(30 days)
Not Found
No
The summary describes a device based on Vibration-Controlled Transient Elastography (VCTE™) technology using ultrasound to measure shear wave speed and estimate tissue stiffness and attenuation. There is no mention of AI, ML, or related concepts in the device description, intended use, or performance studies. The performance evaluation focuses on bias and precision compared to a predicate device, not on the performance of an AI/ML algorithm.
No
Explanation: The device is described as providing measurements and estimates of tissue stiffness and attenuation, which are explicitly stated to be used "as an aid to clinical management" or "as an aid to diagnosis and monitoring." It does not directly provide therapy or treatment.
Yes
The device provides measurements and estimates (shear wave speed, tissue stiffness, and ultrasound coefficient of attenuation) that are explicitly stated to be used "as an aid to diagnosis and monitoring of adult patients with liver disease." This direct mention of aiding diagnosis classifies it as a diagnostic device.
No
The device description explicitly states that the FibroScan® 230 system includes a smaller equipment unit (which includes the hand-held probes) in addition to the FibroScan® application. This equipment unit is a hardware component.
Based on the provided information, the FibroScan® 230 is not an In Vitro Diagnostic (IVD) device.
Here's why:
- Intended Use: The intended use describes the device as providing measurements of shear wave speed, tissue stiffness, and ultrasound coefficient of attenuation in internal structures of the body (specifically the liver). These are physical measurements taken directly from the patient's body.
- Device Description: The device description explains that it uses ultrasound and mechanical vibration to measure properties within the liver.
- Lack of In Vitro Activity: An IVD device is designed to examine specimens derived from the human body (like blood, urine, tissue samples) in vitro (outside the body) to provide information about a physiological state, health, or disease. The FibroScan® 230 does not process or analyze such specimens. It directly interacts with the patient's liver.
Therefore, the FibroScan® 230 is a medical device used for non-invasive physical measurements in vivo (within the living body), not an IVD.
N/A
Intended Use / Indications for Use
The FibroScan® 230 is intended to provide shear wave speed measurements and estimates of tissue stiffness as well as ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body. The Shear wave speed and stiffness measurements may be used as an aid to clinical management of adult patients with liver disease.
The FibroScan® 230 is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as determining a 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).
The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
Shear wave speed and stiffness, and CAP may be used as an aid in the clinical management of pediatric patients with liver disease.
Product codes (comma separated list FDA assigned to the subject device)
IYO, ITX
Device Description
FibroScan® System and its probes is an active non-implantable medical device using ultrasound and based on Vibration- Controlled Transient Elastography (VCTETM) technology. It is designed to perform non-invasive measurements of liver shear wave speed and estimate tissue stiffness. The probe, containing a mechanical vibrator, produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. Ultrasound is used to track the shear (elastic) wave, measure its speed and provide estimated stiffness.
The focus of this submission is the new FibroScan® 230, which separates the FibroScan® System into two parts: a smaller equipment unit (which includes the hand-held probes) and the FibroScan® application, installed on the user's computer. The FibroScan® Application displays the user interface of the system and interacts with the equipment unit through a USB connection. The FibroScan® Application may also interact with the Echosens Cloud through an internet network. Only the FibroScan® 230 equipment and FibroScan® application are included in the FibroScan® 230 system while the Echosens cloud and the end-user computer are not part of the FibroScan® 230 medical device.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Ultrasound
Anatomical Site
Liver
Indicated Patient Age Range
Adult and pediatric patients
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The FibroScan® 230 effectiveness was evaluated in terms of its bias and precision in measuring liver shear wave speed (m/s) and CAP, using the S+, M+ and XL+ probes. Both a comparison with the predicates' documented performance (as presented in K203273) and a direct comparison of the measurement accuracy between the proposed and cleared systems were performed.
Direct comparison refers to testing the predicate and candidate device contemporaneously using the same probes, phantoms, and experimental setup, thereby controlling for environmental conditions (temperature, humidity), imperfections in the materials and manufacturing process leading to sample inhomogeneities, as well as the viscoelastic properties of the materials used to construct the phantoms which may lead to changes in their stiffness values.
