K Number
K211286
Manufacturer
Date Cleared
2021-11-16

(203 days)

Product Code
Regulation Number
868.5260
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pal® Ultipor® U55/U55N Filter is a single use bacterial/viral filter and heat and moisture exchanger (HME) for patient side or machine side installation in breathing systems. It is designed to reduce bacterial/viral transmissions between the patient, the equipment and the environment and to reduce the loss of patient heat and humidity. The Ultipor® U55/U55N breathing circuit filter has >99.999% bacterial and >99.995% viral efficiency.

The filter is for single patient use for adult patients, and is intended for use within breathing systems in healthcare and home environments where ventilation is required and for a maximum duration of 24 hours.

Device Description

The Pall® Ultipor®U55/U55N Breathing Circuit Bacterial Filter and Heat and Moisture Exchanger is a disposable, highly efficient, bi-directional bacterial/viral filter for patient ventilation solutions and is comprised of the following components:

  • . Filter housing - Is comprised of two molded halves, an inlet housing and an outlet housing that are joined and sealed together. The finished filter housing holds the pleated hydrophobic filter media and provides conical (tapered) fittings on opposing sides of the finished housing to provide connection to the conventional equipment used in the breathing circuit. One side of the finished housing has a coaxial conical fitting, with a conical fitting located on the opposing side.
  • Hydrophobic filter media Provides airborne bacterial removal efficiency of >99.999%, . airborne viral removal efficiency of >99.995% and waterborne microbial contaminant removal efficiency of 100%. If the filter is used at the patient end, the filter media also acts as a heat and moisture exchanger (HME) by conserving a proportion of the heat and humidity present in the patient's exhaled air and returning it to the patient on the next inspiration.
AI/ML Overview

The provided text is a 510(k) summary for a medical device called the Pall® Ultipor® U55/U55N Breathing Circuit Bacterial Filter and Heat and Moisture Exchanger. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive comparative effectiveness study against human readers or specific clinical outcomes in the way an AI diagnostic device might.

Therefore, many of the requested categories are not applicable to the information contained in this document.

Here's the information extracted from the provided text, addressing your questions where applicable:

