K Number
K212115
Device Name
Arm-type Fully Automatic Digital Blood Pressure Monitor, Wrist-type Fully Automatic Digital Blood Pressure Monitor
Date Cleared
2022-01-05

(182 days)

Product Code
Regulation Number
870.1130
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Fully Automatic Digital Blood Pressure Monitors are intended to measure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age.
Device Description
The Arm-type and Wrist-type Fully Automatic Digital Blood Pressure Monitor (BPM) series is automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment . The systolic and diastolic pressures are determined using the oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During measurements, an electric pump within the main unit slowly inflates the arm/wrist cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats/min. The pulse rate measurement is compare the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%. Meanwhile, these blood pressure monitor devices can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the "JoyHealth" APP through the embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD of the device displays results. And the device will start to transmit data to the pair-up terminal automatically. This app is only intended to display trend graphs of measured systolic and diastolic blood pressure and pulse rate, which does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user.
More Information

Not Found

No
The description mentions "software algorithms within the microprocessor" for determining blood pressure and pulse rate, and a simple comparison for irregular heartbeat detection. There is no mention of AI, ML, training data, or complex learning processes. The associated app is explicitly stated to only display data and not provide diagnostic or analytical functions.

No.
The device is intended to measure blood pressure and pulse rate for monitoring purposes, and the associated app does not provide diagnostic or treatment functions.

Yes

Explanation: The device measures blood pressure and pulse rate, and indicates irregular heartbeats, which are physiological parameters used to aid in the diagnosis of medical conditions.

No

The device description clearly outlines physical components like a cuff, piezoelectric pump, electric automatic rapid deflation valve, and a microprocessor, indicating it is a hardware device with embedded software. While it can connect to an app, the core measurement functionality resides within the physical unit.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information for diagnosis, monitoring, or screening.
  • Device Function: This device measures blood pressure and pulse rate directly from the patient's body (on the arm or wrist) using a non-invasive method (oscillometric). It does not analyze any biological specimens.
  • Intended Use: The intended use is to measure blood pressure and pulse rate, not to analyze biological samples for diagnostic purposes.

The device is a medical device, specifically a non-invasive blood pressure monitor, but it falls under a different regulatory category than IVDs.

N/A

Intended Use / Indications for Use

The Fully Automatic Digital Blood Pressure Monitors are intended to measure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age.

Product codes

DXN

Device Description

The Arm-type and Wrist-type Fully Automatic Digital Blood Pressure Monitor (BPM) series is automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment . The systolic and diastolic pressures are determined using the oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During measurements, an electric pump within the main unit slowly inflates the arm/wrist cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats/min.

The pulse rate measurement is compare the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%.

Meanwhile, these blood pressure monitor devices can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the "JoyHealth" APP through the embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD of the device displays results. And the device will start to transmit data to the pair-up terminal automatically. This app is only intended to display trend graphs of measured systolic and diastolic blood pressure and pulse rate, which does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making.

Unlimited readings can be stored in the app for archiving and review by the user.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Upper arm, Wrist

Indicated Patient Age Range

adults and adolescents over 12 years of age

Intended User / Care Setting

over-the-counter (OTC) use in home and clinical environment

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Clinical Validation:

  • ISO81060-2:2018+AMD2020 Non-invasive sphygmomanometers --Part 2: Clinical investigation of intermittent automated measurement type.

Arm-type:
In the Arm-type fully automatic digital blood pressure clinical investigation, 85 patients (46 females and 39 males) participated in the clinical study. Same arm sequential method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

Wrist-type:
Even through the wrist circumference size of cuff is the same (13.5-21.5cm)among these subject wrist-type blood pressure monitors, but the the air bag size inside the wrist cuff have tiny difference, therefore we have conduct two clinical accuracy investigation which choose two different representative models.

