was non-toxic. |
| Cleaning and Sterilization Testing | ISO 17665-1:2006 Moist Heat - Req. for the development, validation, and routine control of a sterilization process for medical devices.ISO 15883-2:2006 Req. and tests for washer-disinfectors employing thermal disinfection for surgical instruments (Used for the cleaning and thermal disinfection for reusable surgical instruments)ANSI/AAMI ST79: 2017 Guide to steam sterilization and sterility assurance in health care facilities | No growth at the end of the incubation period. |
| Performance Testing | Custom Grip TestASTM E322-12 Standard Test Method For Analysis Of Low-Alloy Steels And Cast Irons By Wavelength Dispersive X-Ray Fluorescence SpectrometryASTM F1089-10 Standard Test Method for Corrosion of Surgical InstrumentsASTM E18 - 20 Standard Test Methods for Rockwell Hardness of Metallic Materials | Test articles met the acceptance criteria. |
Summary of Clinical Testing
Not applicable.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Medline, a healthcare company. The logo consists of the word "MEDLINE" in a bold, sans-serif font, stacked above a white, stylized starburst symbol. The background is a solid blue color, and there is a gray bar on the right side of the image.
Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093
K212911 Page 5 of 5
Conclusion
In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that the Konig Mogen Clamps are as safe and as effective for their intended use as the predicate device, Medicon Mogen Circumcision Clamp, under K100916.
§ 884.4530 Obstetric-gynecologic specialized manual instrument.
(a)
Identification. An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes.
(2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant.
(3) An umbilical clamp is an instrument used to compress the umbilical cord.
(4) A uterine curette is an instrument used to scrape and remove material from the uterus.
(5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix.
(6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus.
(7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination.
(8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix.
(9) A gynecological cerclage needle is a looplike instrument used to suture the cervix.
(10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus.
(11) A gynecological fibroid screw is an instrument used to hold onto a fibroid.
(12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity.
(13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina.
(14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures.
(15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus.
(16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis.
(17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina.
(18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
(b)
Classification. (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.