(302 days)
Blue Sky Bio Multi One Implant System is intended for surgical placement in the bone of the upper or lower jaw to provide support for prosthetic devices to restore chewing function. Implants may be used with single-stage or two-stage procedures. They can be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Blue Sky Bio Multi One Implants are indicated for multiunit restorations in splinted applications. Blue Sky Bio Multi One Implant System with a 45° angulation are indicated for surgical installation in the pterygoid region only, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function.
Blue Sky Bio Long Implant System is intended for surgical placement in the bone of the upper jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. Implants may be used with single-stage or two-stage procedures, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Blue Sky Bio Long implants can be placed bicortically in cases of reduced bone density. Blue Sky Bio Long implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. Blue Sky Bio Long Implant System with a 45° angulation are indicated for surgical installation in the pterygoid region only, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function.
Blue Sky Bio PEEK Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation.
Blue Sky Bio BIO|MAX MULTI ONE Implants include a one-piece implant/abutment construct provided with body diameters of 3.0 mm to 4.3 mm. in multiple lengths from 10 mm to 20 mm. They are provided with a head angulation of 17°, 30°, or 45°. Multi One Implants with body diameters of 3.5 mm and 4.3 mm, in lengths ranging from 10 mm to 20 mm with a 45° angulation are indicated for the pterygoid region only. Implant body diameters of 3.0 mm and 3.25 mm are only intended for the alveolar ridge and not intended for placement in the pterygoid region.
Blue Sky Bio BIO|LONG implants and BIO|MAX MULTI ONE LONG implants include implant lengths designed for placement into the posterior maxillary region. including the pterygoid plate. The implants have one of three connections, an internal hexagon interface (3.5 mm platform), a tapered internal hexagon interface (NP platform) or a Multi One one-piece design. The implants are provided with a body diameter of 3.7 mm to 5.0 mm in implantable lengths ranging from 20 mm to 25 mm. BIO|MAX MULTI ONE LONG implants are provided with a head angulation of 17°, 30°, or 45°. Blue Sky Bio BIO/LONG implants and BIQIMAX MULTI ONE LONG implants are intended to be used with the TILT procedure in the maxilla as shown in the clinical literature. Implants with 45° angulation are indicated for the pterygoid region only. BIOJINTERNAL HEX LONG implants (3.5 mm Platform) are for use with abutments of up to 30° only.
Details of specific implant diameters and lengths are outlined in the table below. Abutment screws compatible with the BIOJLONG Implants were cleared in K060957. K102034 and K190491.
Blue Sky Bio PEEK abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation. Subject device PEEK abutments are provided with two implant/abutment connections, BIO|INTERNAL HEX (3.5 mm Platform) and BIO|MAX (NP).
The subject device implants are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium- 6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). Previously cleared abutments and abutment screws are made of material conforming to ASTM F136. Temporary PEEK abutments are made from TECAPEEK conforming to ASTM F2026 Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications.
The Blue Sky Bio Implant System (K212785) is a dental implant system. The document does not explicitly state acceptance criteria or a formal study that proves the device meets specific acceptance criteria in the typical sense of a clinical trial with predefined endpoints and statistical analysis. Instead, the submission focuses on demonstrating substantial equivalence to legally marketed predicate devices through non-clinical performance testing and a review of existing clinical literature.
Here's a breakdown of the information based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide a table of explicit acceptance criteria with corresponding reported device performance in terms of clinical outcomes (e.g., success rates, marginal bone loss) derived from a dedicated study of the subject device. Instead, it relies on:
- Non-clinical performance test data demonstrating compliance with ISO standards and FDA guidance for dental implants.
- Clinical literature review to support the safety and effectiveness of similar implant dimensions and placement techniques, concluding that the subject device's dimensions and indications do not raise new questions of safety and effectiveness.
Here's a summary of the non-clinical performance and the basis of equivalence:
| Acceptance Criteria (Implied / Indirect) | Reported Device Performance (as stated in submission) |
|---|---|
| Fatigue Resistance (ISO 14801:2016 for Endosseous Dental Implants) | "Fatigue testing was performed according to the requirements of ISO 14801:2016, Dentistry - Dynamic loading test for Endosseous Dental Implants and following FDA Guidance, Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." (Implies satisfactory performance, no specific values given). |
| Biocompatibility (Cytotoxicity, per ISO 10993-1, 10993-5, 10993-12 for PEEK abutments) | "Cytotoxicity testing was performed on representative Temporary PEEK abutments (ASTM F2026-17) according to ISO 10993-1, ISO 10993-5, ISO 10993-12." (Implies satisfactory performance, no specific values given). |
| Packaging and Shelf-Life Stability | "The packaging for the subject device is the identical to the sponsor's own prior clearances as referenced in K102034 and K073713 in which accelerated aging testing was presented and leveraged for the subject devices." (Leveraged from predicate devices). |
| Sterilization (SAL of 10^-6 for sterile components) | "Test results and Sterilization Validations performed for the sponsor's own reference devices is also leveraged demonstrate suitable sterilization of the subject device sterile components with demonstration of a sterility assurance level (SAL) of 10-6." (Leveraged from predicate devices). |
| Pyrogenicity (LAL Endotoxin Analysis with testing limit of 20 EU/device) | "Pyrogenicity information provided is based on FDA Guidance... The method used to determine the device meets pyrogen limit specifications is LAL Endotoxin Analysis with testing limit of 20 EU/device, based on a blood contacting and implanted device." (Implies compliance). |
| End-user Sterility Validation (ISO 17665-2 for PEEK abutments) | "End-user sterility validation was conducted according to ISO 17665-2 for subject Temporary PEEK Abutments." (Implies compliance). |
| MRI Safety (Magnetically Induced Displacement Force and Torque) | "Non-clinical worst-case MRI review to evaluate the metallic devices in the MRI environment using scientific rationale and published literature... Rationale addressed parameters per the FDA guidance..." (Implies safety based on rationale and literature). |
| Clinical Safety/Effectiveness for Pterygoid Placement and Long Implants (Implied from clinical literature) | Clinical literature supports high success rates (e.g., 90.7% to 99%) for implants with similar dimensions and angulations in the pterygoid region and for the TILT procedure. The subject device's dimensions (10-25mm length, up to 45° angulation) fall within this window and are considered to not raise new questions of safety and effectiveness. |
2. Sample Size Used for the Test Set and the Data Provenance
- Non-clinical tests: The document does not specify sample sizes for the non-clinical tests. For leveraged data (packaging, sterilization, material biocompatibility), the sample sizes would refer to those used for the predicate devices, which are not detailed here. For fatigue, the ISO standard typically specifies a minimum number of samples (e.g., 11 per group for staircase method).
