K Number
K212359
Date Cleared
2022-09-21

(419 days)

Product Code
Regulation Number
878.4780
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Applied Tissue Technologies PWD™ Platform Wound Dressing Negative Pressure System comprised of the; PWD™ Dressing, PWD™ Negative Pressure Wound Therapy Pump, and the PWD™ Wound Exudate Canister is indicated in patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it creates an environment that may promote wound healing by removing excess wound exudate. The Applied Tissue Technologies Platform Wound Dressing is appropriate for use on the following wounds: exudating, chronic, sub-acute and dehisced wounds, partial thickness burns, ulcers (such as diabetic, pressure and venous insufficiencies), flaps, grafts and closed surgical incisions.

PWD Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.

Device Description

The Applied Tissue Technologies PWD™ Platform Wound Dressing Negative Pressure System consists of the PWD™ Dressing, PWD™ Negative Pressure Wound Therapy Pump and the Wound exudate canister. The system provides a negative pressure wound dressing therapy in a moist environment to facilitate the wound healing process. The dressing has an access port through which fluids can be drawn from the wound site. The same port can also be utilized to administer negative pressure wound therapy using the negative pressure pump if required. The negative pressure pump is used with a collection cannister to collect the exudate. The PWD Platform Wound Dressing is supplied in five different shapes and sizes, all having the same intended use.

AI/ML Overview

The provided document describes the PWD™ Platform Wound Dressing Negative Pressure System and its substantial equivalence to predicate devices, focusing on performance data from bench testing.

Here's an analysis based on your requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Test NameDescriptionReference StandardsAcceptance CriteriaReported Device Performance
Water Vapor Transmission RateMeasures the passage of water vapor through the membrane/adhesive barrierASTM 1249Must be at least equivalent to commercially available NPWT dressingPass
Exudate Throughput Properties Under Negative PressureSimulated Use testing using various exudate types to confirm acceptable fluid flow when connected to the Pump and Canister systemN/ASystem must demonstrate the ability to remove exudates of various types from the wound bedPass
Maintenance of Negative PressureSimulated use testing from under moist conditions to demonstrate maintenance of negative pressureNAThe PWD must maintain negative pressure at the wound bed for the test durationPass
UsabilityUsability testing for both healthcare and home use environmentsNAUsers must demonstrate the ability to correctly use the devicePass
Electrical SafetyComplies with applicable safety standardsIEC 60601-1, IEC 60601-1-2, IEC 60601-1-11Requirements for electrical safety per the applicable safety standards is metPass
Electro-magnetic CompatibilityComplies with applicable safety standardsComplies with IEC 60601-2-3Requirements for electromagnetic compatibility per the applicable safety standards is metPass

2. Sample Size Used for the Test Set and Data Provenance:

The document explicitly states that the performance data was generated through "Bench testing" for various characteristics. This type of testing typically uses laboratory samples of the device components or simulated systems rather than human subjects or a clinical test set.

  • Sample Size for Test Set: Not specified, but implied to be laboratory samples of device components.
  • Data Provenance: Not applicable in the context of clinical data. The data is from laboratory bench testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • This information is not provided in the document. For bench testing, ground truth is typically established by engineering specifications, validated test methods, and compliance with industry standards, rather than expert clinicians.

4. Adjudication Method for the Test Set:

  • This information is not applicable as the described tests are bench tests, not clinical studies requiring adjudication of output from human readers or an algorithm.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No, an MRMC comparative effectiveness study was not done. The document describes a 510(k) premarket notification process based on substantial equivalence to predicate devices and bench testing. There is no mention of a clinical trial involving human readers or a comparison of human reader performance with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • No, a standalone algorithm performance study was not done. This device is a physical medical device (Negative Pressure Wound Therapy System), not an AI algorithm. The performance evaluation focuses on the physical and functional aspects of the system.

7. The Type of Ground Truth Used:

  • The ground truth for the bench tests appears to be based on engineering specifications, established industry standards (e.g., ASTM, IEC), and the functional requirements for a negative pressure wound therapy system. For instance, "ability to remove exudates of various types from the wound bed" and "maintain negative pressure at the wound bed for the test duration" serve as the functional ground truths.

8. The Sample Size for the Training Set:

  • Not applicable. This document describes the testing of a physical medical device. There is no mention of an AI component requiring a training set.

