K Number
K211726
Date Cleared
2021-11-05

(154 days)

Product Code
Regulation Number
892.1560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Intraoperative Abdominal, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

The typical examinations performed using the SC2000 Ultrasound System are:

Cardiac Imaging Applications and Analysis

The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating 2D (B), 3D, MMode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system also supports catheters which are intended for intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Vascular Imaging Applications and Analysis

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and deep veins and arteries in the arms and legs and abdomen; and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Superficial Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic capture of ultrasound images.

Intraoperative Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during intraoperative procedures.

Transcranial Imaging Applications

The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence of pathology. The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Device Description

The ACUSON SC2000 Diagnostic Ultrasound System is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display of thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to transmit and receive ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M Mode, Doppler Tissue Mode, Amplitude Doppler Mode, a combination of modes and Harmonic Imaging on a Display. The transducer and catheter-based transducers will follow Track 3 acoustic labeling (AIUM 1004, IEC 2007, AIUM/NEMA 2004a).

AI/ML Overview

The provided text, primarily a 510(k) summary for the ACUSON SC2000 Diagnostic Ultrasound System, does not contain information about the acceptance criteria and the study that proves the device meets the acceptance criteria for an AI/ML-driven medical device.

This document focuses on establishing substantial equivalence of a new version (VC11/v6.1) of an existing ultrasound system to its predicate device (VC10/v6.0) and an earlier clearance (K171766) for the syngo® TrueFusion software feature. The key modification described is an update to the syngo® TrueFusion software from version 1.0 to 2.0 to support "Live Image Overlay."

The document states:

  • "The SC2000 software update to VC11 (v6.1) does not involve any hardware modifications." (Page 11)
  • "The VC11A (v6.1) software update and the SC2000 System software has undergone component, integration and system level verification testing. The software verification and regression testing has been performed successfully to meet their previously determined acceptance criteria consistent with Siemens Product Lifecycle Procedure (PLP) and has passed." (Page 11)
  • "Since the ACUSON SC2000 VC11 (v6.1) Diagnostic Ultrasound System uses the identical technology and principles of use as the existing predicate devices SC2000, VC10 (v6.0) K202683, clinical studies were not required to support substantial equivalence. The SC2000 VC11A (v6.1) software was fully validated on phantoms by clinical personal with training in Ultrasound technology and has passed." (Page 12)
  • "In summary, based on the successful verification and validation testing to the software acceptance criteria, the update for the TrueFusion 2.0 feature does not introduce any new potential safety risks." (Page 12)

This indicates that the modifications were considered minor enough that extensive clinical studies to prove new acceptance criteria were not deemed necessary for the 510(k) clearance, as the device was shown to be substantially equivalent to an already cleared device. The "acceptance criteria" mentioned are likely related to internal software verification and validation, such as functional testing and regression testing, rather than performance metrics for an AI/ML model for clinical diagnosis.

Therefore, I cannot provide the requested information from this document. The document doesn't detail:

  1. A specific table of acceptance criteria and reported device performance for an AI/ML component.
  2. Sample sizes, data provenance, ground truth establishment, expert qualifications, or adjudication methods related to an AI/ML performance study.
  3. MRMC study details or standalone performance of an AI/ML algorithm.

The syngo® TrueFusion feature is described as integrating ultrasound with X-Ray Fluoroscopy for "Live Image Overlay" (page 6). While this involves data processing and fusion, the document does not present it as a new AI/ML diagnostic algorithm requiring a specific performance study with clinical endpoints or detailed AI validation metrics. It's more about system integration and data presentation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Siemens Medical Solutions USA, Inc. % Christine Dunbar Senior Regulatory Affairs Specialist 685 East Middlefield Road MOUNTAIN VIEW CA 94043

November 5, 2021

Re: K211726

Trade/Device Name: ACUSON SC2000 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, IYN, ITX, OBJ, LLZ Dated: September 29, 2021 Received: October 4, 2021

Dear Christine Dunbar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211726

Device Name

ACUSON SC2000 Diagnostic Ultrasound System

Indications for Use (Describe)

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Intraoperative Abdominal, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

The typical examinations performed using the SC2000 Ultrasound System are:

Cardiac Imaging Applications and Analysis

The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating 2D (B), 3D, MMode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system also supports catheters which are intended for intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Vascular Imaging Applications and Analysis

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and deep veins and arteries in the arms and legs and abdomen; and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Superficial Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

Intraoperative Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during intraoperative procedures.