- Clinical Literature Review (not a "test set" in the traditional sense): This is a review of retrospective and prospective studies already published in scientific literature. The data provenance is global, as scientific literature authors are not limited by country. The "test set" here refers to the patient cohorts described in the reviewed articles, which would vary significantly. For example, one article mentions "1053 pterygoid implants in 676 patients" (Article 7).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This is not applicable as there was no dedicated clinical "test set" with a pre-established ground truth. The submission relies on scientific literature published by various researchers and clinicians, implicitly accepted by the regulatory body as supporting evidence.
4. Adjudication Method for the Test Set
Not applicable, as there was no dedicated clinical "test set" requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This device is a dental implant system, not an AI-powered diagnostic or assistive tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable, as this is a physical medical device (dental implant).
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
For the safety and effectiveness claims related to clinical use (e.g., implant success rates, complications, bone loss), the "ground truth" is derived from the clinical outcomes data reported in the published scientific literature that was reviewed. This literature typically involves various forms of clinical follow-up, radiographic assessment, and professional judgment by the treating clinicians/researchers in those studies.
8. The Sample Size for the Training Set
Not applicable. This device does not involve a "training set" in the context of machine learning or AI.
9. How the Ground Truth for the Training Set Was Established
Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Blue Sky Bio, LLC Albert Zickmann Vice President of Product Development 800 Liberty Dr. Libertyville, Illinois 60048
Re: K212785
Trade/Device Name: Blue Sky Bio Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: June 1, 2022 Received: June 3, 2022
Dear Albert Zickmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K212785
Device Name
Blue Sky Bio Implant System
Indications for Use (Describe)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Blue Sky Bio Multi One Implant System is intended for surgical placement in the bone of the upper or lower jaw to provide support for prosthetic devices to restore chewing function. Implants may be used with single-stage or two-stage procedures. They can be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Blue Sky Bio Multi One Implants are indicated for multiunit restorations in splinted applications. Blue Sky Bio Multi One Implant System with a 45° angulation are indicated for surgical installation in the pterygoid region only, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function.
Blue Sky Bio Long Implant System is intended for surgical placement in the bone of the upper jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. Implants may be used with single-stage or two-stage procedures, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Blue Sky Bio Long implants can be placed bicortically in cases of reduced bone density. Blue Sky Bio Long implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. Blue Sky Bio Long Implant System with a 45° angulation are indicated for surgical installation in the pterygoid region only, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function.
Blue Sky Bio PEEK Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Blue Sky Bio, LLC
Blue Sky Bio Implant System – K212785
June 30, 2022
ADMINISTRATIVE INFORMATION
| Manufacturer Name | Blue Sky Bio, LLC800 Liberty Dr.Libertyville, IL 60048 |
|---|---|
| Telephone | +1 718-376-0422 |
| Fax | +1 888-234-3685 |
| Official Contact | Albert Zickmann |
| Vice President of product development | |
| Telephone +1 718-376-0422 | |
| Email: azickmann@blueskybio.com |
DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Blue Sky Bio Dental Implant System |
|---|---|
| Common Name | Endosseous dental implant |
| Endosseous dental implant abutment | |
| Temporary Abutment | |
| Classification Regulations | 21 CFR 872.3640 Product Code |
| DZE and NHA | |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
PREDICATE DEVICE INFORMATION
Primary predicate devices:
K190958, Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários SA;
Reference devices: K061319, Spectra Implant System, Implant Direct, LLC K191250, PEEK Abutments, Southern Implants, Inc K163634, External Hex Implant, Southern Implants, (Pty) Ltd. K190491, Blue Sky Bio Zygomatic Implant System, Blue Sky Bio, LLC; K153064, Blue Sky Bio Zygomatic Implant System, Blue Sky Bio, LLC; K102034, Blue Sky Bio Dental Implant System, Blue Sky Bio, LLC; K060957, Blue Sky Bio Dental Implant System, Blue Sky Bio, LLC;
K160119, NobelSpeedy Groovy, Nobel Biocare AB and
K071235, Zimmer One Piece, 3.0 Angled, Zimmer Dental, Inc.
