K Number
K211966

Validate with FDA (Live)

Device Name
Segment 3DPrint
Manufacturer
Date Cleared
2022-05-06

(316 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Segment 3DPrint is a software for review and segmentation of images from a medical scanner as well as of medical 3D models. Segment 3DPrint is intended to generate 3D models for diagnostic purposes in both paediativ and adult populations in the field of orthopaedic, maxillofacial, and cardiovascular applications. The models can be used for visualisation, measuring, and treatment planning. Output from Segment 3DPrint can be used to fabrical replical replical replical replical replical replical replical replical by use of additive manufacturing methods. Segment 3DPrint is intended to be used by trained professionals in conjunction with expert clinical judgement.

Device Description

Seqment 3DPrint is a software for review and segmentation of images from a medical scanner as well as of medical 3D models. Segment 3DPrint is intended to generate 3D models for diagnostic purposes. The models can be used for visualisation, measuring, and treatment planning. Output from Segment 3DPrint can be used to fabricate physical replicas, by use of additive manufacturing methods.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the Segment 3DPrint device based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / MetricReported Device Performance
Accuracy of final 3D model< 1 mm
Maximum 95th percentile surface distance< 1 mm (for AI bone segmentation)
AI Bone Segmentation - Dice CoefficientMean: 0.96, SD: 0.03
AI Bone Segmentation - Jaccard ScoreMean: 0.92, SD: 0.05
AI Bone Segmentation - Mean Absolute Dist.Mean: 0.23 mm, SD: 0.18 mm
AI Bone Segmentation - Signed Dist. Diff.Mean: 0.03 mm, SD: 0.26 mm
AI Bone Segmentation - 95th PercentileMax: 0.93 mm

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set for AI Bone Segmentation: 21 data sets.
  • Test Set for Print Accuracy (3D Models): 12 models (representing complex structures and worst-case scenarios).
  • Data Provenance: Studies were performed in Europe. The text does not specify exact countries or whether the data was retrospective or prospective. It mentions "a great variety of data (such as scanner model, image quality, and anatomy)" was included.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The text mentions "agreement between reference segmentation by expert readers." However, it does not specify the number of experts used or their specific qualifications (e.g., "radiologist with 10 years of experience").

4. Adjudication Method for the Test Set

The text does not explicitly state the adjudication method used for establishing ground truth (e.g., 2+1, 3+1, none). It only refers to "reference segmentation by expert readers."

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with AI assistance versus without AI assistance was not explicitly described in the provided text. The studies focused on the standalone performance of the device's segmentation and 3D printing accuracy.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance evaluation was largely conducted.

  • The "AI bone segmentation algorithm was trained on 20 data sets, and 21 data sets were used for its validation." The reported metrics (Dice, Jaccard, distances) are measures of this algorithm's performance against ground truth.
  • The "device validation study validates digital models and 3D models from additive manufacturing" with reported accuracy for the generated models.
  • The device is described as a "support tool" for "medically trained professionals" but the performance results are for the algorithm's output directly.

7. The Type of Ground Truth Used

The ground truth for the AI bone segmentation was established by "reference segmentation by expert readers" (expert consensus). For the 3D model accuracy, the text implies that the "validation or application studies using established methods as reference standards" were used, but the exact nature of this reference standard for physical replica accuracy is not fully detailed beyond implying measurement against the physical object vs. the digital model.

8. The Sample Size for the Training Set

  • AI Bone Segmentation: 20 data sets.
  • 3D Models/Print: The text mentions "The device validation study validates digital models and 3D models... In total 12 models... were printed." This appears to be a validation set rather than a training set for print accuracy itself. The training set for the 3D model generation process is not explicitly stated.

9. How the Ground Truth for the Training Set Was Established

The text states that the "AI bone segmentation algorithm was trained on 20 data sets." It can be inferred that the ground truth for these 20 training data sets would have been established in a similar manner to the validation set, likely through "expert readers" creating reference segmentations. However, the document does not explicitly detail the ground truth establishment method for the training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medviso AB % Einar Heiberg, Ph.D. Founder, Vice President, CTO Griffelvägen 3 Lund. SE-22467 SWEDEN

May 6, 2022

Re: K211966

Trade/Device Name: Segment 3DPrint Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: March 30, 2022 Received: April 8, 2022

Dear Einar Heiberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K211966

Device Name Segment 3DPrint

Indications for Use (Describe)

Segment 3DPrint is a software for review and segmentation of images from a medical scanner as well as of medical 3D models. Segment 3DPrint is intended to generate 3D models for diagnostic purposes in both paediativ and adult populations in the field of orthopaedic, maxillofacial, and cardiovascular applications. The models can be used for visualisation, measuring, and treatment planning. Output from Segment 3DPrint can be used to fabrical replical replical replical replical replical replical replical replical by use of additive manufacturing methods. Segment 3DPrint is intended to be used by trained professionals in conjunction with expert clinical judgement.