Transcranial Imaging Applications

The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence of pathology. The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

SECTION 1. 2. 510(K) SUMMARY - K211726

Date:June 01, 2021 / September 28, 2021
Part 1. Sponsor:Siemens Medical Solutions USA, Inc.,Ultrasound Division685 East Middlefield RoadMountain View, California 94043
Contact Person:Christine DunbarTel: (925) 374-2045
Part 2. Device Name:ACUSON SC2000 Diagnostic Ultrasound System - 21 CFR 807.92(a)(1)
Common Name:Diagnostic Ultrasound System with Accessories - 21 CFR 807.92(a)(2)
Classification:Regulatory Class: IIReview Category: Tier IIClassification Panel: 90, Radiology
Ultrasonic Pulsed Echo Imaging System892.156090-IYO
Ultrasonic Pulsed Doppler Imaging System892.155090-IYN
Diagnostic Ultrasound Transducer892.157090-ITX
Diagnostic Intravascular Catheter892.120090-OBJ
Medical Image Management andProcessing System (Optional for SC2000Diagnostic US System)892.205090-LLZ
Manufacturing Site:22010 S. E. 51st Street, Issaquah, WA 98029, USA
Legal Manufacturer:Siemens Medical Solutions USA, Inc.22010 S. E. 51st StreetIssaquah, WA 98029, USA

Part 3. Legally Marketed Predicate Devices - 21 CFR 807.92(a)(3)

The ACUSON SC2000 Diagnostic Ultrasound System is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to the company's own products, the ACUSON SC2000, v6.0 (VC10) under K202683 on October 15, 2021, is the primary predicate device.

The feature under modification is the syngo® TrueFusion 1.0 originally cleared under K171766 on July 12, 2017. The syngo® TrueFusion 1.0 software feature has been subsequently cleared under successive SC2000 Diagnostic Ultrasound device clearances including the most recent under K202683 as an optional feature. Therefore, K171766 serves as a reference predicate since Siemens prefers to continue

{5}------------------------------------------------

to clear the syngo® TrueFusion software application on the SC2000 Diagnostic Ultrasound system device.

Part 4. Device Description - 21 CFR 807.92(a)(4)

The ACUSON SC2000 Diagnostic Ultrasound System is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display of thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to transmit and receive ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M Mode, Doppler Tissue Mode, Amplitude Doppler Mode, a combination of modes and Harmonic Imaging on a Display. The transducer and catheter-based transducers will follow Track 3 acoustic labeling (AIUM 1004, IEC 2007, AIUM/NEMA 2004a).

Part 5. Intended Use and Indications for Use Statements - 21 CFR 807.92(a)(5)

SC2000 Diagnostic Ultrasound System, VC10 (v6.0)

(The Indications for Use remains unchanged as cleared under K202683 (VC10)

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac. Neo-natal and Fetal Cardiac. Pediatic. Transesophageal. Adult Cephalic. Peripheral Vessel. Abdominal. Intraoperative Abdominal. Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are:

Cardiac Imaging Applications and Analysis

The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating 2D (B), 3D, MMode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic capture of ultrasound images. The system also supports catheters which are intended for intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients.

The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Vascular Imaging Applications and Analysis

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and deep veins and arteries in the arms and legs and abdomen; and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Superficial Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic capture of ultrasound images.

Intraoperative Imaging Applications

{6}------------------------------------------------

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Dopler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during intraoperative procedures.

Transcranial Imaging Applications

The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology.

