(447 days)
The Cannulated Compression Device System is indicated for fracture fixation of small and long bones. The system is not intended for spinal use.
The Cannulated Compression Device System is used to aid in the alignment and stabilization of bone fractures. The system consists of the following parts:
- A cannulated compression device body with distal threads for bone engagement and distal portals that allow passage of deployable integral anchors to achieve stabilization distally within the bone. The anchors may be retracted for removal of the device if and when it is necessary. The device will be provided in a pre-assembled condition with the deployable anchors and a distal end cap already installed. The device will be available in a variety of lengths.
- A compression nut will be provided separately in various configurations, including both threaded and non-threaded versions. The compression nut has a proximal head and internal threads. The proximal head engages the bone fragment. The internal threads allow engagement to the cannulated compression device body. As the compression nut is tightened onto the device body compression is achieved across the bone fragments. The rate of compression varies based on the proximal head configuration selected.
ODi-NA will manufacture the implants from implant grade titanium allov (Ti-6AL-4V-ELI) per ASTM F136. The implants will be offered in both sterile (steam) and non-sterile packaging configurations and are intended for single-use.
The provided text is a 510(k) summary for the Cannulated Compression Device System. It describes the device, its indications for use, and a comparison to predicate devices, along with performance data. However, it does not contain the information required to fill in the table and answer all the questions about acceptance criteria and a study proving a device meets them, particularly in the context of an AI/ML algorithm.
This document describes a physical medical device (cannulated compression device for bone fracture fixation), not an AI/ML algorithm or software. Therefore, many of the requested fields are not applicable to the content provided (e.g., sample size for training set, number of experts for ground truth, MRMC study, standalone algorithm performance).
Here's what can be extracted and what cannot:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria (from F543-17 & F1264-16) | Reported Device Performance |
---|---|
Torsional strength (per ASTM F543-17, Annex A1) | Evaluated |
Driving torque (per ASTM F543-17, Annex A2) | Evaluated |
Pullout strength (per ASTM F543-17, Annex A3) | Evaluated |
Static bend strength (per ASTM F1264-16, Annex A1) | Evaluated |
Dynamic bend strength (per ASTM F1264-16, Annex A4) | Evaluated |
Endotoxin limit (20.0 USP Endotoxin Units per device, per ANSI/AAMI ST72) | Meet specified endotoxin limit |
Compression features performance (comparative to predicate) | Comparatively tested |
Note: The document states that the device was "evaluated" and "tested" against these standards and criteria, and that it "meets the specified endotoxin limit." Specific numerical results of these evaluations and tests are not provided in this summary. The "acceptance criteria" here refers to the standards used for mechanical and biological testing of the physical hardware, not the performance of an AI/ML system.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable / Not provided. The study described is for a physical medical device, involving mechanical and biological testing according to ASTM standards. There is no "test set" in the context of software/AI data, nor "data provenance." The tests would involve physical samples of the device and biological samples for pyrogen testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable / Not provided. This information is relevant for AI/ML studies where human experts establish ground truth for image interpretation or diagnosis. For a physical device, ground truth is established by measured physical properties or biochemical assays against established standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable / Not provided. Adjudication methods are used in AI/ML studies to resolve disagreements among human experts when establishing ground truth. This is not relevant for the mechanical and biological testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable / Not provided. MRMC studies are specific to evaluating AI/ML systems, particularly in medical imaging, to compare human performance with and without AI assistance. This document describes a physical bone fixation device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable / Not provided. This pertains to the performance of an AI/ML algorithm without human intervention. The device in question is a physical surgical implant.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the mechanical tests (torsional strength, driving torque, pullout strength, bend strength), the "ground truth" would be the measured physical properties of the device, compared against the specifications defined by the ASTM standards (F543-17 and F1264-16).
- For the pyrogen testing, the "ground truth" is the measured endotoxin level compared against the specified limit of 20.0 USP Endotoxin Units per device, as per ANSI/AAMI ST72 using the Limulus amebocyte lysate (LAL) test.
8. The sample size for the training set
- Not applicable / Not provided. This is relevant for AI/ML algorithms.
9. How the ground truth for the training set was established
- Not applicable / Not provided. This is relevant for AI/ML algorithms.
§ 888.3040 Smooth or threaded metallic bone fixation fastener.
(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.