K Number
K210684
Date Cleared
2021-07-20

(134 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Nitrile Medical Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Device Description

The subject device is powder free nitrile examination gloves. The subject device is white. The subject device is non-sterile.

AI/ML Overview

The provided document is a 510(k) summary for a Disposable Nitrile Medical Examination Glove. It details the non-clinical testing performed to demonstrate substantial equivalence to a predicate device. This document is not for an AI/ML medical device, but rather for a physical examination glove. Therefore, many of the requested criteria regarding AI/ML device acceptance, such as expert consensus, MRMC studies, training/test sets, and ground truth establishment, are not applicable to this submission.

However, I can extract and present the acceptance criteria and test results related to the physical and biocompatibility performance of the glove, as described in the document.

Acceptance Criteria and Device Performance for Disposable Nitrile Medical Examination Glove (K210684)

The device, a "Disposable Nitrile Medical Examination Glove," is classified as Class I and its performance is evaluated against established ASTM and ISO standards for medical gloves, rather than complex AI/ML performance metrics.

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodologyPurposeAcceptance Criteria Reported in DocumentReported Device Performance (Results)
ASTM D6319Physical Dimensions TestLength (mm):- XS/S: ≥220- M/L/XL: ≥230Length:- XS/S: > 220/Pass- M/L/XL: > 230/Pass
Width (mm):- XS: 70±10- S: 80±10- M: 95±10- L: 110±10- XL: 120±10Width:- XS: 78-80 / Pass- S: 85-89 /Pass- M: 95-99/ Pass- L: 105-111/ Pass- XL: 115-120/ Pass
Thickness (mm):- Finger: ≥0.05- Palm: ≥0.05Thickness:- Finger: 0.09-0.13/Pass- Palm: 0.07-0.13/Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125 leaks / Pass
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0 mg0.07 mg/Pass
ASTM D412Physical Properties (Before Aging)Tensile Strength: ≥14 MPaUltimate Elongation: ≥500%Tensile Strength:- XS:29-31/Pass- S:27-32/Pass- M:28-33/Pass- L:25-33/Pass- XL:28-31/PassUltimate Elongation:- XS:520-565/Pass- S:520-575/Pass- M:520-565/Pass- L:510-565/Pass- XL:525-563/Pass
ASTM D412Physical Properties (After Aging)Tensile Strength: ≥14 MPaUltimate Elongation: ≥400%Tensile Strength:- XS:26-31/Pass- S:28-32/Pass- M:27-31/Pass- L:28-33/Pass- XL:28-31/PassUltimate Elongation:- XS:480-520/Pass- S:490-525/Pass- M:485-530/Pass- L:490-545/Pass- XL:480-530/Pass
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells./ Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant/ Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer./ Pass

Non-Applicable Criteria for this Device:

The following requested information is primarily relevant to AI/ML (Artificial Intelligence/Machine Learning) medical devices, which typically involve algorithms processing data (e.g., images, signals) to make predictions or assist in diagnoses. The submitted device, a medical examination glove, is a physical product and its testing is based on established performance standards for materials and manufacturing. Therefore, these items are not applicable to this 510(k) submission:

2. Sample size used for the test set and the data provenance: Not applicable. Tests are standard physical and chemical property evaluations on manufactured gloves, not a data-based test set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth is defined by physical/chemical standards, not expert medical opinion on a data set.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. No AI component.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. No algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The ground truth for this device is based on established international standards (ASTM, ISO) for physical properties and biocompatibility of medical gloves.
8. The sample size for the training set: Not applicable. No training set for an algorithm.
9. How the ground truth for the training set was established: Not applicable.

Summary of Study (Non-Clinical Testing):

The study described is a series of non-clinical tests conducted to verify that the Disposable Nitrile Medical Examination Glove meets design specifications and complies with relevant international standards. The tests cover:

  • Physical Dimensions: Length, width, and thickness for various glove sizes.
  • Physical Properties: Tensile strength and ultimate elongation, both before and after accelerated aging, to assess durability and flexibility.
  • Freedom from Holes: Watertightness testing to ensure barrier integrity.
  • Powder Content: To confirm it is "powder free" as claimed.
  • Biocompatibility: Tests for irritation, sensitization, and cytotoxicity to ensure the material is safe for contact with human skin.

The document reports that all tests demonstrated the proposed device complies with the specified standards and meets the acceptance criteria, leading to the conclusion that it is as safe and effective as the legally marketed predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 20, 2021

Jiangsu Cureguard Glove Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai 200120 China

Re: K210684

Trade/Device Name: Disposable Nitrile Medical Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: June 15, 2021 Received: June 28, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210684

Device Name Disposable Nitrile Medical Examination Glove

Indications for Use (Describe)

The Disposable Nitrile Medical Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K210684)

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: Jiangsu Cureguard Glove Co., Ltd. Address: No.65 Shenzhen Road, The Economic Development Zone, Sugian, Jiangsu, 223800 China. Phone Number: +86-13485097856 Contact: Guo Hua Date of Preparation: July 19, 2021

Designated Submission Correspondent

Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Disposable Nitrile Medical Examination Glove Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): XS,S, M, L, XL

3.0 Classification

Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Ever Global (Vietnam) Enterprise Corp Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422

{4}------------------------------------------------

5.0 Indication for Use

The Disposable Nitrile Medical Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

6.0 Device Description

The subject device is powder free nitrile examination gloves. The subject device is white. The subject device is non-sterile.

