K Number
K211100

Validate with FDA (Live)

Date Cleared
2021-05-14

(31 days)

Product Code
Regulation Number
870.1280
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AcQGuide MAX Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The AcQGuide MAX deflection facilitates catheter positioning.

Device Description

The AcQGuide® MAX Steerable Sheath, Model 900200 is a single use, percutaneous catheter introducer designed to provide additional maneuverability to interventional catheters that are advanced through the sheath and into the right or left chambers of the heart. The distal portion of the sheath is comprised of a composite structured single lumen shaft. At the proximal end, an ergonomic handle provides torque and active deflection, a hemostasis valve allows safe introduction of an interventional catheter, and a side port provides access for aspiration, fluid flushes and fluid/medication infusions. The dilator is designed to introduce the Steerable Sheath into the vasculature and into chambers of the heart. The dilator has a smooth tapered tip and provides a smooth transition to the round edge of the non-traumatic tip of the sheath. The dilator is able to track over a 0.035″ guidewire. The hub section is attached to the shaft. It is a standard female luer made of high-density polyethylene (HDPE) material.

AI/ML Overview

This submission details the AcQGuide MAX ASteerable Sheath (Model 900200), a device for introducing catheters into the vasculature and heart chambers. The submission seeks to demonstrate substantial equivalence to a previously cleared predicate device (AcQGuide Steerable Sheath, Model 900002).

Here's an analysis of the provided information regarding acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state quantitative acceptance criteria in a formal table format. Instead, it describes general performance goals and confirms that testing was conducted to ensure the device "conformed to the design specifications" and "met its established performance specifications." The key performance attributes evaluated are listed, and the conclusion is that the device "meets the specifications necessary for consistent performance."

Thus, the acceptance criteria are implicitly tied to meeting design specifications and established performance specifications, as demonstrated by passing various nonclinical tests.

Acceptance Criteria (Implied)Reported Device Performance
Conformance to design specificationsCollective results of nonclinical testing demonstrate the materials, manufacturing processes, and design meet necessary specifications for consistent performance.
Meeting established performance specificationsDevice met established performance specifications, verified by dimensional testing, visual inspection, leak testing, functional and compatibility testing, mechanical testing, and handle torque testing.
Consistent performance during intended useCollective results of nonclinical testing affirm consistent performance during intended use.
No potential impact on safety or effectivenessCollective bench testing demonstrates no potential impact on safety or effectiveness for percutaneous catheter introduction into the vasculature and heart chambers, compared to the predicate device.
Packaging and shelf-life requirements metStudies performed to confirm acceptable product performance and ability to withstand sterilization; shelf-life difference (12 months vs. 18 months for predicate) does not raise new safety/effectiveness questions.
Sterilization effective and safeTechnical review established the subject device can be adopted into the existing EtO processing cycle validated to a 10^6 sterility assurance level (SAL). Materials are suitable for EtO sterilization and not known to retain high EtO residual levels.
Biocompatibility in accordance with standardsPassed appropriate biocompatibility tests in accordance with ISO 10993-1:2018 and FDA Guidance. Categorized as an external communicating device with limited (≤24 hours) contact with circulating blood.
Non-pyrogenicPerformed material mediated pyrogenicity testing in New Zealand white rabbits per 21 CFR Part 58; considered non-pyrogenic per ISO 10993-11 guidelines.
Dimensional accuracy of componentsDimensional testing ensured all components met required dimensions per specifications.
Absence of defectsVisual inspection ensured all components were free of defects.
Packaging integrityLeak testing ensured packaging meets its specifications. The packaging system does not adversely affect gas ingress or egress, nor does it pose a risk to product heating.
Functional and compatibility requirements metFunctional and compatibility testing ensured requirements are met.
Mechanical strength and durabilityMechanical testing ensured components can withstand expected forces.
Handle torque resistanceHandle torque testing ensured it can withstand expected torques.
Deflection and Reach (Identical to Predicate)Maximum Deflection: 180 degrees; Reach: 5.0 cm at 90 degrees.
Guidewire compatibility (Identical to Predicate)0.035".
Radiopaque marker visibility (Identical to Predicate)Distal sections visible with standard or low-level fluoroscopy.
Hemostasis valve function (Identical to Predicate)Prevents air incursion and minimizes blood loss. The differences in hemostasis valve material do not raise new safety/effectiveness questions.
Flush port function (Identical to Predicate)Provides ability to administer saline flush throughout procedure.
Sheath deflection (Identical to Predicate)Facilitates accurate catheter positioning.
Usable with catheters up to 12 Fr (Improved from Predicate 10.5 Fr)Facilitates use with catheters up to 12 Fr in diameter.

