(91 days)
The SYNERGY is an Over-the-Counter (OTC) device intended for the use in treating wrinkles on the face.
The SYNERGY is an over-the-counter, battery powered, hand-held light emitting diode (LED) device that emits light energy in the red and infrared spectrum for the treatment of wrinkles on the face. The SYNERGY components include an applicator, a charging station and an adaptor. The treatment surface is applied directly to the skin to ensure consistent administration of light during each treatment. The device does not contain any user repairable components. The device is sold for Over the Counter (OTC) use.
The provided text is a 510(k) summary for the "SYNERGY" device, a red/IR light therapy device intended for treating facial wrinkles. It establishes substantial equivalence to a predicate device, "Elevare Plus" (K172909).
However, the provided document does not contain any information about acceptance criteria for device performance related to efficacy (e.g., wrinkle reduction) or any study that proves the device meets such criteria (e.g., a clinical study with human subjects). It focuses primarily on the safety and technological equivalence to a predicate device.
The document details non-clinical testing for electrical safety, EMC, photobiological safety, battery safety, and biocompatibility, as well as software validation and skin temperature testing. These tests ensure the device is safe to use and performs within its design specifications, but they do not address the clinical effectiveness of wrinkle reduction or any acceptance criteria for that purpose.
Therefore, I cannot fulfill your request for:
-
A table of acceptance criteria and reported device performance
- The document does not provide acceptance criteria for clinical efficacy (e.g., quantitative wrinkle reduction) or corresponding performance data. The "performance testing" mentioned refers to software validation and skin temperature, not clinical outcomes.
-
Sample size used for the test set and the data provenance
- No human test set for clinical performance is described. The "test set" for non-clinical testing (e.g., electrical safety, temperature) would refer to the device units tested, not human subjects.
-
Number of experts used to establish the ground truth
- Not applicable, as no clinical efficacy study or ground truth establishment for clinical outcomes is described.
-
Adjudication method
- Not applicable.
-
MRMC comparative effectiveness study
- Not applicable.
-
Standalone performance
- Not explicitly described in terms of clinical efficacy. The standalone performance testing mentioned is for device safety and functional specifications (e.g., skin temperature).
-
Type of ground truth used
- Not applicable for clinical ground truth. For the safety tests, the "ground truth" would be established by the standards themselves (e.g., IEC 60601-1, IEC 62471).
-
Sample size for the training set
- Not applicable, as this is a physical device, not an AI/ML algorithm that requires a training set of data for clinical performance.
-
How the ground truth for the training set was established
- Not applicable.
Summary of Information Available Related to Safety and Technical Performance:
The document primarily focuses on demonstrating substantial equivalence based on technological characteristics and safety standards.
- Acceptance Criteria (Safety/Technical Performance): The device conforms to several international consensus standards for electrical safety (ANSI AAMI ES60601-1, ANSI AAMI IEC 60601-1-2, ANSI AAMI IEC60601-1-11), photobiological safety (IEC 62471), battery safety (IEC 62133-2), and biocompatibility (ISO 10993-5, ISO 10993-10). Additionally, skin temperature testing ensured the device stayed within 41 ± 2℃ and did not elevate skin temperature above 43°C. Software validation was conducted per FDA guidance.
- Study Proving Acceptance Criteria (Safety/Technical Performance):
- The document states that "The SYNERGY has been tested and conforms to international consensus standards" and lists the specific standards.
- For skin temperature, it states "The test results concluded that the SYNERGY was within specification of 41± 2℃ and did not elevate skin temperature above 43°C."
- Sample size and data provenance for these technical tests: Not specified beyond the implication that device units were tested in a non-clinical setting. The data provenance would be from laboratory testing.
- Experts/Ground truth for technical tests: Ground truth is established by the specified international consensus standards for device safety and performance. No "experts" in the sense of clinical reviewers are mentioned for these technical tests; rather, certified testing laboratories and their protocols would be involved.
- MRMC/Standalone (clinical efficacy): Not relevant or discussed in this document.
In conclusion, the provided FDA 510(k) summary addresses the safety and technical specifications of the SYNERGY device, demonstrating its substantial equivalence to a predicate device. It does not include information on clinical efficacy studies or acceptance criteria related to the device's intended effect on wrinkles.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
June 29, 2021
YRS Group Inc. % Doris Dong Senior Regulatory Affairs Project Manager Shanghai CV Technology Co., Ltd. Room 903, No. 19 Dongbao Road, Songjiang Area Shanghai, Shanghai 201613 China
Re: K210954
Trade/Device Name: Synergy Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: March 3, 2021 Received: March 30, 2021
Dear Doris Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K210954
Device Name SYNERGY
Indications for Use (Describe)
The SYNERGY is an Over-the-Counter (OTC) device intended for the use in treating wrinkles on the face.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K210954
510(k) Summary [As required by 21 CFR 807.92]
1. Submission Information 510(k) Number: K210954 Date: June 15th, 2021 Traditional Type of 510(k) Submission: New device Basis for 510(k) Submission: Owner: YRS Group Inc 4480 Delancey Dr Ste 14, Las Vegas, NV, 89103-3753, US Tel: (702) 4268921 Email: usa@opatra.com Doris Dong Contact: [Consultant, from Shanghai CV Technology Co., Ltd.] Add: Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 China E-mail: doris.d@ceve.org.cn Tel: 86 21-31261348 / Fax: 86 21-57712250 2. Device Description Proprietary Name: SYNERGY Model: BP-1818R1 Common Name: Red/IR Light Therapy Device Classification Name: Light Based Over The Counter Wrinkle Reduction Regulation Number: 21 CFR 878.4810 OHS Product Code: Device Class: II Review Panel: General & Plastic Surgery The SYNERGY is an over-the-counter, battery powered, hand-held light Device Description: emitting diode (LED) device that emits light energy in the red and infrared spectrum for the treatment of wrinkles on the face. The SYNERGY components include an applicator, a charging station and an adaptor. The treatment surface is applied directly to the skin to ensure consistent administration of light during each treatment. The device does not contain any user repairable components. The device is sold for Over the Counter (OTC) use. Indications for use: The SYNERGY is an Over-the-Counter (OTC) device intended for the use in treating wrinkles on the face.
