K Number
K210723
Date Cleared
2021-07-12

(124 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nitrile Patient Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Device Description

The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.

AI/ML Overview

The provided document is a 510(k) Summary for a medical device called "Nitrile Patient Examination Glove." It outlines the device's characteristics, intended use, and comparison to a predicate device, along with non-clinical performance testing.

Here's the breakdown of the acceptance criteria and study information:

Acceptance Criteria and Reported Device Performance

ItemAcceptance Criteria (Predicate Device K171422)Reported Device Performance (Subject Device K210723)Remark
MaterialNitrileNitrileSame
Physical Properties
Before AgingTensile Strength: 14MPa, minTensile Strength: 14MPa, minSame
Ultimate Elongation: 500% minUltimate Elongation: 500% minSame
After AgingTensile Strength: 14MPa, minTensile Strength: 14MPa, minSame
Ultimate Elongation: 400% minUltimate Elongation: 400% minSame
ComplianceComply with ASTM D6319Comply with ASTM D6319Same
Freedom from HolesBe free from holes when tested in accordance with ASTMD5151 AQL=2.5Be free from holes when tested in accordance with ASTMD5151 AQL=2.5Same
Powder ContentMeet the requirements of ASTM D6124Meet the requirements of ASTM D6124Same
Biocompatibility
IrritationComply with ISO 10993-10:2010Under the conditions of the study, not an irritantThe subject device met the criteria for not being an irritant, which is considered compliant with ISO 10993-10.
SensitizationComply with ISO 10993-10:2010Under conditions of the study, not a sensitizer.The subject device met the criteria for not being a sensitizer, which is considered compliant with ISO 10993-10.
CytotoxicityComply with ISO 10993-5:2009Under conditions of the study, device extract is not cytotoxicThe subject device met the criteria for not being cytotoxic. While the predicate device column for cytotoxicity has a "/", the remark "Similar" implies an acceptable comparison or implied compliance for the predicate, and the subject device explicitly demonstrates compliance.
Device DimensionsASTM D6319-19 requirements (specific values present for predicate)ASTM D6319-19 requirements (specific values present for subject)The physical dimensions are little different but all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

Study Information

  1. Sample sizes used for the test set and the data provenance:
    The document does not specify exact sample sizes for each test. However, it indicates compliance with recognized standards such as ASTM D6319-19, ASTM D5151-19, ASTM D6124-06 (Reapproved 2017), ISO 10993-10:2010, and ISO 10993-5:2009. These standards inherently define appropriate sampling plans for testing.
    The data provenance is from non-clinical tests conducted to verify design specifications. The company, Jiangxi Zhonghong Pulin Medical Products Co.,Ltd., is located in China. The nature of these tests (e.g., laboratory testing of material properties) implies prospective data collection as part of the device development and regulatory submission process.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This information is not applicable as the evaluation relies on established standardized test methods (e.g., ASTM, ISO standards) for material properties and biocompatibility, rather than expert interpretation of a diagnostic output. The "ground truth" is determined by the objective results of these standardized tests against their defined pass/fail criteria.

  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    This is not applicable. The tests performed are objective, quantitative or qualitative measurements governed by international standards, not subjective interpretations requiring adjudication.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This is not applicable. The device is a patient examination glove, not an AI-powered diagnostic or assistive SaMD (Software as a Medical Device). Therefore, MRMC studies are not relevant.

  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
    This is not applicable. The device is a physical medical glove, not an algorithm.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    The "ground truth" for this device is based on objective, standardized test results against the specifications outlined in relevant ASTM and ISO standards for physical properties, barrier integrity, and biocompatibility. For example:

    • Physical properties (Tensile Strength, Elongation): Measured values compared against minimum requirements specified in ASTM D6319.
    • Freedom from Holes: Determined by testing in accordance with ASTMD5151, with an Acceptance Quality Limit (AQL) of 2.5.
    • Powder Content: Measured against requirements of ASTM D6124.
    • Biocompatibility (Irritation, Sensitization, Cytotoxicity): Determined by laboratory assays compliant with ISO 10993-10 and ISO 10993-5, where the outcome is classified as "not an irritant," "not a sensitizer," or "not cytotoxic."
  7. The sample size for the training set:
    This is not applicable. The device is a physical product, not an AI/ML algorithm that requires a training set.

  8. How the ground truth for the training set was established:
    This is not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.