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not specify exact sample sizes for each nonclinical test (e.g., how many units were tested for dimensional inspection, leak testing, etc.). It generally states that "All necessary testing was conducted" and "The necessary bench testing was performed."
  • Data Provenance: The testing was conducted by or for Acutus Medical, Inc. in the U.S. (Carlsbad, CA). The studies are nonclinical bench tests and "Physician simulated use in an animal model." This would be considered prospective data for device verification.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided. The "Physician simulated use in an animal model" implies expert involvement, but the number and qualifications of these physicians are not detailed. For bench testing, human experts typically set the specifications, but they don't "establish ground truth" for test outcomes in the same way clinical data is adjudicated.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable/not provided. Adjudication methods like 2+1 or 3+1 are typically used for clinical studies involving interpretation of medical images or outcomes where expert consensus is needed to establish ground truth. As this is a nonclinical bench testing and simulated animal model study, such adjudication methods would not be relevant.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. This submission is for a medical device (steerable sheath), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed.

6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done:

This is not applicable. This device is a physical medical instrument, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For the nonclinical bench tests, the "ground truth" is defined by the design specifications and established performance specifications of the device. These are objectively measurable parameters (e.g., dimensions, force resistance, electrical properties, etc.). For biocompatibility and sterilization, the "ground truth" refers to meeting the requirements of established international standards (ISO 10993-1, ISO 11135:2014+A1:2019) and regulations (21 CFR Part 58), which are objectively tested against. The "Physician simulated use in an animal model" would likely have qualitative or quantitative outcomes based on the successful demonstration of the device's function in a physiological setting, but the specific metrics are not detailed here.

8. The sample size for the training set:

This is not applicable. This is a non-AI medical device; therefore, there is no "training set."

9. How the ground truth for the training set was established:

This is not applicable. As there is no training set, there is no ground truth for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

May 14, 2021

Acutus Medical, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K211100

Trade/Device Name: AcQGuide MAX Steerable Sheath Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable catheter Regulatory Class: Class II Product Code: DRA Dated: April 9, 2021 Received: April 13, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh Deoras Acting Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211100

Device Name AcQGuide MAX Steerable Sheath

Indications for Use (Describe)

The AcQGuide MAX Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart.

The AcQGuide MAX deflection facilitates catheter positioning.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K21100

510(k) SUMMARY

510(k) Number: K211100

Date Prepared: May 11, 2021

SUBMITTER INFORMATION [807.92(a)(1)]

Manufacturer: Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008, U.S.A Phone: +1-442-232-6080 Fax: +1-442-232-6081

Contact Person:

Serena Sanginthirath Senior Regulatory Affairs Specialist Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008, U.S.A Phone: +1-442-232-6178 Fax: +1-442-232-6081 E-mail: Serena.Sanginthirath@acutus.com

DEVICE INFORMATION [807.92(a)(2)]

Trade NameAcQGuide® MAX Steerable Sheath
Common NameSteerable Sheath
Classification NameCatheter, Steerable
Regulation21 CFR § 870.1280
Product CodeDRA
Regulatory Classification:Class II
Device Panel:Cardiovascular

The Acutus AcQGuide® MAX Steerable Sheath, Model 900200 is substantially equivalent to the predicate device, AcQGuide Steerable Sheath, Model 900002. Neither of these have been subject to a design-related recall.

{4}------------------------------------------------

PREDICATE DEVICE [807.92(a)(3)]

The predicate device to support substantial equivalence of the AcQGuide® MAX Steerable Sheath, Model 900200 is the Acutus Medical, Inc. AcQGuide Steerable Sheath, Model 900002, cleared under 510(k) K162925.

