(58 days)
Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
The proposed device is Powder Free Nitrile Examination Gloves. The proposed device is blue. The proposed device is non-sterile.
This document is a 510(k) Premarket Notification summary for Powder Free Nitrile Patient Examination Gloves, Blue Color. It does not describe an AI medical device or a study involving human readers. Therefore, I cannot provide details on acceptance criteria or an AI-related study as described in your request.
The provided text details the submission of a medical device (nitrile examination gloves) for regulatory clearance. The core of the submission, and the content of the "Acceptance Criteria" table, relates to physical and chemical properties of the gloves, and biocompatibility, as tested against established ASTM and ISO standards.
Here's an analysis of the existing content in relation to your request, highlighting what's present and what's missing:
1. A table of acceptance criteria and the reported device performance
The provided text contains a comprehensive table on page 4 titled "Features & Description" which functions as the acceptance criteria and reported device performance.
| Features & Description | Acceptance Criteria (from Subject Device) | Reported Performance (from Subject Device) |
|---|---|---|
| Dimension - Length | ≥230mm min (from predicate guidance) | 232 mm min for all sizes |
| Dimension - Width | Small 75-90 mm, Medium 87-102 mm, Large 107-119mm, X large 114-128 mm | Measured values for each size fall within these ranges (implied by "Similar" and comparison to predicate values) |
| Dimension - Thickness | Fingertip ≥0.08mm min, Palm ≥0.08mm min | Fingertip 0.08mm min, Palm 0.08mm min (reported as "Similar" to predicate which implies meeting criteria) |
| Physical Properties - Tensile strength (Before & After aging) | ≥14MPa | Before Aging: 19-25 MPa; After Aging: 17-23 MPa |
| Physical Properties - Elongation (Before aging) | ≥500% | 560-610 % |
| Physical Properties - Elongation (After aging) | ≥400% | 460-570 % |
| Freedom from Pinholes (Water leakage test) | Inspection Level I, AQL2.5, Accept/Reject criteria of 10/11. 5 noncompliance is allowed. | Meets criteria (implied by "Similar" and no reported failure) |
| Residual Powder | <2mg/glove | Mean: 0.1mg/pcs |
| Biocompatibility - Primary Skin Irritation | Not a primary skin irritant | Passes |
| Biocompatibility - Dermal Sensitization | Not a skin sensitizer | Passes |
| Biocompatibility - Cytotoxicity | Non-cytotoxicity to L-929 cells | Pass |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: The sample sizes are generally specified by the ASTM and ISO standards referenced. For instance, "Freedom from Pinholes" specifies "Inspection Level I, AQL2.5, and Accept/Reject criteria of 10/11," which implies a specific sampling plan based on lot size. "Residual Powder" states "Checked on 5pcs sub-samples (N=5)."
- Data Provenance: The document does not explicitly state the country of origin of the data itself, but the applicant company (Anhui Powerguard Technology Co., Ltd.) is from China. The tests are "Non-clinical tests" conducted to verify compliance with design specifications. It is a prospective study as the tests are conducted on the new device for regulatory submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is not applicable as the device is a physical product (gloves) and not an AI or diagnostic device that requires expert interpretation for ground truth. The "ground truth" is established by direct physical, chemical, and biological testing according to validated standard protocols.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable for a physical product; no human adjudication is involved in the performance testing of gloves.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a clearance for a physical medical device (gloves), not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. There is no algorithm in this submission.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
The ground truth is established by objective, standardized laboratory tests (e.g., tensile strength testing, water leakage test, chemical analysis for residual powder, in vitro and in vivo biocompatibility assays) as defined by ASTM and ISO standards.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the ground truth for the training set was established
Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
August 13, 2021
Anhui Powerguard Technology Co., Ltd. % Chu Xiaoan Official Correspondent Beijing Easy-Link Company Rm. F302 Bldg., 41, Jing Cheng Ya Ju, Courtyard 6 of Southern Dou Ge Zhuang, Chaoyang District Beijing, 100121 China
Re: K211860
Trade/Device Name: Powder Free Nitrile Patient Examination Gloves, Blue Color Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: April 16, 2021 Received: June 16, 2021
Dear Chu Xiaoan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K211860
Device Name
Powder Free Nitrile Patient Examination Gloves, Blue Color
Indications for Use (Describe)
Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary
"The assigned 510(k)number is: K211860
This summary of 510(k) is being submitted in accordance with 21 CFR 807.
Premarket Notification [510(k)] Summary
1.0 Submitter:
| Submitter's name : | Anhui Powerguard Technology Co., Ltd. |
|---|---|
| -------------------- | --------------------------------------- |
Submitter's address : Lingbi Economic Development Zone (North), Suzhou City, Anhui Province, 234200,China
Name of contact person: Mr. Nick Dai
Submitter information 0086-557-6868999
Date of preparation : 2021-08-09
2.0 Name of the Device
Proprietary/,Trade name:Powder Free Nitrile Patient Examination Gloves, Blue Color
| Common Name: | Patient Examination gloves |
|---|---|
| Classification Name: | Non-powdered Patient examination glove |
| Device Classification: | I |
| Regulation Number: | 21 CFR 880.6250 |
| Panel: | General Hospital |
| Product Code: | LZA |
3.0 Predicate device
| Device Name: | Powder FreeGlove, Blue Color | Nitrile Patient Examination |
|---|---|---|
| Company name: | Tangshan Zhonghong Pulin Plastic Co., Ltd. | |
| 510(K) Number: | K120970 |
4.0 Device Description:
The proposed device is Powder Free Nitrile Examination Gloves. The proposed device is blue. The proposed device is non-sterile.