{4}------------------------------------------------
INDICATIONS FOR USE
Blue Sky Bio Multi One Implant System is intended for surgical placement in the bone of the upper or lower jaw to provide support for prosthetic devices to restore chewing function. Implants may be used with single-stage or two-stage procedures. They can be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Blue Sky Bio Multi One Implants are indicated for multiunit restorations in splinted applications. Blue Sky Bio Multi One Implant System with a 45° angulation are indicated for surgical installation in the pterygoid region only, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function.
Blue Sky Bio Long Implant System is intended for surgical placement in the bone of the upper jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. Implants may be used with single-stage or two-stage procedures, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Blue Sky Bio Long implants can be placed bicortically in cases of reduced bone density. Blue Sky Bio Long implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. Blue Sky Bio Long Implant System with a 45° angulation are indicated for surgical installation in the pterygoid region only, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function.
Blue Sky Bio PEEK Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation.
DEVICE DESCRIPTION
Blue Sky Bio BIO|MAX MULTI ONE Implants include a one-piece implant/abutment construct provided with body diameters of 3.0 mm to 4.3 mm. in multiple lengths from 10 mm to 20 mm. They are provided with a head angulation of 17°, 30°, or 45°. Multi One Implants with body diameters of 3.5 mm and 4.3 mm, in lengths ranging from 10 mm to 20 mm with a 45° angulation are indicated for the pterygoid region only. Implant body diameters of 3.0 mm and 3.25 mm are only intended for the alveolar ridge and not intended for placement in the pterygoid region.
Blue Sky Bio BIO|LONG implants and BIO|MAX MULTI ONE LONG implants include implant lengths designed for placement into the posterior maxillary region. including the pterygoid plate. The implants have one of three connections, an internal hexagon interface (3.5 mm platform), a tapered internal hexagon interface (NP platform) or a Multi One one-piece design. The implants are provided with a body diameter of 3.7 mm to 5.0 mm in implantable lengths ranging from 20 mm to 25 mm. BIO|MAX MULTI ONE LONG implants are provided with a head angulation of 17°, 30°, or 45°. Blue Sky Bio BIO/LONG implants and BIQIMAX MULTI ONE LONG implants are intended to be used with the TILT procedure in the maxilla as shown in the clinical literature. Implants with 45° angulation are indicated for the pterygoid region only. BIOJINTERNAL HEX LONG implants (3.5 mm Platform) are for use with abutments of up to 30° only.
Details of specific implant diameters and lengths are outlined in the table below. Abutment screws compatible with the BIOJLONG Implants were cleared in K060957. K102034 and K190491.
Blue Sky Bio PEEK abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation. Subject device PEEK abutments are provided with two implant/abutment connections, BIO|INTERNAL HEX (3.5 mm Platform) and BIO|MAX (NP).
The subject device implants are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium- 6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). Previously cleared abutments and abutment screws are made of material conforming to ASTM F136. Temporary PEEK abutments are made from TECAPEEK conforming to
{5}------------------------------------------------
| BIO MAX MULTI ONE(one-piece) | 10 | 11.5 | 13 | 16 | 18 | 20 | |
|---|---|---|---|---|---|---|---|
| Ø 3.0 mm Straight | X | X | X | X | |||
| Ø 3.0 mm 17° | X | X | X | X | |||
| Ø 3.0 mm 30° | X | X | X | X | |||
| Ø 3.25 mm Straight | X | X | X | X | X | X | |
| Ø 3.25 mm 17° | X | X | X | X | X | X | |
| Ø 3.25 mm 30° | X | X | X | X | X | X | |
| Ø 3.5 mm 45° (pterygoid only) | X | X | X | X | X | X | |
| Ø 4.3 mm 45° (pterygoid only) | X | X | X | X | X | X | |
| BIO LONG Implants | |||||||
| BIO INTERNAL HEX LONG | 20 | 22.5 | 25 | ||||
| Ø 3.7 mm | X | X | X | ||||
| Ø 4.3 mm | X | X | X | ||||
| Ø 5.0 mm | X | X | X | ||||
| BIO MAX LONG | |||||||
| Ø 3.7 mm | X | X | X | ||||
| Ø 4.3 mm | X | X | X | ||||
| Ø 5.0 mm | X | X | X | ||||
| BIO MAX MULTI ONE LONG | |||||||
| 20 | 22.5 | 25 | |||||
| Ø 3.7 mm 17° | X | X | X | ||||
| Ø 3.7 mm 30° | X | X | X | ||||
| Ø 3.7 mm 45° (pterygoid only) | X | X | X | ||||
| Ø 4.3 mm 17° | X | X | X | ||||
| Ø 4.3 mm 30° | X | X | X | ||||
| Ø 4.3 mm 45° (pterygoid only) | X | X | X | ||||
| Ø 5.0 mm 17° | X | X | X | ||||
| Ø 5.0 mm 30° | X | X | X | ||||
| Ø 5.0 mm 45° (pterygoid only) | X | X | X |
ASTM F2026 Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications.
NON-CLINICAL PERFORMANCE TEST DATA
Fatigue testing was performed according to the requirements of ISO 14801:2016, Dentistry -Dynamic loading test for Endosseous Dental Implants and following FDA Guidance, Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments.
The subject device titanium implants are manufactured in the same facilities using the same materials, manufacturing processes as used for the sponsor's own implants as previously cleared in K190491, K102034 and K060957.
{6}------------------------------------------------
The subject titanium implant devices have the same nature of body contact, contact duration, material formulation as the sponsor's own implants as previously cleared in K190491, K102034 and K060957.
