K Number
K212287
Device Name
Dayspring Lite
Manufacturer
Date Cleared
2021-09-21

(62 days)

Product Code
Regulation Number
870.5800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Dayspring Lite is a prescription only wearable compression system that is intended for use in a clinic or home setting by medical professionals and patients who are under medical supervision, for the following conditions:

  • Chronic edema
  • Lymphedema
  • Venous insufficiency
  • Wound healing
    Dayspring Lite is developed on a wearable compression technology platform, which is designed to provide mobility for patients.
Device Description

Dayspring Lite consists of two main components: a controller and garment. The garment is powered by an active smart compression technology that is instant-acting, and silent. This technology uses a Nickel Titanium (Ni-Tì) shape-memory alloy that is activated by the controller. A liner is worn as an accessory under the device to prevent direct patient contact with the garment is wrapped around the patient's affected area so that the device fits snugly. The device has up to 14 independently controlled sections in each limb. The controller is pre-programmed to provide sequential compression therapy to the affected area. The device is powered by a rechargeable Lithium-ion battery pack. The device was developed to provide patients with untethered access and a functional range of motion and mobility.

AI/ML Overview

Based on the provided text, the document is a 510(k) Premarket Notification from the FDA regarding the Koya Medical, Inc. Dayspring Lite device. This document primarily details the regulatory review process, device description, comparison to predicate devices, and various performance testing conducted to ensure substantial equivalence for a medical device. It does not describe a study involving an AI algorithm or its performance parameters in a diagnostic or clinical decision support context.

Therefore, I cannot provide a detailed answer to your request regarding acceptance criteria and a study proving an AI device meets those criteria, as the provided text pertains to a compression system and does not mention any AI components or performance studies typical of AI/ML-enabled medical devices.

The sections you've asked for (such as ground truth, expert consensus, MRMC studies, effect size of AI assistance, standalone algorithm performance, training set details) are characteristic of studies conducted for AI/ML medical devices, which are not present in this document.

In summary, the provided document does not contain the information required to answer your prompt because it is not about an AI-powered device or a study involving AI performance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 21, 2021

Koya Medical, Inc. % Alex Chang Regulatory Affairs Consultant BioDesign Regulatory Services, LLC 16185 Los Gatos Blvd. Los Gatos, California 95032

Re: K212287

Trade/Device Name: Dayspring Lite Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: August 23, 2021 Received: August 24, 2021

Dear Alex Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Nicole Gillette Assistant Director (Acting) DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K212287

Device Name Dayspring Lite

Dayspring Lite

Indications for Use (Describe)

Dayspring Lite is a prescription only wearable compression system that is intended for use in a clinic or home setting by medical professionals and patients who are under medical supervision, for the following conditions:

  • · Chronic edema
  • · Lymphedema
  • · Venous insufficiency
  • · Wound healing

Dayspring Lite is developed on a wearable compression technology platform, which is designed to provide mobility for patients.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. 510(K) SUMMARY

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

I. SUBMITTER:Koya Medical, Inc.
2461 Peralta St. Oakland CA 94607 USA
Establishment Registration: 3017424826
CONTACT:Alex Chang
Regulatory Consultant
Phone: 408 839 5826
Fax; 510 217 2340
E-mail: a.chang@biodesign-rac.com
DATE PREPARED:Aug 20, 2021
II. DEVICE:
TRADE NAME:DAYSPRING LITE
CLASSIFICATION NAME:COMPRESSIBLE LIMB SLEEVE (21 CFR 870.5800)
DEVICE CLASSIFICATION:CLASS II
PRODUCT CODE:JOW
III. PREDICATE DEVICES:K143185, K210885

IV. DEVICE DESCRIPTION:

Dayspring Lite consists of two main components: a controller and garment. The garment is powered by an active smart compression technology that is instant-acting, and silent. This technology uses a Nickel Titanium (Ni-Tì) shape-memory alloy that is activated by the controller. A liner is worn as an accessory under the device to prevent direct patient contact with the garment is wrapped around the patient's affected area so that the device fits snugly. The device has up to 14 independently controlled sections in each limb. The controller is pre-programmed to provide sequential compression therapy to the affected area. The device is powered by a rechargeable Lithium-ion battery pack. The device was developed to provide patients with untethered access and a functional range of motion and mobility.

{4}------------------------------------------------

V: INDICATION FOR USE:

Dayspring Lite is a prescription only wearable compression system that is intended for use in a clinic or home setting by medical professionals and patients who are under medical supervision, for the treatment of the following conditions:

  • . Chronic edema
  • . Lymphedema
  • Venous insufficiency .
  • Wound healing

Dayspring Lite is developed on a wearable compression technology platform, which is designed to provide mobility for patients.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE

DEVICE:

FeatureSubject DevicePrimaryPredicate Device(K210885)AdditionalPredicate Device(K143185)
ElectricalRequirementsRechargeable Li-ionBattery Pack, ACCharging AdapterRechargeable Li-ionBattery Pack, ACCharging Adapter100-240 VAC50/60 Hz to ACAdapter withoutput voltage of12.0V DC and 3.0A
OutputSequential gradientPressureSequentialcalibrated gradientPressureSequential gradientPressure
Mechanism ofActionExertion ofsequential pressureto affected areaExertion ofsequential pressureto affected areaExertion ofsequential pressureto affected area
Principles ofOperationLithium-ion batterypowered integratedshape memory alloychannels creatingcompressivepressureLithium-ion batterypowered integratedshape memory alloychannels creatingcompressivepressureElectricallypowered integratedpneumatic airchannels creatingcompressivepressure
ControllerEnclosure MaterialAll plasticconstructionAll plasticconstructionAll plasticconstruction
User InterfacePushbuttons.Also available isBluetooth LowEnergy (BLE)Pushbuttons.Also available isBluetooth LowEnergy (BLE)PushbuttonsMobile applicationor BLE notavailable