Predicate DeviceManufacturerFDA 510(k)
AcQGuide Steerable SheathAcutus Medical, Inc.K162925

DEVICE DESCRIPTION [807.92(a)(4)]

The AcQGuide® MAX Steerable Sheath, Model 900200 is a single use, percutaneous catheter introducer designed to provide additional maneuverability to interventional catheters that are advanced through the sheath and into the right or left chambers of the heart. The distal portion of the sheath is comprised of a composite structured single lumen shaft. At the proximal end, an ergonomic handle provides torque and active deflection, a hemostasis valve allows safe introduction of an interventional catheter, and a side port provides access for aspiration, fluid flushes and fluid/medication infusions.

The dilator is designed to introduce the Steerable Sheath into the vasculature and into chambers of the heart. The dilator has a smooth tapered tip and provides a smooth transition to the round edge of the non-traumatic tip of the sheath. The dilator is able to track over a 0.035″ guidewire. The hub section is attached to the shaft. It is a standard female luer made of high-density polyethylene (HDPE) material.

Some of the key design attributes of the Steerable Sheath include:

  • Deflectability
  • Hemostasis
  • Kink-resistance
  • Visibility under fluoroscopy
  • Easily flushed during the procedure
  • Single-handed operation
  • Biocompatible materials
  • Sterile, single use
  • Combined three-way stopcock

INDICATIONS FOR USE [807.92(a)(5)]

The AcQGuide MAX Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart.

The AcQGuide MAX deflection facilitates catheter positioning.

{5}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject device AcQGuide® MAX Steerable Sheath, Model 900200 has the same intended use and fundamental scientific technology as the predicate device. The purpose of this submission is to introduce the modified version of the Steerable Sheath and eventually replace the cleared AcQGuide® MAX Steerable Sheath, Model 900002 with the subject device. Table 1 provides a comparison of the predicate AcQGuide® MAX Steerable Sheath, Model 900002 classification against the subject device, comparison of the indications for use, and a comparison of the technological characteristics.

{6}------------------------------------------------

TABLE 1: COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Table 1: COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
CharacteristicsPredicate DeviceAcQGuide® MAX Steerable Sheath, Model 900002(K162925)Subject DeviceAcQGuide® MAX Steerable Sheath, Model 900200
510(k) NumberK162925TBD
Classification21 CFR 870.1280Identical
Product CodeDRAIdentical
Indications for UseThe AcQGuide is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The AcQGuide deflection facilitates catheter positioning.Identical indications for use. Adding the product brand name "MAX" do not raise a new or different question of safety and effectiveness.
Device IdentificationThe AcQGuide is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The AcQGuide deflection facilitates catheter positioning.
Key ComponentsEmploys single deflection curve with atraumatic tip. Sheath deflection facilitates accurate catheter positioning. Facilitates use with catheters up to 10.5 Fr in diameter. Hemostasis valve to prevent air incursion and minimizes blood loss. Flush port provides ability to administer saline flush throughout procedure. Usable length of up to 75 cm.Identical key components with regards to sheath deflection, use with catheters up to 12 Fr in diameter, hemostasis valve and flush port. Usable length of up to 70 cm. The minor difference in sheath usable length do not raise a new or different question of safety and effectiveness in the subject device.
Sheath SizeInner Diameter: 12.8 Fr (4.3 mm)Outer Diameter: 16 Fr (5.3 mm)Usable Length: 75 cmTotal Length: 92 cmRadiopaque Marker: distal sections are visible when using standard or low-level fluoroscopyInner Diameter: 12.4 Fr (4.1 mm)Outer Diameter: 15.2 Fr (5.1 mm)Usable Length: 70 cmTotal Length: 85 cmIdentical radiopaque marker for visualization when using fluoroscopy.The minor differences in sheath size do not raise a new or different question of safety and effectiveness in the subject device.
Sheath MaterialShaft: Biocompatible copolymer (Pebax®) with stainless steel, platinum/10% Iridium, PTFEShaft: PTFE Liner for main lumen Pebax 4033 with 20-25% BaSO4, Pantone 422C, Pebax 5533 Pantone 419C, Pebax 3533 Pantone 422C, Pebax 6333 Pantone 422C, Pebax 7233 Pantone 422C.