Cytotoxicity testing was performed on representative Temporary PEEK abutments (ASTM F2026-17) according to ISO 10993-1, ISO 10993-5, ISO 10993-12.
The packaging for the subject device is the identical to the sponsor's own prior clearances as referenced in K102034 and K073713 in which accelerated aging testing was presented and leveraged for the subject devices. Test results and Sterilization Validations performed for the sponsor's own reference devices is also leveraged demonstrate suitable sterilization of the subject device sterile components with demonstration of a sterility assurance level (SAL) of 10-6.
Pyrogenicity information provided is based on FDA Guidance on Submission and Review of Sterility Information in Premarket Notification (510(k)) Submission for Devices Labeled as Sterile, issued on 21 January 2016. The method used to determine the device meets pyrogen limit specifications is LAL Endotoxin Analysis with testing limit of 20 EU/device, based on a blood contacting and implanted device.
End-user sterility validation was conducted according to ISO 17665-2 for subject Temporary PEEK Abutments.
Non-clinical worst-case MRI review to evaluate the metallic devices in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the subject device components and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
CLINICAL LITERATURE
Clinical Literature is submitted in support of the subject implant placement in the pterygoid region and implant dimensions in this submission. The subject device dimensions with Ø3.5 - Ø5.0 mm with lengths of 10 mm - 25 mm are substantially equivalent for pterygoid placement.
Blue Sky Bio BIO|MAX MULTI ONE Implants, Blue Sky Bio BIO|LONG implants, and BIO|MAX MULTI ONE LONG implants are intended to be used with the TILT procedure in the maxilla as shown in the clinical literature and as shown for the predicate devices in publicly available literature. Implants with 45° angulation are indicated for the pterygoid region only.
The table below outlines the clinical literature submitted in the premarket notification for a determination of substantial equivalence in support of the Blue Sky Bio BIO|MAX MULTI ONE Implants, Blue Sky Bio BIO|LONG implants, and BIO|MAX MULTI ONE LONG implants. These articles provide information to show that long implants used clinically in the TILT procedure and in the pterygoid region are clinically safe and effective for their intended use.
Implants from 10 – 25 mm in length placed at angles 15° - 45° are used in both the TILT procedure and in the pterygoid region with a similar success rate to standard implants. Placement in the pterygoid region is preferable to placement of zygomatic implants due to the greater bone to implant contact and the reduced fulcrum force on the bone. Literature also supports maxillary placement with the TILT procedure or pterygoid placement to avoid additional procedures such as sinus lift and grafting in areas of an atrophic maxilla. BIO|MAX MULTI ONE Implants, Blue Sky Bio BIO|LONG implants, and BIO|MAX MULTI ONE LONG implants are provided in lengths of 10 mm - 25 mm and placed at an angle of 45° or less. This is within the window of the successful implants identified in the literature. Nobel Biocare, NobelSpeedy Groovy implants in lengths of 20-25 mm are identified as using the TILT procedure in the article and the
{7}------------------------------------------------
literature identified was used in support of K160119 where implants of 22 mm and 25 mm had no predicate.
Conclusion:
The subject device sizes for long implant Ø3.5 – Ø5.0 mm with lengths of 10 mm – 25 mm do not raise new questions of safety and effectiveness compared to the predicate devices and the clinical literature in the maxilla or pterygoid regions. BIO|MAX MULTI ONE Implants, Blue Sky Bio BIO|LONG implants, and BIO|MAX MULTI ONE LONG implants are provided in lengths of 10 mm - 25 mm and placed at an angle of 45° or less. This is within the window of the successful implants identified in the literature. They are shorter than the lengths placed in the zygoma which have less bone to implant contact and greater fulcrum force. Therefore, Blue Sky Bio Long Implants do not present greater risk than zygomatic implants.
{8}------------------------------------------------
Clinical Literature Summary Table
| Article | Diameter | Length | Angle | Comments |
|---|---|---|---|---|
| 1 | Refers to ReferenceArticles | Refers to ReferenceArticles | 45° | A systematic review in 2011 indicated a 92% short-term survival rate.However, these baseline findings included data on machined (obsolete older surfaces)and modern micro-roughened surfaces. A more recent study focusing exclusively onimplants with micro-roughened surfaces reported a 3-year survival rate of 99%.9 Thiswas similar to another study in 2005 that reporteda survival rate of 98.6%. |
| 2 | 3.75 - 4.0 mm | 16 - 20 mm | 15 - 60° | "A retrospective study was performed of patients with an atrophic posterior maxillarehabilitated with pterygoid implants between 1999 and 2010 and followed for at least36 months after implant loading. A total of 238 titanium implants (172 anterior and 66pterygoid) were placed in 56 patients. The 3-year overall pterygoid implant survivalrate was 99%" |
