K Number
K212641
Device Name
StrataMR II Valves and Shunts
Date Cleared
2021-09-16

(27 days)

Product Code
Regulation Number
882.5550
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The StrataMR II Valves and shunts are designed to provide continuous Cerebrospinal Fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design enables the physician to non-invasively adjust valve pressure/performance level pre-and post-implantation by using magnetic adjustment tools without the need of radiographic confirmation.
Device Description
The StrataMR™ II adjustable valves are single-use implantable devices which provide a noninvasive method to address changing patient needs in the management of hydrocephalus. The valves and their associated catheters drain cerebrospinal fluid (CSF) from ventricles in the brain into the peritoneal cavity or the right atrium of the heart, where it is absorbed by the body. StrataMR™ II valve incorporates a ball and cone pressure valve in series with a normally closed siphon control mechanism. Flow control is achieved, and retrograde flow is prevented by combined resistance of the ball and cone and siphon control diaphragm. Pre- and postimplantation, the performance level of the valve can be modified by StrataMR™ Adjustment Tools: Locator tool, guider tool, indicator tool and adjustment tool cleared under K181622.
More Information

No
The device description and performance studies focus on mechanical and hydraulic properties, with no mention of AI/ML algorithms for analysis, adjustment, or prediction. The adjustment is performed manually by a physician using magnetic tools.

Yes
The device is designed to treat hydrocephalus by draining cerebrospinal fluid, which is a therapeutic function.

No

The device is a shunt system designed to drain cerebrospinal fluid. Its function is to treat hydrocephalus by regulating fluid flow, not to diagnose a condition.

No

The device description clearly states it is an implantable device (valves and shunts) and describes physical components and their function. It also mentions associated catheters and adjustment tools, all of which are hardware.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to provide continuous cerebrospinal fluid (CSF) flow from the brain to other parts of the body. This is a therapeutic intervention, not a diagnostic test performed on samples outside the body.
  • Device Description: The device is an implantable valve and shunt system designed to manage hydrocephalus by draining CSF. This is a medical device used for treatment, not for diagnosing a condition.
  • Lack of Diagnostic Function: The device's function is to regulate fluid flow, not to analyze biological samples or provide diagnostic information about a patient's condition.
  • No Mention of In Vitro Testing: The description of the device and its testing focuses on its mechanical performance, biocompatibility, and ability to be adjusted. There is no mention of testing biological samples outside the body.

IVD devices are specifically designed to examine specimens derived from the human body (like blood, urine, tissue) to provide information for the diagnosis, monitoring, or treatment of a disease or condition. This device does not fit that description.

N/A

Intended Use / Indications for Use

The StrataMR II Valves and shunts are designed to provide continuous Cerebrospinal Fluid (CSF) flow from the ventricles of the brain into the right atrium of the peritoneal cavity. The design enables the physician to noninvasively adjust valve pressure/performance level pre-and post-implantation by using magnetic adjustment tools without the need of radiographic confirmation.

Product codes (comma separated list FDA assigned to the subject device)

JXG

Device Description

The StrataMR™ II adjustable valves are single-use implantable devices which provide a noninvasive method to address changing patient needs in the management of hydrocephalus. The valves and their associated catheters drain cerebrospinal fluid (CSF) from ventricles in the brain into the peritoneal cavity or the right atrium of the heart, where it is absorbed by the body. StrataMR™ II valve incorporates a ball and cone pressure valve in series with a normally closed siphon control mechanism. Flow control is achieved, and retrograde flow is prevented by combined resistance of the ball and cone and siphon control diaphragm. Pre- and postimplantation, the performance level of the valve can be modified by StrataMR™ Adjustment Tools: Locator tool, guider tool, indicator tool and adjustment tool cleared under K181622.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

