K Number
K211229
Date Cleared
2021-07-29

(97 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vinyl compound examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Device Description

The proposed device, Vinyl compound examination gloves are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. The proposed device is Powder Free Vinyl Patient Examination Gloves. The proposed device is Blue. The design of proposed device is addressing the standards as ASTM D6124. ASTM D5151, and ASTM D5250. The proposed device is not provided as sterilized. The proposed device is made of vinyl chloride.

AI/ML Overview

This document is a 510(k) premarket notification for "Vinyl compound examination gloves." It focuses on demonstrating substantial equivalence to a predicate device, not on proving new clinical effectiveness or diagnostic accuracy of an AI device. Therefore, many of the requested categories related to AI studies, ground truth, expert adjudication, and comparative effectiveness are not applicable.

Here's an analysis based on the provided text, focusing on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodPurposeAcceptance Criteria (from D5250)Reported Device Performance
ASTM D5151Testing for Freedom from holesFreedom from holesNo water leakage is inspected from 200 samples
ASTM D6124Determine the powder residue for powder free gloves<2.0 mg per gloveResidual Powder of Size XS: Average 0.28 mg;Residual Powder of Size S: Average 0.25 mg;Residual Powder of Size M: Average 0.31 mg;Residual Powder of Size L: Average 0.25 mg;Residual Powder of Size XL: Average 0.35mg;
ASTM D412, ASTM D573Testing for Physical property characteristicsTensile Strength: 11 MPa min.Ultimate Elongation: 300% min.Tensile Strength: ≥ 12 MPa;Ultimate Elongation: ≥ 321%.
ASTM D3767Testing For physical dimensions specificationLength: 230 mm min. for all sizes (XS, S, M, L, XL);Width: 75±5 mm for XS; 85±5 mm for S;95±5 mm for M; 105±5 mm for L; 115±5 mm for XL.Finger Thickness: ≥0.08 mm;Palm Thickness: ≥0.08 mm;All acceptance criteria above meet the requirements in Table 2 Dimensions and Tolerances of ASTM D5250Length of Size XS: ≥ 237 mm;Width of Size XS: 75±1 (74-76) mm;Palm Thickness of Size XS: ≥0.08 mm;Finger Thickness of Size XS: ≥0.09 mm.Length of Size S: ≥ 235 mm;Width of Size S: 85±1 (85-86) mm;Palm Thickness of Size S: ≥0.08 mm;Finger Thickness of Size S: ≥0.09 mm.Length of Size M: ≥ 235 mm;Width of Size M: 95±2 (95-97) mm;Palm Thickness of Size M: ≥0.08 mm;Finger Thickness of Size M: ≥0.09 mm.Length of Size L: ≥ 235 mm;Width of Size L: 105±2 (105-107) mm;Palm Thickness of Size L: ≥0.08 mm;Finger Thickness of Size L: ≥0.09 mm.Length of Size XL: ≥ 234 mm;Width of Size XL: 115±3 (115-118) mm;Palm Thickness of Size XL: ≥0.08 mm;Finger Thickness of Size XL: ≥0.09 mm.
ISO 10993-5Evaluate the endpoint of Cytotoxicity for biocompatibilityThe test article showed "negative" cytotoxicityUnder the conditions of the study, the test article showed "negative" cytotoxicity.
ISO 10993-10 (Irritant)Evaluate the endpoint of irritant for biocompatibilityThe response of the test article extract is negligible.The test result showed that the response of the test article extract was categorized as negligible under the test condition.
ISO 10993-10 (Sensitization)Evaluate the endpoint of sensitization for biocompatibilityThe test article showed no evidence of causing delayed dermal contact sensitization.The test article showed no evidence of causing delayed dermal contact sensitization in the guinea pig.

2. Sample size used for the test set and the data provenance

  • ASTM D5151 (Freedom from holes): 200 samples. The data provenance is not specified beyond being "bench tests conducted." Given the company location (China), it's likely the tests were conducted in China. It is a prospective test, as it's part of the device's design verification.
  • Other tests (D6124, D412, D573, D3767, ISO 10993-5, ISO 10993-10): Specific sample sizes for each characteristic within these tests are not explicitly stated, but the results are reported as averages or meeting minimums across sizes. It is implied that adequate samples per standard test method were used. Data provenance is similar to D5151.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This document describes safety and performance testing for a physical medical device (gloves) against established international standards (ASTM, ISO). There is no "ground truth" derived from expert interpretation (like medical image analysis) involved in these non-clinical tests. The tests themselves provide objective measurements against defined criteria.

