(170 days)
PCI Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous coronary intervention (PCI) and percutaneous transluminal angioplasty (PTA), including use in crossing de novo coronary chronic total occlusions (CTO).
The PCI Guide Wires are not to be used in the neurovasculature.
The ASAHI PCI Guide Wire ASAHI CONFIANZA PRO 8-20 (Hereafter "ASAHI CONFIANZA PRO 8-20") is a steerable guide wire with a maximum diameter of 0.014 inches (0.36mm) and available in 190cm, 235cm and 300cm length. The extension wire is connected to the end of the guide wire outside the body for 190cm and 235cm wire. The guide wire is constructed from a stainless-steel core wire with a platinum-nickel coil. The coil is soldered to the core wire. The coil has radiopacity to achieve visibility and can be made to bend easily with the vessel curve. A hydrophilic coating is applied to the distal portion of the guide wire.
Here's an analysis of the provided text regarding the acceptance criteria and study for the device:
Device: ASAHI PCI Guide Wire ASAHI CONFIANZA PRO 8-20
The information provided describes a guide wire, not an AI/ML powered device, so several of the requested sections (MRMC study, standalone AI performance, ground truth for training set, training sample size) are not applicable. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical bench testing.
1. Table of Acceptance Criteria and Reported Device Performance
The document states that the ASAHI CONFIANZA PRO 8-20 met all acceptance criteria. However, it does not explicitly list the specific numerical acceptance criteria for each test. It only lists the types of tests performed. Without the specific thresholds for "Tensile Strength," "Torque Strength," etc., a detailed table with specific acceptance criteria is not possible from the provided text.
Here's a table based on the types of tests and the general conclusion:
| Test Type | Acceptance Criteria (Not explicitly detailed in source) | Reported Device Performance (Summary) |
|---|---|---|
| Tensile Strength | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Torque Strength | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Torqueability | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Tip Flexibility | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Coating Adhesion / Integrity | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Catheter Compatibility | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Visual Inspection | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Corrosion Resistance | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Kink Resistance | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Radiodetectability | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Dimensional Verification | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
| Coating Integrity/Particulate | (Not specified) | Met all acceptance criteria / Performed similarly to predicate |
Note: The document explicitly states: "The in vitro bench tests demonstrated that the ASAHI CONFIANZA PRO 8-20 met all acceptance criteria and performed similarly to the predicate devices." This is the highest level of detail provided regarding performance.
2. Sample Size Used for the Test Set and Data Provenance
The document only states that "in vitro bench tests" were conducted. It does not specify:
- The exact sample size (e.g., number of guidewires tested for each parameter).
- The data provenance (e.g., country of origin for the data).
- Whether the study was retrospective or prospective, although bench testing is typically prospective by nature.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This section is not applicable as the study described is non-clinical bench testing, not a clinical study involving experts establishing ground truth for diagnostic or prognostic purposes. The "ground truth" here is determined by objective physical measurements against defined performance specifications.
4. Adjudication Method for the Test Set
This section is not applicable for non-clinical bench testing. Adjudication methods (like 2+1 or 3+1) are used to resolve discrepancies among multiple human readers or evaluators in clinical studies. For bench testing, results are typically objective measurements.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is not applicable. The device is a physical medical device (guidewire), not an AI-powered diagnostic or therapeutic tool for which an MRMC study would be relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This section is not applicable. The device is a physical medical device (guidewire), not an algorithm or AI system.
7. The Type of Ground Truth Used
For this physical device, the "ground truth" for the non-clinical testing is based on pre-defined engineering specifications and performance benchmarks established for similar medical devices (the predicate device and industry standards/guidance documents). The tests measure physical and mechanical properties against these objective standards.
8. The Sample Size for the Training Set
This section is not applicable. As a physical guidewire, there is no "training set" in the context of machine learning. The device design and manufacturing processes are developed through engineering and materials science principles, not machine learning training.
9. How the Ground Truth for the Training Set Was Established
This section is not applicable for the reasons mentioned in point 8.
{0}------------------------------------------------
December 8, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Asahi Intecc Co., Ltd. Cynthia Valenzuela Director, Regulatory Affair 3002 Dow Avenue, Suite 212 Tustin, California 92780
Re: K211898
Trade/Device Name: ASAHI PCI Guide Wire ASAHI CONFIANZA PRO 8-20 Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: November 11, 2021 Received: November 23, 2021
Dear Cynthia Valenzuela:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
ASAHI PCI Guidewire ASAHI CONFIANZA PRO 8-20
Indications for Use (Describe)
PCI Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous coronary intervention (PCI) and percutaneous transluminal angioplasty (PTA), including use in crossing de novo coronary chronic total occlusions (CTO).
The PCI Guide Wires are not to be used in the neurovasculature.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary
[as required by 21CFR § 807.92(c)]
ASAHI INTECC CO.,LTD.
