K Number
K210933
Manufacturer
Date Cleared
2024-09-30

(1281 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The YSI 2900C Biochemistry Analyzer is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of venous whole blood samples, fingerstick capillary whole blood samples, serum and plasma in the laboratory.

For in vitro diagnostic use.

Glucose measurements from the YSI 2900C Biochemistry Analyzer are used in the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell tumors. The YSI 2900C Biochemistry Analyzer is not intended for use in the screening or quantitative analysis on neonates. The YSI 2900C Biochemistry Analyzer is not intended for point of care (POC) use.

This test is for prescription use only.

Device Description

The device under review, the 2900C Biochemistry Analyzer, is a laboratory instrument and In Vitro Diagnostic Device for determining glucose in human whole blood, plasma and serum taken from venous or capillary samples. The 2900C Biochemistry Analyzer is a semi-automated electronic device that incorporates fluidics for sampling, calibrating, and flushing, a membrane-immobilized enzyme-coupled electrochemical detection system with digital electronic control and has graphical user and data interfacing. It is designed for ambient indoor use in a technical laboratory environment.

The system is not intended to be a point of care product and is intended only for professional laboratory use.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the YSI 2900C Biochemistry Analyzer, which is a device for quantitative testing of glucose. It aims to demonstrate substantial equivalence to a predicate device, the YSI 2300 STAT PLUS Analyzer.

Here's a breakdown of the acceptance criteria and study information, based on the provided document:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of "acceptance criteria" alongside specific numerical "reported device performance" in the format typically seen for algorithm performance (e.g., sensitivity, specificity, AUC thresholds). Instead, the studies aim to demonstrate substantial equivalence to a predicate device. The performance metrics discussed are related to the analytical performance of a biochemistry analyzer.

However, we can extract the precision acceptance criterion and its comparison to the predicate device:

Performance MetricAcceptance Criteria (from Predicate Device: YSI 2300 STAT PLUS)Reported Device Performance (YSI 2900C Biochemistry Analyzer)
Precision±2% of the reading or 2.5 mg/dL (25 mg/L, 0.2 mmol/L), whichever is larger2% or 2.0 mg/dL whichever is greater

It's important to note that for other tests like Method Comparison, Linearity, Specificity/Interference, etc., the "acceptance criteria" would be met by demonstrating that the new device's performance is comparable to the predicate, often through statistical equivalence testing within defined analytical limits, rather than fixed numerical thresholds for a specific algorithm output. The document states: "The results of the performance testing confirmed that the YSI 2900C Biochemistry Analyzer demonstrates substantial equivalence to the YSI 2300 StatPlus Analyzer."

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document provides details on the types of studies performed but does not specify the sample sizes used for the test sets in any of the performance studies (Method Comparison, Precision/Reproducibility, Linearity, Specificity/Interference, etc.).

There is no information provided regarding data provenance (e.g., country of origin, retrospective or prospective nature of the samples). The samples are stated to be "human whole blood, plasma and serum taken from venous or capillary samples."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable in the context of this device. The YSI 2900C Biochemistry Analyzer is an in vitro diagnostic device that quantitatively measures glucose in biological samples. The "ground truth" for its performance is established by comparing its measurements against reference methods or the predicate device, not by expert interpretation of images or clinical data. Therefore, there's no mention of experts establishing a ground truth for a test set in the way one would for an AI-powered diagnostic imaging device.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This section is not applicable for this type of device. Adjudication methods like "2+1" or "3+1" are typically used in studies involving expert readers interpreting medical images or clinical data, where consensus is needed to establish a "ground truth" when individual expert opinions might differ. For a biochemistry analyzer, the validation involves analytical comparisons to reference methods or a predicate device, not human interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable to this device. An MRMC study is designed to evaluate how AI assists human interpretation, typically in diagnostic imaging. The YSI 2900C Biochemistry Analyzer is a laboratory instrument that provides quantitative measurements; it does not involve human "readers" interpreting output in the same way, nor does it provide "AI assistance" for human interpretation. Its performance is evaluated analytically against established methods.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device itself is a "standalone" system in its operation for measuring glucose. It performs the analysis and provides quantitative results without direct human "interpretation" of a processed image or complex data set. The performance testing described (Method Comparison, Precision, Linearity, etc.) represents the device's inherent analytical capabilities, which is analogous to "standalone" performance for an IVD. There is no "human-in-the-loop" component for the actual glucose measurement process that would affect its analytical performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" in these performance studies is typically established by:

  • Reference Methods: Highly accurate and precise analytical methods (e.g., isotope dilution mass spectrometry) that are considered the gold standard for glucose measurement, or
  • Predicate Device: For demonstrating substantial equivalence, the performance of the new device (YSI 2900C) is compared against that of the legally marketed predicate device (YSI 2300 STAT PLUS) across various analytical parameters. The predicate device's established performance serves as the comparative "truth" for demonstrating equivalence.
  • Known Concentrations: For linearity and limit studies, samples with precisely known glucose concentrations are used.