ventricles of the brain, right atrium of the heart, peritoneal cavity

Indicated Patient Age Range

Not Found

Intended User / Care Setting

physician

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance Data – Bench
Study Type: Bench Testing
Test Summary:
Resistance to Leakage: Resistance to leakage was measured using air. The valves were required to show no leakage for 5 minutes with a differential pressure from inside to outside of 1 m H₂O.
Reservoir dome needle puncture: Reservoir domes were required to show no leakage when repeatedly punctured with a non-coring needle under pressure.
Dynamic Breaking Strength: Tension was applied in the flow direction, leading to an elongation of the shunt of 10% or a maximum force of 5 N. Shunts were required to exhibit no break, rupture or disconnection after 100,000 cycles at frequency 1.0 ± 0.2 Hz.
Opening Pressure: The test demonstrates valve flow following valve manufacture, the valve was to show patency at a pressure that meets manufacturer's specifications
Pressure/Flow: Pressure/flow performance was tested in accordance to BS EN ISO 7197:2009. The measured pressure has to remain inside the manufacturer's specifications
Posture Effect: The test compared valve pressure at -50 cm hydrostatic pressure with valve pressure at 0 cm hydrostatic pressure. The difference was required to meet manufacturer's specifications
Ability to withstand Overpressure: After application of positive pressure of 1 m water to the open shunt in accordance with BS EN ISO 7197:2009, valves were required to meet pre-established pressure/flow specifications.
Bursting Pressure: After application of positive pressure of 2 m water in accordance with BS EN ISO 7197:2009, valves were required to meet pre-established pressure/flow specifications.
Long Term Stability: Valves were placed in a water bath with temperature of 37℃ ± 5° while pumping water through the valves at an average flow rate of 20 mL/h for 28 days. During this time, valves were required to maintain pre-established pressure/flow specifications.
Flushing capacity: Flushing volume that is produced from valve reservoir compressions was tested.
Identification of shunts in vivo: Valves underwent X-ray imaging. The valve identification markers must be visible, and the valve setting must be readable in the X-ray images.
Post-MRI Functional Testing: Valves were exposed to multiple MRI exposures in clinically relevant orientations. After MRI exposure, valves were required to: (1) maintain the pre-conditioning pressure setting, (2) be able to be read and adjusted, and (3) meet pre-established pressure/flow specifications.
MRI Safety Information: Magnetically induced displacement force testing per ASTM F2052-15; Magnetically induced torque testing per ASTM F2213-06; Radio Frequency induced heating testing per ASTM F2182-11a; Image artifact testing per ASTM F2119-07.

Key Results:
All valves met acceptance criteria for Resistance to Leakage, demonstrating no concerns regarding valve integrity/leakage relative to the predicate device.
All valves met acceptance criteria for Opening Pressure, demonstrating no concerns regarding valve opening pressure (patency).
All valves met acceptance criteria for Ability to withstand Overpressure and Bursting Pressure, demonstrating no concerns regarding pressure/flow performance relative to the predicate device.
All valves met acceptance criteria for Flushing capacity, demonstrating no concerns regarding reservoir volume relative to the predicate device.
All valves met acceptance criteria for Identification of shunts in vivo, demonstrating no concerns regarding identification of the valve via X-ray relative to the predicate device.
All valves met acceptance criteria for Post-MRI Functional Testing, demonstrating no concerns regarding performance after MRI exposure relative to the predicate device.
Test results demonstrated that StrataMR™ II valves are MR conditional and that there are not MRI safety concerns relative to the predicate device when scanned according to the MR conditions specified in the labeling.

Biocompatibility Testing
Study Type: Biocompatibility testing per ISO 10993-1
Test Methodologies:
Cytotoxicity- MEM Elution
Irritation-Intracutaneous study in rabbits
Sensitization- Guinea Pig Maximization
Systemic toxicity - Acute systemic toxicity study in mice
Material Mediated Pyrogenicity- USP Rabbit Pyrogen Study
Genotoxicity/ Carcinogenicity
Implantation
Hemocompatibility

Key Results:
Non-cytotoxic
Non-irritant
Non-sensitizer
Non-toxic (Systemic toxicity and Implantation)
Non-pyrogenic
Non-mutagenic
Non-hemolytic

Performance Data - Animal
No animal testing was conducted.

Performance Data - Clinical
No clinical testing was conducted.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K152700

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K060681

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 882.5550 Central nervous system fluid shunt and components.

(a)
Identification. A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The FDA acronym is in a blue box, and the full name, "U.S. Food & Drug Administration," is in blue text.