The shear wave speed bias measurement and the precision measurement was evaluated and compared to the documented corresponding value reported for the predicate device. Results show that, for liver examination, the bias and precision measurements of the candidate FibroScan® 230 system and the predicate FibroScan® 530 Compact system documented similar performance. Specifically, the ranges of the bias values documented for the candidate are substantially equivalent to the ranges of the bias values documented for the predicate device (4.9% - 14.6% for the candidate and 0.5% - 12.6% for the precision values documented for the candidate are very similar to the ranges of the precision values documented for the predicate device.
Direct comparison of the CAP results, using dynamic testing (>200 measurements) on multiple spots demonstrated the range and mean values of CAP bias were substantially equivalent to the bias of the CAP in the candidate device (range of 1.4% - 11.1% for the candidate CAP dynamic testing and 1.8% – 10.1% for the predicate). In addition, the range and mean values of CAP precision values were equivalent to the precision of the CAP in the candidate device.
An additional direct comparison between the candidate and predicate devices showed that the bias values for the shear wave speed and CAP were shown to be comparable (under 5% difference) between FibroScan® 230 and FibroScan® 530 Compact. Therefore, the proposed systems have substantially equivalent effectiveness in measuring shear wave speed and CAP bias.
Furthermore, the stiffness and CAP measurement ranges, as well as the SmartDepth ranges, have been verified on the FibroScan® 230 system and are the predicate device.
In summary, the bias and the precision of the shear wave speed and the CAP measured by the FibroScan® 230 are within the same ranges as those of the predicate FibroScan® 530 Compact cleared in K203273 and, therefore, considered substantially equivalent.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1560 Ultrasonic pulsed echo imaging system.
(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
July 30, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Echosens % Zvi Ladin, Ph.D. Principal Boston MedTech Advisors Inc. 990 Washington Street, Suite #204 DEDHAM MA 02026
Re: K212035
Trade/Device Name: FibroScan® 230 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, ITX Dated: June 25, 2021 Received: June 30, 2021
Dear Dr. Ladin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
1
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known)
Device Name FibroScan® 230
Indications for Use (Describe)
The FibroScan® 230 is intended to provide shear wave speed measurements and estimates of tissue stiffiness as well as ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body. The Shear wave speed and stiffness measurements may be used as an aid to clinical management of adult patients with liver disease.
The FibroScan® 230 is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as determining a 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).
The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
Shear wave speed and stiffness, and CAP may be used as an aid in the clinical management of pediatric patients with liver disease.
Type of Use (Select one or both, as applicable) | |
---|---|
Prescription Use (Part 21 CFR 801 Subpart D) | |
Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 1 of 1
3
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
| Manufacturer: | Echosens
6 rue Ferrus, Paris, France, 75014
Telephone: +33 1 44 82 78 56
Fax: +33 1 44 82 78 60 |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Zvi Ladin,
Ph.D.
Principal
Boston MedTech Advisors, Inc.
990 Washington Street; Suite
#204
Dedham, MA 02026
Telephone: (781) 407 0900 x104
Fax: (781) 407 0901
Email: zladin@bmtadvisors.com |
| Date Prepared: | June 25, 2021 |
Name of Device and Name/Address of Sponsor
Trade/Proprietary Name: | FibroScan® 230 |
---|---|
Common Name: | Diagnostic Ultrasound System and Accessories |
Classifications:
Classification Name | Regulation | Product Code |
---|---|---|
Ultrasonic Pulsed Echo Imaging System | 21 CFR §892.1560 | IYO |
Diagnostic Ultrasonic Transducer | 21 CFR §892.1570 | ITX |
| Manufacturing Facility: | Echosens
6 rue Ferrus, Paris, France, 75014
Telephone: +33 1 44 82 78 56
Fax: +33 1 44 82 78 60 |
|-------------------------|----------------------------------------------------------------------------------------------------------|
| Establishment: | |
Registration Number: | 3010258456 |
---|
Predicate Device
This submission claims substantial equivalence to Echosens's FibroScan® 530 Compact, cleared on March 25, 2021 (#K203273).
4
Device Description
FibroScan® System and its probes is an active non-implantable medical device using ultrasound and based on Vibration- Controlled Transient Elastography (VCTE™) technology. It is designed to perform non-invasivemeasurements of liver shear wave speed and estimate tissue stiffness. The probe, containing a mechanical vibrator, produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. Ultrasound is used to track the shear (elastic) wave, measure its speed and provide estimated stiffness.
