K Number
K211953
Date Cleared
2022-05-25

(336 days)

Product Code
Regulation Number
878.4460
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Device Description

The proposed device, Disposable Sterilized Latex Surgical Gloves is a sterilized and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, per standard ASTM D3577-09(15), the rubber surgical gloves classification is: "Type 1-gloves compounded primarily from nature rubber latex." The gloves are powder-free and available in white in sizes 6, 6.5, 7, 7.5, 8, and 8.5. The proposed device is provided EO sterilized to achieve the sterility Assurance Level (SAL) of 106. The disposable sterilized latex surgical glove shelf life is 3 years.

AI/ML Overview

This document is a 510(k) summary for Disposable Sterilized Latex Surgical Gloves, demonstrating substantial equivalence to a predicate device (K171550). It focuses on non-clinical performance data and does not involve AI or clinical studies as typically defined in medical imaging or diagnostic device contexts. Therefore, many of the requested points are not applicable.

Here's an analysis of the provided information in the context of the requested questions:

Acceptance Criteria and Device Performance

This product is a medical device (Disposable Sterilized Latex Surgical Gloves), not an AI algorithm. The "acceptance criteria" and "reported device performance" are based on physical, chemical, and biological properties of the gloves, compared against recognized ASTM and ISO standards.

Table of Acceptance Criteria and Reported Device Performance:

Test Method (Standard)Test PurposeAcceptance CriteriaResult
Dimension
ASTM D3577-09(Reapproved 2015)Determine length of gloves265mm, minPass
ASTM D3577-09(Reapproved 2015)Determine width of glovesSize 6: 76 ±6 mm; Size 6.5: 83 ± 6 mm; Size 7: 89 ± 6 mm; Size 7.5: 95 ± 6 mm; Size 8: 102 ± 6 mm; Size 8.5: 108 ± 6 mmPass
ASTM D3577-09(Reapproved 2015)Determine thickness of glovesPalm: 0.10mm, min; Finger: 0.10mm, min; Cuff: 0.10mm, minPass
Physical Properties
ASTM D3577-09(Reapproved 2015)Determine physical properties (Tensile strength)Before Ageing Tensile Strength: 24Mpa Minimal for all sizes; After Ageing Tensile Strength: 18Mpa Minimal for all sizesPass
ASTM D3577-09(Reapproved 2015)Determine physical properties (Ultimate Elongation)Before Ageing Ultimate Elongation: 750% Min for all sizes; After Ageing Ultimate Elongation: 560% Min for all sizesPass
Watertight test
ASTM D5151-19Determine watertightness of glovesSample size: 200 pcs; Inspection level: GI; AQL 1.5; Criteria: ≤8 piecesPass
Residual powder
ASTM D6124-06 (Reapproved 2017)Determine residual powder in gloves2 mg per glove or lessPass
Protein content
ASTM D5712 - 15Determine extractable protein in gloves200 µg/dm² Max for all sizesPass
Biocompatibility
ISO 10993-10Evaluate potential of a test article to cause skin sensitizationUnder the conditions of the study, not a sensitizerPass
ISO 10993-10Evaluate potential skin irritation caused by test article contact with skin surface of rabbitsUnder the condition of study, not an irritantPass
ISO 10993-11:2017Evaluate whether leachables extracted from the test article would cause acute systemic toxicity following injection into miceUnder the conditions of study, the device extracts do not pose a systemic toxicity concernPass
USP 43-NF 38<151>Evaluate risks of febrile reaction in the rabbit to administration by injectionUnder the condition of study, not a pyrogenPass
USP 43 <85>Bacterial Endotoxin Test<20 EU/ pair of glovesPass
Sterilization Residuals
ISO 10993-7:2008Determine if Ethylene Oxide residues of test article are within requirements≤10ug/cm²Pass
ISO 10993-7:2008Determine if Ethylene Chlorohydrin residues of test article are within requirements≤5mg/cm²Pass

