K Number
K211684

Validate with FDA (Live)

Date Cleared
2021-08-27

(87 days)

Product Code
Regulation Number
876.1300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PillCam SB 3 capsule is intended for visualization of the small bowel mucosa.

  • It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.
  • It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
  • It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy.
    The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas.
    The PillCam SB 3 capsule may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age.
Device Description

The PillCam SB 3 Capsule endoscopy system is a device that provides visualization of the small bowel mucosa. The device was previously cleared under 510(k) submission K123864 for the SB 3 Capsule, PillCam DR3 recorder and accessories, and K170210 incorporated the PillCam Software 9.0 to the device is comprised of four main subsystems further described below.
Ingestible PillCam SB 3 Capsule: The disposable, ingestible PillCam SB 3 Capsule is designed to acquire images during the natural propulsion through the digestive system. The capsule transmits the acquired images via a RF communication channel to the PillCam DR3 recorder located outside the body.
PillCam DR3 Recorder: The PillCam DR3 recorder is an external receiving/recording unit that receives and stores the acquired images from the capsule.
PillCam Software: The PillCam Software 9.0E is a software application that is utilized to process, analyze, store, and view the acquired images collected from the PillCam DR3 recorder to create a video of the images. The software also includes a reporting function to create detailed clinical reports, in-service training videos, and patient instruction forms. PillCam Software 9.0E supports PillCam capsule endoscopy of the gastrointestinal (GI) tract with all PillCam capsules (SB, COLON, UGI and SBC (renamed Crohn's)).
Workstation and Accessories: The Workstation is a user-provided modified standard personal computer, that is the operational platform for the PillCam software. The Sensor array or sensor belt receive data from the PillCam capsule and transfer the data to the PillCam Recorder DR3.

AI/ML Overview

The provided text does not contain detailed acceptance criteria for standalone device performance and specific study data to prove that the device meets these criteria. Instead, it focuses on demonstrating substantial equivalence to predicate devices, particularly regarding a new remote administration option.

However, based on the information provided, here's a breakdown of what can be inferred about the device's performance characteristics and the supporting evidence:

1. Table of Acceptance Criteria (Inferred) and Reported Device Performance

Since explicit acceptance criteria for a "new" device performance study are not provided, we must infer that the accepted performance is based on the previously cleared predicate devices. The document explicitly states: "Prior testing provided in support of the Given PillCam SB 3 capsule endoscopy system (K123864) and PillCam Desktop Software 9.0 (cleared in K170210 with SBC capsule endoscopy system), verified the device's components, as well as the performance."

Acceptance Criteria (Implied)Reported Device Performance (Inherited from Predicate)
Visualization of small bowel mucosaProvides visualization of the small bowel mucosa
Detection of lesions indicating Crohn's diseaseCan be used in visualization and monitoring of lesions indicating Crohn's disease
Detection of lesions causing obscure bleedingCan be used in visualization and monitoring of lesions causing obscure bleeding
Detection of lesions causing IDACan be used in visualization and monitoring of lesions causing iron deficiency anemia (IDA)
Suspected Blood Indicator (SBI) feature functionalityIntended to mark frames suspected of containing blood or red areas
Usability for remote administrationUsability testing demonstrated the device was easy to use and healthcare providers could instruct/monitor participants. No user-related hazards recorded.
Safety with remote administration35 procedures performed via remote administration with no complications reported

2. Sample Size Used for the Test Set and Data Provenance

  • For remote administration usability and safety:

    • Test Set Sample Size: 35 procedures.
    • Data Provenance: Commercial centers, since October 2020. This indicates prospective real-world use data. The country of origin is not explicitly stated, but the submitter is an Israeli company.
  • For core device performance (visualization, lesion detection):

    • The document implies that the performance data for these aspects come from the predicate device clearances (K123864 and K170210). The sample sizes and data provenance for those original studies are not provided in this document.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • For remote administration usability: The document mentions "usability testing... consisted of simulated-use sessions with the device and a healthcare provider." It does not specify the number of healthcare providers or their qualifications.
  • For remote administration safety: The 35 procedures involved patients, and "no complications reported" would likely be based on clinical observation by medical staff. No independent experts establishing "ground truth" are mentioned for this aspect, as it relates to complication rates rather than diagnostic accuracy.
  • For core device performance: This information would be within the predicate device submissions (K123864, K170210) and is not provided in this document.