Global Headquarters and R&D Center 3-100 Akatsuki-cho, Seto-shi, Aichi 489-0071 Japan TEL : +81-561-48-5551 FAX : +81-561-48-5552 http://www.asahi-intecc.co.jp/
ASAHI PCI Guide Wire ASAHI CONFIANZA PRO 8-20
510(K)
| DATE PREPARED: | 27OCT2021 |
|---|---|
| APPLICANT: | ASAHI INTECC CO., LTD. |
| 3-100 Akatsuki-cho, Seto | |
| Aichi 489-0071, Japan | |
| PRIMARY CONTACT: | Mrs. Cynthia Valenzuela |
| Director, Regulatory Affairs | |
| ASAHI INTECC USA, INC. | |
| 3002 Dow Avenue, Suite 212 | |
| Tustin, California 92780 | |
| Phone: (714) 442 0575 | |
| Fax: (949) 377 3255 | |
| Email: cynthiav@asahi-intecc-us.com | |
| ALTERNATE CONTACT: | Mr. Hiroshi Obara |
| Manager, Regulatory Affairs | |
| ASAHI INTECC CO., LTD. | |
| 3-100 Akatsuki-cho | |
| Seto, Aichi, Japan 489-0071 | |
| Email: hiroshi.obara@asahi-intecc.com | |
| TRADE NAME: | ASAHI PCI Guide Wire ASAHI CONFIANZA PRO 8-20 |
| DEVICE CLASSIFICATION: | Class II, 21CFR § 870.1330 |
| CLASSIFICATION NAME: | Catheter Guide Wire |
| PRODUCT CODE: | DQX - Wire, Guide, Catheter |
| PREDICATE DEVICE(S): | ASAHI PTCA Guide Wire ASAHI CONFIANZA PRO 12 (K171933) |
| REFERENCE DEVICE(S): | ASAHI PTCA Guide Wire ASAHI Gaia Next (K192599) |
| ASAHI Neurovascular Guide Wire ASAHI CHIKAI black (K141751) |
Intended Use/Indications for Use:
PCI Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous coronary intervention (PCI) and percutaneous transluminal angjoplasty (PTA), including use in crossing or assisting in crossing de novo coronary chronic total occlusions (CTO).
The PCI Guide Wires are not to be used in the neurovasculature.
{4}------------------------------------------------
Description:
The ASAHI PCI Guide Wire ASAHI CONFIANZA PRO 8-20 (Hereafter "ASAHI CONFIANZA PRO 8-20") is a steerable guide wire with a maximum diameter of 0.014 inches (0.36mm) and available in 190cm, 235cm and 300cm length. The extension wire is connected to the end of the guide wire outside the body for 190cm and 235cm wire. The guide wire is constructed from a stainless-steel core wire with a platinum-nickel coil. The coil is soldered to the core wire. The coil has radiopacity to achieve visibility and can be made to bend easily with the vessel curve. A hydrophilic coating is applied to the distal portion of the guide wire.
Comparison with Predicate Device:
Predicate Device:
ASAHI PTCA Guide Wire ASAHI CONFIANZA PRO 12 (K171933)
Comparisons of the ASAHI CONFIANZA PRO 8-20 and predicate devices show that the technological characteristics of the ASAHI CONFIANZA PRO 8-20 such as components, design, materials, sterilization method, shelf life and operating principle are identical or similar to the currently marketed predicate device. A tabular comparison of the specific technological characteristics between the predicate device and subject device is provided below.
| Name of Device | ASAHI PCI Guide WireASAHI CONFIANZA PRO 8-20 | ASAHI PTCA Guide WireASAHI CONFIANZA PRO 12 |
|---|---|---|
| 510(K) | -- | K171933 |
| Intended Use andIndications | PCI Guide Wires are intended tofacilitate the placement of balloondilatation catheters duringpercutaneous coronaryintervention (PCI) andpercutaneous transluminalangioplasty (PTA), including use incrossing or assisting in crossing denovo coronary chronic totalocclusions (CTO).The PCI Guide Wires are not to beused in the neurovasculature. | ASAHI PTCA Guide Wires areintended to facilitate the placementof balloon dilatation cathetersduring percutaneous transluminalcoronary angioplasty (PTCA) andpercutaneous transluminalangioplasty (PTA), including use incrossing or assisting in crossing denovo coronary chronic totalocclusions (CTO).The ASAHI PTCA Guide Wires arenot to be used in theneurovasculature. |
| Sterilization | Ethylene Oxide to SAL 10-6 | |
| Shelf Life | 3 Years | 3 Years |
| Target Body Location | Coronary, Peripheral | |
| Outer Coil Material | Platinum Nickel | Platinum Nickel |
| Core Wire Material | Stainless Steel | Stainless Steel |
| Distal Tip Shape | Straight, Pre-shape | Straight |
| Overall Length | 190cm, 235cm, 300cm | 180cm, 300cm |
| Distal Section Coating length | 17cm | 20cm |
| Outside Diameter of Wire | Distal 0.20mm/Proximal 0.36mm | Distal 0.23mm/Proximal 0.36mm |
{5}------------------------------------------------
Non Clinical Testing / Performance Data:
The substantial equivalence of the ASAHI CONFIANZA PRO 8-20 line extension was evaluated in bench testing that followed the recommendations in the FDA guidance document; Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling, 15JUN2018.
- Tensile Strength ●
- Torque Strength
- . Torqueability
- Tip Flexibility
- Coating Adhesion / Integrity ●
- Catheter Compatibility ●
- . Visual Inspection
- Corrosion Resistance ●
- Kink Resistance
- Radio detectability ●
- Dimensional Verification ●
- Coating Integrity/Particulate .
The in vitro bench tests demonstrated that the ASAHI CONFIANZA PRO 8-20 met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the device functions as intended and has a safety and effectiveness profile that is similar to the predicate device.
BIOCOMPATIBILITY:
The ASAHI CONFIANZA PRO 8-20 was compared to the predicate devices. Based on similarities of the materials used in the subject device to its predicates / reference devices, the biocompatibility of the ASAHI CONFIANZA PRO 8-20 was verified to be the same as those of the predicates / reference devices.
CONCLUSION:
The ASAHI CONFIANZA PRO 8-20 has similar intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate and reference device. Performance data demonstrates that the device functions as intended.
Therefore, the ASAHI CONFIANZA PRO 8-20 is substantially equivalent to the predicate device.
§ 870.1330 Catheter guide wire.
(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.