September 16, 2021

Medtronic Neurosurgery Sudeep Nambiar Regulatory Affairs Manager 5290 California Avenue Irvine, California 92617

Re: K212641

Trade/Device Name: StrataMR II Valves and Shunts Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: August 17, 2021 Received: August 20, 2021

Dear Sudeep Nambiar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Adam D. Pierce -S

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K212641

Device Name StrataMR II Valves and Shunts

Indications for Use (Describe)

The StrataMR II Valves and shunts are designed to provide continuous Cerebrospinal Fluid (CSF) flow from the ventricles of the brain into the right atrium of the peritoneal cavity. The design enables the physician to noninvasively adjust valve pressure/performance level pre-and post-implantation by using magnetic adjustment tools without the need of radiographic confirmation.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) Summary

  • I. Medtronic, Inc. Company: Medtronic Neurosurgery 5290 California Ave Irvine, California 92617 USA
    • Swara Vashi Contact: Regulatory Affairs Specialist swara.vashi@medtronic.com Telephone Number: 267-638-8778
    • Alternate Contact: Sudeep Nambiar Regulatory Affairs Manager sudeep.nambiar(@medtronic.com Telephone Number: 607-349-9048

Date of Summary Prepared: 19 August 2021

  • II. Proprietary Trade Name: StrataMR™ II Valves and Shunts
  • III. Regulatory Class: II

IV. Primary Classification:

Name: Central nervous system fluid shunt and components Product Code: JXG Regulation: 882.5550

V. Product Description

The StrataMR™ II adjustable valves are single-use implantable devices which provide a noninvasive method to address changing patient needs in the management of hydrocephalus. The valves and their associated catheters drain cerebrospinal fluid (CSF) from ventricles in the brain into the peritoneal cavity or the right atrium of the heart, where it is absorbed by the body. StrataMR™ II valve incorporates a ball and cone pressure valve in series with a normally closed siphon control mechanism. Flow control is achieved, and retrograde flow is prevented by combined resistance of the ball and cone and siphon control diaphragm. Pre- and postimplantation, the performance level of the valve can be modified by StrataMR™ Adjustment Tools: Locator tool, guider tool, indicator tool and adjustment tool cleared under K181622.

VI. Indications for Use

The StrataMR™ II Valves and shunts are designed to provide continuous Cerebrospinal Fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design enables the physician to non-invasively adjust valve pressure/performance level pre-and post-implantation by using magnetic adjustment tools without the need of radiographic confirmation.

4

VII. Summary of the Technological Characteristics

The subject device and the predicate device have the same indications for use, intended use, fundamental technology and operating principle. The primary difference between the subject device StrataMR™ II and the predicate StrataMR™ is in the cap material. The cap material of the subject device StrataMR™ II is same as the reference device, Strata™ II that was cleared under K042465 and K060681. An additional tantalum marker has been included on the valves to noninvasively differentiate between the subject StrataMR™ II and predicate StrataMR™ valves. Outer sterile packaging barrier is changed from Mylar to Nylon with Tyvek.

VIII. Identification of Legally Marketed Predicate Devices

Primary Predicate device: StrataMR™ Valves and Shunts Medtronic (K152700) Reference Predicate: Strata™ II Valve and Shunt (K060681)