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

--- Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (k) Summary 5

Submitter 5.1

Medviso AB Griffelvägen 3 SE-224 67 Lund, Sweden +46(0)761-836442

Date Prepared: May 4, 2022 Contact Person: Einar Heiberg, PhD, CTO, +46(0)761-836442, einar@medviso.com

5.2 Device

Device Trade Name: Segment 3DPrint Device Common Name: Image processing system Classification Name: Class II - System, Image Processing Regulation Number: 892.2050(Medical image management and processing system) Product Code: LLZ

Predicate Device 5.3

Mimics Medical (K183105) Materialise N.V. Technologielaan 15 3001 Leuven Belgium

5.4 Device Description

Device Description

Seqment 3DPrint is a software for review and segmentation of images from a medical scanner as well as of medical 3D models. Segment 3DPrint is intended to generate 3D models for diagnostic purposes. The models can be used for visualisation, measuring, and treatment planning. Output from Segment 3DPrint can be used to fabricate physical replicas, by use of additive manufacturing methods.

Image /page/3/Picture/14 description: The image shows the word "MEDVISO" in a light blue color. The word is surrounded by a pattern of small, light blue squares that form a semi-circular shape above and below the word. The squares are arranged in a grid-like pattern and appear to be fading out towards the edges of the semi-circle.

Page 5-1 of 5

{4}------------------------------------------------

Medical images and 3D models may be imported from various sources, including images stored on portable media, network storage devices, and other vendor systems.

Segment 3DPrint meets the identification criteria LLZ 892.2050 - Picture archiving and communications system. Segment 3DPrint is a software with the capability to import and display medical images and to perform digital processing of a rendered 3D object. Segment 3DPrint is a support tool with the means of generating 3D models and should be used by medically trained professionals in conjunction with expert clinical judgement.

The user will interact with Seqment 3DPrint through a graphical user interface on a standard PC platform, using Windows operating system.

5.5 Intended Use

Segment 3DPrint is a software for review and segmentation of images from a medical scanner as well as of medical 3D models. Segment 3DPrint is intended to generate 3D models for diagnostic purposes in both paediatric and adult populations in the field of orthopaedic, maxillofacial, and cardiovascular applications. The models can be used for visualisation, measuring, and treatment planning. Output from Segment 3DPrint can be used to fabricate physical replicas, by use of additive manufacturing methods. Segment 3DPrint is intended to be used by trained professionals in conjunction with expert clinical judgement.

Comparison with Predicate Device

Both Segment 3DPrint and the predicate device Mimics Medical are support tools which provide the healthcare professional(s) with relevant clinical data to support clinical decisions by analyzing the generated 3D models.

Both Seament 3DPrint and Mimics Medical can be used for the fabrication of physical replicas of the output file using additive manufacturing methods.

The intended use for Segment 3DPrint is substantially equivalent to the intended use of the predicate device Mimics Medical.

Technological Characteristics 5.6

Seament 3DPrint and the predicate device are both software packages that can be used for visualization and segmentation of medical images. Both Segment 3DPrint and the predicate device provide a user interface with items for selecting images and adjusting image viewing and can be operated from a personal computer. The subject device and predicate device render segmentations of the region of interest either semiautomatically, manually, or in combination, providing digital 3D models. The technological difference between the subject device and the predicate device is that different algorithms are used for the semi-automatic segmentation approaches. There might be slight differences in features and menu, but these differences between the

Image /page/4/Picture/13 description: The image shows the word "MEDVISO" in a light blue color. The word is surrounded by a pattern of small, light blue squares that form a semi-circular shape above and below the word. The squares appear to be pixelated, giving the image a digital or technological feel. The background is white.

{5}------------------------------------------------

predicate device and the proposed device are not significant since they do not raise any new or potential safety risks to the user or patient and questions of safety or effectiveness. Based on the results of software validation and verification tests, we conclude that Segment 3DPrint is substantially equivalent to the predicate device.

