K Number
K212068
Date Cleared
2021-11-08

(129 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Drill Stop Kit is intended to be used to enclose drill stops that are to be sterlized by a health care provider. Drill Stop Kit is intended to allow sterilization of the enclosed medical devices. Drill Stop Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.)

Drill Stop Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum)- Exposure at 132 °C for 4 minutes and 20 minutes dry time. Drill Stop Kit is intended for sterilization of non-porous loads.

Drill Stop Kit is recommended not to be stacked during sterilization.

The combined weight of the Drill Stop Kit and the associated devices is 199.8 grams.

The weight of the empty Drill Stop Kit is 192.6 grams.

Prosthetic Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Prosthetic Kit is intended to allow sterilization of the enclosed medical devices.

Prosthetic Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.)

Prosthetic Kit is to be enclosed in a sterlization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Prosthetic Kit is intended for sterilization of non-porous loads.

Prosthetic Kit is recommended not to be stacked during sterilization.

The combined weight of the Prosthetic Kit and the associated instruments is 178.8 grams.

The weight of the empty Prosthetic Kit is 141.8 grams

Healing Screws Kit is intended to be used to enclose healing screws that are to be sterilized by a health care provider. Healing Screws Kit is intended to allow sterilization of the enclosed medical devices. Healing Screws Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Healing Screws Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) - Exposure at 132 °C for 4 minutes and 20 minutes dry time.

Healing Screws Kit is intended for sterilization of non-porous loads.

Healing Screws Kit is recommended not to be stacked during sterilization.

The combined weight of the Healing Screws Kit and the associated instruments is 203.1 grams.

The weight of the empty kit Healing Screws Kit is 168.9 grams.

Kit for Healing Screws Kontat Perio Level is intended to be used to enclose healing screws that are to be sterilized by a health care provider. Kit for Healing Screws Kontat Perio Level is intended to allow sterilization of the enclosed medical devices. Healing Screws Kit, Perio Level requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Kit for Healing Screws Kontat Perio Level in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes dry time. Kit for Healing Screws Kontat Perio Level is intended for sterilization of non-porous loads.

Kit for Healing Screws Kontat Perio Level is recommended not to be stacked during sterilization.

The combined weight of the Kit for Healing Screws Kontat Perio Level and the associated instruments is 200.0 grams. The weight of the empty Kit for Healing Screws Kontat Perio Level is 169.0 grams.

Scanbodies Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Scanbodies Kit is intended to allow sterilization of the enclosed medical devices.

Scanbodies Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.)

Scanbodies Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Scanbodies Kit is intended for sterilization of non-porous loads.

Scanbodies Kit is recommended not to be stacked during sterilization.

The combined weight of the Scanbodies Kit and the associated instruments is 259.7 grams.

The weight of the empty Scanbodies Kit is 233.9 grams.

Pick-up Impression Coping Cica Kit is intended to be used to enclose pick-up impression copings that are to be sterilized by a health care provider. Pick-up Impression Coping Cica Kit is intended to allow sterilization of the enclosed medical devices. Pick-up Impression Coping Cica Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.)

Pick-up Impression Coping Cica Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Pick-up Impression Coping Cica Kit is intended for sterilization of non-porous loads.

Pick-up Impression Coping Cica Kit is recommended not to be stacked during sterilization.

The combined weight of the Pick-up Impression Coping Cica Kit and the associated instruments is 338.0 grams.

The weight of the empty tray Pick-up Impression Coping Cica Kit is 291.4 grams.

Pop-up Impression Coping Cica Kit is intended to be used to enclose pop-up impression copings that are to be sterilized by a health care provider. Pop-up Impression Coping Cica Kit is intended to allow sterilization of the enclosed medical devices. Pop-up Impression Coping Cica Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.)

Pop-up Impression Coping Cica Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) = Exposure at 132 °C for 4 minutes and 20 minutes dry time. Pop-up Impression Coping Cica Kit is intended for sterilization of non-porous loads.

Pop-up Impression Coping Cica Kit is recommended not to be stacked during sterilization.

The combined weight of the Pop-up Impression Coping Cica Kit and the associated instruments is 317.1 grams.

The weight of the empty kit Pop-up Impression Coping Cica Kit is 270.9 grams.

Multi Kontact Surgical Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Multi Kontact Surgical is intended to allow sterilization of the enclosed medical devices. Multi Kontact Surgical requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.) Multi Kontact Surgical is to be enclosed in a sterlization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes dry time.

Multi Kontact Surgical is intended for sterilization of non-porous loads.

Multi Kontact Surgical is recommended not to be stacked during sterilization.

The combined weight of the Multi Kontact Surgical and the associated instruments is 580.7 grams.

The weight of the empty Multi Kontact Surgical is 449.6 grams.

Try-in Abutment Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Try-in Abutment Kit is intended to allow sterilization of the enclosed medical devices. Try-in Abutment Kit requires the use of a double pouch to maintain the sterlity of the enclosed devices (pouch must be FDA cleared in the U.S.). Try-in Abutment Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes dry time.

Trv-in Abutment Kit is intended for sterilization of non-porous loads.

Try-in Abutment Kit is recommended not to be stacked during sterilization.

