(143 days)
For in vitro diagnostic and laboratory professional use.
VITROS Chemistry Products PHBR Slides quantitatively measure phenobarbital (PHBR) concentration in serum and plasma (lithium heparin) using the automated VITROS 5600 Integrated System.
Measurements obtained by this device are used as an aid in the diagnosis and treatment of phenobarbital use or overdose and in monitoring levels of phenobarbital to help ensure appropriate therapy.
The VITROS PHBR Slide is a multilayered, analytical element coated on a polyester support. The phenobarbital assay is based on an enzymatic heterogeneous, competitive immunoassay format. Immobilized anti-phenobarbital antibody and phenobarbital-peroxidase conjugate are present in the spreading layer.
A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Phenobarbital in the sample competes with the phenobarbitalperoxidase conjugate for a limited number of antibody binding sites during Incubation 1. The subsequent addition of 12 µL of VITROS Immuno-Wash Fluid to the slide removes unbound phenobarbital-peroxidase conjugate from the read area, while also providing a substrate for the enzyme mediated oxidation of leuco dye.
The rate of dye formation, as monitored by reflectance spectrophotometry for Incubation 2, is inversely proportional to the phenobarbital concentration in the sample. To determine if an adequate wash has occurred, a wash detection dye is read at 540 nm during Incubation 2.
Here's an analysis of the provided FDA 510(k) summary, specifically focusing on the acceptance criteria and study proving the device meets those criteria, formatted as requested:
Device: VITROS Chemistry Products PHBR Slides
Device Type: In vitro diagnostic device for quantitative measurement of phenobarbital (PHBR) concentration in serum and plasma.
1. Table of Acceptance Criteria and Reported Device Performance
The FDA 510(k) summary outlines several analytical performance characteristics that serve as acceptance criteria for the VITROS Chemistry Products PHBR Slides. The document doesn't explicitly state "acceptance criteria" for all metrics in the form of numerical thresholds before the study results, but rather presents the study results as meeting acceptable performance for substantial equivalence. For some, like LoQ, a specific goal is mentioned.
| Performance Characteristic | Acceptance Criterion (Implicit/Explicit) | Reported Device Performance |
|---|---|---|
| Method Comparison | Demonstrate substantial equivalence to predicate device (ARCHITECT iPhenobarbital Assay) via Passing-Bablok Regression. Implied criteria for acceptable slope and correlation coefficient close to 1, and low intercept. | N=142 samples. Slope: 0.92, Correlation Coefficient: 0.983, Intercept: -2.0 µg/mL, Sv.x: 2.6. Range of Samples: 5.6 - 76.1 µg/mL. |
| Precision | Demonstrate acceptable repeatability and within-lab precision (low %CV and SD). No explicit numerical criterion stated, but implied to be within industry standards for clinical chemistry assays of this type for therapeutic drug monitoring. | Repeatability: PHBR conc. (µg/mL) / SD / %CV5.0 / 0.20 / 3.9%9.1 / 0.27 / 2.9%11.0 / 0.30 / 2.7%23.0 / 0.50 / 2.2%25.1 / 0.59 / 2.4%38.1 / 0.79 / 2.1%59.3 / 1.50 / 2.5%Within Lab: PHBR conc. (µg/mL) / SD / %CV5.0 / 0.25 / 5.0%9.1 / 0.33 / 3.6%11.0 / 0.44 / 4.0%23.0 / 0.65 / 2.8%25.1 / 0.81 / 3.2%38.1 / 1.09 / 2.9%59.3 / 2.11 / 3.6% |
| Detection Limits (LoD) | Demonstrate a limit of detection low enough for clinical utility. No explicit numerical criterion stated prior to reporting. | LoD: 1.3 µg/mL |
| Detection Limits (LoQ) | Allowable error goal for LoQ: ≤ 1.5 µg/mL. | Claimed LoQ: 3.0 µg/mL. The study found the claimed LoQ to be acceptable within the ≤ 1.5 µg/mL total error goal. |
| Linearity | Deviation from linearity within allowable limits: ± 1.3 µg/mL at phenobarbital concentrations <10 µg/mL, and max. concentration-dependent allowable deviation of ± 11.5 µg/mL at phenobarbital concentrations >10 µg/mL. | Demonstrated linearity over the measuring range of 3.0 - 80.0 µg/mL. LLLI: 0.5 µg/mL, ULLI: 83.5 µg/mL. Reported deviation from linearity was within the specified allowable deviations. |
| Specificity (Interference) | Bias > 1.8 µg/mL at approx. 15 µg/mL PHBR or bias > 5.2 µg/mL at approx. 50 µg/mL PHBR considered significant interference. Implied acceptance: most common substances should not interfere, or known interferences clearly identified and characterized. | Nine (9) substances showed interference (bias exceeding criteria) at specified concentrations relative to PHBR concentrations of 15 µg/mL or 50 µg/mL. Sixty-one (61) other test substances did not cause significant bias. |
| Cross-Reactivity | Characterize the cross-reactivity of structurally related compounds and common co-prescribed drugs. Implied acceptance: cross-reactivity is understood and characterized. | Cross-reactivity of 12 substances (e.g., amobarbital, mephobarbital, phenytoin) was evaluated. Results are provided for reference in the customer instructions for use (specific numerical results not provided in this summary). |
| Measuring Range | 3.0 – 80.0 µg/mL | 3.0 – 80.0 µg/mL |
2. Sample Sizes and Data Provenance
- Method Comparison: 142 serum samples.