July 12, 2021

Jiangxi Zhonghong Pulin Medical Products Co.,Ltd. % Boyle Wang Shanghai Truthful Information Technology Co., Ltd. RM.608,No.738,Shangcheng Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K210723

Trade/Device Name: Nitrile Patient Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA Dated: June 15, 2021 Received: June 24, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ryan Ortega Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210723

Device Name Nitrile Patient Examination Glove

Indications for Use (Describe)

The Nitrile Patient Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K210723)

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: Jiangxi Zhonghong Pulin Medical Products Co.,Ltd. Address: Yinshawan Park,High-tech Industrial Park,Hukou County, Jiujiang City,Jiangxi Province,China,332500 Phone Number: +86-19931401105 Contact: Zhou Ziyu Date of Preparation: Jun.15,2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Nitrile Patient Examination Glove Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): XS,S, M, L, XL

3.0 Classification

Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Ever Global (Vietnam) Enterprise Corp Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422

{4}------------------------------------------------

5.0 Indication for Use

The Nitrile Patient Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

6.0 Device Description

The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.

7.0 Technological Characteristic Comparison Table

ItemSubject Device(K210723)Predicated Device(K171422)Remark
Product CodeLZALZASame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseThe Nitrile PatientExamination Gloveis a non-steriledisposable deviceintended formedical purposesthat is worn on theexaminer's handsor finger to preventcontaminationbetween patientand examiner.The Nitrile PowderFree patientexamination gloveis a non-steriledisposable deviceintended formedical purposesthat is worn on theexaminer's handsor finger to preventcontaminationbetween patientand examiner.Same
Powdered orPowered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
Labeling InformationSingle-useindication, powderfree, device color,device name,glove size andquantity, NitrileGlove PowderSingle-useindication, powderfree, device color,device name, glovesize and quantity,Disposable PowderFree NitrileSame

Table1-General Comparison

{5}------------------------------------------------

Free Blue, Non-SterileExamination Glove, Non-Sterile
----------------------------------------------------------
PredicateDevice(K171422)DesignationSizeTolerance
Length, mmXSSMLXLmin
Width, mm758595105115±5
Thickness, mm:
Finger0.05min
Palm0.05min
Subject Device(K210723)DesignationSizeTolerance
Length, mmXSSMLXLmin
Width, mm708095110120±10
Thickness, mm:
Finger0.05min
Palm0.05min
RemarkSIMILAR

Table2 Device Dimensions Comparison

Analysis: The physical dimensions are little different with that of the predicate, but they all meet the requirements of ASTM D6319-19,so the differences do not raise any new safety or performance questions.

Table3 Performance Comparison

ItemSubject device(K210723)Predicated device(K171422)Remark
ColorantBlueWhite/ Blue/ Black/ PinkSame
PhysicalPropertiesBefore AgingTensile Strength14MPa, min14MPa, minSame
Ultimate Elongation500% min500% minSame
After AgingTensile Strength14MPa, min14MPa, minSame
Ultimate Elongation400%min400%minSame
Comply with ASTM D6319Comply with ASTM D6319Same
Freedom from HolesBe free from holes when tested inBe free from holes when tested inSame

{6}------------------------------------------------

accordancewithASTMD5151AQL=2.5accordancewithASTMD5151AQL=2.5
Powder ContentMeet therequirementsof ASTMD6124Meet therequirementsof ASTMD6124Same

Table4 Safety Comparison

ItemSubjectdevice(K210723)Predicateddevice(K171422)Remark
MaterialNitrileNitrileSame
BiocompatibilityIrritation (ISO10993-10:2010BiologicalEvaluation ofMedical Devices -Part 10: Tests ForIrritation And SkinSensitization)Under theconditions ofthe study, notan irritant
Sensitization(ISO 10993-10:2010BiologicalEvaluation ofMedical Devices -Part 10: Tests ForIrritation And SkinSensitization)Underconditions ofthe study, nota sensitizer.Comply withISO10993-10Same
Cytotoxicity (ISO10993-5:2009BiologicalEvaluation ofMedical Devices -Part 5: Tests ForIn VitroCytotoxicity)Underconditions ofthe study,deviceextract is notcytotoxic/Similar

8.0 Discussion of Non-clinical and Performance Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device

{7}------------------------------------------------

complies with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

9.0 Discussion of Clinical and Performance Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.