In the Wrist-type fully automatic digital blood pressure clinical investigation, Model DBP-8278B was selected as representative for testing. 85 patients (43 females and 42 males) participated in the clinical study. Same wrist sequential method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

In the Wrist-type fully automatic digital blood pressure clinical investigation, Model DBP-8276H was selected as representative for testing. 85 patients (41 females and 44 males) participated in the clinical study. Same wrist sequential method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K161886, K182127, K182166

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Jan 5, 2022

Joytech Healthcare Co.,ltd Ren Yunhua General Manager No.365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou city Hangzhou, Zhejiang 311100 China

Re: K212115

Trade/Device Name: Arm-type Fully Automatic Digital Blood Pressure Monitor, Wrist-type Fully Automatic Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 27, 2021 Received: December 7, 2021

Dear Ren Yunhua:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K212115

Device Name

Fully Automatic Digital Blood Pressure Monitor Including Arm-type: DBP-6179, DBP-6173, DBP-6173, DBP-6273B, DBP-6175, DBP-6277B, DBP-6177, DBP-6181, DBP-6281B, DBP-6182, DBP-6282B, DBP-6185, DBP-6285B, DBP-6186, DBP-6286B, DBP-6191 Wrist-type: DBP-8276H, DBP-8176H, DBP-8278B, DBP-8178, DBP-8288B, DBP-8188, DBP-8189

Indications for Use (Describe)

The Fully Automatic Digital Blood Pressure Monitors are intended to measure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age.

Type of Use (Select one or both, as applicable)

□ Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1/1

PSC Publishing Services (301) 443-6740 EF

3

Image /page/3/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. The word "JOYTECH" is in blue, except for the "T", which is in red. To the right of the word "JOYTECH" are three Chinese characters in blue.

510(k) Summary

The assigned 510(k) number is:

K212115

2.1 Subiectter's Identification:

Name: JOYTECH Healthcare Co., Ltd.

Add.: No. 365. Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100

Zhejiang, China.

Contact Person: Yunhua Ren

Phone: +86-571-81957767

Fax: +86-571-81957750

Email: renyh@sejoy.com

2.2 Name of the Device:

Trade Name: Fully Automatic Digital Blood Pressure Monitor

| Arm-type | DBP-6279B, DBP-6179, DBP-6173, DBP-6273B, DBP-6275B, DBP-6175,
DBP-6277B, DBP-6177, DBP-6181, DBP-6281B, DBP-6182, DBP-6282B,
DBP-6185, DBP-6285B, DBP-6186, DBP-6286B, DBP-6191 |
|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Wrist-type | DBP-8276H, DBP-8176H, DBP-8278B, DBP-8178, DBP-8288B, DBP-8188,
DBP-8189 |

Common Name: Blood Pressure Monitor

Classification name: Non-invasive blood pressure measurement System

21 CFR 870-1130, Class II, 74-DXN.

2.3 Classification Information:

Regulation Number: 870.1130

Product Code: DXN

Device Class: II

Panel: 74 Cardiovascular

2.4 Predicate Device Information:

4

Image /page/4/Picture/2 description: The image shows the logo for JOYTECH-I. The logo is in blue and red. The text "JOYTECH" is in blue, with a red arrow pointing to the right above the "O" and "Y". The "I" is red. To the right of "JOYTECH-I" is Chinese text in blue.

The Arm-Type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the following predicate device:

510(k) numberPredicate device modelProduct codeManufacturer
K161886H-BP100SBPDXNGuangdong Transtek Medical Electronics Co.,
Ltd.

The Wrist-Type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the following prdicate devices:

510(k) numberPredicate device modelProduct codeManufacturer
K182127BP6100DXNOMRON HEALTHCARE Co., Ltd.
K182166BP4350DXNOMRON HEALTHCARE Co., Ltd.

2.5 Device Description:

The Arm-type and Wrist-type Fully Automatic Digital Blood Pressure Monitor (BPM) series is automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment . The systolic and diastolic pressures are determined using the oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During measurements, an electric pump within the main unit slowly inflates the arm/wrist cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats/min.

The pulse rate measurement is compare the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%.

Meanwhile, these blood pressure monitor devices can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the "JoyHealth" APP through the embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD of the device displays results. And the device will start to transmit data to the pair-up terminal automatically. This app is only intended to display trend graphs of

5

measured systolic and diastolic blood pressure and pulse rate, which does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making.

Unlimited readings can be stored in the app for archiving and review by the user.