{7}------------------------------------------------

Table 1: COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
CharacteristicsPredicate DeviceAcQGuide® MAX Steerable Sheath, Model 900002(K162925)Subject DeviceAcQGuide® MAX Steerable Sheath, Model 900200
Both devices are constructed of biocompatible materials and PTFE Linerfor main lumen (Pebax). The differences in color additives do not raise anew or different question of safety and effectiveness in the subjectdevice.
Hemostasis ValveMaterial: Shinetsu KEG-2000 50 A/BMaterial: Silicone .080 NVR 50D (100578). The differences in hemostasisvalve material do not raise a new or different question of safety andeffectiveness in the subject device.
Radiopaque MarkerTotal Length: .040" (1.0 mm)Total Length: .039" (0.99 mm)
DilatorTotal Length: 95.5 cmOuter Diameter: 4.1 mmInner Diameter: .039" (0.99 mm)Total Length: 87.4 cmOuter Diameter: 4 mmInner Diameter: .039" (0.99 mm)The differences in total length and outer diameter of the dilator do notraise a new or different question of safety and effectiveness in thesubject device.
Steerable SheathTotal Weight69.2 g75 g. The difference in sheath total weight do not raise a new ordifferent question of safety and effectiveness in the subject device.
Deflection andReachMaximum Deflection: 180 degreesReach: 5.0 cm at 90 degreesIdentical
CompatibilityGuidewire compatibility: .035"Identical
Method of UsePercutaneous catheter introduction into the vasculature and into thechambers of the heart. The sheath deflection facilitates catheterpositioning.Same method of use.
PackagingPouch: Tyvek® 1073B Tyvek / 30655-O FilmPouch Card Tray: .035" HDPE, WhiteStrap: .028" HDPE, WhitePouch: Tyvek® 1073B Tyvek / 30655-O FilmPouch Card Tray: .035" HDPE, WhiteStrap: .028" HDPE, White
Table 1: COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
CharacteristicsPredicate DeviceAcQGuide® MAX Steerable Sheath, Model 900002(K162925)Subject DeviceAcQGuide® MAX Steerable Sheath, Model 900200
EO Process Indicator Label: NAMSA radiation CPIs are non-odorous andnon-hazardous to useBlank Label: Avery 8126 or equivalent6 Pack Box: 200C Raft RSCCatheter Box (single): 200-E #W-BSEO Process Indicator Label: NAMSA radiation CPIs are non-odorous andnon-hazardous to useBlank Label: Avery 8126 or equivalent6 Pack Box: 200C Raft RSCCatheter Box (single): 200-E #W-BSIdentical packaging system.
SterilizationEthylene Oxide (EtO)Identical
Shelf Life18 months12 months. The difference in shelf-life labeling do not raise a new ordifferent question of safety and effectiveness in the subject device.

{8}------------------------------------------------

{9}------------------------------------------------

PERFORMANCE DATA [807.92(b)]

All necessary testing was conducted on the AcQGuide® MAX Steerable Sheath, Model 900200 to support a determination of substantial equivalence to the predicate device.

NONCLINICAL TESTING SUMMARY - BENCH [807.92(b)(1)]

The necessary bench testing was performed on the AcQGuide® MAX Steerable Sheath, Model 900200 in order to ensure that it conformed to the design specifications and to support a determination of substantial equivalence to the predicate device. The bench testing performed on the AcQGuide® MAX Steerable Sheath, Model 900200 consisted of the following:

  • Packaging and shelf-life
  • Sterilization
  • Biocompatibility
  • Design Verification and Validation studies, including:
    • Dimensional inspection of Sheath and Dilator O
    • Visual inspection for any defects O
    • Leak testing O
    • O Functional and compatibility testing
    • Mechanical testing O
    • o Handle torque testing
  • Physician simulated use in an animal model ●

The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the AcQGuide® MAX Steerable Sheath, Model 900200 meets the specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the AcQGuide® MAX Steerable Sheath, Model 900200 does not have any potential impact on the safety or effectiveness for percutaneous catheter introduction into the vasculature and into chambers of the heart, when compared to the predicate device.