IX. Device Comparison

The tables below provide a summary of the device

CharacteristicSubject DevicePredicate DeviceReference Predicate Device
Device NameStrataMR TM II Valves
and ShuntsStrataMR TM Valves and
ShuntsStrata TM II Valves and
Shunts
510k NumberK212641K152700K060681
Indications for UseThe StrataMR TM II
valves and shunts are
designed to provide
continuous cerebrospinal
fluid (CSF) flow from
the ventricles of the
brain into the right
atrium of the heart or the
peritoneal cavity. The
design enables the
physician to non-
invasively adjust valve
pressure/performance lev
el pre-and post-
implantation by using
magnetic adjustment
tools without the need
for radiographic
confirmation.The Medtronic StrataMR TM
valves and shunts are
designed to provide
continuous cerebrospinal
fluid (CSF) flow from the
ventricles of the brain into
the right atrium of the heart
or the peritoneal cavity. The
design enables the physician
to non-invasively adjust
valve pressure/performance
level pre- and post-
implantation by using
magnetic adjustment tools
without the need for
radiographic information.The Medtronic PS
Medical Strata TM II Valve
and Shunt
Assemblies with and
without BioGlide are
shunt components
designed to provide
continued CSF flow from
the ventricles of the brain
into the right atrium of
the heart or the peritoneal
cavity. The design allows
the physician to non-
invasively adjust valve
pressure/performance lev
el pre-and post-
implantation tool without
the need for radiographic
confirmation.
Physical Characteristics/ Principal of operation
DimensionMaximum Small: 1.93"
Length Regular: 2.29"Small: 1.93"
Regular: 2.29"Small: 1.89"
Regular: 2.25"
Maximum Small: 0.37"
Height Regular :0.42"Small: 0.35"
Regular :0.40"Small: 0.30"
Regular :0.35"
Maximum Small: 0.59"
Width Regular: 0.70"Small: 0.59"
Regular: 0.70"Small: 0.53"
Regular: 0.67"
Pressure settings5 discrete pressure
settingsSameSame

5

| Characteristic | Subject Device | Predicate Device | Reference Predicate
Device |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Adjustment Tools | Adjustment tool set
contains locator tool,
indicator tool,
adjustment tool
and guider tool cleared
under K181622 and
K152700 | Same | Strata™ II
adjustment kit contains
locator tool, indicator
tool and Adjustment tool
cleared under K060681 |
| Shunt system
components | Valves used with
existing, legally
marketed shunt
components and
accessories | Same | Same |
| Patient contact and
duration | Tissue/bone contact with
long term duration (>30
days) | Same | Same |
| Shelf Life | 22 months | 3 years | 3 years |
| Sterilization Method | Ethylene Oxide | Same | Same |
| Packaging material | Valves and shunts are
packaged in plastic tray
sealed with Tyvek lid.
Sealed tray is placed in a
Tyvek-Nylon peel pouch
which is then placed in
carton. | Valves and shunts are
packaged in plastic
tray sealed with Tyvek lid.
Sealed tray is placed in a
Tyvek-Mylar peel pouch
(StrataMR valves model no.
42955 & 42965) or Tyvek-
Nylon pouch (StrataMR
shunt assembly model no.
46955 & 46965) which is
then placed in carton | Valves and shunts are
packaged in plastic tray
sealed with Tyvek lid.
Sealed tray is placed in a
Tyvek-Nylon peel pouch
which is then placed in
carton. |