4. Adjudication method for the test set

  • Not Applicable. As mentioned above, these are objective physical and chemical tests, not involving human interpretation or adjudication. The results are quantitative measurements against predefined thresholds.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not an AI device, and therefore, no MRMC study or AI assistance evaluation was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is not an AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Standardized Test Methods and Criteria: The "ground truth" for these tests are the objective, pre-defined acceptance criteria specified by the relevant international standards (ASTM D5250, D5151, D6124, ISO 10993-5, ISO 10993-10). These standards themselves represent a consensus of expert knowledge in materials science, medical device design, and biocompatibility, but not in the context of diagnostic interpretation.

8. The sample size for the training set

  • Not Applicable. This is not an AI device.

9. How the ground truth for the training set was established

  • Not Applicable. This is not an AI device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.

July 29, 2021

Hebei Titans Hongsen Medical Technology Co., LTD. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm. 912, Building 15, XiYueHui, No. 5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China

Re: K211229

Trade/Device Name: Vinyl compound examination gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: July 13, 2021 Received: July 19, 2021

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K211229

Device Name

Vinyl compound examination gloves

Indications for Use (Describe)

The Vinyl compound examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number: K211229

    1. Date of Preparation: 07/28/2021
    1. Sponsor

HEBEI TITANS HONGSEN MEDICAL TECHNOLOGY CO., LTD.

Eastern Industrial Zone, NANGONG, HEBEI, 051800, CHINA Contact Person: Feng Shuangyan Position: Quality Management Representative Tel: +86-13733285059 Email: STQAFSY@126.COM

    1. Submission Correspondent

Beijing Believe-Med Technology Service Co., Ltd.

Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China,102401 Contact Person: Ray Wang Position: General Manager Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang(@believe-med.com

    1. Proposed Device Identification
      Trade Name: Vinyl compound examination gloves Common Name: Vinyl Patient Examination Gloves (Powder Free)

Regulatory Information: Classification: I Product Code: LYZ Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

{4}------------------------------------------------

Indication For Use Statement:

The Vinyl compound examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    1. Predicate Device Identification
      510(k) Number: K182043

Product Name: Single-use medical poly (vinyl chloride) examination glove (Clear, Non-colored) Manufacturer: HEBEI TITANS HONGSEN MEDICAL TECHNOLOGY CO., LTD

6. Device Description

The proposed device, Vinyl compound examination gloves are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

The proposed device is Powder Free Vinyl Patient Examination Gloves. The proposed device is Blue. The design of proposed device is addressing the standards as ASTM D6124. ASTM D5151, and ASTM D5250.

The proposed device is not provided as sterilized

The proposed device is made of vinyl chloride.

DesignationSizeTolerance
XSSMLXL
Length, mm230230230230230min
Width, mm758595105115±5
Thickness, mm:
Finger0.08min
Palm0.08min

Table 1 Device Size Specifications

Table 2 Performance and Physical Specifications

Before AgingAfter AgingPinhole AQL
TensileStrengthUltimateElongationTensileStrengthUltimateElongation
11 MPa, min300 % min11 MPa, min300 % min4.0

The above data of size, performance, and physical specifications of proposed gloves meet all the

{5}------------------------------------------------

current specifications listed in the ASTM standard D5250.

7. Comparison of Technological Characteristics

ITEMProposed DevicePredicate Device (K182043)Remark
Vinyl compound examination gloves(K211229)Single-use medical poly (vinyl chloride)examination glove (Clear, Non-colored)
Product CodeLYZLYZSAME
Regulation No.21 CFR 880.625021 CFR 880.6250SAME
ClassIISAME
Intended UseThe Vinyl compound examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.The Single-use medical poly (vinyl chloride) examination glove (Clear, Non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.SAME
Powdered orPowered freePowdered freePowdered freeSAME
Table 1 General Comparison
Table 2 Device Dimensions Comparison
------------------------------------------------
DesignationSizeTolerance
XSSMLXL
Proposed Device (K211229)Length, mm230230230230230min
Width, mm758595105115$\pm$ 5
Thickness, mm:
Finger0.08min
Palm0.08min
DesignationSize
XSSMLXL
Predicate Device ((K182043)Length, mm240240240240240min
Width, mm808595105115$\pm$ 5
Thickness, mm:
Finger0.07min
Palm0.08min
RemarkAnalysis 1

Analysis 1:

The proposed device has different size specification to the predicate device, but all proposed devices are meet the specifications of ASTM D 5250.

So we consider this as the proposed device is as safe, as effective, and performs as well as the predicate

{6}------------------------------------------------

device.

ITEMProposed DeviceVinyl compound examination gloves(K211229)Predicate Device (K182043)Single-use medical poly (vinylchloride) examination glove (Clear,Non-colored)Remark
ColorantBlueClear, Non-coloredAnalysis 2
PhysicalPropertiesBeforeAgingTensileStrength11 MPa, min15 MPa, minAnalysis3
UltimateElongation300 % min350 % min
AfterAgingTensileStrength11 MPa, min15 MPa, minSAME
UltimateElongation300 % min350 % min
Freedom from HolesBe free from holes when tested inaccordance with ASTM D5151
Powder ContentLess than 2 mg per glove when testedin accordance with ASTM D6124
Table 3 Performance Comparison

Analysis 2:

The proposed device has different color to the predicate device, this different may causes potential biocompatibility risk, for this risk we conducted the biocompatibility test according to the ISO 10993-5 and ISO 10993-10, the test results showed that the proposed devices with blue colorant did not raise biocompatibility risk.