9. How the Ground Truth for the Training Set was Established:

  • Not applicable. As there is no AI component or training set mentioned, this information is not relevant to the provided document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the agency's name on the right. The symbol on the left is a stylized depiction of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "FOOD & DRUG" are larger than the other words.

September 21, 2022

Applied Tissue Technologies LLC % Ms. Michele Lucey President Lakeshore Medical Device Consulting LLC 128 Blye HIll Landing Newbury, New Hampshire 03255

Re: K212359

Trade/Device Name: PWD Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: August 19, 2022 Received: August 23, 2022

Dear Ms. Lucey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K212359

Device Name

PWD™ Platform Wound Dressing Negative Pressure System

Indications for Use (Describe)

The Applied Tissue Technologies PWD™ Platform Wound Dressing Pressure System comprised of the; PWD™ Dressing, PWD™ Negative Pressure Wound Therapy Pump, and the PWD™ Wound Exudate Canister is indicated in patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it creates an environment that may promote wound healing by removing excess wound exudate. The Applied Tissue Technologies Platform Wound Dressing is appropriate for use on the following wounds: exudating, chronic, sub-acute and dehisced wounds, partial thickness burns, ulcers (such as diabetic, pressure and venous insufficiencies), flaps, grafts and closed surgical incisions.

PWD Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K212359 Traditional 510(k) SUMMARY

Submitter Information

Submitters' Name:Applied Tissue Technologies LLC
Address:99 Derby Street, Suite 200
Hingham, MA 02043
Telephone:781-366-3848
Contact Person:Ms. Michele Lucey
President, Lakeshore Medical Device Consulting LLC
Address128 Blye Hill Landing,
Newbury, NH 03255
Telephone603-748-1374
Emaillucey_m@msn.com
Date Prepared:September 20, 2022
Device Trade Name:PWD™ Platform Wound Dressing Negative PressureSystem
Classification:Class II
Product Code(s):OMP
Regulation Number(s):21 CFR 878-4780 Powered Suction Pump
Predicate Device (Primary):PICO 7 Single Use Negative PressureWound Therapy System K172005
Predicate Device (Secondary)PWD™ Platform Wound Device K191460

Indications for Use:

The Applied Tissue Technologies PWD™ Platform Wound Dressing Negative Pressure System comprised of the; PWD™ Dressing, PWD™ Negative Pressure Wound Therapy Pump, and the PWD™ Wound Exudate Canister is indicated in patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it creates an environment that may promote wound healing by removing excess wound exudate. The Applied Tissue Technologies Platform Wound Dressing is appropriate for use on the following wounds: exudating, chronic, acute,

{4}------------------------------------------------

K212359

traumatic, sub-acute and dehisced wounds, partial thickness burns, ulcers (such as diabetic, pressure and venous insufficiencies). flaps, grafts and closed surgical incisions.

PWD Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.

Device Description/Technological Characteristics:

The Applied Tissue Technologies PWD™ Platform Wound Dressing Negative Pressure System consists of the PWD™ Dressing, PWD™ Negative Pressure Wound Therapy Pump and the Wound exudate canister. The system provides a negative pressure wound dressing therapy in a moist environment to facilitate the wound healing process. The dressing has an access port through which fluids can be drawn from the wound site. The same port can also be utilized to administer negative pressure wound therapy using the negative pressure pump if required. The negative pressure pump is used with a collection cannister to collect the exudate. The PWD Platform Wound Dressing is supplied in five different shapes and sizes, all having the same intended use. The five sizes are identified as follows.

PWD TM Platform Wound Device Dressing Sizes
DescriptionCatalog/REF #
1" RoundAT1070-01
2" RoundAT1071-01
3" RoundAT1072-01
1" x 3" OblongAT1073-01
3" x 5" OblongAT1074-01

Performance Data:

Bench testing of the PWD™ Platform Wound Dressing Negative Pressure System was performed to evaluate device vapor transmission characteristics, ability to maintain negative pressure, and the ability to remove exudate. Biocompatibility testing was conducted on the dressing component according to the ISO 10993 standard series. The Pump and the Cannister are not tissue contacting. Test results of all nonclinical testing were acceptable and demonstrate that the device is safe and effective for its intended use. A summary of the Non-clinical Bench tests is provided in the following Table:

Performance Test Summary
Test NameDescriptionReferenceStandardsAcceptance CriteriaResults
Water VaporTransmissionRateMeasures thepassage of watervapor through themembrane/adhesivebarrierASTM 1249Must be at leastequivalent tocommerciallyavailable NPWTdressingPass
ExudateSimulated UseN/ASystem mustPass

{5}------------------------------------------------

Performance Test Summary
Test NameDescriptionReferenceStandardsAcceptance CriteriaResults
ThroughputPropertiesUnderNegativePressuretesting using variousexudate types toconfirm acceptablefluid flow whenconnected to thePump and Canistersystemdemonstrate theability to removeexudates of varioustypes from the woundbed
Maintenance ofNegativePressureSimulated use testingfrom under moistconditions todemonstratemaintenance ofnegative pressureNAThe PWD mustmaintain negativepressure at the woundbed for the testdurationPass
UsabilityUsability testing forboth health care andhome useenvironmentsNAUsers mustdemonstrate theability to correctly usethe devicePass
ElectricalSafetyComplies withapplicable safetystandardsIEC 60601-1, IEC60601-1-2,IEC 60601-1-11Requirements forelectrical safety perthe applicable safetystandards is metPass
Electro-magneticCompatibilityComplies withapplicable safetystandardsComplieswith IEC60601-2-3Requirements forelectromagneticcompatibility per theapplicable safetystandards is metPass

Substantial Equivalence:

The Indications for Use statement for the PWD™ Platform Wound Dressing Negative Pressure System is the same as the PICO 7 Single Use Negative Pressure Wound Therapy System.

The following table provides the evidence to further support the claim for substantial equivalence:

{6}------------------------------------------------

Substantial Equivalence Comparison Chart
Feature /SpecificationPWD™ PlatformWound DressingNegative PressureSystemPICO 7 SingleUse NegativePressure WoundTherapy SystemPWDTMPlatform WoundDressingComparison
RegulatoryClearance/ApprovalReferenceN/APrimaryK172005SecondaryK191460N/A
Product CodeOMPOMPOMPSame
RegulationNumber21 CFR 878.478021 CFR878.478021 CFR 878.4780Same
Regulation NamePowered SuctionPumpPowered SuctionPumpPowered SuctionPumpSame
AnatomicallocationBody surface wherewound is presentBody surface wherewound is presentBody surfacewhere wound ispresentSame
Where used(environment)The Negative PressureWound Therapy Systemis suitable for use inboth hospital andhomecaresettings.The NegativePressure WoundTherapy System issuitable for use inboth hospitaland homecaresettings.The NegativePressure WoundTherapy System issuitable for use inboth hospital andhomecare settings.Same
Indicationsfor UseThe Applied TissueTechnologies PWD™Platform WoundDressing NegativePressure Systemcomprised of the;PWD™ Dressing,PWD™ NegativePressure WoundTherapy Pump, and thePWD™ WoundExudate Canister isindicated in patientswho would benefit froma suction device(Negative PressureWound Therapy) as itcreates an environmentthat may promotewound healing byremoving excess woundPICO 7 isindicated forpatients who wouldbenefit from asuction device(Negative PressureWound Therapy) asit may promotewound healing viaremoval of low tomoderate levels ofexudate andinfectiousmaterials.Appropriate woundtypes include:ChronicAcuteTraumaticSubacute anddehisced woundsThe AppliedTissueTechnologiesPWDTM PlatformWound Dressing isintended to beused inconjunction withthe Invia MotionNegative PressureWound Therapy(NPWT) systemand is indicated inpatients whowould benefitfrom a suctiondevice (NPWT) asit creates anenvironment thatmay promotewound healing by