The detail comparisons among the Arm-Type series are listed in table below:

| Features
Models | A | B | C | D | E | F | G | H | I (mm) | J
(cm) | K(mm) | L | M | N | O | P | Q | R | S | T |
|--------------------|---|---|------------------|---|---|---|---|---|-----------------|-----------|-----------|---|---|---|---|---|---|---|---|---|
| DBP-6179 | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 142.5107.244 | 22-36 | 11262 | O | O | N | N | N | N | O | Y | Y |
| DBP-6279B | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 142.5
107.244 | 22-36 | 11262 | O | O | N | N | Y | N | O | Y | Y |
| DBP-6173 | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 142.5107.244 | 22-36 | 95.554.5 | O | O | N | N | N | N | O | Y | Y |
| DBP-6273B | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 142.5
107.244 | 22-36 | 95.554.5 | O | O | N | N | Y | N | O | Y | Y |
| DBP-6175 | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 128130144 | 22-36 | 11865 | O | O | N | N | N | N | O | Y | Y |
| DBP-6275B | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 128
130144 | 22-36 | 11865 | O | O | N | N | Y | N | O | Y | Y |
| DBP-6177 | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 149.880.445.7 | 22-36 | 83.153.1 | O | O | N | N | N | N | O | Y | Y |
| DBP-6277B | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 149.8
80.445.7 | 22-36 | 83.153.1 | O | O | N | N | Y | N | O | Y | Y |
| DBP-6181 | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 155.5108.575 | 22-36 | 87.585 | O | O | N | N | N | N | O | Y | Y |
| DBP-6281B | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 155.5
108.575 | 22-36 | 87.585 | O | O | N | N | Y | N | O | Y | Y |
| DBP-6182 | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 15010865 | 22-36 | 87.585 | O | O | N | N | N | N | O | Y | Y |
| DBP-6282B | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 150
10865 | 22-36 | 87.585 | O | O | N | N | Y | N | O | Y | Y |
| DBP-6185 | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 15010865 | 22-36 | 87.585 | O | O | N | N | N | N | O | Y | Y |
| DBP-6285B | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 150
10865 | 22-36 | 87.585 | O | O | N | N | Y | N | O | Y | Y |
| DBP-6186 | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 15010865 | 22-36 | 87.585 | O | O | N | N | N | N | O | Y | Y |
| DBP-6286B | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 150
10865 | 22-36 | 87.585 | O | O | N | N | Y | N | O | Y | Y |
| DBP-6191 | Y | O | 60
Memories×2 | Y | N | Y | Y | Y | 13694.557 | 22-36 | 50*62 | O | O | N | N | N | N | O | Y | Y |

A = Powered by 3 AAA size batteries

  • B= Powered by AC adaptor
  • C = Memory Size
  • D = Time & Date
  • E = Results Average in Three way
  • F = WHO (World Health Organization) Classification Indicator

G = Last 3 Results Average

6

Image /page/6/Picture/0 description: The image shows the logo for JOYTECH-I. The logo is in blue and red. The text is in a sans-serif font. The logo is for a company that specializes in technology.

H = Irregular Heartbeat Detection I = Outside Demission (L x W x H in unit mm) J = Cuff Size K = LCD Size (Viewing Area in unit mm) L = LCD Backlight M= Voice N= PC connecter O=Music P=Bluetooth Function Q=mmHg/kPa Switchable R=Beep buzzer S= Cuff Loose Detection T= Arm Shake Detection Note: Y= Yes N = No O= Optional function depending on clients' needs

The detail comparisons among the Wrist-Type series are listed in table below:

| Features

ModelsABCDEFGH (mm)I (cm)J(mm)KLMNOPQRST
DBP-8176HY60 Memories×2YYYYY6255.21913.5-21.534*34.6YNNNNYOYNO
DBP-8276HY60 Memories×2YYYYY6255.21913.5-21.534*34.6YNNYNYOYNO
DBP-8178Y60 Memories×2YYYYY736627.513.5-21.544*42OONNNOOYNN
DBP-8278BY60 Memories×2YYYNY736627.513.5-21.544*42OONYNOOYNN
DBP-8188Y60 Memories×2YYYYY84622513.5-21.543.7*40OONNNOOYNN
DBP-8288BY60 Memories×2YYYYY84622513.5-21.543.7*40OONYNOOYNN
DBP-8189Y60 Memories×2YYYYY846223.613.5-21.532*40.5ONNNNYOYNN

Note: DBP-8276H, DBP-8176H are equipped with lithium battery. Other models equipped with alkaline battery.

