K Number
K212295

Validate with FDA (Live)

Device Name
21HQ513D
Date Cleared
2021-12-03

(134 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

Device Description

The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images

AI/ML Overview

The provided text describes the 510(k) summary for the LG Electronics Inc. device, 21HQ513D, a medical monitor. The document asserts the device's substantial equivalence to a predicate device (K191864, 21HK512D) rather than detailing a specific study to prove it meets acceptance criteria with quantitative performance metrics for diagnostic accuracy.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly defined as pass/fail thresholds in a table. Instead, the document focuses on demonstrating that the proposed device, 21HQ513D, has similar technological characteristics and performance to the predicate device, 21HK512D, through comparative measurements. The "Reported Device Performance" listed below are the characteristics of the proposed device that were compared against the predicate.

Measurement / CharacteristicDescriptionProposed Device (21HQ513D) PerformancePredicate Device (21HK512D) PerformanceSE Note (Comparison to Predicate)
Technological Characteristics
Power ConsumptionElectrical power usage of the device.MAX. 120W, Off Mode ≤ 0.3WMAX. 85WDifference (not affecting safety/reliability)
LCD ScreenDisplay panel technology.TFT LCDTFT LCDSame
Pixel PitchDistance between the centers of two adjacent pixels.0.2115 x 0.2115 mm0.2115 x 0.2115 mmSame
ResolutionNumber of distinct pixels in each dimension that can be displayed.1,536 x 2,048 pixels1,536 x 2,048 pixelsSame
Horizontal FrequencyRate at which the display scans horizontally.30 kHz to 130 kHz63 kHz to 96 kHzDifference (not affecting safety/reliability)
Vertical FrequencyRate at which the display refreshes the image vertically.56 Hz to 61 Hz50 Hz to 75 HzDifference (not affecting safety/reliability)
Input Video SignalsTypes of video input connections supported.DisplayPort x 2, DVI-IN x 1DisplayPort x 1, DVI-IN x 1Difference (not affecting safety/reliability)
Calibration ToolSoftware used for display calibration.PerfectLum 4.0PerfectLum 3.9Difference (updated version with UI/UX, OS, functional improvements, validated to ISO 62304)
Performance Items (Non-Clinical Test Summary)
Spatial resolutionMeasurements of information transfer from image data to luminance fields at different spatial frequencies, using TG18 QC Test Pattern.Not explicitly quantified values, but "Same" implying comparable performance to predicate.Assumed comparable to predicateSame
Pixel defectsMeasurements (count, types, locations) of pixel defects.Not explicitly quantified values, but "Equivalent" implying acceptable limits comparable to predicate.Assumed comparable to predicateEquivalent
ArtifactsEvaluation for image artifacts such as ghosting and/or image sticking from displaying a fixed test pattern (5x5 mosaic pattern, 64 Gray / 127 Gray judgment).Not explicitly quantified values, but "Same" implying comparable performance to predicate.Assumed comparable to predicateSame
Temporal responseMeasurements of the temporal behavior of the display in responding to changes in image values from frame to frame (rise and fall time constants).Not explicitly quantified values, but "Equivalent" implying acceptable limits comparable to predicate.Assumed comparable to predicateEquivalent
LuminanceMeasurements of the maximum and minimum luminance the device outputs under recommended conditions.Not explicitly quantified values, but "Same" implying comparable performance to predicate.Assumed comparable to predicateSame
Color trackingChromaticity at different luminance levels of primary colors and color gamut (e.g., CIE u'v').Not explicitly quantified values, but "Equivalent" implying acceptable limits comparable to predicate.Assumed comparable to predicateEquivalent
Conformance to a grayscale-to-luminance functionMeasurements of the mapping between image values and the luminance output following a target model response for 256 or more levels.Not explicitly quantified values, but "Equivalent" implying acceptable limits comparable to predicate.Assumed comparable to predicateEquivalent

2. Sample Size Used for the Test Set and the Data Provenance

The document does not describe a clinical test set with a specific sample size. The testing performed was primarily non-clinical, comparing the technical characteristics and performance measures of the single proposed device (21HQ513D) against the predicate device (21HK512D).

  • Sample Size: Not applicable for a typical clinical test set as this was a non-clinical comparison of device specifications.
  • Data Provenance: The tests were conducted by LG Electronics Inc., which is based in South Korea. The nature of these tests (e.g., in-house lab testing) suggests the data provenance is internal to the manufacturer. The document doesn't specify if it's retrospective or prospective, but for non-clinical performance testing of a monitor, it would typically be prospective testing of the manufactured device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. This submission is for a medical monitor, not an AI diagnostic algorithm. The "test set" here refers to the monitor's performance characteristics, not a dataset of medical images requiring expert ground truth for diagnostic accuracy.