The focus of this submission is the new FibroScan® 230, which separates the FibroScan® System into two parts: a smaller equipment unit (which includes the hand-held probes) and the FibroScan® application, installed on the user's computer. The FibroScan® Application displays the user interface of the system and interacts with the equipment unit through a USB connection. The FibroScan® Application may also interact with the Echosens Cloud through an internet network. Only the FibroScan® 230 equipment and FibroScan® application are included in the FibroScan® 230 system while the Echosens cloud and the end-user computer are not part of the FibroScan® 230 medical device.
Comparison of Technological Characteristics
The FibroScan® 230 system is substantially equivalent to the FibroScan® 530 Compact system cleared by 510(k) #K203273. Both systems provide 50Hz shear wave speed measurements and estimates of tissue stiffness in the liver using the S+, M+, and XL+ transducers. Additionally, both systems encompass a Controlled Attenuation Parameter (CAP) designed to estimate the ultrasound attenuation (forward and return paths) at the frequency of 3.5 MHz.
The FibroScan® 230 modification relates to its design configuration and therefore includes structural differences in system design software architecture and electronic architecture (e.g., no integrated touchscreen). The FibroScan® 230 system liver measurement is otherwise identical to the predicate FibroScan® 530 Compact system as related to the intended use and indications for use, same operating principle and materials, same S+, M+, and XL+ probes, emits the same energy, same examination procedure, acquires the same information, applies the same signal processing and imaging modes, displays the same information, performance measurements, and manufacturing process. Therefore, the design changes related to FibroScan® 230 do not impact the liver examination. Table 1 provides a detailed comparison of the candidate and predicate devices.
As with the predicate device, the CAP (previously called CAPc) and SmartDepth options are enabled with the 'SmartExam' capability. The SmartDepth automatically adapts the depth of the liver stiffness measurement (LSM) based on the anatomy of the patient and is available for M+ and XL+ probes while CAP is available for S+, M+, and XL+ probes. The LSM and CAP calculated with the SmartDepth capability uses the same algorithm as the predicate system.
5
Recognized Consensus Standards Used
Non-clinical testing to assure compliance with acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety were performed and have been found to conform to applicable standards. The system complies with the following standards:
- . IEC 60601-2-37: Medical Electrical Equipment - Part 2-37: Particular Requirements for The Basic Safety And Essential Performance Of Ultrasonic Medical Diagnostic And Monitoring Equipment; Edition 2.1 2015.
- NEMA UD: Acoustic Output Measurement Standard for Diagnostic Ultrasound . EquipmentRevision 3; 2-2004 (R2009).
- IEC 62127-1: Ultrasonics -- Hydrophones -- Part 1: Measurement and Characterization . ofMedical Ultrasonic Fields Up To 40 Mhz; Edition 1.1 2013-02.
- IEC 62127-2: Ultrasonics -- Hydrophones -- Part 2: Calibration for Ultrasonic Fields Up . To40 Mhz; Edition 1.1 2013-02.
- IEC 62127-03: Ultrasonics -- Hydrophones -- Part 3: Properties of Hydrophones for . Ultrasonic Fields Up To 40 Mhz; Edition 1.1 2013-05.
- IEC 61161: Ultrasonics -- Power Measurement -- Radiation Force Balances and . Performance Requirements; Edition 3.0 2013-01.
- . AAMI / ANSI ES60601-1: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (IEC 60601-1:2005, Mod); 2005/(R) 2012
- IEC 60601-1-2: Medical Electrical Equipment Part 1-2: General Requirements For . Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests; Edition 4: 2014-02.
- IEC 60601-1-6: Medical Electrical Equipment Part 1-6: General Requirements For . BasicSafety And Essential Performance - Collateral Standard: Usability; Edition 3.1 2013-10
- . IEC 62366-1 Edition 1.0 2015-02: Medical Devices - Application Of Usability Engineering To Medical Devices.
- IEC 62304: Medical Device Software Software Life Cycle Processes; First Edition . 2006-05, Equivalent to IEC 62304: 2006/A1:2016.
- . ISO 14971 Second: Medical Devices - Application Of Risk Management To Medical Devices: Edition 2007-03-01.
6
Intended Use / Indications for Use
The FibroScan® 230 is intended to provide shear wave speed measurements and estimates of tissue stiffness as well as ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body. The shear wave speed and stiffness measurements may be used as an aid to clinical management of adult patients with liver disease.