A = Powered by AAA Batteries / Lithium battery

(DBP-8276H, DBP-8176H equipped with rechargeable lithium battery)

  • B= Memory Size
  • C= Time & Date
  • D = WHO (World Health Organization) Classification Indicator
  • E = Last 3 Results Average
  • F = Irregular Heartbeat Detection
  • G = Low Voltage detection
  • H= Outside Dimension (L x W x H in unit mm)
  • I = Cuff Size
  • J = LCD Size (Viewing Area in unit mm)
  • K= LCD Backlight
  • L= Voice
  • M= PC connecter
  • N=Bluetooth function
  • O=mmHg/kPa Switchable
  • P=Beep buzzer
  • Q=Heart Zone indicator

7

Image /page/7/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. The word "JOYTECH" is in blue, except for the "T", which is in red. To the right of the word "JOYTECH" are three Chinese characters in blue.

R= Arm Shake indicator S= Cuff Loose indicator T=Powered by AC adaptor Note: Y= Yes N = No O= Optional function depending on clients' needs

The devices are all designed and manufactured according to AAMI/ANSV/IEC80601-2-30:2018, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.

2.6 Intended Use:

The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age.

The Wrist-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age.

2.7 Comparison of Technological Characteristics with predicate device:

2.7.1 Arm-type

The arm-type blood pressure monitor manufactured by JOYTECH have the same arm cuff type, features and specifications with the Welch Allyn blood pressure monitor H-BP100SBP which 510k number is K161886, therefore we choose the device act as the predicate device. The detail comparison of technical characteristic as below:

Comparison itemSubject device in present applicationPredicate device K161886 (model:H-BP100SBP)Comparison result / Explanation
The trade nameArm-type Fully Automatic
Digital Blood Pressure MonitorWelch Allyn Blood Pressure
Device/
ManufacturerJOYTECH Healthcare Co.,
Ltd.Guangdong Transtek Medical
Electronics Co., Ltd/
Recommended
classification
regulation21CFR 870.1130,
Noninvasive Blood
Pressure Measurement
System21CFR 870.1130,
Noninvasive Blood
Pressure Measurement
SystemIdentical
Regulatory classIIIIIdentical
Panel74 Cardiovascular74 CardiovascularIdentical
Product codeDXNDXNIdentical
Indication for useThe Arm-type Fully AutomaticWelch Allyn Blood PressureSimilar
Blood Pressure Monitors are
intended to measure blood
pressure (systolic and diastolic)
and pulse rate of adults and
adolescents over 12 years of
age.Device H-BP100SBP is digital
monitor intended for use in
measuring blood pressure and
heartbeat rate in adult patient
population with arm
circumference ranging from 15
cm to 54 cm (about 6-21 inches).
These devices detect the
appearance of irregular
heartbeats during measurement
and give a warning signal with
readings.Note1
Measuring
principleOscillometric methodOscillometric methodIdentical
Measurement
typeDetermined during inflationDetermined during inflationIdentical
Cuff locationUpper armUpper armIdentical
Specification
Measuring rangeSystolic Pressure: 60mmHg ~
260 mmHg
Diastolic Pressure:
30mmHg~200 mmHg
Pulse:30~180 Beats/MinuteSystolic Pressure: 50mmHg ~
260 mmHg
Diastolic Pressure:
25mmHg~220 mmHg
Pulse:40~199 Beats/MinuteSimilar.
Note2
Max cuff pressure299 mmHg300mmHgSimilar
AccuracyStatic Pressure: $\pm$ 3mmHg
Pulse: $\pm$ 5%Pressure: $\pm$ 0.4kPa (3mmHg)
Pulse value: $\pm$ 5%Identical
InflationBy air pumpBy air pumpIdentical
Pressure releaseBy solenoid valveBy solenoid valveIdentical
Operating
Temp. &
humidityTemp.: 10℃~40℃
Humidity: 15~93%RH
Atmospheric:
800hPa~1060hPaTemp.:5℃~40℃
Relative humidity 85%RH
Atmospheric
pressure:86kPa to 106kPaSimilar
Note 3
Storage Temp.
& humidityTemp.: -25℃~55℃
Humidity:
$\le$ 93% RHTemp.:-20℃~60℃
Relative Humidity:
10% RH - 93% RH
Atmospheric
Pressure: 50kPa to
106 kPaSimilar
Note 4
Cuff
circumferenceFits arm circumference 22-36
cmThe three preformed cuff units,
which are
applicable to arm circumference
between 15-24cm, 22-42cm and
40-54cm to meet different
Population's need.Similar.
Note 5
Supply
power
source3x1.5V AAA battery or 2MOPP
Medical AC adaptor (DC 5V,
1A, recommend but not
provide)powered by four AA alkaline
batteries or by a AC adapter (DC
6V,1A)(not include)Similar
The different in the
specification is
documented
and tested
PCB or Electrical
schemeBP73PCB
BP75PCBUnknownSimilar
The different in the
BP77PCB
BP81PCB
BP91PCBspecification is
documented and tested
Display
component1.Device LCD;
2.iOS device or Android device1.Device LCD;
2.iOS device or Android deviceIdentical
Bluetooth
Module versionV5.0V4.0Similar.
Note6
System
requirementBluetooth 4.0 or later/
Android 5.0 or later/
IOS 9.0 or laterPlatform which support
Bluetooth 4.0 technologySimilar
Note 7
Bluetooth
featuresTransmit data to
smart phone via
BluetoothTransmit data to
smart phone via
BluetoothIdentical
Function on APPStorage test result/
Announce test result/ Read
memories in App/ Empty
memories in App/ Draw curve/
Set the personal
information in appUnknownDifferent.
Note8
Wireless
communicationFrequency range: 2.4 GHz
(2400 - 2483.5 MHz)
Modulation type: GFSK
Antenna gain: 0.5dBiFrequency range: 2.4 GHz (2402
  • 2480 MHz)
    Modulation type: GFSK
    Antenna gain: 0dBi | Equivalent
    The difference have
    verified by FCC test. |
    | Sterilization | Not applicable | Not applicable | Identical |
    | Features
    Irregular
    heart
    beat | Irregular heart beat is displayed
    on the LCD. | Irregular heart beat is displayed
    on the LCD. | Identical |
    | Blood
    Pressure
    categories | WHO classification | Blood pressure high and low
    distinguish | Similar
    Note9 |
    | Arm
    shake
    indicator | Yes | Yes | Identical |
    | Cuff
    loose
    indicator | Yes | None | Different.
    Note10 |
    | Memory | 2*60 Memories in Two Groups
    with Date and Time | Have memory function | Similar.
    Note11 |
    | Material | ABS for the case and plastic
    foil for the labels of the device.
    Bio-compatibility materials are
    used for the applied parts (Cuff
    containing a interal bladder) | ABS is used to outer housing of
    the main unit. | Similar.
    Note12 |