1. Table of Acceptance Criteria and Reported Device Performance

MethodologyPurposeAcceptance CriteriaReported Device Performance
BiocompatibilityEvaluate device's biological safety for the intended use, in accordance with ISO 10993-1 and FDA's corresponding guidance document• Cytotoxicity (L929 MEM elution) per ISO 10993-5;• Sensitization and intracutaneous injection per ISO 10993-10;• Acute systemic toxicity per ISO 10993-11 with both polar and non-polar solvents (in lieu of testing to ISO 18562-4); and• Material-mediated pyrogenicity per ISO 10993-11.All results were acceptable.
Microbial retention (aerosol bacterial and viral)Evaluate aerosol bacterial and viral removal>99.999% effectiveness for bacteria removal and >99.995% for virus removalUnaged and aged (5 years) filters demonstrated bacterial effectiveness of >99.999% and viral effectiveness of >99.995%.
Microbial retention (liquid bacterial)Evaluate liquid bacterial removal100% retentionNo bacteria were recovered from the water placed on the machine side following the challenge for any unaged or aged (3 & 5 years) filters. This testing also supports filter media integrity and hydrophobicity.
Poly-alpha-olefin (PAO) RemovalEvaluate filtration efficiencyPenetration of ≤0.09%All unaged and 5-year aged filters had a penetration of 0.09%.
Sodium Chloride Particulate RemovalEvaluate filtration efficiencyPre- and post-conditioning penetrations of <1%Unaged and 5-year aged filters had acceptable performance for pre-conditioning and post-conditioning penetration.
Filter Media Integrity with 50 cm hydrostatic headEvaluate hydrophobicity (resistance to penetration by a measured column of liquid)No water breakthrough of the filter media for 60 seconds after application of a 50 cm head of sterile waterNo unaged or 3-year aged filter showed water breakthrough.
Air Flow Resistance (Pressure Drop) Pre- and Post-hydrostatic headEvaluate any changes in the filter's resistance to air flow before and after being subjected to a 15 cm hydrostatic head.At 60 L/min, average inspiratory/expiratory flow resistance should be <2.5 cm H2O pre-hydrostatic head and ≤6.0 cm H2O post-hydrostatic head, and no evidence of water breakthrough within 60 seconds of the 15 cm hydrostatic head.The pre-hydrostatic head and post-hydrostatic maximum flow resistance were passing across all unaged, 3-year aged, and 5-year aged filters.
Heat and Moisture Exchange efficiencyEvaluate moisture loss for tidal volumes of 250 mL, 500 mL, 750 mL, and 1000 mL<12 mg/L (10-12 mg H2O/L) at 250 ml tidal volume;<16 mg/L (13-16 mg H2O/L) at 500 ml tidal volume;<20 mg/L (19 mg H2O/L) at 750 ml tidal volume;<22 mg/L (19-22 mg H2O/L) at 1000 ml tidal volume.The average moisture loss met the acceptance criteria for the specified tidal volumes for unaged, and aged (1, 3, & 5 years filters.
Pressure drop pre and post moisture loss testingEvaluate pressure drop pre and post moisture loss testingThe inspiratory/expiratory flow resistance should be <2.5 cm H2O at 60 L/minFor unaged as well as 1, 3 & 5 year aged filters, the maximum inspiratory and expiratory flow resistance were passing, showing that the pressure drop does not increase when conditioned for the length of filter use or when challenged with liquid.
Leakage (pressure decay) pre and post moisture loss testingEvaluate leakage (pressure decay) under an applied pressure pre and post moisture loss testingISO 9360-1 Section 6.4 does not specify acceptance criteria, only to record the volume of air required; however, Pall has specified an internal acceptance criteria for leakage of <25 mL/min at 7 ± 0.5 kPa.The maximum leakage values for unaged as well as for 1, 3 & 5 year aged filters met predefined criteria both pre- and post-moisture loss.
Compliance in accordance with ISO 9360 pre- and post-moisture loss testingEvaluate compliance in accordance with ISO 9360 pre- and post- moisture loss testingISO 9360-1 Section 6.5 does not specify an acceptance criterion, only to record the volume of air required.There was no significant increase in flow rate following filter conditioning. All test values for unaged and aged (1, 3, & 5 years) filters were acceptable.
Housing Integrity (for pressure decay)Evaluate filter leakage at two different pressuresLeakage <4 ml/min for unaged filters at 70 mbar (7 kPa) and 150 mbar (15 kPa); leakage <25 ml/min at 70 mbar (7 kPa) for aged filters.Unaged as well as 1 & 3 year aged filters had leakage values meeting the predefined acceptance criteria at the evaluated pressures.
Connector ComplianceDemonstrate inlet and outlet ports comply with ISO 5356-1All ports must comply with ISO 5356-1For both 15 mm and 22 mm connectors, axial force and rotation for unaged and 1 & 3 year aged filters were in accordance with the standard.
NebulizationEvaluate changes in air flow resistance across filter when used with nebulized drugs at patient and machine ends of breathing circuitAir flow resistance <6 cm H2O at 60 L/min airflow after 24 hours of nebulizationUnaged & 3 year aged filters: Highest flow resistance values were within specification for both filter locations and for drugs delivered by air-driven and ultrasonic nebulizers as well as drugs delivered by metered dose inhalers, as applicable.
Particulate AnalysisAssess if airborne particulate is emitted into the gas stream in accordance with ISO 18562Particulate category PM2.5 and PM10 acceptable concentrations: 12 µg/m3 and 150 µg/m3, respectivelyThe minimum, maximum, and average particulate concentrations are all below acceptable limits.
Volatile Organic Compounds (VOCs) AnalysisAssess if airborne VOCs are emitted from the filter into the gas stream per ISO 18562Threshold of Toxicological Concern (TTC) of 360 µg/d for the first 24 hours and a limit of 120 µg/d for the remaining 24 hoursExposure of the individual VOCs released are all below pre-defined TTC.
Shelf life (stability / maintained performance)Evaluate shelf life3 years from the date of manufacture based on the stability testingStability of device confirmed from performance testing conducted on 3-year real-time aged and 5-year accelerated aged devices.

2. Sample size used for the test set and the data provenance

The document states, "Testing was performed on the subject device or on representative filters reflecting the same fundamental design and materials, and the same manufacturing methods." It mentions testing unaged and aged filters (e.g., 1, 3, or 5 years aged). However, it does not specify the exact number of filters (sample size) used for each test. The data provenance is related to the manufacturing process of the device itself and aging protocols, not clinical patient data from a specific country.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable. The tests performed are laboratory-based, non-clinical performance evaluations of the device's physical and biological properties (e.g., filtration efficiency, pressure drop, biocompatibility). These do not involve human readers establishing a "ground truth" for interpretations like in a medical imaging study.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable, as the tests are non-clinical laboratory evaluations and do not involve human interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. This submission is for a breathing circuit filter, not an AI-assisted diagnostic device. Therefore, no MRMC comparative effectiveness study was performed, nor is there any AI assistance involved.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. As explained above, this is not an AI algorithm but a physical medical device (filter).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" in this context is defined by standardized test methodologies and established performance specifications for medical filters (e.g., ISO standards for biocompatibility, filtration efficiency, pressure drop, moisture exchange, connector compliance, VOC emissions). The results are quantitative measurements against these criteria.

