K Number
K212898

Validate with FDA (Live)

Date Cleared
2021-12-16

(97 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs per ASTM D6978-05(2019) Standard Practice for assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

Blue Nitrile Powder Free Patient Examination Glove, Non-Sterile Tested for use with Chemotherapy Drugs meet all requirements of ASTM standard D6319 and FDA 21 CFR 880.6250. The powder free nitrile examination glove is manufactured from nitrile rubber. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder on the glove surface. The glove is ambidextrous, i.e., can be worn on right hand or left hand.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device (Blue Nitrile Powder Free Patient Examination Glove) and does not describe a study involving an AI/Machine Learning enabled device. Therefore, it does not contain the information requested in your prompt regarding acceptance criteria, sample sizes, expert involvement, or MRMC studies for an AI device.

The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing of physical properties, dimensions, watertightness, residual powder, and permeation to chemotherapy drugs.

However, to answer your request to the best of my ability based on the provided document and assuming the "device" refers to the Blue Nitrile Powder Free Patient Examination Glove, here's an interpretation of the relevant information provided:


Since this document pertains to a medical glove and not an AI/Machine Learning enabled device, the requested information regarding AI-specific criteria (e.g., human-in-the-loop, expert consensus for AI ground truth, MRMC studies) is not applicable and hence, not present in the provided text.

The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the performance standards and non-clinical tests conducted to demonstrate the safety and effectiveness of the medical glove.

1. Table of Acceptance Criteria and Reported Device Performance:

The acceptance criteria are generally based on ASTM standards (e.g., ASTM D6319, ASTM D5151, ASTM D6124) and outlined in the "Summary of Non-Clinical Testing 7.0" section, particularly in the tables on pages 11-14.

CharacteristicStandardAcceptance CriteriaReported Device Performance (Current Blue)Status
Physical PropertiesASTM D412
Before Aging: Tensile Strength-Min 14.0 MPa31.2 MPaPass
Before Aging: Ultimate Elongation-Min 500%594%Pass
After Aging: Tensile Strength-Min 14.0 MPa34.7 MPaPass
After Aging: Ultimate Elongation-Min 400%553%Pass
Dimensions (Example: X-Small)ASTM D3767
Length (X-Small)-Min 240 mm250 mmPass
Width (X-Small)-70 ± 10 mm78.0 mmPass
Thickness Finger (X-Small)-Min 0.05mm0.10 mmPass
Thickness Palm (X-Small)-Min 0.05mm0.07 mmPass
WatertightASTM D5151Sample size: 315 pcs, Inspection level: G1, AQL: 1.5, Acceptance No. 101 piece found with leaks (within acceptance criteria)Pass
Residual PowderASTM D6124Less than 2 mg per glove0.26 mg/glovePass
Chemotherapy Drug PermeationASTM D6978-05(2019) (Minimum Breakthrough Detection Time in Minutes)Varies by drug type (Implicitly, the device must meet or exceed predicate performance or established safety thresholds)Values as listed in "Indications for Use" and "Table 1" (e.g., Carmustine: 23.0 min, Thiotepa: 98.8 min, others > 240 min)Pass (compared to predicate)
BiocompatibilityISO 10993-10, ISO 10993-5, ISO 10993-11Non-irritant, non-sensitizer, and no systemic toxicity. Cytotoxicity may be noted but compensated by other tests.Non-irritant, non-sensitizer, no systemic toxicity. Cytotoxic effect noted, but acute systemic toxicity test passed.Similar to predicate / Pass

2. Sample Size Used for the Test Set and Data Provenance:

The document clearly states the sample sizes for certain tests:

  • Watertight Test: 315 pieces (gloves) were used for this specific test.
  • Residual Powder Test: 5 pieces (gloves) were used.
  • Physical Properties & Dimensions: Specific sample sizes are not explicitly stated for these tests, but standards like ASTM D412 and D3767 would define the required sample sizes.
  • Chemotherapy Drug Permeation: The specific number of gloves tested per drug type is not listed, but the test methodology (ASTM D6978-05(2019)) typically specifies this. The results are average breakthrough times.

Data Provenance: The tests were conducted by Rimba Glove Sdn Bhd, located in Perak, Malaysia (Submitter information on page 5). The data is non-clinical performance data for the device itself. It is inherently prospective testing of manufactured samples.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This question is not applicable as the "ground truth" for this device is established by physical measurement and chemical analysis against specified ASTM (American Society for Testing and Materials) standards. There are no human experts involved in "establishing ground truth" in the way a radiologist reviews images for an AI algorithm. The tests are objective laboratory measurements.