| 3 | Stated as consideredbut not identified inthe publication | 13 - 20 mm | 18 - 20 mm @ 15-60°13 - 18 mm @ 60-90°15 - 20 mm @ 45° | A total of 634 patients received 1.893 pterygoid implants, with a meanimplant survival rate of 94.87%. The mean prevalence of implant failure was 0.056with a 95% CI of 0.04 - 0.077. This study demonstrates that pterygoid implants can besuccessfully used in patients with atrophic posterior maxilla. |
| 4 | 3.75 - 4.2 mm | 18 - 25 mm | 30 - 55° | "Referenced literature: patient population not specifically stated in this review.Presented are 17 studies from 1992 to 2015, with total number of pterygoid implants -2525 and average success of 91.88%." |
| 5 | 3.75 - 4.2 mm | 18 - 25 mm | 30 - 55° | Referenced literature: patient population not specifically stated in this review."Pterygoid implant provides a reasonable alternative to 3D maxillaryreconstruction, sinus lifts, and bone augmentation technique. Many authors havereported success rates of pterygoid implants ranging from 90% to 100% after follow-up period ranging from 1 to 12 years with minimal complications. Avoidance of aprosthetic distal cantilever with good stability fit for immediate loading is possiblewith this technique". |
| 6 | 3.75 mm | 13 - 22 mm | not specified | 3 Case reports: 37yr (F), 42 yr (M), 61 yr (F), plus literature review.Referenced literature: patient population not specifically stated in this review."Pterygoid implants were intended to pass through maxillary tuberosity, the pyramidalprocess of palatine bone and the pterygoid process of the sphenoid bone. In theliterature, success rates of pterygoid implants for maxillary rehabilitation have beenmore than 94 to 99%." |
| 7 | 3.75 - 4.0 mmnot all sizesidentified | 10 - 22 mm | 35 - 55° | Thirteen articles were included, reporting a total of 1053 pterygoid implants in 676patients. The weighted average success of pterygoid implants was 90.7%;bone loss evaluated radiographically ranged between 0 and 4.5 mm. No additionalcomplications compared with conventional implants were found, and patientsatisfaction level with the prosthesis was high. Pterygoid implants have high successrates, similar bone loss levels to those of conventional implants, minimal complicationsand good acceptance by patients, being therefore an alternative to treat patients withatrophic posterior maxilla. |
| Article | Diameter | Length | Angle | Comments |
| 8 | 3.5 & 3.75 mmlisted | 16 - 57 mm: (Thepreferred length in thepterygoid is 22-25 mm) | 30 - 50° | Referenced literature: patient population not specifically stated in this review." Pterygoid implants were introduced by Tulasne which were intended to pass throughmaxillary tuberosity, the pyramidal process of palatine bone and engaging thepterygoid process of sphenoid bone. It overcomes the need for sinus lift procedures,grafting process, and, at times, even the invasive zygoma implants. The success rates oftilted and pterygoid implants for maxillary rehabilitation have been more than 98%and 94%-99% as documented in the literature, respectively. Engagement of basalbicortical bone is done which is highly mineralized and least resistant to resorption forachieving good torque. Bicortical engagement of implants has proven to achieve betterstabilization with less stresses on the crestal bone and implants." |
| 9 | 4.0 - 5.0 mm | 20 - 25 mm: (K160119Implants used in thisstudy did not have apredicate with 22 & 25mm Lengths per the510(k) Summary) | 15 - 45° drill protocol with 30°abutment | "In this prospective cohort study, 18 patients (9 male and 7 female) with an averageage of 62 years (age range: 33 to 82 years) with completeedentulous arches (n = 17).There were no implant failures, rendering a cumulative implant survival rate of100%." [at 24 months] Several studies using the sameimplant design reported survival rates on the short-termoutcome between 97.6% and 100%." |
{9}------------------------------------------------
Blue Sky Bio Dental Implant System
{10}------------------------------------------------
EQUIVALENCE TO MARKETED DEVICE
Blue Sky Bio, LLC submits the following information in this Premarket Notification to demonstrate, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the legally marketed predicate devices listed above.
A comparison of the technological characteristics of the subject device and the primary predicate devices is provided in the following tables.
| Subject Device | Reference Device | |
|---|---|---|
| Comparison | Blue Sky Bio, LLCBlue Sky Bio Implant System | Southern Implants, Inc. Southern ImplantsPEEK AbutmentsK191250 |
| Abutment | PEEK Abutment | |
| Indication for Use | Blue Sky Bio PEEK Abutments arepremanufactured prosthetic componentsdirectly connected to endosseous dentalimplants and are intended for provisionaluse up to 180 days as an aid in prostheticrehabilitation. | The Southern Implants PEEK Abutments arepremanufactured prosthetic componentsdirectly connected to endosseous dentalimplants and are intended for provisional useup to 180 days as an aid in prostheticrehabilitation |
| Abutment Design | Screw retainedInternal Connection | Screw retainedEngaging and Non-Engaging |
| Platform Diameter | 3.5 mm, NP (3.5) | 3.35 mm – 4.00 mm |
| Abutment Angle | Straight | Straight |
| Materials | ||