5.0 Indications for Use Statement:
Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
{4}------------------------------------------------
| Features & Description | Predicate Device (K120970) | Subject Device (K211860) | Comparison | Test Methodology | Purpose | Acceptance Criteria | Results | |
|---|---|---|---|---|---|---|---|---|
| Product name | Powder Free Nitrile Patient Exa mination Glove, Blue Color | Powder Free Nitrile Patient Examination Gloves, Blue Color | Same | ASTM D 6319-06(Reapproved2015). | Dimension | Length | ≥230mm | |
| Regulation Number | 21CFR880.6250 | 21CFR880.6250 | Same | Width | Small | 75-90 mm | ||
| Product Code | LZA | LZA | Same | Medium | 87-102 mm | |||
| Color | Blue | Blue | Same | Large | 107-119mm | |||
| Size | Small/ Medium/ Large/X large | Small/ Medium/ Large/X large | Same | X large | 114-128 mm | |||
| Indications for Use | Powder Free Nitrile Patient Examination Glove, Blue Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Powder Free Nitrile Patient Examination Gloves, Blue Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Same | Thickness | Fingertip | ≥0.08mm | ||
| Device Description and Specifications | Meets ASTM D6319-10 | Meets ASTM D6319-10 (Reapproved 2015) | Same | Palm | ≥0.08mm | |||
| Dimensions --LengthILS-2AQL4.0 | Meets ASTM D6319-10>=230mm min | 232 mm min for all sizes | Similar | ASTM D 6319-06(Reapproved2015). | PhysicalProperties | Tensile strength(Before & Afteraging) | ≥14MPa | 17-25 |
| Dimensions -- WidthIL S-2AQL4.0 | Meets ASTM D6319-10Small 70-90 mmMedium 85-105mmLarge 100-120mmX large 110-130 mm | Small 75-90 mmMedium 87-102 mmLarge 107-119mmX large 114-128 mm | Similar | Before agingElongation | ≥500% | 560-610 | ||
| Dimensions --ThicknessIL S-2AQL4.0 | Meets ASTM D6319-10Finger 0.05mm min.Palm 0.05mm min. | Thickness (mm) min.Finger 0.08Palm 0.08 | Similar | After agingElongation | ≥400% | 460-570 | ||
| Physical PropertiesIL S-2AQL4.0 | Meets ASTMDD6319-10 | Aging Before After | Similar | |||||
| Before aging/after agingTensile Strength≥ 14MPaBefore agingElongation≥500%After agingElongation ≥400% | Elongation (% 560-610 460-570Tensile Strength (MPa)19-25 17-23 | ● 21 CFR 800.20● ASTM D 6319-06(Reapproved2015).● ASTM D5151-19 | Freedom frompinholes | Water leakage test:Inspection LevelI, AQL2.5, andAccept/Reject criteria of 10/11. | 5noncomplianceis allowed. | |||
| Freedom from PinholesInspection Level IAQL2.5 | Meets• 21 CFR 800.20• ASTM D6319-10 | 1) Inspection Levell AQL2.5,and Accept/Reject criteria of 10/112) Water leakage test:5 noncomplianceis allowed. | Similar | ● ASTM D6319-10(Reapproved2015) | PowderResidual | Meets<2mg/glove | Mean:0.1mg/pcs | |
| Residual Powder | Meets ASTM D 6124-06 (Reaffirmation 2011) | 1) Checked on 5pcs sub-samples (N=5). | Similar | ● ASTM D6124-06 (Reapproved2017), | Pass | |||
| below 2mg of residualpowder | Mean: 0.1mg/pcs | Primary SkinIrritation in rabbitsISO 10993-10:2010-08-01 | Biocompatibility | Under the conditions of the study,the subject device is not aprimary skin irritant. | Passes | |||
| Materials used tofabricate the devices | Nitrile | Nitrile | Same | Dermalsensitization in theguinea pig ISO10993-10: 2010-08-01 | Under the conditions of the study,the subject device is not a skinsensitizer. | Passes | ||
| Single Patient Use | Single Patient Use | Single Patient Use | Same | The test article wasadded to L929cells measured byMTT assay ISO10993-5:2009 | Under the conditions of thisstudy, the test article was non-cytotoxicity to L-929 cells. | Pass | ||
| Biocompatibility | Under the conditions ofthis study, the testarticle was a non-irritant or non-sensitizerSKIN IRRITATIONDERMAL andSENSITIZATIONSTUDIES Meets ISO10993-10:2002/Amd.1:2006 | Under the conditions of thisstudy, the test article was anon- irritant or non-sensitizer and non-potentialtoxicity to L-929 cells.SKIN IRRITATIONDERMAL andSENSITIZATION STUDIESMeets ISO 10993-10: ThirdEdition 2010-08-01.Cytotoxicity study meetsISO 10993-5 Third edition2009-06-01 | Similar | |||||
| Labeling for thelegally marketeddevice to whichsubstantialequivalence isclaimed. | -Powder Free-Patient ExaminationGlove-Single Use Only-Manufactured For:-Lot-Blue color-Non sterile | -Powder Free-Patient Examination Glove-Single Use Only-Manufactured For:-Lot-Blue color-Non sterile | Same |
6.0 Technological Characteristic Comparison:
{5}------------------------------------------------
7.0 Discussion of Non-clinical and Performance Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6319-10(Reapproved 2015), Standard Specification for Nitrile Examination Gloves for Medical Application.
{6}------------------------------------------------
8.0 Discussion of Clinical and Performance Testing
Clinical testing is not needed for this device.
9.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.