{7}------------------------------------------------

Substantial Equivalence Comparison Chart
Feature /SpecificationPWD™ PlatformWound DressingNegative PressureSystemPICO 7 SingleUse NegativePressure WoundTherapy SystemPWDTMPlatform WoundDressingComparison
exudate. The AppliedTissue TechnologiesPlatform WoundDressing is appropriatefor use on the followingwounds: exudating,chronic, acute,traumatic, sub-acuteand dehisced wounds,partial thickness burns,ulcers (such as diabetic,pressure and venousinsufficiencies), flaps,grafts and closedsurgical incisions.PWD NegativePressure WoundTherapy System aresuitable for use both ina hospital and homecaresetting.Partial-thicknessburns, Ulcers (suchas diabetic orpressure)Flaps and graftsClosed surgicalincisionsPICO Single UseNegative PressureWound TherapySystem is suitablefor usein both a hospitaland homecaresetting.removing excesswound exudate.The AppliedTissueTechnologiesPlatform WoundDressing isappropriate foruse on thefollowingwounds:exudating,chronic, acute,traumatic, sub-acute anddehisced wounds,partial-thicknessburns, ulcers(such as: diabetic,pressure andvenousinsufficiencies),flaps and grafts
MaximumDuration ofa single dressing7 Days7 Days3 DaysSame as theprimarypredicate
NegativePressure to thewound surfacerequirements inmmHgFixed setting at -80mmHg (nominal) to thewound surfaceFixed setting at -80mmHg to the woundsurface-80 mmHgSame
Multiple DressingShapes for WoundSizes (area = L xW)1-3" Round1 x 3" Oblong3 x 5" Oblong10 x 20 cm to 25 x25 cm1-3" Round1 x 3" Oblong3 x 5" OblongPWD issmaller thanthe predicatefor overalldimensions.The dressingis sized forthe wound

{8}------------------------------------------------

Substantial Equivalence Comparison Chart
Feature /SpecificationPWD™ PlatformWound DressingNegative PressureSystemPICO 7 SingleUse NegativePressure WoundTherapy SystemPWD™Platform WoundDressingComparison
size. Same asthe SecondaryPredicate
DressingConfigurationClear polyurethaneembossed occlusivedrape with negativepressure/instillationtubingClear polyurethaneocclusive drapedressing with foamor gauze filler,negativepressure/instillationtubing - filler is cutto sizeClearpolyurethaneembossedocclusive drapewith negativepressure/instillation tubingSame as thesecondarypredicate
Access PortFor attachment to thenegative pressure pump.For removal of fluids.For attachment to thenegative pressurepump. For removalof fluidsFor attachment tothe negativepressure pump.For removal offluids.Same
Collection PackThe dressing tubing isconnected to thecanister/tubing setduring negative pressurewound therapy forcollection of fluids. Thefluids are removed fromthe wound site andsecured within thecannister withoutrequiring additionaldressing changesNo Cannister, relieson dressing changesfor removal ofexudate, frequencydependent on thelevel of exudateThe dressingtubing isconnected to thecanister/tubing setduring negativepressure woundtherapy forcollection offluids. The fluidsare removed fromthe wound siteand securedwithin thecannister withoutrequiringadditionaldressing changesSame as thesecondarypredicate.
PumpPWD™ NegativePressure WoundTherapy PumpPICO 7NASame
Pump TypeCustom designed pumpcontrolled by amicroprocessorCustom designedpump controlled bya microprocessorNASame

{9}------------------------------------------------

Substantial Equivalence Comparison Chart
Feature /SpecificationPWD™ PlatformWound DressingNegative PressureSystemPICO 7 SingleUse NegativePressure WoundTherapy SystemPWD™Platform WoundDressingComparison
How SuppliedPWD Dressing issupplied Sterile SingleUse Only. The pumpand cannister aresupplied non-sterilesingle use onlyThe system issupplied sterile,single use onlyPWD Dressing issupplied SterileSingle Use Only.The PWDpump andcannister aresuppliednonsterile –similar toother systemswith the sameintended use.Thisdifferencedoes not raisenew questionsof safety.
TissueContactMaterialsPolymers and AdhesivePolymers andAdhesivePolymers andAdhesiveSame as thesecondarypredicate
Additives -antimicrobial,animal originNoNoNoSame
SterilizationEto at an SAL 10-6EtoEto at an SAL 10-6Same.
Electrical SafetyTestingComplies with IEC60601-1, IEC 60601-1-2, IEC 60601-1-11Complies with IEC60601-1, IEC 60601-1-2, IEC 60601-1-11NASame
ElectromagneticCompatibilityComplies with IEC60601-2-3Complies with IEC60601-2-3NASame
MR SafetyPump is MR UnsafePump is MR UnsafeNASame

Conclusion:

When considering design, feature, materials, and performance characteristics the PWD™

PNegPlatform Wound Dressing Negative Pressure System is substantially equivalent to the predicate device, the PICO 7 Single Use Negative Pressure Wound Therapy System K172005

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.