| Abutments | Polyetheretherketone (PEEK) | Polyetheretherketone (PEEK) |
| Abutment Screws | NA | Ti-6Al-4V |
{11}------------------------------------------------
| Subject Device | Primary Predicate | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | |
|---|---|---|---|---|---|---|---|
| Comparison | Blue Sky Bio, LLCBlue Sky Bio Implant System | JJGC Indústria e Comércio deMateriais Dentários SANeodent Implant SystemK190958 | Implant DirectSpectra Dental Implant SystemK061319 | Southern Implants (Pty) Ltd.External Hex ImplantsK163634 | Blue Sky Bio, LLCBlue Sky Bio ZygomaticImplant SystemK190491K153064 | Nobel Biocare ABNobelSpeedy GroovyK160119 | Blue Sky Bio, LLCBlue Sky Bio Dental ImplantSystem K060957K102034 |
| Implants | BIO MAX MULTI ONE LONG | ||||||
| Indications for Use | Blue Sky Bio Long Implant System isintended for surgical placement in the boneof the upper jaw to provide support forprosthetic devices, such as artificial teeth, torestore chewing function. Implants may beused with single-stage or two-stageprocedures, and may be loaded immediatelywhen good primary stability is achieved andwith appropriate occlusal loading. Blue SkyBio Long implants can be placed bicorticallyin cases of reduced bone density. Blue SkyBio Long implants are only indicated formultiple unit restorations in splintedapplications that utilize at least two implantsBlue Sky Bio Long Implant System with a45° angulation are indicated for surgicalinstallation in the pterygoid region only, incases of severe jaw resorption, in order torestore patient esthetics and chewingfunction. | The Neodent Implant System is intended tobe surgically placed in the bone of theupper or lower jaw to provide support forprosthetic devices, such as artificial teeth, torestore chewing function. It may be usedwith single-stage or two-stage procedures,for single or multiple unit restorations, andmay be loaded immediately when goodprimary stability is achieved and withappropriate occlusal loading. The NeodentGM Helix LG implants can be placedbicortically in cases of reduced bonedensity. The Neodent GM Helix LGimplants are only indicated for multiple unitrestorations in splinted applications thatutilize at least two implants | Blue Sky Bio ZygomaticImplant System is indicated forsurgical installation in thezygoma region, in cases ofsevere jaw resorption, in orderto restore patient esthetics andchewing function. Blue Sky Biozygomatic implants arerecommended for the posterior(pre-molar/molar) region, oneimplant on each side, with atleast two standard dentalimplants in the anterior regionto support a fixed restoration.Blue Sky Bio zygomaticimplants may be loadedimmediately when goodprimary stability is achievedand with appropriate occlusalloading. | ||||
| Design | One-piece implant/abutment | Threaded root-form implant tobe used with mating abutments | Threaded root-form implant tobe used with matingabutments, or one-pieceimplant/abutment | Threaded root-formimplant to be used withmating abutments | Threaded root-form implantto be used with matingabutments | Threaded root-form implantto be used with matingabutments | Threaded root-form implant tobe used with mating abutments,or one-piece implant/abutment |
| Platform Ø, mm | 3.5, NP (3.5) | 3.75; 4.0 | 3.0, 3.5, 3.7, 4.3, 4.5, 4.7 5.0,5.7, 6.0, 6.5 | 3.25, 3.75, 4.0, 4.7, 5.0,5.7, 6.0 | K153064Internal hex connection: 4.7,Tapered internal hexconnection: 4.3, 5.0 | 4.0 | 3.5, 4.3, 4.5 6.5 |
| Implant Body Ø,mm | 3.7 4.3, 5.0 | 3.75; 4.0 | 3.0, 3.5, 3.7, 4.3, 4.5, 4.7, 5.0,5.7, 6.0 | 3.25, 3.75, 4.0, 4.7, 5.0,5.7, 6.0 | K1530643.5, 4.1, 4.3, 4.7, 5.0 | 4.0 | 3.0 - 8.0 |
| Implantable Length,mm | 20, 22.5, 25 | 20; 22.5; 25 | 8, 10, 13, 16 | 6, 7, 8.5, 10, 11.5, 13, 15,16, 18, 20 | K15306435, 37.5, 40, 42.5, 45, 47.5,50, 52.5, 55 | 20; 22.5; 25 | 6 - 16 |
| Implant/AbutmentInterface Type | One-piece | Morse taper | Internal hexExternal HexTri-Lobe; One Piece | External Hex | K153064Internal hex with 12° taperInternal hex with 45° bevel | External Hex | Internal hex with 12° taperInternal hex with 45° bevelSquare hex; One Piece |
| One-pieceabutment angle | Straight, 17°, 30°, 45° (45° inpterygoid placement only) | NA | Straight, 12° | NA | NA | NA | Straight |
| Subject Device | Primary Predicate | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | |
| Abutment Angle | NA | Straight, 17°, 30° | Straight, 15°, 30° | Straight, 12°, 17°, 24°, 30° | 17°, 30°, 45° | NA | 15°, 25°, 30° |
| Implant Surface | Grit blasted and acid etched | Acqua™ Hydrophilic Surface | HA RBM or HA Coated | Grit blasted and acid etched | Grit blasted and acid etched | TiUnite | Grit blasted and acid etched |
| Materials | |||||||
| Implants | Ti-6Al-4V | F67 unalloyed titanium, Gr 4 | Titanium Alloy | Unalloyed titanium | Ti-6Al-4V | CPTi | Ti-6Al-4V |
{12}------------------------------------------------
| Subject Device | Primary Predicate | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | |
|---|---|---|---|---|---|---|---|
| Comparison | Blue Sky Bio, LLCBlue Sky Bio Implant System | JJGC Indústria e Comércio deMateriais Dentários SANeodent Implant SystemK190958 | Implant DirectSpectra Dental Implant SystemK061319 | Southern Implants (Pty) Ltd.External Hex ImplantsK163634 | Blue Sky Bio, LLCBlue Sky Bio ZygomaticImplant SystemK190491K153064 | Nobel Biocare ABNobelSpeedy GroovyK160119 | Blue Sky Bio, LLCBlue Sky Bio Dental ImplantSystem K060957K102034 |