The FibroScan® 230 is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as determining a 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).
The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
Shear wave speed and stiffness, and CAP may be used as an aid in the clinical management of pediatric patients with liver disease.
Performance Data
The FibroScan® 230 effectiveness was evaluated in terms of its bias and precision in measuring liver shear wave speed (m/s) and CAP, using the S+, M+ and XL+ probes. Both a comparison with the predicates' documented performance (as presented in K203273) and a direct comparison of the measurement accuracy between the proposed and cleared systems were performed.
Direct comparison refers to testing the predicate and candidate device contemporaneously using the same probes, phantoms, and experimental setup, thereby controlling for environmental conditions (temperature, humidity), imperfections in the materials and manufacturing process leading to sample inhomogeneities, as well as the viscoelastic properties of the materials used to construct the phantoms which may lead to changes in their stiffness values.
The shear wave speed bias measurement and the precision measurement was evaluated and compared to the documented corresponding value reported for the predicate device. Results show that, for liver examination, the bias and precision measurements of the candidate FibroScan® 230 system and the predicate FibroScan® 530 Compact system documented similar performance. Specifically, the ranges of the bias values documented for the candidate are substantially equivalent to the ranges of the bias values documented for the predicate device (4.9% - 14.6% for the candidate and 0.5% - 12.6% for the precision values documented for the candidate are very similar to the ranges of the precision values documented for the predicate device.
Direct comparison of the CAP results, using dynamic testing (>200 measurements) on multiple spots demonstrated the range and mean values of CAP bias were substantially equivalent to the bias of the CAP in the candidate device (range of 1.4% - 11.1% for the candidate CAP dynamic testing and 1.8% – 10.1% for the predicate). In addition, the range and mean values of CAP precision values were equivalent to the precision of the CAP in the candidate device.
7
An additional direct comparison between the candidate and predicate devices showed that the bias values for the shear wave speed and CAP were shown to be comparable (under 5% difference) between FibroScan® 230 and FibroScan® 530 Compact. Therefore, the proposed systems have substantially equivalent effectiveness in measuring shear wave speed and CAP bias.
Furthermore, the stiffness and CAP measurement ranges, as well as the SmartDepth ranges, have been verified on the FibroScan® 230 system and are the predicate device.
In summary, the bias and the precision of the shear wave speed and the CAP measured by the FibroScan® 230 are within the same ranges as those of the predicate FibroScan® 530 Compact cleared in K203273 and, therefore, considered substantially equivalent.
Clinical Data
No clinical data was required for this submission.
Substantial Equivalence Discussion
The FibroScan® 230 includes the same intended use and indications for use, uses the same operating principle and materials, emits the same energy and acquires the same information as the predicate FibroScan® 530 Compact device (K203273). The separation of the system unit from the FibroScan® application, installed on the user's computer, do not raise new or different questions of safety or efficacy.
The candidate device was demonstrated to be substantially equivalent to its predicate device FibroScan® 530 (K203273) based on the aforementioned verification and validation studies.
Conclusion
The conclusions drawn from the clinical and nonclinical information provided in this submission demonstrate that FibroScan® 230 is as safe, as effective, and performs as well as the legally marketed device predicate. The FibroScan® 230 raises no new or different issues of safety or effectiveness and is substantially equivalent to its predicate device.