{5}------------------------------------------------

Module forcommunication withmobile applicationon mobile deviceModule forcommunicationwith mobileapplication onmobile device
Software/HardwareAnalog and digitalelectronic withmicroprocessorAnalog and digitalelectronic withmicroprocessorAnalog and digitalelectronic withmicroprocessor
Garment MaterialNylon fabric withvelcro strapsNylon fabric withvelcro strapsNylon fabric withvelcro straps
Stockinette/LinerClass Ibiocompatible linerprovided with theunitClass Ibiocompatible linerprovided with theunitClass Ibiocompatible linerprovided with theunit

VII: PERFORMANCE DATA:

The following performance data were provided in support of substantial equivalence determination.

Biocompatibility Testing

The subject device is considered a surface contacting device with prolonged exposure duration considering potential cumulative use. The surface contacting material described was previously evaluated for biocompatibility per ISO 10993-1, ISO 10993-5, and ISO 10993-10.

Sterilization

The subject device is non-sterile, and components are unlikely to deteriorate with age. Accelerated 1 year accelerated shelf-life testing was performed and shown to support shelf stability.

Electrical Safety and Electromagnetic Compatibility (EMC)

The subject device was evaluated based on the following applicable performance and safety standards: IEC 60601-1:2012, IEC 60601-1-11:2015 and IEC 60601-1-2:2014. Results demonstrated that the subject device was compliant to all applicable performance and safety standards.

Software Verification and Validation Testing

The subject device includes embedded firmware in the controller which has the ability to connect with custom software installed on a mobile device. The system software exhibits a moderate level of concern.

{6}------------------------------------------------

Software lifecycle planning and documentation as well verification and validation testing were performed in accordance with IEC 62304:2015 and as recommended by the following FDA Guidance documents for Industry and FDA Staff:

  • . Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
  • . Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices
  • . Guidance for Industry and Food and Drug Administration Staff Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
  • Guidance for Industry, FDA Reviewers and Compliance on Post market . Management of Cybersecurity in Medical Devices

In accordance with IEC 62304:2015 and FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, the following software lifecycle documentation has been developed. Testing demonstrated that the subject device met all software requirements.

DocumentStatus
Software Development PlanCompleted
Software Requirements Specifications (SRS)Completed
Software Design Specification (SDS)Completed
Off-the-Shelf Software ManagementCompleted
Software Configuration ManagementCompleted
Software ArchitectureCompleted
Cybersecurity AnalysisCompleted
Software Coding StandardCompleted
Traceability AnalysisCompleted
Software Verification and ValidationCompleted / Pass

Summary of Performance Testing

The subject device was evaluated based on the following benchtop performance tests.

{7}------------------------------------------------

Performance StudyOverviewStatus
Storage and Shelf-Life Stability1 Year aging guided by Q10Theory & ASTM F1980Completed / Pass
Transportation SimulationASTM D4169Completed / Pass
Pressure Verification TestingPressure capable ofdelivering 0-100 mmHgcompression pressures.Completed / Pass

Packaging, Shelf Life and Transport Stability Testing

The subject device was packaged in a corrugated shipper. The packaging configuration was evaluated based on ASTM D4169. Prior to running the transit simulation, the test sample underwent 1 year accelerated aging based on ASTM F1980 - Standard Guide for Accelerated Aging of Medical Device Packages and was subjected to environmental conditioning to confirm shelf-life stability. Post-simulation, functional testing of the subject device was performed and showed that it continued to meet all functional specifications.

Benchtop Pressure Testing

Benchtop pressure testing was performed to verify the pressure range applied by the subject device was equivalent to pressure applied by the K143185 and K210885 predicate systems.

Test Summary

The subject device has been investigated and tested against and complies with the following voluntary standards:

StandardsStandardsOrganizationStandards Title
60601-1:2005 +CORR: 1:2006 +CORR. 2:2007 +A1:2012 (Edition 3.1)IECMedical electrical equipment - Part 1: Generalrequirements for basic safety and essentialperformance
60601-1-11:2015IENRequirements for medical electrical equipment andmedical electrical systems used in the homehealthcare environment

{8}------------------------------------------------

60601-1-2:2014(Edition 4.0)IECMedical electrical equipment – Part 1-2: Generalrequirements for basic safety and essentialperformance - Collateral standard: Electromagneticdisturbances – Requirements and tests
62304:2015(Edition 1.1)IECMedical devices software –software life cycleprocesses
10993-1:2018ISOBiological evaluation of medical devices — Part 1:Evaluation and testing within a risk managementprocess
10993-5: 2009ISOBiological evaluation of medical devices - Part 5:Tests for in vitro cytotoxicity
10993-10: 2010ISOBiological evaluation of medical devices - Part 10:Tests for irritation and skin sensitization
D4169-16ASTMStandard Practice for Performance Testing ofShipping Containers and Systems
14971:2013ISOMedical devices – Application of risk managementto medical devices

Testing has been performed and all components, subassemblies and/or full devices and systems have met the required specifications for the completed tests.

Animal Study

Animal performance testing was not required to demonstrate safety and effectiveness of the device.

Clinical Study

Clinical testing was not required to demonstrate the safety and effectiveness of the Koya Dayspring Lite. Instead, substantial equivalence is based upon benchtop performance testing.

VIII. CONCLUSION:

The data included in this submission demonstrates that Dayspring Lite is substantially equivalent to the cleared and marketed primary predicate for its intended use.

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).