4. Adjudication Method for the Test Set

  • The document does not describe any formal adjudication method for the remote administration usability and safety data. For usability, errors were "captured," but the process for classifying or adjudicating these errors (e.g., by multiple experts) is not detailed.
  • For the core device performance, information on adjudication from the predicate studies is not provided in this document.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a MRMC comparative effectiveness study was not done for this submission. The submission is primarily focused on demonstrating substantial equivalence for a labeling change (remote administration) based on existing device performance. There is no mention of human readers improving with or without AI assistance, as the "AI" component here (SBI feature) is already part of the predicate and not being newly evaluated in a comparative effectiveness study against human readers.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • The document refers to the "Suspected Blood Indicator (SBI) feature" which "is intended to mark frames of the video suspected of containing blood or red areas." This feature, while part of the software, is an assistive tool for human readers.
  • The document does not report a standalone performance study (i.e., algorithm only without human-in-the-loop performance) for the SBI feature or any other automated detection capability. Its performance is implicitly considered acceptable through its inclusion in the predicate device.

7. Type of Ground Truth Used

  • For remote administration usability: Ground truth was established through observing and recording user performance and errors in simulated use, likely against pre-defined success criteria for the procedure steps.
  • For remote administration safety: Ground truth was the absence of reported complications in real-world clinical use.
  • For core device performance (e.g., lesion detection): This information would be in the predicate device submissions and is not provided in this document. Typically for such devices, ground truth might involve endoscopic findings, histological confirmation (pathology), or clinical follow-up/outcomes data.

8. Sample Size for the Training Set

  • The document does not provide any information on the sample size used for training any algorithms within the PillCam Software 9.0E, including the SBI feature. This is likely because the software itself (including the SBI feature) existed in the predicate device.

9. How the Ground Truth for the Training Set Was Established

  • As no information on a training set is provided, the method for establishing its ground truth is also not available in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

August 27, 2021

Given Imaging Ltd. (Medtronic) % Randy Prebula Partner Hogan Lovells US LLP 555 13th Street. NW Washington, DC 20004

K211684 Re:

Trade/Device Name: PillCam SB 3 capsule endoscopy system, PillCam Software 9.0E Regulation Number: 21 CFR 876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: Class II Product Code: NEZ Dated: June 1, 2021 Received: June 1, 2021

Dear Randy Prebula:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K211684 Device Name

PillCam SB 3 Capsule Endoscopy System PillCam Software 9.0E Indications for Use (Describe)

The PillCam SB 3 capsule is intended for visualization of the small bowel mucosa.

  • . It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.
  • It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
  • It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency ● anemia (IDA) not detected by upper and lower endoscopy.

The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas.

The PillCam SB 3 capsule may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K211684 Page 1 of 5

510(k) SUMMARY

Medtronic's PillCam

SB 3 capsule endoscopy system

PillCam Software 9.0E

Submitter

Given Imaging Ltd. (a Medtronic company) 2 Hacarmel St. New Industrial Park, PO Box 258 Yoqneam, 20692 Israel

Phone: +972-73-2501577

Contact Person: Efrat Shamgar, Sr. Regulatory Affairs Specialist

Date Prepared: June 1, 2021

Name of Device:

PillCam SB 3 Capsule endoscopy system with PillCam Software 9.0E

Common or Usual Name:

Ingestible telemetric gastrointestinal capsule imaging system

Classification Name:

Ingestible telemetric gastrointestinal capsule imaging system

Regulatory Class:

Class II, 21 CFR 876.1300

Product Code:

NEZ

Predicate Devices

Given PillCam SB 3 capsule endoscopy system and Given PillCam® endoscopy system with RAPID 8.0 (K123864) - Primary Predicate

Given Imaging PillCam SBC capsule endoscopy system and PillCam Desktop Software 9.0 (K170210) - Secondary Predicate

{4}------------------------------------------------

Device Description

The PillCam SB 3 Capsule endoscopy system is a device that provides visualization of the small bowel mucosa. The device was previously cleared under 510(k) submission K123864 for the SB 3 Capsule, PillCam DR3 recorder and accessories, and K170210 incorporated the PillCam Software 9.0 to the device is comprised of four main subsystems further described below.