3. Substantial Equivalence to Predicate device
Detailed comparison data is included in "Section 10 - Substantial Equivalence Discussion" of this 510(k) submission.
Table 1-
| Parameters | New Device | Predicate Device | Remark |
|---|---|---|---|
| 510(k) number | K210954 | K172909 | -- |
| Device name | SYNERGY | Elevare Plus | -- |
| Product code | OHS | OHS | Same |
{4}------------------------------------------------
| Class | II | II | Same |
|---|---|---|---|
| Indication for use | The SYNERGY is anOver-the-Counter (OTC) deviceintended for the use in treatingwrinkles on the face. | The Elevare Plus is anOver-the-Counter (OTC) deviceintended for the use in treatingwrinkles on the face. | Same |
| Anatomical site | Entire Face | Entire Face | Same |
| Handheld | Yes | Yes | Same |
| wavelengths | 610, 630, 660, 850nm ±5nm | 610, 630, 660, 850nm ±5nm | Same |
| Irradiance source | LED | LED | Same |
| Visible light LEDs | Yes | Yes | Same |
| LED Array | 16 LEDs, Over 17cm² | 25 LEDs, Over 17cm² | SimilarNote 1 |
| Energy Level(mW/cm²) | 65 mW/cm² (± 5 mW/cm²) | 65 mW/cm² (± 5 mW/cm²) | Same |
| Power Supply | Li-Ion rechargeable batteries | Li-Ion rechargeable batteries | Same |
| Treatment Time | 3 minutes each area each mode, 5days per week for 8 weeks | 3 minutes daily, 5 days per weekfor 8 weeks | SimilarNote 2 |
| Target population | Individuals with wrinkles on theface | Individuals with wrinkles on theface | Same |
| Location for use | OTC | OTC | Same |
| Main material | ABS and stainless steel | ABS and stainless steel | Same |
Summary of the technological characteristics of the device compared:
Note 1:
The subject device is slightly different from the predicate device in terms of LEDs and appearance. The subject device has 16 LEDs while the predicate device has 25 LEDs,but they have the same energy density, and both of them have passed the tests of IEC 60601-1 and IEC 62471, these differences will not raise any new issues of safety or effectiveness.
Note 2:
The treatment time of the proposed device is slightly different from the predicate device uses 4 wavelengths of LEDs to work simultaneously for 3 minutes, while the proposed device sets 4 wavelengths in two modes (Mode 1: 610nm & 630nm, Mode 2: 660nm & 850nm), each mode is treated for 3 minutes. Considering both of them have the same treatment time of each wavelength, the energy produced by these wavelengths, power density and maintenance time are also the same, it can be concluded that the same treatment effect as the predicate device can be achieved. In addition, the proposed device has passed the tests of IEC 60601-1 and IEC 62471. Based on above analysis, these differences will not cause any safety and effectiveness issues.
4. Non-clinical Testing
The conclusions drawn from the non-clinical testing below demonstrate that the SYNERGY is substantially equivalent to the predicate devices K172909. The SYNERGY has been tested and conforms to international consensus standards:
Electrical safety:
- ANSI AAMI ES60601-1: 2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 ● (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD);
EMC:
{5}------------------------------------------------
K210954
-
ANSI AAMI IEC 60601-1-2:2014, Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances --Requirements And Tests;
Additional safety testing: -
ANSI AAMI IEC60601-1-11 2015 Medical Electrical Equipment -- Part 1-11: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment;
-
IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems; ●
-
IEC 62133-2, Edition1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems;
Biocompatibility testing:
- ISO 10993-5:2009/(R) 2014, Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)
- ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. (Biocompatibility)
The SYNERGY has been tested to ensure the device meets specifications:
Performance testing
-
Software Validation Testing ●
The SYNERGY's software was tested and validated in accordance with FDA's "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices." -
Skin Temperature Testing
Temperatures of treatment surface were recorded every 1 minutes, using a calibrated digital temperature sensor. The test results concluded that the SYNERGY was within specification of 41± 2℃ and did not elevate skin temperature above 43°C.
5. Conclusions
Based upon comparison to the predicate devices, the SYNERGY has the same intended uses, with similar technological characteristics as predicate device SYNERGY is substantially equivalent to the predicate device K172909.
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.