(274 days)
EVA Ophthalmic Surgical System (K142877)
Visalis V500, Visalis S500 (K193376)
No
The summary describes a standard phacoemulsification system with fluidic, ultrasound, and pneumatic modules controlled by the surgeon via a GUI and foot pedal. There is no mention of AI, ML, image processing, or data-driven decision-making. The software testing follows standard medical device software lifecycle processes (IEC 62304) and usability standards (IEC 62366), not AI/ML-specific validation.
Yes
The device is intended for surgical procedures like phacoemulsification and anterior vitrectomy, which treat medical conditions (cataracts), thus qualifying it as a therapeutic device.
No
Explanation: The QUATERA 700 is described as a phacoemulsification system intended for emulsification and removal of cataracts and anterior segment vitrectomy. Its functions are surgical, not diagnostic.
No
The device description explicitly mentions "fluidic, ultrasound and pneumatic modules" and a "Foot Control Panel," indicating the presence of hardware components beyond just software. The performance testing also includes electrical safety, biocompatibility, and sterility testing, which are relevant to hardware.
Based on the provided information, the QUATERA 700 is not an IVD (In Vitro Diagnostic) device.
Here's why:
- Intended Use: The intended use clearly states that the device is for "emulsification and removal of cataracts and anterior segment vitrectomy" and is used "in anterior segment surgery." This describes a surgical procedure performed directly on a patient's eye.
- Device Description: The description details a "mobile phacoemulsification system" used in "ophthalmic surgical operating rooms" to perform procedures like "irrigation and/or aspiration, phacoemulsification of crystalline lens, anterior vitrectomy, and bipolar coagulation." These are all surgical interventions.
- Lack of In Vitro Activity: An IVD device is designed to examine specimens derived from the human body (like blood, urine, tissue) in vitro (outside the body) to provide information for diagnosis, monitoring, or screening. The QUATERA 700 does not perform any such analysis of biological specimens. It is a surgical tool used in vivo (within the body).
The device is a surgical system used for ophthalmic procedures, not for diagnostic testing of biological samples.
N/A
Intended Use / Indications for Use
QUATERA 700 is intended for the emulsification and removal of cataracts and anterior segment vitrectomy. In combination with various required components and accessories, the device is designed for use in anterior segment surgery. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation, bipolar coagulation and anterior vitrectomy.
Product codes
HQC, HQE
Device Description
QUATERA 700 is a mobile phacoemulsification system designed for use in the ophthalmic surgical operating rooms during surgery of the anterior eye segment. When QUATERA 700 is used with compatible components and accessories, the system will perform the following surgical procedures: irrigation and/or aspiration, phacoemulsification of crystalline lens, anterior vitrectomy, and bipolar coagulation.
QUATERA 700 has fluidic, ultrasound and pneumatic modules for emulsification and aspiration of the cataractous lens from eye and maintain the pressure and volume of the eye intraoperatively. The required values are pre-set via a Graphical User Interface and controlled directly by the surgeon using the Foot Control Panel of the device and delivered in the eye via a range of accessories. The systems control mechanism verifies the output values and the pre-set values.
QUATERA 700 has the following functions:
- Irrigation and Aspiration
- . Ultrasound Capability
- Diathermy
- Anterior Vitrectomy ●
- . Reflux
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
anterior eye segment
Indicated Patient Age Range
Adults
Intended User / Care Setting
QUATERA 700 is intended to be used within a clinic(s)/hospital(s)/surgical practice network.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-Clinical Performance Testing: Non-clinical system testing provided an evaluation of the performance of the system relevant to each of the system specifications. The functional and system level testing showed that the defined specifications.
Biocompatibility & Animal Testing: Biocompatibility assessments and testing for patient contacting components/accessories was performed in accordance with the following standards: ISO 10993-1, ISO 10993-7, ISO 10993-10, ISO 10993-11, and ISO 10993-12, USP 34 . Specifications regarding intracutaneous reactivity or acute ocular irritation tested in rabbits (ISO 10993-10), material mediated pyrogenicity tested in rabbits (USP 34 ), acute systemic cytotoxicity tested in mice (ISO 10993-11), kligman maximization (i.e., skin sensitization) tested in guinea pigs (ISO 10993-10) have been met. Testing passed.
Sterility Testing: The sterilization cycle B-01 /B-01x was validated to achieve a sterility assurance level (SAL) of 10-9 according to ISO 11135:2014. Additionally, visual inspection, micro-biological dusting, seal strength, dye penetration and bubble leak testing was performed to ensure micro-biological barrier function met specifications. Testing passed. After the sterilization process was completed, testing on the patient contacting components for the accessories to measure ethylene oxide (EO) and ethylene chlorohydrin residual (ECH) levels, as well as Limulus Amebocyte Lysate (LAL) levels was conducted. Testing passed.
