K Number
K210615
Date Cleared
2022-05-13

(438 days)

Product Code
Regulation Number
882.4800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Thompson Brain Retractor Table Mounted and Skull Clamp Mounted Systems are indicated for self-retaining retraction of soft tissue during neurological procedures to provide surgical access and exposure. The Thompson Brain Retractor Systems also allow attachment for applicable accessories and function as a hand-rest for the surgeon.

Device Description

Thompson Surgical Instruments Brain Retractor is a reusable manual instrument made from stainless steel and aluminum, as indicated. It is composed of materials that can be reused with steam sterilization methods. It is provided to the customer non-sterile. The Thompson Brain Retractor is a retractor system used to create adequate exposure of the cranial region to assist surgeons in accessing areas for surgery. The system is comprised of a stainless-steel frame with 1/2-inch diameter frame components, which create a plane over the access point in the skull of the patient. The system uses stainless steel flexible arms that are attached to the frame to hold blades (stainless steel), which ultimately retract the tissue for the needed exposure. The frame also acts as a rest for the surgeons' wrists/hands during the procedure, as well as an attachment point for various accessories to aid the surgeon (ex: pattie tray).

The system uses clamps, cam actuation, and screws to construct and support the frame. The flexible arms are attached to the frame via screw down mechanism and the flexible arms are tightened into a semi-rigid state also using a screw mechanism, very similar to the function of all the predicate devices.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Thompson Brain Retractor

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.04.33
Silver Spring, MD 20993
www.fda.gov

May 13, 2022

Thompson Surgical Instruments Inc.
Tiffany Bush
Regulatory Engineer
10341 E Cherry Bend Road
Traverse, Michigan 49684

Re: K210615
Trade/Device Name: Thompson Brain Retractor Table Mounted (TM), Thompson Brain Retractor Skull Clamp Mounted (SCM)
Regulation Number: 21 CFR 882.4800
Regulation Name: Self-Retaining Retractor For Neurosurgery
Regulatory Class: Class II
Product Code: GZT
Dated: April 11, 2022
Received: April 12, 2022

Dear Tiffany Bush:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

Page 2

K210615 - Tiffany Irwin Page 2

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce -S

Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 3

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K210615

Device Name:
Thompson Brain Retractor Skull Clamp Mounted
Thompson Brain Retractor Table Mounted

Indications for Use (Describe):
The Thompson Brain Retractor Table Mounted and Skull Clamp Mounted Systems are indicated for self-retaining retraction of soft tissue during neurological procedures to provide surgical access and exposure. The Thompson Brain Retractor Systems also allow attachment for applicable accessories and function as a hand-rest for the surgeon.

Type of Use (Select one or both, as applicable):
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number"

FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 4

510(k) Summary

K210615

Submitter Name: Thompson Surgical Instruments, Inc.

Submitter Address: 10341 E Cherry Bend Rd
Traverse City, MI 49684

Contact Person: Tiffany Bush, Regulatory Engineer

Phone Number: 231-303-1854
Fax: 231-922-0174
Email: tiffany.bush@thompsonsurgical.com

Date Prepared: May 10, 2022

Device Trade Name: Thompson Brain Retractor Table Mounted (TM), Thompson Brain Retractor Skull Clamp Mounted (SCM)

Classification Name: Self-retaining retractor for neurosurgery

Device Class: Class II

Classification Regulation: 21 CFR 882.4800

Product Code: GZT

Predicate Device: Greenberg Retractor and Handrest (pre-amendment device)

Reference Devices:

  • Phantom Fukushima Neurological Holding Systems, Phantom Mastability Neurological Holding Systems (K161318)
  • Budde-Halo Retractor (K830332)

Indications for Use: The Thompson Brain Retractor Table Mounted and Skull Clamp Mounted Systems are indicated for self-retaining retraction of soft tissue during neurological procedures to provide surgical access and exposure. The Thompson Brain Retractor Systems also allow attachment for applicable accessories and function as a hand-rest for the surgeon.

Device Description: Thompson Surgical Instruments Brain Retractor is a reusable manual instrument made from stainless steel and aluminum, as indicated. It is composed of materials that can be reused with steam sterilization methods. It is provided to the customer non-sterile. The Thompson Brain Retractor is a retractor system used to create

Page 5

K210615, Page 2 of 7

adequate exposure of the cranial region to assist surgeons in accessing areas for surgery. The system is comprised of a stainless-steel frame with 1/2-inch diameter frame components, which create a plane over the access point in the skull of the patient. The system uses stainless steel flexible arms that are attached to the frame to hold blades (stainless steel), which ultimately retract the tissue for the needed exposure. The frame also acts as a rest for the surgeons' wrists/hands during the procedure, as well as an attachment point for various accessories to aid the surgeon (ex: pattie tray).

