(144 days)
Not Found
No
The device is a disposable medical examination glove, and the summary focuses on physical and biological performance characteristics, with no mention of AI or ML.
No
The device, a disposable nitrile medical examination glove, is intended for contamination prevention and examination purposes, not for treating or preventing disease or injury.
No
The device description indicates that it is a "Disposable Nitrile Medical Examination Glove," which is used to prevent contamination, not to diagnose a medical condition.
No
The device is a physical medical device (gloves) and does not involve any software component.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to detect diseases, conditions, or infections. These tests are performed outside the body (in vitro).
- Device Description and Intended Use: The description clearly states the device is a "Disposable Nitrile Medical Examination Glove" intended to be "worn on the examiner's hands to prevent contamination between patient and examiner." This is a barrier device used for protection and hygiene, not for performing diagnostic tests on patient samples.
The information provided focuses on the physical properties and biocompatibility of the glove, which are relevant for a medical device worn on the skin, but not for an IVD.
N/A
Intended Use / Indications for Use
The Disposable Nitrile Medical Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Product codes (comma separated list FDA assigned to the subject device)
LZA
Device Description
The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
hands
Indicated Patient Age Range
Not Found
Intended User / Care Setting
examiner
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
Key results:
Physical Dimensions Test (ASTM D6319):
Lot1: Length: >230; Width: XS: 78-80; S: 85-86; M: 95-96; L: 105-106; XL: 115-116; Finger Thickness: 0.11-0.13; Palm Thickness: 0.06-0.07. Pass.
Lot2: Length: >230; Width: XS: 79-80; S: 85-86; M: 95-96; L: 105-106; XL: 115-116; Finger Thickness: 0.11-0.13; Palm Thickness: 0.06-0.07. Pass.
Lot3: Length: >230; Width: XS: 78-80; S: 85-86; M: 95-96; L: 105-106; XL: 115-116; Finger Thickness: 0.11-0.13; Palm Thickness: 0.07. Pass.
Watertightness Test for Detection of Holes (ASTM D5151):
Lot1: 0/125 leaks. Pass.
Lot2: 0/125 leaks. Pass.
Lot3: 0/125 leaks. Pass.
Powder Content (ASTM D6124):
Lot1: 0.09mg/glove. Pass.
Lot2: 0.06mg/glove. Pass.
Lot3: 0.08mg/glove. Pass.
Physical properties (ASTM D412):
Before Aging:
Tensile Strength: Lot 1: 28-36; Lot2: 28-33; Lot3: 23-33. Pass.
Ultimate Elongation: Lot1: 510-570; Lot 2: 480-565; Lot3: 515-566. Pass.
After Aging:
Tensile Strength: Lot 1: 27-33; Lot2: 28-33; Lot3: 28-32. Pass.
Ultimate Elongation: Lot1: 485-535; Lot 2: 480-530; Lot3: 485-533. Pass.
Cytotoxicity (ISO 10993-5):
Under conditions of the study, did not show potential toxicity to L-929 cells. Pass.
Irritation (ISO 10993-10):
Under the conditions of the study, not an irritant. Pass.
Sensitization (ISO 10993-10):
Under conditions of the study, not a sensitizer. Pass.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 30, 2021
Jiangsu Cureguard Glove Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China
Re: K210691
Trade/Device Name: Disposable Nitrile Medical Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: June 15, 2021 Received: June 22, 2021
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K210691
Device Name Disposable Nitrile Medical Examination Glove
Indications for Use (Describe)
The Disposable Nitrile Medical Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable) | |
---|---|
------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary (K210691)
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
1.0 Submitter's Information
Name: Jiangsu Cureguard Glove Co., Ltd. Address: No.65 Shenzhen Road, The Economic Development Zone, Sugian , Jiangsu, 223800 China. Phone Number: +86-13485097856 Contact: Guo Hua Date of Preparation: Jun.30,2021
Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn
2.0 Device Information
Trade name: Disposable Nitrile Medical Examination Glove Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): XS,S, M, L, XL
3.0 Classification
Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital
4.0 Predicate Device Information
Manufacturer: Ever Global (Vietnam) Enterprise Corp Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422
4
5.0 Indication for Use
The Disposable Nitrile Medical Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
6.0 Device Description
The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.