8. The sample size for the training set

This question is not applicable. As this is not an AI device, there is no "training set."

9. How the ground truth for the training set was established

This question is not applicable, as there is no "training set" for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

November 16, 2021

Pall Corporation % Randy Prebula Partner Hogan Lovells U.S. LLP 555 Thirteenth Street NW Washington, District of Columbia 20004

Re: K211286

Trade/Device Name: Ultipor U55/U55N Breathing Circuit Filter and Heat and Moisture Exchanger Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: Class II Product Code: CAH Dated: October 18, 2021 Received: October 18, 2021

Dear Randy Prebula:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K211286

Device Name

Ultipor® U55/U55N Breathing Circuit Bacterial Filter and Heat and Moisture Exchanger

Indications for Use (Describe)

The Pal® Ultipor® U55/U55N Filter is a single use bacterial/viral filter and heat and moisture exchanger (HME) for patient side or machine side installation in breathing systems. It is designed to reduce bacterial/viral transmissions between the patient, the equipment and the environment and to reduce the loss of patient heat and humidity. The Ultipor® U55/U55N breathing circuit filter has >99.999% bacterial and >99.995% viral efficiency.

The filter is for single patient use for adult patients, and is intended for use within breathing systems in healthcare and home environments where ventilation is required and for a maximum duration of 24 hours.

Type of Use (Select one or both, as applicable)

図 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY - K211286

Pall Corporation's Ultipor® U55/U55N Breathing Circuit Bacterial Filter and Heat and Moisture Exchanger

Submitter:Pall Corporation25 Harbor Park DrivePort Washington, NY 11050, USA
Primary Contact:Brian GoetzSenior Manager, Regulatory AffairsTelephone Number: 516-801-9267Fax Number: 516-801-9643
Date Prepared:November 2, 2021
Trade Names:Pall® Ultipor® U55/U55N Filter
Common Name:Filter, Bacterial, Breathing-Circuit
Classification Regulation:21 C.F.R. § 868.5260, Breathing circuit bacterial filter
Regulatory Class:Class II
FDA Product Code:CAH
Predicate Device:Intersurgical, Inc. Hydro-Guard Mini Breathing Filter and Heatand Moisture Exchanger (Model 1745) (K092409)

Intended Use/Indications for Use

The Pall® Ultipor® U55/U55N Filter is a single use bacterial/viral filter and heat and moisture exchanger (HME) for patient side or machine side installation in breathing systems. It is designed to reduce bacterial/viral transmissions between the patient, the equipment and the environment and to reduce the loss of patient heat and humidity. The Ultipor® U55/U55N breathing circuit filter has

99.999% bacterial and >99.995% viral efficiency.

The filter is for single patient use for adult patients, and is intended for use within breathing systems in healthcare and home environments where ventilation is required and for a maximum duration of 24 hours.

Device Description

The Pall® Ultipor®U55/U55N Breathing Circuit Bacterial Filter and Heat and Moisture Exchanger is a disposable, highly efficient, bi-directional bacterial/viral filter for patient ventilation solutions and is comprised of the following components:

  • . Filter housing - Is comprised of two molded halves, an inlet housing and an outlet housing that are joined and sealed together. The finished filter housing holds the pleated hydrophobic filter media and provides conical (tapered) fittings on opposing sides of the finished housing to provide connection to the conventional equipment used in the breathing circuit. One side of the finished housing has a coaxial conical fitting, with a conical fitting located on the opposing side.

{4}------------------------------------------------

  • Hydrophobic filter media Provides airborne bacterial removal efficiency of >99.999%, . airborne viral removal efficiency of >99.995% and waterborne microbial contaminant removal efficiency of 100%. If the filter is used at the patient end, the filter media also acts as a heat and moisture exchanger (HME) by conserving a proportion of the heat and humidity present in the patient's exhaled air and returning it to the patient on the next inspiration.

Summary of Technological Characteristics

The subject and predicate devices have the same fundamental design, consisting of the filter media and housing. In addition, the subject and predicate filters both operate based on size exclusion (direct interception filtration) as the primary filtration mechanism, supplemented by effects of inertial impaction, and diffusion impaction, to prevent certain particles from passing through the filter to the other side.