Software Testing: Subject Device was tested according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005). In addition, the software testing also followed the Carl Zeiss Meditec internal software development procedure that follows the IEC 62304:2006+AC:2015 - Medical device software - Software life cycle processes. Validation has been conducted according to IEC 62366. Testing passed.
Electrical Safety/EMC Testing: Electrical, Thermal, EMC testing was conducted in accordance with IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-2 standards. Testing passed.
Additional Bench Testing: Conformance to IEC 80601-2-58 and IEC 60529 was demonstrated by testing for occlusion break response, fluidics assessments, spring-eye model, and emersion testing was performed. Testing passed.
Clinical Performance Testing: Clinical studies were not required since the system does not introduce a new methodology or new/expanded clinical claims.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
EVA Ophthalmic Surgical System (K142877)
Reference Device(s)
Visalis V500, Visalis S500 (K193376)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 886.4670 Phacofragmentation system.
(a)
Identification. A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font size.
April 19, 2022
Carl Zeiss Meditec AG % Maria Golovina Head of Regulatory Affairs - USA Carl Zeiss Meditec Inc 5300 Central Parkway Dublin, California 94568
Re: K212241
Trade/Device Name: QUATERA 700 Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System Regulatory Class: Class II Product Code: HQC, HQE Dated: March 9, 2022 Received: March 10, 2022
Dear Maria Golovina:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Anjana Jain, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K212241
Device Name QUATERA 700
Indications for Use (Describe)
QUATERA 700 is intended for the emulsification and removal of cataracts and anterior segment vitrectomy. In combination with various required components and accessories, the device is designed for use in anterior segment surgery. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation, bipolar coagulation and anterior vitrectomy.
Type of Use (Select one or both, as applicable) | |
---|---|
✓ Prescription Use (Part 21 CFR 801 Subpart D) | ❍ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/0 description: The image shows the word "ZEISS" in white text on a blue background. The text is in a bold, sans-serif font. The background is a solid, dark blue color. The logo is simple and modern.
In accordance with 21 CFR 807.92 the 510(k) Summary for the QUATERA 700 is provided below.
1. SUBMITTER
Applicant: | Carl Zeiss Meditec AG |
---|---|
Goeschwizer Strasse 51-52 | |
D-07745 Jena | |
Germany | |
Primary Correspondent | Maria Golovina |
Head of Regulatory Affairs - USA | |
Carl Zeiss Meditec, Inc. | |
5300 Central Parkway Dublin, CA 94568 | |
(925) 216-1078 Phone (925) 557-4259 Fax | |
E-mail: maria.golovina@zeiss.com (preferred) |
Date Prepared:
April 18, 2022
4
Image /page/4/Picture/0 description: The image shows the ZEISS logo. The logo consists of the word "ZEISS" in a bold, sans-serif font, set against a solid blue background. The bottom of the blue background has a curved shape, giving it a distinctive appearance.
2. SUBJECT DEVICE
Device Trade Name: | QUATERA 700 |
---|---|
Classification: | 21CFR886.4670 Phacofragmentation System |
Regulatory Class: | II |
Product Code: | HQC, HQE |
PREDICATE DEVICE AND REFERENCE DEVICE 3.
3.1. PREDICATE DEVICE
Predicate Device: | EVA Ophthalmic Surgical System (K142877) |
---|---|
Classification: | 21 CFR 886.4670 Phacofragmentation System |
Regulatory Class: | II |
Product Code: | HQC, HQE, HQF |
3.2. REFERENCE DEVICE
Predicate Device: | Visalis V500, Visalis S500 (K193376) |
---|---|
Classification: | 21 CFR 886.4670 Phacofragmentation System |
Regulatory Class: | II |
Product Code: | HQC |
5
Image /page/5/Picture/0 description: The image features the word "ZEISS" in a bold, sans-serif font. The word is displayed in white against a solid blue background. The logo is simple and modern, with a focus on the brand name.
DEVICE DESCRIPTION 4.
QUATERA 700 is a mobile phacoemulsification system designed for use in the ophthalmic surgical operating rooms during surgery of the anterior eye segment. When QUATERA 700 is used with compatible components and accessories, the system will perform the following surgical procedures: irrigation and/or aspiration, phacoemulsification of crystalline lens, anterior vitrectomy, and bipolar coagulation.
QUATERA 700 has fluidic, ultrasound and pneumatic modules for emulsification and aspiration of the cataractous lens from eye and maintain the pressure and volume of the eye intraoperatively. The required values are pre-set via a Graphical User Interface and controlled directly by the surgeon using the Foot Control Panel of the device and delivered in the eye via a range of accessories. The systems control mechanism verifies the output values and the pre-set values.