| Item | Subject Device
(K210691) | Predicated Device
(K171422) |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | LZA | LZA |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 |
| Class | I | I |
| Intended Use | Disposable Nitrile Medical
Examination Glove is a
disposable device intended
for medical purposes that is
worn on the examiner's
hands to prevent
contamination between
patient and examiner. | The Disposable Powder Free
Nitrile Examination Glove,
White/ Blue/ Black/ Pink Color
is a disposable device
intended for medical purposes
that is worn on the examiner's
hands to prevent
contamination between patient
and examiner. |
| Material | Nitrile | Nitrile |
| Powdered or
Powered free | Powdered free | Powdered free |
| Design Feature | Ambidextrous | Ambidextrous |
| Colorant | Blue | White/ Blue/ Black/ Pink |
| Labeling Information | Single-use
indication,
powder free, device color,
device name, glove size and
quantity,
Non-Sterile | Single-use indication, powder
free, device color, device
name, glove size and quantity,
Non-Sterile |
| Dimensions(mm) | Length:
XS/S: $\geq$ 220;
M/L/XL: $\geq$ 230;
Width:
XS: 70 $\pm$ 10;
S: 80 $\pm$ 10; | Length: $\geq$ 230;
Width:
XS: 75 $\pm$ 5;
S: 85 $\pm$ 5;
M: 95 $\pm$ 5;
L: 105 $\pm$ 5; |
7.0 Technological Characteristic Comparison Table
5
M: 95±10; | XL: 115±5 | ||||
---|---|---|---|---|---|
L: 110±10; | |||||
XL: 120±10 | |||||
Thickness(mm) | Finger: ≥0.05; | Finger: ≥0.05; | |||
Palm: ≥0.05 | Palm: ≥0.05 | ||||
Physical | |||||
Properties | Before | ||||
Aging | Tensile | ||||
Strength | 14MPa, min | Tensile | |||
Strength | 14MPa, min | ||||
Ultimate | |||||
Elongation | 500% min | Ultimate | |||
Elongation | 500% min | ||||
After | |||||
Aging | Tensile | ||||
Strength | 14MPa, min | Tensile | |||
Strength | 14MPa, min | ||||
Ultimate | |||||
Elongation | 400%min | Ultimate | |||
Elongation | 400%min | ||||
Freedom from | |||||
Holes | Be free from holes when | ||||
tested in accordance with | |||||
ASTMD5151 AQL=2.5 | Be free from holes when tested | ||||
in accordance with | |||||
ASTMD5151 AQL=2.5 | |||||
Powder Content | Meet the requirements of | ||||
ASTM D6124 | |||||
Biocompatibility | ISO 10993-10; | ||||
Under the conditions of the | |||||
study, not an irritant or a | |||||
sensitizer | |||||
ISO 10993-5 | |||||
Under conditions of the | |||||
study, device extract is not | |||||
cytotoxic | Comply with | ||||
ISO 10993-10 |
Comply with ISO 10993-5 | |
8.0 Discussion of Non-clinical and Performance Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
6
ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
| Test
Methodology | Purpose | Acceptance Criteria | Results |
---|---|---|---|
ASTM D6319 | Physical | ||
Dimensions | |||
Test | Length(mm): | ||
XS/S:≥220; | |||
M/L/XL: ≥230; | Lot1: | ||
Length: >230 | |||
Width(mm): | |||
XS: 70±10; | |||
S: 80±10; | |||
M: 95±10; | |||
L: 110±10; | |||
XL: 120±10 | Width: | ||
XS: 78-80; | |||
S: 85-86 | |||
M: 95-96 | |||
L: 105-106 | |||
XL: 115-116 | |||
Lot2: | |||
Length: >230 | |||
Width: | |||
XS: 79-80 | |||
S: 85-86 | |||
M: 95-96 | |||
L: 105-106 | |||
XL: 115-116 | |||
Lot3: | |||
Length: >230 | |||
Width: | |||
XS: 78-80 | |||
S: 85-86 | |||
M: 95-96 | |||
L: 105-106 | |||
XL: 115-116 | |||
Pass | |||
Thickness (mm) : | |||
Finger: ≥0.05 | |||
Palm: ≥0.05 | Lot1: | ||
Finger: 0.11-0.13 | |||
Palm: 0.06-0.07 | |||
Lot2: | |||
Finger: 0.11-0.13 | |||
Palm: 0.06-0.07 | |||
Lot3: |
Table 2: Performance Characteristic Comparison
7
| | | | | Finger: 0.11-0.13
Palm: 0.07
Pass | |
|------------|--------------------------------------------|---------------------------------------------|---------------------|------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| ASTM D5151 | Watertightness Test for Detection of Holes | Meet the requirements of ASTM D5151 AQL 2.5 | | Lot1:
0/125 leaks
Lot2:
0/125 leaks
Lot3:
0/125 leaks
Pass | |
| ASTM D6124 | Powder Content | Meet the requirements of ASTM D6124