4. Adjudication Method for the Test Set:

Not applicable. This concept applies to subjective assessments, often in an AI context where multiple human readers disagree. Here, the tests are objective laboratory measurements (e.g., tensile strength, water leak detection, chemical permeation time).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No. This is a medical device (glove), not an AI/Machine Learning enabled device. MRMC studies are specific to evaluating diagnostic accuracy, often comparing human performance with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This is not an AI/Machine Learning enabled device. The "standalone performance" is the performance of the glove itself against the defined physical and chemical standards.

7. The Type of Ground Truth Used:

The "ground truth" for the device's performance is based on objective, measurable criteria defined by international standards (ASTM, ISO) for medical gloves and their material properties. This includes:

  • Physical properties (tensile strength, elongation)
  • Dimensional measurements (length, width, thickness)
  • Integrity (watertightness)
  • Chemical residue (residual powder)
  • Permeation resistance to chemotherapy drugs.

8. The Sample Size for the Training Set:

Not applicable. This is not an AI/Machine Learning device that requires a training set. The device manufacturing processes are subject to Quality System Regulations (21 CFR Part 820), not machine learning training.

9. How the Ground Truth for the Training Set was Established:

Not applicable, as there is no "training set" for this type of medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 8, 2022

Rimba Glove SDN BHD % Michael Woude U.S Agent Emergo Global Representation LLC 2500 Bee Cave Road Bldg 1, Suite 300 Austin, Texas 78746

Re: K212898

Trade/Device Name: Blue Nitrile Powder Free Patient Examination Glove, Non-Sterile, Tested For Use With Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA, LZC

Dear Michael Woude:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 16, 2021. Specifically, FDA is updating this SE Letter to address a typo in the sponsor name as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Dr. Clarance Murray, OHT4: Office of Surgical and Infection Control Devices, 301-796-0870, Clarence.Murray@fda.hhs.gov.

Image /page/0/Picture/8 description: The image shows the text "Sincerely," followed by the name "Liqun Zhao -S". The text is written in a clear, legible font and is likely part of a signature or closing statement. The name is slightly larger than the word "Sincerely,". The image appears to be a cropped portion of a larger document.

For Clarence W. Murray, III, Ph.D Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 16, 2021

Rimba Glove SND BHD % Michael Woude U.S Agent Emergo Global Representation LLC 2500 Bee Cave Road Bldg 1. Suite 300 Austin, Texas 78746

Re: K212898

Trade/Device Name: Blue Nitrile Powder Free Patient Examination Glove, Non-Sterile, Tested For Use With Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: September 4, 2021 Received: September 10, 2021

Dear Michael Woude:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Liqun Zhao -S

For Clarance W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K212898

Device Name

BLUE NITRILE POWDER FREE PATIENT EXAMINATION GLOVE, NON STERILE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs per ASTM D6978-05(2019) Standard Practice for assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugBreakthrough Detection Time in Minutes
*Carmustine (BCNU)23.0
Cisplatin> 240
Cyclophosphamide (Cytoxan)> 240
Dacarbazine> 240
Doxorubicin HCI> 240
Etoposide> 240
Fluorouracil> 240
Ifosfamide> 240
Methotrexate> 240
Mitomycin C> 240
Mitoxantrone> 240
Paclitaxel> 240
*Thiotepa98.8
Vincristine Sulfate> 240

*CAUTION: Testing showed an average breakthrough time of 23.0 minutes for Carmustine and an average breakthrough time of 98.9 minutes for Thiotepa.

** WARNING: DO NOT USE WITH CARMUSTINE.

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

1.0 Submitter:

Name:Zahari Bin Darus
Address:Rimba Glove Sdn BhdPlot 37, Medan Tasek, Tasek Industrial Estate, 31400 Ipoh,Perak, Malaysia
Phone No.:+605-5483688
Fax No.:+605-5481688

Date of Summary Prepared: 20th August 2021 (REVISED DATE: 7th Dec 2021)

Identification of the subject device: 2.0

Trade Name: Blue Nitrile Powder Free Patient Examination Glove, Non-Sterile, Tested for use with Chemotherapy Drugs.
Common Name: Patient Examination Gloves
Classification Name: Patient Examination Gloves
Device Classification: 1
Regulation Number: 21 CFR 880.6250
Product Code: LZA, LZC

3.0 Predicate Device:

K210944

Powder Free Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs (Blue).

Company: Harbour Health LLC.

4.0 Description of The Device:

Blue Nitrile Powder Free Patient Examination Glove, Non-Sterile Tested for use with Chemotherapy Drugs meet all requirements of ASTM standard D6319 and FDA 21 CFR 880.6250.

