K Number
K210957

Validate with FDA (Live)

Date Cleared
2021-08-31

(154 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Kubtec XPERT 80 is a Cabinet x-ray system that is used to provide film and/or digital x-ray images of harvested specimens from various regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure. It is not used for mammography

Device Description

The XPERT 80 Specimen Radiography System is a Cabinet X-ray System specifically designed to provide high detail radiographic imaging of surgically excised medical specimens. The exceptionally high magnification capability (up to 5X) from the 0.02 mm focal spot with optimized cabinet geometry and the superior contrast available from the low kV capability provides enhanced film and/or digital imaging performance. This device supports radiographic film sizes up to 30 x 35 cm and can be configured to acquire high resolution, DICOM compliant, digital x-ray images through the use of an integrated camera and Kubtec DIGICOM Specimen Radiography software. It has been designed to comply with 21 CFR 1020.40.

AI/ML Overview

The provided text is a 510(k) Summary for the Kubtec XPERT 80 Specimen Radiography System. It details the device, its intended use, and its substantial equivalence to predicate devices, but does not contain information about the acceptance criteria or a study proving the device meets those criteria, especially in the context of AI performance or human-in-the-loop studies.

The document discusses:

  • Device Description: The XPERT 80 Specimen Radiography System is a Cabinet X-ray System for high-detail radiographic imaging of surgically excised medical specimens.
  • Comparison to Predicate Devices: The key difference from the primary predicate (K071233) is the addition of an Amorphous Selenium detector. The document asserts that this new detector's performance has been verified and cleared via 510(k) K193166 as incorporated into the Siemens MAMMOMAT Revelation.
  • Standards Compliance: The device successfully completed internal and external safety testing to various IEC, ISO, NEMA, and CFR standards related to medical electrical equipment, software life cycle processes, risk management, and radiation safety.
  • Conclusion: The device is considered as safe and effective as the predicate device, with technological differences limited to the detector type, and no new indications for use.

Therefore, I cannot provide the requested information from the given text as it does not describe:

  1. A table of acceptance criteria and reported device performance.
  2. Sample size used for a test set or data provenance for such a test.
  3. Number of experts or their qualifications.
  4. Adjudication methods.
  5. MRMC comparative effectiveness study or human-AI improvement effect size.
  6. Standalone (algorithm-only) performance.
  7. Type of ground truth used (expert consensus, pathology, outcomes data, etc.).
  8. Sample size for the training set.
  9. How ground truth for the training set was established.

The document primarily focuses on demonstrating substantial equivalence based on technical specifications and compliance with safety standards rather than a performance study involving AI or human interpretation of images.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a caduceus, which is a traditional symbol of medicine. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The logo is used to identify the FDA and its role in regulating food and drugs in the United States.

KUB Technologies, Inc. % Chester Lowe, Ph.D. Chief Technology Officer 111 Research Drive STRATFORD CT 06615

Re: K210957

Trade/Device Name: Kubtec Xpert 80 Specimen Radiography System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MWP Dated: August 1, 2021 Received: August 11, 2021

Dear Dr. Lowe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

August 31, 2021

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara
For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210957

Device Name KUBTEC XPERT 80 Specimen RADIOGRAPHY SYSTEM

Indications for Use (Describe)

The Kubtec XPERT 80 is a Cabinet x-ray system that is used to provide film and/or digital x-ray images of harvested specimens from various regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure.

It is not used for mammography

Type of Use (Select one or both, as applicable)

✓ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains two logos for KUBTEC. The first logo is for "KUBTEC MEDICAL IMAGING", with "KUBTEC" in a larger, bold, blue font and "MEDICAL IMAGING" in a smaller, lighter blue font underneath. The second logo is for "KUBTEC SCIENTIFIC IMAGING", with "KUBTEC" in a larger, bold, yellow font and "SCIENTIFIC IMAGING" in a smaller, lighter yellow font underneath.