The system uses clamps, cam actuation, and screws to construct and support the frame. The flexible arms are attached to the frame via screw down mechanism and the flexible arms are tightened into a semi-rigid state also using a screw mechanism, very similar to the function of all the predicate devices.

Technological and Performance Characteristics: The Thompson Brain Retractor was found to be similar to the predicate. Three fundamental similarities are:

  1. Basic design: Both the Thompson Brain Retractor and the predicate device consist of stainless-steel frames for surrounding the cranial region with flexible arms and blades that retract tissue during surgery. General size, shape, weight, and materials are not identical but are very similar.

  2. Sterilization and Biocompatibility: Both the Thompson Brain Retractor and predicate device are designed for repeat use and can be reprocessed with cleaning and steam sterilization.

  3. Use: Both the Thompson Brain Retractor and predicate device are manipulated using common mechanisms to lock and unlock the components from one another. The devices are to be used in a healthcare facility by a trained health care personnel or medical professional such as a surgeon, PA circulating nurse, surgical tech, etc.

Primary differences are related to minor differences in size, shape, and overall look. However, performance data demonstrates that the differences do not raise new questions of safety and effectiveness. Please see further comparison below.

Page 6

K210615, Page 3 of 7

Comparison to Predicate Device:

Table 5.1: Predicate and Reference Device Comparison

DescriptionThompson Brain Retractor SystemK161318 TeDan Fukushima Reference DevicePre-amendment Codman Greenberg (sold by Symmetry, Mediflex, Integra) Predicate DeviceK830332 Budde-Halo Reference Device
Indications for UseThe Thompson Brain Retractor Table Mounted and Skull Clamp Mounted Systems are indicated for self-retaining retraction of soft tissue during neurological procedures to provide surgical access and exposure. The Thompson Brain Retractor Systems also allow attachment for applicable accessories and function as a hand-rest for the surgeon.SimilarSimilarSimilar
Intended UseThe Thompson Retractor is intended for use during surgical procedures in order to provide surgical access and exposure.SameSameSame
Product CodeGZTSameSameSame
Target PopulationIntended for surgeon use on individuals, adult or pediatric, during surgerySameSameSame
Anatomical SiteCranial RegionSameSameSame
Where UsedHospital, health facilitySameSameSame
Human FactorsHighly trained surgeon or other health professional intended user with general knowledge of similar types of deviceSameSameSame
DesignMitigation of use related hazards---
MaterialStainless Steel and Anodized AluminumStainless Steel, Anodized Aluminum, PolymerStainless SteelStainless Steel, Composite Carbon Fiber, Titanium Arms

Page 7

K210615, Page 4 of 7

DescriptionThompson Brain Retractor SystemK161318 TeDan Fukushima Reference DevicePre-Amendment Codman Greenberg (sold by Symmetry, Mediflex, Integra)K830332 Budde-Halo Reference Device
SizesVarious blade sizes, and various frame component sizes, lengths, and configurationsSameSameSame
FrameTable Mounted and Skull Clamp MountedTable Mounted and Skull Clamp MountedSkull Clamp Mounted, can be Table MountedSkull Clamp Mounted, can be Table Mounted
FrameCircular or CurvedCurved ArmsStraight ArmsCircular
Flex Arm Length9"14.5", 16.5" and 6.5"9.5"9", optional 6" and 12"
BladesFlat Malleable Tapered Blades: Length: 9" Widths: 1/4" x 3/4", 5/32" x 5/8", 1/8" x 1/2", 3/32" x 3/8", 1/16"x 1/4" Straight Malleable Blades with 1/8" rod attachment: 6mm x 102mm, 9mm x 102mm, 16mm x 102mm, 19mm x 102mm, 25mm x 102mmSimilarSameSimilar, click-in mechanism blades
Classification21 CFR 882.4800SameSameSame
Device DescriptionThe Thompson Brain Retractor is a retractor system used to create adequate exposure of the cranial region to assist surgeons in accessing areas for surgery. The system is comprised of a stainless-steel frame with 1/2-inch diameter frame components, which create a plane over the access point in the skull ofSimilar, knobs and pattie tray are anodized aluminum, instrument holder is polymerSimilar, Uses straight frame componentsSimilar, also comes in a radiolucent version. Frame components are not 1/2-inch diameter