- Precision: 88 observations (2 replicates per run, 2 runs per day over 22 days) using patient pools and quality control materials.
- Linearity: Fourteen proportionally related admixtures of low and high concentration fluids were tested, each in duplicate.
- Specificity (Interference) & Cross-Reactivity: Number of individual samples tested for each interferent/cross-reactant not explicitly stated, but the studies describe adding substances to serum samples.
- Training Set Sample Size: Not applicable based on the provided document. This is an IVD device measuring analyte concentration, not an AI/ML device requiring a training set for algorithm development in the traditional sense. The phrase "training set" is typically used for machine learning models.
- Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective/prospective). Standard laboratory validation protocols (CLSI guidelines listed) imply controlled, often prospective, collection for such studies.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not applicable. The "ground truth" for this device (a quantitative diagnostic for phenobarbital concentration) is established by direct chemical measurement in the form of a reference method (the predicate device) or by known concentrations of prepared standards/controls. It does not involve expert interpretation or consensus in the way a diagnostic imaging AI might.
4. Adjudication Method for the Test Set
- Not applicable. This specific study does not involve human readers or interpretations that would require adjudication. Measurements are quantitative and compared to a reference method or known concentrations.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No. An MRMC study is typically performed for AI-assisted diagnostic imaging or similar scenarios where human readers make subjective interpretations. This document describes the analytical performance of an in vitro diagnostic device for measuring a chemical analyte, which is not subject to human reader variability in the same way.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Yes, indirectly. The entire document describes the standalone analytical performance of the VITROS Chemistry Products PHBR Slides on the VITROS 5600 Integrated System. The device provides a quantitative measurement directly, without human interpretation in the loop that would alter the result itself. The measurements obtained are then used by laboratory professionals to aid in diagnosis and treatment, but the device's performance itself is evaluated as a standalone analytical instrument.
7. Type of Ground Truth Used
- Reference Method / Known Concentrations:
- For Method Comparison, the ground truth is established by the results from the legally marketed predicate device (ARCHITECT iPhenobarbital Assay) on the same patient samples.
- For Precision, Detection Limits, Linearity, Specificity, and Cross-Reactivity, the ground truth is established by using prepared samples with known, controlled concentrations of phenobarbital and/or interfering/cross-reacting substances.
8. Sample Size for the Training Set
- Not applicable. As explained in section 2, this is an IVD device for quantitative measurement, not an AI/ML device that requires a "training set" for algorithm development.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. No training set in the AI/ML sense. For analytical validation, ground truth is established by precisely prepared standards and controls, or by comparison to a validated reference method (like the predicate device).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
August 13, 2021
Ortho-Clinical Diagnostics, Inc. Darlene Phillips Manager, Regulatory Affairs 100 Indigo Creek Drive Rochester, New York 14626-5101
Re: K210858
Trade/Device Name: VITROS Chemistry Products PHBR Slides Regulation Number: 21 CFR 862.3660 Regulation Name: Phenobarbital Test System Regulatory Class: Class II Product Code: DLZ Dated: March 22, 2021 Received: March 23, 2021
Dear Darlene Phillips:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K210858
Device Name VITROS Chemistry Products PHBR Slides
Indications for Use (Describe)
Rx Only
For in vitro diagnostic and laboratory professional use.