The document explicitly states: "A mirror of this testing was also performed concurrently on the predicate device, the YSI 2300 STAT PLUS Biochemistry Analyzer, for demonstration of substantial equivalence." This indicates the predicate device serves as the primary comparative "ground truth" for the equivalence assessment.

8. The sample size for the training set

This section is not applicable in the context of this traditional analytical device. The YSI 2900C Biochemistry Analyzer uses an "enzyme-mediated electrochemical detection" system, not a machine learning or AI algorithm that requires a "training set" of data. Its underlying technology is based on established biochemical and electrochemical principles.

9. How the ground truth for the training set was established

As there is no "training set" for an AI or machine learning algorithm in this device, this question is not applicable. The device's operational parameters and calibration are based on physical chemistry principles and factory calibration/validation, not a data-driven training process.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 30, 2024

YSI Incorporated Eric Lute Director QA 1725 Brannum Lane Yellow Springs, Ohio 45387

Re: K210933

Trade/Device Name: YSI 2900C Biochemistry Analyzer Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: CGA Dated: May 5, 2023 Received: May 15, 2023

Dear Eric Lute:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula V. Caposino -S

Paula Caposino, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210933

Device Name

YSI 2900C Biochemistry Analyzer

Indications for Use (Describe)

The YSI 2900C Biochemistry Analyzer is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of venous whole blood samples, fingerstick capillary whole blood samples, serum and plasma in the laboratory.

For in vitro diagnostic use.

Glucose measurements from the YSI 2900C Biochemistry Analyzer are used in the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell tumors. The YSI 2900C Biochemistry Analyzer is not intended for use in the screening or quantitative analysis on neonates. The YSI 2900C Biochemistry Analyzer is not intended for point of care (POC) use.

This test is for prescription use only.

Type of Use (Select one or both, as applicable) Exempt Activities (Part 21 CFR 201.24 Subject to Over-The-Counter Use Exempt Activities (Part 21 CFR 201.24 Subject to Over-The-Counter Use
Exempt Activities (Part 21 CFR 201.24 Subject to
Over-The-Counter Use

|X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitted by:

YSI Incorporated 1725 Brannum Lane Yellow Springs, OH 45387 USA

510(k) Owner: YSI Incorporated FDA Registration #: 1550101 510(k) # - K210933

Contact Person

Eric Lute Director, QA Ph: (937) 749-0587 Fax: 937-767-8058 Date Prepared: 26-Sep-2024

Device Name: YSI 2900C Biochemistry Analyzer

Common Name: Glucose Analyzer

Classification Name: Glucose Test System (21 CFR 862.1345, Product Code CGA)

Predicate Device/Claims of Equivalence against YSI marketed MODEL 23A GLUCOSE ANALYZER; YSI 2300 STAT PLUS (referenced throughout as 2300 STAT PLUS). Reference 510(k): K913806.

There have been no prior applications or submissions for the review/approval of the YSI 2900C BIOCHEMISTRY ANALYZER.

Description of Device

The device under review, the 2900C Biochemistry Analyzer, is a laboratory instrument and In Vitro Diagnostic Device for determining glucose in human whole blood, plasma and serum taken from venous or capillary samples. The 2900C Biochemistry Analyzer is a semi-automated electronic device that incorporates fluidics for sampling, calibrating, and flushing, a membrane-immobilized enzyme-coupled electrochemical detection system with digital electronic control and has graphical user and data interfacing. It is designed for ambient indoor use in a technical laboratory environment.

The system is not intended to be a point of care product and is intended only for professional laboratory use.

The intended use and methods are shown to be substantially equivalent to the process and chemistry for glucose analysis as used in the 2300 STAT PLUS instrument (K913806). Lactate is not included in this submission as the feature has been disabled in the YSI 2900C Biochemistry Analyzer version of the system. There is no effect on glucose analysis in the system as the removal of lactate as the two assays are mutually exclusive.

Page 1 | 5 K210933

{4}------------------------------------------------

Indications for Use of the YSI 2900C Biochemistry Analyzer

The YSI 2900C Biochemistry Analyzer is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of venous whole blood samples, fingerstick capillary whole blood samples, serum, and plasma in the laboratory.

For in vitro diagnostic use.

Glucose measurements from the YSI 2900C Biochemistry Analyzer are used in the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell tumors. The YSI 2900C Biochemistry Analyzer is not intended for use in the screening or quantitative analysis on neonates. The YSL 2900C Biochemistry Analyzer is not intended for point of care (POC) use.

This test is for prescription use only.