The combined weight of the Try-in Abutment Kit and the associated devices is 469.5 grams.

The weight of the empty Try-in Abutment Kit is 463.8 grams

AtlaSurgery Guided Surgery Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. AtlaSurgery Guided Surgery Kit is intended to allow sterilization of the enclosed medical devices. AtlaSurgery Guided Surgery Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

AtlaSurgery Guided Surgery Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) = Exposure at 132 °C for 4 minutes and 20 minutes dry time.

AtlaSurqery Guided Surgery Kit is intended for sterilization of non-porous loads.

AtlaSurgery Guided Surgery Kit is recommended not to be stacked during sterilization.

The combined weight of the AtlaSurgery Guided Surgery Kit and the associated instruments is 689.2 grams.

The weight of the empty AtlaSurgery Guided Surgery Kit is 476.0 grams.

Pilot Drills Kit for Guided Surgery is intended to be used to group other medical devices that are to be sterilized by a health care provider. Pilot Drills Kit for Guided Surgery is intended to allow sterlization of the group medical devices. Pilot Drills Kit for Guided Surgery requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Pilot Drills Kit for Guided Surgery is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes dry time.

Pilot Drills Kit for Guided Surgery is intended for sterilization of non-porous loads.

Pilot Drills Kit for Guided Surgerv is recommended not to be stacked during sterilization.

The combined weight of the Pilot Kit for Guided Surgery and the associated instruments is 39.4 grams.

The weight of the empty Pilot Drills Kit for Guided Surgery is 36.8 grams.

Device Description

The subject device trays are reusable rigid containers, comprising a base (bottom), a removable inner trav (if applicable), and a lid (cover). The subject device trays are to be used to organize and protect the dental healing screws, try-in abutments, instruments, and accessories that are sterilized in the trays by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior contents during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared pouch to maintain sterility.

The subject device includes a total of 10 sizes (same lid and base and enclosed volumes), and a total of 11 tray configurations. The subject device has seven (7) trays that are 2-piece plastic trays manufactured from Polyphenylsulfone (Radel® R-5000) for the lid and Polypropylene (PPHS) or Polysulfone for the base. The subject device has three (3) trays that are 3-piece plastic trays manufactured from Polyphenylsulfone (Radel R-5000) with silicone grommets. The subject device has one (1) tray that is a 2-piece metal tray manufactured from stainless steel.

AI/ML Overview

This document is a 510(k) Premarket Notification for Biotech Dental Kits, which are instrument sterilization trays. It details the device's indications for use, technological characteristics, and non-clinical testing performed to demonstrate substantial equivalence to predicate devices.

Acceptance Criteria and Device Performance Study for Biotech Dental Kits

The review focuses on the "Summary of Non-Clinical Testing" section, specifically the table and the text indicating how the device meets acceptance criteria.

1. Table of Acceptance Criteria and Reported Device Performance:

Test MethodologyPurposeAcceptance CriteriaReported Device Performance
Automated Cleaning Validation (FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued March 2015)To validate that the cleaning instructions provided in the Instructions for Use appropriately clean the tray, and to ensure the sterilization cycle will be effective.Protein assay: less than 6.4 µg/cm² (Assay quantitation limits: 5.0 µg/mL) TOC assay: less than 2.010 mg/L (Assay quantitation limits: 2.010 mg/L)Pass
Sterilization Validation (including sterilant penetration and dry time validation) (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2)To validate that the sterilization instructions listed in the Instructions for Use appropriately sterilize the tray and contents.3 consecutive half-cycles performed for each of the two sizes of trays will demonstrate complete inactivation of all biologic indicators; A minimum SAL of 10⁻⁶ is achieved if the Instructions for Use are followedPass
Dry timeTo validate that the sterilization instructions listed in the Instructions for Use appropriately dry the pouched tray for storage.Using pre-cycle and post-cycle weights, the weight gain after drying will not exceed ± 3%Pass
Life Cycle / Simulated Use Life Validation (FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued March 2015)To validate the service life of the trays as stated in the Instructions for Use.Visual inspection, component dimensional fit verification, functional closure/latch verification for 100 use cyclesPass
Biocompatibility of Subject Device (by cytotoxicity testing) (ANSI/AAMI/ISO 10993-5, ANSI/AAMI/ISO 10993-12)To evaluate the cytotoxicity potential of the test article using an in vitro cell culture assay.Non-cytotoxic if ≤ 30% reduction of NRU cell viability after exposure to extract of the devicePass

2. Sample size used for the test set and data provenance:

The document describes non-clinical performance testing.

  • Cleaning Validation: The sample size isn't explicitly stated as a numerical count of individual units tested, but it refers to the cleaning validation of "the tray." The number of cycles performed for the cleaning validation is not specified in the summary.
  • Sterilization Validation: "3 consecutive half-cycles performed for each of the two sizes of trays." This implies a total of 6 cycles across two representative sizes (one representing the most challenging to sterilize).
  • Dry Time: Not explicitly stated, but performed in conjunction with sterilization validation.
  • Life Cycle / Simulated Use Life Validation: "100 use cycles" are performed for this validation.
  • Biocompatibility: The acceptance criteria reference "extract of the device," implying testing on material samples representative of the device.