XPERT 80 Premarket Notification 510(k) Summary K210957

Date: August 20, 2021

Identification of Devices:

Proprietary/Trade Name:Kubtec XPERT 80 Specimen Radiography System
Classification name:Cabinet X-ray System
Classification:Class II
CFR Section:21 CFR 892.1680
Product Codes:MWP
Common Name:Stationary X-ray System

Applicant:

KUB Technologies, Inc. 111 Research Drive Stratford, CT 06615 USA

Contact Person:

Name:Chester Lowe, Ph.D.
Title:Chief Technology Officer
Telephone:+1.203.364.8544
Facsimile:+1.203.255.7494
Email:CLowe@kubtec.com

Establishment Number: KUB Technologies, Inc. # 3006051164

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains two logos for KUBTEC. The first logo is for KUBTEC Medical Imaging, with "KUBTEC" in large blue letters and "MEDICAL IMAGING" in smaller blue letters underneath. The second logo is for KUBTEC Scientific Imaging, with "KUBTEC" in large gold letters and "SCIENTIFIC IMAGING" in smaller gold letters underneath. The logos are stacked vertically, with the medical imaging logo on top and the scientific imaging logo on the bottom.

SUBSTANTIAL EQUIVALENCE

The proposed Kubtec XPERT 80 Specimen Radiography System is substantially equivalent to the following currently cleared devices:

    1. 510(k) Number: K071233 Trade Name: Kubtec XPERT 40/80/80L Specimen Radiography System (primary predicate)
    1. 510(k) Number: K 111508 Trade Name: Hologic TRIDENT SPECIMEN RADIOGRAPHY
    1. 510(k) Number: K 193166 Trade Name: Siemens MAMMOMAT Revelation

The only difference between the predicate to the Kubtec XPERT 40/80/80L Specimen Radiography System K071233 is the addition of the Amorphous Selenium detector and imaging qualities are exhibited by predicates 2 and 3.

DEVICE DESCRIPTION

The XPERT 80 Specimen Radiography System is a Cabinet X-ray System specifically designed to provide high detail radiographic imaging of surgically excised medical specimens. The exceptionally high magnification capability (up to 5X) from the 0.02 mm focal spot with optimized cabinet geometry and the superior contrast available from the low kV capability provides enhanced film and/or digital imaging performance. This device supports radiographic film sizes up to 30 x 35 cm and can be configured to acquire high resolution, DICOM compliant, digital x-ray images through the use of an integrated camera and Kubtec DIGICOM Specimen Radiography software. It has been designed to comply with 21 CFR 1020.40.

Xpert 80 (Predicate)Xpert 80 (Proposed)
TUBE POTENTIAL90kV90kV
FOCAL SPOT<5μm<5μm
TUBE CURRENT.178.178
INPUT POWER90-250v AC, 50/60Hz, 500VA90-250v AC, 50/60Hz, 500VA
DETECTOR TYPECMOSAmorphous Selenium
DETECTOR SIZE22.8 x 29.1 cm23.9 x 30.5 cm
FIELD OF VIEW45 degree45 degree
DETECTOR RESOLUTION49.5 μm85 μm
DETECTOR PIXELS4608 x 58902816 x 3584
INTERIOR CHAMBER SIZE35.6 W x 43.2 D x 45.7 H cm35.6 W x 43.2 D x 45.7 H cm

DEVICE TECHNICAL DESCRIPTION

111 Research Drive, Stratford, CT 06615 USA. Tel: +1(203) 364-8544; Fax:+1 (203) 255-7494 KUBTEC is a registered trademark of KUB Technologies, Inc.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows two logos for KUBTEC. The first logo is for "KUBTEC MEDICAL IMAGING" with the word "KUBTEC" in a large, bold, blue font and the words "MEDICAL IMAGING" in a smaller font below it. The second logo is for "KUBTEC SCIENTIFIC IMAGING" with the word "KUBTEC" in a large, bold, yellow font and the words "SCIENTIFIC IMAGING" in a smaller font below it. The logos are stacked vertically, with the medical imaging logo on top and the scientific imaging logo on the bottom.