8

Image /page/8/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. To the right of the word JOYTECH is Chinese characters.

9

Image /page/9/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. The word "JOYTECH" is in blue, with a red triangle above the "O". To the right of "JOYTECH" are some Chinese characters in blue.

Notel: The intended use of subject device and predicate device are the same. The clinical accuracy of subject device applied to adolescents over 12 years of age population have been verified and validation , which comply with the requirements of ISO 81060-2. And the clinical safety also have been validated as for without any AE or side-effect in clinical investigation.

10

Image /page/10/Picture/2 description: The image shows the logo for JOYTECH-I. The logo is in blue and red. The text "JOYTECH-I" is in English, and there are three Chinese characters to the right of the English text.

Note2: The measuring range of the subject device have been verified by IEC60601-1 and IEC 80601-2-30 standard test.

Note3: The operation environment of the subject device have been verified by IEC 60601-2-30 standard test.

Note4: The storage environment of the subject device have been verified by IEC 60601-1 and IEC 80601-2-30 standard test.

Note5: The safety and effective for fits arm circumference 22-36cm of subject device have been verified and validation by bench test according to IEC 60601-1-11 and IEC 80601-2-30. Meanwhile, the clinical accuracy of cuff size 22-36cm have been validation according to the ISO 81060-2:2018, which cannot raise new safety and effective issue.

Note 6: The subject device adopt new BT version and it could compatible for BT 4.0. The cybersecurity risk has been considered and the overall residual risk is acceptable.

Note7: All over of subject models have passed FCC, EN301489-17, EN 300328 and wireless coexistence test.

Note8: The change in the specification is documented, and the cybersecurity risk has been considered and the overall residual risk is acceptable. The change does not affect the intended use or the fundamental scientific technology.

Note9: The blood pressure WHO classification function have been verified by software verified and validation report.