1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set:
    • For the Watertight test (ASTM D5151-19), the sample size was 200 pieces.
    • Other tests (e.g., dimensions, physical properties, residual powder, protein content, biocompatibility, sterilization residuals) refer to specific ASTM/ISO standards. While the standards themselves define sampling plans, the document does not explicitly state the exact sample size used for each of these specific tests beyond the watertight test.
  • Data Provenance: The document does not specify the country of origin of the data (other than the manufacturer being in China) or whether the studies were retrospective or prospective. Given the nature of mechanical/chemical testing of medical devices, these are typically prospective laboratory tests conducted to specific standard protocols.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable. This is a medical device (surgical gloves) undergoing non-clinical performance and biocompatibility testing, not an AI algorithm requiring expert ground truthing of images or clinical data. The "ground truth" here is defined by physical, chemical, and biological measurement standards (ASTM, ISO, USP).

3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. As above, this is non-clinical device testing, not a study involving human interpretation or consensus. The "adjudication" is based on meeting pre-defined numerical or qualitative criteria from recognized standards.

4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This document pertains to the clearance of a physical medical device (surgical gloves), not an AI algorithm. Therefore, no MRMC study was conducted.

5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not Applicable. This document pertains to the clearance of a physical medical device (surgical gloves), not an AI algorithm.

6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the non-clinical tests, the "ground truth" is defined by the established criteria within the cited ASTM, ISO, and USP standards. For example:
    • Dimensions are compared to specified minimums/ranges.
    • Tensile strength is compared to specified minimal MPa.
    • Watertightness is determined by meeting an AQL (Acceptance Quality Limit) of 1.5, meaning a certain percentage of defects is acceptable.
    • Biocompatibility (sensitization, irritation, systemic toxicity, pyrogenicity, endotoxin) is determined by standardized biological test methods with defined pass/fail criteria (e.g., "not a sensitizer," "not an irritant," "no systemic toxicity concern," "<20 EU/pair of gloves").
    • Protein and residual chemical content are compared to specified maximum limits.

7. The sample size for the training set

  • Not Applicable. This is not an AI algorithm; there is no "training set."

8. How the ground truth for the training set was established

  • Not Applicable. As there is no training set, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

May 25, 2022

Jiangxi Kemei Medical Apparatus & Instruments Group Co., Ltd % Alice Huang RA Manager Shanghai Mind-Link Business Consulting Co., Ltd. Room A08, Floor 14th, No 699, Jiaozhou Road, Jingan District Shanghai. 200040 China

Re: K211953

Trade/Device Name: Disposable Sterilized Latex Surgical Gloves Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO Dated: April 15, 2022 Received: April 25, 2022

Dear Alice Huang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K211953

Device Name Disposable Sterilized Latex Surgical Gloves

Indications for Use (Describe)

The Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

K211953

I. SUBMITTER:

Jiangxi Kemei Medical Apparatus & Instruments Group Co., Ltd. No. 316 National road, Xinju, Changshan Town, Jinxian County, Nanchang City, 331724, Jiangxi Province, China. Contact Person: Jenny Huang

TITLE: International Trade Supervisor

TEL. : +86 17702094798

Email: sales@jxkmjt.com

Submission Correspondent: Alice Huang Email: alice.huang(@mind-link.net Tel:+86 15618536177 Shanghai Mind-link Business Consulting Co., Ltd. Room A08, Floor 14th, No 699, Jiaozhou Road, Jingan District, Shanghai

Summary prepared: 05/23/2022

II. DEVICE 510(k) Number: K211953 Name of Device: Disposable Sterilized Latex Surgical Gloves Regulation Number: 21 CFR878.4460 Common Name: Surgeon's Gloves Classification Name: Non-powdered surgeon's glove Regulatory Class: I Product Code: KGO

III. PREDICATE DEVICE

510(k) Number: K171550 Product Name: Sterile Latex Surgical Gloves, Powder Free Manufacture: Sanrea Healthcare Pvt Ltd

{4}------------------------------------------------

IV. DEVICE DESCRIPTION

The proposed device, Disposable Sterilized Latex Surgical Gloves is a sterilized and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination.