Ingestible PillCam SB 3 Capsule

The disposable, inqestible PillCam SB 3 Capsule is designed to acquire images during the natural propulsion through the digestive system. The capsule transmits the acquired images via a RF communication channel to the PillCam DR3 recorder located outside the body.

PillCam DR3 Recorder

The PillCam DR3 recorder is an external receiving/recording unit that receives and stores the acquired images from the capsule.

PillCam Software

The PillCam Software 9.0E is a software application that is utilized to process, analyze, store, and view the acquired images collected from the PillCam DR3 recorder to create a video of the images. The software also includes a reporting function to create detailed clinical reports, in-service training videos, and patient instruction forms. PillCam Software 9.0E supports PillCam capsule endoscopy of the gastrointestinal (GI) tract with all PillCam capsules (SB, COLON, UGI and SBC (renamed Crohn's)).

Workstation and Accessories

The Workstation is a user-provided modified standard personal computer, that is the operational platform for the PillCam software. The Sensor array or sensor belt receive data from the PillCam capsule and transfer the data to the PillCam Recorder DR3.

Intended Use / Indications for Use

The PillCam SB 3 capsule is intended for visualization of the small bowel mucosa.

  • It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.
  • It may be used in the visualization and monitoring of lesions that may be a source of ● obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
  • It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy.

The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas.

{5}------------------------------------------------

The SB 3 capsule may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age.

Summary of Technological Characteristics

The PillCam SB 3 capsule endoscopy system and PillCam Software 9.0E components are identical to the components of the currently marketed predicate devices. Both the subject device and the primary and secondary predicate devices, use the same mechanisms to fulfil their identical intended use, respectively, and share the same technological characteristics. Each of the components of the subject device operates using the same technology and technological characteristics as the currently marketed predicate devices, respectively.

The only difference between the subject device and the currently marketed proposed predicates is a change to the labeling to allow remote administration of the subject device's SB 3 Capsule as an option.

A table comparing the key features of the subject and predicate devices is provided below.

Performance Data

Prior testing provided in support of the Given PillCam SB 3 capsule endoscopy system (K123864) and PillCam Desktop Software 9.0 (cleared in K170210 with SBC capsule endoscopy system), verified the device's components, as well as the performance. The subject device is almost identical to the currently marketed predicate devices; therefore, the previously completed non-clinical tests remain applicable.

The remote administration procedure has been performed at 3 commercial centers since October of 2020. At the time of this submission, 35 procedures have been performed with no complications reported. Additionally, usability testing was conducted in accordance with IEC 62366-1:2015 to evaluate the remote administration procedure of the device. This simulateduse testing consisted of simulated telehealth sessions with the device and a healthcare provider. The testing demonstrated that the device was easy to use and the HCP was able to instruct and monitor participants as needed during the procedure. No user-related hazards were recorded, with the errors captured not posing any additional risk.

Conclusions

The PillCam SB 3 Capsule endoscopy system with PillCam Software 9.0E is as safe and effective as the PillCam SB 3 endoscopy system (K123864), and the PillCam SBC capsule endoscopy system, PillCam Desktop Software 9.0 (K170210). The PillCam SB 3 capsule endoscopy system and PillCam Software 9.0E has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor differences in the environment of use to incorporate the remote administration procedure do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled.