The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Part 6. Summary of Technological Characteristics - 21 CFR 807.92(a)(6)

6.1 The ACUSON SC2000 VC11 (v6.1) Diagnostic Ultrasound System is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to the company's own products, the ACUSON SC2000 VC10 (v6.0) (K202683) with regard to both the intended use and technological characteristics. Both the subject ultrasound system and the predicate ultrasound system function in the same manner as similar diagnostic ultrasound systems and transducers including the cardiac catheters such as the ACUSON V and Volume ICE catheters and support for third party ICE catheters; the SoundStar™ 8F and 10F and the SoundStar™ eco 8F and eco 10F versions of the Diagnostic Ultrasound Catheters, which are developed, manufactured and distributed by Biosense Webster. Inc. under multiple clearances.

The syngo® TrueFusion 1.0 software feature that is currently cleared as an optional software application on the SC2000 Diagnostic Ultrasound system has remained essentially unchanged except for the minor modification to serial communications to support Live Image Overlay which is the combined imaging of the SC2000 Ultrasound system and Siemens-Healthineers Axios™ Fluoroscopy system imaging in realtime. The syngo® TrueFusion software is comprised of the Siemens ACUSON SC2000 Diagnostic Ultrasound System used in conjunction with Siemens X-Ray Fluoroscopy systems called ArtisTM.

This new Live Image Overlay feature will be described in design descriptions for the syngo® TrueFusion 2.0 software feature.

The foundation of the ACUSON SC2000 VC11 (v6.1) with syngo® TrueFusion 2.0 software is the SC2000 VC10 (v6.0) release with syngo® TrueFusion 1.0 software. All other imaging features and transducers integrated with the ACUSON SC2000 Diagnostic Ultrasound system hardware and the software developed for SC2000 platform remain unchanged.

The SC2000 (WP) workplace system software includes the same currently cleared software applications supports the viewing of the syngo® TrueFusion acquired images, however the WP does not acquire Ultrasound images and is not present in the surgical / interventional suite.

It is Siemens' opinion that the ACUSON SC2000 VC11 (v6.1) is substantially equivalent to the predicate device, the ACUSON SC2000 VC10 (v6.0) with regards to both intended use, indications for use and technological characteristics.