VITROS Chemistry Products PHBR Slides quantitatively measure phenobarbital (PHBR) concentration in serum and plasma (lithium heparin) using the automated VITROS 5600 Integrated System.
Measurements obtained by this device are used as an aid in the diagnosis and treatment of phenobarbital use or overdose and in monitoring levels of phenobarbital to help ensure appropriate therapy.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K210858
Submitter's Information
Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, New York 14626-5101 Phone: (585) 453-4253 Fax: (585) 453-4110
Contact Person:
Darlene J Phillips, RAC Manager, Regulatory Affairs
Date of Preparation:
22-March-2021
Device Proprietary Name(s):
VITROS Chemistry Products PHBR Slides
Common Names:
Phenobarbital assay
Classification Names
| ProductCode | Class | Regulation Section | Panel |
|---|---|---|---|
| DLZ | II | 21 CFR 862.3660Phenobarbital test system | Clinical Chemistry (75) |
Predicate Device(s)
| Predicate Device | FDA 510(k) Number |
|---|---|
| ARCHITECT iPhenobarbital Assay | K081231 |
Intended Use Statement(s)
VITROS Chemistry Products PHBR Slides
Rx Only
For in vitro diagnostic and laboratory professional use.
VITROS Chemistry Products PHBR Slides quantitatively measure phenobarbital (PHBR) concentration in serum and plasma(lithium heparin) using the automated VITROS 5600
{4}------------------------------------------------
Integrated Systems. Measurements obtained by this device are used as an aid in the diagnosis and treatment of phenobarbital use or overdose and in monitoring levels of phenobarbital to help ensure appropriate therapy.
Device Description
The VITROS PHBR Slide is a multilayered, analytical element coated on a polyester support. The phenobarbital assay is based on an enzymatic heterogeneous, competitive immunoassay format. Immobilized anti-phenobarbital antibody and phenobarbital-peroxidase conjugate are present in the spreading layer.
A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Phenobarbital in the sample competes with the phenobarbitalperoxidase conjugate for a limited number of antibody binding sites during Incubation 1. The subsequent addition of 12 µL of VITROS Immuno-Wash Fluid to the slide removes unbound phenobarbital-peroxidase conjugate from the read area, while also providing a substrate for the enzyme mediated oxidation of leuco dye.
The rate of dye formation, as monitored by reflectance spectrophotometry for Incubation 2, is inversely proportional to the phenobarbital concentration in the sample. To determine if an adequate wash has occurred, a wash detection dye is read at 540 nm during Incubation 2.
Comparison to Predicate Devices
Indications for use/intended use of candidate and predicate devices are the same. The following tables show similarities and differences between the new and predicate devices.
| Summary of the technological characteristics of the device compared to the predicate device | ||
|---|---|---|
| DeviceCharacteristic | New DeviceVITROS PHBR Slide(New) | Predicate DeviceARCHITECT iPhenobarbital Assay[K081231] |
| Intended Use | For in vitro diagnostic use only.Same | For in vitro diagnostic use only.Quantitative measurement ofphenobarbital in human serum orplasma |
| DeviceDescription | Multilayered, analytical elementcoated on a polyester support | Two ready-to-use solutions |
| Basic Principle | Multi-point immunorate assay | Chemiluminescent MicroparticleImmunoassay (CMIA) |
| Sample type | Serum and lithium heparin plasma | Serum and plasma |
| Sample Volume | 11 μL | 20 μL |
| Assay Range | 3.0 – 80.0 μg/mL | 1.10 – 80.00 μg/mL |
| Calibrators | VITROS Chemistry ProductsCalibrator Kit 9 | ARCHITECT iPhenobarbital CalibratorKit |
| Controls | VITROS Chemistry ProductsTDM Performance Verifiers | Abbott Immunoassay-MCC (Liquid) orother commercial controls |
| Instrumentation | VITROS 5600 Integrated System | ARCHITECT i System |
{5}------------------------------------------------
Non-Clinical Testing Analytical Performance
Method Comparison
Method Comparison testing followed CLSI Protocol EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples. Serum samples were evaluated on the VITROS Chemistry Products PHBR Slides using the VITROS 5600 Integrated System and the ARCHITECT iPhenobarbital assay. Passing-Bablok Regression was performed to determine the correlation.