The data provenance is not explicitly stated as country of origin, nor is it categorized as retrospective or prospective. However, given that these are non-clinical performance tests (e.g., cleaning, sterilization efficacy, material compatibility, and durability) conducted as part of a premarket submission, they are inherently prospective tests performed in a controlled laboratory environment to evaluate the device's design and manufacturing parameters.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document states that no clinical data were included in this submission. Therefore, no human experts (e.g., radiologists) were used to establish ground truth in the context of clinical interpretation or diagnosis. The ground truth for these non-clinical tests is based on established scientific and engineering principles, and validated laboratory methodologies and standards (e.g., ISO, AAMI, FDA guidance). The "experts" involved would be the engineers, microbiologists, and other technical personnel who perform these tests and interpret the results against the defined acceptance criteria. Their qualifications are implicit in the adherence to the referenced standards.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

No adjudication method is relevant or described as these are non-clinical, objective performance tests against established quantitative and qualitative criteria (e.g., protein residue levels, sterility assurance level, weight gain, visual inspection for integrity). There is no subjective human reading or interpretation of "cases" that would require adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No MRMC comparative effectiveness study was done. This device is a medical instrument sterilization tray, not an AI-powered diagnostic or assistive technology for human readers. The submission explicitly states, "No clinical data were included in this submission."

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This is not applicable. The device is a physical sterilization tray, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

As these are non-clinical tests, the "ground truth" is defined by:

  • Established industry standards: ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2, ANSI/AAMI/ISO 10993-5, ANSI/AAMI/ISO 10993-12.
  • FDA guidance documents: "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (issued March 2015)."
  • Validated laboratory methodologies: Protein assays (Micro BCA Protein Assay), TOC assays, biologic indicator inactivation, weight measurements, visual inspections.

The ground truth is therefore based on meeting specific, measurable, and objective physical, chemical, and microbiological endpoints defined by regulatory and consensus standards.

8. The sample size for the training set:

Not applicable. This device is a physical product and does not involve machine learning or a "training set."

9. How the ground truth for the training set was established:

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 8, 2021

Biotech Dental, SAS % Melissa Burbage RA Sr Specialist PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K212068

Trade/Device Name: Biotech Dental Kits Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: October 7, 2021 Received: October 8, 2021

Dear Melissa Burbage:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801: medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental Drill Stop Kit

Indications for Use (Describe)

Drill Stop Kit is intended to be used to enclose drill stops that are to be sterlized by a health care provider. Drill Stop Kit is intended to allow sterilization of the enclosed medical devices. Drill Stop Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.)

Drill Stop Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum)- Exposure at 132 °C for 4 minutes and 20 minutes dry time. Drill Stop Kit is intended for sterilization of non-porous loads.

Drill Stop Kit is intended for sterilization of non-porous loads.
Drill Stop Kit is recommended not to be stacked during sterilization.

Drill Stop Kit is recommended not to be stacked during sterilization. The combined weight of the Drill Stop Kit and the associated devices is 199.8 grams.

The weight of the empty Drill Stop Kit is 192.6 grams.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental Prosthetic Kit

Indications for Use (Describe)

Prosthetic Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Prosthetic Kit is intended to allow sterilization of the enclosed medical devices.

Prosthetic Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.)

Prosthetic Kit is to be enclosed in a sterlization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Prosthetic Kit is intended for sterilization of non-porous loads.

Prosthetic Kit is recommended not to be stacked during sterilization.

The combined weight of the Prosthetic Kit and the associated instruments is 178.8 grams.

The weight of the empty Prosthetic Kit is 141.8 grams

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental Healing Screws Kit

Indications for Use (Describe)

Healing Screws Kit is intended to be used to enclose healing screws that are to be sterilized by a health care provider. Healing Screws Kit is intended to allow sterilization of the enclosed medical devices. Healing Screws Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Healing Screws Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) - Exposure at 132 °C for 4 minutes and 20 minutes dry time.

Healing Screws Kit is intended for sterilization of non-porous loads.

Healing Screws Kit is recommended not to be stacked during sterilization.

The combined weight of the Healing Screws Kit and the associated instruments is 203.1 grams.

The weight of the empty kit Healing Screws Kit is 168.9 grams.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental Kit for Healing Screws Kontact Perio Level

Indications for Use (Describe)

Kit for Healing Screws Kontat Perio Level is intended to be used to enclose healing screws that are to be sterilized by a health care provider. Kit for Healing Screws Kontat Perio Level is intended to allow sterilization of the enclosed medical devices. Healing Screws Kit, Perio Level requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Kit for Healing Screws Kontat Perio Level in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes dry time. Kit for Healing Screws Kontat Perio Level is intended for sterilization of non-porous loads.

Kit for Healing Screws Kontat Perio Level is recommended not to be stacked during sterilization.

The combined weight of the Kit for Healing Screws Kontat Perio Level and the associated instruments is 200.0 grams. The weight of the empty Kit for Healing Screws Kontat Perio Level is 169.0 grams.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{6}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental Scanbodies Kit

Indications for Use (Describe)

Scanbodies Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Scanbodies Kit is intended to allow sterilization of the enclosed medical devices.