| Implants | BIO LONG Implants(BIO INTERNAL HEX LONG;BIO MAX LONG) | ||||||
| Indications for Use | Blue Sky Bio Long Implant System isintended for surgical placement in the boneof the upper jaw to provide support forprosthetic devices, such as artificial teeth, torestore chewing function. Implants may beused with single-stage or two-stageprocedures, and may be loaded immediatelywhen good primary stability is achieved andwith appropriate occlusal loading. Blue SkyBio Long implants can be placed bicorticallyin cases of reduced bone density. Blue SkyBio Long implants are only indicated formultiple unit restorations in splintedapplications that utilize at least two implants.Blue Sky Bio Long Implant System with a45° angulation are indicated for surgicalinstallation in the pterygoid region only, incases of severe jaw resorption, in order torestore patient esthetics and chewingfunction. | The Neodent Implant System is intended to besurgically placed in the bone of the upper orlower jaw to provide support for prostheticdevices, such as artificial teeth, to restorechewing function. It may be used with single-stage or two-stage procedures, for single ormultiple unit restorations, and may be loadedimmediately when good primary stability isachieved and with appropriate occlusal loading.The Neodent GM Helix LG implants can beplaced bicortically in cases of reduced bonedensity. The Neodent GM Helix LG implants areonly indicated for multiple unit restorations insplinted applications that utilize at least twoimplants | Blue Sky Bio ZygomaticImplant System is indicated forsurgical installation in thezygoma region, in cases ofsevere jaw resorption, in orderto restore patient esthetics andchewing function. Blue Sky Biozygomatic implants arerecommended for the posterior(pre-molar/molar) region, oneimplant on each side, with atleast two standard dentalimplants in the anterior regionto support a fixed restoration.Blue Sky Bio zygomaticimplants may be loadedimmediately when goodprimary stability is achievedand with appropriate occlusalloading. | ||||
| Design | Threaded root-form implant to beused with mating abutments | Threaded root-form implant to beused with mating abutments | Threaded root-form implant tobe used with matingabutments, or one-pieceimplant/abutment | Threaded root-formimplant to be used withmating abutments | Threaded root-form implantto be used with matingabutments | Threaded root-form implantto be used with matingabutments | Threaded root-form implant tobe used with mating abutments,or one-piece implant/abutment |
| Platform Ø, mm | 3.5, NP (3.5) | 3.75; 4.0 | 3.0, 3.5, 3.7, 4.3, 4.5, 4.7 5.0,5.7, 6.0, 6.5 | 3.25, 3.75, 4.0, 4.7, 5.0,5.7, 6.0 | K153064Internal hex connection: 4.7,Tapered internal hexconnection: 4.3, 5.0 | 4.0 | 3.5, 4.3, 4.5 6.5 |
{13}------------------------------------------------
Blue Sky Bio Dental Implant System
| Subject Device | Primary Predicate | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | |
|---|---|---|---|---|---|---|---|
| Implant Body Ø,mm | 3.7 4.3, 5.0 | 3.75; 4.0 | 3.0, 3.5, 3.7, 4.3, 4.5,4.7, 5.0,5.7, 6.0 | 3.25, 3.75, 4.0, 4.7, 5.0,5.7, 6.0 | K1530643.5, 4.1, 4.3, 4.7, 5.0 | 4.0 | 3.0 - 8.0 |
| Implantable Length,mm | 20, 22.5, 25 | 20; 22.5; 25 | 8, 10, 13, 16 | 6, 7, 8.5, 10, 11.5, 13, 15,16, 18, 20 | K15306435, 37.5, 40, 42.5, 45, 47.5,50, 52.5, 55 | 20; 22.5; 25 | 6 - 16 |
| Implant/AbutmentInterface Type | Internal hex with 12° taperInternal hex with 45° bevel | Morse taper | Internal hexExternal HexTri-Lobe; One Piece | External Hex | K153064Internal hex with 12° taperInternal hex with 45° bevel | External Hex | Internal hex with 12° taperInternal hex with 45° bevelSquare hex; One Piece |
| Abutment Angle | Straight, 12°, 17°, 24°, 30°, 45°(45° used with BIOMAX LONGin pterygoid placement only) | Straight, 17°, 30° | Straight, 15°, 30° | Straight, 12°, 17°, 24°, 30° | 17°, 30°, 45° | NA | 15°, 25°, 30° |
| Implant Surface | Grit blasted and acid etched | Acqua™ Hydrophilic Surface | HA RBM or HA Coated | Grit blasted and acid etched | Grit blasted and acid etched | TiUnite | Grit blasted and acid etched |
| Materials | |||||||
| Implants | Ti-6Al-4V | F67 unalloyed titanium, Gr 4 | Titanium Alloy | Unalloyed titanium | Ti-6Al-4V | CPTi | Ti-6Al-4V |
| Subject Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | |
|---|---|---|---|---|---|---|
| Comparison | Blue Sky Bio, LLCBlue Sky Bio Implant System | Implant DirectSpectra Dental Implant SystemK061319 | Southern Implants (Pty) Ltd.External Hex ImplantsK163634 | Blue Sky Bio, LLCBlue Sky Bio ZygomaticImplant SystemK190491K153064 | Blue Sky Bio, LLCBlue Sky Bio Dental ImplantSystem K060957K102034 | Zimmer Dental, Inc.Zimmer One-PieceImplant, 3.0mm, AngledK071235 |
| Implants | BIO MAX MULTI ONE | |||||