| N | Slope | CorrelationCoefficient | Range of Samplesug/mL | Interceptug/mL | Sv.x |
|---|---|---|---|---|---|
| 142 | 0.92 | 0.983 | 5.6 - 76.1 | -2.0 | 2.6 |
Precision
Precision was evaluated with patient pools and quality control materials following CLSI Protocol EP05-A3, Evaluation of Precision Performance of Quantitative Methods; Approved Guideline-Third Edition, using the VITROS Chemistry Products PHBR Slides on the VITROS 5600 Integrated System. The test included 88 observations (2 replicates per run, 2 runs per day over 22 days).
The data presented are a representation of test performance and are provided as a guideline. Variables such as sample handling and storage, reagent handling and storage, laboratory environment, and system maintenance can affect reproducibility of test results.
| Mean | Conventional Units (µg/mL) | |||||
|---|---|---|---|---|---|---|
| PHBRConc. | Repeatability* | Within Lab** | No. ofDays | No. ofObs. | ||
| SD | %CV | SD | %CV | |||
| 5.0 | 0.20 | 3.9% | 0.25 | 5.0% | 22 | 88 |
| 9.1 | 0.27 | 2.9% | 0.33 | 3.6% | 22 | 88 |
| 11.0 | 0.30 | 2.7% | 0.44 | 4.0% | 22 | 88 |
| 23.0 | 0.50 | 2.2% | 0.65 | 2.8% | 22 | 88 |
| 25.1 | 0.59 | 2.4% | 0.81 | 3.2% | 22 | 88 |
| 38.1 | 0.79 | 2.1% | 1.09 | 2.9% | 22 | 88 |
| 59.3 | 1.50 | 2.5% | 2.11 | 3.6% | 22 | 88 |
*Repeatability was determined using two replicates per run.
** Within Lab precision was determined using a single lot of slides and a single calibration
Detection Limits
The Limit of Detection (LoD) and Limit of Quantitation (LoQ) were determined for the VITROS PHBR Slides assay consistent with CLSI document EP17.: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures. The total allowable error goal used to accept the LoQ was ≤ 1.5 µg/mL.
The limit of detection (LoD) for the VITROS PHBR Slides assay is 1.3 ug/mL.
{6}------------------------------------------------
The claimed limit of quantification (LoQ) is 3.0 µg/mL. The Total Error goal used to accept the LoQ was ≤ 1.5 µg/mL.
Linearity
Linearity studies were performed according to CLSI EP06-A. Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approved Guideline (2003). VITROS PHBR Slides were tested on the VITROS 5600 Integrated System. A series of fourteen proportionally related admixtures of low and high concentration fluids were tested to verify linearity of the VITROS PHBR Slides test; each sample was tested in duplicate.
| Lower Limit of Linearity Interval (LLLI) | Upper Limit of Linearity Interval (ULLI) |
|---|---|
| 0.5 µg/mL | 83.5 µg/mL |
The deviation from linearity was within the allowable deviation of ± 1.3 µg/mL at phenobarbital concentrations <10 ug/mL and within a maximum concentration-dependent allowable deviation of ± 11.5 µg/mL at phenobarbital concentrations
10 ug/mL.
The VITROS Chemistry Product PHBR Slides assay is linear over the measuring range of 3.0 - 80.0 µg/mL.
Serum and plasma samples with values greater than the VITROS PHBR Slides test measuring range may be diluted with 1part sample and 1part diluent. VITROS Chemistry Products Specialty Diluent is an acceptable diluent for the VITROS PHBR Slides test.
Expected Values
The expected values of the VITROS PHBR Slides are based on the therapeutic range* of 10.0 to 40.0 µg/mL.
*Burtis CA, Bruns DE. Tietz Fundamentals of Clinical Chemistry and Molecular Diagnostics. 7th ed. St. Louis, MO: Elsevier Saunders; 2015:pg 545.
Each laboratory should confirm the validity of these intervals for the population it serves.