Scanbodies Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.)

Scanbodies Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Scanbodies Kit is intended for sterilization of non-porous loads.

Scanbodies Kit is recommended not to be stacked during sterilization.

The combined weight of the Scanbodies Kit and the associated instruments is 259.7 grams.

The weight of the empty Scanbodies Kit is 233.9 grams.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{7}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental Pick-up Impression Coping Cica Kit

Indications for Use (Describe)

Pick-up Impression Coping Cica Kit is intended to be used to enclose pick-up impression copings that are to be sterilized by a health care provider. Pick-up Impression Coping Cica Kit is intended to allow sterilization of the enclosed medical devices. Pick-up Impression Coping Cica Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.)

Pick-up Impression Coping Cica Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Pick-up Impression Coping Cica Kit is intended for sterilization of non-porous loads.

Pick-up Impression Coping Cica Kit is recommended not to be stacked during sterilization.

The combined weight of the Pick-up Impression Coping Cica Kit and the associated instruments is 338.0 grams.

The weight of the empty tray Pick-up Impression Coping Cica Kit is 291.4 grams.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{8}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental Pop-up Impression Coping Cica Kit

Indications for Use (Describe)

Pop-up Impression Coping Cica Kit is intended to be used to enclose pop-up impression copings that are to be sterilized by a health care provider. Pop-up Impression Coping Cica Kit is intended to allow sterilization of the enclosed medical devices. Pop-up Impression Coping Cica Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.)

Pop-up Impression Coping Cica Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) = Exposure at 132 °C for 4 minutes and 20 minutes dry time. Pop-up Impression Coping Cica Kit is intended for sterilization of non-porous loads.

Pop-up Impression Coping Cica Kit is recommended not to be stacked during sterilization.

The combined weight of the Pop-up Impression Coping Cica Kit and the associated instruments is 317.1 grams.

The weight of the empty kit Pop-up Impression Coping Cica Kit is 270.9 grams.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{9}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental Multi Kontact Surgical Kit

Indications for Use (Describe)

Multi Kontact Surgical Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Multi Kontact Surgical is intended to allow sterilization of the enclosed medical devices. Multi Kontact Surgical requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.) Multi Kontact Surgical is to be enclosed in a sterlization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes dry time.

Multi Kontact Surgical is intended for sterilization of non-porous loads.

Multi Kontact Surgical is recommended not to be stacked during sterilization.

The combined weight of the Multi Kontact Surgical and the associated instruments is 580.7 grams.

The weight of the empty Multi Kontact Surgical is 449.6 grams.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{10}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental Try-in Abutment Kit

Indications for Use (Describe)

Try-in Abutment Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Try-in Abutment Kit is intended to allow sterilization of the enclosed medical devices. Try-in Abutment Kit requires the use of a double pouch to maintain the sterlity of the enclosed devices (pouch must be FDA cleared in the U.S.). Try-in Abutment Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes dry time.

Trv-in Abutment Kit is intended for sterilization of non-porous loads.

Try-in Abutment Kit is recommended not to be stacked during sterilization.

The combined weight of the Try-in Abutment Kit and the associated devices is 469.5 grams.

The weight of the empty Try-in Abutment Kit is 463.8 grams

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{11}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental AtlaSurgery Guided Surgery Kit

Indications for Use (Describe)

AtlaSurgery Guided Surgery Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. AtlaSurgery Guided Surgery Kit is intended to allow sterilization of the enclosed medical devices. AtlaSurgery Guided Surgery Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

AtlaSurgery Guided Surgery Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) = Exposure at 132 °C for 4 minutes and 20 minutes dry time.

AtlaSurqery Guided Surgery Kit is intended for sterilization of non-porous loads.

AtlaSurgery Guided Surgery Kit is recommended not to be stacked during sterilization.

The combined weight of the AtlaSurgery Guided Surgery Kit and the associated instruments is 689.2 grams.

The weight of the empty AtlaSurgery Guided Surgery Kit is 476.0 grams.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{12}------------------------------------------------

510(k) Number (if known)

K212068

Device Name

Biotech Dental Pilot Drills Kit for Guided Surgery

Indications for Use (Describe)

Pilot Drills Kit for Guided Surgery is intended to be used to group other medical devices that are to be sterilized by a health care provider. Pilot Drills Kit for Guided Surgery is intended to allow sterlization of the group medical devices. Pilot Drills Kit for Guided Surgery requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Pilot Drills Kit for Guided Surgery is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes dry time.

Pilot Drills Kit for Guided Surgery is intended for sterilization of non-porous loads.

Pilot Drills Kit for Guided Surgery is recommended not to be stacked during sterilization.

The combined weight of the Pilot Kit for Guided Surgery and the associated instruments is 39.4 grams.