| Indication | Blue Sky Bio Multi One Implant System is intended for surgicalplacement in the bone of the upper or lower jaw to providesupport for prosthetic devices to restore chewing function.Implants may be used with single-stage or two-stage procedures.They can be loaded immediately when good primary stability isachieved and with appropriate occlusal loading. Blue Sky BioMulti One Implants are indicated for multi-unit restorations insplinted applications. Blue Sky Bio Multi One Implant Systemwith a 45° angulation are indicated for surgical installation in thepterygoid region only, in cases of severe jaw resorption, in orderto restore patient esthetics and chewing function. | The Spectra Dental Implant System consists of one-piece or two-piece implants forsingle-stage or two- stage surgical procedures that are intended for use in partiallyor fully edentulous mandibles and maxillae, in support of single or multiple-unitrestorations and terminal or intermediate abutment support for fixed bridgework.They may be placed in immediate function if initial implant stability can beestablished.The ScrewDirect 3.0 mm implant is indicated for:1.An artificial root structure for single tooth replacement of mandibular central andlateral incisors and maxillary lateral incisors.2.Multiple tooth replacements or denture stabilization. | Blue Sky Bio Zygomatic ImplantSystem is indicated for surgicalinstallation in the zygoma region, incases of severe jaw resorption, in orderto restore patient esthetics and chewingfunction. Blue Sky Biozygomatic implants are recommendedfor the posterior (pre-molar/molar)region, one implant on each side, withat least two standard dental implants inthe anterior region to support a fixedrestoration. Blue Sky Bio zygomaticimplants may be loaded immediatelywhen good primary stability isachieved and with appropriate occlusalloading. | |||
| Design | One-piece implant/abutment | Threaded root-form implant to be used with mating abutments, orone-piece implant/abutment | Threaded root-form implant tobe used with mating abutments | Threaded root-form implant tobe used with mating abutments | Threaded root-form implantwith one-pieceimplant/abutment design |
{14}------------------------------------------------
Blue Sky Bio Dental Implant System
| Subject Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | |
|---|---|---|---|---|---|---|
| Platform Ø, mm | 3.5, NP (3.5) | 3.0, 3.5, 3.7, 4.3, 4.5, 4.7, 5.0, 5.7, 6.0, 6.5 | 3.25, 3.75, 4.0, 4.7, 5.0, 5.7, 6.0 | K153064Internal hex connection: 4.7,Tapered internal hexconnection: 4.3, 5.0 | 3.5, 4.3, 4.5 and 6.5 | NA |
| Implant Body Ø, mm | 3.0, 3.25, 3.5, 3.7 4.3 | 3.0, 3.5, 3.7, 4.3, 4.5, 4.7, 5.0, 5.7, 6.0 | 3.25, 3.75, 4.0, 4.7, 5.0, 5.7 and6.0 | K1530643.5, 4.1, 4.3, 4.7, 5.0 | 3.0 mm to 8.0 | 3.0 |
| Implantable Length,mm | 10, 11.5, 13, 16, 18, 20 | 8, 10, 13, 16 | 6, 7, 8.5, 10, 11.5, 13, 15, 16,18, 20 | K153064: 35, 37.5, 40, 42.5,45, 47.5, 50, 52.5, 55 | 6 to 16 | 10, 11.5, 13, 16 |
| Implant/AbutmentInterface Type | One Piece | Internal hexExternal HexTri-Lobe; One Piece | External Hex | K153064Internal hex with 12° taperInternal hex with 45° bevel | Internal hex with 12° taperInternal hex with 45° bevelSquare hex; One Piece | One-Piece |
| One-pieceabutment angle | Straight, 17°, 30°, 45° | Straight, 12° | NA | NA | Straight | 17° |
| Abutment Angle | NA | Straight, 15°, 30° | Straight, 12°, 17°, 24°, 30° | 17°, 30°, 45° | 15°, 25°, 30° | NA |
| Implant Surface | Grit blasted and acid etched | HA RBM or HA Coated | Grit blasted and acid etched | Grit blasted and acid etched | Grit blasted and acid etched | MTX – HA RBM |
| Materials | ||||||
| Implants | Ti-6Al-4V | Titanium Alloy | Unalloyed titanium | Ti-6Al-4V | Ti-6Al-4V | Ti-6Al-4V |
{15}------------------------------------------------
DISCUSSION:
The indications for use for the primary predicate K190958 do not include pterveoid placement and do not contain language to state that the implants with 45° angulation are indicated for the pterygoid region only. Thus, the indications for use for the subject device was updated to be in line with the K190491 indications for use which contain language specific to zygomatic placement and supported by clinical data.
The subject device sizes for implants Ø3.5 - Ø5.0 mm with lengths of 10 mm - 25 mm do not raise new questions of safety and effectiveness compared to the predicate devices and the clinical literature. BIO|MAX MULTI ONE Implants, Blue Sky Bio BIO|LONG implants, and BIO|MAX MULTI ONE LONG implants are provided in lengths of 10-25 mm and placed at an angle of 45° or less. This is within the window of the successful implants identified in the literature and the predicate devices. They are shorter than the lengths placed in the zygoma which have less bone to implant contact and greater fulcrum force. Therefore, BIO|MAX MULTI ONE Implants, Blue Sky Bio BIO|LONG implants and BIO|MAX MULTI ONE LONG implants do not present greater risk than zygomatic implants.
CONCLUSION
The subject device and the predicate devices have substantially equivalent indications for use, including restoration of patient esthetics and chewing function, and immediate loading. The implants of the subject device and the predicate devices have similar designs and dimensions, including lengths appropriate for pterygoid placement supported by the referenced literature.
The subject device implants and abutments have similar packaging and are sterilized using the same materials and processes as the sponsor's own predicate devices.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
§ 872.3640 Endosseous dental implant.
(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.