4. Adjudication Method for the Test Set

Not applicable. There was no clinical test set requiring ground truth adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was done, as this device is a medical monitor, not an AI-assisted diagnostic tool. No AI component is described as being part of the device's functionality.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

Not applicable. This device is a medical monitor, not a standalone AI algorithm. It displays images for human practitioners.

7. The Type of Ground Truth Used

Not applicable. The "ground truth" in this context refers to established engineering standards (e.g., IEC standards) and guidance documents (e.g., FDA's Display Devices for Diagnostic Radiology guidance) against which the monitor's physical and technical performance (spatial resolution, luminance, etc.) are measured.

8. The Sample Size for the Training Set

Not applicable. As noted, this is a medical monitor, not an AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. No training set was used.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA acronym is in a blue square, and the full name of the agency is in blue text.

LG Electronics Inc. % DoGyun Im Senior Researcher GMS Consulting 4th Floor, Digital Cube, 34, Sangamsan-ro Seoul, Mapo-gu 03909 Korea, South

December 7, 2021

Re: K212295

Trade/Device Name: 21HO513D Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: PGY Dated: November 3, 2021 Received: November 4, 2021

Dear DoGyun Im:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 3, 2021. Specifically, FDA is updating this SE Letter as an administrative correction. The missing information for the "Dated" field in the original letter should have been November 3, 2021. Additionally, the trade name should have been listed as 21HQ513D.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Dr. Jessica Lamb, Assistant Director, Office of in vitro Diagnostics and Radiological Health at 301-796-6167 or Jessica.Lamb@fda.hhs.gov.

Sincerely

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Mammography Ultrasound and Imaging Software Branch Division of Radiological Health OHT7: Office of in vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

LG Electronics Inc. % Dogyun Im Senior Researcher GMS Consulting 4th Floor, Digital Cube, 34, Sangamsan-ro Seoul, Mapo-gu 03909 Korea, South

Re: K212295

Trade/Device Name: 21hq513d Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: PGY Dated: [NOTE: Use date of most recent supplement] Received: November 4, 2021

Dear Dogyun Im:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

December 3, 2021

{2}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb Assistant Director, Mammography Ultrasound and Imaging Software Branch Division of Radiological Health OHT7: Office of in vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K212295

Device Name 21HQ513D

Indications for Use (Describe)

This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

[As Required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(a)]

July 20, 2021

.

2. Submitter's Information [21 CFR 807.92(a)(1)]

Name of Sponsor:LG Electronics Inc.77, Sanho-daero, Gumi-si, Gyeongsangbuk-do, 39381,Republic of Korea
Name of Manufacturer:LG Electronics Inc.77, Sanho-daero, Gumi-si, Gyeongsangbuk-do, 39381,Republic of Korea
  • Contact Name: Jinhwan Jun / Chief Research Engineer Telephone No.: +82-31-8066-5641 -
  • Email Address: jinhwan.jun@lge.com

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

  • Trade Name: ●
    21HQ513D

  • Common Name: ●
    Medical Monitor

  • Classification: ●

Classification NameMedical image management and processing system
Classification Number21 CFR 892.2050
Product CodePGY
Device ClassII
Review PanelRadiology

{5}------------------------------------------------

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]

The identified predicate devices within this submission are shown as follow;

Predicate Device

  • 510(k) Number:
  • Applicant:
  • Classification Name:
  • Trade Name:

LG Electronics Inc. Medical image management and processing system 21HK512D

5. Description of the Device [21 CFR 807.92(a)(4)]

The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images

K191864

6. Indications for use [21 CFR 807.92(a)(5)]

This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

7. Intended Use [21 CFR 807.92(a)(5)]

The 21HQ513D is a prescription device, and the display is not intended for mammography.

  • -The display is indicated for use to display radiological images for review, analysis, and diagnosis.
  • -The display is indicated for use by trained medical practitioners.
  • -The display is indicated to provide high-resolution color and grayscale medical imaging for PACS and Radiology systems.