8
FibroScan® 530 Compact | FibroScan® 230 | |
---|---|---|
510(k) # (Clearance) | K203273 | |
(March 25, 2021) | Candidate Device | |
TBD | ||
Manufacturer | Echosens | Echosens |
Indications for Use | The FibroScan® is intended to provide shear wave speed measurements and estimates of tissue | |
stiffness as well as ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in | ||
internal structures of the body. The Shear wave speed and stiffness measurements may be used as an | ||
aid to clinical management of adult patients with liver disease. | ||
The FibroScan® is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and | ||
estimates of stiffness as well as determining a 3.5 MHz ultrasound coefficient of attenuation (CAP: | ||
Controlled Attenuation Parameter). | ||
The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of | ||
adult patients with liver disease, as part of an overall assessment of the liver. | ||
Shear wave speed and stiffness, and CAP may be used as an aid in the clinical management of | ||
pediatric patients with liver disease. | The FibroScan® is intended to provide shear wave speed measurements and estimates of tissue | |
stiffness as well as ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in | ||
internal structures of the body. The Shear wave speed and stiffness measurements may be used as an | ||
aid to clinical management of adult patients with liver disease. | ||
The FibroScan® is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and | ||
estimates of stiffness as well as determining a 3.5 MHz ultrasound coefficient of attenuation (CAP: | ||
Controlled Attenuation Parameter). | ||
The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of | ||
adult patients with liver disease, as part of an overall assessment of the liver. | ||
Shear wave speed and stiffness, and CAP may be used as an aid in the clinical management of | ||
pediatric patients with liver disease. | ||
Application | Abdominal | Abdominal |
Imaging Modes | A-mode / M-mode | |
Transient Elastography/ Shear Wave / (CAP™) | A-mode / M-mode | |
Transient Elastography/ Shear Wave / (CAP™) | ||
Ultrasound | ||
Probes | Piezoelectric ultrasound source | |
M+-probe (3.5 MHz) | ||
XL+ probe (2.5 MHz) | ||
S+ probe (5 MHz) | ||
(single element ultrasound transducer) | Piezoelectric ultrasound source | |
M+-probe (3.5 MHz) | ||
XL+ probe (2.5 MHz) | ||
S+ probe (5 MHz) | ||
(single element ultrasound transducer) | ||
Depth Analysis Method | Fixed Depth: | |
S1 exam : 15-40 mm | ||
S2 exam : 20-50 mm | ||
Adaptive Depth (SmartDepth): | ||
M exam: 25-65 / 30-70 mm | ||
XL exam: 35-75 mm/ 40-80 mm/ 45-85 mm | Fixed Depth: | |
S1 exam : 15-40 mm | ||
S2 exam : 20-50 mm | ||
Adaptive Depth (SmartDepth): | ||
M exam: 25-65 / 30-70 mm | ||
XL exam: 35-75 mm/ 40-80 mm/ 45-85 mm | ||
VCTE™ Mode | Shear wave speed measurements and tissue | |
stiffness | Shear wave speed measurements and tissue | |
stiffness | ||
VCTE™ Range | Shear wave speed (0.8-5.0 m/s) | |
Stiffness (2.0-75 kPa) | Shear wave speed (0.8-5.0 m/s) | |
Stiffness (2.0-75 kPa) | ||
VCTE™ Display | Shear wave speed and stiffness medians and | |
IQR/median ratio | Shear wave speed and stiffness medians and | |
IQR/median ratio | ||
Attenuation Mode | Controlled Attenuation Parameter (CAP) | Controlled Attenuation Parameter (CAP) |
Attenuation Range | CAP value (100-400 dB/m) | CAP value (100-400 dB/m) |
Attenuation Display (Second | ||
generation/ Enhanced CAP) | CAP mean and standard deviation | CAP mean and standard deviation |
Attenuation Display - Probes | ||
compatibility | S+ Probe | |
M+ Probe | ||
XL Probe | S+ Probe | |
M+ Probe | ||
XL Probe | ||
Size and Weight | 460 mm x 360 mm x 250 mm (H x W x D) | |
10kg with accessories | 265.5mm x 157mm x 200mm | |
(H x W x D) | ||
4.4kg with accessories | ||
Power supply | 100-240 V ~ 50-60 Hz | 100-240 V ~ 50-60 Hz |
Core Component | Elastography engine | |
Analog front end High frequency (US): PV3 | ||
Analog front end Low frequency (servo control): | ||
PV3 | Elastography engine | |
Analog front end High frequency (US): PV3 | ||
Analog front end Low frequency (servo control): | ||
PV3 | ||
Operating system | Windows 10 Embedded | Windows 10 Embedded |
Screen | Color LCD touch screen, 15-inch. | User's computer (Screen resolution req: 1024 x |
-
|
| Internet | Internet connection included as a feature | An Internet connection is required prior to first
usage |
| Battery | ARTS Energy (ref. 4 INR19/66-2)
Part number 806957 / M300002 | N/A |
| Accessories | Roll Stand 530 | N/A |
Table 1. Comparison of FibroScan® 230 to FibroScan® 530 Compact Predicate
9
The following abbreviations are used in the table: US – Interquartile Range, VCTE™ – Vibration Controlled Transient Elastography