X. Discussion of the Performance testing

Performance Data – Bench

TestTest SummaryResults
Resistance to
LeakageResistance to leakage was measured using air. The
valves were required to show no leakage for 5 minutes
with a differential pressure from inside to outside of 1
m H₂O.All valves met acceptance
criteria, demonstrating that
there are no concerns
regarding valve
Reservoir dome
needle punctureReservoir domes were required to show no leakage
when repeatedly punctured with a non-coring needle
under pressure.integrity/leakage relative to
the predicate device.
TestTest SummaryResults
Dynamic
Breaking
StrengthTension was applied in the flow direction, leading to
an elongation of the shunt of 10% or a maximum force
of 5 N. Shunts were required to exhibit no break,
rupture or disconnection after 100,000 cycles at
frequency 1.0 $\pm$ 0.2 Hz.
Opening PressureThe test demonstrates valve flow following valve
manufacture, the valve was to show patency at a
pressure that meets manufacturer's specificationsAll valves met acceptance
criteria, demonstrating that
there are no concerns
regarding valve opening
pressure (patency).
Pressure/FlowPressure/flow performance was tested in accordance to
BS EN ISO 7197:2009. The measured pressure has to
remain inside the manufacturer's specifications
Posture EffectThe test compared valve pressure at -50 cm hydrostatic
pressure with valve pressure at 0 cm hydrostatic
pressure. The difference was required to meet
manufacturer's specifications
Ability to
withstand
OverpressureAfter application of positive pressure of 1 m water to
the open shunt in accordance with BS EN ISO
7197:2009, valves were required to meet pre-
established pressure/flow specifications.All valves met acceptance
criteria, demonstrating that
there are no concerns
regarding pressure/flow
performance relative to the
predicate device.
Bursting PressureAfter application of positive pressure of 2 m water in
accordance with BS EN ISO 7197:2009, valves were
required to meet pre-established pressure/flow
specifications.
Long Term
StabilityValves were placed in a water bath with temperature
of 37℃ $\pm$ 5° while pumping water through the valves
at an average flow rate of 20 mL/h for 28 days. During
this time, valves were required to maintain pre-
established pressure/flow specifications.
Flushing capacityFlushing volume that is produced from valve reservoir
compressions was tested.All valves met acceptance
criteria, demonstrating that
there are no concerns
regarding reservoir volume
relative to the predicate
device.
Identification of
shunts in vivoValves underwent X-ray imaging. The valve
identification markers must be visible, and the valve
setting must be readable in the X-ray images.All valves met acceptance
criteria, demonstrating that
there are no concerns
regarding identification of the
valve via X-ray relative to the
predicate device.
TestTest SummaryResults
Post-MRI
Functional
TestingValves were exposed to multiple MRI exposures in
clinically relevant orientations. After MRI exposure,
valves were required to: (1) maintain the pre-
conditioning pressure setting, (2) be able to be read
and adjusted, and (3) meet pre-established
pressure/flow specifications.All valves met acceptance
criteria, demonstrating that
there are no concerns
regarding performance after
MRI exposure relative to the
predicate device.
MRI Safety
InformationMagnetically induced displacement force testing
• per ASTM F2052-15
Magnetically induced torque testing per ASTM
• F2213-06
Radio Frequency induced heating testing per
• ASTM F2182-11a
Image artifact testing per ASTM F2119-07
Test results demonstrated that
StrataMR ™ II valves are MR
conditional and that there are
not MRI safety concerns
relative to the predicate device
when scanned according to
the MR conditions specified
in the labeling.

6

7

StrataMR™ II valves and shunts met applicable pre-established acceptance criteria and raised no concerns regarding safety and effectiveness relative to the predicate device. Therefore, bench testing summarized above supports the substantial equivalence of StrataMR™ II and the predicate.

Biocompatibility Testing

The following Biocompatibility tests were conducted per ISO 10993-1 endpoints for tissue/bone/CSF implant device with long-term contact.

TestTest Method SummaryResults
Cytotoxicity- MEM ElutionCell culture treated with test sample exhibited no signs of toxicity (Grade 0)Non-cytotoxic
Irritation-Intracutaneous study in rabbitsAnimals treated with test sample exhibited no dermal reactions (Grade 0)Non-irritant
Sensitization- Guinea Pig MaximizationAnimals treated with test sample exhibited no dermal reactions.Non-sensitizer
Systemic toxicity - Acute systemic toxicity study in miceAnimals treated with test sample exhibited no mortality or evidence of systemic toxicityNon-toxic
Material Mediated Pyrogenicity- USP Rabbit Pyrogen StudyIndividual animals treated with test sample exhibited no temperature rise above 0.5°CNon-pyrogenic
Genotoxicity/ CarcinogenicityThe test article did not show a significantly greater biological reaction than the controlsNon-mutagenic
ImplantationThe test did not show a significantly greater biological reaction than theNon-toxic

8

TestTest Method SummaryResults
control (Grade 0 - no
reaction)
HemocompatibilityThe test article exhibited no
hemolysis and was within the
control valuesNon-hemolytic

StrataMR™ II valves and shunts passed biocompatibility testing demonstrating that StrataMRTM II material change does not raise biocompatibility concerns relative to predicate device. Therefore, biocompatibility testing summarized above supports the substantial equivalence of StrataMRTM II to the predicate device.

Performance Data - Animal

No animal testing was conducted. The differences in technological characteristics do not raise new questions of safety and effectiveness as demonstrated through non-clinical bench testing using well established acceptable scientific methods.

Performance Data - Clinical

No clinical testing was conducted. The differences in technological characteristics do not raise new questions of safety and effectiveness as demonstrated through non-clinical bench testing using well established acceptable scientific methods.

XI. Conclusion

Based on indications for use, design, technological characteristics, and performance testing conducted on the proposed device, it can be concluded that the StrataMR™ II valves and Shunts are substantially equivalent to StrataMR™ valves and Shunts cleared under K152700.