K Number
K210420
Device Name
Goblin and Goblin LS Pedicle Screw Systems
Manufacturer
Date Cleared
2021-04-19

(67 days)

Product Code
Regulation Number
888.3070
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Goblin and Goblin LS Pedicle Screw Systems are intended for posterior, non-cervical fixation in skeletally mature patients as an adjunct to fusion for the following indications: - · Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); - · Spondylolisthesis; - · Trauma (i.e., fracture or dislocation); - Spinal stenosis; - · Curvatures (i.e., scoliosis, kyphosis and/or lordosis); - · Tumor and pseudarthrosis
Device Description
The Goblin and Goblin LS Pedicle Screw Systems are top-loading multiple component, posterior spinal fixation systems which consist of pedicle screws, rods, set screws, connectors, and transverse (cross) linking mechanisms. The implant components are supplied non-sterile, single use and fabricated from titanium alloy (Ti-6Al-4V ELI) as specified in ASTM F136. Various sizes of these implants are available.
More Information

No
The document describes a mechanical spinal fixation system and does not mention any AI or ML components or functionalities.

No.
Therapeutic devices are generally defined as those used for treatment, rehabilitation, or symptom management. This device is a pedicle screw system used for fixation, which aids fusion but is not explicitly defined as therapeutic in the provided text.

No

The device is described as a "Pedicle Screw System" used for "posterior, non-cervical fixation" as an "adjunct to fusion" for various spinal conditions. This indicates it is an implantable surgical device intended for treatment, not for diagnosing medical conditions.

No

The device description explicitly states it consists of physical components like pedicle screws, rods, and set screws, fabricated from titanium alloy. This indicates a hardware-based medical device, not a software-only one.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly describes a surgical implant system for posterior spinal fixation. This is a device used in vivo (within the body) to provide structural support and aid in fusion.
  • Device Description: The description details physical components like screws, rods, and connectors made of titanium alloy. These are physical implants, not reagents, instruments, or systems used to examine specimens in vitro (outside the body).
  • Lack of IVD Characteristics: There is no mention of analyzing biological specimens (blood, urine, tissue, etc.), detecting specific analytes, or providing diagnostic information based on laboratory testing.

IVD devices are used to perform tests on samples taken from the human body to provide information for diagnosis, monitoring, or screening. This device is a surgical implant used to treat a physical condition.

N/A

Intended Use / Indications for Use

The Goblin and Goblin LS Pedicle Screw Systems are intended for posterior, non-cervical fixation in skeletally mature patients as an adjunct to fusion for the following indications:

  • Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
  • Spondylolisthesis;
  • Trauma (i.e., fracture or dislocation);
  • Spinal stenosis;
  • Curvatures (i.e., scoliosis, kyphosis and/or lordosis);
  • Tumor and pseudarthrosis

Product codes (comma separated list FDA assigned to the subject device)

NKB

Device Description

The Goblin and Goblin LS Pedicle Screw Systems are top-loading multiple component, posterior spinal fixation systems which consist of pedicle screws, rods, set screws, connectors, and transverse (cross) linking mechanisms. The implant components are supplied non-sterile, single use and fabricated from titanium alloy (Ti-6Al-4V ELI) as specified in ASTM F136. Various sizes of these implants are available.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

non-cervical

Indicated Patient Age Range

skeletally mature patients

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-clinical testing was performed to demonstrate that the subject Goblin and Goblin LS Pedicle Screw Systems are substantially equivalent to the predicate device. The following testing was performed in accordance with the ASTM F1717:

  • Static compression
  • Dynamic compression
  • Static Torsion
    The nonclinical tests demonstrate that the Goblin and Goblin LS Pedicle Screw Systems are as safe, as effective, and performs as well as or better than the legally marketed predicate devices.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K061202 and K093294, K111362 and K173131, K160731

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K171808, K190830

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

0

April 19, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medynus Inc. % Jeena Mathai President Eerkie Corporation 4027 Runnymeade Drive Collegeville, Pennsylvania 19426