QUATERA 700 has the following functions:
- Irrigation and Aspiration
- . Ultrasound Capability
- Diathermy
- Anterior Vitrectomy ●
- . Reflux
QUATERA 700 is intended to be used within a clinic(s)/hospital(s)/surgical practice network.
INDICATIONS FOR USE న్.
QUATERA 700 is intended for the emulsification and removal of cataracts and anterior segment vitrectomy. In combination with various required components and accessories, the designed for use in anterior segment surgery. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation/aspiration, bipolar coagulation and anterior vitrectomy.
This device is Prescription Use (Rx) only.
SUBSTANTIAL EQUIVALENCE TO PRIMARY PREDICATE 6.
Table 1. Subject to Predicate Device Comparison Table - Indications for Use
| Subject Device | Primary Predicate Device
(K142877) | Equivalency Analysis |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| QUATERA 700 is intended for the
emulsification and removal of cataracts and
anterior segment vitrectomy. In combination
with various required components and | The EVA Ophthalmic Surgical
System is indicated for both
anterior segment (i.e.) | The indications for use are
equivalent as basis of the
medical context. The
different wording leads to |
6
Image /page/6/Picture/0 description: The image contains the word "ZEISS" in a bold, sans-serif font. The word is white and is set against a solid blue background. The logo is simple and modern, with a clean and professional look.
Page 4 of 7
| Subject Device | Primary Predicate Device
(K142877) | Equivalency Analysis |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| accessories, the device is designed for use in
anterior segment surgery. It provides
capabilities for phacoemulsification, coaxial
and bimanual irrigation/aspiration, bipolar
coagulation and anterior vitrectomy. | phacoemulsification and
removal of cataracts) and
posterior segment (i.e.,
vitreoretinal)
ophthalmic surgery. | equivalent therapy options,
however the indications
provided by QUATERA
700 is clearer. |
Table 2. Subject to Predicate Device Comparison Table – Technical Characteristics
| Attribute | Subject Device | Primary Predicate Device
(K142877) | Equivalency
Analysis |
|--------------------------------|-----------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Device name | QUATERA System
(QUATERA 700) | EVA Ophthalmic Surgical
System | N/A |
| Manufacturer | Carl Zeiss Meditec AG | D.O.R.C. Dutch Ophthalmic
Research Center (International)
B.V. | N/A |
| 510(k) | K212241 | K142877 | N/A |
| Classification Product
Code | HQC, HQE | HQC, HQE, HQF | Equivalent |
| Regulation # | 21CFR886.4670 | 21CFR886.4670 | Identical |
| Application | Phacofragmentation System
Ophthalmic Surgery | Phacofragmentation System
Ophthalmic Surgery | Identical |
| Combination Device | No | No | Identical |
| Patient Population | Adults | Adults | Identical |
| System Procedures | - Irrigation / Aspiration
- Ultrasound
- Diathermy
- Anterior Vitrectomy | - Irrigation / Aspiration
- Ultrasound
- Diathermy
- Vitrectomy
- Illumination
- Air tamponade
- Silicone tamponade
- Laser | Identical,
narrower scope |
| Accessories Provided Sterile | Yes | Yes | Identical |
| Method of Sterilization | Ethylene Oxide | Ethylene Oxide | Identical |
| User Interface | Foot Control Panel,
Graphical User Interface,
Handpiece | Foot Control Panel, Graphical
User Interface, Handpiece | Identical |
| IRRIGATION and ASPIRATION | | | |
| Aspiration Pump Type | Flow Control & Vacuum
Control
Quattro Pump | Peristaltic | Different |
| Irrigation | Yes | Yes | Identical |
| Attribute | Subject Device | Primary Predicate Device
(K142877) | Equivalency
Analysis |
| Adjustable pump ramp | Yes | Yes | Identical |
| Continuous Irrigation | Yes | Yes | Identical |
| PHACOEMULSIFICATION | | | |
| Handpiece type | Piezoelectrical | Piezoelectrical | Identical |
| Range of frequency | 40 kHz | 40 kHz | Identical |
| Control | fixed or linear | fixed or linear | Identical |
| Tip stroke | up to 100µm | 100 µm | Equivalent |
| Incision type Co-Mix | 1.8mm | 1.8 mm | Identical |
| Pulse Mode/Duration | 0 - 250 pps | 0 - 250 pps | Identical |
| Phaco Tip Movement | longitudinal | longitudinal | Identical |
| DIATHERMY | | | |
| Operating Frequency | 2 MHz (± 20%) | 1MHz ± 10% | Equivalent |
| Foot Control Panel | Yes | Yes | Identical |
| ANTERIOR VITRECTOMY | | | |
| Pneumatic | Yes | Yes | Identical |
| Cutting mode | Back and Forth | Back and Forth | Identical |
| Cutter control | Linear or fixed | Linear or fixed | Identical |
| Single Cut | Yes | Yes | Identical |
7
Image /page/7/Picture/0 description: The image shows the word "ZEISS" in white text on a blue background. The text is bold and sans-serif. The background is a solid blue color, and there is a white curved line at the bottom of the image.