Summary of Instrument Operation

Underlying Technology of YSI Biochemistry Analyzers - Operation

A blood or plasma sample, contained in vials or tubes of various forms, is placed in a configured holder on the instrument. A robotic sampling sipper/injector travels to the sample and precisely withdraws a sample from it. The glucose containing sample is transferred to and precisely injected into a reaction chamber. Within the reaction chamber the sample is mixed to homogeneity with the co-contained system buffer within the precise volume of the chamber. The reaction chamber fluid is in direct contact with three electrodes - two sense electrodes and an auxiliary electrode. The technology of the device is described as an electrochemical detection system or more explicitly as coulometry or potentiostatic amperometry. The glucose in the reaction chamber fluid diffuses into an enzyme-containing membrane that overlays a platinum button electrode. Within the membrane the immobilized enzyme glucose oxidase catalytically reacts the glucose with dissolved oxygen. A product of this reaction, hydrogen peroxide, then diffuses to the platinum electrode. The platinum sense electrode, potentiostatically held at +0.7 volts, re-oxidizes the peroxide back into oxygen. This reoxidation creates an electric current which is detected and quantitated by the electronics in the instrument. The current detected is exactly proportional to the amount of glucose in the sample. The current created by the sample is compared to the current created by a calibrator solution of known exact glucose concentration and the glucose in the sample is computed by direct proportionality.

Principle of Measurement:

Glucose measurement is based on the level of H2O2 produced during the enzymatic reaction between glucose and oxygen molecules in the presence of the glucose oxidase enzyme. At a constant potential of 0.70 volts, electroactive H2O2 is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the surface of the platinum electrode is proportional to the glucose concentration of the sample.

Performance Testing Included with this Submission

The analytical performance of the YSI 2900C Biochemistry Analyzer is tested across the entire range of the instrument from 0.0 mg/dL to 900 mg/dL of glucose in whole blood and plasma. This span includes all four generally recognized medical decision points: 54, 70, 125, and 250 mg/dL, and includes other range intermediates and extrema. The data gained is used for Linearity Evaluation and comparison with the predicate device. A mirror of this testing was also performed concurrently on the predicate device, the YSI 2300 STAT PLUS Biochemistry Analyzer, for demonstration of substantial equivalence.

Nonclinical Tests Page 2 | 5 K210933

{5}------------------------------------------------

Testing to confirm instrument performance and establish system limits included:

  • Limit of Blank ●
  • Limit of Detection ●
  • Limit of Quantification .
  • Environmental Conditions (Temperature and Humidity) .
  • Temperature Compensation Algorithm .

Analytical and Clinical Validation

Testing was completed to show that the YSI 2900C Biochemistry Analyzer demonstrates substantial equivalence to the YSI 2300 StatPlus Analyzer. The performance testing included:

  • Method Comparison Studies
  • Precision/Reproducibility Studies
  • Linearity Testing
  • Specificity / Interference Testing ●
  • Anticoagulant Matrix Equivalencies/Suitabilities ●
  • . Hematocrit Bias
  • Stat Interrupt Biases
    • Stat Position Sampling o
    • o Syringe Position Sampling

The results of the performance testing confirmed that the YSI 2900C Biochemistry Analyzer demonstrates substantial equivalence to the YSI 2300 StatPlus Analyzer.

{6}------------------------------------------------

Table 1. Predicate comparison table to outline differences and similarities between the sample and predicate devices.

DescriptionCandidate Device2900C BIOCHEMSITRYANALYZERPredicate Device2300 STAT PLUS
Indications for UseThe 2900C Biochemistry Analyzer is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of venous, or capillary whole blood, serum and plasma in the laboratory.Glucose measurements from the 2900C Biochemistry Analyzer are used in the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell tumors. The 2900C Biochemistry Analyzer is not intended for use in the screening or quantitative analysis on neonates. The system is intended for prescription use only.Determine glucose and lactate level in human whole blood or plasma
SizeDesktopSame
Battery or mains poweredMains poweredSame
Sample Volume25 µLSame
Method of AnalysisEnzyme mediated electrochemical detectionSame
Sample IntroductionAutomated robotic sampler armSame
Installed Membrane Life21 daysSame
Reagent/Test TrackingPrinted Data / Internal DatabasePrinted Data
Precision2% or 2.0 mg/dL whichever is greater±2% of the reading or 2.5 mg/dL (25 mg/L, 0.2 mmol/L), whichever is larger
QC MaterialYSI product 1531 Glucose standard, 9.00 g/mLSame
CalibrationAutomatic, interval determined by user, recommended maximum is after 5 samples or 15 minutes since previous calibrationSame
User InterfaceTouchscreen GUImembrane switch panel and 2 line low resolution monochrome alphanumeric display
OS SoftwareWindows Embedded Standard 7in Kiosk (read only) modeNone
Working Range15–35°C, 10–75% humidity(non-condensing)Same
Non-Interfering Anticoagulantsheparin sodium, dipotassiumEDTA, sodiumfluoride/potassium oxalate,tripotassium EDTA, LithiumHeparinSame
InterferentsListed in user manualSame

{7}------------------------------------------------

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.