The weight of the empty Pilot Drills Kit for Guided Surgery is 36.8 grams.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{13}------------------------------------------------

510(k) Summary

K212068

Biotech Dental Kits

Biotech Dental, SAS

October 29, 2021

ADMINISTRATIVE INFORMATION

Manufacturer NameBiotech Dental, SAS
305, Allées de Craponne
13300 Salon de Provence
France
Telephone:+33 4 90 44 60 60
Fax:+33 4 90 44 60 61
Official ContactDelphine Mercier, Vice President, Compliance
Representative/ConsultantMelissa Burbage, Senior Regulatory Specialist
Kevin A. Thomas, PhD
Floyd G. Larson, MS, MBA
PaxMed International, LLC
12264 El Camino Real, Suite 400
San Diego, CA 92130
Telephone: +1 858-792-1235
Fax: +1 858-792-1236
Email: mburbage@paxmed.com
kthomas@paxmed.com, flarson@paxmed.com

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary NameBiotech Dental Kits
Common NameInstrument sterilization trays
Regulation Number21 CFR 880.6850
Regulation NameSterilization Wrap Containers, Trays, Cassettes & Other Accessories
Regulatory ClassClass II
Product CodeKCT
Classification PanelGeneral Hospital
Reviewing OfficeOffice of Surgical and Infection Control Devices (OHT4)
Reviewing DivisionDivision of Infection Control and Plastic Surgery Devices(DHT4B)

PREDICATE DEVICE INFORMATION

The primary predicate device is K191566 Avinent Surgical Tray, Avinent Implant Systems, S.L.U. The additional predicate devices are K182865, Neodent Instrument Kits, JJGC Indústria e Comércio de Materiais Dentários S.A., and K173391, OsteoMed MMF Sterilization Tray, OsteoMed, LLC.

{14}------------------------------------------------

INDICATIONS FOR USE STATEMENT

Indications for Use for Drill Stop Kit

Drill Stop Kit is intended to be used to enclose drill stops that are to be sterilized by a health care provider. Drill Stop Kit is intended to allow sterilization of the enclosed medical devices. Drill Stop Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Drill Stop Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time.

Drill Stop Kit is intended for sterilization of non-porous loads.

Drill Stop Kit is recommended not to be stacked during sterilization.

The combined weight of the Drill Stop Kit and the associated devices is 199.8 grams.

The weight of the empty Drill Stop Kit is 192.6 grams.

Indications for Use for Prosthetic Kit

Prosthetic Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Prosthetic Kit is intended to allow sterilization of the enclosed medical devices.

Prosthetic Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Prosthetic Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Prosthetic Kit is intended for sterilization of non-porous loads.

Prosthetic Kit is recommended not to be stacked during sterilization.

The combined weight of the Prosthetic Kit and the associated instruments is 178.8 grams.

The weight of the empty Prosthetic Kit is 141.8 grams

Indications for Use for Healing Screws Kit

Healing Screws Kit is intended to be used to enclose healing screws that are to be sterilized by a health care provider. Healing Screws Kit is intended to allow sterilization of the enclosed medical devices. Healing Screws Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Healing Screws Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time.

Healing Screws Kit is intended for sterilization of non-porous loads.

Healing Screws Kit is recommended not to be stacked during sterilization.

The combined weight of the Healing Screws Kit and the associated instruments is 203.1 grams.

The weight of the empty kit Healing Screws Kit is 168.9 grams.

Indications for Use for Kit for Healing Screws Kontact Perio Level

Kit for Healing Screws Kontact Perio Level is intended to be used to enclose healing screws that are to be sterilized by a health care provider. Kit for Healing Screws Kontact Perio Level is intended to allow sterilization of the enclosed medical devices. Kit for Healing Screws Kontact Perio Level requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Kit for Healing Screws Kontact Perio Level is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Kit for Healing Screws Kontact Perio Level is intended for sterilization of non-porous loads.

Kit for Healing Screws Kontact Perio Level is recommended not to be stacked during sterilization.

The combined weight of the Kit for Healing Screws Kontact Perio Level and the associated instruments is 206.0 grams.

The weight of the empty Kit for Healing Screws Kontact Perio Level is 169.0 grams.

{15}------------------------------------------------

510(k) Summary Page 3 of 8

Indications for Use for Scanbodies Kit

Scanbodies Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Scanbodies Kit is intended to allow sterilization of the enclosed medical devices.

Scanbodies Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Scanbodies Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Scanbodies Kit is intended for sterilization of non-porous loads.

Scanbodies Kit is recommended not to be stacked during sterilization.

The combined weight of the Scanbodies Kit and the associated instruments is 259.7 grams.

The weight of the empty Scanbodies Kit is 233.9 grams.

Indications for Use for Pick-up Impression Coping Cica Kit

Pick-up Impression Coping Cica Kit is intended to be used to enclose pick-up impression copings that are to be sterilized by a health care provider. Pick-up Impression Coping Cica Kit is intended to allow sterilization of the enclosed medical devices. Pick-up Impression Coping Cica Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Pick-up Impression Coping Cica Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Pick-up Impression Coping Cica Kit is intended for sterilization of non-porous loads.

Pick-up Impression Coping Cica Kit is recommended not to be stacked during sterilization.

The combined weight of the Pick-up Impression Coping Cica Kit and the associated instruments is 338.0 grams. The weight of the empty tray Pick-up Impression Coping Cica Kit is 291.4 grams.

