K Number
K211152
Manufacturer
Date Cleared
2022-05-11

(387 days)

Product Code
Regulation Number
884.5300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Durex Condom with Benzocaine is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).

Device Description

The Durex Condom with Benzocaine is a natural rubber latex-based condom used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs). This is a textured condom with a silicone-based lubricant. Durex Condom with Benzocaine has a nominal length of 195 mm and width of 56 mm with a thickness of 65 um. The Durex Condom with Benzocaine is packaged with 480 mg of silicone lubricant. In addition, the Benzocaine male genital desensitizer paste on the condom helps in temporarily prolonging the time until ejaculation. This condom is a surface device that comes into contact with skin and mucosal membranes. Durex Condom with Benzocaine is packaged in individually sealed flexible laminate foils made of polyethylene terephthalate, polyethylene, and aluminum. The foils come packaged in an outer consumer cardboard carton. The Durex Condom with Benzocaine is intended for over-the-counter (OTC) use. These condoms conform with FDA-recognized standards ASTM D3492-16 and ISO 4074:2015.

AI/ML Overview

The provided document is a 510(k) summary for the Durex Condom with Benzocaine, which is a medical device. This type of document is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving effectiveness in the same way a clinical trial for a drug or a novel medical device would.

Therefore, the study described here is about confirming that the condom meets specific performance standards and is biocompatible, rather than conducting a comparative effectiveness study with human readers or standalone algorithm performance.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard)Reported Device Performance
Physical Performance:
ASTM D3492-16 (Standard Specification for Rubber Contraceptives)Met all requirements of ASTM D3492-16
ISO 4074:2015 (Natural rubber latex male condoms – Requirements and test methods)Met all requirements of ISO 4074:2015
Air Burst Pressure: > 1.0 kPa (per ISO 4074:2015/ASTM D3492-16)> 1.0 kPa
Air Burst Volume: $\geq 18.0$ L (per ISO 4074:2015/ASTM D3492-16)$\geq 18.0$ L
Nominal Width: 56 $\pm$ 2 mm56 $\pm$ 2 mm
Nominal Length: 195 $\pm$ 10 mm195 $\pm$ 10 mm
Nominal Thickness: 0.065 $\pm$ 0.01 mm0.065 $\pm$ 0.01 mm
Lubricant Quantity: 480 $\pm$ 50 mg480 $\pm$ 50 mg
Benzocaine Quantity: $140 \pm 20$ mg (5% Benzocaine)$140 \pm 20$ mg (5% Benzocaine)
Biocompatibility:
Cytotoxicity (ISO 10993-5:2009/R 2014)Device is non-cytotoxic (implied by meeting general biocompatibility conclusion)
Sensitization (ISO 10993-10:2010/R 2014)Device is non-sensitizing
Vaginal Irritation (ISO 10993-10:2010/R 2014)Device is non-irritating
Acute Systemic Toxicity (ISO 10993-11:2017)Device is not acutely, systemically toxic
Shelf Life:
Accelerated stability evaluations (per 21 CFR 801.435)All samples met predefined acceptance criteria for a 5-year shelf life

2. Sample Size for the Test Set and Data Provenance

The document does not explicitly state the exact sample sizes (number of condoms) used for each specific test (e.g., air burst, dimensions, biocompatibility assays). It generally states "All samples met predefined acceptance criteria" for shelf-life testing and that the device "was tested and met all the requirements" of the specified standards.

Data provenance is not detailed in terms of country of origin or whether it was retrospective/prospective. These are standard performance tests for condom manufacturing and regulatory submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This type of information is not relevant to the evaluation of a condom's physical and biological performance tests. The "ground truth" here is defined by the established international and national standards (ASTM D3492-16, ISO 4074:2015, ISO 10993 series). The tests are objective and measurable, not based on expert interpretation of complex data like medical images.

4. Adjudication Method for the Test Set

Not applicable. The performance criteria are objective measurements against defined physical and biological specifications in recognized standards. There is no subjective interpretation requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, an MRMC comparative effectiveness study was not done. This device is a condom, and the studies described are performance and safety (biocompatibility) tests against industry standards, not studies involving human readers or comparative effectiveness in a clinical diagnostic sense. The concept of "human readers improve with AI vs without AI assistance" does not apply here.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Not applicable. This is not an AI-powered diagnostic or interpretive device.

7. The Type of Ground Truth Used

The "ground truth" is established by:

  • International and National Standards: Specifically, ISO 4074:2015 and ASTM D3492-16 for physical performance, and the ISO 10993 series for biocompatibility.
  • Defined Specifications: The specific nominal dimensions, lubricant quantities, and benzocaine quantities (e.g., 56 $\pm$ 2 mm nominal width, 480 $\pm$ 50 mg lubricant).
  • Regulatory Guidance: FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and 21 CFR 801.435 for shelf life.

8. The Sample Size for the Training Set

Not applicable. This is not a machine learning or AI device that requires a "training set." The tests are empirical and based on physical and chemical properties.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set in this context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 11, 2022

RB Health (US) LLC Kyle Prince Regulatory Specialist 399 Interpace Parkway Parsippany, NJ 07054

Re: K211152

Trade/Device Name: Durex Condom with Benzocaine Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: April 7, 2022 Received: April 8, 2022

Dear Kyle Prince:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211152

Device Name Durex Condom with Benzocaine

Indications for Use (Describe)

The Durex Condom with Benzocaine is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summarv K211152 Durex Condom with Benzocaine

1. Submitter Information

Applicant:RB Health (US) LLC
Address:399 Interpace Parkway ParsippanyNJ 07054-0024
Phone:(862) 325-0012

2. Correspondent Information

Company:RB Health (US) LLC
Contact:Kyle Prince
Phone:(973) 404-2600
Email:kyle.prince@rb.com

3. Date prepared: May 9, 2022

4. Device Information

Device Name:Durex Condom with Benzocaine
Common Name:Male Natural Rubber Latex Condom
Regulation Number:21 CFR 884.5300
Regulation Name:Condom
Product Code:HIS (Condom)
Regulatory Class:Class II

5. Predicate Device Information

Device Name:Durex Latex Condom with Male Genital Desensitizer Lubricant
510(k) Number:K020659
Sponsor:Ssl Americas, Inc.