Note10: The cuff loose indicator function of subject device have been verified and validation through Software verified and validation report. And this function adding cannot bring new risks.

Note 11: The quantity of storage readings is different between the subject device and predicate device. The change in the specification is documented, and The change does not affect the intended use or the fundamental scientific technology.

Note12:The materials of cuff from subject device have been tested and found to comply with requirements of ISO 10993-5: 2009 and ISO 10993-10:2010, which cannot result in cytotoxicity, sensitization or intracutaneous reacticity.

2.7.2 Wrist-type

The wrist-type blood pressure monitor manufactured by JOYTECH have the same wrist cuff type, features and specifications with the Omron wrist blood pressure monitor BP6100 which 510k number is K182127, therefore we choose the device act as the predicate device. While BP6100 model has no bluetooth function, Omron wrist blood pressure monitor BP4350 (K182166) was chosen to be predicate device 2 with bluetooth function. The detail comparison of technical characteristic as below:

| Comparison
item | Subject device in
present application | Predicate device 1
K182127 | Predicate device 2
K182166 | Comparison
result / |

----------------------------------------------------------------------------------------------------------------------------------------------------

11

Image /page/11/Picture/0 description: The image shows the logo for JOYTECH-I. The logo is in blue and red. The text is in a sans-serif font.

(Model:BP6100)(Model: BP4350)Explanation
The trade nameWrist-type
Automatic
Blood
Pressure
MonitorWrist Blood Pressure
Monitor Model BP6100Wrist Blood Pressure
Monitor Model BP4350/
ManufacturerJOYTECH
Healthcare Co., Ltd.OMRON HEALTHCARE
Co., LtdOMRON
HEALTHCARE Co., Ltd/
Recommended
classification
regulation21CFR 870.1130,
Noninvasive Blood
Pressure
Measurement
System21CFR 870.1130,
Noninvasive Blood
Pressure Measurement
System21CFR 870.1130,
Noninvasive Blood
Pressure Measurement
SystemIdentical
Regulatory
classIIIIIIIdentical
Panel74 Cardiovascular74 Cardiovascular74 CardiovascularIdentical
Product codeDXNDXNDXNIdentical
Indication for
useThe Fully Automatic
Blood Pressure
Monitors are
intended to measure
blood pressure
(systolic and
diastolic) and pulse
rate of adults and
adolescents over 12
year of age.The device is a digital
monitor intended for use
in measuring blood
pressure and pulse
rate in adult patient
population with wrist
circumference ranging
from 5.3 inches to
8.5 inches (13.5 cm to
21.5 cm).
The device detects the
appearance of irregular
heartbeats during
measurement and gives a
warning signal with
readings.The device is a digital
monitor intended for use
in measuring blood
pressure and pulse
rate in adult patient
population with wrist
circumference ranging
from 5.3 inches to 8.5
inches (13.5cm to
21.5cm). The device
detects the appearance of
irregular heartbeats
during measurement and
gives a warning signal
with readings.Similar
Note1
Measuring
principleOscillometric
methodOscillometric methodOscillometric methodIdentical
Measurement
typeDetermined during
inflationDetermined during
inflationDetermined during
inflationIdentical
Cuff locationWristWristWristIdentical
Specification
Measuring
rangeSystolic Pressure:
60mmHg~260
mmHg
Diastolic Pressure:
30mmHg~200
mmHg
Pulse:30~180
Beats/minSystolic Pressure:
60mmHg~260 mmHg
Diastolic Pressure:
40 mmHg~215 mmHg

Pulse:40180 Beats/min | Systolic Pressure:
60mmHg
260 mmHg
Diastolic Pressure:
40 mmHg~215 mmHg