SystemSegment 3DPrintMimics Medical
Version3.223.0.2
ManufacturerMedviso ABMaterialise NV
510(k) numberK211966K183105
Classification892.2050LLZ, Class II892.2050LLZ, Class II
Intended usesamesame
Patient populationAll with images frommedical scanner.Unspecified
Graphical user interfaceYesYes
PlatformPCPC
Operating systemMS Windows 10MS Windows 10
Image display monitorUnspecifiedResolution of 1280x1024 orhigher
Report display monitorNoNo
Patient demographicsYesYes
NetworkingTCP/IPTCP/IP
DICOM compliant imagecompressionLosslessLossless
Image communicationYesYes
Image processing annotationsYesYes
Linear measurement toolsYesYes
Automatic and manualsegmentation of objectYesYes
Automatic filling andsmoothing toolsYesYes
Local and remote imagestorageYesYes
Type of software – customintegratedYesYes
ViewingYesYes
Safety - For use only by alicensed professionalYesYes

Image /page/5/Picture/4 description: The image shows the word "MEDVISO" in a light blue color. The word is surrounded by a pattern of small, light blue squares that form a semi-circular shape above and below the word. The squares appear to be pixelated, giving the image a digital or technological feel. The overall design is simple and clean, with a focus on the text and the surrounding pattern.

{6}------------------------------------------------

5.7 Performance Data

The following performance data were provided in support of the substantial equivalence determination.

The accuracy of the final 3D model generated by Segment 3DPrint is < 1 mm.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documented according to FDA's Guidance "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices". It was concluded that Segment 3DPrint was considered to be of "moderate" level of concern. Extensive testing of the software package is performed by an automated test suite prior to commercial release. As a complement to this, manual testing is performed by Application specialists and the software is evaluated at one beta test site.

Bench and Clinical Studies

The features for Segment 3DPrint have been clinically evaluated using bench and clinical studies. The studies are validation or application studies using established methods as reference standards, and were performed in Europe.

The device validation study validates digital models and 3D models from additive manufacturing (low force stereolithography), representative of the three different application areas. In total 12 models, representing the most complex structures and worst-case scenarios, were printed. The patient characteristics for the validation of the print accuracy included four females and eight males. Mean age was 26 ± 29 years, range 15 days - 79 years. Five cases were maxillofacial, three cases were orthopaedic, and four cases were cardiovascular. This yielded replicas with an accuracy of <1 mm, well suited for clinical use in all intended patient population groups.

Al bone segmentation algorithm was trained on 20 data sets, and 21 data sets were used for its validation. There was no overlap of data between the two sets, and care was taken to include a great variety of data (such as scanner model, image quality, and anatomy). The maximum 95th percentile surface distance between the ground truth segmentation and the resulting image was <1 mm.

The patient characteristics for the validation of the Al bone segmentation includes ten females, four males and seven of unknown sex. Mean age was 33 ± 26 years, range 15 days - 76 years. Table 1 below presents the agreement between reference segmentation by expert readers and the segmentation by the automatic segmentation algorithm.

Image /page/6/Picture/12 description: The image shows the word "MEDVISO" in a light blue color. The word is surrounded by a pattern of small, light blue squares that form a curved shape above and below the word. The squares are more concentrated near the word and become more sparse as they move away from it, creating a gradient effect.

Page 5-4 of 5

{7}------------------------------------------------

SetDiceJaccDist [mm]Dist[mm]95th percentile[mm]
Mean0.960.920.230.03-
SD0.030.050.180.26-
Max----0.93

Table 1 - Dice is dice coefficient, Jacc is Jaccard score, IIDistII is mean±SD absolute distance, Dist is signed distance difference, and 95th is the 95th percentile of the absolute distance.

The results of the studies show that the values from the evaluated features in Segment 3DPrint were in good agreement with values from the reference method.

No adverse events, or complications, associated with the subject device were observed in the studies. Based on the clinical performance as documented in the performance studies, Segment 3DPrint was found to have a safety and effectiveness profile that is similar to the predicate device.

5.8 Conclusion

We conclude that the subject device Segment 3DPrint is as safe and effective as the predicate device. All identified hazards for Segment 3DPrint have been mitigated to acceptable levels, and the overall residual risk evaluation concluded that the residual risk of Segment 3DPrint is acceptable. The risks associated with the use of Segment 3DPrint are acceptable when weighed against the benefits for the patient. Segment 3DPrint performs in accordance with its intended use as well as the predicate device. ldentical to the predicate device, Segment 3DPrint does not in any way alter the imaging data in the analytical process. Segment 3DPrint provides assistance to a medically trained professional and all of the information is subject to his/her oversight, control, and clinical judgement. Medviso AB considers the features of Segment 3DPrint to be substantially equivalent to the subset of features in the predicate device Mimics Medical.

MEDVISO AB Griffelvägen 3 SE-224 67 Lund, Sweden tel: +46-76-1836442 www.medviso.com

Image /page/7/Picture/8 description: The image shows the word "MEDVISO" in a light blue sans-serif font. The word is surrounded by a pattern of small, light blue squares that form a semi-circular shape above and below the word. The squares are arranged in a grid-like pattern, and they appear to be pixelated. The overall effect is a modern and technological look.

Page 5-5 of 5

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).