6.2 - List of Technological Characteristics and SE Comparison Table

{7}------------------------------------------------

Feature / CharacteristicACUSON SC2000 v6.1 IncludingTransducers, Catheters &syngo® TrueFusion 2.0This SubmissionACUSON SC2000 6.0K202683 Including TransducersAcuNav Volume ICE Catheter &3rd Party Catheters &syngo® TrueFusion 1.0 option
Indications for Use:
Indications for Use - DeviceUnchanged for VC11 (v6.1)
Fetal
Abdominal
Pediatric
Small Organ
Cardiac
Peripheral vessel
Musculo-skeletal(conventional)
Musculo-skeletal(superficial)
Frequencies Supported:(1.7MHz~10MHz)(1.7MHz~10MHz)
Modes:Unchanged for VC11 (v6.1)
B
M
PWD (Pulsed WaveDoppler)
CWD (Continuous WaveDoppler)
PW DTI (Doppler TissueImage)
Color Doppler
Color Power Doppler(CPD)
Combined (BMDC)
Transducers/ CathetersUnchanged for VC11 (v6.1)
Z6Ms - TEE Transducer
4Z1c - Phased Array Transducer
4Z1c - Phased Array Transducer –Imaging Preset update
10V4 - Phased Array Transducer
9L4 - Linear Array Transducer
9L4 - Linear Array Transducer –update imaging presat
4V1c - Phased Array Transducer
V5Ms - TEE Transducer
8V3 - Phased Array Transducer
V7M - TEE Transducer
6C1HD - Curved array transducer
CW2 - Continuous Wave Transducer
AcuNav 8F - Phased Array UltrasoundCatheter
Feature / CharacteristicACUSON SC2000 v6.1 IncludingTransducers, Catheters &syngo® TrueFusion 2.0This SubmissionACUSON SC2000 6.0K202683 Including TransducersAcuNav Volume ICE Catheter &3rd Party Catheters &syngo® TrueFusion 1.0 option
AcuNav 10F Intracardiac Transducer
AcuNav™ V ICE Catheter
AcuNav™ Volume ICECatheter (aka P6)
AcuNav™ Volume ICECatheter Imaging Presetupdate
Supports Third party USCatheters with Swift-Linkconnector:
SoundStar™ eco 8 Frenchcatheter¹
SoundStar™ eco 10 Frenchcatheter
SoundStar™ 8 Frenchcatheter²
SoundStar™ 10 Frenchcatheter
SwiftLink Connector
Features on SC2000 System:Unchanged for VC11 (v6.1)
Patient RegistrationFields
Change/Edit PatientInformation Active exam
Native™ tissue harmonicimaging (2D Brightnessmode)
TEQ™ ultrasoundtechnology
Volume ICE Package
Support for AcuNavVolume ICE Catheterconnected by SwiftLinkcable.
ICE Catheter Auto-reduce Mechanical Index(MI)
2D ICE Package
Feature / CharacteristicACUSON SC2000 v6.1 Including Transducers, Catheters & syngo® TrueFusion 2.0This SubmissionACUSON SC2000 6.0 K202683 Including Transducers AcuNav Volume ICE Catheter & 3rd Party Catheters & syngo® TrueFusion 1.0 option
True Volume Imaging Support, AcuNav Volume ICE Catheter
Volume Color Doppler
Fetal Echo Calculations
Fetal Imaging Presets
Cardiac Imaging physiological signal display
eSie Measure on TEE
Advanced SieClear™ spatial compounding
Clarify™ vascular enhancement technology
syngo® Velocity Vector Imaging (eSie VVI)
syngo® Velocity Vector Imaging (eSie VVI)Changed to be consistent with v5.1 implementation.
eSie Valves Advanced Measurement Package
eSie LVA: 5 Cardiac Cycles
eSie PISA
eSie Left Heart Measurement Package
Volume Right Ventricular Analysis (RVA)
Stress Echo Package
Rapid Stress Volume Stress Echo App.
Cardiac Measurements and Calculations
syngo® TrueFusion v1.0
syngo® TrueFusion v2.0Updated to support Live Imaging OverlayTrueFusion 1.0
syngo® TrueFusion v2.0Update communication protocol for True Fusion 2.0.TrueFusion 1.0
Septal Guide – Orthogonal Guidelines
VR Measurement Tools (Volume)
Reference LinesOne-click MPR
Feature / CharacteristicACUSON SC2000 v6.1 IncludingTransducers, Catheters &syngo® TrueFusion 2.0ACUSON SC2000 6.0K202683 Including TransducersAcuNav Volume ICE Catheter &3rd Party Catheters &syngo® TrueFusion 1.0 option
This Submission
• Volume Reference LineProjections• eSie Slice / eSie Lines
• One-click MPR A/B Alignon Volume Review (VR)
• Trace erase, back upbehavior with Trackball:B-Mode
• Single V
• Contrast Agent ImagingLVO
• Edit patient data onactive exam.
• Zoom & Pan
• eSieScan(Protocols)
• InFocus CoherentTechnology
• Circle Tool
• DICOM
• DICOM SR(Structure Reports)
• DICOM GSDF (Grayscale Standard DisplayFunction).
• DICOM Tags (Teamplay)• DICOM support theexport of DICOM tags forSystem Data Collection
MS Windows 10 OSWindows 10 plus updatesWindows 10
Wireless - enabled
WiFi EAP-TLSauthentication protocol
Cybersecurity Features
Security - User Accounts
User Accounts \local userauthentication
Table Side Remote ControlJoy-Stick
Service Save logs
Service Save logsinclude save log locationService Logs available
Feature / CharacteristicACUSON SC2000 v6.1 IncludingTransducers, Catheters &syngo® TrueFusion 2.0This SubmissionACUSON SC2000 6.0K202683 Including TransducersAcuNav Volume ICE Catheter &3rd Party Catheters &syngo® TrueFusion 1.0 option
Security Hot-fix Installationinstall a hotfix update via RUH(remote update handling) or USBdrive.install a hotfix update via RUH(remote update handling) or USBdrive.
Study Back-up and Restore
Monitor: 21" FPD
Output Display Standard(Track 3)
Patient Contact MaterialsTested to ISO 10993-1Tested to ISO 10993-1
UL 60601-1 Certified

{8}------------------------------------------------

¹ SoundStar and SoundStar eco catheters are products of Biosense Webster, Inc. (A Johnson company) holds the 510(k) clearances for these devices.