EXTERIOR CABINETDIM WITH CART76.2 W x 61.0 D x 152.4 H cm76.2 W x 61.0 D x 152.4 H cm
WEIGHT250 lbs250 lbs
WEIGHT WITH CART400 lbs400 lbs
MAXIMUMGEOMETRICMAGNIFICATIONMORE THAN 5 TIMESMORE THAN 5 TIMES
SOFTWAREDIGICOM 11DIGICOM 11
OPERATING SYSTEMWINDOWS 10 PROWINDOWS 10 PRO
SYSTEM DESCRIPTION
DETECTORS
Currently ApprovedKUBTEC K071233ProposedHOLOGIC APPROVEDK111508
ManufacturerManufacturerManufacturer
Rad-IconAnalogicHologic
MFR Model #2329AXS-2430V21214
KUBTECSKUDV266DV270n/a
TypeActive PixelCMOSAmorphousSeleniumAmorphous Selenium
DetectionMethodIn-directDirectDirect
Size (cm)22.8 x 29.123.9 x 30.512 x 14
Resolution49.5μm85μm70μm
Pixel size4608 x 58902816 x 35842285 x 2571
DQE1 Lp/mm>55%>70%>55%
MTF1 Lp/mm>85%>95%>90%
Bit Depth16 bits16 bits14 bits

Radiographic Magnification: 1.0x to 10x Exposure Control: Automatic or Manual X-ray Duty Cycle: 100%

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains two logos for KUBTEC. The first logo is in blue and reads "KUBTEC MEDICAL IMAGING". The second logo is in yellow and reads "KUBTEC SCIENTIFIC IMAGING". The word "KUBTEC" is in a larger font than the words "MEDICAL IMAGING" and "SCIENTIFIC IMAGING".

Radiation Safety - 21CFR1020.40 Radiation shielded cabinet Compartment door equipped with dual safety interlocks

Radiation leakage: Less than 0.1 mR/hr at 5 cm from any exterior surface at maximum kV.

INDICATIONS FOR USE

The Kubtec XPERT 80 is a Cabinet x-ray system that is used to provide film and/or digital x-ray images of harvested specimens from various regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure.

It is not used for mammography

COMPARISON TO PREDICATE DEVICE

The Kubtec XPERT80 Specimen Radiography System equipped with the Amorphous Selenium detector is substantially equivalent to the following currently cleared devices:

510(k) Number: K071233 Trade Name: Kubtec XPERT 40/80/80L Specimen Radiography System (primary predicate) 510(k) Number: K 111508 Trade Name: Hologic TRIDENT SPECIMEN RADIOGRAPHY 510(k) Number: K 193166 Trade Name: Siemens MAMMOMAT Revelation

Predicate Device:

Device Name:Kubtec XPERT 40/80/80L SPECIMEN RADIOGRAPHYSYSTEM
510(k) Number:K071233
ClassificationClass: II
Regulation Number:21 CFR 892.1680
Product Code:MWP
Regulation Name:Stationary X-ray System
Device Name:Hologic TRIDENT SPECIMEN RADIOGRAPHY SYSTEMModel:RC
510(k) Number:K111508
ClassificationClass: II
Regulation Number:21 CFR 892.1680
Product Code:MWP
Regulation Name:Stationary X-ray System
Device Name:Siemens MAMMOMAT Revelation

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows two logos for KUBTEC. The top logo is for KUBTEC Medical Imaging, with the word "KUBTEC" in blue and "MEDICAL IMAGING" in a smaller font below it. The bottom logo is for KUBTEC Scientific Imaging, with the word "KUBTEC" in yellow and "SCIENTIFIC IMAGING" in a smaller font below it. The logos are simple and modern, with a focus on the company name.

510(k) Number: Classification Regulation Number: Product Code: Regulation Name:

K193166 Class: II 21 CFR 892.1715 MUE Full-field digital mammography system

The proposed and predicate devices utilize similar technology and materials, comparable safety and effectiveness features, and is similar in design and construction.

The Indications for Use and labeling are virtually the same or similar and our labeling contain the required Cautions, Warnings and Contraindications consistent to those required for similar cleared devices.

Both systems produce digital images which can be sent to hardcopy printers, softcopy diagnostic workstations and/or stored in archive.