Re: K210420

Trade/Device Name: Goblin and Goblin LS Pedicle Screw Systems Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: February 22, 2021 Received: February 23, 2021

Dear Jeena Mathai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K210420

Device Name

Goblin and Goblin LS Pedicle Screw Systems

Indications for Use (Describe)

The Goblin and Goblin LS Pedicle Screw Systems are intended for posterior, non-cervical fixation in skeletally mature patients as an adjunct to fusion for the following indications:

  • · Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
  • · Spondylolisthesis;
  • · Trauma (i.e., fracture or dislocation);
  • Spinal stenosis;
  • · Curvatures (i.e., scoliosis, kyphosis and/or lordosis);
  • · Tumor and pseudarthrosis
Type of Use (Select one or both, as applicable)
-------------------------------------------------

|X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) SUMMARY

Medynus's Goblin and Goblin LS Pedicle Screw Systems

| Sponsor: | Manufacturer | Medynus Inc
18 Technology Dr. Ste 109
Irvine, CA 92618 |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| | Official Contact | David Shin |
| | Phone: | 9499320847 |
| | Fax: | 9499320851 |
| | Contact Person | Jeena Mathai
mgsharemg@gmail.com |
| | Date: | April 19, 2021 |
| Device Name: | Goblin and Goblin LS Pedicle Screw Systems | |
| Common Name: | Pedicle Screw Spinal Fixation System | |
| Classification Name: | Thoracolumbosacral Pedicle Screw System | |
| Classification
Number: | 21 CFR 888.3070 | |
| Product
Code/Classification: | NKB, class II | |
| Description: | The Goblin and Goblin LS Pedicle Screw Systems are top-loading multiple
component, posterior spinal fixation systems which consist of pedicle
screws, rods, set screws, connectors, and transverse (cross) linking
mechanisms. The implant components are supplied non-sterile, single
use and fabricated from titanium alloy (Ti-6Al-4V ELI) as specified in
ASTM F136. Various sizes of these implants are available. | |
| Indications For Use: | The Goblin and Goblin LS Pedicle Screw Systems are intended for posterior,
non-cervical fixation in skeletally mature patients as an adjunct to fusion for
the following indications: | |
| | Degenerative disc disease (defined as back pain of discogenic origin with
degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e., fracture or dislocation); Spinal stenosis; Curvatures (i.e., scoliosis, kyphosis and/or lordosis); Tumor and pseudarthrosis | |
| Performance Data: | Non-clinical testing was performed to demonstrate that the subject
Goblin and Goblin LS Pedicle Screw Systems are substantially equivalent
to the predicate device. The following testing was performed in
accordance with the ASTM F1717:

  • Static compression
  • Dynamic compression
  • Static Torsion
    The nonclinical tests demonstrate that the Goblin and Goblin LS Pedicle
    Screw Systems are as safe, as effective, and performs as well as or better
    than the legally marketed predicate devices. | |
    | Predicate Device: | Primary predicate: Globus Medical – REVERE Stabilization System (K061202
    and K093294)
    Additional predicates: Huvexel – Rexious Spinal Fixation System (K111362
    and K173131) and CG Bio. Co. Ltd. - LumFix Spinal Fixation System
    (K160731). | |
    | Reference Devices: | K171808 - TDM Small Locking Plate and Screw System
    K190830 - TDM Screw System | |
    | Technological Charact
    eristics | The Goblin and Goblin LS Pedicle Screw Systems were shown to be
    substantially equivalent and have equivalent technological characteristics
    to its predicate and reference devices through comparison in areas
    including design, labeling/intended use, material composition, function,
    range of sizes, and packaging. | |
    | Performance and SE
    Determination: | The Goblin and Goblin LS Pedicle Screw Systems have been
    demonstrated to be substantially equivalent to the predicate system(s)
    with respect to technical characteristics, performance, and intended use.
    The information provided within this premarket notification supports
    substantial equivalence of the subject device to the predicate device(s). | |

4