² SoundStar and SoundStar eco catheters are products of Biosense Webster, Inc. (A Johnson & Johnson company) holds the 510(k) clearances for these devices.

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

Part 7. A brief discussion of nonclinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence. 21 CFR 807.92(b)(1)

The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and have been found to conform to applicable medical device safety standards. The SC2000 software update to VC11 (v6.1) does not involve any hardware modifications. The VC11A (v6.1) software update and the SC2000 System software has undergone component, integration and system level verification testing. The software verification and regression testing has been performed successfully to meet their previously determined acceptance criteria consistent with Siemens Product Lifecycle Procedure (PLP) and has passed. The ACUSON SC2000 Ultrasound system complies with the following standards:

  • AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and . A2:2010/(R)2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
  • IEC 60601-1-2: 2007(Third Edition), Medical electrical equipment Part 1-2: General ● requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • IEC 60601-2-37:2007+A1:2015, Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasound medical diagnostic and monitoring equipment
  • IEC 60601-2-18:2009. Medical electrical equipment Part 2: Particular requirements . for the safety of endoscopic equipment
  • IEC 62304:2006 Medical Device Software Software Life Cycle Processes ●
  • AIUM/NEMA UD-3:2004. Standard for Real Time Display of Thermal and Mechanical Acoustic ● Output Indices on Diagnostic Ultrasound Equipment
  • AIUM/NEMA UD-2:2004, Acoustic Output Measurement Standard for Diagnostic ● Ultrasound Equipment
  • ISO 10993-1:2009, Biological evaluation of Biological evaluation of Medical Devices ●
  • IEC 62366:2014, Medical Devices Application of Usability ●
  • IEC 62359:2010, Ultrasonics Field characterization Test Methods for the ● determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields.

{12}------------------------------------------------

Part 8. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence. 21 CFR 807.92(b)(2)

Since the ACUSON SC2000 VC11 (v6.1) Diagnostic Ultrasound System uses the identical technology and principles of use as the existing predicate devices SC2000, VC10 (v6.0) K202683, clinical studies were not required to support substantial equivalence. The SC2000 VC11A (v6.1) software was fully validated on phantoms by clinical personal with training in Ultrasound technology and has passed.

Substantial Equivalence to Predicates (21 CFR §807.92(b)(1))

The verification testing to the Customer requirements (CRS) for the SC2000 Diagnostic Ultrasound System including component, integration and regression testing for TrueFusion 2.0 and validation of the intended use is intended to support the claim of substantial equivalence to the following predicates:

The ACUSON SC2000, v6.0 (VC10) under K202683 on October 15, 2021, is the primary predicate device. The VC10 software includes the syngo® TrueFusion 1.0 software feature as an option.

The feature under modification is the syngo® TrueFusion 1.0 which was originally cleared under K171766 on July 12, 2017, this clearance serves as a reference predicate.

Part 9. Summary - 21 CFR 807.92(b)(3)

The Intended uses and other key features are consistent with traditional clinical practice and FDA guidelines. The design and development process of the manufacturer conforms to 21 CFR 820 Quality System Regulation and ISO 13485:2016 quality system standards. The product is designed to conform to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. The SC2000 Diagnostic ultrasound systems have accumulated a long history of safe and effective performance.

In summary, based on the successful verification and validation testing to the software acceptance criteria, the update for the TrueFusion 2.0 feature does not introduce any new potential safety risks. It is the opinion of Siemens Medical Solutions USA, Inc. that the ACUSON SC2000 Diagnostic Ultrasound system with the VC11 (v6.1) software update is substantially equivalent with respect to safety and effectiveness and performs as well as the currently cleared predicate device.

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.