{6}------------------------------------------------

Subject DevicePredicate DevicesComparison
Device NamePillCam SB 3 capsuleendoscopy system and PillCamSoftware 9.0E (includingRemote procedure)PillCam SB 3 capsuleendoscopy system with PillCamSoftware 9.0E (K123864,K170210)
Product CodeNEZNEZ, PGDSimilar
Indications for UseThe PillCam SB 3 capsule isintended for visualization of thesmall bowel mucosa.It may be used in the visualizationand monitoring of lesions that mayindicate Crohn's disease notdetected by upper and lowerendoscopy. It may be used in the visualizationand monitoring of lesions that maybe a source of obscure bleeding(either overt or occult) notdetected by upper and lowerendoscopy. It may be used in the visualizationand monitoring of lesions that maybe potential causes of irondeficiency anemia (IDA) notdetected by upper and lowerendoscopy. The Suspected Blood Indicator(SBI) feature is intended to markframes of the video suspected ofcontaining blood or red areas. The PillCam SB 3 capsule may beused as a tool in the detection ofabnormalities of the small boweland is intended for use in adultsand children from two years ofage.The PillCam SB 2/ SB 3 capsule isintended for visualization of thesmall bowel mucosa.It may be used in the visualizationand monitoring of lesions that mayindicate Crohn's disease notdetected by upper and lowerendoscopy. It may be used in the visualizationand monitoring of lesions that maybe a source of obscure bleeding(either overt or occult) notdetected by upper and lowerendoscopy. It may be used in the visualizationand monitoring of lesions that maybe potential causes of irondeficiency anemia (IDA) notdetected by upper and lowerendoscopy. The Suspected Blood Indicator(SBI) feature is intended to markframes of the video suspected ofcontaining blood or red areas. The PillCam SB 2/ SB 3 capsulemay be used as a tool in thedetection of abnormalities of thesmall bowel and is intended foruse in adults and children fromtwo years of age.Identical*
Environment ofUseInpatient, outpatient settings,remote useInpatient and outpatient settingsSimilar
ObtainedParametersimagesimagesIdentical
Table 1: Substantial Equivalence Table
------------------------------------------

{7}------------------------------------------------

Subject DevicePredicate DevicesComparison
Device NamePillCam SB 3 capsuleendoscopy system and PillCamSoftware 9.0E (includingRemote procedure)PillCam SB 3 capsuleendoscopy system with PillCamSoftware 9.0E (K123864,K170210)
SystemComponentsThree main subsystems:PillCam SB capsulesThe PillCam DR3 RecorderThe PillCam Software Workstation(WS) (supplied by the user) withPillCam Software Application(Version 9.0E)Three main subsystems:PillCam SB capsulesThe PillCam DR3 RecorderThe PillCam Software Workstation(WS) with PillCam SoftwareApplication (Version 9.0E)Identical
Method ofRemovalNatural excretionNatural excretionIdentical
Capsule length26.2 mm26.2 mmIdentical
Capsule diameter11.4 mm11.4 mmIdentical
Weight3.0 g3.0 gIdentical
MaterialBiocompatible plasticBiocompatible plasticIdentical
Capsule PowerSource BatteryType2 Mercury free Silver Oxidebatteries2 Mercury free Silver OxidebatteriesIdentical
# of optical heads11Identical
Optical Field ofView156 ° (ISO-8600-3)156 ° (ISO-8600-3)Identical
SignalTransmissionRF signalRF signalIdentical
Effective VisibilityDistance0-30 mm0-30 mmIdentical
Frame Rate2 fps or 2 to 6 fps2 fps or 2 to 6 fpsIdentical
Operating TimeMinimum of 8 hoursMinimum of 8 hoursIdentical
ImagerGILO6GILO6Identical
Number of pixels320x320320x320Identical
Minimaldetectable objectAt least 0.07 mmAt least 0.07 mmIdentical
Programmingof/and operationmodeAdaptive frame rate beginning atpairingAdaptive frame rate beginning atpairingIdentical

*The full IFU statement of K170210 contains more statements due to the Colon Capsule, which is unrelated to the subject device. The only information of note is for the SB3.

§ 876.1300 Ingestible telemetric gastrointestinal capsule imaging system.

(a)
Identification. An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.(b)
Classification. Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”