Indications for Use for Pop-up Impression Coping Cica Kit

Pop-up Impression Coping Cica Kit is intended to be used to enclose pop-up impression copings that are to be sterilized by a health care provider. Pop-up Impression Coping Cica Kit is intended to allow sterilization of the enclosed medical devices. Pop-up Impression Coping Cica Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Pop-up Impression Coping Cica Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Pop-up Impression Coping Cica Kit is intended for sterilization of non-porous loads.

Pop-up Impression Coping Cica Kit is recommended not to be stacked during sterilization.

The combined weight of the Pop-up Impression Coping Cica Kit and the associated instruments is 317.1 grams. The weight of the empty kit Pop-up Impression Coping Cica Kit is 270.9 grams.

Indications for Use for Multi Kontact Surgery Kit

Multi Kontact Surgical Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Multi Kontact Surgical Kit is intended to allow sterilization of the enclosed medical devices. Multi Kontact Surgical Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Multi Kontact Surgical Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Multi Kontact Surgical Kit is intended for sterilization of non-porous loads.

Multi Kontact Surgical Kit is recommended not to be stacked during sterilization.

The combined weight of the Multi Kontact Surgical Kit and the associated instruments is 580.7 grams.

The weight of the empty Multi Kontact Surgical Kit is 449.6 grams.

{16}------------------------------------------------

510(k) Summary Page 4 of 8

Indications for Use for Try-in Abutment Kit

Try-in Abutment Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Try-in Abutment Kit is intended to allow sterilization of the enclosed medical devices. Try-in Abutment Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Try-in Abutment Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Try-in Abutment Kit is intended for sterilization of non-porous loads.

Try-in Abutment Kit is recommended not to be stacked during sterilization.

The combined weight of the Try-in Abutment Kit and the associated devices is 469.5 grams.

The weight of the empty Try-in Abutment Kit is 463.8 grams.

Indications for Use for AtlaSurgery Guided Surgery Kit

AtlaSurgery Guided Surgery Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. AtlaSurgery Guided Surgery Kit is intended to allow sterilization of the enclosed medical devices. AtlaSurgery Guided Surgery Kit requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

AtlaSurgery Guided Surgery Kit is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. AtlaSurgery Guided Surgery Kit is intended for sterilization of non-porous loads.

AtlaSurgery Guided Surgery Kit is recommended not to be stacked during sterilization.

The combined weight of the AtlaSurgery Guided Surgery Kit and the associated instruments is 689.2 grams.

The weight of the empty AtlaSurgery Guided Surgery Kit is 476.0 grams.

Indications for Use for Pilot Drills Kit for Guided Surgery

Pilot Drills Kit for Guided Surgery is intended to be used to group other medical devices that are to be sterilized by a health care provider. Pilot Drills Kit for Guided Surgery is intended to allow sterilization of the group medical devices. Pilot Drills Kit for Guided Surgery requires the use of a double pouch to maintain the sterility of the enclosed devices (pouch must be FDA cleared in the U.S.).

Pilot Drills Kit for Guided Surgery is to be enclosed in a sterilization double pouch that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

In the U.S. only, dynamic air removal (pre-vacuum) – Exposure at 132 °C for 4 minutes and 20 minutes dry time. Pilot Drills Kit for Guided Surgery is intended for sterilization of non-porous loads.

Pilot Drills Kit for Guided Surgerv is recommended not to be stacked during sterilization.

The combined weight of the Pilot Drills Kit for Guided Surgery and the associated instruments is 39.4 grams. The weight of the empty Pilot Drills Kit for Guided Surgery is 36.8 grams.

SUBJECT DEVICE DESCRIPTION

The subject device trays are reusable rigid containers, comprising a base (bottom), a removable inner trav (if applicable), and a lid (cover). The subject device trays are to be used to organize and protect the dental healing screws, try-in abutments, instruments, and accessories that are sterilized in the trays by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior contents during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared pouch to maintain sterility.

The subject device includes a total of 10 sizes (same lid and base and enclosed volumes), and a total of 11 tray configurations. The subject device has seven (7) trays that are 2-piece plastic trays manufactured

{17}------------------------------------------------

from Polyphenylsulfone (Radel® R-5000) for the lid and Polypropylene (PPHS) or Polysulfone for the base. The subject device has three (3) trays that are 3-piece plastic trays manufactured from Polyphenylsulfone (Radel R-5000) with silicone grommets. The subject device has one (1) tray that is a 2-piece metal tray manufactured from stainless steel.

TECHNOLOIGCAL CHARACTERISTIC COMPARSION TO MARKETED DEVICES

The subject device and the predicate devices have intended use, the same product classification and product code (KCT) and have similar Indications for Use statements. The subject device and the predicate devices are reusable rigid containers used to organize and protect the instruments that are sterilized by the healthcare provider. The subject device and the predicate devices are to be sterilized by moist heat (steam) and require the use of an FDA cleared pouch to maintain sterility.

Differences between the IFUS for the subject device and the predicate devices include structure of language, the names of the kits, and the weights of the individual kits (loaded and empty). The slight differences in language of the IFUS do not affect the intended use.