7.0 Technological Characteristic Comparison Table

Table1-General Comparison
ItemSubject Device(K210684)Predicate Device(K171422)Remark
Product CodeLZALZASame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseDisposable NitrileMedicalExamination Gloveis a disposabledevice intended formedical purposesthat is worn on theexaminer's hands topreventcontaminationbetween patient andexaminer.The DisposablePowder Free NitrileExamination Glove,White/ Blue/ Black/Pink Color is adisposable deviceintended for medicalpurposes that is wornon the examiner'shands to preventcontaminationbetween patient andexaminer.Similar
Powdered or Powered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
Labeling InformationSingle-useindication, powderfree, device color,device name, glovesize and quantity,Non-SterileSingle-use indication,powder free, devicecolor, device name,glove size andquantity, Non-SterileSimilar

Table1-General Comparison

Table2 Device Dimensions Comparison

{5}------------------------------------------------

Size
PredicateDevice(K171422)DesignationXSSMLXLTolerance
Length, mm230230230230230min
Width, mm758595105115±5
Thickness, mm:
Finger0.05min
Palm0.05min
SubjectDevice(K210684)DesignationXSSMLXLTolerance
Length, mm220220230230230min
Width, mm708095110120±10
Thickness, mm:
Finger0.05min
Palm0.05min
RemarkSimilar

Table3 Performance Comparison

ItemSubject device(K210684)Predicateddevice(K171422)Remark
ColorantWhiteWhite/ Blue/Black/ PinkSimilar
PhysicalPropertiesBeforeAgingTensileStrength14MPa, min14MPa, minSame
UltimateElongation500% min500% minSame
AfterAgingTensileStrength14MPa, min14MPa, minSame
UltimateElongation400%min400%minSame
Comply with ASTM D6319Comply withASTM D6319Comply withASTM D6319Same
Freedom from HolesBe free fromholes whentested inaccordance withASTMD5151AQL=2.5Be free fromholes whentested inaccordancewithASTMD5151AQL=2.5Same
Powder ContentMeet therequirements ofASTM D6124Meet therequirements ofASTM D6124Same

{6}------------------------------------------------

ItemSubject device(K210684)Predicateddevice(K171422)Remark
MaterialNitrileNitrileSame
BiocompatibilityIrritation (ISO 10993-10:2010BiologicalEvaluation ofMedical Devices -Part 10: Tests forIrritation And SkinSensitization)Under theconditions ofthe study, notan irritant
Sensitization (ISO10993-10:2010BiologicalEvaluation ofMedical Devices -Part 10: Tests forIrritation And SkinSensitization)Underconditions ofthe study, not asensitizer.Comply withISO10993-10Similar
Cytotoxicity (ISO10993-5:2009BiologicalEvaluation ofMedical Devices -Part 5: Tests For InVitro Cytotoxicity)Underconditions ofthe study,device extractis not cytotoxic/Comply withISO 10993-5Similar

Table4 Safety Comparison

8.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

{7}------------------------------------------------

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

Test MethodologyPurposeAcceptance CriteriaResults
ASTM D6319PhysicalDimensionsTestLength(mm):XS/S:≥220;M/L/XL:≥230;Width(mm):XS:70±10;S: 80±10;M: 95±10;L: 110±10;XL: 120±10;Thickness (mm):Finger: ≥0.05Palm: ≥0.05Length:XS/S: > 220/Pass;M/L/XL: > 230/Pass;Width:XS: 78-80/ PassS: 85-89 /PassM: 95-99/ PassL: 105-111/ PassXL: 115-120/ PassFinger: 0.09-0.13/PassPalm: 0.07-0.13/Pass
ASTM D5151WatertightnessTest forDetection ofHolesMeet the requirements of ASTM D5151 AQL 2.50/125 leaks / Pass
ASTM D6124PowderContentMeet the requirements of ASTM D6124 < 2.0mg0.07 mg/Pass
ASTM D412PhysicalpropertiesBefore Aging Tensile Strength ≥14MPaXS:29-31/PassS:27-32/PassM:28-33/PassL:25-33/PassXL:28-31/Pass
Ultimate Elongation ≥500%XS:520-565/PassS:520-575/PassM:520-565/PassL:510-565/PassXL:525-563/Pass
After Aging Tensile Strength ≥14MPaXS:26-31/PassS:28-32/PassM:27-31/PassL:28-33/PassXL:28-31/Pass
Ultimate Elongation ≥400%XS:480-520/PassS:490-525/Pass
M:485-530/PassL:490-545/PassXL:480-530/Pass
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of thestudy, did not showpotential toxicity to L-929 cells./ Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions ofthe study, not an irritant/Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditions of thestudy, not a sensitizer./Pass

Table 5 Summary of non-clinical performance testing

{8}------------------------------------------------

9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated device in K171422.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.