Specificity
The VITROS Chemistry Products PHBR Slide test was screened for interfering substances following CLSI document EP07-A3, Interference Testing in Clinical Chemistry. The supplemental tables in CLSI document EP37 were referenced for recommended testing concentrations for analytes and endogenous substances that may interfere in clinical chemistry measurement procedures.
Point estimates of the effects of potential interferents on the VITROS PHBR Slides were made using the paired difference method. Dose-response analysis was conducted to characterize the degree of interference for each substance that exceeded the acceptance criterion in the initial screening test, and expected bias was reported at the lowest test level which did not meet acceptance criteria for bias as shown in the product claims.
{7}------------------------------------------------
Nine (9) substances were found to interfere with the VITROS Chemistry Products PHBR Slides, bias > 1.8 ug/mL at a phenobarbital concentration of approximately 15 ug/mL or bias > 5.2 ug/mL at a phenobarbital concentration of approximately 50 ug/mL.
The substances listed in the table, when tested at the concentrations indicated, caused the bias shown. The bias is an estimate of the maximum bias observed.
| Interferent | Conventional Units | ||
|---|---|---|---|
| PHBR Conc.(µg/mL) | InterferentConcentration | Bias(µg/mL) | |
| Conjugated bilirubin | 15 | 13.7 mg/dL | 2.1 |
| Ethamsylate | 4.5 mg/dL | 3.4 | |
| Hemoglobin | 300 mg/dL | -2.3 | |
| Mephobarbital | 0.79 mg/dL | 2.6 | |
| Thiamylal | 7.4 mg/dL | 3.1 | |
| Thiopental | 2.5 mg/dL | 4.2 | |
| Total protein | 13.0 g/dL | -2.8 | |
| Valproic acid | 18.4 mg/dL | 2.3 |
| Interferent | Conventional Units | ||
|---|---|---|---|
| PHBR Conc.(µg/mL) | InterferentConcentration | Bias(µg/mL) | |
| Conjugated Bilirubin | 27.4 mg/dL | 6.3 | |
| Ethamsylate | 3.0 mg/dL | 7.2 | |
| Ethanol | 6.0 mg/mL | 7.7 | |
| Hemoglobin | 50 | 225 mg/dL | -8.9 |
| Mephobarbital | 0.79 mg/dL | 6.6 | |
| Thiamylal | 7.4 mg/dL | 8.5 | |
| Thiopental | 10.1 mg/dL | 9.1 | |
| Total protein | 13.0 g/dL | -13.5 |
These results are representative. It is possible that other interfering substances may be encountered in the patient population. The degree of interference at concentrations other than those listed might not be predictable.
Sixty-one (61) test substances, when tested at the concentrations indicated, were found not to interfere with the VITROS PHBR Slides, not to cause bias > 1.8 µg/mL at a phenobarbital concentration of approximately 15 µg/mL or bias > 5.2 µg/mL at a phenobarbital concentration of approximately 50 ug/mL. These will be listed in the customer instructions for use.
{8}------------------------------------------------
Cross-Reactivity
The most common drugs concurrently prescribed with anti-convulsant therapies were reviewed for selection for testing, as well as compounds structurally related to phenobarbital. Substances tested included amobarbital, aprobarbital, butabarbital, butalbital, hexobarbital, mephobarbital, phenylethyl-malonamide (pema), phenytoin, primidone, secobarbital, thiamylal, and thiopental. The cross-reactivity of the VITROS Chemistry Products PHBR Slides method was evaluated by adding the substances to serum samples containing phenobarbital concentrations of approximately 15 and 50 µg/mL. Results are provided for reference in the customer instructions for use.
Other Limitations
Certain drugs and clinical conditions are known to alter phenobarbital concentrations in vivo.
Conclusion
The conclusions drawn from the nonclinical tests (discussed above) demonstrate the VITROS Chemistry Products PHBR Slides for use on the VITROS 5600 Integrated System is as safe, effective, and performs as well as the predicate device. The information submitted in the premarket notification is complete and supports a substantial equivalence decision.
§ 862.3660 Phenobarbital test system.
(a)
Identification. A phenobarbitol test system is a device intended to measure phenobarbital, an antiepileptic and sedative-hypnotic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of phenobarbital use or overdose and in monitoring levels of phenobarbital to ensure appropriate therapy.(b)
Classification. Class II.