(88 days)
The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
The subject device is powder free nitrile patient examination gloves. The subject device is white color. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The subject device is non-sterile.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Device: Disposable Nitrile Examination Glove
1. Table of Acceptance Criteria and Reported Device Performance
| Test Methodology | Purpose | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| ASTM D6319 | Physical Dimensions Test | Length (mm):XS/S: ≥ 220;M/L/XL: ≥ 230;Width:XS: 70±10;S: 80±10;M: 95±10;L: 105±10;XL: 115±10 | Length (mm): >230 (all sizes)Width (mm):XS: 73-76;S: 80-83M: 95-97L: 110-114XL: 118-120Pass |
| ASTM D6319 | Thickness | Finger: ≥ 0.05 mmPalm: ≥ 0.05 mm | XS:Finger: 0.07-0.11 mmPalm: 0.08-0.10 mmS:Finger: 0.08-0.10 mmPalm: 0.08-0.11 mmM:Finger: 0.08-0.11 mmPalm: 0.08-0.11 mmL:Finger: 0.08-0.12 mmPalm: 0.09-0.12 mmXL:Finger: 0.08-0.12 mmPalm: 0.08-0.11 mmPass |
| ASTM D5151 | Watertightness Test for Holes | Meet the requirements of ASTM D5151 AQL 2.5 | XS: 0/125 leaksS: 0/125 leaksM: 0/125 leaksL: 1/125 leaksXL: 1/125 leaksPass |
| ASTM D6124 | Powder Content | < 2.0 mg | XS: 0.04 mgS: 0.06 mgM: 0.08 mgL: 0.08 mgXL: 0.09 mgPass |
| ASTM D412 | Physical Properties (Before Aging) | Tensile Strength: ≥14 MPaUltimate Elongation: ≥500% | XS:Tensile Strength: 15.5-17.9 MPaUltimate Elongation: 524-565%S:Tensile Strength: 16.1-18.1 MPaUltimate Elongation: 520-567%M:Tensile Strength: 15.4-17.9 MPaUltimate Elongation: 528-567%L:Tensile Strength: 15.3-17.7 MPaUltimate Elongation: 521-566%XL:Tensile Strength: 15.4-17.7 MPaUltimate Elongation: 525-567%Pass |
| ASTM D412 | Physical Properties (After Aging) | Tensile Strength: ≥14 MPaUltimate Elongation: ≥400% | XS:Tensile Strength: 15.3-17.4 MPaUltimate Elongation: 524-568%S:Tensile Strength: 15.4-17.8 MPaUltimate Elongation: 528-563%M:Tensile Strength: 15.5-17.8 MPaUltimate Elongation: 527-570%L:Tensile Strength: 15.5-17.5 MPaUltimate Elongation: 534-563%XL:Tensile Strength: 15.9-17.8 MPaUltimate Elongation: 530-569%Pass |
| ISO 10993-5 | Cytotoxicity | Non-cytotoxic | Pass: Under conditions of the study, did not show potential toxicity to L-929 cells. |
| ISO 10993-10 | Irritation | Non-irritating | Pass: Under the conditions of the study, not an irritant. |
| ISO 10993-10 | Sensitization | Non-sensitizing | Pass: Under conditions of the study, not a sensitizer. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The sample sizes for some tests are explicitly mentioned:
- ASTM D5151 (Watertightness Test): 125 gloves per size (XS, S, M, L, XL).
- Other tests (Physical Dimensions, Thickness, Powder Content, Physical Properties Before/After Aging, Cytotoxicity, Irritation, Sensitization): The exact number of samples tested for each of these is not explicitly stated in the provided text, beyond the aggregated "Results" sections.
Data Provenance: The document does not specify the country of origin of the testing data or whether the studies were retrospective or prospective. It only states that the testing was "Non-clinical tests." The submitter is Inner Mongolia Cureguard Medical Technology Co., Ltd. from China.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a medical glove, and the "ground truth" is established by adherence to recognized international and national standards (ISO and ASTM) through laboratory testing, not by expert consensus on diagnostic interpretations.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is a non-clinical device performance study based on standardized laboratory testing, not a diagnostic study requiring expert adjudication of results.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a medical glove and does not involve AI assistance or human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is a medical glove and does not involve an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth used for this device is based on pre-defined specifications and limits set by international and national standards (ASTM D6319, ASTM D5151, ASTM D6124, ASTM D412, ISO 10993-5, ISO 10993-10). These standards define acceptable ranges for physical dimensions, strength, elongation, watertightness, powder content, and biocompatibility.
8. The sample size for the training set
Not applicable. This is a physical medical device (glove) and does not involve a training set as would be used for machine learning.
9. How the ground truth for the training set was established
Not applicable. This device does not use a training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
November 19, 2021
Inner Mongolia Cureguard Medical Technology Co., Ltd. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161, East Lujiazui Rd., Pudong Shanghai, Shanghai 200120 China
Re: K212661
Trade/Device Name: Disposable Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: July 26, 2021 Received: August 23, 2021
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K212661
Device Name Disposable Nitrile Examination Glove
Indications for Use (Describe)
The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K212661
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
1.0 Submitter's Information
Name: Inner Mongolia Cureguard Medical Technology Co.,Ltd. Address: Room 326, Management Committee of New Industrial Park, Tumote youqi, Baotou, Inner Mongolia Autonomous Region 014100, China. Phone Number: +86-13485097856 Contact: Guo Hua Date of Preparation: Jul.26,2021
Designated Submission Correspondent
Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn
2.0 Device Information
Trade name: Disposable Nitrile Examination Glove Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove XS,S, M. L. XL Model(s):
3.0 Classification
Production code: LZA Requlation number: 21CFR880.6250 Classification: Class I Panel: General Hospital
4.0 Predicate Device Information
Ever Global (Vietnam) Enterprise Corp Manufacturer: Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422
5.0 Indication for Use
{4}------------------------------------------------
The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
6.0 Device Description
The subject device is powder free nitrile patient examination gloves. The subject device is white color. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The subject device is non-sterile.