The powder free nitrile examination glove is manufactured from nitrile rubber. Inner surface of gloves underqoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder on the glove surface. The glove is ambidextrous, i.e., can be worn on right hand or left hand.

5.0 Indication for use:

A patient examination qlove is a disposable device made of nitrile rubber intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device has also been tested for use with Chemotherapy Drugs per ASTM D 6978-5(2019). WARNING: DO NOT USE WITH CARMUSTINE.

6.0 Comparison of the Technological Characteristics of the Device:

The Blue Nitrile Powder Free Patient Examination Glove. Non-Sterile tested for use with chemotherapy drugs are summarized with the following technological characteristics compared to ASTM D6319 or equivalent standards as shown in Table 1.

{6}------------------------------------------------

Table 1

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCECOMPARISON ANALYSIS
PREDICATEBLUECURRENTBLUE
510(k) Number-K210944K212898
Manufacturer(s)-Harbour Health LLC, New Jersey07059.Rimba Glove Sdn Bhd
MaterialASTM D6319NitrileNitrileSame
Color-BlueBlueSame
Texture-Finger textureFinger TextureSame
Physical PropertiesASTM D6319
Before AgingTensile Strength:Ultimate Elongation:≥ 14Mpa500%31.2Mpa594%Different butwithin the ASTMstandard
After AgingTensile Strength:Ultimate Elongation:≥ 14Mpa400%34.7Mpa553%Different butwithin the ASTMstandard
Thickness:- Finger- PalmASTM D63190.05mm0.05mm0.10mm0.07mmDifferent butwithin the ASTMstandard
Powder FreeASTM D6124≤ 2 mg/glove0.26 mg/gloveDifferent butwithin the ASTMstandard
DEVICE PERFORMANCE
CHARACTERISTICSSTANDARDSPREDICATECURRENTCOMPARISONANALYSIS
BLUEBLUE
Primary Skin Irritation- ISO 10993-10:2010(E) & ConsumerProduct SafetyCommission Title 16.Chapter II, Part 1500Under the condition of study,the device is a non-irritantUnder the conditions of study,the test material did not causean irritant response. ThePrimary Irritant ResponseCategory is deemed 'Negligible'Similar
Dermal Sensitization-ISO 10993-10: 2010(E) & ConsumerProduct SafetyCommission, Title 16,Chapter II, Part1500.3 (c) (4)Under the conditions of thestudy, the device is a non-sensitizerUnder the conditions of study,the test material did notproduce a skin sensitizationeffect in the guinea pigs.Similar
BiocompatibilityCytotoxicity - MEMElution, ISO 10993-5:2009 (E)Under the conditions of thestudy, the device is potentiallycytotoxic.Under the conditions of study,the test material demonstrateda cytotoxic effect under thecondition of this study.Additional test i.e. AcuteSystemic Toxicity was tested.Similar
DEVICE PERFORMANCE
CHARACTERISTICSSTANDARDSPREDICATECURRENTCOMPARISONANALYSIS
BLUEBLUE
BiocompatibilityAcute SystemicToxicity,ISO 10993-11:2017 (Ε)Under the conditions of study, thedevice does not elicit a systemictoxicity response in the modelanimal.Under the conditions of study,the test item did not induceany systemic toxicity in Swissalbino mice.Similar
Watertight (1000ml)ASTM D 6319-19ASTM D5151-19Complies with ASTM D6319-19 andASTM D5151-19, G-1, AQL 2.5Gloves passed AQL 1.5Different but wellwithin the ASTMD6319-19
Intended use-The Harbour Health Powder FreeNitrile Examination Glove, Blue(Tested for use with ChemotherapyDrugs) is a disposable deviceintended for medical purposes that isworn on the examiner's hand toprevent contamination betweenpatient and examiner. The proposeddevice was tested for use withChemotherapy drugs per ASTMD6978-05(2019), Standard Practicefor Assessment of Medical Gloves toPermeation by Chemotherapy drugs.A patient examination glove isa disposable device made ofnitrile rubber intended formedical purposes that is wornon the examiner's hand orfinger to prevent contaminationbetween patient and examiner.The device has also beentested for use withChemotherapy Drugs per ASTMD 6978-5(2019)Same
SizeMedical GloveGuidanceManual -LabelingSmallMediumLargeExtra LargeExtra SmallSmallMediumLargeExtra LargeSimilar
Single useMedical GloveGuidanceManual -LabelingSingle UseSingle UseSame
CHARACTERISTICSSTANDARDSDEVICE PERFORMANCECOMPARISON ANALYSIS
PREDICATECURRENT
BLUEBLUE
Chemotherapy DrugsPermeation TestMinimum BreakthroughDetection Times in MinutesMinimum BreakthroughDetection Times in Minutes
Carmustine (BCNU) -3.3 mg/ml14.523.0Different in theresult
Cisplatin - 1.0 mg/ml>240>240Same
Cyclophosphamide (Cytoxan)- 20.0 mg/ml>240>240Same
Dacarbazine – 10.0 mg/ml>240>240Same
Doxorubicin HCI – 2.0 mg/ml>240>240Same
Etoposide - 20.0 mg/ml>240>240Same
Fluorouracil – 50.0 mg/mlASTM D6879-5(2019)>240>240Same
Ifosfamide – 50.0 mg/ml>240>240Same
Methotrexate - 25 mg/ml>240>240Same
Mitomycin C - 0.5 mg/ml>240>240Same
Mitoxantrone – 2 mg/ml>240>240Same
Paclitaxel - 6.0 mg/ml>240>240Same
Thiotepa – 10 mg/ml47.498.8Different in theresult
Vincristine Sulfate - 1 mg/ml>240>240Same