The AcQGuide® MAX Steerable Sheath, Model 900200 was tested to verify that the device met its established performance specifications. This was done by utilizing dimensional testing to ensure that all components met the required dimensions per specifications, visual inspection to ensure that all components were free of defects, leak testing to ensure the packaging meets its specifications, functional and compatibility testing to ensure the functional and compatibility requirements are met, mechanical testing to ensure the components are able to withstand the expected forces, and handle torque testing to ensure it can withstand the expected torques.

{10}------------------------------------------------

CLINICAL TESTING SUMMARY [807.92(b)(2)]

This section is not applicable. No clinical testing is being submitted to support review of this 510(k) premarket notification.

BIOCOMPATIBILITY

The AcQGuide® MAX Steerable Sheath, Model 900200 is manufactured from materials similar to the predicate AcQGuide® MAX Steerable Sheath, Model 900002, and have passed all of the appropriate biocompatibility tests. Biocompatibility testing for the AcQGuide® MAX Steerable Sheath, Model 900200 was conducted in accordance with ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process and the FDA Guidance Document titled, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" issued on September 4, 2020. The AcQGuide® MAX Steerable Sheath, Model 900200 has three components: a composite structured single lumen shaft, an ergonomic handle to que and active deflection, and a hemostasis valve with side flush port. Each component was evaluated per the above standards and guidance documents to determine the appropriate biocompatibility testing. Per ISO 10993-1:2018, the AcQGuide® MAX Steerable Sheath is categorized as an external communicating device, which comes into contact with circulating blood for a limited (≤24 hours) contact duration.

STERILIZATION

Acutus Medical, Inc. conducted a technical review to compare the subject device AcQGuide® MAX Steerable Sheath, Model 900200 and the predicate device AcQGuide® MAX Steerable Sheath, Model 900002, to establish that the subject device can be adopted into the existing AcQGuide EtO processing cycle number ACU-537 at Steris Corporation. The predicate device AcQGuide® MAX Steerable Sheath, Model 900002 has been validated to an ethylene oxide sterilization process for a 106 sterility assurance level (SAL). The following aspects were used to perform the evaluation in accordance with ISO 11135:2014+A1:2019, packaging system, product design/materials, sterile barrier system, load configuration and bioburden resistance for the subject and predicate devices. Both the subject AcQGuide® MAX Steerable Sheath, Model 900200 and the predicate AcQGuide® MAX Steerable Sheath, Model 900002 product packaging systems use the same pouch, which consists of a sealed EtO compatible 1073B Tyvek pouch, 0.035" HDPE backer card and a secondary packaging (catheter shelf box). Additionally, both the subject and predicate devices are packaged in a corrugated cardboard shipper box, and uses the same packaging system, which contains (n=6) sheaths per shipper. The product packaging system does not adversely affect gas ingress or egress, nor does it pose a risk to product heating. The device is labeled and marketed as "Sterile" by Acutus Medical, Inc.

Shelf life study was performed to confirm acceptable product performance and ability to withstand sterilization. All materials identified were suitable for EtO sterilization and were not known to retain high EtO residual levels, or adversely impact the ability of EtO process

{11}------------------------------------------------

parameters to be met. In addition, Acutus Medical, Inc. performed material mediated pyrogenicity testing, to evaluate its potential to produce a pyrogenic response when tested in New Zealand white rabbits, per 21 CFR Part 58 Compliance Good Laboratory Practice for Nonclinical Laboratory Studies. The subject AcQGuide® MAX Steerable Sheath, Model 900200 is considered non-pyrogenic per the requirements of ISO 10993-11 guidelines.

ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY (EMC)

This section is not applicable as the AcQGuide® MAX Steerable Sheath, Model 900200 is non powered and does not contain any electrical components.

conclusions [807.92(p)(3)]

Extensive nonclinical testing has been performed on the AcQGuide® MAX Steerable Sheath, Model 900200 to evaluate the overall performance of the subject device is substantially equivalent to the predicate device.

§ 870.1280 Steerable catheter.

(a)
Identification. A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.(b)
Classification. Class II (performance standards).