{6}------------------------------------------------

8. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]

The table below presents comparisons between the subject device (21HQ513D) and the legally marketed predicate device (K191864):

Proposed DevicePredicate Device
K NumberNot knownK191864
ManufacturerLG Electronics Inc.LG Electronics Inc.
Model Name21HQ513D21HK512D
Classification NameMedical image management andprocessing systemMedical image management andprocessing system
Classification Number21 CFR 892.205021 CFR 892.2050
Indications for UseThis Medical Monitor is indicatedfor use in displaying radiologicalimages for review, analysis, anddiagnosis by trained medicalpractitioners. The display is notintended for mammography.This Medical Monitor is intendedto be used by trained medicalpractitioners for displaying,reviewing, and analysis ofmedical images. The display isnot intended for mammography.
Power ConsumptionMAX. 120WOff Mode ≤ 0.3WMAX. 85W
LCD ScreenTFT LCDTFT LCD
Pixel Pitch0.2115 x 0.2115 mm0.2115 x 0.2115 mm
Resolution1,536 x 2,048 pixels1,536 x 2,048 pixels
Horizontal Frequency30 kHz to 130 kHz63 kHz to 96 kHz
Vertical Frequency56 Hz to 61 Hz50 Hz to 75 Hz
Input video signalsDisplayPort x 2DVI-IN x 1DisplayPort x 1DVI-IN x 1
Calibration ToolPerfectLum 4.0PerfectLum 3.9

The comparison table shows that the subject device (21HQ513D) has the same indications for use the predicate one. Although the devices have some different technological characteristics (power consumption, horizontal & vertical frequency, input video signals), these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does. And there is a difference in the calibration tool. 21HQ513D uses the updated version of the PerfectLum. PerfectLum 4.0 only have differences in UI/UX improvements, supported OS extensions, functional improvements for remote control server. So it does not affect the calibration function of the PerfectLum. And PerfectLum 4.0 was also validated according to the ISO 62304. Therefore, there are no significant differences in the technological characteristics of the subject device. All the differences between the subject and predicate device do not raise different questions of safety and effectiveness. It is substantially equivalent to a predicate device in indications for use and technology characteristics.

{7}------------------------------------------------

9. Non-Clinical Test summary

    1. Electrical Safety and Electromagnetic Compatibility
    • The test results demonstrated that the proposed device complies with the following standards: · Electrical Basic Safety and Essential Performance reguirements in accordance with IEC 60601-
    • 1:2005/AMD1:2012 • Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2 Edition 4.0:2014
    • · Medical Electrical Equipment Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability in accordance with IEC 60601-1-6:2010/A1:2013
    1. Software Validation

The 21HQ513D contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated. The software information is provided in accordance with FDA quidance: The content of premarket

    1. Guidance
      Display Devices for Diagnostic Radiology – Guidance for Industry and Food and Drug Administration Staff, issued on October 2, 2017
MeasurementsDescriptionSE Note
a.Spatial resolutionMeasurements of the transfer ofinformation from the image data to theluminance fields at different spatialfrequencies of interest typically done byreporting the modulation transferfunction. Non-isotropic resolutionproperties should be characterizedproperly by providing two-dimensionalmeasurements or measurements alongat least two representative axes.(Using TG18 QC Test Pattern)Same
b.Pixel defectsMeasurements (count, types (e.g., sub-pixel or entire pixel, always-on, always-off), and locations (map) of pixeldefects. This is typically provided as atolerance limit. Pixel defects caninterfere with the visibility of smalldetails in medical images.Equivalent
c.ArtifactsEvaluate for image artifacts such asghosting and/or image sticking fromdisplaying a fixed test pattern for aperiod of time.(Using 5x5 mosaic pattern, 64Gray /127 Gray judgment)Same
d.Temporal responseMeasurements of the temporal behaviorof the display in responding to changesin image values from frame to frame.Since these transitions are typically notsymmetric, rise and fall time constantsEquivalent

According to the quidance, we test following performance items:

submissions for software contained in medical devices, on May 11, 2005.

{8}------------------------------------------------

MeasurementsDescriptionSE Note
are needed to characterize the system.Slow displays can alter details andcontrast of the image when large imagestacks are browsed or in video,panning, and zooming modes.
e.LuminanceMeasurements of the maximum andminimum luminance that the deviceoutputs as used in the application underrecommended conditions and theachievable values if the device is set toexpand the range to the limit.Same
f.Color trackingChromaticity at different luminancelevels of primary colors as indicated bythe color coordinates in an appropriateunits system (e.g., CIE u'v') and thecolor gamut enveloped by the primarycolors.Equivalent
g.Conformance to agrayscale-to-luminancefunctionMeasurements of the mapping betweenimage values and the luminance outputfollowing a target model response for256 or more levels.Equivalent

According to the above test results, there are no significant performance differences between 21HQ513D and the predicate device that would adversely affect the use of the product. It has substantially equivalent performance compared to the predicate device.

Clinical Test Summary:

No clinical studies were considered necessary and performed.

10. Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification LG Electronics, concludes that the 21HQ513D is substantially equivalent in safety and effectiveness to the predicate devices as described herein.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).