The subject device is provided in 10 sizes and 11 configurations; the primary predicate device K191566 is provided in one (1) size and one (1) configuration, the additional predicate device K182865 is provided in four (4) sizes and four (4) configurations, and the additional predicate device K173391 is provided in one (1) size and one (1) configuration. The subject device and the predicate devices have similar overall dimensions, enclose similar volumes, and have similar vent to volume ratios. Differences in the dimensions and vent to volume ratios between the subject device and the predicate devices are mitigated by the sterilization validation performed.

The subject device has seven (7) trays that are 2-piece plastic trays similar to the additional predicate

device K182865 in that both include components manufactured from polyphenylsulfone (Radel R-5000) for the lid and polypropylene (PPHS) or polysulfone for the base. Both materials have a history of biocompatibility and clinical use for the intended or cleared indications.

The subject device has three (3) trays that are 3-piece plastic trays similar to the primary predicate device K191566 in that both include components manufactured from polyphenvlsulfone (Radel R-5000) with silicone grommets. Both materials have a history of biocompatibility and clinical use for the intended or cleared indications.

The subject device has one (1) tray that is a 2-piece metal tray similar to the additional predicate device K173391 in that both include components manufactured from metal (stainless steel or aluminum). This material also has a history of biocompatibility and clinical use for the intended or cleared indications.

The subject device and the predicate devices are to be used according to similar labeling, including the use of an FDA cleared pouch and similar sterilization processes and cycles.

{18}------------------------------------------------

SUMMARY OF NON-CLINICAL TESTING

Provided below is a summary of the non-clinical performance testing included in this submission, including the test methods, purpose, acceptance criteria, and results.

Test MethodologyPurposeAcceptance CriteriaResults
Automated Cleaning ValidationFDA Guidance ReprocessingMedical Devices in Health CareSettings: Validation Methodsand Labeling(issued March 2015)The purpose of this test is tovalidate that the cleaninginstructions provided in theInstructions for Useappropriately clean the tray,and to ensure thesterilization cycle will beeffective.Protein assay - Micro BCA Protein AssayAcceptance criterion: less than 6.4 µg/cm²Assay quantitation limits: 5.0 µg/mLPass
and Labeling(issued March 2015)appropriately clean the tray,and to ensure thesterilization cycle will beeffective.TOC assayAcceptance criterion: less than 2.010 mg/LAssay quantitation limits: 2.010 mg/LPass
Sterilization Validationincluding sterilant penetrationand dry time validationANSI/AAMI/ISO 17665-1ANSI/AAMI/ISO 17665-2The purpose of this test is tovalidate that the sterilizationinstructions listed in theInstructions for Useappropriately sterilize thetray and contents.Acceptance criterion:3 consecutive half-cycles performed foreach of the two sizes of trays willdemonstrate complete inactivation of allbiologic indicators;A minimum SAL of $10^{-6}$ is achieved if theInstructions for Use are followedPass
Dry timeThe purpose of this test is tovalidate that the sterilizationinstructions listed in theInstructions for Useappropriately dry thepouched tray for storage.Acceptance criterion:Using pre-cycle and post-cycle weights,the weight gain after drying will notexceed ± 3%Pass
Life Cycle / Simulated Use LifeValidationFDA Guidance ReprocessingMedical Devices in Health CareSettings: Validation Methodsand Labeling(issued March 2015)The purpose of this test is tovalidate the service life ofthe trays as stated in theInstructions for Use.Acceptance criteria:Visual inspection, component dimensionalfit verification, functional closure/latchverification for 100 use cyclesPass
Biocompatibility of SubjectDevice (by cytotoxicity testing)ANSI/AAMI/ISO 10993-5ANSI/AAMI/ISO 10993-12The purpose of this test is toevaluate the cytotoxicitypotential of the test articleusing an in vitro cell cultureassay.Acceptance criterion:Non-cytotoxic if ≤ 30% reduction of NRUcell viability after exposure to extract ofthe devicePass

Summary of Nonclinical Testing

In summary, the nonclinical testing provided for these devices met the acceptance criteria for each standard and test methodology used to evaluate the devices as shown in the table above

No clinical data were included in this submission.

CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices.