The predicate device has not been subject to a design-related recall.

6. Device Description

The Durex Condom with Benzocaine is a natural rubber latex-based condom used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs). This is a textured condom with a silicone-based lubricant. Durex Condom with Benzocaine has a nominal length of 195 mm and width of 56 mm with a thickness of 65 um. The Durex Condom with Benzocaine is packaged with 480 mg of silicone lubricant. In addition, the Benzocaine male genital desensitizer paste on the condom helps in temporarily prolonging the time until ejaculation. This condom is a surface device that comes into contact with skin and mucosal membranes. Durex Condom with Benzocaine is packaged in individually sealed flexible laminate foils made of polyethylene terephthalate, polyethylene, and aluminum. The foils come packaged in an outer consumer cardboard carton. The Durex Condom with Benzocaine is intended for over-the-counter (OTC) use. These condoms conform with FDA-recognized standards ASTM D3492-16 and ISO 4074:2015.

{4}------------------------------------------------

Device specifications are listed in Table 1 below.

7. Indications for Use Statement

The Durex Condom with Benzocaine is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).

8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

Table 1. below includes a comparison of the intended use and technological characteristics of the subject and predicate devices.

Subject DevicePredicate Device
Durex Condom withBenzocaineK211152Durex Latex Condom WithMale Genital DesensitizerLubricantK020659
Device & Predicate DeviceDurex Condom withBenzocaineDurex Latex Condom WithMale Genital DesensitizerLubricant
510(K) NumberK211152K020659
Product CodeHISHIS
Regulation Number21 CFR 884.530021 CFR 884.5300
Regulation NameCondomCondom
Indications for UseThe Durex Condom withBenzocaine is used forcontraception and forprophylactic purposes (tohelp reduce the risk ofpregnancy and thetransmission of sexuallytransmitted infections, STIs).The DUREX® LatexCondom with Male GenitalDesensitizer Lubricant is usedfor contraception and forprophylactic purposes (tohelp reduce the risk ofpregnancy and thetransmission of sexuallytransmitted infections, STIs.The male genital desensitizerlubricant on the condomhelps in temporarilyprolonging the time untilejaculation.
Prescription or Over-The-Counter UseOver-The-CounterOver-The-Counter
Condom MaterialNatural Rubber LatexNatural Rubber Latex
Nominal Width56 $\pm$ 2 mm52 $\pm$ 2 mm
Nominal Length195 $\pm$ 10 mm180 $\pm$ 10 mm
Nominal Thickness0.065 $\pm$ 0.01 mm0.065 mm minimum
LubricantSilicone gel withTransatak odormaskerSilicone gel withZeus odor masker
Lubricant Quantity480 $\pm$ 50 mg400 $\pm$ 50 mg

{5}------------------------------------------------

Benzocaine Quantity$140 \pm 20$ mg5% Benzocaine140 mg 4.75 – 5.25%Benzocaine
Air Burst Pressure> 1.0 kPa> 1.0 kPa
Air Burst Volume$\geq 18.0$ L$\geq 18.0$ L
SterilizationNon-SterileNon-sterile
TextureNipple ended, texturedcondomStraight wall, nipple ended,smooth condom
Shelf Life5 Years5 Years

Table 1. Comparison of intended use and technological characteristics of the subject and predicate device

The subject and predicate device have similar indications for use and have the same intended use - for contraception and for prophylactic purposes. The technological characteristics of the subject device and predicate device are similar in that they are natural rubber latex-based, include Benzocaine in similar quantities, are lubricated with silicone, and have the same shelf-life duration. The subject and predicate devices have different technological characteristics, including different dimensions, form, and specifications (e.g., burst volume). However, the different technological characteristics of the subject and predicate devices do not raise different questions of safety and effectiveness.

9. Summary of Non-Clinical Performance Testing

Biocompatibility:

Biocompatibility studies, including Acute Systemic Toxicity, Vaginal Irritation Testing, Cytotoxicity and Sensitization testing were performed in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993-1:2009 as follows:

  • · Cytotoxicity (ISO 10993-5:2009/R 2014)
  • · Sensitization (ISO 10993-10:2010/R 2014)
  • · Vaginal Irritation (ISO 10993-10:2010/R 2014)
  • · Acute Systemic Toxicity (ISO 10993-11:2017)

The results of testing demonstrate that the subject device is non-irritating, non-sensitizing, and not acutely, systemically toxic.

Physical Performance Testing:

The Durex Condom with Benzocaine was tested and met all the requirements of ISO 4074:2015 -Natural rubber latex male condoms – Requirements and test methods and ASTM D3492-16 - Standard Specification for Rubber Contraceptives (Male Condoms).

Shelf Life:

The Durex Condom with Benzocaine has a five-year shelf life based on the results of accelerated stability evaluations conducted as required in 21 CFR 801.435. All samples met predefined acceptance criteria.

{6}------------------------------------------------

10. Conclusion

The results of the performance testing described above demonstrate that the Durex Condom with Benzocaine is as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.5300 Condom.

(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.