| Item | Subject Device(K212661) | Predicate Device(K171422) |
|---|---|---|
| Product Code | LZA | LZA |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 |
| Class | I | I |
| Intended Use | The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. | The Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. |
| Material | Nitrile | Nitrile |
| Powdered orPowered free | Powdered free | Powdered free |
| Design Feature | Ambidextrous | Ambidextrous |
| Colorant | White | White/ Blue/ Black/ Pink |
| Labeling Information | Single-use indication, powder free, device color, device name, glove size and quantity,Non-Sterile | Single-use indication, powder free, device color, device name, glove size and quantity, Non-Sterile |
| Dimensions(mm) | Length:XS/S: $\ge$ 220;M/L/XL: $\ge$ 230;Width:XS: 70 $\pm$ 10;S: 80 $\pm$ 10;M: 95 $\pm$ 10; | Length:XS/S: $\ge$ 220; M: $\ge$ 235;L/XL: $\ge$ 245Width:XS: 75 $\pm$ 5;S: 85 $\pm$ 5;M: 95 $\pm$ 5; |
7.0 Technological Characteristic Comparison Table
{5}------------------------------------------------
| L: 105±10; | L: 105±5; | ||
|---|---|---|---|
| XL: 115±10 | XL: 115±5 | ||
| Thickness(mm) | Finger: ≥0.05; | Finger: ≥0.05; | |
| Palm: ≥0.05 | Palm: ≥0.05 | ||
| PhysicalProperties | BeforeAging | TensileStrength14MPa, minUltimateElongation500% min | TensileStrength14MPa, minUltimateElongation500% min |
| AfterAging | TensileStrength14MPa, min | TensileStrength14MPa, min | |
| UltimateElongation400%min | UltimateElongation400%min | ||
| Freedom fromHoles | |||
| Powder Content | Meet the requirements ofASTM D6124 <2.0mg | Meet the requirements ofASTM D6124 | |
| Biocompatibility | ISO 10993-10;Under the conditions of thestudy, not an irritant or asensitizerISO 10993-5Under conditions of thestudy, device extract is notcytotoxic | Comply withISO10993-10/ |
Analysis: The physical dimensions are little different with that of the predicate, but they all meet the requirements of ASTM D6319.
8.0 Summary of Non-clinical Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
{6}------------------------------------------------
ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
| TestMethodology | Purpose | Acceptance Criteria | Results |
|---|---|---|---|
| ASTM D6319 | PhysicalDimensionsTest | Length(mm):XS/S: $\ge$ 220;M/L/XL: $\ge$ 230;Width:XS: 70±10;S: 80±10;M: 95±10;L: 105±10;XL: 115±10 | Length(mm):>230Width(mm):XS: 73-76;S: 80-83M: 95-97L: 110-114XL: 118-120Pass |
| Thickness (mm) :Finger: $\ge$ 0.05Palm: $\ge$ 0.05 | XS:Finger: 0.07-0.11Palm: 0.08-0.10S:Finger: 0.08-0.10Palm: 0.08-0.11M:Finger: 0.08-0.11Palm: 0.08-0.11L:Finger: 0.08-0.12Palm: 0.09-0.12XL:Finger: 0.08-0.12Palm: 0.08-0.11Pass | ||
| ASTM D5151 | WatertightnessTest forDetection ofHoles | Meet the requirements of ASTM D5151 AQL 2.5 | XS:0/125 leaksS:0/125 leaksM:0/125 leaksL: 1/125 leaksXL: 1/125 leaksPass |
| ASTM D6124 | PowderContent | Meet the requirements of ASTM D6124 < 2.0mg | XS:0.04mgS:0.06mg |
Table 2: Summary of Non-clinical Testing Table
{7}------------------------------------------------
| M:0.08mg | |||||
|---|---|---|---|---|---|
| L:0.08mg | |||||
| XL:0.09mg | |||||
| Pass | |||||
| ASTM D412 | Physicalproperties | BeforeAging | TensileStrength | ≥14MPa | XS:15.5-17.9S:16.1-18.1M: 15.4-17.9L:15.3-17.7XL:15.4-17.7Pass |
| UltimateElongation | ≥500% | XS:524-565S:520-567M: 528-567L:521-566XL:525-567Pass | |||
| AfterAging | TensileStrength | ≥14MPa | XS:15.3-17.4S:15.4-17.8M:15.5-17.8L:15.5-17.5XL:15.9-17.8Pass | ||
| UltimateElongation | ≥400% | XS:524-568S:528-563M:527-570L:534-563XL:530-569Pass | |||
| ISO 10993-5 | Cytotoxicity | Non-cytotoxic | Under conditionsof the study, didnot show potentialtoxicity to L-929cells.Pass | ||
| ISO 10993-10 | Irritation | Non-irritating | Under theconditions of thestudy, not anirritant.Pass |
{8}------------------------------------------------
| ISO 10993-10 | Sensitization | Non-sensitizing | Under conditionsof the study, not asensitizer.Pass |
|---|---|---|---|
| -------------- | --------------- | ----------------- | ---------------------------------------------------------------- |
9.0 _Summary of Clinical Testing
Clinical testing is not needed for this device.
10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device, Disposable Nitrile Examination Glove, is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.