So we consider this as the proposed device is as safe, as effective, and performs as well as the predicate device.

Analysis 3:

The proposed device has different Tensile Strength before aging specification to the predicate device, but all proposed device meets the specification requirements of ASTM D 5250.

So we consider this as the proposed device is as safe, as effective, and performs as well as the predicate device.

ITEMProposed DevicePredicate Device (K182043)Remark
Vinyl compound examinationgloves (K211229)Single-use medical poly (vinylchloride) examination glove (Clear,Non-colored)

Table 4 Safety Comparison

{7}------------------------------------------------

MaterialVinyl chlorideVinyl chlorideSAME
BiocompatibilityCytotoxicityUnder conditions of the study, not a cytotoxicity.Under conditions of the study, not a cytotoxicity.SAME
IrritationUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritantSAME
SensitizationUnder conditions of the study, not a sensitizer.Under conditions of the study, not a sensitizer.SAME

Summary of Non-Clinical Tests 8.

Bench tests were conducted to demonstrate that the proposed device complies with the following standards:

ASTM D5250-19 Standard Specification for Poly (vinyl chloride) Gloves for Medical Application. ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. ASTM D6124-17, Standard Test Method for Residual Powder on Medical Gloves. ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity. ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

Test MethodPurposeAcceptance CriteriaResults
ASTM D5151Testing for Freedom from holesFreedom from holesNo water leakage is inspected form 200 samples
ASTM D6124Determine the powder residue for powder free gloves<2.0 mg per gloveResidual Powder of Size XS: Average 0.28 mg;Residual Powder of Size S: Average 0.25 mg;Residual Powder of Size M: Average 0.31 mg;Residual Powder of Size L: Average 0.25 mg;Residual Powder of Size XL: Average 0.35mg;
ASTM D412ASTM D573Testing for Physical property characteristicsTensile Strength: 11 MPa min.Ultimate Elongation: 300% min.Tensile Strength: ≥ 12 MPa;Ultimate Elongation: ≥ 321%.
ASTM D3767Testing For physical dimensions specificationLength: 230 mm min. for all size (XS, S, M, L, XL);Width: 75±5 mm for XS; 85±5 mm for S;95±5 mm for M; 105±5 mm for L; 115±5 mm for XL.Finger Thickness: ≥0.08 mm;Palm Thickness: ≥0.08 mm;All acceptance criteria above meet theLength of Size XS: ≥ 237 mm;Width of Size XS: 75±1 (74-76) mm;Palm Thickness of Size XS: ≥0.08 mm;Finger Thickness of Size XS: ≥0.09 mm.Length of Size S: ≥ 235 mm;Width of Size S: 85±1 (85-86) mm;

Table 5 Performance Test Results Summary

{8}------------------------------------------------

requirements in Table 2 Dimensions andPalm Thickness of Size S: ≥0.08 mm;
Tolerances of ASTM D5250Finger Thickness of Size S: ≥0.09 mm.
Length of Size M: ≥ 235 mm;
Width of Size M: $95±2$ (95-97) mm;
Palm Thickness of Size M: ≥0.08 mm;
Finger Thickness of Size M: ≥0.09 mm.
Length of Size L: ≥ 235 mm;
Width of Size L: $105±2$ (105-107) mm;
Palm Thickness of Size L: ≥0.08 mm;
Finger Thickness of Size L: ≥0.09 mm.
Length of Size XL: ≥ 234 mm;
Width of Size XL: $115±3$ (115-118) mm;
Palm Thickness of Size XL: ≥0.08 mm;
Finger Thickness of Size XL: ≥0.09 mm.
ISO 10993-5Evaluate the endpoint ofCytotoxicity forbiocompatibilityThe test article showed "negative"cytotoxicityUnder the conditions of the study, the test articleshowed "negative" cytotoxicity.
ISO 10993-10Evaluate the endpoint ofirritant forbiocompatibilityThe response of the test article extract isnegligible.The test result showed that the response of thetest article extract was categorized as negligibleunder the test condition.
Evaluate the endpoint ofsensitization forbiocompatibilityThe test article showed no evidence ofcausing delayed dermal contactsensitization.The test article showed no evidence of causingdelayed dermal contact sensitization in theguinea pig.

9. Summary of Clinical Test

No clinical study is included in this submission.

10. Conclusion

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Proposed Device Vinyl compound examination gloves cleared under K182043.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.