Pulse:40180 Beats/min | Similar.
Note 2 |
| Max
pressure | cuff 299 mmHg | 299 mmHg | 299 mmHg | Identical |
| | | | | |
| circumference | circumference 13.5-
21.5cm | 13.5-21.5cm | 13.5-21.5cm | |
| Accuracy | Static Pressure: ±
3mmHg
Pulse: ± 5% | Static Pressure: ± 3mmHg
Pulse: ± 5% of display
reading | Static Pressure: ±
3mmHg
Pulse: ± 5% of display
reading | Identical |
| Inflation | By air pump | By air pump | By air pump | Identical |
| Pressure release | By solenoid valve | By solenoid valve | By solenoid valve | Identical |
| Operating
Temp. &
humidity | Temp.: 10°C
40°C
Humidity:
1593%RH
Atmospheric:
800hPa
1060hPa | Temp.: 10°C40°C
Humidity: 15
90%RH
Atmospheric:
800hPa1060hPa | Temp.: 10°C40°C
Humidity: 1590%RH
Atmospheric:
800hPa
1060hPa | Similar
Note 3 |
| Storage Temp.
& humidity | Temp.: -25°C55°C
Humidity:
≤93% RH | Temp.:-20°C
60°C
Relative Humidity:
10% RH - 90% RH
Atmospheric | Temp.:-20°C~60°C
Relative Humidity:
10% RH - 90% RH
Atmospheric | Similar
Note 4 |
| Supply
power
source | For models DBP-
8178, DBP-8278B,
DBP-8188, DBP-
8288B, DBP-8189:
2x1.5V AAA battery
DBP-8276H and
DBP-8176H are
powered by 3.7 V
rechargeable lithium
battery or 2MOPP
Medical AC adaptor
(DC 5V,1A,
recommend but not
provide ) | Two "AAA" batteries 1.5v | Two "AAA" batteries
1.5v | Similar.
Note 5 |
| Display
component | 1.Device LCD;
2.iOS device or
Android device | Device LCD | 1.Device LCD;
2.iOS device or Android
device | Identical
to
predicate
device 2. |
| Bluetooth
Module version | V5.0 | None | Unknown | Note 6 |
| Wireless
communication | Frequency range: 2.4
GHz (2400 - 2483.5
MHz)
Modulation type:
GFSK
Antenna gain:
0.5dBi | None | Frequency range: 2.4
GHz (2400 - 2483.5
MHz)
Modulation: GFSK
Effective radiated power:
Features | | | | |
| Irregular heart beat | Irregular heart beat is displayed on the LCD. | Irregular heart beat is displayed on the LCD. | Irregular heart beat is displayed on the LCD. | Identical |
| Body movement detection | Yes | Yes | Yes | Identical |
| Cuff loose detection | No | Yes | Yes | Different
Note 8 |
| Heart zone indicator | Prompt the user whether the cuff is located in correct height when measurement. | Yes | Yes | Identical |
| WHO /ACC AHA classification indicator | Yes | Yes | Yes | Identical |
| Memory | 260 Memories in Two Groups with Date and Time | 60 memories in one groups | 2100 Memories in Two Groups with Date and Time | Similar
Note 9 |

12

Image /page/12/Picture/0 description: The image shows the logo for JOYTEC-I. The logo is in blue and red. The text "JOYTEC" is in blue, and the "I" is in red. There are some Chinese characters to the right of the "I".

13

Notel: The intended use of subject device and predicate device are the same. The clinical accuracy of subject device applied to adolescents over 12 years of age population have been verified and validation , which comply with the requirements of ISO 81060-2. And the clinical safety also have been validated as for without any AE or side-effect in clinical investigation.

Note 2: The measuring range of the subject device have been verified by IEC 60601-1 and IEC 80601-2-30 standard test.

Note 3: The operation environment of the subject device have been verified by IEC 60601-1 and IEC80601-2-30 standard test.

Note 4: The storage environment of the subject device have been verified by IEC 60601-2-30 standard test.

Note 5: The supplied power of subject device model DBP-8278B, DBP-8288B, DBP-8188, DBP-8188, DBP-8189 are the same as the predicate device 1 and 2. The model DBP-8276H and DBP-8176H equipped with rechargeable lithium battery or powered by 2MOPP medical AC adaptor (Recommend but not provide ). The lithium battery have pass the IEC 62133-2:2017 test. The Adaptor interface have been verified through safety and EMC test. The risk from lithium battery have been analysis and evaluation in risk management report , and the overall residual risk is acceptable.

Note 6: Although the Bluetooth module version of predicate device BP4350 was unknown, The subject device adopt new BT version 5.0 and it could compatible for BT 4.0. The cybersecurity risk has been considered and the overall residual risk is acceptable.

Note 7: The change in the specification is documented, and the cybersecurity risk has been considered and the overall residual risk is acceptable. The change does not affect the intended use or the fundamental scientific technology.