Subject DevicePredicate Device
ElementPall® Ultipor® U55 and U55NBreathing Circuit Bacterial Filter andHeat and Moisture ExchangerIntersurgical 1745 Hydro-GuardMini Breathing Filter and Heatand Moisture Exchanger(K092409)
Indications for UseThe Pall® Ultipor® U55/U55N Filter is asingle use bacterial/viral filter and heat andmoisture exchanger (HME) for patient sideor machine side installation in breathingsystems. It is designed to reducebacterial/viral transmissions between thepatient, the equipment and the environmentand to reduce the loss of patient heat andhumidity. The Ultipor® U55/U55N breathingcircuit filter has >99.999% bacterial and>99.995% viral efficiency.The filter is for single patient use for adultpatients, and is intended for use withinbreathing systems in healthcare and homeenvironments where ventilation is requiredand for a maximum duration of 24 hours.The Intersurgical Hydro-Guard MiniFilter and Heat and MoistureExchanger is for use at the patientand equipment connections. It isdesigned to reduce bacterial/viraltransmissions between the patientand equipment and to reduce theloss of patient heat and humidity.The filter is for single patient usefor an adult target population, isintended for use within breathingsystems in healthcare and homeenvironments where ventilation isrequired and a maximum durationof 24 hours.
Installation positionsPatient and machine endsPatient and machine ends
Filtration efficiency(%)>99.999% bacterial; >99.995% viral>99.999%
Filtration efficiency -waterborne bacterialremoval (%)100Not available
Minimum tidalvolume (mL)150 (See Note 1 below)>200
Maximum tidalvolume (mL)10001000
Pressure drop(cmH2O) @ 60 L/min2.52.9
Heat and moistureexchanger loss H2O@ 500 mL tidalvolume (mg/L)1523

{5}------------------------------------------------

Compressible volume(dead space) (mL)approx.5063
Gas Leakage(mL/min)< 4<5
Compliance (ml/kPa)<1<1
Connections patientside (mm)22 male/15 female22 male/15 female
Connectionsmachine side (mm)22 female22 female/15 male
Dry weight (g)5030
Filter Life (hrs)2424

NOTE: 1. Lowest tidal volume while providing adequate patient ventilation should be determined by the physician by clinical assessment, inter volume, ventilation settings and breathing circuit configuration.

Performance Data

Testing was performed on the subject device or on representative filters reflecting the same fundamental design and materials, and the same manufacturing methods. The following non-clinical tests were performed to demonstrate the performance of the subject Ultipor® U55/U55N Filter and support the proposed indications for use:

MethodologyPurposeAcceptance CriteriaResults
BiocompatibilityEvaluate device'sbiological safetyfor the intendeduse, in accordancewith ISO 10993-1and FDA'scorrespondingguidancedocument• Cytotoxicity (L929 MEMelution) per ISO 10993-5;• Sensitization andintracutaneous injection perISO 10993-10;• Acute systemic toxicity perISO 10993-11 with both polarand non-polar solvents (inlieu of testing to ISO 18562-4); and• Material-mediatedpyrogenicity per ISO 10993-11.All results were acceptable.
MicrobialretentionEvaluate aerosolbacterial and viralremoval>99.999% effectiveness forbacteria removal and>99.995% for virus removalUnaged and aged (5 years) filtersdemonstrated bacterialeffectiveness of >99.999% andviral effectiveness of >99.995%.
MicrobialretentionEvaluate liquidbacterial removal100% retentionNo bacteria were recovered fromthe water placed on the machineside following the challenge forany unaged or aged (3 & 5 years)filters. This testing also supportsfilter media integrity andhydrophobicity.
Poly-alpha-olefin (PAO)Removal1Evaluate filtrationefficiencyPenetration of ≤0.09%,All unaged and 5-year aged filtershad a penetration of 0.09%.
SodiumChlorideParticulateRemovalEvaluate filtrationefficiencyPre- and post-conditioningpenetrations of <1%Unaged and 5-year aged filtershad acceptable performance forpre-conditioning and post-conditioning penetration.
MethodologyPurposeAcceptance CriteriaResults
Filter MediaIntegrity with50 cmhydrostaticheadEvaluatehydrophobicity(resistance topenetration by ameasured column ofliquid)No water breakthrough of thefilter media for 60 secondsafter application of a 50 cmhead of sterile waterNo unaged or 3-year aged filtershowed water breakthrough.
Air FlowResistance(PressureDrop) Pre- andPost-hydrostaticheadEvaluate anychanges in thefilter's resistanceto air flow beforeand after beingsubjected to a 15cm hydrostatichead.At 60 L/min, averageinspiratory/expiratory flowresistance should be <2.5 cmH2O pre-hydrostatic head and≤6.0 cm H2O post-hydrostatichead, and no evidence of waterbreakthrough within 60 secondsof the 15 cm hydrostatic head.The pre-hydrostatic head andpost-hydrostatic maximum flowresistance were passing acrossall unaged, 3-year aged, and 5-year aged filters.
Heat andMoistureExchangeefficiencyEvaluate moistureloss for tidalvolumes of 250mL, 500 mL, 750mL, and 1000 mL<12 mg/L (10-12 mg H2O/L) at250 ml tidal volume;<16 mg/L (13-16 mg H2O/L) at500 ml tidal volume;<20 mg/L (19 mg H2O/L) at750 ml tidal volume;<22 mg/L (19-22 mg H2O/L) at1000 ml tidal volume.The average moisture loss metthe acceptance criteria for thespecified tidal volumes forunaged, and aged (1, 3, & 5years filters.
Evaluate pressuredrop pre and postmoisture losstestingThe inspiratory/expiratory flowresistance should be <2.5 cmH2O at 60 L/minFor unaged as well as 1, 3 & 5year aged filters, the maximuminspiratory and expiratory flowresistance were passing, showingthat the pressure drop does notincrease when conditioned for thelength of filter use or whenchallenged with liquid.
Evaluate leakage(pressure decay)under an appliedpressure pre andpost moisture losstestingISO 9360-1 Section 6.4 doesnot specify acceptancecriteria, only to record thevolume of air required;however, Pall has specified aninternal acceptance criteria forleakage of <25 mL/min at 7 ±0.5 kPa.The maximum leakage values forunaged as well as for 1, 3 & 5year aged filters met predefinedcriteria both pre- and post-moisture loss.
Evaluatecompliance inaccordance withISO 9360 pre- andpost- moisture losstestingISO 9360-1 Section 6.5 doesnot specify an acceptancecriterion, only to record thevolume of air required.There was no significant increasein flow rate following filterconditioning. All test values forunaged and aged (1, 3, & 5years) filters were acceptable.
HousingIntegrity (forpressuredecay)Evaluate filterleakage at twodifferent pressuresLeakage <4 ml/min for unagedfilters at 70 mbar (7 kPa) and150 mbar (15 kPa); leakage<25 ml/min at 70 mbar (7 kPa)for aged filters.Unaged as well as 1 & 3 yearaged filters had leakage valuesmeeting the predefinedacceptance criteria at theevaluated pressures.
ConnectorComplianceDemonstrate inletand outlet portscomply with ISO5356-1All ports must comply with ISO5356-1For both 15 mm and 22 mmconnectors, axial force androtation for unaged and 1 & 3year aged filters were inaccordance with the standard.
MethodologyPurposeAcceptance CriteriaResults
NebulizationEvaluate changesin air flowresistance acrossfilter when usedwith nebulizeddrugs at patientand machine endsof breathing circuitAir flow resistance <6 cm H2Oat 60 L/min airflow after 24hours of nebulizationUnaged & 3 year aged filters:Highest flow resistance valueswere within specification for bothfilter locations and for drugsdelivered by air-driven andultrasonic nebulizers as well asdrugs delivered by metered doseinhalers, as applicable.
ParticulateAnalysisAssess if airborneparticulate isemitted into thegas stream inaccordance withISO 18562Particulate category PM2.5and PM10 acceptableconcentrations: 12 µg/m3 and150 µg/m3, respectivelyThe minimum, maximum, andaverage particulateconcentrations are all belowacceptable limits
VolatileOrganicCompounds(VOCs)AnalysisAssess if airborneVOCs are emittedfrom the filter intothe gas stream perISO 18562Threshold of ToxicologicalConcern (TTC) of 360 µg/d forthe first 24 hoursand a limit of 120 µg/d for theremaining 24 hoursExposure of the individual VOCsreleased are all below pre-definedTTC
Shelf life(stability /maintainedperformance) 1Evaluate shelf life3 years from the date ofmanufacture based on thestability testingStability of device confirmed fromperformance testing conductedon 3-year real-time aged and 5-year accelerated aged devices.

{6}------------------------------------------------

{7}------------------------------------------------

1 Testing also performed during filter manufacturing process.

² The subject filter is not supplied sterile and therefore does not have an expiry date based on validation of the integrity of a sterile barrier seal.

Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, Intersurgical Inc.'s Model 1745 Hydro-Guard Mini Breathing Filter and Health and Moisture Exchanger (K092409).

§ 868.5260 Breathing circuit bacterial filter.

(a)
Identification. A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.(b)
Classification. Class II (performance standards).