Page 5 of 7
8
Image /page/8/Picture/0 description: The image shows the logo for Zeiss. The logo consists of the word "ZEISS" in a bold, sans-serif font. The word is white and is set against a dark blue background. The logo is simple and modern.
7. SUMMARY OF STUDIES
Non-Clinical Performance Testing
Non-clinical system testing provided an evaluation of the performance of the system relevant to each of the system specifications. The functional and system level testing showed that the defined specifications. Substantial equivalence was additionally demonstrated by using equivalent testing methods applied during testing of the subject device (such as biocompatibility, sterility, software, electrical safety, and software) as compared to the reference device. The totality of the performance data provided supports substantial equivalence and is summarized below.
Biocompatibility & Animal Testing
Biocompatibility assessments and testing for patient contacting components/accessories was performed in accordance with the following standards: ISO 10993-1, ISO 10993-7, ISO 10993-10, ISO 10993-11, and ISO 10993-12, USP 34 . Specifications regarding intracutaneous reactivity or acute ocular irritation tested in rabbits (ISO 10993-10), material mediated pyrogenicity tested in rabbits (USP 34 ), acute systemic cytotoxicity tested in mice (ISO 10993-11), kligman maximization (i.e., skin sensitization) tested in guinea pigs (ISO 10993-10) have been met.
Additionally, patient contacting components referenced in this 510(k) are identical to the device components referenced in K193376. These components are I/A HANDPIECE 21G ANGLED 45°, VA HANDPIECE 21G CURVED. IRR & ASP HANDPIECE BIMANUAL FRONT. IRR & ASP HANDPIECE BIMANUAL. PHACO SET PLUS 21G BICONICAL 15°, PHACO SET PLUS 21G FLARED 30°, PHACO SET PLUS 21G BICONICAL 30°, tips for PHACO SET (flared tip 30°, biconical tip 15°, biconical tip 30°) Sleeve, test chamber set plus 21G, DIATHERMY FORCEPS.
Sterility Testing
The sterilization cycle B-01 /B-01x was validated to achieve a sterility assurance level (SAL) of 10 9 according to ISO 11135:2014. Additionally, visual inspection, micro-biological dusting, seal strength, dye penetration and bubble leak testing was performed to ensure micro-biological barrier function met specifications. Testing passed.
Additionally, after the sterilization process was completed, testing on the patient contacting components for the accessories to measure ethylene oxide (EO) and ethylene chlorohydrin residual (ECH) levels, as well as Limulus Amebocyte Lysate (LAL) levels was conducted. Testing passed.
The following standards were followed: ISO 11135, ISO 14644-2, ASTM F1608, ISO 11607-1, ASTM F1886F1886M, ASTM F88/F88M and EN 868-5, ASTM F2096, AAMI TIR 30, AAMI TIR 12, AAMI TIR 34, AAMI TIR ST81, ISO 15883-1.
Software Testing
Subject Device was tested according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005). In addition, the software testing also followed the Carl Zeiss Meditec internal software development procedure that follows the IEC 62304:2006+AC:2015 - Medical device software - Software life cycle processes. Validation has been conducted according to IEC 62366. Testing passed.
9
Image /page/9/Picture/0 description: The image features the word "ZEISS" in bold, white capital letters against a solid blue background. The word is positioned at the top of the frame, and the blue background extends below it in a curved shape, creating a sense of depth. The overall design is clean and modern, typical of corporate branding.
Electrical Safety/EMC Testing
Electrical, Thermal, EMC testing was conducted in accordance with IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-2 standards. Testing passed.
Additional Bench Testing
In addition to testing mentioned above, conformance to IEC 80601-2-58 and IEC 60529 was demonstrated by testing for occlusion break response, fluidics assessments, spring-eye model, and emersion testing was performed. Testing passed.
Clinical Performance Testing
Clinical studies were not required since the system does not introduce a new methodology or new/expanded clinical claims.
CONCLUSION 8.
The indications for use are equivalent to the indications for use of the predicate device; are deemed to be substantially equivalent.
The technological characteristics and risk profile of the subject device are equivalent to the predicate device; and therefore, are deemed to be substantially equivalent.
Testing methods are equivalent to those of the predicate device and reference device; and therefore, are deemed to be substantially equivalent.
Therefore, the subject device meets the requirements for substantial equivalence as compared to the proposed predicate device.