14

JOYTECH-I健拓石

Note 8: Comparing with the predicate device, cuff loose indicate function does not affect the the intended use or the fundamental scientific technology. And this function have been verification and validation report. This function adding cannot bring new risks.

Note 9: The quantity of storage readings is different between the subject device and predicate device. The change in the specification is documented, and The change does not affect the intended use or the fundamental scientific technology.

2.8. Performance Data:

Testing information demonstrating safety and effectiveness of the device in the intended environment of use is supported by testing that was conducted.

The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices.

The following National and International Standards were utilized for testing the subject device.

Electrical Safety and performance requirements:

  • ES60601-1:2005/(R)2012 AAMI And A2:2010/(R)2012 Medical Electrical Equipment.
  • IEC 80601-2-30:2018, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.

Home-used medical equipment requirements and environmental test:

  • IEC 60601-1-11:2015 General requirements for basic safety and essential performance --Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
    Electromagnetic compatibility requirements:

  • IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests

  • EN300328:Electromagnetic compatibility and Radio spectrum Matters (ERM); Wideband transmission systems; Data transmission equipment operating in the 2,4 GHz ISM band and using wide band modulation techniques;

  • EN ETSI 301489-1: Electromagnetic compatibilityand Radio spectrum Matters(ERM);ElectroMagnetic Compatibility (EMC)standard for radio equipment and services;Part 1:Common technical requirements;

  • ETSI EN 301489-17: Electromagnetic compatibility and Radio spectrum Matters (ERM); ●

15

ElectroMagnetic Compatibility (EMC) standard for radio equipment; Part 17: Specific conditions or Broadband Data Transmission Systems;

Bio-compatibility Evaluation for patient contacting components:

  • ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
  • ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and ● skin sensitization

FCC Test

  • FCC Part15 Subpart C ●
  • RF Exposure Evaluation

Guidance Document:

The software/firmware verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005.

The test result all meet or exceed the requirement of these standards.

2.10. Discussion of Clinical Tests Performed:

Clinical Validation:

  • ISO81060-2:2018+AMD2020 Non-invasive sphygmomanometers --Part 2: Clinical investigation of intermittent automated measurement type.

Arm-type:

In the Arm-type fully automatic digital blood pressure clinical investigation, 85 patients (46 females and 39 males) participated in the clinical study. Same arm sequential method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

Wrist-type:

Even through the wrist circumference size of cuff is the same (13.5-21.5cm)among these subject wrist-type blood pressure monitors, but the the air bag size inside the wrist cuff have tiny difference, therefore we have conduct two clinical accuracy investigation which choose two

16

different representative models.

In the Wrist-type fully automatic digital blood pressure clinical investigation, Model DBP-8278B was selected as representative for testing. 85 patients (43 females and 42 males) participated in the clinical study. Same wrist sequential method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

In the Wrist-type fully automatic digital blood pressure clinical investigation, Model DBP-8276H was selected as representative for testing. 85 patients (41 females and 44 males) participated in the clinical study. Same wrist sequential method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

2.11. Conclusions:

This submitted arm-type series manufactured by JOYTECH Healthcare Co., Ltd. have been found to be respectively substantially equivalent to the predicate device (Welch Allyn Blood Pressure device H-BP100SBP) manufactured by Guangdong Transtek Medical Electronics Co., Ltd .(K161886). Based on the information provided in this submission, differences between the subject device and predicate device have been tested and documented, it has been demonstrated that the subject devices are as safe and effective as the predicate device.

This submitted wrist-type series manufactured by JOYTECH Healthcare Co., Ltd. have been found to be respectively substantially equivalent to the predicate device 1 (Omron wrist blood pressure monitor BP6100) manufactured by OMRON HEALTHCARE Co., Ltd (K182127) and the predicate device 2 (Omron wrist blood pressure monitor BP4350) manufactured by OMRON HEALTHCARE Co., Ltd (K182166) . Based on the information provided in this submission, differences between the subject device and predicate devices have been tested and documented, it has been demonstrated that the subject devices are as safe and effective as the predicate device.

In the other words, the difference between the subject and the predicate devices does not: (1)

17

Image /page/17/Picture/6 description: The image shows the logo for JOYTECH. The logo is in blue and red. To the right of the word JOYTECH is Chinese text.

affect the intended use or (2) alter the fundamental scientific technology of the device, and the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.

--- End of this section---