{19}------------------------------------------------

Table of Technological

FeatureSubject DevicePrimary Predicate DeviceAdditional Predicate DeviceAdditional Predicate DeviceComparison
Biotech Dental KitsBiotech Dental SASK191566Avinent Surgical TrayAvinent Implant Systems, S.L.U.K182865Neodent Instrument KitsJJGC Indústria e Comércio de MateriaisDentários S.A.K173391OsteoMed MMF Sterilization TrayOsteoMed LLC
Product ModelNumbersFull: KBK, KPK, KVCK, KPLVCK, KSCANK,KTIPK, KTOPK, TROUSSEKONTMULTI,KPJFKIT, KATLAK, KSFE20KITEmpty: KBKV, KPKV, KVCKV, KPLVCKV,KSCANKV, KTIPKV, KTOPKV, 6000,KPJFKITV, KATLAKV, KSFE20KITVNot listed in summary100.287, 110.294, 110.295, 110.296,Not listed in summary
Product CodeKCTKCTKCTKCTSame
Intended Useorganize and protect the devices, instruments, andaccessories that are sterilized by health careproviderorganize and protect the devices, instruments, andaccessories that are sterilized by health careproviderorganize and protect the devices, instruments, andaccessories that are sterilized by health careproviderorganize and protect the devices, instruments, andaccessories that are sterilized by health careproviderSame
Design2-piece plastic traysRigid polymer base and lid3-piece plastic traysRigid polymer base, lid, and removableinner tray2-piece metal traysRigid metal base, lid, and inner tray.Rigid polymer base, lid, and removable inner trayRigid polymer base, lid, and removable inner trayRigid metal base, lid, and removable inner tray2-piece plastic traysSimilar to K1915663-piece plastic traysSimilar to K1828652-piece metal traysSimilar to K173391
Materials2-piece plastic traysPolyphenylsulfone (PPSU) (RadelR5000) [lid]Polypropylene (PPHS) [base]3-piece plastic traysPolyphenylsulfone (PPSU) (RadelR5000) [lid, base, insert]Medical grade silicone [grommets]2-piece metal traysStainless Steel AISI 304LMedical grade silicone [grommets]Polyphenylsulfone (Radel R-5000) [lid, base, tray]Medical grade silicone [grommets/holders]Polysulfone [base, tray]Polyphenylsulfone (Radel R5000) [lid]Medical grade silicone [grommets]Titanium alloy [retention fixtures]Aluminum [metal lid and base]2-piece plastic traysSimilar to K1915663-piece plastic traysSimilar to K1828652-piece metal traysSimilar to K173391
MaterialsCompatible withSterilizationMethodYesYesYesYesSame
PerforatedYes; allows moist heat (steam) penetration toachieve sterilizationYes; allows moist heat (steam) penetration toachieve sterilizationYes; allows moist heat (steam) penetration toachieve sterilizationYes; allows moist heat (steam) penetration toachieve sterilizationSame
Number of OverallSizes10141Similar
Number ofConfigurations11241Similar
OverallDimensionsLength x Width x HeightVent to VolumeRatioLength x Width x HeightVent to VolumeRatioLength x Width x HeightVent to VolumeRatioLength x Width x HeightVent to VolumeRatioSimilar to K191566
andVent to VolumeRatio150 mm x 65 mm x 25 mm$0.2259 cm^2/cm^3$207.5 mm x 157.5 mm x 71.7mm$0.0043 cm^2 / cm^3$(0.011 in $^2$ / in $^3$ )264 mm x 163 mm x 54 mm$0.0102 cm^2 / cm^3$241.3 mm x 127 mm x 38.1mm$0.0197 cm^2 / cm^3$(0.050 in $^2$ / in $^3$ )
150 mm x 65 mm x 25 mm$0.1472 cm^2/cm^3$195 mm x 90 mm x 54 mm$0.0191 cm^2 / cm^3$
150 mm x 65 mm x 25 mm$0.3694 cm^2/cm^3$195 mm x 90 mm x 44 mm$0.0247 cm^2 / cm^3$
150 mm x 65 mm x 25 mm$0.3932 cm²/cm³$264 mm x 90 mm x 58 mm$0.0093 cm²/cm³$
125 mm x 127 mm x 24.5 mm$0.1224 cm²/cm³$
125 mm x 127 mm x 36.5 mm$0.0346 cm²/cm³$
125 mm x 127 mm x 32.5 mm$0.0567 cm²/cm³$
162.2 mm x 170 mm x 55 mm$0.0049 cm²/cm³$
162.2 mm x 170 mm x 55 mm$0.0049 cm²/cm³$
162.2 mm x 170 mm x 65 mm$0.0041 cm²/cm³$
40 mm x 50 mm x 20 mm$0.9500 cm²/cm³$
ReusableYesYesYesYesSame
Use Life TestingReusable up to 100 cyclesAssembled, sterilizedVisual inspectionComponent dimensional fit verificationFunctional closure (lid-base latch) verificationReusable up to 100 cyclesAssembled, sterilizedVisual inspectionReusable up to 100 cyclesAssembled/disassembled, cleaned, sterilizedVisual inspectionComponent dimensional fit verificationFunctional closure (lid-base latch) verificationNot listed in 510(k) SummarySimilar
Sterilization Method
SterilantMoist heat (steam)Moist heat (steam)Moist heat (steam)Same
CyclesFractionated vacuum (pre-vacuum)Exposure at 132 oC (270 oF) for 4 minutes with 20minutes drying time.Fractionated vacuum (pre-vacuum)Exposure at 132 oC (270 oF) for 4 minutes with 30minutes drying time.Gravity displacement and fractionated vacuum(pre-vacuum)Fractionated vacuum (pre-vacuum) - Exposure at132 oC for 4 minutes, 20 minute dry timeGravity displacement – Exposure at 132 oC for 15minutes, 20 minute dry timeFractionated vacuum (pre-vacuum)Exposure at 132 oC (270 oF) for 4 minutes with 30minutes drying time.Same
Sterile BarrierSterilization pouch, FDA cleared for indicatedmethod and cycleSterilization wrap, FDA cleared for indicatedmethod and cycleSterilization wrap, FDA cleared for indicatedmethod and cyclesSterilization wrap, FDA cleared for indicatedmethod and cycleSame

{20}------------------------------------------------

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).