K Number
K210566
Date Cleared
2021-03-25

(27 days)

Product Code
Regulation Number
888.3565
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Total Knee components are indicated for rheumatoid arthritis, osteoarthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartment, or total knee replacement. LEGION Porous CR Narrow Femoral Components are indicated for use without bone cement and are single use devices.

Device Description

The subject devices of this premarket notification are the LEGION Porous CR Narrow Femoral Components, which include catalog item offerings with and without HA Coating. These Femoral Components are indicated for use without bone cement applications. The subject devices are a line extension of the predicate devices listed, which are cleared under 510(k) submission numbers K073325 and K091543, with a modification in their femoral components' outer dimensions making them Narrow Femoral Components (similar to the cleared reference device: LEGION Narrow Oxinium CR and PS Femoral Components and Device Specific Instruments cleared under K112941). The overall design of the subject device is geometrically identical to the predicate devices LEGION Porous Femoral Components (K073325, K091543) with their outer dimensional changes making them Narrow Femoral Components (Reference device K112941).

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, specifically LEGION Porous CR Narrow Femoral Components. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials as would be required for a PMA. As such, the information you're requesting regarding acceptance criteria and performance studies in the context of AI/algorithms, human readers, and ground truth is not applicable to this document.

This document pertains to a Class II orthopedic implant (knee joint prosthesis), which is a physical device, not a software or AI-driven diagnostic tool. The performance testing conducted is engineering bench testing to ensure mechanical equivalence, not statistical performance metrics typically associated with AI.

Here's why the specific questions are not answered by the provided text:

  • 1. A table of acceptance criteria and the reported device performance: The document states that "the bench performance testing met the predefined acceptance criteria," but it does not provide the specific acceptance criteria or the quantitative results of the tests (e.g., maximum stress, fatigue cycles, contact area measurements). It lists the types of tests conducted (Tibiofemoral constraint analysis, Tibiofemoral contact area analysis, Patellofemoral resistance to lateral subluxation testing).
  • 2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable. This refers to bench testing of physical components, not data-driven studies.
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. "Ground truth" for this device is established through engineering specifications and physical measurements, not expert human interpretation.
  • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. There are no human adjudicators for the mechanical performance of this implant.
  • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
  • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For bench testing, the "ground truth" would be the engineering specifications and physical measurements obtained from calibrated testing equipment. The document does not detail these specifics.
  • 8. The sample size for the training set: Not applicable. There is no training set for this type of physical device.
  • 9. How the ground truth for the training set was established: Not applicable.

In summary: The provided document is for a medical device clearance based on substantial equivalence to predicate devices, supported by engineering bench testing. It does not involve AI, image analysis, or clinical studies with human readers, and therefore, the questions you've posed are not addressed by this specific regulatory submission.

{0}------------------------------------------------

March 25, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Smith & Nephew, Inc. Michelle Huettner Director, Regulatory Affairs 1450 Brooks Road Memphis, Tennessee 38116

Re: K210566

Trade/Device Name: LEGION Porous CR Narrow Femoral Components Regulation Number: 21 CFR 888.3565 Regulation Name: Knee Joint Patellofemorotibial Metal/Polymer Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: MBH Dated: February 22, 2021 Received: February 26, 2021

Dear Michelle Huettner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ting Song Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210566

Device Name

LEGION Porous CR Narrow Femoral Components

Indications for Use (Describe)

Total Knee components are indicated for rheumatoid arthritis, osteoarthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartment, or total knee replacement. LEGION Porous CR Narrow Femoral Components are indicated for use without bone cement and are single use devices.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Submitted by:Smith & Nephew, Inc.Orthopaedics Division1450 East Brooks RoadMemphis, Tennessee 38116
Date of Summary:March 18, 2021
Contact Person and Address:Michelle HuettnerDirector, Regulatory AffairsPhone: 765-426-6070Michelle.Huettner@smith-nephew.com
Name of Device:LEGION Porous CR Narrow FemoralComponents
Common Name:Orthopedic Devices
Device Classification Name andReference:21 CFR 888.3565: Knee joint patellofemorotibialmetal/polymer porous-coated uncementedprosthesis
Device Class:Class II
Panel Code:Orthopaedics/87
Product Code:MBH
Predicate Device:Primary Predicate – K073325 - LEGION PorousPrimary Femoral Components
Reference Device – K091543 - LEGION PorousPlus HA Primary Femoral Components
Reference Device (Narrow) – K112941 - LEGIONNarrow Oxinium CR and PS FemoralComponents and Device Specific Instruments

