(231 days)
The DRPx Locking Distal Radius Plate System is indicated for the fixation of intra- and extra-articular fractures and osteotomies of the distal radius.
The DRPx Locking Distal Radius Plate System consists of titanium plates for the distal radius, a dorsal spanning plate, bone pegs, bone screws and instruments to facilitate implantation. The Distal Radius plates come in a variety of sizes and are pre-contoured to match the anatomy of the distal radius. They accept 2.0mm locking pegs, 2.3mm locking screws, and 2.6mm and 3.5mm locking and non-locking bone screws range in length from 6mm to 30mm. The system also includes instruments used to implant the plates. The dorsal spanning plate is being added to the system in this submission.
This document (K212601) is a 510(k) premarket notification for a medical device called the DRPx Locking Distal Radius Plate System. The information provided focuses on demonstrating substantial equivalence to predicate devices, rather than a clinical study evaluating the device's performance against specific acceptance criteria in a clinical setting.
Therefore, the requested information concerning acceptance criteria, device performance, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details for a study proving the device meets acceptance criteria cannot be fully provided from this document.
This document describes mechanical performance testing rather than clinical performance.
Here's an analysis based on the provided text, highlighting what can be extracted and what cannot:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Mechanical Performance (Based on ASTM F382) | |
| Single cycle bend testing for metallic bone plates | "found to have acceptable mechanical characteristics for the intended uses" |
| Bending fatigue testing for metallic bone plates | "found to have acceptable mechanical characteristics for the intended uses" |
| Clinical Performance | Not applicable; no clinical acceptance criteria or performance reported in this document. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not specified for the mechanical tests. ASTM F382 typically defines the number of samples required for such tests.
- Data Provenance: The mechanical tests were performed by Orthopaedic Implant Company (OIC). The country of origin of the data is implicitly the USA (where OIC is located) if the testing was done in-house or by a US-based laboratory. The testing described is retrospective in the context of device submission, meaning it was completed prior to the submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. This document describes mechanical testing to industry standards (ASTM F382) and does not involve human experts establishing ground truth for a clinical test set. The "ground truth" for mechanical testing is adherence to the specified test parameters and achieving acceptable results according to the standard.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This pertains to clinical studies involving human interpretation or assessment, which is not the nature of the described mechanical testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. An MRMC study is relevant for AI-powered diagnostic/interpretive devices. This submission concerns a physical orthopaedic implant, not an AI system assisting human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This pertains to AI algorithms. The DRPx Locking Distal Radius Plate System is a physical medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Mechanical Test Standards: For the mechanical performance testing, the "ground truth" is defined by the technical specifications and acceptance criteria of the ASTM F382 standard for metallic bone plates. There is no clinical "ground truth" in this context.
8. The sample size for the training set
- Not applicable. This concept applies to machine learning models, not to the design and testing of a physical orthopaedic implant.
9. How the ground truth for the training set was established
- Not applicable. See point 8.
In summary: K212601 is a 510(k) submission for a physical medical device. The "study" mentioned refers to mechanical performance testing performed to demonstrate the device's physical properties meet established engineering standards (ASTM F382). It does not involve clinical trials with human subjects, evaluation of clinical outcomes, or the use of AI. Therefore, most of the requested fields related to clinical studies, human experts, and AI are not applicable to the information contained in this document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA) along with the Department of Health & Human Services. The FDA logo features a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below. To the left of the FDA logo is the symbol of the Department of Health & Human Services, which consists of a stylized human figure.
April 5, 2022
Orthopaedic Implant Company Douglas Fulton Quality Assurance Manager 770 Smithridge Dr. #400 Reno, Nevada 89502
Re: K212601
Trade/Device Name: DRPx Locking Distal Radius Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC, HTN Dated: August 13, 2021 Received: August 17, 2021
Dear Douglas Fulton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
DRPx Locking Distal Radius Plate System
Indications for Use (Describe)
The DRPx Locking Distal Radius Plate System is indicated for the fixation of intra- and extra-articular fractures and osteotomies of the distal radius.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
510(k) Summary
Prepared 2/16/2022
Name and Address of Manufacturer: The Orthopaedic Implant Company (OIC) 770 Smithridge Drive. Suite 400 Reno, NV 89502
Contact: Douglas Fulton Quality Assurance Manager Telephone: 775-636-8281 Fax: 775-636-8284 Email: doug@orthoimplantcompany.com
Device Identification: Trade Name: DRPx Locking Distal Radius Plate System Common Name: Plate, Fixation, Bone Classification Name: Plate, Fixation, Bone (21 CFR 888.3030)(Primary); Screw, Fixation, Bone (21 CFR 888.3040); and Washer, Bolt Nut (21 CFR 888.3030) Classification: Class II Panel: Orthopedic Product Code: HRS (Primary), HWC, and HTN
Indications for Use: The DRPx Locking Distal Radius Plate System is indicated for the fixation of intra-and extra-articular fractures and osteotomies of the distal radius.
Device Description:
The DRPx Locking Distal Radius Plate System consists of titanium plates for the distal radius, a dorsal spanning plate, bone pegs, bone screws and instruments to facilitate implantation. The Distal Radius plates come in a variety of sizes and are pre-contoured to match the anatomy of the distal radius. They accept 2.0mm locking pegs, 2.3mm locking screws, and 2.6mm and 3.5mm locking and non-locking bone screws range in length from 6mm to 30mm. The system also includes instruments used to implant the plates. The dorsal spanning plate is being added to the system in this submission.
Comparison of Technological Characteristics (Substantial Equivalence):
Primary predicate device: Orthopaedic Implant Company K202971 DRPx Locking Distal Radius System
Additional predicate devices: Skeletal Dynamics K150675 Dorsal Spanning Plate
K082807 Synthes (USA) 3.5mm and 4.5mm Locking Compression Plate (LCP) System with Expanded Indications
The DRPx Locking Distal Radius Plate System has the following similarities to those which previously received 510(k) concurrence:
· has the same indicated use,
- · uses the same operating principle,
- · incorporates a very similar design, and
- · incorporates the same materials
Performance Testing:
Single cycle bend testing and bending fatigue testing were performed on the OIC Dorsal Spanning Plate and the Synthes Locking Compression Plate per ASTM F382, "Standard Specification and test Method for Metallic Bone Plates". The plate was found to have acceptable mechanical characteristics for the intended uses.
Conclusion:
The DRPx Locking Distal Radius Plate System described in this substantially equivalent to the predicate devices.
§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.
(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.