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

There are no significant differences between the two products and they are the same or similar in terms of intended use, materials design, physical properties, thickness and biocompatibility test.

Summary of Non-Clinical Testing 7.0

The performance test data of the non-clinical test for this powder free nitrile examination glove is summarized as per below.

{11}------------------------------------------------

Test MethodStandardPurpose of TestingAcceptance CriteriaResultsStatus
Before agingAfter agingBefore agingAfter aging
Physical PropertiesASTM D412(Standard Test Method forVulcanized Rubber and ThermoplasticElastomers-Tension)To evaluate the tensile (tension) properties of glove.Tensile strengthMin 14.0 MPaMin 14.0 MPa31.2Mpa34.7MpaPass
Ultimate elongationMin 500%Min 400%594%553%Pass

{12}------------------------------------------------

TestMethodStandardPurpose ofTestingGloveSizeAcceptance CriteriaResultsStatus
DimensionASTM D3767Standard Practice forRubber—Measurementof DimensionsTo measure thelength, widthand thicknessof gloveX-SmallLengthMin 240 mmLength250 mmPass
Width$70 \pm 10$ mmWidth78.0 mmPass
ThicknessFingermin 0.05mmThickness0.10mmPass
Palmmin 0.05mm0.07 mm
SmallLengthMin 240 mmLength250 mmPass
Width$80 \pm 10$ mmWidth88.0 mmPass
ThicknessFingermin 0.05mmThickness0.10mmPass
Palmmin 0.05mm0.07 mm
MediumLengthMin 240 mmLength250 mmPass
Width$95 \pm 10$ mmWidth98.0 mmPass
ThicknessFingermin 0.05mmThickness0.10mmPass
Palmmin 0.05mm0.07 mm
LargeLengthMin 240 mmLength250 mmPass
Width$110 \pm 10$ mmWidth108 mmPass

{13}------------------------------------------------

ThicknessFinger – min 0.05mmThickness0.10 mmPass
Palm – min 0.05mm0.07 mm
X-LargeLengthMin 240 mmLength250 mmPass
Width$120 \pm 10$ mmWidth118 mmPass
ThicknessFinger – min 0.05mmThickness0.10 mmPass
Palm – min 0.05mm0.07 mm

{14}------------------------------------------------

TestMethodStandardPurpose ofTestingAcceptance CriteriaResultsStatus
WatertightASTM D5151(Standard TestMethod forDetection of Holesin Medical Gloves)To detect holesthat leak waterand therebycompromise theusefulness of theglove.Sample size: 315 pcsInspection level: G1AQL: 1.5, Acceptance No. 10The batch size for thissampling is 150,001 to500,000. Hence,according to the singlesampling plan GI, thesample to be drawn isunder code Mequivalent to 315pieces with accept 10and reject 11 to beaccepted under AQL1.5. During the test, 1piece was found withleaks. Hence it fallswithin the acceptancecriteria.Pass
TestMethodStandardPurpose ofTestingAcceptance CriteriaResultsStatus
ResidualPowderASTM D6124(Standard TestMethod for ResidualPowder on MedicalGloves)To determine theamount ofresidual powderand non-powdersolids found ongloves.Less than 2 mg per gloveSample size : 5 pcsRequirement : <2mg/gloveResult :0.26mg/glovePass

{15}------------------------------------------------

8.0 Summary of Clinical Testing:

No clinical study is included in this submission.

9.0 Conclusion

The conclusion drawn from the non-clinical tests demonstrate that the subject Blue Nitrile Powder Free Patient Examination Glove, Non-Sterile is as safe, as effective, and performs as well as or better than the legally marketed predicate devices K210944.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.