Device Description

The subject devices of this premarket notification are the LEGION Porous CR Narrow Femoral Components, which include catalog item offerings with and without HA Coating. These Femoral Components are indicated for use without bone cement applications. The subject devices are a line extension of the predicate devices listed, which are cleared under 510(k) submission numbers K073325 and K091543, with a modification in their femoral components' outer

{4}------------------------------------------------

dimensions making them Narrow Femoral Components (similar to the cleared reference device: LEGION Narrow Oxinium CR and PS Femoral Components and Device Specific Instruments cleared under K112941). The overall design of the subject device is geometrically identical to the predicate devices LEGION Porous Femoral Components (K073325, K091543) with their outer dimensional changes making them Narrow Femoral Components (Reference device K112941).

Design Changes:

The Narrow versions of the Predicate Femoral Component sizes 3, 4, 5, 6, and 7 will have their Medial/Lateral Width and Anterior/Posterior Dimension reduced making them Narrow versions of the cleared predicate devices. The full list of the catalog item numbers in scope for the subject device is included in the Device Description of the submission.

Indication for Use:

Total Knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement, or total knee replacement. LEGION Porous CR Narrow Femoral Components, with and without HA Coating, are indicated for use without bone cement and are single use devices.

Note, the Indications for Use for the subject devices (LEGION Porous CR Narrow Femoral Components) are the same as the cleared indications for the Predicate devices for the LEGION Porous CR Femoral Components, with and without HA Coating (K073325 and K091543 respectively).

Technological Characteristics

A review of the technological characteristics indicates that the LEGION Porous CR Narrow Femoral Components are equivalent to existing, legally marketed predicate devices with regards to mechanical performance and that there are no new issues raised related to the safety and effectiveness of the subject devices from Risk Management Plan.

{5}------------------------------------------------

A review of this design verification testing of the modified Femoral components also demonstrate that the subject devices are equivalent to their cleared predicate devices in their intended use and no new issues/risks raised from these design changes.

Substantial Equivalence

The subject devices are substantially equivalent in function, intended use, indications for use, and material composition, and they are very similar in overall design to the cleared predicate devices.

ManufacturerDescriptionSubmissionNumberClearanceDate
PrimaryPredicateSmith & Nephew, Inc.LEGION Porous Primary Femoral ComponentK07332512/20/2007
ReferenceDeviceSmith & Nephew, Inc.LEGION Porous Plus HA Primary Femoral ComponentK09154312/21/2009
ReferenceDeviceSmith & Nephew, Inc.LEGION Narrow Oxinium CR and PS Femoral Components and Device Specific Instruments (Narrow)K11294112/20/2011

Table 5.1: Substantially Equivalent Predicates

Performance Testing:

To further support a determination of substantial equivalence, various types of bench testing were conducted or leveraged for the subject, implantable devices in comparison against one or more of the previously cleared predicate devices. The specific types of bench testing included:

  • . Tibiofemoral constraint analysis testing
  • . Tibiofemoral contact area analysis testing
  • Patellofemoral resistance to lateral subluxation testing .

{6}------------------------------------------------

There was no necessary animal pre-clinical testing or clinical testing in scope for the subject devices. In conclusion, the bench performance testing met the predefined acceptance criteria and the results demonstrated substantial equivalence to the cleared predicate.

Conclusion

As previously noted, this Special 510(k) Premarket Notification is being submitted to request clearance for the LEGION Porous CR Narrow Femoral Components, which include catalog item offerings with and without HA Coating. Based on the similarities to the predicate devices, the subject devices are line extensions and substantially equivalent to the commercially available cleared predicate devices listed above.

§ 888.3565 Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis.

(a)
Identification. A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.(b)
Classification. Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.