The proposed XPERT 80 Specimen Radiography System, utilizes the predicate devices.XPERT 80 Specimen Radiography System, with the same 90 kVp monobloc Xray tube installed in the enclosed head of the cabinet as the predicate and both the Hologic TRIDENT(Predicate 2) and the proposed XPERT 80 Specimen Radiography System utilize the Amorphous Silicon detector mounted stationary in the bottom of the same shielded cabinet x-ray unit with proprietary software installed into an off the shelf personal computer, Microsoft Windows 10 Operating System, and a 2 megapixel or greater portrait type monitor.

The performance of the proposed AXS-2430V2 detector has been verified and cleared via 510k K193166 as incorporated into the Siemens MAMMOMAT Revelation.

Both systems utilize the same DIGICOM software.

The "Level of Concern" of XPERT 80 Specimen Radiography System software (DIGICOM) is "moderate"

SUMMARY OF STUDIES AND SAFETY

Kubtec successfully completed internal and external safety testing requirements to the following standards The software and verification testing was also performed.

The Proposed XPERT 80 Specimen Radiography System with a-Se detector was successfully tested by INTERTEK to IEC 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements as well to EMC Directive 2004/108/EC

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains two logos for KUBTEC. The first logo is in blue and says "KUBTEC MEDICAL IMAGING". The second logo is in yellow and says "KUBTEC SCIENTIFIC IMAGING".

Design controls and design transfer with appropriate risk management and analysis have been maintained per ISO 13485:2016, KUBTEC's Standard Operating Procedures via Quality and Regulatory Assurance (QARA) and Engineering, 21 CFR Part 820, and to 21 CFR 1020.40.

AgencyStandardRecognitionNumberDescription
IEC60601-1-219-36Medical electrical equipment - Part 1-2:General requirements for basic safety andessential performance - Collateral Standard:Electromagnetic disturbances -Requirements and tests
IEC60601-1-1019-37Medical electrical equipment - Part 1-10:General requirements for basic safety andessential performance - Collateral Standard:Requirements for the development ofphysiologic closed-loop controllers
IEC60601-1-1119-38Medical electrical equipment - Part 1-11:General requirements for basic safety andessential performance - Collateral Standard:Requirements for medical electricalequipment and medical electrical systemsused in the home healthcare environment
IEC60601-1-1219-39Medical electrical equipment - Part 1-12:General requirements for basic safety andessential performance - Collateral Standard:Requirements for medical electricalequipment and medical electrical systemsintended for use in the emergency medicalservices environment
IEC61010-119-34Safety requirements for electrical equipmentfor measurement, control, and laboratoryuse - Part 1: General requirements
IEC62304 1st Ed13-79Medical device software - Software lifecycle processes
ISO14971 2nd Ed5-40Medical devices - Application of riskmanagement to medical devices
ISO15223-15-117Medical devices - Symbols to be used withmedical device labels, labelling, andinformation to be supplied - Part 1: Generalrequirements
Medical electrical equipment - Radiationdose documentation - Part 1: Radiation dosestructured reports for radiography andradioscopy
IEC61910-1 1st Ed12-290
NEMAPS 3.1 - 3.2012-300Digital Imaging and Communications inMedicine (DICOM) Set
CFR21 CFR 1020.40N/ACabinet X-Ray Systems

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains two logos for KUBTEC. The first logo is for "KUBTEC MEDICAL IMAGING", with "KUBTEC" in blue and "MEDICAL IMAGING" in a smaller font size below it. The second logo is for "KUBTEC SCIENTIFIC IMAGING", with "KUBTEC" in gold and "SCIENTIFIC IMAGING" in a smaller font size below it. The logos are stacked vertically, with the medical imaging logo on top and the scientific imaging logo on the bottom.

CONCLUSION:

The Kubtec proposed KUBTEC XPERT 80 Specimen Radiography System equipped with an Amorphous Selenium Detector is as safe and effective as the predicate device, the technological differences amount to the change of the type of detector to produce the 2-D x-ray image. It has no new indications for use, thus rendering it substantially equivalent to the predicate device and conforms to applicable medical device safety standards.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.