Page 8

K210615, Page 5 of 7

DescriptionThompson Brain Retractor SystemK161318 TeDan Fukushima Reference DevicePre-Amendment Codman Greenberg (sold by Symmetry, Mediflex, Integra)K830332 Budde-Halo Reference Device
the patient. The system uses stainless steel flexible arms that are attached to the frame to hold blades (stainless steel), which ultimately retract the tissue for the needed exposure. The frame also acts as a rest for the surgeons' wrists/hands during the procedure, as well as an attachment point for various accessories to aid the surgeon (ex: pattie tray).
Compatibility with other devicesSCM – Compatible with most skull clamps, test attachment with skull clamp before use. TM – Compatible with operating room table.SimilarCompatible with many Skull Clamps or may attach to the side rails of the operating room table.Compatible with MAYFIELD Skull Clamps, has separate side rail supports to facilitate use without skull clamp.
SterilityNon-Sterile, no single use componentsNon-Sterile, has some single-use componentsNon-SterileNon-Sterile

Page 9

K210615, Page 6 of 7

Non-Clinical Performance Studies:

A brief discussion of tests used to support the conclusion of substantial equivalence with the predicate device is provided below.

TestTest Method SummaryResult
SterilizationThe Thompson Brain Retractor and predicate device are provided non-sterile and can be cleaned and reprocessed with similar guidelines. The Thompson Brain Retractor can be reprocessed following the Thompson Brain Retractor IFU (brifu) through comparative analysis of the design features with previously validated components. The Thompson Brain Retractor contains no single-use components.Device met acceptance criteria in the comparative analysis for achieving a minimum sterility assurance level (SAL) of 10-6 when compared to similar previously validated components.
BiocompatibilityThe blade components of the Thompson Brain Retractor Systems are intended to come into direct contact with human tissue during use and were evaluated and tested per the requirements of ISO 10993-1 and the recommendations of the FDA Guidance, "Use of ISO-10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process." The battery of testing included the following endpoint tests:Device met the acceptance criteria for all tests. Passing results in all endpoint testing demonstrates the patient contacting devices are biocompatible and meet the standard requirements for ISO 10993-1.
Cytotoxicity (MEM Elution)Pass – Non-cytotoxic
Sensitization (Guinea Pig Maximization Test)Pass – Non-sensitizer
Irritation (Intracutaneous Reactivity)Pass – Non-irritant
Acute Systemic ToxicityPass
Material Mediated PyrogenicityPass
Hemolysis (Indirect)Pass
MechanicalVarious strength and life cycle testing was performed on components of the Brain Retractor Systems in comparison to the predicate device. The tests specified below were to compare strength and life cycle of key components the subject device with those of the predicate. The mechanical testing specified for key components included:Devices passed all testing per the acceptance criteria and performed as well as the predicate device.
Blades – Life cycle bend comparison to predicatePass – met acceptance criteria performed as well as the predicate.
Flexible Arm Holder – Life cycle of screw-down mechanismPass – met acceptance criteria with no new questions of safety and effectiveness.

Page 10

K210615, Page 7 of 7

TestTest Method SummaryResult
Flexible Arm Holder – Flex Arm Strength and comparison to predicate failure modePass – met acceptance criteria and performed as well as the predicate.
Flexible Arm Holder – Life cycle of blade attachment mechanismPass – met acceptance criteria with no new questions of safety and effectiveness.
Frame Components – Strength of assembled framePass – met acceptance criteria with no new questions of safety and effectiveness.
Skull Clamp Mount – Screw-down strength comparison to predicatePass – met acceptance criteria and performed as well as the predicate.
Skull Clamp Mount – Cycle testing of half inch rod connectionPass – met acceptance criteria with no new questions of safety and effectiveness.
FunctionalValidation with a neurosurgeon in a simulated clinical environment covering the clinical workflow with the subject devices and a comparison predicate device was conducted to assure no new issues of safety or effectiveness were raised.Device met acceptance criteria and performed as well as the comparison predicate device.

Clinical Performance Studies: This premarket notification report does not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

Conclusion: The Thompson Brain Retractor Table Mounted and Thompson Brain Retractor Skull Clamp Mounted are substantially equivalent to the Greenberg Retractor and Handrest (pre-amendment device). The differences between the Thompson Brain Retractor and the predicate do not change the intended use or raise new questions of safety and effectiveness.

§ 882.4800 Self-retaining retractor for neurosurgery.

(a)
Identification. A self-retaining retractor for neurosurgery is a self-locking device used to hold the edges of a wound open during neurosurgery.(b)
Classification. Class II (performance standards).