The proposed device is made of natural rubber latex, per standard ASTM D3577-09(15), the rubber surgical gloves classification is:

"Type 1-gloves compounded primarily from nature rubber latex." The gloves are powder-free and available in white in sizes 6, 6.5, 7, 7.5, 8, and 8.5.

The proposed device is provided EO sterilized to achieve the sterility Assurance Level (SAL) of 106. The disposable sterilized latex surgical glove shelf life is 3 years.

V. INDICATIONS FOR USE

The Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Disposable Sterilized Latex Surgical Gloves are compared with the predicate device (Sterile Latex Surgical Gloves, Powder Free, K171550). The results are shown below in the Technological Characteristics Comparison Table:

DEVICESubject DevicePrimary Predicate Device(K171550)Comparison
Indications For UseThe Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.The Sterile Latex Surgical Gloves, Powder free, is a disposable device made of natural rubber, intended for medical purposes that is worn by operating room personnel to protect a surgical wound from contamination.Same
Intended UseThe Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.Intended for medical purposes that is worn by operating room personnel to protect a surgical wound from contamination.Same
ClassificationProduct CodeKGOKGOSame
Regulation No.21 CFR 878.446021 CFR 878.4460Same
ClassIISame
Powdered orPowdered freePowdered freePowdered freeSame
MaterialNatural Rubber LatexNatural Rubber LatexSame
DimensionsLength:Size 6: 265mm, minSize 6.5: 265mm, minSize 7: 265mm, minSize 7.5: 265mm, minSize 8: 265mm, minSize 8.5: 265mm, minWidth:Size 6: 76 ±6 mmSize 6.5: 83 ± 6 mmSize 7: 89 ± 6 mmSize 7.5: 95 ± 6 mmSize 8: 102 ± 6 mmSize 8.5: 108 ± 6 mmPalm Thickness: 0.10mm, minFinger Thickness: 0.10mm, minCuff Thickness: 0.10mm, minsize 7.5 :Length : 300mmsize 7.5 :Width : 95mmPalm Thickness: 0.19mmFinger Thickness: 0.22mmCuff Thickness: 0.14mmSimilarMeetingrequirementof ASTMD 3577
BiocompatibilityUnder the conditions of the study, not a sensitizer.Under the conditions of the study, not an irritant.Under the conditions of the study, no evidence of acute systemic toxicity.Non-sensitizingNon-irritatingSimilar
Tensile strengthBefore Aging 24MPa, minAfter Aging 18MPa, minBefore Aging 31.22MPaAfter Aging 24.43MPaSimilarMeeting
UltimateElongationBefore Aging 750%, minAfter Aging 560%,minBefore Aging 843.09%After Aging 762.28%requirementof ASTMD 3577
Freedom fromMeets ASTM D5151-06(2015)Meets ASTM D 5151 -06Same
HolesAQL 1.5(2011)
Protein ContentMeets ASTM D5712-2015Meets ASTM D5712Same
Powdered residueMeets ASTM D6124-06(Reapproved 2017)Meets ASTM D 6124-06(2011)Same
SterilityEOEOSame
EtO and ECH residualsMeets ISO 10993-7:2008Meets ISO 10993-7:2008Same
Shelf life3 Years3 YearsSame
Single UseYESYESSame
Label and labelingMeet FDA's RequirementMeet FDA's RequirementSame

{5}------------------------------------------------

{6}------------------------------------------------

VII. SUMMARY OF NON-CLINICAL TESTING

The gloves have the same technological characteristics compared to ASTM or equivalent standards as given below:

  • ASTM D5151-06(Reapproved 2015) Standard Test Method for Detection of Holes in Medical Gloves
  • ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
  • ASTM D3577-09 (Reapproved 2015) Standard Specification for Rubber Surgical Gloves
  • . ASTM D5712-15 Standard Test Method for Analysis of Aqueous Extractable Protein in Natural Rubber in Latex, Natural Rubber, and Elastomeric Products Using the Modified Lowry Method
  • ISO 10993-1 : 2009 Biological evaluation of medical devices - Part1: Evaluation and testing within a risk management process
  • ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
  • USP 43-NF 38<151> Pyrogen Test
  • USP 43<85> Bacterial Endotoxin Test
  • ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials

{7}------------------------------------------------

  • ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
  • . ASTM D7160 Standard Practice for Determination of Expiration Dating for medical Gloves
TestMethodStandardTest PurposeCriteriaResult
DimensionASTM D3577-09(Reapproved2015) Standard Specification forRubber Surgical GlovesTo determine the lengthof the gloves265mm, minPass
ASTM D3577-09(Reapproved2015)Standard Specification forRubber Surgical GlovesTo determine the widthof the glovesSize 6: 76 ±6 mmSize 6.5: 83 ± 6 mmSize 7: 89 ± 6 mmSize 7.5: 95 ± 6 mmSize 8: 102 ± 6 mmSize 8.5: 108 ± 6 mmPass
ASTM D3577-09(Reapproved2015) Standard Specification forRubber Surgical GlovesTo determine thethickness of the glovesPalm: 0.10mm, minFinger: 0.10mm, minCuff: 0.10mm, minPass
PhysicalPropertiesASTM D3577-09(Reapproved2015) Standard Specification forRubber Surgical GlovesTo Determine thephysical propertiesTensile strengthBefore AgeingTensile Strength24Mpa Minimal for allsizesAfter AgeingTensile Strength18Mpa Minimal for allsizesPass
ASTM D3577-09(Reapproved2015) Standard Specification forRubber Surgical GlovesTo Determine thephysical propertiesUltimate ElongationBefore AgeingUltimate Elongation750% Min for all sizesAfter AgeingUltimate Elongation560% Min for all sizesPass
WatertighttestASTM D5151-19 Standard TestMethod for Detection of Holes inMedical GlovesTo determine thewatertightness of the testglovesSample size: 200 pcsInspection level : GIAQL 1.5Criteria:≤8 piecesPass
ResidualpowderASTM D6124-06 (Reapproved2017) Standard Test Method forTo determine theresidual powder in the2 mg per glove or lessPass
Residual Powder on MedicalGlovesgloves
ProteincontentASTM D5712 - 15, Standard TestMethod for Analysis of AqueousExtractable Protein in NaturalRubber.To determine theextractable protein in thegloves.$200 \mu g/dm^2$ Max forall sizesPass
SkinSensitizationISO 10993-10Biological Evaluation ofMedical Devices Test forIrritation and SkinSensitization. Test donefor irritationThe test was designed toevaluate the potential ofa test article to causeskin sensitization.Under the conditionsof the study not asensitizerPass
IrritationISO 10993-10Biological Evaluation ofMedical Devices Test forIrritation and SkinSensitization. Test donefor irritationTo evaluate the potentialskin irritation caused bytest article contact withthe skin surface ofrabbits.Under the condition ofstudy not an irritantPass
SystemictoxicityISO 10993-11:2017 biologicalevaluation of medical devices - part11, tests for systemic toxicityThe test article wasevaluated to determinewhether leachablesextracted from the testarticle would cause acutesystemic toxicityfollowing injection intomice.Under the conditionsof study the deviceextracts do not pose asystemic toxicityconcernPass
PyrogenUSP 43-NF 38<151> Pyrogen TestThe test article wasevaluated for the risks offebrile reaction in therabbit to theadministration byinjection.Under the condition ofstudy not a pyrogenPass
BacterialEndotoxinUSP 43 <85> BacterialEndotoxinTestBacterial EndotoxinTest<20 EU/ pair of glovesPass
EO ResidueISO 10993-7:2008Determine if theEthylene Oxide residuesof test article is withinthe requirements.≤10ug/cm²Pass
ECHResidueISO 10993-7:2008Determine if theEthylene Chlorohydrin≤5mg/cm²Pass
residues of test article iswithin the requirements.

{8}------------------------------------------------

{9}------------------------------------------------

VIII. SUMMARY OF CLINICAL TESTING No clinical testing is included in this submission.

IX. CONCLUSION

The conclusions drawn from the non-clinical performance data demonstrate that the subject device, Disposable Sterilized Latex Surgical Glove, is as safe, as effective, and performs as well as or better than the legally marketed predicate device K171550.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).